nvlad
CALIBRATION
NVLAE CAB CODE wosEvE-0
Certificate of Calibration
Fluke Park Laboratory
Description: Field Metrology Well with Process Readout Certificate Number: 86407046
Manufacturer: Fluke Date of Calibration: 07 oct 2016
Model: 9242 Date Due:
Serial Number: a6A868 ‘Temperature: 20.010 26.0°C
Status: As-Found: New Relative Humidity: 20 1070 %RH
As-Left: In Tolerance Pressure: 95 to 103 kPa
Calibration: = Full Issue Date: 07 Oct 2016
Procedure: HCT301-1
Customer: INTERWORLD HIGHWAY LLC
JONESTOWN PA,
PONumber: 9126031
‘This calibration is traceable to the Si through recognized national measurement institutes (NIST, PTB, NPL. NIM,
NRC, etc.) ratiometric techniques, or natural physical constants and is in compliance with ISQAEC?7026-2008
and ANSUNCSL 2540,1. The calibration nas been completed in accordance with the Fluke Corporate Quality
System document OSD 111.0. Calibration certificates without identification of the authorzing person are not cate
This certificate apples to only the item identified and shall not be reproduced other than in full without the specific
wnitten approval by Fluke Corporation
This calibration certificate may contain data that is not covered by the Scope of Accreditation The unaccredited
{est points, where appiicable, are indicated by an asterisk (*). or confined to clearly marked sections The
cortficate shall not be used to claim product certification, approval, or endorsement by NVLAP. NIST. or any
agency of the U.S. Government.
Measurement uncertainties at the time of test are given where applicable. They are calculates in accordance with
the method described in the ISO Guide to the Expression of Uncertainty in Measurement. The reported expanded
{inceriainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage
factor k such that the coverage probability corresponds to approximately 86 %
‘Comments:
es SEY
cere +. seag70es Be Electronically signed by
meer ee oere Rg Approved Signatory
Sr aes0s8 i Ivars tkstrums
7H satan Metrologist
Fluke Corporation Telephone internet ___ Page 1 of 4
6920 Seaway Blvd 877.385.3225, www flukecal.com Rev 20160922"
Everett, WA 98203 USACertificate Number: 86407046 Date of Calibratio:
Standards Used
Description Serial Number Due-pate
35866 Digital therroreter 26032 R
1560 Therroneter, “Black Stack” Base Unit 808223
2562-H Precision Digite] Thennoneten Boso7e ov-2016
2562-H Precision Digital Thermonete> Bian? tow-2036
5628 Plotinu Resistance Tnernonete~ 178e
5628 Platinun Resistance Thermometer 2247 V7 emar-2027
1529-8 Thermoneter, Chud-E4, 4-PRT/Thenaistor 54889 “way-2027
5616 Therristor Prove 890201 26-Nov-2626
5610 Thermistor Probe 8222908 26-Nov-2016
5780A OC Reference Stancare 9595607 23-wov-2016
7498 Stendarc Resistor Set 80002 25-Nov-2676
Field Metrology well Test station 5 nck
Quality Manuals
This calibration has been completed in accordance with’
+ The Fluke Corporate Quality Manual, QSD 111.0, Revision 118, Dated August, 2014 and/or
+ The Fluke 17025 Quality Manual, QSD 111.41, Revision 005, Dated Sept. 2014
The instrument described herein consists of a heat source component and a built-in process readout component
This calibration pertains to both components.
The heat source component was calibrated by direct measurement of generated temperatures using the pertinent
reference standards listed in the "Test Equipment” section of this report. The calioration was performed using test
insert Model 914x-INST a5 described in the user manual. This insert is similar to insert “C° but is designed to
accommodate the test PRTS and aid in the performance of the axial gradient calibration. The calibration data,
internal calibration constants, and uncertainties are shown on the following page(s) of this report. The temperature
‘accuracy testis self-explanatory. The axial differential temperature test is more complex. Due to the nature of the
axial differential temperature characteristic and the influence of the test equipment on the test result, this test
utilizes tolerances which do not precisely match the instrument specification. However, the unique tolerances used
are intended to determine the axial differential temperature tolerance status based on the published specifications,
‘The temperature observations were performed in both increasing and decreasing directions.
The process readout component was calibrated directly using laboratory transfer standards as listed in the "Test
Equipment” section of this report. The calibration data, internal calibration constants, and uncertainties are shown
con the following page(s) of this report.
The calibration uncertainties are shown at a coverage factor of 2 (k=2). All known significant sources of uncertainty
have been considered. Any limitations or remarks pertaining to this instrument and/or calibration are shown below,
Additionally, out of tolerance indications, if any, are identified along with the corresponding data on the data pages
of this report, Calibration uncertainties have been taken into account in the determination of tolerance status using
risk analysis algorithms. When using the instrument in a catibration process, tis recommended that the instrument
specifications be used as the contribution of the instrument rather than the calibration uncertainties, The
instrument tolerances are shown on the report at a confidence interval of 95%
The sections labeled Temperature Stability, Axial Differential Temperature and/or Maximum Hysteresis contain data
that are not covered by the NVLAP Scope of Accreditation
Page 2 of 4Model 9142
Serial No., SBA868
Certificate No 88407046
‘6 Found Data
No As Found Data Reauired
Certificate of Calibration
[— As Left Data
Data 1D: B6280115507327
Calioration Constants
TEMP? 0.099
TEMP2 0.194
TEMP3 0.342
GRADi -0.014
GRAD2 0.016
GRAD3 0.000
_ Control Constants
TEMPPS 2.0
TEMPINT 300
TEMPOER 4.0
Temperature Accuracy
Set-point °C Actual
~~ -25:000 24,984
0.000 0.007
50.000 49.995
100.000 100.013
160,000 149,995
‘Temperature Stability
Observed *C
Setpoint "S| (2 Sigma)
“25.000 0.001
160.000 0.002
Axial Differential Temperature
Sor-point"C
_Target°C Actual °C.
‘0.000 0.000
0.000 0.003
0.010 0.004
0.010 0.015
Etror°C
0.016
0.007
0.008
0.013
0.008
Tolerance °C
20.010
30.010
Error °C.
0.000
0.003
0.006
0.005:
0.200
$0,200
20.200
Uncenamty
+£0.0040
20.0055
PassiFai)
e
Pp
Tolerance *C_Uncertainty
£0,040
40.040
40.040
20.040
20.020
+0020
40.025
20.030
PassiFail
veveD
Passail_
>
Pe
Pp
P
This report shall not be reproduced except i fll without writen approval of Flake Calibration Page 3 ot 4