Вы находитесь на странице: 1из 10

Clinical Chemistry 52:7

1346 –1355 (2006) Drug Monitoring and


Toxicology

Liquid Chromatography–Tandem Mass


Spectrometry for Detection of Low
Concentrations of 21 Benzodiazepines,
Metabolites, and Analogs in Urine:
Method with Forensic Applications
Oscar Quintela,1,2 François-Ludovic Sauvage,1 Fabienne Charvier,1
Jean-Michel Gaulier,1 Gérard Lachâtre,1 and Pierre Marquet1*

Background: Commonly used methods for detecting suppression was seen except for alprazolam, for which
benzodiazepines (BZPs) and BZP-like substances, such baseline decreased by almost 20%. In the forensic urine
as zolpidem and zopiclone, may not detect low concen- sample, the method detected alprazolam (3.5 ␮g/L) and
trations of these drugs. We developed a liquid chro- its characteristic metabolite, ␣-hydroxyalprazolam
matographic–tandem mass spectrometric method for (0.17 ␮g/L).
identifying these drugs and their relevant metabolites. Conclusion: This method measured low concentrations
Methods: We extracted BZPs from urine by solid-phase of BZPs and BZP-like substances and might be useful
extraction with a mixed-mode phase (OASIS® HLB for analyses of urine in suspected drug-facilitated sex-
cartridges). Chromatographic separation was performed ual assault cases.
with a Waters XTerra MS C18 [150 ⴛ 2.1 mm (i.d.); bead © 2006 American Association for Clinical Chemistry
size, 5 ␮m] reversed-phase column with deuterated
analogs of the analytes as internal standards (IS). Detec- The benzodiazepines (BZPs)3 and BZP-like substances
tion was performed with a triple-quadruple mass spec- (e.g., zopiclone and zolpidem) are relatively safe drugs
trometer that monitored 2 specific transitions per com- that are frequently prescribed for the treatment of stress,
pound in the electrospray, positive-ion selected-reaction panic disorders, sleep disorders, muscle spasms, alcohol
monitoring mode. We tested this technique on urine withdrawal, and seizures (1 ). Because of the high con-
samples from 12 healthy volunteers and 1 forensic sumption of these substances world-wide, they are in-
sample obtained in a case of alleged drug-facilitated volved in many forensic cases. The influence of BZPs on
sexual assault. driving and in jobs in which the presence of high serum
Results: Chromatographic separation was achieved concentrations in employees could be dangerous has been
within 18 min. The linear dynamic ranges extended documented (2– 6 ). These drugs also play a role in drug-
from 0.02 or 0.1 ␮g/L (depending on the drug or metab- facilitated sexual assault (DFSA) (7, 8 ).
olite) to 50 ␮g/L. Extraction recovery (range) was 77%– Zopiclone and zolpidem belong to a new generation of
110%. Limits of detection were <0.05 ␮g/L. No ion hypnotics and sedatives that are structurally quite differ-
ent from the BZPs (9 ). They have a rapid onset of action
and a short half-life and, unlike BZPs, have weak myore-
1
Department of Pharmacology-Toxicology, Limoges University Hospital,
Limoges, France.
2
Forensic Toxicology Department, Institute of Legal Medicine, University
3
of Santiago de Compostela, C/San Francisco, Spain. Nonstandard abbreviations: BZP, benzodiazepine; DFSA, drug-facili-
* Address correspondence to this author at: Department of Pharmacology- tated sexual assault; LC-MS, liquid chromatography–mass spectrometry; MS/
Toxicology, Limoges University Hospital, 87042 Limoges, France. Fax 33-555- MS, tandem MS; ESI electrospray ionozation; IS, internal standard; SPE,
05-61-62; e-mail pierre.marquet@unilim.fr. solid-phase extraction; SRM, single reaction monitoring; LLOQ, lower limit of
Received December 21, 2005; accepted May 2, 2006. quantification; ULOQ, upper limit of quantification; MRE, mean relative error;
Previously published online at DOI: 10.1373/clinchem.2005.065631 RT, retention time; and GC, gas chromatography.

