INSTRUCTION MANUAL DENSIMAT
INSTRUCTION MANUAL &
60.50.371 - Ed. 10 – 09/2015
INSTRUCTION MANUAL
INSTRUCTION MANUAL ALCOMAT 2
INSTRUCTION MANUAL
INSTRUCTION MANUAL
INSTRUCTION MANUAL
Tutte le informazioni riportate in questo manuale sono quelle disponibili al momento della stampa.
Il costruttore si riserva il diritto di apportare modifiche al prodotto in qualsiasi momento senza preavviso.
Si consiglia di verificare eventuali aggiornamenti.
All the information given in this manual is what available at the time of printing.
The manufacturer reserves the right to make changes to the product at any time without notice.
It is advisable to verify if there are any updates.
- Power consumption: 10 VA
- Power source: 220 V - 15% / + 10 % ; optional: 110 V
- Operating temperature range: 15 - 30 °C
THERMOMETER SECTION
- 2 floaters (plungers) calibrated for weight (50 g) and volume (20 ml)
- 2 double wall 70 ml cylinders
- 1 power supply
The volume of floaters, on request, can be certified by the competent European Institutes such as the
L.N.E. of Paris, the I.M.G.C. of Turin, the PTB in Germany etc.
1
Realized with an antimagnetic material, inoxidizable and a diameter lower or equal of 0,20 mm
2
DENSIMAT & ALCOMAT-2 ENGLISH
2. DESCRIPTION
The instrument consists of an electronic hydrostatic balance DENSIMAT, with a sensitivity of fifth
decimal points, which can be connected to an ALCOMAT 2 module that receives and processes the
data.
NOTE: if the Densimat stops working because of an incorrect operation on the keyboard, the
plug must be removed and connected again to the socket while pressing the “+” key. When the
instrument is switched on again the following message appears
Pnor
which indicates the return to use of the parameters of standard functioning.
Figure 1
3
ENGLISH DENSIMAT & ALCOMAT-2
Figure 2
• Insert the connector of the thermo probe in the place indicated in figure 1 (10), keeping the locking
pin in an upright position and pushing it till its complete insertion: the connector is in the correct
position if, trying to pull it out, it does not come. To pull it out press the locking pin.
• Connect (figure 3) the circular socket of the transformer indicated in figure 3 with (6) to the connector
of Densimat indicated with (3) keeping the flat side of pin upwards. Connect the transformer to the
socket (fig. 3 n. 5) of the electric mains.
1- Connection Alcomat 2
2- Serial number
3- Transformer socket
4- Adjustable supports
5- Mains plug
6- Transformer plug
4
DENSIMAT & ALCOMAT-2 ENGLISH
The Densimat keyboard (fig. 1 n. 4) has two displays and three keys. On the left display it appears the
temperature read by thermo probe; on the right display it appears:
a. specific gravity, with the parameters of standard functioning;
b. messages for user (see Appendix A: List of principal messages visualised by Densimat);
c. this is selectable by means of three keys “Mode”, “+” and “-“. The activated functions through
these keys are described in Appendix B: Selecting options from the Densimat keyboard.
directly:
• the specific gravity d20/20 of a liquid
• the density of a liquid at 20°C (by a simple calculation)
N.B. Indications on the best conditions to carry out the measurements with Densimat are indicated in
paragraph 4. The behaviour of the instrument when switching it on and the conditions for optimal
measurement.
2
See R. Bechetti, “Methods of analysis of wines and spirit drinks”, 1999 pub. Gibertini Elettronica
5
ENGLISH DENSIMAT & ALCOMAT-2
It is operated as follows:
1. degas if necessary the samples rich a CO2 by agitation or vacuum filtration or through filter paper;
2. fill the double coated cylinder with the liquid, whose density is to be measured, up to the level (about
70 ml);
3. immerse the thermal probe into the liquid;
4. immerse the clean and dry floaters carefully;
Thermostatic
bath
Gibertini
Figure 4 Elettronica
5. slowly move the floater for a few seconds, with a circular movement, into the cylinder, in order to
obtain a uniform temperature of the liquid (be carefully not to create air bubbles which might falsify
the measurement);
6. once 20°C has been reached (left display) the reading is carried out: e.g. 0.98480
Figure 5
3
For this scope a thermostatic bath (manufactured by Gibertini Elettronica), where the vessel containing the
liquid under examination is be put, is useful.
