03 - Ionization chambers

Jaroslav Adam

Czech Technical University in Prague

Version 2

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 1 / 130


Principle of operation

Charged particle getting through a volume of a gas or noble liquid


Interaction proceed through ionization and end excitation of the molecules, electron-ion pairs
are created
Ion can be created directly by the incident particle, or by the δ-electrons, when the energy
from the primary particle if first transferred to the electron which acquires enough energy to
make further ionization
Electric field is applied by the electrodes, electrons and ions drift to them
If the field is high enough, drifting electrons can also ionize the gas (proportional counters)
After further increase of the field strength, electrons emit UV light on the anode
(Geiger-Muller counter)
Electronic signal at the output, pulsed or current regime
Position sensitivity by segmenting one of the electrodes (xy ) and by timing measurement (z)

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 2 / 130


Ionization detectors without amplification

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 3 / 130


Number of ion pairs

Minimum energy W to be transferred from the incident particle to create at least one ion pair
Quantified as average energy loss to create the pair
Given by the least tightly bound shell, W = 10 - 25 eV
Non-ionizing energy loss (excitation) makes number of pairs lower
Fully stopped 1 MeV particle produces 30 000 ion pairs
Number of pairs is important for resolution

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 4 / 130


Energy dissipation per ion pair (the W-Value)

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 5 / 130


Fano factor

Fluctuation in number of pairs affect the energy resolution



The simplest approach postulate Poisson statistics for number of ion pairs (σ = N)
Fano factor makes correction to predicted variance to get observed variance
Fano factor = 0 if all incident energy converted into pairs, no statistical fluctuation

In gases Fano factor < 1, Poisson distribution valid but fluctuations are smaller than N
Significant in pulse mode

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 6 / 130


Principle of ionization chamber

Figure : Planar ionization chamber

Suppose two metallic electrodes at distance D covering volume of a gas or noble liquid
Voltage V applied to anode (thousands of kV)
Number of electrons n− given by the number of minimum-ionizing-particles (mip)

n dE
n− = Dρ (1)
W dx

ρ is the density and dE/dx is energy loss per g cm−2

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 7 / 130


Charge carriers in gas/liquid volume

Several processes applies to the ion pairs


Drift movement by external electric field
Diffusion due to random thermal movement
Charge transfer
Recombination

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 8 / 130


Drift movement

Electrostatic force moves the charges, positive ions opposite to electrons and negative ions
Drift of electrons characterized by drift velocity in electric field E = V /D

dx µe V µe
vd (e) = = E= (2)
dt p pD

Electron mobility µe given in unit of bar cm2 V−1 s−1 , p is the gas pressure
Mobility of ions is about 1000 less than of electrons
Description with mobility provides calculation of readout times
ms for ions, µs for electrons

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 9 / 130


Saturation of the drift velocity

Figure : Electron velocity vs. filed

No increase in drift velocity after reaching it’s maximum, only in some gases
Hydrocarbons, argon-hydrocarbon mixtures
In non-saturation gases, E/p proportionality holds up to high fields

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 10 / 130


Diffusion

Thermal movement with mean free path of about 10−6 - 10−8 m


More important for electrons since their thermal velocity is bigger
Point-like collection of electrons form Gaussian spatial distribution widening with time
Widening in one direction x, y or z given by diffusion coefficient D


σ= 2Dt (3)

D is given by kinetic gas theory or the process is described by a transport model

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 11 / 130


Charge transfer collisions

Electron transfer from neutral gas molecule to positive ion in mutual collision
Significant in mixtures, net positive charge transfered to species with lowest ionization energy
Negative ion can be formed by capturing of free electron (oxygen)

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 12 / 130


Recombination

Free electron captured by positive ion making ordinary neutral atom


Positive and negative ions recombine, most probable compared to electron and ion case
Original charge is lost, no contribution to final signal
Recombination rate given by density of positive and negative species n+ and n− and
recombination coefficient α

dn+ dn−
= = −αn+ n− (4)
dt dt

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 13 / 130


Columnar (initial) recombination

Electron-ion pairs created in column along particle trajectory


Recombination of electron with it’s parent ion
Mainly for heavy ionizing particles (low energy α), minimal for mip
Independent of interaction rate
Recombination may occur with neighboring ion when electrons are drifting in electric field
Depends on the angle between incident particle and electric field