1346
Clinical Chemistry 52, No. 7, 2006 1347

laxant and anticonvulsant effects. They are prescribed as 3 months under these conditions. Working solutions of
hypnotics instead of BZPs (10 ). mixtures of the analytes at concentrations of 0.0004, 0.001,
Many widely used screening and confirmation meth- 0.002, 0.01, 0.04, 0.2, and 1 mg/L were freshly prepared
ods do not detect lower concentrations of these drugs, and each day by appropriate dilution in methanol–water
in some cases, the laboratory may not have the capability (50:50 by volume). In the same way, a 0.1 mg/L working
to detect a particular drug (11 ). New ionization tech- solution was prepared from all the IS stock solutions.
niques have made liquid chromatography–mass spec-
trometry (LC-MS) and LC-tandem MS (LC-MS/MS) ex- sample preparation
tremely effective for the analysis of BZPs and BZP-like We added 1 mL of Sorensen buffer, 50 ␮L of working IS
drugs. We developed an LC electrospray ionization (ESI)- solution, and 100 ␮L of the appropriate working solutions
MS/MS method for screening and quantification of zopi- of the parent drugs and metabolites to 2 mL of urine to
clone, zolpidem, and several BZPs and their relevant obtain the following calibration concentrations: 0, 0.02,
metabolites, with a focus on the analysis of urine samples 0.05, 0.1, 0.5, 2, 10, and 50 ␮g/L. The tubes were then
from DFSA cases. vortex-mixed for 10 s. Solid-phase extraction (SPE) was
performed with Oasis HLB cartridges (3 mL/60 mg)
Materials and Methods previously conditioned with 2 mL of methanol and equil-
chemicals, reagents, and solutions ibrated with 2 mL of water. The mixtures of sample and
The calibrators used were 1-hydroxymidazolam, 4-hy- buffer were loaded on the cartridge and slowly passed
droxymidazolam, midazolam, 7-aminoclonazepam, des- through the bed without vacuum. The SPE column was
methylflunitrazepam, flunitrazepam, and 7-aminofluni- rinsed sequentially with 3 mL of deionized water and
trazepam (all from Lipomed); alprazolam, bromazepam, 3 mL of 0.5% ammonia in methanol–water (40:60 by
clonazepam, triazolam, lorazepam, 3-hydroxyflunitraz- volume). Reduced pressure was then applied to the
epam, zolpidem, and zopiclone (all from Sigma); fluraz- column (maximum of ⫺60 kPa) for 15 min to dry the
epam, desalkylflurazepam, ␣-hydroxyalprazolam, ␣-hy- column. The retained drugs were eluted under gravity
droxytriazolam, and the deuterated internal standards with 3 mL of dichloromethane–propanol-2 (75:25 by vol-
(IS) 7-aminoflunitrazepam-d7, 7-aminoclonazepam-d4, ume) into 5-mL round-bottomed glass tubes. The eluate
alprazolam-d5, triazolam-d4, ␣-hydroxyalprazolam-d5, was evaporated under nitrogen at 40 °C in a Zymark
desmethylflunitrazepam-d4, ␣-hydroxytriazolam-d4, de- TurboVap® LV Concentration Workstation. The dried
salkylflurazepam-d4, and flunitrazepam-d7 (all from extracts were reconstituted in 50 ␮L of mobile phase, and
Ceriliant); lormetazepam (kindly provided by Schering, 15 ␮L was injected into the chromatographic system.
Levallois-Perret, France); and loprazolam (kindly pro-
vided by Sanofi-Aventis, Paris, France). hplc conditions
Propanol-2 (Normapur), dichloromethane (Norma- The HPLC system consisted of a Shimadzu LC 10 ADvp
pur), and ammonia (Normapur) were purchased from pump with an SIL-Ht autoinjector and a CTO-10ASvp
Prolabo. HPLC-grade acetonitrile (Normapur) and meth- column oven. The chromatographic separation was per-
anol (Normapur) were obtained from Carlo-Erba. Formic formed on a Waters XTerra MS C18 [150 ⫻ 2.1 mm (i.d.);
acid and ammonium formate (⬎99% pure for both) were bead size, 5 ␮m] reversed-phase column maintained at
purchased from Sigma. Deionized water was prepared 30 °C. The flow rate was 0.2 mL/min. The mobile phase
on a Direct-Q laboratory plant (Millipore). For sample was a gradient of a mixture of acetonitrile and 2 mmol/L
extraction, Waters Oasis® HLB extraction cartridges (3 (pH 3) ammonium formate (90:10 by volume; solvent A)
mL/60 mg) were used. and 2 mmol/L (pH 3) ammonium formate (solvent B),
We prepared a pH 7.4 Sorensen buffer (66.7 mmol/L) programmed as follows: initial, 30% A, increased to 36%
by dissolving 9.07 g of potassium dihydrogen phosphate in 4 min, maintained at 36% for 8.5 min, then increased to
into 1 L of deionized water and by dissolving 11.6 g of 90% A in 3.5 min, and finally, decreased to 30% A in the
disodium hydrogen phosphate anhydrous into 1 L of last 1.5 min. All chromatographic solvents were degassed
deionized water. The disodium hydrogen phosphate so- with helium.
lution was then used to adjust the potassium dihydrogen A divert valve was set to direct the LC flow initially to
phosphate solution to pH 7.4. A solution of 0.5% ammonia the waste before 1.3 min and after 18 min to protect the
in methanol–water (40:60 by volume) was prepared by spectrometer from salts, polar compounds, and lipids.
adding 1.25 mL of concentrated ammonia to a mixture of
99.5 mL of methanol and 149.25 mL of water. ms detection
Stock solutions of desalkylflurazepam, ␣-hydroxytria- Detection was carried out with a Thermo-Electron TSQ
zolam, and ␣-hydroxyalprazolam were prepared at 10 Quantum Discovery MS/MS system equipped with an
mg/L in methanol; all stock solutions of the deuterated orthogonal ESI source and controlled by the Xcalibur
standards were prepared at 1 mg/L, and those of the computer program. Ionization was achieved with electro-
remaining drugs at 1 g/L in methanol. All solutions were spray in the positive ionization mode. The following
kept at ⫺20 °C in the dark. The drugs are stable for at least optimized conditions were used: spray voltage, 4 kV;
1348 Quintela et al.: LC-MS/MS Assay for BZPs