6
DENSIMAT & ALCOMAT-2 ENGLISH
a. push the cylinder against of centering plate (figure 1 n.8 or figure 5 n. 2);
b. wait that the instrument has brought itself to zero (if this should not occur it carry out a zeroing
manual operation by pressing the key “ - “ of keyboard): on the right display appears
000
c. hang the floater in suspension (figure 6 n.1) and verify that it not touch the internal walls of the
cylinder or the thermoprobe. If necessary, correct the position of the centering plate loosening its
fastening knobs and repositioning it carefully (fig.5 – n. 1).
d. read the specific gravity d20/20 of the liquid on the right display;
0.98480
N.B.- At the end of the determination, always replace the floater in use, clean and dry, on the
appropriate hooks (figure 6 n.2).
Figure 6
1 – Suspension
2 – Hooks to replace
the floaters not in use
7
ENGLISH DENSIMAT & ALCOMAT-2
It is operated as follow:
• determine the specific gravity d20/20 of the distilled water or a water-alcohol solution with a known
titre (procedure mentioned in the precedent paragraph);
• if the read value is 1,00000 (theoretical d20/20 H2O) or coincides with the d20/20 indicated on the bottle
label of the water-alcohol solution then we are in conditions of ideal reading: the measurements can
be carried out;
• if the read value is various(different) from 1.00000 or from the theoretical value d20/20 on the bottle
label then it actives the procedure of " Compensation of floater" since in this situation the readings
are not reliable.
000
hang floater in the suspension, read the value of specific gravity d20/20 of the distilled water or of the
water-alcohol solution with a known titre, on the display appears as for example:
0.99990.
the point on the right of the display shows that the CorrP option is active;
the keys " + " and " - " increase and decrease respectively the value of density read to 5 units at a
time: press the keys until making to coincide the value read on the display with the theoretical value
d20/20 of the distilled water or of the water-alcohol solution used like reference;
the compensation remains memorized even when the instrument is switched off;
remove the suspension and exit from the procedure " Compensation of floater ": press the key
MODE: “CorrP "appears on the display; press the key " - ".
N.B.- When the water-alcohol solution with a known titre is used, it is advisable, in order to stabilize the
value, to wait a few seconds more before reading the specific gravity.
8
DENSIMAT & ALCOMAT-2 ENGLISH
At the moment when Densimat is switched on, the following message appears on the right display:
dx-xx
where x-xx are the numerical values that indicate the installed release on instrument, giving therefore
an useful indication for technical service if it is necessary.
After around two seconds the following message appears:
HELLO
and immediately after this all segments of both displays for the test are switched on and flashing for a
few seconds.
In this case the following message appears:
rECAL
successively flashing for a few seconds, to indicate that it is carrying out the first phase of a calibration.
The appearance of the following messages on the display
LOAd e CALC
indicates the passage to the second phase of calibration.
At the end of calibration and in the conditions of rest, the right display shows:
000
to indicate that the instrument is in zero condition, or
- 000 e + 000
to indicate that it is near zero.
At the same time on the left display it appears firstly the message “HI”, and after this the value of
temperature (in °C) recorded by thermo probe.
The instrument automatically, after approximately 10 minutes of rest, starts the calibration procedure: on
the display of right appears that as already described previously. During the calibration cannot be use
the instrument.
At the moment of switching on of the Densimat a period of ten minutes of thermal stabilization is
needed4 during which the instrument carries out automatically a calibration each minute and, even
if possible, it is not advisable to carry out measurements.
After this period of thermal stabilization Densimat it is in the better conditions to work.
