Академический Документы
Профессиональный Документы
Культура Документы
u1 Standard uncertainty
associated with the
Uniform
(Square)
repeatability of the balance
response to the reference
weight set.
reference thermometer.
reference balance.
u1 Standard uncertainty of
the readability and
Uniform
(Square)
resolution of the working
volumetric instrument
on Table for Balances
Estimated Value (NISTIR 6919)
Expanded uncertainty on the
calibration certificate of the
weights divided by 2 (coverage
factor – k)
0.001
9-Nov-09
(|Diff – Mean|)2
-
-
-
0.0000010
-
0.0000010
-
0.0000020
0.0005774
0.0005774
0.0012
Middle Weight
0.001
9-Nov-09
(|Diff – Mean|)2
0.000000
0.000001
0.000000
0.000001
0.000000
0.000001
0.000000
0.000003
0.000282
0.000282
0.0006
Heaviest Weight
0.01
9-Nov-09
(|Diff – Mean|)2
0.00007
0.00013
0.00000
0.00013
0.00000
0.00007
0.00007
0.00049
0.00367
0.00367
0.00735
Thermometer Calibration Worksheet
ur was changed to a value more representative of a reference thermometer. This changed the values for combined and
standard uncertainty.
0.5
10-Nov-09
(|Diff – Mean|)2
0.0002
0.0000
0.0029
0.0012
0.0000
0.0013
0.0021
0.0078
0.0360
0.0469
0.09
Please note that the value in Cell H8 is an experimental evaporation rate (in mL for 30 seconds) determined by
liquid being dispensed under the laboratory conditions used for the pipette calibration (loss measured after 30
worksheet presents example data for 1 mL under the conditions used by the laboratory for pipette calibrations
experimentally determined by the laboratory for each volume being dispensed (calibrated) under the lab condi
pipetted, the correction may be negligible. The actual correction factor applied in Cell H9 (Cell H8 divided by f
time (6 seconds) required for the balance to stabilize following the dispensing of each aliquot for weighing.
Please note that a minor error in the spreadsheet has been corrected. This error did not significantly affect un
calibrations conditions greatly vary from 20 C. The temperature correction in cell H7 should not be an absolute
The formula has been corrected in this version of the spreadsheet to remove the absolute value function.
To facilitate understanding of calculations, table values have been expanded to 4 decimal places as would be e
balance.
Calibration Worksheet
Serial / Lab number
SN00099387
Smallest Disp Vol 0.10
Date: 23-Oct-09
mperature Correction contribution in ml per ml 0.0005
ap Loss Corr in millilitres (at 30 seconds) 0.0600
(per reading) 0.0120
Diff (Bal – Corr) |Diff – Mean| |(Diff - Mean)|^2
Vol (ml) Vol (ml) Vol (ml)
(0.0125) 0.0418 0.0017
(0.0034) 0.0328 0.0011
0.0056 0.0237 0.0006
0.0158 0.0136 0.0002
0.0151 0.0142 0.0002
0.0252 0.0041 0.0000
0.0444 0.0150 0.0002
0.0543 0.0250 0.0006
0.0735 0.0442 0.0020
0.0754 0.0460 0.0021
0.2934 Sum s: 0.0087
0.0293 sp = Std Dev 0.0311
0.03
0.06
rate (in mL for 30 seconds) determined by the laboratory for the specific volume of
pipette calibration (loss measured after 30 seconds). It is not a percentage. This
d by the laboratory for pipette calibrations. The evaporation rate needs to be
dispensed (calibrated) under the lab conditions. Depending on conditions and volume
tor applied in Cell H9 (Cell H8 divided by five) represents the evaporation in the typical
ispensing of each aliquot for weighing.
d. This error did not significantly affect uncertainty, but may affect a bias calculation if
rection in cell H7 should not be an absolute value.