Академический Документы
Профессиональный Документы
Культура Документы
net/publication/320640480
CITATIONS READS
2 1,407
6 authors, including:
Andrei Blasko
Novartis
92 PUBLICATIONS 1,268 CITATIONS
SEE PROFILE
Some of the authors of this publication are also working on these related projects:
All content following this page was uploaded by Andrei Blasko on 06 November 2017.
a r t i c l e i n f o a b s t r a c t
Article history: A Total Organic Carbon (TOC) based analytical method to quantitate trace residues of clean-in-place
Received 9 August 2017 ® ®
(CIP) detergents CIP100 and CIP200 on the surfaces of pharmaceutical manufacturing equipment was
Received in revised form
developed and validated. Five factors affecting the development and validation of the method were
18 September 2017
Accepted 22 October 2017
identified: diluent composition, diluent volume, extraction method, location for TOC sample preparation,
Available online 26 October 2017 and oxidant flow rate. Key experimental parameters were optimized to minimize contamination and
to improve the sensitivity, recovery, and reliability of the method. The optimized concentration of the
Keywords: phosphoric acid in the swabbing solution was 0.05 M, and the optimal volume of the sample solution
Manufacturing equipment was 30 mL. The swab extraction method was 1 min sonication. The use of a clean room, as compared to
Cleaning optimization an isolated lab environment, was not required for method validation. The method was demonstrated to
Cleaning verification be linear with a correlation coefficient (R) of 0.9999. The average recoveries from stainless steel surfaces
Detergent residues at multiple spike levels were >90%. The repeatability and intermediate precision results were ≤5% across
Coupons the 2.2–6.6 ppm range (50–150% of the target maximum carry over, MACO, limit). The method was also
TOC shown to be sensitive with a detection limit (DL) of 38 ppb and a quantitation limit (QL) of 114 ppb.
The method validation demonstrated that the developed method is suitable for its intended use. The
methodology developed in this study is generally applicable to the cleaning verification of any organic
detergents used for the cleaning of pharmaceutical manufacturing equipment made of electropolished
stainless steel material.
© 2017 Elsevier B.V. All rights reserved.
https://doi.org/10.1016/j.jpba.2017.10.021
0731-7085/© 2017 Elsevier B.V. All rights reserved.
34 X. Li et al. / Journal of Pharmaceutical and Biomedical Analysis 149 (2018) 33–39
aerosol detection [22] must be used. All these component specific ing System Suitability 1,4-Benzoquinone Standard Solution (RSS),
detection techniques are not applicable when chemical changes Working Standard Sucrose (RS), and Reagent Water (RW)). Clean-
® ®
occur during the cleaning process, as is always the case of biologics’ ing in place solutions (CIP100 and CIP200 ) were obtained from
degradation during cleaning of manufacturing equipment. Steris Corporation (Mentor, OH, USA). The sampling swabs with low
Assessing detergent residues by TOC has been reported [18–21]. carbon, certified <50 ppb, part number TX714K, were from Texwipe
Compared to the conventional HPLC based methods, TOC meth- (Kernersville, NC, USA) and were used without pretreatment. The
ods are faster, simpler, and shorter; thus, the analytical method sample diluent was 0.05 M phosphoric acid in water prepared by
development is simpler. The TOC technique is a nonspecific, total dilution of concentrated (85%, w/w) phosphoric acid (EMD Milli-
sample combustion analysis that measures the total organic carbon pore, Darmstadt, Germany) in water. Each TOC vial (VWR (Radnor,
content from all sources, including that from residual API, excipi- PA, USA), 89094-204, TOC <10 ppb) was filled with 30 mL of the
ents, detergents, and contaminants. The analyzer uses UV radiation diluent. Pipettes were from Rainin (Oakland, CA, USA). Ultrapure
and a chemical oxidizing agent to oxidize the organic compounds water was obtained using the in-house Millipore Milli-Q Advan-
in the sample to form carbon dioxide (CO2 ). The generated CO2 tage Water Purification System (Darmstadt, Germany). HPLC grade
is measured using a sensitive and selective membrane-based con- methanol and phosphoric acid were from EMD (Billerica, MA, USA).
ductivity detection technique [23], and the TOC result is reported Stainless steel coupons were 316L SS with a roughness average of
as concentration of organic carbon in the sample (ppm or ppb). 20, from GlobePharma, Inc. (New Brunswick, NJ, USA).
Despite the above advantages, a general, systematic, and prac-
tical approach to method development has not been reported.