Other periods for thermal stabilization are not requested until the instrument remains connected to the
mains/electricity, also in the case when module ALCOMAT 2 is switched off. It is advised therefore to
maintain the Densimat always electrically connected.
ATTENTION:
• carry out always the measurements in the zero condition of the instrument;
• for eventual anomalies found during the installation of the instrument (as for example on display
appear the flashing messages "- HI -" or "UnLd" or "Lo") consult the Appendix A: List of the principal
messages showed by Densimat.
4
The necessary time for the thermal stabilization varies in function of the difference of temperatures between
Densimat and the ambience
9
ENGLISH DENSIMAT & ALCOMAT-2
The fixed messages warn the operator that some process is in progress. No intervention is needed.
The flashing messages warn of an anomalous condition or they ask for an intervention of the
operator.
---- Zero setting phase. It is activated by pressing the key " - ".
Weight not stable during the zero setting phase. Press the key " - " to cancel. The
previous zero value is valid.
Er18 If it occurs at the switching on of the instrument, it is not possible to cancel it.
Switch off for some second and switch on again; if it appears again, the
instrument is broken or exists an excessive ventilation.
Weight not stable during a calibration phase. Press the key " - " to cancel. The
Er19
previous calibration value is valid. The cause can be an excessive ventilation.
Calibration error. Press the key " - " to cancel. The previous calibration value is
ErrC
valid.
Suspension overloaded. Remove the cause to cancel.
- HI - If the suspension is engaged during the 2 phases: rECAL and LOAd (see below)
the display shows - HI -. Remove the floater. in the end shows ErrC.
Indicates the phase during which the internal calibration mass is unloaded.
UnLd
It can indicates an overload on the suspension; remove the cause to continue.
Zero setting phase during a calibration. Flashing message and subsequently fixed
RECAL
for a few seconds.
LOAd Indicates the phase during which the internal calibration mass is loaded.
Indicates the phase during which the internal calibration mass is weighed and the
CALC calibration factor is being calculated. After this, appears " 2,49995 " follow from
“000” or the ErrC message (see above).
Fuori The internal diagnostic of the instrument detects an anomaly that could impair its
SCALA accuracy. It is advisable to do a careful check. Press the key " - " to continue.
This is appeared when the instrument is switched on again holding(keeping)
Pnor pressed the key " + " to indicate the return to use of the parameters of standard
functioning, as previously explained.
10
DENSIMAT & ALCOMAT-2 ENGLISH
NOTE: if the Densimat stops working because of an incorrect manoeuvre on the keyboards,
remove the plug and connect it again in the socket keeping pressed the key “+”. As soon as the
instrument is switched on again the following message appears:
Pnor
11
ENGLISH DENSIMAT & ALCOMAT-2
7.1 INSTALLATION
• Put in place the second centre-cylinder over the one already existent and fix it with the relative
knobs.
• Fix the support of floater with the two screws (Figure 8).
The function of support is to hold the floater when it is not used, allowing therefore the Densimat to carry
out the automatic calibration.
• Fix the cylinder on the Densimat with the spring as shown in figure 7.
• Connect the tubes to the cylinder for measurements in continuo.
• Insert into the inside of the cylinder the appropriate "net to break bubbles", pushing it to the bottom.
The "net to break bubbles", as the name suggests, has the function to avoid the formation of
bubbles.
• Insert the tail pipe (Figure 7 n.6) in an appropriate container.
• Verify that the floater moves easily in the centre of the cylinder.
12
DENSIMAT & ALCOMAT-2 ENGLISH
Figure 8
• Put the floater in place, to the inside of the cylinder for measurements in continuous, on the "support
for floater".
• Close the tap (Figure 7 n. 5)
• Fill up the cylinder pouring the liquid into the funnel. The excess liquid is eliminated through the
“overflow” tube (Figure 7 n. 4)
• The liquids under examination must have the same temperature as the floater. Being careful not to
create air bubbles adhering to the floater or to the wire of the floater which might falsify the
measurement.
• Hang the floater in suspension and read the specific gravity on the display.