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 14 / 130


Volume recombination

Recombination during drift towards electrodes


Unlike initial/columnar recombination, this depends on irradiation rate
Suppressed by fast charge separation and collection -> high electric field

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 15 / 130


Ionization current, DC ion chamber

Constant irradiation rate creates constant formation of ion pairs


Steady-state current is measure of the rate
Supposing negligible recombination and efficient charge collection

Figure : Planar ionization chamber

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 16 / 130


Current-voltage characteristics

Electric field by external voltage


Current in the circuit equal to ionization current at equilibrium
Increasing voltage begins to separate the charges that would recombine
High electric field makes recombination negligible
After ion saturation, all charges are collected, on increase in current when increasing the
voltage
Standard operation of ion chambers, current in the circuit is an indication of the rate of ion
pairs formation

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 17 / 130


Saturation current
Recombination, especially columnar recombination require high voltage
Volume recombination important at high irradiation intensity

Higher voltage required to get true saturation current


More important in neutron measurement, where heavy fragments are detected
With chambers filled by ambient air, recombination depends on humidity

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 18 / 130


Perturbations in current due to diffusion

Imbalance in steady-state situation supposing uniform production of ions within the chamber
Larger concentration of positive ions close to cathode, opposite for electrons close to anode
Gradient in concentration formed, diffusion opposite to drift
Perturbation in measured current in planar chamber given by

∆I kT
− = (5)
I eV

 is ratio of average energy of charge carrier, kT /e ≈ 2.5 × 10−2 V at room temperature


 close to one for ions, but several hundreds for electrons

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 19 / 130


Losses of saturation current due to diffusion

 close to one for ions, but much larger, several hundreds for electrons
Minimized by high voltage
Columnar recombination not fully eliminated
Separate measurements of ionization current as a function of voltage
1/I as a function of 1/V to determine true saturation current

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 20 / 130


Operation of DC ion chamber

No special requirement on gas since negative charge can be collected as free electrons as
well as negative ions
Only recombination could affect the amount of charge, suppressed by high enough voltage
Few centimeters and tens of hundreds of volts sufficient to reach saturation
Air for gamma-ray exposure, denser gases like Argon to increase ionization density
Pressure about 1 atmosphere, higher to increase sensitivity

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 21 / 130


Geometry of ionization chamber

Electric field given by geometry of electrodes


Uniform electric field with planar geometry
U0
Cylindrical geometry with electrical field as E(r ) = r ln(ra /ri )

Figure : Cylindrical ionization chamber

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 22 / 130


Insulators

Small values of ionization current, < 10−12 A


Resistance of insulator at least 1016 Ω to keep leakage current below 1% for U = 100 V
Leakage by moisture absorbed on surface suppressed by the guard rings and smooth surface
of insulators
Plastics or ceramic for higher irradiation

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 23 / 130


Measurement of gamma-ray exposure with ion chambers

Amount of charge in air-filled ionization chamber


Charge is measure of exposure, ionization current gives exposure rate
Requires to measure ionization of all secondary electrons, mean free pairs several meters
compensation for secondary electrons needed

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 24 / 130


Compensation by free-air ionization chamber

Collimated X-rays, compensation in horizontal direction


Compensation of electrons escaping the sensitive volume by electrons emerging elsewhere
Up to gamma-ray of 100 keV

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 25 / 130


Air equivalent compensation

Reduces space requirements for higher gamma-rays energies, wall thickness less than 1 cm
Exposure rate given by saturation current Is and mass M in active volume

Is
R= (6)
M

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 26 / 130


DC chambers for radiation survey

Radiation monitoring
Saturation current in closed volume of several cm3
Dose measured by charge integration
Initial charge on the chamber, drop in voltage measures total integrated ionization charge

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 27 / 130


Ionization dosimeter

Cylindrical air capacitor, initially charged at some voltage V0


Discharged by the radiation
Voltage reduction measures absorbed dose

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 28 / 130


Calibration of radiation source

Saturation current depends only on geometry, long time stability


Current from unknown source compared with standard under same geometry
Chambers with thousands of cm3
Pressured gas for higher sensitivity