Typically, there are several commercial products used for the 2.2. Instrumentation
®
cleaning of manufacturing equipment [16,17]. CIP100 (base) and
®
CIP200 (acid) are the most commonly used for CIP systems or auto- The TOC analysis was conducted using a Sievers TOC ana-
mated cleaning systems. In a typical cleaning process, the surface lyzer, Model 900 with Autosampler (GE Analytical Instruments;
of the manufacturing equipment is cleaned by the application of Boulder, CO, USA). Key instrument parameters were “Acid” flow
hot water for injection (WFI), followed by the sequence of diluted rate (0.8 L/min) and “Oxid” flow rate (1.4 L/min for blanks;
® ®
CIP100 , hot WFI, diluted CIP200 , and hot WFI. Stainless steel 6.8 L/min for the standard and sample solutions). Other parame-
coupons, with the same surface/finish as the manufacturing equip- ters were default values set by the instrument vendor [22]. System
ment, are used to simulate equipment surfaces during laboratory suitability was tested at the beginning of each run according to the
method validation studies of the cleaning verification methods [15]. vendor’s instructions using the system suitability standard set. The
The material of construction (MOC) of the coupons was not the background TOC values for the diluent blanks and reagent water
subject of this study. Considering the stainless steel material, MOC should not be more than 250 ppb.
might indirectly affect the recovery by not being able to be cleaned
properly. If MOC were a factor, after properly cleaning the coupons,
2.3. Method development
different types of molecules (small, large) with different electronic
surface density would interact differently with the metal surface,
2.3.1. Concentration of phosphoric acid in the diluent
which we demonstrated was not the case [15].
To determine the appropriate concentration of the phosphoric
Previous reports addressed the problem associated with the
acid in the diluent, three concentrations of phosphoric acid were
difficulty of recovering the sample from the equipment which
prepared (0.05 M, 0.25 M, and 0.5 M) and their TOC values were
depended on the tools and the skills of the analyst [24–27]. Stud-
measured. Additionally, two clean swabs were added to 15 mL of
ies have reported challenges with using TOC such as low speed,
the diluents at each concentration. The swab blank extractions
lack of specificity, limitation of use to water soluble compounds,
were performed, and the TOC values of the resulting solutions were
variability from one instrument to another, and inaccuracies that
determined.
may lead to false positives [24–26]. We believe that a majority of
the reported unreliability with TOC measurements is due to the
sample preparation and handling along with the precise control 2.3.2. Volume of the diluent
of the instrumental parameter. These two issues, which may have Cleaning sample solutions at the proposed maximum allow-
driven analysts to use techniques such as IR absorption and mass able carry over (MACO) limit were prepared with 15 mL or 30 mL
spectrometry instead of TOC [27], are addressed in this paper. of the diluent, respectively. To determine the impact of vol-
Parameters such as temperature on oxidation efficiency and car- ume/concentration of the sample solution on carryover, water
rier gas flow rate are known to affect the accuracy, precision, limit blanks were injected both before and after the injection of sample
of detection, and calibration of the TOC analysis [28]. This paper solutions.
describes the influence of different parameters affecting the devel-
opment and validation of a TOC based analytical method to quantify
® ®
trace residues of CIP100 and CIP200 on the surface of manu- 2.3.3. Extraction method
facturing equipment. The methodology developed in this study is Each TOC vial containing two swab heads that were spiked at the
generally applicable to the cleaning verification/quantitation of any MACO limit was either sonicated or vortexed for 1 min, and then
organic detergents used for the cleaning of pharmaceutical manu- shaken by hand to ensure thorough mixing. The extracted swab
facturing equipment using TOC analysis based on the UV/persulfate heads were removed from the vials using a clean disposable pipette
oxidation principle. tip, and the vial was re-capped.
Table 1
Calibration curve data.
Linearity Solution Target concentration of CIP200 (g/mL) Theoretical TOC value (ppm) Actual TOC reading (ppm) Correcteda TOC reading (ppm)
Table 2
®
Preparation of the CIP200 detergent simulated residue.
0% (blank) 0 0
50% 101.1 202.2
100% 202.2 404.4
150% 303.3 606.6
a
The range covers from 50% to 150% of the target limit (333 g dried detergent
®
content, equivalent to 6066 g of the commercial CIP200 solution applied to a
50 cm2 surface area, corresponding to 4.4 ppm as TOC value).