• At the end of the measurement replace the floater on the support for floater (Figure 7 n.1) and open
the tap to empty the cylinder.
13
ENGLISH DENSIMAT & ALCOMAT-2
Figure 9
• The weight of the pan (10 g) must be zeroset: in this case the available capacity is 50 g.
• Carry out the measurements at 10, 20, 30, 40 and 50 g (maximum tolerance ± 1 mg).
• To read again the density on display:
press MODE until the display shows "PAr" and confirm with key " + "; on the display appears
"P4=06"
press the key " + ": on the display appears " P4=00 ".
press the key " - "; on the display appears " 000 ".
IMPORTANT ! - When on the Densimat display the value in grams is visualized, the instrument does
not carry out the automatic calibration that therefore must be carried out by the keyboard.
For that purpose press the key MODE until the display shows " CAL " and confirm with the key "+".
14
DENSIMAT & ALCOMAT-2 ENGLISH
The use of the Densimat hydrostatic balance, to determine the Alcoholic Strength by Volume (alcoholic
strength), is included. in the utilizations of the "weighing instruments " that involve the obligation of CE
seal (art. 2 subsection 2 letter a) of the EEC 90/384 Directive acknowledged with the Legislative Decree
517/92).
The mod. Densimat hydrostatic balances are sealed " CE ": the Metric Central Office of Rome has
issued an approval certificate CE of the type n. the 96-011, in virtue of this the Gibertini company
provides the affixing of the nameplate " CE " with the Lead Seal of the Metric Office ("seal") on the
balance.
The balance must be verified periodically by the Metric Office of the competent local Chamber of
Commerce .
15
ENGLISH DENSIMAT & ALCOMAT-2
• Absorbed power: 10 VA
• Mains voltage: 220 V -15% +10%; on request: 110 V
• Ambience temperature of use: 15 ÷ 30 °C
• Fuse: 160 mA T
• Alphanumeric display of liquid crystals
• Visualization of messages in different languages (I, D, F, E, GB)
• Printer output
• Menu functions shown on display
MUSTS EVALUATION
Additional equipment
• 1 serial cable
• 1 Alcomat 2 mains cable
• Supplementary output RS232
Optional accessories
• Printer
16
DENSIMAT & ALCOMAT-2 ENGLISH
4
3
Connect ALCOMAT2 to the mains introducing the cable of the transformer into the place indicated with 4
(figure 10); switch on the instrument with the switch (Fig. 10 n.6). The cable for the printer must be
connected to the connector indicated with 1 in figure 10. Make sure that the screws are screwed to the
bottom.
N.B.- After some minutes, in which ALCOMAT 2 must not be used, the display is switched off
automatically. Pressing [ ON ] the display is switched on again.
17
ENGLISH DENSIMAT & ALCOMAT-2
N.B.- For ease of writing, in the text that follows, we use the generic function key corresponding to a
determined text present on the display with the corresponding text; as for example on the PRINCIPAL
MENU the key corresponding to [ 1 ] is indicated with [ 1 ].
operating to a temperature also different from 20 °C, provided that it is in the interval between 15
and 25 °C.
18
DENSIMAT & ALCOMAT-2 ENGLISH
It is operated as follows:
• press the key [ 1 ] to select “Alcoholic Strength by Volume”. On the display of ALCOMAT 2 appears:
Temperature
read from
thermo probe
of Densimat
• fill up the double coated cylinder (see figure 4) with distillate or the water-alcohol solution, to carry
out the measurements of specific gravity at 20 °C and the ASV, until reaching the mark (70 ml);
• place(put) the cylinder on the Densimat;
• immerse the thermo probe into the liquid;
• immerse carefully the floater clean and dry;
• agitate (for a few second) floater into the liquid (figure 5), slowly, with circular movement, in order to
uniform the temperature (being careful not to create air bubbles adhering to the floater or the wire of
floater which might falsify the measurement);
• carry out the reading:
a. push the cylinder against the appropriate centre-cylinders (figure 1 n.8 or figure 5 n.2);
b. attend that display of ALCOMAT 2 shows the “zero condition” or the “zero” of the hydrostatic
balance (figure 13); if this should not occur (+000 or -000) carry out a manual operation of
zeroing:
press the key [ zero ] (figure 13);
on the display appears “zero” therefore it is returned to the precedent view
c. hang the floater in suspension (figure 6 n.1) and verify that it not touch the internal walls of the
cylinder or the thermo probe. Correct the position of the centre-cylinders, if it is necessary,
loosening the knobs that blocks it(7 figure 1n.7 or figure 5 n. 1) and put in place it again in an
appropriate mode;
d. read the ASV value on display (figure 14):
The ASV and the specific gravity at 20 °C appear subsequently on the display to the read
specific gravity; the figures(numbers) change until when the data not reach the stability.