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 29 / 130


Geometry for calibration of gamma-ray sources

Calibration for gamma-ray sources, ionization current for small displacement of the source

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 30 / 130


Measurement of radioactive gases

Radioactive gas as constituent of fill gas, sampled as continuous flow


Ionization current I given by average energy deposition per gas disintegration Ē and total
activity α

Ēαe
I= (7)
W

Principle of smoke detectors where ionization current from internal alpha source decreases
due to presence of the smoke in sampled air

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 31 / 130


Remote sensing of ionization

Positive and negative ions in the air survive for minutes before recombination
Flow of air is transported to a chamber outside the source of radiation
LRAD - Long Range Alpha Detector:
Sample of alpha-contaminated material in container, air flowing though it
Positive and negative ions carried into an ion chamber

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 32 / 130


Pulse mode operation

Each radiation quantum provides with distinguishable signal pulse


Application in radiation spectroscopy
Alpha spectrometry
Neutron detection

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 33 / 130


Equivalent circuit of ion chamber

Chamber is represented by it’s capacity C


Voltage V0 on load resistor R
Drifting charges from ionization create induced charge on electrodes
Voltage drops from equilibrium V0 , giving output pulse VR
Slow return to equilibrium according time constant RC
If RC > time to collect all charges, amplitude of the pulse measures the original charge

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 34 / 130


Equivalent circuit of ion chamber
The ion chamber behaves as parallel connection of resistor and capacity, called RC circuit

Time dependence of voltage V given by


capacity C and resistivity R

dV V
C =I= (8)
dt R

Response to initial charge giving voltage V0


is

V (t) = V0 e−t/RC (9)

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 35 / 130


Electron sensitive mode

Time to collect electrons ∼ µs, time for ions ∼ ms


Collection of all charges would require RC of orders of >ms
This would mean low pulse rate only and interference from microphonic signals
For electron sensitivity, RC between electron and ion collection time
Negative ions no longer allowed

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 36 / 130


Pulse shape in planar chamber

Constant electric field E = V /d


Uniform equipotentials parallel to electrode surfaces
Ions supposed to be formed at equal distance x where electric potential is Ex

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 37 / 130


Pulse shape of electron and ion collection

Derivation of VR (t) based on energy conservation


RC long enough to collect all charges
Energy in capacitance 1/2CV02 is energy source to move the charges
Drift velocity v − for electrons and v + for ions, n0 is number of ion pairs

1 1
CV02 = n0 eEv + t + n0 eEv − t + CVch 2
(10)
2 2
Vch V0
V0 + Vch ' 2V0 and ' (11)
d d

n0 e +
VR = (v + v − )t (12)
dC

After collecting the electrons (time t − ≡ x/v − )

n0 e +
VR = (v )t (13)
dC

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 38 / 130


Pulse shape of electron and ion collection

After collecting the ions (time t + ≡ (d − x)/v + )

n0 e
VR = [(d − x) + x] (14)
dC
n0 e
VR = (15)
C

If RC  t + , maximum amplitude of the pulse is

n0 e
Vmax = (16)
C

Independent of position of incident ionization

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 39 / 130


Pulse shape of electron and ion collection

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 40 / 130


Pulse shape in electron sensitive operation

Pulse shape given by electrons


Amplitude sensitive to position of incident radiation

n0 e x
V |elec = · (17)
C d

Monoenergetic radiation produces a range of pulses


Removed by dividing the ion chamber by adding a grid

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 41 / 130


Gridded ion chamber

Designed to remove amplitude position dependency


All interactions between the grid (Frisch grid) and the cathode
Grid at intermediate potential and transparent for electrons
Grid-anode voltage drops after electrons pass through the grid

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 42 / 130


Electron signal in gridded ion chamber

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 43 / 130


Electron signal in gridded ion chamber

With RC larger than electron collection time and grid-anode spacing d, signal on load resistor
is

n0 e −
VR = v t (18)
dC

Maximum voltage is

n0 e
Vmax = (19)
C

Allows to operate at short RC


Pulse independent of position of incident radiation
But signals of order of < 10−5 V
Can not be measured directly, amplification needed