2.3.5. Flow rate of the oxidant Fig. 1. TOC value of diluent with different concentrations of phosphoric acid and
Cleaning sample solutions at the MACO limit were analyzed at the sample solution (diluent + swab) after extraction from two clean swabs.
various oxidant flow rates (1.4, 3.2, 6.8, 9.2, and 12.2 L/min), and
the recovery was determined for each experiment.
3. Results and discussion
Before spiking, the coupons need to be cleaned, following the 3.1.1. Concentration of phosphoric acid in the diluent
procedures as previously reported [15,29]. The spiked coupons Once the swabbing procedure is executed, the swab residues
were prepared by applying the appropriate volumes of the stock need to be extracted in a proper diluent. Water has been used as
®
CIP200 solutions (Table 2) onto the coupons. Each concentra- the extraction solution in previous reports [18,20]. However, in this
tion level was spiked in triplicate onto separate, freshly cleaned study, diluted phosphoric acid solution was chosen as the diluent
coupons. The coupons were allowed to evaporate to dryness at for sample preparation. Phosphoric acid (6 M) is also used as the
ambient conditions and swabbed as previously described [15]. The “Acid” reagent in the TOC analyzer to remove the inorganic car-
swabbing procedure was performed as previously reported [15]. bon from the sample solution [23]. An acidic diluent solution helps
After swabbing, the swab heads are separated from the swab han- to effectively extract the analyte from the swab; however, if the
dle at the pre-scored notch by forcefully bending the swab against acid concentration is too high, there may be undesirable extraction
lip of the vial, avoiding extraneous contact and contamination. As of organic carbon from the sampling swabs. To select the opti-
the swab breaks, the swab heads fall into the TOC vial containing mal acid concentration, we evaluated the blank swab extractions
the sample diluent, and the vial is securely capped. Each TOC vial at three concentrations of phosphoric acid, 0.05, 0.25, and 0.5 M.
containing two swabs was sonicated for 1 min, and then shaken by The results are summarized in Fig. 1. The TOC values were trending
hand to ensure thorough mixing. The extracted swab heads were higher as the phosphoric acid concentration increased. To minimize
pulled from the vial using a clean disposable pipette tip, again tak- the measured TOC background, 0.05 M of phosphoric acid in water
ing care to avoid extraneous contact and contamination, and the was chosen as the sample diluent for coupon recovery studies and
vial is re-capped. swab extraction. The value for 0.05 M corresponds to a TOC of the
36 X. Li et al. / Journal of Pharmaceutical and Biomedical Analysis 149 (2018) 33–39
Table 3
Influence of the volume of diluent.
15 mL diluent 30 mL diluent
Before sample injection After sample injection Before sample injection After sample injection
1 25 187 25 54
2 25 165 25 64
3 24 200 25 48
Fig. 3. TOC reading of a clean stainless steel surface swab sampled under various
environmental conditions. The error bars represent the standard deviation of three
Fig. 2. The notch on the swab handle where it is broken to allow the swab head to sample preparations (three coupons) in each case.
fall into the TOC vial.
Table 4 precision of the sonication method was much better than that of
Influence of the extraction method. the vortexing method, while the percent recoveries were within
Extraction method Average Recovery (%) Srel (%) the experimental variability. Therefore, sonication for 1 min was
chosen as the sample extraction method.
Vortexing 79 8
Sonication 76 3
3.1.4. Location of sample preparation
The TOC analysis is extremely sensitive and cannot differentiate
swab extraction of 0.4 ppm, which is well below the cleaning limit the different sources of carbon (e.g., sample, contamination from