19
ENGLISH DENSIMAT & ALCOMAT-2
Specific gravity of
the distillate at 19.4
°C measured by
Densimat
Specific gravity at 20 °C (d20/20) of distillate
Figure 14 – ASV and d20/20 of a distillate
At the moment when the data reaches stability, the value of ASV is visualized with
contrasting characters compared to the background and the instrument emits a short
acoustic signal. The data remains fixed on display until the removal of floater from
suspension.
The read specific gravity is corrected at 20°C in accordance with the tables given by the
official method contained in the EEC Regulation n. 2676/90 and n. 2870/00). The correction
is reliable only for the water-alcohol mixtures and in a temperature interval of the liquid under
examination comprising between 15 and 25 °C. Outside the mentioned interval the results
may have variations.
• At the end of the determination, always replace the floater in use on the appropriate hook (figure 6
n.2).
Temperature of
liquid
Specific gravity
of liquid at 19.4°C;
operating at 20 °C
the READ
SPEC:GRAV. is the
d20/20 of liquid
Figure 15 – Specific gravity of a liquid
20
DENSIMAT & ALCOMAT-2 ENGLISH
ASV or alcoholic
strength
calculated from
the read specific
gravity
Specific gravity of
water-alcoholic
mixture read with
Densima
21
ENGLISH DENSIMAT & ALCOMAT-2
• fill the cylinder with 70 ml of distilled water in a temperature comprises between 15 and 25°C,
immerse the thermo probe, the floater and agitate;
• from principal Menu, by pressing (1) (figure 12), select “Alcoholic Strength by Volume”, then press
CORREC to enter in option “Correction”, press(H2O) (figure 16);
• wait the “zero” of the balance and hang the floater on the suspension;
• the values of the factor of current correction, the specific gravity that the water should have at the
temperature measured by thermo probe and the specific gravity of the water measured effectively by
the instrument appear on the display (figure 17);
• if these two values are equal then the instrument is in condition of ideal reading; in the opposite case
it is necessary to change(vary) the factor of correction of floater.
Press the keys [+] and [–] taking in consideration that key [+] increase the value of specific gravity
read and the key[–] decrease it.
• At the end of the determination, always replace the floater in use on the appropriate hook (figure 6
n.2).
Specific gravity
of water
measured with
Densimat
at 19.5°C
Specific gravity
of water
at 19.5 °C
from table
22
DENSIMAT & ALCOMAT-2 ENGLISH
During the printing, the following flashing message appears on the ALCOMAT 2 display:
Date: 06-03-2001 Hour : 16:07 Pr. 012 Date: 06-03-2001 Hour : 15:20 Pr. 001
Alcoholic .strength. [%VOL] : 13.24
N.B. - On the printed label appears an identification number indicated with “Pr”, that is “progressive
label”. When the instrument is switched on, the numbering starts from 1. Every time a label is printed, the
numbering is automatically increased.
It is possible to modify the numbering ; for more details see paragraph 11.4.1 Modification of the
progressive label.
ZEROSET IMPOSSIBLE
23
ENGLISH DENSIMAT & ALCOMAT-2
The utilization of Alcomat 2 with a Densimat allows one to determine the value of the total dry extract of
a wine or a must for specific gravity including values from 0.990 and 1.160, in a temperature range of 15
°C and 25 °C. Under 15 °C and over 25 °C the results are not completely reliable.