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 44 / 130


Statistical limit to energy resolution

Number of charges n0 and fano factor F gives statistical limit


Supposing fully stopped α at energy E = 5.5 MeV in gas with W = 30 eV/ion pair and
F = 0.15

Ed
n0 = = 1.83 × 105 ion pairs (20)
W

Standard deviation as square root of variance


p
σn0 = Fn0 = 166 (21)

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 45 / 130


Statistical limit to energy resolution

FWHM of Gaussian distribution for n0 and particle energy

FWHM(n0 ) = 2.35σn0 = 390 (22)


FWHM(E) = 2.35σn0 · W = 11.7 keV (23)

Relative energy resolution to the deposited energy

2.35σn0 W
R= = 0.123% (24)
Ed

Good theoretical resolution can not be achieved due to electronic noise

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 46 / 130


Charged particle spectroscopy with pulse-type ion chamber

Some advantages vs. semiconductor or scintillator detectors


Arbitrary size and geometry
Gas dielectric cable 3500 m long was used as a beam-loss monitor at SLAC
Gas pressure to adjust stopping power or effective thickness
Radiation resistant, simple design
Low level alpha particles, cross sectional area 500 cm2
Bragg curve spectroscopy: analysis of pulse shape from incident particle parallel to the field,
can distinguish particle species

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 47 / 130


Alpha spectroscopy with ionization chamber

Fisch grid ionization chamber filled by argon allows to separate two uranium isotopes

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 48 / 130


Proportional counters

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 49 / 130


Charge multiplication in the gas

Late 1940s, effect of gas multiplication


More ion pairs compared to ionization chamber
Low energy X-rays
Neutron detection
Pulse mode most common operation

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 50 / 130


Avalanche formation

Kinetic energy of the electron sufficient to make secondary ionization


Field at least 106 V m−1
Gas multiplication: secondary electrons also ionize after acceleration
Townsend avalanche

dn
= αdx (25)
n

α is the first Townsend coefficient of the gas


Depends on electric field
For constant field (parallel plate geometry), solution is exponential increase in electron
density:
n(x) = n(0) exp(αx) (26)

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 51 / 130


Townsend coefficient vs. field

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 52 / 130


Regions of operation

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 53 / 130


Cylindrical geometry

Field with anode wire radius a, cathode inner radius b and applied voltage V is

V
E(r ) = (27)
r ln(b/a)

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 54 / 130


Field as a function of radius

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 55 / 130


Avalanche from a single electron, MC transport model

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 56 / 130


Simulation of avalanche

Winkler et al. Am. J. Phys., Vol. 83, 733-740 (2015)


Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 57 / 130
Development of the avalanche

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 58 / 130


Construction design

Figure : Cross sectional view, dimensions in cm

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 59 / 130


Beverage can design, Winkler et al. Am. J. Phys., Vol. 83, 733-740
(2015)

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 60 / 130


2π proportional counter

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 61 / 130


4π proportional counter

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 62 / 130


Fill gas

Gases without significant electron attachment


Oxygen or electronegative impurities must be removed in flow-counters
Sealed counters more convenient to use, but may have limited lifetime
Polyatomic gas (methane) to prevent photon-induced effects, component called quench gas
Most common proportional gas is 90% argon + 10% methane (P-10 gas)
Penning effect - component with ionization energy lower than excitation energy of principal
gas

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 63 / 130


Quenching in proportional detectors

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 64 / 130


Gas multiplication factor

Single electron response by avalanche created by one electron outside multiplication region
Total charge by n0 original pairs with gas multiplication factor M is

Q = n0 eM (28)

In cylindrical geometry, Townsend coefficient α depends on the radius (and on the field)
Z rc
ln M = dr α(r ) (29)
a

α is higher closer to the anode where the field is stronger


Parametrizations of α on the field use experimental data

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 65 / 130


Gas multiplication factor vs. voltage

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 66 / 130


Space charge effects

Reduction of the field near anode due to positive ions


Self induced effects resulting from high gas gain, does not depend on pulse rate
General space charge effect is cumulative for different avalanches, also for lower gain
M should just satisfy signal-to-noise requirement to prevent space charge effects
Simple check by varying the voltage and looking for suppression of large pulse amplitudes

