VER 6.0
DOCUMENT NO.:F_E1305004
Declaration
This document is confidential and is the property of Saimo Electric. This
document may not be copied or reproduced in any way without the express
written consent of Saimo Electric. This document also is unpublished work of
Saimo Electric. Saimo Electric intends to and is maintaining the work as
confidential information. Saimo Electric also may seek to protect this work as
published version. In the event of either inadvertent or deliberate publication,
Saimo Electric intends to enforce its rights to this work under the copyright
laws as a published work. Those having access to this work may not copy, use
or disclose the information in this work unless expressly authorized by Saimo
Electric.
Saimo Electric will not make any guarantee on this manual, and be free liability
on any print mistake or misunderstanding by user.
If there is any difference from equipment ordered in this manual, Saimo
Electric reserves its final explanation right.
If any changes made on this manual, which will not be informed individually.
6105-S5 Feeder Controller Operating and Service Manual
Please read and observe the following safety precautions found throughout
this manual and be strict in execution, which will help to protect equipment
from damage, prolong equipment life time, and guarantee your normal usage
and maintenance and person safety.
· Please read through all the documents dispatched together with the
equipments!
· For the sake of personal safety and normal operation of the equipments, it is
imperative to assure that only the designed AC power supply is to be used.
Reliable 3-pin earthed power outlets must be used to avoid electric shock
when touching the enclosure. Ensure that the rated voltage of the display or
other peripheral devices is compatible with your local AC power requirements.
· The equipment must be protected against thunderstrike.
· Disconnect all the cables from the equipment before you attempt to move the
equipment. Do not attempt to move any components on the equipment when
equipment is powered on.
· Do not attempt to change any components or external connections on the
equipment before the equipment is powered off.
· Avoid inadvertent power interruption during normal operation.
· Only qualified technicians can be allowed to open the enclosure of the
equipment.
· keep the equipment from moisture. Shut off power from the equipment
immediately if it is splashed with water or other liquids occasionally.
6105-S5 Feeder Controller Operating and Service Manual
CATALOGUE OF CHART.................................................................................................................. VI
CHAPTER 1 INTRODUCTION.......................................................................................................... 1
1.1 GENERAL........................................................................................................................................ 1
1.2 TECHNICAL SPECIFICATIONS.......................................................................................................... 2
1.3 FEATURES...................................................................................................................................... 3
1.3.1 Measuring Function............................................................................................................. 4
1.3.2 Monitoring Function............................................................................................................. 5
1.3.3 Printing Function.................................................................................................................. 5
1.3.4 Communications (Optional)................................................................................................ 6
1.4 WARRANTY..................................................................................................................................... 7
1.5 UNPACKING AND INSPECTION........................................................................................................ 8
1.6 STORAGE....................................................................................................................................... 8
1.7 SYMBOL.......................................................................................................................................... 8
1.8 HARDWARE INSTRUCTION.............................................................................................................. 8
1.8.1 Housing................................................................................................................................. 8
1.8.2 Ambient Condition................................................................................................................ 8
1.8.3 Power Requirements........................................................................................................... 8
1.8.4 Speed sensor input.............................................................................................................. 9
1.8.5 Digital I/O Board(Figure 1-5)....................................................................................... 10
1.8.6 Load cell input.................................................................................................................... 13
1.8.7 Analog I/O Board................................................................................................................ 13
1.8.8 Comm. Board...................................................................................................................... 13
2.1 GENERAL...................................................................................................................................... 14
2.2 6105-S5 CONTROLLER INSTALLATION........................................................................................ 14
2.2.1 Mounting.............................................................................................................................. 14
2.2.2 Safety Precautions............................................................................................................. 18
2.2.3 Wiring (External Power Supply)....................................................................................... 19
2.2.4 Input voltage setup............................................................................................................ 20
2.2.5 Serial Communication....................................................................................................... 20
2.2.6 Load cell.............................................................................................................................. 22
2.2.7 Controller analog board..................................................................................................... 22
2.2.8 Controller CPU board........................................................................................................ 24
2.2.9 Digital input/output board.................................................................................................. 25
2.2.10 Wiring................................................................................................................................. 26
CHAPTER 3 OPERATION................................................................................................................. 27
Catalogue of Chart
FIGURE 1-1 6105-S5 FEEDER CONTROLLER –PANEL MOUNT.............................................................. 1
FIGURE 1-2 6105-S5 FEEDER CONTROLLER –FIELD MOUNT............................................................... 2
FIGURE 1-3 6105-S5 FEEDER CONTROLLER FRONT PANEL................................................................. 4
FIGURE 1-4 SPEED SENSOR INNER WIRING DRAWING.......................................................................... 10
FIGURE 1-5 DIGITAL INPUT INNER WIRING DRAWING............................................................................. 11
FIGURE 1-6 DIGITAL OUTPUT INNER WIRING DRAWING......................................................................... 12
FIGURE 1-7 DC SOLID STATE RELAY WIRING DRAWING........................................................................ 12
FIGURE 1-8 AC STATE RELAY WIRING DRAWING................................................................................... 12
FIGURE 1-9 ANALOG INPUT INTERIOR WIRING DRAWING...................................................................... 13
Chapter 1 Introduction
1.1 General
Saimo belt scale in feeder is specially designed for weighing material which is
continuously passing through the conveyor. Automatic operation is achieved without
intervention of the operator.
The electronic belt scale is typically comprised of three parts: load cell, weigh frame and
secondary electronics. The 6105-S5 controller serves as one of the Saimo customized
secondary electronics.
This manual provides the information necessary for installation, operation, maintenance
and service of the 6105-S5 controller.
The 6105-S5 controller is available in panel mount and field mount. By processing the
weight signal from the load cell and the speed signals from the speed sensor, It calculates
and displays flow rate and total of the material fed.
· Measuring
· Monitoring
· Printing
· Communication
Allows the operator to perform the calibration without the need of applying
test weights or test chain on the weighbridge. The electronic calibration
checks all the circuitry including the load cell, and is performed by
unbalancing the load cell bridge using a precision resistor. The calibration
constant is calculated on the basis of the load cell and the scale data.
· Test Weight Calibration
Requires the positioning of test weights on the weighbridge
· Chain Calibration
Requires the application of calibrated chains on the belt. This method is the
nearest to actual operating conditions.
· Material Calibration
Is performed by running a known quantity of material on the scale and
weighing that quantity on a static scale of known accuracy.
4. Zero Tracking (AZT)
Under a preset minimum flowrate when enabled, the instrument makes
subsequent automatic zero calibrations with the following sequence:
· performs automatic zero for one test duration.
· continuously repeats above zero calibration as long as feed rate remains
below AZT preset value.
“Zero Tracking” has Max. “Tracking range”, which is setup with % in “System
data” Menu .
5. Current Output Signals
An optional current output/analog input board is available. Each current output
may be programmed via the keyboard to deliver one of the following signals:
· flowrate
· belt loading
· belt speed
1.3.2 Monitoring Function
· Serial Communications
The communication protocol allows a remote intelligent device to read and
eventually write the contents of the registers.
During the communication activity, the 6105-S5 controller always acts as a
Slave, meaning it responds to a request from a Master device on the line, but
never attempts to send messages out.
6105-S5 Feeder Controller Operating and Service Manual
1.4 Warranty
Said warranty shall not apply if the equipment shall not have been operated
and maintained in accordance with seller’s written instructions applicable to
such equipment, or if such equipment shall have been repaired or altered or
modified without seller’s approval.