Proceed as follows:
• by pressing [2] (figure 12), select TOTAL DRY EXTRACT from the page of Principal Menu. The
instrument awaits the measurement of specific gravity of a given wine/must(valid sample with a
specific gravity comprised between 0.99 e 1.16) with a flashing message:
Figure 18
• wait the “zero” of balance or press [ZERO] to a manual zeroing operation, hang then the floater on
suspension;
• on the display appear: temperature, specific gravity read, specific gravity corrected at 20°C and the
density at 20°C of the sample of must/wine under examination (figure 19).
Density of a
given
wine/must
Recall the sample Memorize the sample Pass directly to the calculation of
of wine/must of wine/must DRY EXTRACT
Figure 19
24
DENSIMAT & ALCOMAT-2 ENGLISH
ALCOMAT 2 provides automatically to bring the specific gravity read to the corresponding value at
20°C.
On reaching the stable value, the value of corrected specific gravity at 20°C (expressed to a fourth
decimal point ) appears on the display with black characters on a white ground and with a short acoustic
signal.
The value of density is obtained by multiplying the value of corrected specific gravity at 20 °C with
0.998203 g/ml (density of water at 20 °C).
To recall a sample:
Press [ RECALL. ] or [ RECALLS ],
by means of keys [ + ], [ - ], [ +10 ], [ - 10 ] select the identification number of sample to recall,
confirm the choice pressing [ RECALL. ] and pass to the determination of the dry extract pressing
[ CALCULATION ].
Specific gravity
of wine/must as
soon as read,
otherwise
“000” appears
Wait for the “zero” or make manual zeroing before the reading of
distillate
Figure 20
wait the "000" of the balance, hang therefore floater on the suspension; on the display appear:
temperature, read specific gravity and Alcoholic Strength by Volume of the distillate together with the
value of total dry extract in g/l of saccharose (figure 21).
ALCOMAT 2 automatically shows the ASV of the distillate, using the valid algorithm of correction for
water-alcohol mixtures, and subsequently the same specific gravity d20/20.
ALCOMAT 2 applies the formula of Tabarié to determine the specific gravity d20/20 of the dealcoholized
wine/must, then it gives the corresponding tabular value5 of total dry extract.
5
For further information see R. Becchetti, “Method of analysis of wine and spirit drinks”, 1999 pub. Gibertini
elettronica
25
ENGLISH DENSIMAT & ALCOMAT-2
On reaching the stable value, the value of total dry extract appears on the display with black characters
on a white ground and with a short acoustic signal.
Figure 21
• Once the value of total dry extract is calculated it is possible to print the relative label (see the
following paragraph) and/or to cancel from the memory the sample of wine/ must as soon as
examined by pressing [ CANCEL. ].
• To exit from the function of total dry extract press [ EXIT ] until the PRINCIPAL MENU appears on the
display.
• At the end of the determination, always replace the floater in use on the appropriate hook (figure 6
n.2).
If the printer is not connected, the following flashing message appears on display :
If the printer is connected but the stable value of dry extract is not reached the following message
appears on display:
26
DENSIMAT & ALCOMAT-2 ENGLISH
The use of ALCOMAT 2 with Densimat allows the automatic determination of the density and the value
of the degrees Baumé, Babo, Brix and Oechsle of a must, for specific gravity d20/20 comprised between
1.04 and 1,16, in a range of temperature comprised between 15 °C and 25 °C. Under 15 °C and over the
25 °C the results are not completely reliable.
It operates as follows:
• In the case when the must is not limpid, filtration is necessary.
• Select MUSTS EVALUATION from the PRINCIPAL MENU pressing [ 3 ] (figure 12). Wait for the "
zero " of the balance and then hang the floater in suspension. On the display appears: temperature,
read specific gravity, corrected specific gravity at 20 °C (with four figures) and density at 20 °C of the
sample of must under examination (valid sample for d20/20 comprised between 1.04 and 1.16, figure
22).