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 67 / 130


Energy resolution

Fluctuations in n0 and in multiplication


Variance of pulse amplitude given by ion pair fluctuation and single-electron multiplication
variations
Furry distribution of number of electrons in a given avalanche, probability of ionization
depends only on E
Polya distribution for higher fields, ionization depends on electron’s history, not only on E
Distribution of pulse amplitude approaches Gaussian for large n0 (more than 20)

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 68 / 130


Energy resolution
Low energy photons have resolution close to statistical limit

Fractional standard deviation of the peak (FWHM/2.35), P-10 gas at 1 atm pressure
Open circles: standard tungsten anode
Full circles: improved uniformity

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 69 / 130


Systematic variations in gas multiplication factor

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 70 / 130


Time characteristics of the signal pulse
Output pulse is sum of drift time of free electrons to reach multiplication region and
multiplication time required for the avanlache
Drift time greater than multiplication time

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 71 / 130


Equivalent circuit of proportional counter seen by developing signal

Derivation of pulse shape utilizes energy conservation of charge moving across the capacitor
Most of the pulse amplitude from positive ion drift
Collection time of all ions long, hundreds ms, but large fraction of signal developed early of
ion drift, fraction of µs
Spread in electron drift times (ionizations along radius) makes spread in rise time of output
pulse

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 72 / 130


Time development of output pulse

Solid curve represents initial ionization at fixed radius (constant drift time)
Dashed line is for uniform ionization along the radius
Rise time minimized by high electric field in drift region and gases with high electron drift
velocity

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 73 / 130


Shaping of the output pulse

Pulse shape with different time constants


T is total drift time of positive ions from anode to cathode

Sauli, Principles of operation of multiwire proportional and drift chambers, CERN 77-09 (1977)
Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 74 / 130
Electron drift velocity

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 75 / 130


Variations of pulse rise time

Pulse shaping with short time constant removes slow component of rise time
Amplitude less than of full collection, effect of balistic deficit
Shape of pulse would vary with radial position of original ion pairs
Minimized by shaping time larger than variation in rise time
Can be used to separate signal α events from electron background

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 76 / 130


Spurious pulses

Satellite afterpulses from secondary processes following desired signal


Removed by amplitude discrimination since correspond to single-electron avalanche
Effect increases with high value of gas multiplication
Optical photons by excited atoms within the avalanche, low energy electron at cathode
surface through photoeffect

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 77 / 130


Operating voltage

Signal pulse for every particle with enough energy deposit


Sensitivity to more than one avalanche to suppress space charge effects and spurious pulses
Lower values of multiplication factor, some number of original ion pairs to create detectable
signal
Counting curve to select appropriate operating voltage (counting rate vs. voltage at same
source condition), looking for flat region

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 78 / 130


Alpha counting

Monoenergetic charged particles of range less than dimensions of the chamber

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 79 / 130


Beta counting

Range greater than chamber dimensions, smaller pulses compared to alpha


4π counter for absolute beta activity
Greater range advantage for mechanical source backing in 4π counter

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 80 / 130


Mixed alpha and beta sources

Two readout channels for alpha and beta


Separation by amplitude
Background from cosmic rays or ambient gamma (most sources) does not affect alpha
window

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 81 / 130


X-ray and gamma-ray sources

Fraction of incident photons absorbed in 5.08 cm of proportional gases at 1 atm pressure


Full absorption of low energy X-ray photons, full energy peaks in pulse spectrum
Small signal (∼ keV photons, W = 25 − 35 eV) enhanced by internal gas multiplication

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 82 / 130


Response function for low-energy X-rays

Characteristic X-rays after photoabsorption of primary radiation may escape without


interaction
Suppressed with K-shell energy above incident X-ray energy
Requirements to entrance window, beryllium or aluminum 50 - 250 µm

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 83 / 130


Fluorescence gating

Electronic separation of photoelectric absorption of incident photon and absorption of


consequent characteristic X-ray
Proportional counter as series of independent cells, X-ray absorbed in different cell
Two pulses in time coincidence corresponding to full absorption, escape peak avoided