Field Service
Purchaser agrees to underwrite the cost of any labor required for replacement;
including time, travel and living expenses of Saimo Field Service engineer at
closest factory base, altered or modified
1.6 Storage
6105-S5 controller can be safely stored with cover latches secured and hole
plugs installed, between -40℃—+70℃. The units should be protected against
moisture.
1.7 Symbol
Symbol Instruction
Symbol Description
AC
Grounding end
Shielded grounding
Warn, power
Warn
2.2.1 Mounting
The 6105-S5 field mount feeder controller can be vertically secured to the rigid and
flat wall through the 4 mounting holes at the back of the enclosure. Care should be
taken to insure the mounting surface is flat so as not to twist or warp the fiberglass
enclosure when tightening the mounting holes.
The 6105-S5 panel mount controller should have a clearance of at least 5cm at the
top and bottom to allow air circulation. In case that the controller must be positioned
under the heat sources, proper space around the controller should be reserved.
Sufficient space should be kept behind the controller to facilitate wiring and changing
fuse. The side clearance should be easy for securing the holding brackets after the
6105-S5 has been put in place. Disconnect power supply from the controller when
14
6105-S5 Feeder Controller Operating and Service Manual
mounting the bracket. From the back, insert the holding brackets on both sides of the
controller firstly, then screw down tightly to secure the controller and prevent dust
ingress. Care should be taken to insure the mounting surface is flat so as not to twist
or warp the fiberglass enclosure when tightening the mounting holes.
Figure 2-2 Metal housing 6105-S5 Field mounting controller outline dimension and
mounting dimension
15
6105-S5 Feeder Controller Operating and Service Manual
Figure 2-3 Plastic housing 6105-S5 Field mounting controller outline dimension and
mounting dimension
16
6105-S5 Feeder Controller Operating and Service Manual
17
6105-S5 Feeder Controller Operating and Service Manual
Figure 2-4 6105-S5 Panel mounting controller outline dimension and mounting
dimension
Note:
1. See figure 2-4 for Panel mounting dimensions.
2. Take fixing bracket clamp out from the package when installation.
3. Remove bracket clamp and push housing in from the front of opening hole, then
mount bracket clamp and fasten the bolt accordingly.
18
6105-S5 Feeder Controller Operating and Service Manual
19
6105-S5 Feeder Controller Operating and Service Manual
Note: Check controller input voltage to be same as input voltage prior to getting
controller powered on, refer to section 2.2.4 for input voltage setup.
Procedures for Saimo 6105-S5 :
A.Contrarotate bolt mounted at the left lower corner, open the door.
B.Put power supply cable into tube at the right of housing bottom. (See figure 2-2).
C.Connect safe ground wiring plate located on the one side of chassis.
D.Connect fire wire to L.
E.Connect zero wire to N.
F.Close front chassis cover, then contra rotate the bolt at the left lower corner till
cover is locked, validation gate is locked.
20
6105-S5 Feeder Controller Operating and Service Manual
J3 J5
COMM 1# RS-232 RS-232
COMM 1# RS-485 RS-485
COMM 2# RS-232 RS-232
COMM 2# RS-485 RS-485
Note: Terminal resistance RT: 120 OHM, cable should be twisted-pair shielded
cable with R 120OHM.
21
6105-S5 Feeder Controller Operating and Service Manual
22
6105-S5 Feeder Controller Operating and Service Manual
23
6105-S5 Feeder Controller Operating and Service Manual
24
6105-S5 Feeder Controller Operating and Service Manual
25
6105-S5 Feeder Controller Operating and Service Manual
2.2.10 Wiring
26
6105-S5 Feeder Controller Operating and Service Manual
Chapter 3 Operation
3.1 Run Menu
3.1.1 General
Controller 6105-S5 is a microprocessed instrument, which can receive speed and
weight signal, provide visual signal and electric output signal for usage of toal and
flow rate display. Controller provides up to 8 strain resistance type load cell with
10VDC excitement voltage and provides 6 core load cell with compensation signal
terminal.
When conveyor is running with empty loading, auto zero track function let belt scale
system auto zero. Cause such function is not necessary for some application, so
zuto zero track function can be selected via menu. Belt feeder control output can be
PID control.
If power supply is stopped, RAM data storage will last over 20 years.
If any error occurs in initial calibration, it must be checked for reason and corrected
accordingly. In normal operation, error means that there is fault in system or incorrect
operation is made.
27
6105-S5 Feeder Controller Operating and Service Manual
2、“MENU” – enter the main menu, see section 3.4 for details.
3、Up , Down arrow – scroll down & up in selected menu
4、Soft Keys - execute displayed function directly above the key
5、Numeric keys 1~0—input the data parameter
6、Decimal“.”—input the decimal
7、“DEL” – delete the digit or punctuations prior to pressing ENTER.
8、“CHK” –self-check function (disabled)
9、”ENT” - confirm enterin
LCD displays actual run info. or menu input info.
3.1.4 Quick setup method
Description Setup range Default Remark
Measure unit Metric,British, Mix Metric
Total unit Kg、t Kg
Length unit m m
Rate unit Kg/h、t/h、kg/m、t/m t/h
Loadcell unit Kg Kg
Scale capacity 1-1000000 100t/h
0.1、0.2、0.5、1、2、5、
Scale division 10、20、50、0.01、0.02、 0.001
0.05、0.001、0.002、0.005
Scale code 0-255 1
Loadcell no. 1-8 2
Idler space 0-10000mm 36mm
Belt incline angle -25 - 25 0
Sensor capacity 0-1000000kg 50kg
Sensitivity 0-1000000 3mv/v
1# resistance 0-1000000 350
2# resistance 0-1000000 0
3# resistance 0-1000000 0
4# resistance 0-1000000 0
5# resistance 0-1000000 0
6# resistance 0-1000000 0
channel1、channel2、
Weigh channel Channel 1
double channel
Simulated input, actual
Speed mode Simulated input
input
Channel 1,2,3,
Speed chanel Channel 1
Three channel
P to L 0-1000000mm 32mm
Idler No. 1-4 1
P to 1st 0-1000000mm 24mm
P to 2nd 0-1000000mm 0mm
28
6105-S5 Feeder Controller Operating and Service Manual
29
6105-S5 Feeder Controller Operating and Service Manual
System running!
8000000 t
888 t/h
1m/s
System running!
Press ENT key, clear total!
Press TOT RSET soft key, the controller will display ”Press ENT to reset total.” Then
Press ENT key to reset the total.
If ENT key is pressed, the reset total will be zero, but the master total remains not
changed.
To reset the master total, enter “System Password Setup” menu to get the access to
this function.See section “password setup” for details.
30
6105-S5 Feeder Controller Operating and Service Manual
6. ALARM SETUP
7. LOADOUT SETUP
8. PASSWORD SETUP
9. DIAGNOSTICS
10.STATISTICS
PREV PAGE
RUN UP
System word: [ ]
Operate word: [ ]
MAIN PAGE
MENU DOWN
Select “System passoword” and enter password setup in numeric keyboard, press
“ENT” to display “ENTER INPUT”, press “ENTER INPUT” in lower side of screen to
display the follows:
2006-3-21 12:00:00 Scale#1
Correct!
If clear main Tot, press TOT RSET!