• After a few seconds, on reaching the stable value of the specific gravity, the numerical values
corresponding to the degrees Baumé, Babo, Brix and Oechsle of must appear on the display,
preceded by a short acoustic signal.
Figure 22
6
For further information see R. Becchetti, “Method of analysis of wine and spirit drinks”, 1999 pub. Gibertini
elettronica
27
ENGLISH DENSIMAT & ALCOMAT-2
11.3.1 Print of the value of the degrees Baumé, Babo, Brix and Oechsle of a must
For having a relative print to the evaluation of a must enough to press [ PRINT ]; at the moment of
determination the option appears on the display.
During the print the following flashing message appears on the ALCOMAT 2 display:
If the printer is not connected , the flashing message appears on the display:
PRINTING LABEL PLEASES WAIT
PRINTER ERROR
PRESS ANY KEY TO CONTINUE
If the printer is connected but the stable value of specific gravity is not reached the following message
appears on display :
28
DENSIMAT & ALCOMAT-2 ENGLISH
Figure 23
11.4.5 Language
To change the language with which all the messages that appear on the display are visualized, select "
Settings" from the PRINCIPAL MENU by pressing [ 4 ] (figure 12), then select "Language" pressing [ 6 ].
It then selects the desired language (figure 24) from Italian, English, French, Spanish and German by
pressing the corresponding key. Press [ 1 ] to exit.
Figure 24
All the messages appeared on the display will be visualized in this new language. The language
selection will be maintained even when the instrument is switched off.
29
ENGLISH DENSIMAT & ALCOMAT-2
followed by
During the first and second phases of calibration the instrument cannot be used.
30
DENSIMAT & ALCOMAT-2 ENGLISH
Figure 25
Figure 26
At this situation any key pressed (figure 26) brought to the PRINCIPAL MENU, that is that ALCOMAT 2
connected to a balance in phase of thermal stabilization cannot be used to carry out the
measurements, while the Densimat keyboard remains active.
31
ENGLISH DENSIMAT & ALCOMAT-2
If the keyboard is not touched, the flashing message remains on the display until the Densimat is in
phase of thermal stabilization (about 10 minutes).
At the end of this phase(thermal stabilization) the following flashing message appears on the ALCOMAT
2 display
followed by
At this point the display of ALCOMAT 2 is switched off, while the keyboard of Densimat continues to
remain active.
Pressing 3 times the first key on the left of the ALCOMAT 2 keyboard, the displays and the keys of the
Densimat become deactivated and all the functions of the instrument can be carried out only by the
keyboard of ALCOMAT 2.
The PRINCIPAL MENU appears on the display of ALCOMAT 2.
In order to work only with the Densimat, to reactivate the displays and keyboard it is necessary to switch
off and to switch on again the instrument keeping ALCOMAT 2 switched off.
N.B. - If the Densimat at the switching on shows the flashing message "Fuori SCALA", press the key "-"
on the Densimat to continue. ALCOMAT 2, connected with it, shows the flashing message "Balance in
phase of thermal stabilization. Wait". At the end of thermal stabilization ALCOMAT 2 with Densimat it is
operating. In case that the message for thermal stabilization persists for beyond 15 minutes press again
the key " - " and have the DensiAlcomat 2 checked by a technician.
32
DENSIMAT & ALCOMAT-2 ENGLISH
The symbol of a crossed box applied on the apparatus or on the packaging indicates that the product
must be collected separately from other waste materials at the end of its useful life.
The separate waste collection of the apparatus which has reached the end of its useful life is organised
and managed by the producer. The user who desires to get rid of present apparatus must therefore
contact the seller and follow the given instructions.
Suitable separate waste collection for future sending of the disused apparatus for recycling, treatment
and environmentally friendly disposal, contributes towards preventing any possible negative effects on
the environment and on health and encourages the reuse and recycling of the materials the apparatus is
made of.
Unauthorised disposal of the product by the user will lead to payment of the administrative sanctions in
force in the country were it is put on the market.
33