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 84 / 130


Parallel plate avalanche counters

Heavy charged particles which would impose radiation damage to solid state detectors
Parallel plate electrodes, proportional gas at low pressure
Particle species separation by specific energy loss
For gap of 1 mm, fast component from electrons in ns with high field, time resolution 160 ps

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 85 / 130


Position-sensitive proportional counters

Avalanche occurs around small portion of anode wire length


Position of avalanche is indicator of axial position in cylindrical geometry

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 86 / 130


Charge division method

Anode with significant resistance per unit length


Two amplifiers at the ends, sum related to conventional pulse, proportion to the position of the
interaction
Alternatively relative rise time could be analyzed, longer rise time for pulse far from amplifier

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 87 / 130


Multiwire proportional counters (MWPC)

Cathode plates and anode wires positive vs. cathode


Electrons are collected at one of the wires, avalanche at a particular wire
Large negative pulse at anode where avalanche was collected, smaller positive pulses at
neighboring anodes

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 88 / 130


Electric field lines in MWPC

Field by equidistant grid of anode wires between two parallel cathode plates
High field region only in immediate vicinity of the anodes

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 89 / 130


Enlarged view of the field around the anode wires in MWPC

Wire spacing 2 mm, wire diameter 20 µm

Sauli, Principles of operation of multiwire proportional and drift chambers, CERN 77-09 (1977)
Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 90 / 130
Readout of MWPC

Each wire is connected to preamp and processed independently



Theoretical position resolution is spacing/ 12, but with pulse amplitude for each wire around
trajectory of the particle, it’s possible to reconstruct centre-of-gravity and improve the
resolution

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 91 / 130


Two-dimensional MWPC

Cathodes divided into perpendicular strips


Centroid of discharge located by center-of-gravity technique

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 92 / 130


Multicell proportional counter

Cylindrical independent cells


Coincidence measurement for process selection or background rejection
Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 93 / 130
Microstrip gas chamber (MSGC)

Narrow anode metallic strips (10 µm) on substrate


Quick collection of ions, clearing the space charge

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 94 / 130


Two-dimensional MSGC

Ionization in fill gas between anodes and drift plane, avalanches by electrons close to anode
Most of signal generated by drift of positive ions away from the anode

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 95 / 130


Gas electron multiplier (GEM)

Preamplification in gas to allow lower voltage readout of microstrip detector

GEM foil of insulating material, both surfaces covered by metallic cladding


Hole density 50 per mm2
Holes of 140 µm pitch and 70 µm diameter

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 96 / 130


GEM foil

Insulating material, both surfaces covered by metallic cladding

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 97 / 130


Simulation of GEM

Light lines - electrons, dark lines - ions, spots - ionization occurence

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 98 / 130


Cascade of GEM foils

Gain up to 104 - 105 , sensitivity to single electron

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 99 / 130


Micromegas (micro-mesh gaseous structure)

Two-stage parallel plate avalanche chamber


Ionization above the mesh which is transparent to electrons
Amplification below the mesh in a high field
Avalanche ions collected by the mesh, fast clearing of the space charge

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 100 / 130
Resistive Plate Chamber (RPC)

Thin gas volume between electrodes of high electrical resistivity

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 101 / 130
Construction of RPC

Spacers between plates each 10 cm against electrostatic attraction (constant gap size)

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 102 / 130
Readout of position sensitive RPC

Capacitive coupling of strips to preamps

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 103 / 130
Time Projection Chamber (TPC) of the ALICE experiment

3D position and dE/dx using volume of 90 m3 of Ne/CO2 /N2 (90/10/5) divided into two
sectors of 2.5 m
Voltage 100 kV on central electrode, drift field 400 V m−1
Electron drift velocity 2.7 cm µs−1 , maximum drift time 92 µs
Readout by MWPC with cathode pad readout mounted in trapezoidal sectors at each end
plate, active area of 32.5 m2

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 104 / 130
MWPC readout of ALICE TPC, cross sectional view along the wires