MAIN TOT PAGE
MENU RSET DWON
31
6105-S5 Feeder Controller Operating and Service Manual
Then press “Reset main total” to clear main total. If password is incorrect, the
following will be displayed:
MAIN PAGE
MENU DOWN
ALARM DETAILS
ALARM NAME: [ LO RATE ]
ALARM DATE: [ 2005-06-08 ]
ALARM TIME: [ 12:00:00 ]
ALARM ACTION: [ NEW ]
System running!
32
6105-S5 Feeder Controller Operating and Service Manual
800 t/h
SET RATE : [ 800 t/h ]
OUTPUT : [ 100 % ]
Press ↑↓key, then press ENT!
ZERO PREV
CAL RUN
800 t/h
SET RATE : [800]
OUTPUT : [ 100 % ]
Press ↑↓key, then press ENT!
33
6105-S5 Feeder Controller Operating and Service Manual
1. INITIAL SETUP
2. CALIBRATION
3. SYSTEM DATA
4. I/O DEFINITION
5. COMMUNICATION
PREV PAGE
RUN DOWN
Press a numeric key (1-5) to select the submenu. Press “RUN” to return to Run
Menu.
Use the Press “PAGE DOWN” soft key, the second page of the Main Menu is
displayed as follows:
2006-3-21 12:00:00 Scale#1
6. ALARM SETUP
7. LOADOUT SETUP
8. PASSWORD SETUP
9. DIAGNOSTICS
10.STATISTICS
PREV PAGE
RUN UP
Press numeric keys (6-10) to select the submenu. Press “PREV RUN” to return to
Run Menu. Press “PAGE UP” soft key to return to the first page of the Main Menu.
Press numeric key “0” to enter report statistic.
34
6105-S5 Feeder Controller Operating and Service Manual
INITIAL SETUP 1
Measure Units: [ METRIC ]
Total Units: [ t ]
Length Units: [ METERS ]
Rate Units: [ t/h]
PREV PAGE
RUN DOWN
Move the cursor to corresponding item with the up/down scroll keys; then press ENT,
the following screen is displayed.
2006-3-21 12:00:00 Scale#1
INITIAL SETUP 1
Measure Units: METRIC
Total Units: [ t ]
Length Units: [ METERS ]
Rate Units: [ t/h]
In this menu, if Up/Down scroll keys are pressed, the default unit “Metric” is accepted
and the cursor moves to other variables.
Otherwise, if “ENTR” soft key is pressed, the measure unit value changes among
“English”, “Metric ” and “Mixed”. When the desired option is displayed, press “SURE”
soft key to input the value. At this time the screen returns to the first page of Initial
Setup 1 menu.
You can use the same method to setup other units listed in the menu.
Measure Units: English, Metric, Mixed
Total Units: t, kg
Length Units: Meters
Flow Rate Units: t/h, kg/h, t/m, kg/m
35
6105-S5 Feeder Controller Operating and Service Manual
Press “PAGE DOWN” soft key from the first page of the Initial Setup 1 Menu, the first
page of the Initial Setup 2 Menu is shown as follows:
INITIAL SETUP 2
Loadcell Units: [ kg]
Scale Capacity: [ 500 kg/h ]
Scale Division: [ 1 ]
Scale Code: [ 1 ]
Use Up/Down scroll keys to move the cursor to the desired variable, and press ENT
to enter the following screen.
2006-3-21 12:00:00 Scale#1
INITIAL SETUP 2
Loadcell Units: [ kg]
Scale Capacity: 500 kg/h
Scale Division: [ 1 ]
Scale Code: [ 1 ]
If a digital value is entered, and press “SURE” soft key to confirm. the screen return
to the first page of Initial Setup 2 menu. You can use the same method as described
in setup 1 to setup other parameters listed in the menu.
36
6105-S5 Feeder Controller Operating and Service Manual
Idler no. 1 4
38
6105-S5 Feeder Controller Operating and Service Manual
Press “PAGE DOWN” soft key from the Initial Setup 2 Menu or the previous menu to
enter ”Initial Setup 3 “Menu as follows:
2006-3-21 12:00:00 Scale#1
INITIAL SETUP 3
# of Loadcells: [ 1 ]
Idler Spacing: [ 100mm ]
Conveyor Angle: [ 10 ]
Loadcell Cap: [ 100 kg ]
Press “PAGE DOWN” soft key from the Initial Setup 3 Menu to enter the first page of
Initial Setup 4 Menu as follows:
2006-3-21 12:00:00 Scale#1
INITIAL SETUP 4
Sensitivity : [ 3 mv/v ]
1#Resistor : [ 350 Ohms]
2#Resistor : [ 350 Ohms]
3#Resistor : [ 350 Ohms]
39
6105-S5 Feeder Controller Operating and Service Manual
3#Resistance 0 1000000
Press “PAGE DOWN” soft key from the Initial Setup 4 Menu to enter the fifth page of
Initial Setup Menu 5 as follows:
2006-3-21 12:00:00 Scale#1
INITIAL SETUP 5
4#Resistor : [ 350 Ohms]
5#Resistor : [ 350 Ohms]
6#Resistor : [ 350 Ohms]
Weight Channel : [ CHANNEL 1]
Note: When Dual channels are used for the weight channel, the two load cells should
be connected to the channel 1 and channel 2 separately on the A/D board.
Data setup in this section depends on mechanical
Belt scale data(Figure 3—2)
Before turning on belt running switch or belt scale power supply switch, complete the
following work, refer to your system Data Sheet in the front of this manual.
1. Scale capacity
Record scale capacity as per T/H:(Example:400.0)
(Tons/hour)
2. Belt scale parameters
When R-Cal is executed in initial programme, if Test chain, test weight cal is used,
enter weight value in “Cal Data Scroll” after initial programme is completed.
40
6105-S5 Feeder Controller Operating and Service Manual
41
6105-S5 Feeder Controller Operating and Service Manual
(Idler no.)
st
c、Pivot to 1 weigh idler(Figure 3—4)
As indicated on the following drawing, measure the distance from the pivot centerline
to the first weigh idler’s centerline to the nearest 1mm, record the distance below:
Note: In programming process, each scale bridge can be input up to six distances
between pivot to weigh ider. The method is same as d.
e、Pivot to 1st , 2nd, 3rd, forth idler distance(Figure 3—6)
As indicated on the following drawing, measure the distance from the pivot centerline
to weigh idler’s centerline to the nearest 1mm, record the distance below:
mm ( Pivot to weigh idler
distance)
42
6105-S5 Feeder Controller Operating and Service Manual
43
6105-S5 Feeder Controller Operating and Service Manual
44
6105-S5 Feeder Controller Operating and Service Manual
45
6105-S5 Feeder Controller Operating and Service Manual
Press “PAGE DOWN” soft key from the Initial Setup 5 Menu, the sixth page of Initial
Setup Menu is displayed as follows:
INITIAL SETUP 6
Speed Mode: [SIMULATE]
Speed Channel: [CHANNEL 1]
The setting value of the speed mode are “Simulated” and “Actual”.
When “Simulated” is selected, the controller will generate an analog signal for speed
input. In case that “Actual” is selected, the speed input is an actual speed signal.
Simulation allows operation without “actual input”.