capillary temperature, 325 °C; sheath gas pressure (N2), 35 the extracted samples with those of the unextracted
kPa; auxiliary gas pressure (N2), 25 kPa. solutions, which represented 100% recovery.
For selected reaction monitoring (SRM) of the individ- We evaluated within-run precision and recovery by
ual compounds and their respective deuterated analogs, analyzing 6 samples in the same batch, enriched with all
the pseudomolecular ions [M⫹H]⫹ were selected in the substances studied, for 3 or 4 concentrations of the cali-
first quadrupole, and the collision energy was adjusted to bration set, depending on the linear dynamic range of the
optimize the signal for the most abundant product ions. A compound. We evaluated between-run precision and re-
quantification and a confirmation transition were chosen covery, as well as the linearity of the method, by prepar-
for each compound, except for the deuterated analogs, for ing and analyzing 1 set of calibrators each day for 5 days.
which only 1 transition was chosen (Table 1). The chro- Required specifications for within- and between-batch
matographic run was divided into 3 segments to limit the experiments were a CV ⬍15% (precision) and a mean
number of transitions monitored at a time in MS/MS relative error (MRE) ⬍15% from the nominal value at
mode and to ensure sensitivity for quantification pur- every concentration studied, except for the LLOQ, where
poses. In each segment, the compounds were monitored 20% was acceptable for both values (12–14 ). The limit of
according to their chromatographic separation (Table 1). detection was defined as the lowest concentration of the
The total LC-ESI-MS/MS method was optimized to drug giving a signal-to-noise ratio ⬎3:1 and was deter-
detect 32 substances (including 10 IS). mined for all compounds by analysis of a blank urine
enriched with decreasing amounts of the analytes. The
validation procedures LLOQ was the lowest concentration of the calibration
We performed assay validation according to the guide- curve that could be measured with an imprecision and
lines of the US Food and Drug Administration (12 ). We deviation from target concentration ⬍20%, in terms of CV
evaluated selectivity by analyzing urine samples from 10 and MRE, respectively (12–14 ).
healthy volunteers who did not take any of the targeted We investigated ion suppression by injecting 5 extracts
of different blank human urine samples into the system
compounds for several days before urine sampling and
while a mixture of the BZPs in mobile phase was contin-
checked for the absence of the compounds of interest by
uously infused in parallel at 50 ␮L/min in the ionization
analyzing the samples with the present technique before
source through a PEEK tee-junction. We investigated the
using them as blanks. We investigated assay linearity by
stability of the extracts in the autosampler by extracting,
constructing calibration curves (n ⫽ 5) using analysis data
in quadruplicates, a urine sample enriched with the 21
from a blank urine sample and 8 urine samples enriched
compounds at 10 ␮g/L. The second extract was injected
with the analytes at concentrations of 0.02, 0.05, 0.1, 0.5, 2,
6 h after the first, the third extract 12 h after, and the
10, 25, and 50 ␮g/L. Stable isotope IS were used to correct
fourth 15 h after.
for variability in the extraction procedure and the effect of
the ion suppression produced by the matrix. When it was
applications
not possible to obtain a commercially available deuterated This technique was used in a clinical trial approved by
analog for any compound, the deuterated analog of our local ethics committee. Twelve healthy volunteers
another close compound was selected. We determined the who gave informed consent were administered a single
best fit for accuracy for the calibration curves by applying dose of bromazepam, clonazepam, flunitrazepam, loraz-
linear and quadratic regression and different types of epam, zolpidem, or zopiclone (2 patients per drug) at the
weighting (1/x, 1/x2). lowest dosage available. Urine samples were collected
Extraction recoveries were determined in triplicate at 3 every 12 h for 1 week after dosing. The technique was also
concentrations for all compounds studied: the lower limit applied to a forensic sample obtained in a case of alleged
of quantification (LLOQ) was 0.02 or 0.05 ␮g/L, depend- DFSA.
ing on the molecule; the upper limit of quantification
(ULOQ) was 50 ␮g/L; and the intermediate limit was Results
2 ␮g/L. In the case of bromazepam and ␣-hydroxyalpra- Separation of the 21 compounds and their respective IS
zolam, the extraction recoveries were studied only at 2 was achieved in 18 min. A 6-min equilibration time was
and 50 ␮g/L because the first concentration (0.02 or necessary at the end of each run. The extracted ion
0.05 ␮g/L) was outside their calibration ranges. For each chromatograms of the 21 BZPs obtained for a urine
concentration, 3 urine samples were enriched with the sample enriched at 2 ␮g/L are presented in Fig. 1. The
appropriate amounts of drugs and with the IS, and in 3 selectivity of the method was acceptable in terms of
other samples, only the IS was added. After extraction absence of interference in the blank urine samples ana-
and evaporation, the first 3 samples were reconstituted lyzed. On the total ion chromatogram, not all of the
with 50 ␮L of mobile phase and the last 3 samples with chromatographic peaks resolved, but this was not a
50 ␮L of mobile phase containing a nominal amount of the problem because quantification and specific identification
compounds of interest. The relative extraction efficiency were performed for each compound with the correspond-
was calculated as the quotient of the peak-area ratios of ing transitions in the SRM mode. The combination of
Clinical Chemistry 52, No. 7, 2006 1349

