Service Manual ddRCompact™ Series

Chapter 6: Covers

Covers

Caution: US Federal Law restricts this device to use by or on the


order of a physician.

Mise en garde: Les lois fédérales des États-Unis limitent l'utilisation


de cet appareil à un médecin ou à un tiers agissant
sur son ordre.

Revision Date By Reason for Change

00a 2009-01-07 SRCH / WOL First Draft

This document is the English original version, edited and supplied by Swissray Medical AG.
All rights reserved by Swissray Medical AG.

2009-01-07 / WOL SHT_37_102_006_00a_E 1/6


Service Manual ddRCompact™ Series
Chapter 6: Covers

TABLE OF CONTENTS

Section Page

6 Covers................................................................................................................ 3
6.1 Before Assembling the Covers ....................................................................... 3
6.2 Cover Installation of the HD-3000................................................................... 3
6.3 Cover Installation of the Stand........................................................................ 4
6.4 Cover Installation of the Tube......................................................................... 6

2009-01-07 / WOL SHT_37_102_006_00a_E 2/6


Service Manual ddRCompact™ Series
Chapter 6: Covers

6 Covers

6.1 Before Assembling the Covers

Before assembling all covers the stand shall be fully tested in its functions, fully
configurated and fully calibrated.
Flatfield calibration should be done after assembling the detector top cover.

Following parts are pre-assembled with covers:


• C-arm
• Tube holder arm

6.2 Cover Installation of the HD-3000

Place the top cover on the detector and connect


the plugs.

Mount the top cover.

Mount the 2 side covers.

2009-01-07 / WOL SHT_37_102_006_00a_E 3/6


Service Manual ddRCompact™ Series
Chapter 6: Covers

6.3 Cover Installation of the Stand

Mount the right side corner bar cover.

Slide in the 2 socket trim covers.

Slide in the 2 light panels and connect them


carefully on the base of the stand
Right panel to X2.4
Left panel to X2.2

Assemble the top cover to the stand using the 5


fixation screws.

2009-01-07 / WOL SHT_37_102_006_00a_E 4/6


Service Manual ddRCompact™ Series
Chapter 6: Covers

Connect the infrared sensor on the top cover to


X30.

Assemble the 2 plastic side cover, the large


one to the monitor side, the small one to the
rear side.
The covers are attached using Velcro.

Assemble the 2 plastic cover, lower and upper


part over the electro cabinet. The covers are
slight in mounted.

2009-01-07 / WOL SHT_37_102_006_00a_E 5/6


Service Manual ddRCompact™ Series
Chapter 6: Covers

6.4 Cover Installation of the Tube

Mount the tube cover with 3 screws.

2009-01-07 / WOL SHT_37_102_006_00a_E 6/6