Speed channel is used to select external speed signal to be connected to which one
speed channel. “Channel 1”, “channel 2” are located in the first digital I/O board;
channel 3 is located in the second digital I/O board. There are total 4 options:
channel 1, channel 2, channel 3 and Three-channel. Three-channel represents that
all three speed sensor should be connected to controller to tell which speed sensor
faulty and then to select both speed sensors with same output for usage.
Press “PAGE DOWN” soft key from the Initial Setup 6 Menu, the seventh page of
Initial Setup Menu is displayed as follows:
This submenu is test duration establishment. It is similar to run menu, but be noted
with the difference lower side of display screen.
Feeder electric belt scale is dynamic measurement equipment, the accuracy of
measurement depends on conveyor system directly. Cause conveyor system is in
running, Select a certain period of running time to measure the tare and calculate the
average value.
The test duration should be set to a value equivalent to over three belt revolution or
greater than 6min.
2006-3-21 12:00:00 Scale#1
8000000 t
888 t/h
1m/s
Establish test duration.
BGIN PAGE
DURA UP
Enter belt length; the revolutions of belt runnin in test duration, running time of belt.
Press “BGIN DURA” on above menu to enter below menu:
8000000 t
888 t/h
1m/s
Establish test duration.
[ 10 ]
BGIN EXIT PAGE
DURA DURA UP
8000000
Full Length: 100
t
8000000 t
888 t/h
1m/s
Establish test duration.
Auto test duration can use the full duration and part duration, two methods. Full
duration uses belt integer revolution and belt length to make test duration; Part
duration uses part of belt length to make test duration. It is recommended to use full
duration. If the belt is too long to measure, then part duration is available.
Press “FULL DURA” or “PART DURA”, the following screen is displayed.
48
6105-S5 Feeder Controller Operating and Service Manual
8000000
Revolutions:1
t
Running Time:1
8000000 t
888 t/h
1m/s
Weight auto span mode!
[ 100 ]
MAIN
MENU
49
6105-S5 Feeder Controller Operating and Service Manual
8000000 t
888 t/h
1m/s
Scale calibrated, press RUN to start!
MAIN
MENU
1. INITIAL SETUP
2. CALIBRATION
3. SYSTEM DATA
4. I/O DEFINITION
5. COMMUNICATION
PREV PAGE
RUN DOWN
CALIBRATION
MAIN
MENU
50
6105-S5 Feeder Controller Operating and Service Manual
Move the cursor to the desired submenu with Up/Down scroll keys and press ENT
key to select it.
8000000 t
888 t/h
1m/s
Zero calibration mode.
Press “AUTO ZERO “ soft key soft key for the following screen:
2010-11-09 12:00:00 Scale#1
8000000 t
888 t/h
1m/s
Run belt empty, select an option.
BGIN EXIT
ZERO ZERO
Press the “BGIN ZERO“ soft key on the following screen to start the zero test:
51
6105-S5 Feeder Controller Operating and Service Manual
8000000 t
888 t/h
1m/s
Auto zero in progress!
[ 100 ]
BGIN EXIT
ZERO ZERO
In case that the belt is interrupted during the zero calibration procedure, the following
screen will be displayed.
2010-11-09 12:00:00 Scale#1
8000000 t
888 t/h
1m/s
Auto zero in progress!
Belt stops! [10]
BGIN EXIT
ZERO ZERO
8000000 t
Old zero:[ 888 ]
New zero:[ 1000 ]
Main: Auto zero complete!
Zero error :[ -11.2% ]
BGIN EXIT ENTR
ZERO ZERO ZERO
52
6105-S5 Feeder Controller Operating and Service Manual
0.00 t
0.00t/h
1m/s
Main: Run belt empty, select an option.
BGIN EXIT
ZERO ZERO
Note: Duringg auto zeroing process, belt must be running, cause complete zeroing
process needs one revolution of belt running. If there is error in connection of sensor,
or signal vibrates too large, then below menu will be displayed(please check for fault)
8000000 t
t/h
1m/s
Zero failure, exceed zero range!
BGIN EXIT
ZERO ZERO
If press “MANL ZERO” soft key to enter the “Manual Zeroing” menu.
“MANL ZERO” allows direct input via key board and changing scale zero
8000000 t
888 t/h
1m/s
Run belt empty, select an option.
8000000 t
Old zero:[ 888 ]
New zero:
Press numeric keys to enter the new zero and press “SURE ZERO” to validate it.
There are two different system calibration modes: simulated calibration and material
calibration.
Three simulated span calibration methods are available: R-cal, test weights and test
chains. Simulated calibration needs caculating “Simulated constant ”, see section
3.3.10 for its calculation in details. When performing span calibration, the calibration
method should be firstly selected, and the relevant resistor or test weights / chain
weights parameters should be entered into the controller.
The following Span Calibration menu is accessible by selecting “SPAN
CALIBRATION” from the Calibration Menu, then press “ENT”:
CALIBRATION
MAIN
MENU
54
6105-S5 Feeder Controller Operating and Service Manual
Before span calibration, standard weights must be loaded: in “Cal. data setup 1”
under system main menu “3. system setup”, modify calibration method, select
“weight cal” in simulated calibration, and enter weights weight in “weights weight” as
shown in below screen:
2010-11-09 12:00:00 Scale#1
CALIBRATION DATA 1
Simulated: ELECTRIC
Resistor: [ 165000Ohms ]
Weights: [ 100kg ]
Chain: [ 100kg/m ]
Press ↑↓key, then press ENT!
8000000 t
888 t/h
1m/s
Auto span mode!
There are two spanning modes available: automatic spanning and manual spanning.
Press the “AUTO SPAN” soft key under 3.2.2.6 menu for the following screen:
2010-11-09 12:00:00 Scale#1
8000000 t
888 t/h
1m/s
Auto span in progress!
[ 100 ]
BGIN EXIT
CAL CAL
55
6105-S5 Feeder Controller Operating and Service Manual
In case that the belt is stopped during the calibration procedure, the following screen
will be displayed.
2010-11-09 12:00:00 Scale#1
8000000 t
888 t/h
1m/s
Auto span in progress!
Belt stops! [10]
BGIN EXIT
CAL CAL
55.00 t
Old Span:[3751850 ]
New Span:[ 3751850 ]
When the automatic calibration is finished, the screen returns to the first menu of
Auto Span.
Press “Manual Cal.” to enter “Manual cal” menu.
“Manual Cal” allows input and change span value via key board.
2010-11-09 12:00:00 Scale#1
8000000 t
Old Span:[ 888 ]
New Span:
56
6105-S5 Feeder Controller Operating and Service Manual
Press numeric keys to enter the new span and press “SURE SPAN” to validate it.
CALIBRATION
MAIN
MENU
8000000 t
888 t/h
1m/s
Material calibration!
BGIN EXIT
CAL CAL
8000000 t
888 t/h
1m/s
Material calibrating, please wait!
[8]
BGIN EXIT DONE
CAL CAL SPAN
57
6105-S5 Feeder Controller Operating and Service Manual
Pressing “DONE CAL” soft key enters the following screen: Pressing “DONE CAL”
soft key enters the following screen:
2006-3-21 12:00:00 Scale#1
8000000 t
Old Span:[ 888 ]
New Span:
Material calibration, please wait!
Refer weight:
BGIN RE- ENTR
CAL INPT
When material calibration is finished, input the standard weight value and press
“SURE” soft key to validate it and enter the following screen.
2006-3-21 12:00:00 Scale#1
8000000 t
Old Span:[ 888 ]
New Span:[ 1000 ]
Material calibration complete!