(4.4 ppm). environment or handling, including respiratory exhalations). To
assess the influence of the lab environment we conducted spiking
3.1.2. Volume of the diluent and coupon swabbing experiments under different lab conditions.
In trace analysis, it is desirable to make the sample concentration We swabbed clean stainless steel coupons (50 cm2 ) in normal lab
as high as possible to improve the limit of quantitation/detection. environment with normal foot traffic and conversations of other
The TOC vial capacity is 40 mL. Initially, 15 mL diluent was selected lab personnel nearby, in an isolated lab where entrance/exit was
for the swab extraction. This volume provides ample volume (four restricted and the analysts’ talking (i.e., respiration) was minimal-
injections) of the cleaning sample solution in each analysis. In addi- ized during the swabbing, and, finally, in a clean room (ISO 8) also
tion, the notch in the swab handle stays above the surface of the with minimal talking. As shown in Fig. 3, the TOC reading of the
solution when the swab head is submerged into the diluent. This coupon blank sample prepared under the normal lab condition was
avoids contamination of the sample solution from the swab handle the highest, suggesting appreciable contamination from the envi-
where the analyst hand makes contact during sample preparation. ronment. When the swabbing was conducted in an isolated lab,
By using the 15 mL sample volume, however, we observed that the background TOC was significantly reduced. The contamination
the TOC value of the blank injections made after sample injection from the environment was the least when the swabbing was con-
were relatively high, suggesting carryover was taking place. The ducted in a clean room. However, because of the limited availability
carryover significantly decreased when 30 mL of diluent was used of the clean room (for method validation), we chose to conduct
(Table 3). Therefore, 30 mL was chosen as the diluent volume. With the experiments for method validation in an isolated lab environ-
this volume of diluent, the notch in the handle touches the solution ment. Compared to the proposed CL of 4.4 ppm, the TOC from the
when the swab is submerged in the TOC vial. To avoid contamina- environment, approximately 0.2 ppm, was deemed insignificant.
tion from the handle, the swab head should be separated from the
handle (Fig. 2). 3.1.5. Flow rate of the oxidant
In a TOC experiment, the elemental carbon content of the ana-
3.1.3. Extraction method lyte is oxidized to carbon dioxide using ammonium persulfate as
Vortexing and sonication are two commonly used extraction oxidizing agent [23]. The flow rate of the oxidant dictates the
methods. They were compared for extraction accuracy (% recovery) amount of the oxidant in the reaction and should be optimized. Too
and precision (repeatability) (Table 4). The recovery was calcu- low of a flow rate will lead to an incomplete oxidation of the ana-
lated by comparing the measured TOC values with the theoretical lyte; too high a flow rate could result in a lower measured TOC value
TOC values. The recoveries for the vortex and sonication extrac- [30]. To optimize the oxidant flow rate, the recovery of the clean-
tions were 79% and 76%, and the relative standard deviations (Srel ) ing sample (prepared at the MACO limit) was determined at various
were 8% and 3%, respectively. Comparing the two methods, the oxidant flow rates. As shown in Fig. 4, at 1.4 L/min the oxidation
X. Li et al. / Journal of Pharmaceutical and Biomedical Analysis 149 (2018) 33–39 37
Table 5
Accuracy at different spiking levels.
Level Expected TOC value (ppm) Actual TOC reading (ppm) Corrected TOC valuea (ppm) Recovery (%) Average recovery (%) Srel (%)
® ® ®
CIP100 (3.35%, w/w) or of a mixture of CIP100 + CIP200 is higher
®
than that of CIP200 (2.173%, w/w). The lower carbon content in
®
CIP200 makes for the more stringent calculated CL in TOC value
corresponding to the MACO, as detailed below.
In this study, the method was validated for the CL correspond-
®
ing to 333 g detergent/50 cm2 (MACO). The commercial CIP200
solution is a complex mixture. According to the formulation of
®
CIP200 provided by the vendor, there are three key components in
®
CIP200 that contribute to the total carbon content of a given sam-
®
ple. Thus, MACO expressed in g of CIP200 commercial solution
2
per 50 cm is calculated using Eq. (1).
333
MACOCIP200 = = 6066g (1)
A1 × B1 + A2 × B2 + A3 × B3
where,
Fig. 4. % Recovery versus flow rate of the oxidizing reagent (ammonium persulfate)
of a sample prepared at the proposed MACO limit. The % Recovery is calculated by Ai % w/w of component i
comparing the measured TOC with the theoretical (expected) TOC at the CL. Bi % purity of component i
The values of A1 , A2 , A3 , B1 , B2 , and B3 were provided by the
vendor.
is incomplete and less than 60% of the carbon content in the sample Since the MACO of 6066 g per 50 cm2 of commercial CIP200
®
was oxidized. There were no appreciable differences in the recov- solution is swabbed and sampled into a 30 mL volume, the TOC CL
eries at flow rates ≥ 3.2 L/min, indicating that 3.2 L/min oxidant is 4.4 ppm, calculated using Eq. (2).