Grid of anode wires above pad plane, cathode wire plane and gating grid
Wire geometry and pad size dependent on radial track density, 560 000 readout pads
Anode to cathode distance 2 and 3 mm, gain up to 20 000
Gate opens for electrons from the drift volume by the collision trigger (6.5 µs after the
collision) for the duration of drift time, 92 µs, prevents the space charge of positive ions from
drifting back from multiplication region

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 105 / 130
MWPC readout, cross sectional view perpendicular to the wires

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 106 / 130
Particle identification by dE/dx in TPC

Simultaneous measurement of ionization losses and momentum


Parametrization of Bethe-Bloch formula for several particle species
Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 107 / 130
Upgrade of TPC

Expected rate 50 kHz of Pb-Pb, 5 interactions within maximal electron drift time, 10 events
superimposed from past+future time window
Untriggered readout without use of gating grid
Readout by GEM foils instead of MWPC, field 50 kV/cm in the hole
Suppression of back flowing ions
One side segmented to reduce total charge on the foil, segments coupled by resistors

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 108 / 130
Geiger-Mueller Counters

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 109 / 130
G-M counter

Introduced in 1928, one of the oldest detectors


Higher field than in proportional counter
One avalanche can create at least one another avalanche, chain reaction
All pulses have same amplitude, 109 - 1010 ion pairs
Pulse amplitude in volts, no need for external amplification
Counting rate limited by dead time

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 110 / 130
Geiger discharge

Excited molecules of neutral gas in Townsend avalanche, together with positive ions
They emit visible or UV light, may be absorbed by photoeffect at any place of the tube, new
free electron created
Or free electron created by absorption at cathode
New avalanche from the free electron after reaching the multiplication region around anode
Subcritical proportional tube: n00 p  1 with n00 number of excited molecules in typical
avalanche and p the probability of photoabsorption
Geiger discharge with single avalanche multiplication 106 - 100 has n00 bigger, criticality
achieved as n00 p ≥ 1

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 111 / 130
G-M region

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 112 / 130
Geiger discharge propagation

New avalanches created close to the original one, new electrons have to reach anode region
Spread of Geiger discharge along the anode wire with propagation velocity 2 − 4 cm µs−1
over entire length of anode, few µs after initiating event
Avalanches at random positions around anode, electrons collected by anode, secondary
positive ions around the multiplication region
Field reduced by space charge of the ions, multiplication reduced, Geiger discharge
terminated
Amount of needed space charge independent of initial ionization, the reason for same pulse
amplitude

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 113 / 130
Amplitude of Geiger discharge

Amplitude increases with higher applied voltage


More space charge needed to reduce the field below criticality
Overvoltage defined as voltage above minimum required for Geiger discharge initiation

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 114 / 130
Fill gases

Same requirements as for proportional counters since charge multiplication is needed (no
negative ions...)
Noble gases, helium, argon, quenching component
n00 increases with E/p (electric field to gas pressure), G-M as sealed tube with pressure less
than atmospheric
Voltage 500 - 2000 V for anode of 0.1 mm diameter

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 115 / 130
Quenching

Slow drift of positive ions after Geiger discharge termination to the cathode, neutralization
with electron from cathode surface
Energy needed = sum of gas ionization energy and energy to extract the electron from
cathode (work function)
If (gas ionization energy) = 2 × (work function), another free electron emerges from the
cathode surface
At least one free electron from cathode with large amount of positive ions
Will drift towards anode to make new Geiger discharge
Cycle would repeat, continues output of multiple pulses
No problem for proportional counters, number of ions smaller, only spurious pulses

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 116 / 130
External quenching

Needed any method to prevent excessive multiple pulsing


Reduction of high applied voltage for some fixed time after the pulse below gas multiplication
Time greater than transit of positive ions to the cathode (hundreds of ms) and transit time of
free electron (few µs)
Value of R selected high enough (108 Ω), so RC is of ms
Several ms for anode to return to nominal operating voltage, low counting rate

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 117 / 130
Internal quenching