Span err:[ -11.2% ]
BGIN EXIT MATL ADD ENTR
CAL CAL FACR REFR SPAN
58
6105-S5 Feeder Controller Operating and Service Manual
8000000 t
Old factor:[ 0% ]
New factor:[1.0 ]
Sure new factor?
8000000 t
Old Span:[ 888 ]
New Span:[ 1000 ]
Material calibration complete!
Span err:[ -11.2% ]
BGIN EXIT MATL ADD ENTR
CAL CAL FACR REFR SPAN
8000000 t
888 t/h
1m/s
Material calibration complete!
Span err:[ -11.2% ]
BGIN MATL
CAL FACR
59
6105-S5 Feeder Controller Operating and Service Manual
Belt scale calibration procedures consist of two parts: 1) initial calibration, 2) rutine
calibration
Initial calibration need only be done once, various constants are needed for actual
operation of the system must be installed during initial calibration.
Routine calibration is performed by usning ZERO CAL and SPAN CAL procedures
listed under Main Menu 2.
These procedures are the fine adjustments used to compensate for material build-up,
mechnical wear, and uncontrollable conveyor parameters.
Since most intallations differ in characteristics, Saimo recommends that a record be
kept of calibration results. This will help to determine the frequency of calibration
checks necessary to maintain your accuracy requirements. See calibration report at
the end of this section.
Date
Test people
Scale capacity
Belt length
Test length
60
6105-S5 Feeder Controller Operating and Service Manual
Test time
Calibration Constant
R-Cal.
Test chain
Cal. mode
—reserve value
Span—initial cal.
value
—reserve value
1. INITIAL SETUP
2. CALIBRATION
3. SYSTEM DATA
4. I/O DEFINITION
5. COMMUNICATION
PREV PAGE
RUN DOWN
DISPLAY DATA 1
Zero Dead: [ 0% ]
Supplementary: [ SPEED ]
Filt Constant: [ 10 ]
Language: [ English ]
Press ↑↓key, then press ENT!
MAIN PAGE
MENU DOWN
61
6105-S5 Feeder Controller Operating and Service Manual
Press the Up/Down scroll keys to scroll the cursor to the submenu desired, use
numeric key or “Select input” to input value.
Description Range
Zero dead 0% 100%
Supplementary Speed Load
Filter Constant 0 100
Language{} Chinese ENGLISH
[Zero Dead]:During a run period, a total lower than the preset Zero Dead will not be
added into the master total.
Pressing “PAGE DOWN” soft key enters the second page of the System Data Menu.
DATA SETUP
Year: [6]
Month: [3]
Date: [ 21 ]
Press ↑↓ key to select submenu, use numeric key to input the value desired.
Pressing “PAGE DOWN” soft key enters the third page of the System Data Menu.
TIME SETUP
Hour: [ 12 ]
Minute: [0]
Second: [0]
62
6105-S5 Feeder Controller Operating and Service Manual
Press ↑↓ key to select submenu, use numeric key to input the value desired.
Pressing “PAGE DOWN” soft key enters the forth page of the System Data Menu.
2006-3-21 12:00:00 Scale#1
SCALE DATA
Max. Speed: [3 m/s ]
Press ↑↓ key to select submenu, use numeric key to input the value desired.
Press “PAGE DOWN” soft key, the first page of the Calibration Data Menu is
displayed.
2006-3-21 12:00:00 Scale#1
CALIBRATION DATA 1
Simulated: [ ELECTRIC ]
Resistor: [ 165000Ohms ]
Weights: [ 100kg ]
Chain: [ 100kg/m ]
Press ↑↓key, then press ENT!
63
6105-S5 Feeder Controller Operating and Service Manual
CALIBRATION DATA 2
Constant : [ 800 ]
Matl. Factor : [ 2% ]
Zero Track : [ 3% ]
Zero Range : [ 4% ]
Press ↑↓key, then press ENT!
CALIBRATION DATA 3
Linear : [ NO ]
0 ~ 30% : [ 100% ]
30%~ 70% : [ 100% ]
70%~ 100%: [ 100% ]
Press ↑↓key, then press ENT!
64
6105-S5 Feeder Controller Operating and Service Manual
Press “PAGE DOWN” soft key, enter the first page of the CONTROL DATA menu.
CONTROL DATA 1
Control Mode : [ MANU ]
Setting Mode : [ LOCAL ]
High Limit : [ 80% ]
Low Limit : [ 20% ]
Press ↑↓key, then press ENT!
When control mode is selected as Manual,the PID control output mode of 6105-S5
feeder controller will be manual mode, when control mode is selected as Auto, the
PID control output mode of 6105-S5 feeder controller will be auto mode. The control
high and low limit are used to limit the range of the PID control output.
Press “PAGE DOWN” soft key in above screen, enter the second page of the
CONTROL DATA menu
CONTROL DATA 2
Proportional : [ 200 % ]
Integral : [ 2 sec ]
Derivative : [ 0 sec ]
PID Dead : [ 1% ]
Press ↑↓key, then press ENT!
Press “PAGE DOWN” soft key in above screen, enter the third page of the
CONTROL DATA menu.
CONTROL DATA 3
Volume : [ NO ]
Difference : [ 10% ]
Delay : [ 5 sec ]
The 6105-S5 controller has a volumetric measurement function. Under normal operating
conditions, the controller will compare the two independent weighing signal in real time, if the two
weighint signals are within the normal setting range, the controller will calculate the
instantaneous flow and total with the two weighing signals and speed signal. When the weighing
sensor failed, i.e., the two independent weighing signalis beyond the normal range of settings, it
will convert to volumetric metering function.
The working principle of volumetric metering is: the controller automatically records the flow and
speed of current product within 1 hour, the sum of flow in 1 hour divided by the sum of speed in 1
hour, obtain the line weight in 1 hour (weight per unit length). This line weight is multiplied by the
66
6105-S5 Feeder Controller Operating and Service Manual
average speed in 1 hour, then get the average flow rage in 1 hour. When the load cell fault occurs,
the controller will use the current line weight to take place of wiehging signal, and multiple it by
the current speed to get the current flow.
Select YES in VOLUME mode will enable the volumetric metering function.
Select NO in VOLUME mode will disable the volumetric metering function.
The setting of Difference is used to determine the difference threshold between the two load
cells.
The setting of Delay is used to define the delay time to convert to volumetric metering mode.
When the volumetric metering mode is enabled, if the difference of load cell exceeds the preset
value
, and keep the difference until the delay time, the controller will covert into the state of
the volumetric measurement. When the difference is less than the difference value , the
8000000 t
888 t/h
1m/s
Establish test duration!
67
Press “FULL DURA” or “PART DURA” to enter establish test duration menu.
8000000
Running Mark :
t
Running Time :
press “BGIN DURA”, after that each time it reaches the mark, press “MARK CNT”,
when it reaches proper revolutions, press ”DONE DURA” then press “EXIT DURA”.