flow rate is needed to ensure complete oxidation of the analyte. To
ensure consistent, quantitative analyte conversion to carbon diox- CCCIP200 × MACOCIP200 g
CL = = 4.4 = 4.4ppm (2)
ide, a higher value (6.8 L/min) was chosen as the oxidant flow VT mL
rate. where,
®
CCCIP200 % carbon content in the CIP200 solution: 2.173% [16]
®
3.1.6. Calculation of the cleaning limit in TOC MACOCIP200 Maximum Allowable Carry Over of CIP200 solu-
2
tion per 50 cm : 6066 g
The MACO dictates the threshold for the level of residual deter-
gent above which the cleaning process would be considered as a VT Total sample volume: 30 mL
failure [15]. According to EPA (40CFR156.62) toxicity standards,
® ®
CIP100 and CIP200 are considered slightly to moderately toxic 3.2. Method validation
based on LD50 (rat) data. This toxicity data justifies the amount
of residue detergents allowed to carryover of 100 ppm in the 3.2.1. Linearity
subsequent batch. The MACO is calculated based on this value, The linearity between the theoretical TOC value and the back-
®
subsequent batch size, and equipment surface area. In our case, ground corrected TOC value of CIP200 was examined over a range
®
the 100 ppm MACO limit corresponds to 333 g detergent/50 cm2 from 4.0 to 404.4 g/mL of CIP200 (88 ppb − 8.8 ppm of TOC val-
surface area. During the cleaning process, the surface of the manu- ues), corresponding to approximately 2% to 200% of the nominal CL
facturing equipment is cleaned by the application of hot water for concentration of 202.2 g/mL, or 4.4 ppm TOC value. One prepara-
®
injection, followed by the rinse sequence of diluted CIP100 , hot tion of each linearity level solution was analyzed to determine the
®
water for injection, diluted CIP200 , and hot water for injection. TOC value. The TOC reading of the linearity samples were corrected
The volume of the hot water for injection used for the final rinse for the background reading of the blank. The correlation coefficient
®
is typically six times the volume of the diluted CIP100 solution. was 0.9999, and the y-intercept relative to response at 100% nomi-
®
With the multiple rinses with hot water and CIP200 solution, it nal CL concentration was 0.5%. The linearity data is summarized in
®
is reasonable to assume that CIP100 has been completely rinsed Table 1.
from the equipment surface at the end of the cleaning process and
®
that the detergent residue remaining only comes from CIP200 . 3.2.2. Accuracy
® ®
The TOC analysis of CIP200 presents the worst case scenario for Accuracy was demonstrated by the % recovery of CIP200 deter-
calculation of the TOC limit (ppm) because the carbon content of gent spiked in triplicate at 3 different levels onto the 50 cm2
38 X. Li et al. / Journal of Pharmaceutical and Biomedical Analysis 149 (2018) 33–39
Table 6 The method was evaluated for linearity, accuracy, precision, and
Pooled TOC data to assess intermediate precision.
sensitivity. The results demonstrated that the method is reliable
Level Pooled Average Pooled Standard Pooled Relative Standard and suitable for the cleaning assessment of residual detergents of
Recovery (%) Deviation (%) Deviation (Srel,pooled ) (%) the manufacturing equipment. The method was found to be lin-
50% 91 3 3 ear (R = 0.9999), accurate (recovery ≥ 90%), precise (Srel ≤ 4%), and
100% 92 3 3 highly sensitive (DL: 38 ppb, QL: 114 ppb). The methodology devel-
150% 96 5 5 oped in this study is generally applicable to the cleaning verification
of any organic detergents used for the cleaning of pharmaceutical
Table 7 manufacturing equipment made of electropolished stainless steel
Validation characteristics of the TOC analytical methods for cleaning assessment of material.
detergents.
[26] B. Wallace, R. Stevens, M. Purcell, Implementing total organic carbon analysis [28] J. Qian, K. Mopper, Automated high-Performance, high-Temperature
for cleaning validation, Pharm. Tech. Aseptic Processing (2004) combustion total organic carbon analyzer, Anal. Chem. 68 (1996) 3090–3097.
40–43. [29] I.A. Haidar Ahmad, A. Blasko, Failure of cleaning verification in
[27] H. Nakagawa, T. Tajima, M. Kano, S. Kim, S. Hasebe, T. Suzuki, H. Nakagami, pharmaceutical industry due to uncleanliness of stainless steel surface, J. Vis.
Evaluation of infrared-Reflection absorption spectroscopy measurement and Exp. 126 (2017) e56175.
locally weighted partial least-Squares for rapid analysis of residual drug [30] Sievers 900 Series TOC Analyzers–Autoreagent Feature GE Technical Bulletin,
substances in cleaning processes, Anal. Chem. 84 (2012) 3820–3826. 2005 https://www.gewater.com.