Component of quench gas added to primary fill gas, concentration 5 - 10 %


Smaller ionization potential, more complex molecular structure
Multiple pulsing prevented by charge transfer collisions, positive charge of primary gas ion
transfered to quench gas molecule, original ion neutralized before reaching cathode
Ions of quench gas at cathode with some concentration of quench component
At cathode, the excess energy dissociates the more complex molecule, emission of free
electron suppressed, no additional avalanches
Organic molecules, ethyl alcohol, ethyl formate
Limit of 109 counts due to consumption of quench gas
Problem avoided by halogens as quench gas (chorine, bromine), spontaneous recombination
after dissociation
Limit to lifetime then by contamination of reaction products from the discharge and polymer
deposition on anode surface

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 118 / 130
Shape of the output pulse

Signal of multiple avalanches from secondary electrons


Spread in time for them to reach multiplication region
Rise time of few µs
Electric field distorted from the space charge of positive ions
Fast rise and slow rise from ions drift, time constant less than 100 µs to eliminate slow
component
Loss in amplitude tolerated due to large amount of charge
Delay in response by drift of original electrons to multiplying region, applies for timing with
G-M as time uncertainty
Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 119 / 130
Dead time

Drift of positive ions outwards, recovery of field in multiplying region


After drift to some distance, the field permits new Geiger discharge, lower intensity in the
beginning
First pulses reduced in amplitude
Full field and full amplitude after all ions neutralized at the cathode
Dead time = time between initial pulse and second Geiger discharge, regardless of the
magnitude, 50 - 100 µs
Resolving time = time for pulse of amplitude high enough to be registered, dead time may
include also recovery time in practical application
Recovery time = time to return to the original state, pulse of full amplitude

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 120 / 130
Dead time behavior

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 121 / 130
Geiger counting plateau

Dependence of measured counting rate on voltage with constant rate from radiation source
Counter threshold Hd , no pulses if amplitude is below

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 122 / 130
Onset of continues discharge

Corona discharge by irregularities on anode wire or quenching failure


Decrease of voltage needed
Operating voltage just sufficient for the flat region, reduces quench gas consumption
Low amplitude tail causes some slope of the counting curve
Hysteresis in counting curve due to charges on insulators
Linearity of the plateau measures quality of G-M
Organic quenched tubes have slope of 2 - 3 % with change in voltage by 100 V
Halogen quenched tubes have greater slope, but larger lifetime and lower operating voltage

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 123 / 130
End-window Geiger tube

Fine anode wire, less requirements on uniformity of the wire and of the electric field
Cylindrical cathode of metal or glass with metallized inner surface
Thin window for short-range particles, differential pressure
More robust window for beta or gamma

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 124 / 130
Other design of Geiger tube

Wire loop anode in cathode of arbitrary volume


Needle counters - sharply pointed needle, field as 1/r 2 (1/r for cylindrical geometry), used
for smaller active volume
Source in counting volume for low-energy particles
14 C detection by gas mixture of CO2 with counter gas for high counting efficiency

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 125 / 130
Counting electronics

R is load resistance for signal development


Cs is capacitance of the tube and wiring, parallel with R
RCs ∼ few µs for fast-rising component
Coupling capacitor Cc to block high voltage on the tube, only signal pulse transmitted, RCc
larger than pulse duration

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 126 / 130
Counting efficiency for charged particles

100 % for any charged particle entering the active volume since a single ion pair triggers full
Geiger discharge
Effective efficiency given by the window of the tube, absorption or backscattering
For alpha counting window with 1.5 mg/cm2

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 127 / 130
Efficiency for neutrons

Rare application for neutron detection


Low capture cross section for Geiger gases
Gamma-ray discrimination not possible, all pulses have same size
Neutron-induced reactions detected by proportional counters, large amplitude from such
reactions

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 128 / 130
Counting efficiency for gamma rays

Interaction in solid wall of the counter, secondary electron needs just to emerge into active
volume
1 - 2 mm of innermost layer of the wall
Efficiency increased with high-Z material of the cathode (bismuth Z = 83)
Low-energy gamma may interact with fill gas, xenon and krypton (high-Z) enhance the
efficiency up to 100 % for photon energies 10 keV
Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 129 / 130
Counting efficiency for gamma rays

Jaroslav Adam (CTU, Prague) DPD_03, Ionization chambers Version 2 130 / 130