BGIN PAGE
DURA UP
Press ↑↓ to select submenu, use numeric key to input the value desired and then
68
6105-S5 Feeder Controller Operating and Service Manual
2) For test weights calibration mode, the following method should be used:
TestWeight s(Kg)
a) The Equivalent Load (Kg/m) =
WeighSpanL ength(D)
Test Weights = Static weight applied on the weigh idlers
Weigh Span Length = Idler spacing X Number of weigh idlers
For example:
Kg = 100
D = 2.4m
Kg/m = 41.67 kg/m
69
6105-S5 Feeder Controller Operating and Service Manual
b) Simulated Constant For Test Weights can be calculated using the following formula:
(kg / m)BeltLengthforTestDuration(m)
The Simulated Constant =
1000
For example:
The Belt Length for Test Duration = 180m
The Simulated Constant = 41.67180 = 7.5t
1000
3) For test chains calibration mode, the following method should be used:
a) The Equivalent Load (Kg/m) is specified at the tail of the test chains
b) Simulated Constant For Test Chains can be calculated using the following formula:
(kg / m)BeltLengthforTestDuration(m)
The Simulated Constant =
1000
For example:
The Belt Length for Test Duration = 180m
Kg/m = 30
The Simulated Constant = 30180 = 5.4t
1000
Note: The system setup is intended to program some basic parameters which determine whether
correct result can be achieved from the belt scale. These parameters include scale number and
scale capacity etc.
The following is an explanation to a part of these parameters.
[Zero Tracking]: Zero may drift due to various influence factors. When Zero Tracking is set,
automatic zeroing will be implemented in case of a total lower than the preset value.
[Zero Dead]: During a run period, a total lower than the preset Zero Dead will not be added into
the master total.
[Accumulated Pulses]: When the total is up to the preset Accumulated Pulses, an accumulated
pulse signal will be emitted.
[Pulse Width]: This option defines the ratio of high voltage level [1] and low voltage level [0] of
pulse width.
1. INITIAL SETUP
2. CALIBRATION
3. SYSTEM DATA
4. I/O DEFINITION
5. COMMUNICATION
PREV PAGE
RUN DOWN
Pressing “4” from Main Menu enters the first page of the I/O Definition Menu.
70
6105-S5 Feeder Controller Operating and Service Manual
INPUTPORT SETUP 1
In port 1: [ NO DEFINE ] [ OPEN ]
In port 2: [ NO DEFINE ] [ OPEN ]
In port 3: [ NO DEFINE ] [ OPEN ]
In port 4: [ NO DEFINE ] [ OPEN ]
Press ↑↓key, then press ENT!
MAIN PAGE
MENU DOWN
INPUTPORT SETUP 1
In port 1: NO DEFINE [ OPEN ]
In port 2: [ NO DEFINE ] [ OPEN ]
In port 3: [ NO DEFINE ] [ OPEN ]
In port 4: [ NO DEFINE ] [ OPEN ]
Press ENTR to select, then press SURE!
As above menu shown, press “Select input”, input port 1 will switch between “batch
ready”, “batch run”, “belt run”, “reset total”, “reset alarm”, “auto zero”, “flow”,
“departure”, “jam”, “high temp.”, “remote/local”, “VFD fault”, “Print”, “No define”, when
proper input port 1 is selected such as “Batch ready”, press “SURE” to confirm. For
the other items’s selection they are same as above operation.
Pressing Right scroll key from the first page of I/O Definition menu enters the
following screen.
2006-3-21 12:00:00 Scale#1
INPUTPORT SETUP 1
71
In port 1: [ NO DEFINE ] [ OPEN ]
In port 2: [ NO DEFINE ] [ OPEN ]
In port 3: [ NO DEFINE ] [ OPEN ]
In port 4: [ NO DEFINE ] [ OPEN ]
6105-S5 Feeder Controller Operating and Service Manual
INPUTPORT SETUP 1
In port 1: [ NO DEFINE ] OPEN
In port 2: [ NO DEFINE ] [ OPEN ]
In port 3: [ NO DEFINE ] [ OPEN ]
In port 4: [ NO DEFINE ] [ OPEN ]
Press ENTR to select, then press SURE!
As above menu, press “SLCT”, input port 1 switches between “normal open” and
“normal close”, when the proper input port1, for example “NO” is selected, then press
“ENTR” to confirm. The others are same as above in selection and operation.
Pressing “PAGE DOWN” soft key from the first page of I/O Definition 1 Menu enters
the second page of “the I/O Definition Menu”. This is the second page for Input
Definition.
INPUTPORT SETUP 2
In port 5: [ NO DEFINE ] [ OPEN ]
In port 6: [ NO DEFINE ] [ OPEN ]
In port 7: [ NO DEFINE ] [ OPEN ]
In port 8: [ NO DEFINE ] [ OPEN ]
Press ↑↓key, then press ENT!
72
6105-S5 Feeder Controller Operating and Service Manual
OUTPUTPORT SETUP 1
Out port 1: [ NO DEFINE ] [ OPEN ]
Out port 2: [ NO DEFINE ] [ OPEN ]
Out port 3: [ NO DEFINE ] [ OPEN ]
Out port 4: [ NO DEFINE ] [ OPEN ]
Press ↑↓key, then press ENT!
Pressing “PAGE DOWN” soft key enters the fourth page of the I/O Definition Menu.
2006-3-21 12:00:00 Scale#1
OUTPUTPORT SETUP 2
Out port 5: [ NO DEFINE ] [ OPEN ]
Out port 6: [ NO DEFINE ] [ OPEN ]
Out port 7: [ NO DEFINE ] [ OPEN ]
Out port 8: [ NO DEFINE ] [ OPEN ]
Press ↑↓key, then press ENT!
73
6105-S5 Feeder Controller Operating and Service Manual
OUTPUTPORT SETUP 3
Out port 9: [ NO DEFINE ] [ OPEN ]
Out port 10: [ NO DEFINE ] [ OPEN ]
Out port 11: [ NO DEFINE ] [ OPEN ]
Out port 12: [ NO DEFINE ] [ OPEN ]
Press ↑↓key, then press ENT!
SIMULATED OUTPUT #1
Type: [ PID OUTPUT ]
Range: [ 4-20 mA ]
Delay: [ 0 sec ]
Filter: [ 0 ]
Press ↑↓key, then press ENT!
Move the cursor to “Delay” with Up/Down scroll keys and press ENT key, the
74
6105-S5 Feeder Controller Operating and Service Manual
SIMULATED OUTPUT #1
Type: [ PID OUTPUT ]
Range: [ 4-20 mA ]
Delay:
Filter: [ 0 ]
Press numeric key, then press SURE!
Press numeric keys to input the desired value, and press “ENTR” soft key to confirm.
Use the same method to set the Output Filter.
In the first page of Simulated Output #1 menu, move the cursor to “Type” with
Up/Down scroll keys and press ENT key, the following screen appears:
SIMULATED OUTPUT #1
Type: PID OUTPUT
Range: [ 4-20 mA ]
Delay: [ 0 sec ]
Filter: [ 0 ]
Press ENTR to select, then press SURE!
75
6105-S5 Feeder Controller Operating and Service Manual
Output delay 1s 10
Output filter 0 10
When proper output type such as “No define” is selected, press “ENT” to confirm.
“Output range” operation is same as above.
From the first page of Simulated Output #1, press “PAGE DOWN” soft key to enter
the seventh page of the I/O Definition Menu.
SIMULATED OUTPUT #2
Type: [ PID OUTPUT ]
Range: [ 4-20 mA ]
Delay: [ 0 sec ]
Filter: [ 0 ]
Press ↑↓key, then press ENT!
Set the simulated output port 2 in the same way as mentioned above.
Pressing “PAGE DOWN” soft key enters the eighth page of the I/O Definition Menu.
2006-3-21 12:00:00 Scale#1
SIMULATED INPUT
Input #1: [ MOISTURE ]
Input #2: [ INCLINATION ]
Move ↑↓ key to select submenu, The simulated input value changes among “No
Define”, “Moisture ” and “Inclination”.
Pressing “PAGE DOWN” soft key enters the ninth page of the I/O Definition Menu.
2006-3-21 12:00:00 Scale#1
COMPENSATION EFFECT
Inclination: [ NO ]
Moisture: [ NO ]
76
Move ↑↓ key to select submenu, When press “ENTR” soft key, the output type value
changes between “YES” and “NO ”. When the desired option is displayed, press
“SURE” soft key to input the value.
COMPENSATION EFFECT
Inclination: NO
Moisture: [ NO ]
Pressing “PAGE DOWN” soft key enters the tenth page of the I/O Definition Menu.
2006-3-21 12:00:00 Scale#1
ANGLE COMPENSATION
Min. Angle: [ - 25 ]
Max. Angle: [ 25 ]
ANGLE COMPENSATION
Min. Angle: - 25
Max. Angle: [ 25 ]
77
Use the numeric keys to input the settings firstly, then press “SURE” soft key to
confirm.“+/-”means positive / negtive. “Max. Angle” operation is same as above. (as
per the nineth page of submenu “I/O definition Menu”, “angle compensation” is
selected.)
Pressing “PAGE DOWN” soft key enters the eleventh page of the I/O Definition
Menu.
2006-3-21 12:00:00 Scale#1
MOISTURE COMPENSATION
Min. Moisture: [ 10 % ]
Min. Voltage: [ 2 V ]
Max. Moisture: [ 80 % ]
Max. Voltage: [ 5 V ]
Press ↑↓key, then press ENT!
(as per the nineth page of submenu “I/O definition Menu”, “Moisture compensation”
is selected.)
Pressing “PAGE DOWN” soft key enters the twelfth page of the I/O Definition Menu
OUTPUT DATA
Totalizer: [ 1t ]
Width: [ 0.1sec ]
78
Press ↑↓key, then press ENT!
Press↑↓ key to select submenu, and use numeric key to input value, and then press
ENT to confirm.
Setup range see below table:
Description Range
Totalizer 0 1000000
Width 0 1000000
Pressing “PAGE DOWN” soft key enters the thirteenth page of the I/O Definition
Menu.
MAIN PAGE
MENU UP
Belt clip detector can decrease flow rate error caused by metal plate at the belt
coupler. Normally clip detector is a proximity switch. When it detects clip in manual
method, clip detector keeps flow rate constant, and stable in condition of desired belt
length without influence on totalizing. In automatic method, controller automatically
checks tare weight change and makes compensation accordingly. Auto track must
79
6105-S5 Feeder Controller Operating and Service Manual
be effecitive, cause auto zero track store tare weight in memory for tare weight
compensation of fluctuation point by point.
1. INITIAL SETUP
2. CALIBRATION
3. SYSTEM DATA
4. I/O DEFINITION
5. COMMUNICATION
PREV PAGE
RUN DOWN
COMMPORT 1 SETUP
Function: [ MODBUS ]
Baud Rate: [ 115200 ]
Word Length: [ 8BIT ]
Stop Bit: [ 1BIT ]
Press ↑↓key, then press ENT!
MAIN PAGE
MENU DOWN
Press ENTR soft key to view the optional settings and press SURE soft key until the
desired value is displayed.
Setup range see below table:
Description Range
Function NO DEFINE, MODBUS, PRINT
Baud Rate 4800、9600、19200、38400、115200
Word Length 8 BIT, 7 BIT
Stop Bit 1 BIT, 2 BIT
Pressing “PAGE DOWN” soft key enters the second page of the Communication
Menu.
80
6105-S5 Feeder Controller Operating and Service Manual
COMMPORT 1 SETUP
Parity: [ NONE ]
Address: [1]
Press↑↓ key to select submenu. Press ENTR soft key to view the optional settings
and press SURE soft key until the desired value is displayed.
Description Range
Parity No define, MODBUS, Print
Address 0-255
Pressing “PAGE DOWN” soft key enters the third page of the Communication Menu.
2006-3-21 12:00:00 Scale#1
COMMPORT 2 SETUP
Function: [ MODBUS ]
Baud Rate: [ 115200 ]
Word Length: [ 8BIT ]
Stop Bit: [ 1BIT ]
Press ↑↓key, then press ENT!
Pressing “PAGE DOWN” soft key enters the fourth page of the Communication
Menu.
81
6105-S5 Feeder Controller Operating and Service Manual
COMMPORT 2 SETUP
Parity: [ NONE ]
PRINT SETUP
CONTINUE: [ NO ]
SPAN TIME: [ 10 min ]
SPECIFY: [ NO ]
82
6105-S5 Feeder Controller Operating and Service Manual
selected, controller can set up to 4 regular print point, for specific operations, refer to
the follows:.
Pressing “PAGE DOWN” for further setup, the following 4 menus are timeing print
setup for set print time individually. (time is 24H system).
6. ALARM SETUP
7. LOADOUT SETUP
8. PASSWORD SETUP
9. DIAGNOSTICS
10.STATISTICS
PREV PAGE
RUN UP
ALARM SETUP 1
Alarm 1: [ NO DEFINE ]
Alarm 2: [ NO DEFINE ]
Alarm 3: [ NO DEFINE ]
Alarm 4: [ NO DEFINE ]
MAIN PAGE
MENU DOWN
6105-S5 Feeder Controller Operating and Service Manual
Press ↑↓key to select submenu. Press select input to switch options between:
The options of output port function include: “High Rate”, “Low Rate”, “High Load”,
“Low Load”, “High Speed”, “Low Speed”, “Control error”.
Pressing “PAGE DOWN” soft key enters the second page of the Alarm Setup Menu.
ALARM SETUP 2
Alarm 5: [ NO DEFINE ]
Alarm 6: [ NO DEFINE ]
Alarm 7: [ NO DEFINE ]
Alarm 8: [ NO DEFINE ]
84
6105-S5 Feeder Controller Operating and Service Manual
Alarm delay 1 10
Pressing “PAGE DOWN” soft key enters the fourth page of the Alarm Setup Menu.
85
6105-S5 Feeder Controller Operating and Service Manual
Description Range
High speed 0 1000000
Alarm delay 1 10
Low speed 0 1000000
Alarm delay 1 10
Pressing “PAGE DOWN” soft key enters the sixth page of the Alarm Setup Menu.
MAIN PAGE
MENU UP
86
6105-S5 Feeder Controller Operating and Service Manual
6. ALARM SETUP
7. LOADOUT SETUP
8. PASSWORD SETUP
9. DIAGNOSTICS
10.STATISTICS
PREV PAGE
RUN UP
LOADOUT SETUP 1
Weight: [ 50 t ]
Correct: [ AUTO ]
Pre Value: [ 5 t]
Pre Range: [ 10% ]
87
6105-S5 Feeder Controller Operating and Service Manual
Pressing “PAGE DOWN” soft key enters the second page of the Loadout Setup
Menu.
LOADOUT SETUP 2
Pre Length: [ 5 ]
Delay: [ 2 sec ]
MAIN PAGE
MENU UP
6. ALARM SETUP
7. LOADOUT SETUP
8. PASSWORD SETUP
9. DIAGNOSTICS
88
10.STATISTICS
MAIN PAGE
MENU DOWN
Press the Up/Down keys to move the cursor to the item that to be set, followed by
[ENT] key. Use numeric keys to input the password settings and press [ENT] to
confirm.
Prompt: The defaulted password is null. Enter “0” or just leave it blank to access
enter the next page.
Inputting the correct passwords enters the second page of the Password Menu.
2006-3-21 12:00:00 Scale#1
Correct!
If clear main Tot, press TOT RSET.
MAIN TOT PAGE
MENU RSET DOWN
Pressing “PAGE DOWN” soft key enters the second page of the Password Menu.
89
6105-S5 Feeder Controller Operating and Service Manual
Pressing “PAGE DOWN” soft key enters the third page of the Password Menu.
2006-3-21 12:00:00 Scale#1
Pressing “PAGE DOWN” soft key enters the fourth page of the Password Menu.
2006-3-21 12:00:00 Scale#1
The third and forth page of “Password setup” menu used for password changing and
operation. Firstly, input initial system password in the first page of “password setup”
menu, then input new system password and sure password, if both words are same,
then press “”Save word”, it proves successful.
System password and operate word are similar in method of password change.
2006-3-21 12:00:00 Scale#1
6. ALARM SETUP
7. LOADOUT SETUP
8. PASSWORD SETUP
9. DIAGNOSTICS
10.STATISTICS
PREV PAGE
RUN UP
DIAGNOSTICS 1
Weight: [ 1000000 ]
Speed: [ 200 ] 91
Pulses: [ 10000 ]
6105-S5 Feeder Controller Operating and Service Manual
Referring value:
Description Range value in normal
Weigh channel 1 1048575
Speed channel 0 2000Hz
Total Pulses
6. ALARM SETUP
7. LOADOUT SETUP
8. PASSWORD SETUP
9. DIAGNOSTICS
10.STATISTICS
PREV PAGE
RUN UP
CLASS 1 SETUP
Start Hour: [0]
Start Min: [0]
End Hour: [5]
End Min: [ 59 ]
MAIN PAGE
MENU DOWN
92
6105-S5 Feeder Controller Operating and Service Manual
Press the Up/Down keys to move the cursor to the item that to be set, followed by
[ENT] key. Use numeric keys to input the settings and press [ENT] to confirm.
Pressing “PAGE DOWN” soft key enters the second page of the Statistics Menu.
CLASS 2 SETUP
Start Hour: [6]
Start Min: [0]
End Hour: [ 11 ]
End Min: [ 59 ]
Press the Up/Down keys to move the cursor to the item that to be set, followed by
[ENT] key. Use numeric keys to input the settings and press [ENT] to confirm.
Pressing “PAGE DOWN” soft key enters the third page of the Statistics Menu.
2006-3-21 12:00:00 Scale#1
CLASS 3 SETUP
Start Hour: [ 12 ]
Start Min: [0]
End Hour: [ 17 ]
End Min: [ 59 ]
Press the Up/Down keys to move the cursor to the item that to be set, followed by
[ENT] key. Use numeric keys to input the settings and press [ENT] to confirm.
Pressing “PAGE DOWN” soft key enters the forth page of the Statistics Menu.
2006-3-21 12:00:00 Scale#1
CLASS 4 SETUP
Start Hour: [ 18 ]
Start Min: [0]
93
End Hour: [ 23 ]
End Min: [ 59 ]
Press the Up/Down keys to move the cursor to the item that to be set, followed by
[ENT] key. Use numeric keys to input the settings and press [ENT] to confirm.
Pressing “PAGE DOWN” soft key enters the fifth page of the Statistics Menu.
REPORT STATISTIC
Class: [1]
Date: [1]
Press the Up/Down keys to move the cursor to the item that to be set, followed by
[ENT] key. Use numeric keys to input the settings and press [ENT] to confirm.
Pressing “CLAS STAT” soft key enters the sixth page of the Statistics Menu.
2006-3-21 12:00:00 Scale#1
CLASS STATISTIC
Total: [ 100 ]
If report print function is appled, set “PRINT” in COMM FUNCTION menu and main
menu 5, refer to COMM SETUP section for setup details. As shown in below menu:
2006-3-21 12:00:00 Scale#1
COMMPORT 1 SETUP 94
Function: [ PRINT ]
Baud Rate: [ 115200 ]
Word Length: [ 8BIT ]
6105-S5 Feeder Controller Operating and Service Manual
95
6105-S5 Feeder Controller Operating and Service Manual
Integrator
Maintenance info. covered by this manual will meet your requirement. If technical
support is needed, please dial 086-0516-87885888(87885858)for assistance. To
respond to your questions quickly, please wirte down your machine model and
relative data.
4.2 Daily Maintenance Routines of 6105-S5 Controller
Often observing the operation situation of the device can find potential faults in time,
if indeed there is a problem, please follow the troubleshooting steps to check:
A. Check power supply
· Check that the voltage selector switch is set to the correct position.
· Check that the fuse is good.
· Check that the power is switched ON to energize the 6105-S5.
96
6105-S5 Feeder Controller Operating and Service Manual
· Display board
· Power output board
6105-S5
f il
Visual inspection
No Bad
Display illuminates Check power supply Repair power supply
Yes Normal
Touchkeys act correctly Set up the 6101 again Repair CPU board
Yes Yes
Display fault Record and reset
error
No No
Internal test is normal
Yes
Flowrate indication is No Bad
Check speed input
normal
Normal
Normal Check load cell input
No
Total indication is normal
Yes
No Yes No
Remote Total is normal Check relay actions
Yes No
Current output is normal Repair current board
Yes
End
97
6105-S5 Feeder Controller Operating and Service Manual
probes of the AC voltmeter. Check that the measurement is in line with that specified
on the transformer.
4.3.2 Speed Input
Start
Repair or replace
Speed input
part of CPU
Speed input is normal. board fails
End
See figure 4-2, flow chart of troubleshooting the Speed Input. Get powerd on, when
check speed input, belt is running in normal speed. Speed signal is pulse series sent
by speed sensor, speed pulses can be observed in system diagnostic menu of
controller.
Check that the weight indication displayed in Diagnostics menu of the 6105-S5
controller is between 1 and 16777216. The weight indication should be changed in
the same direction as the sense of desired result. If the weight indication is beyond
the given range, there may be problems with load cell, wire connection or CPU board.
In this case, press RUN key to stop the examination of load cell input.
98
6105-S5 Feeder Controller Operating and Service Manual
Start
No
Normal
yes
The voltage between signal leads of
Damaged load cell
the load cell should be lower than 30mV
99
6105-S5 Feeder Controller Operating and Service Manual
100
6105-S5 Feeder Controller Operating and Service Manual
Start
Connect power supply to the system. Run the belt empty at the normal speed, the
flowrate indication should be zero. Remove the connections on the signal output
terminals which are then connected to a milli-amperometer.
Check the flowrate indication at current output of 4mA, which should be zero.
Otherwise, perform zero calibration.
No Bad current
output board
Yes
The current output is normal – 4mA
Check the current output at a known flowrate. Apply reference weights on the scale
and press RUN key to determine the flowrate at a percentage of the scale capacity.
End
101
6105-S5 Feeder Controller Operating and Service Manual
102
6105-S5 Feeder Controller Operating and Service Manual
103