Table 1. MS and chromatographic characteristics for tested BZPs.


Segment of MS experiment Collision
Compound (I, II, or III) SRM transition,a m/z energy, V RT, min
Zopiclone I 389.13245.1 26 2.3
389.13217.2 42
7-Aminoclonazepam I 286.23222.1 30 3.5
286.23250.1 26
Zolpidem I 308.03263.1 34 2.8
308.03235.1 30
7-Aminoflunitrazepam I 284.03135.0 34 4.3
284.03226.0 38
4-Hydroxymidazolam I 342.13325.0 26 3.9
342.13297.0 30
1-Hydroxymidazolam I 342.13324.0 26 5.8
342.13203.0 34
Loprazolam I 465.43408.1 30 4.5
465.43252.2 46
Midazolam I 326.13291.2 32 4.8
326.13249.2 40
Bromazepam I 316.03181.1 30 6.1
316.03209.2 34
␣-Hydroxytriazolam/4-Hydroxytriazolam II 359.03331.0 32 8.4
359.03250.0 42
␣-Hydroxyalprazolam II 325.13297.1 32 8.5
325.13216.1 44
3-Hydroxyflunitrazepam II 330.03284.1 24 9.4
330.03238.1 36
Desmethylflunitrazepam II 300.03254.1 30 10.0
300.03225.1 40
Alprazolam II 309.13281.2 28 11.1
309.13205.2 46
Lorazepam II 321.03275.1 28 11.2
321.03303.1 20
Triazolam II 343.03308.0 32 12.1
343.03239.0 48
Clonazepam II 316.03270.0 30 12.2
316.03241.0 40
Desalkylflurazepam III 289.03140.1 36 14.3
289.03226.1 34
Flurazepam I 388.23315.1 26 4.8
388.23317.1 24
Flunitrazepam III 314.03268.0 32 14.9
314.03239.0 40
Lormetazepam III 335.03289.1 26 16.6
335.03177.1 46
7-Aminoclonazepam-d4 I 290.03226.2 30 3.5
7-Aminoflunitrazepam-d7 I 291.03138.1 34 4.2
␣-Hydroxytriazolam-d4 II 363.13176.1 32 8.2
␣-Hydroxyalprazolam-d5 II 330.03302.1 30 8.3
Desmethylflunitrazepam-d4 II 304.13258.1 28 9.8
Alprazolam-d5 II 314.03286.1 32 10.8
Triazolam-d4 II 346.93312.1 30 11.8
Desalkylflurazepam-d4 III 293.13230.0 32 14.1
Flunitrazepam-d7 III 321.13246.2 30 14.5
a
The selected quantification transitions are in bold.
1350 Quintela et al.: LC-MS/MS Assay for BZPs

Fig. 1. Analysis of blank urine samples enriched with 2 ␮g/L each of the studied compounds.
TIC, total ion chromatogram.

retention time (RT) and 2 transitions (and their relative tration range. After assaying different types of regression
abundances) provided high specificity for all of the com- and weighing factors, we constructed the calibration
pounds. However, with this analytical method, we could curves using a 1/x weighted quadratic regression to
not distinguish ␣-hydroxytriazolam from 4-hydroxytria- obtain the best fit across the calibration range, based on
zolam because we obtained exactly the same RT and SRM the standard error of the fit and minimization of bias of
transitions for both, with the same intensity ratio. We the low calibrators. The squared correlation coefficient
therefore introduced ␣-hydroxytriazolam, but not 4-hy- (r2), y-intercept, and slope of the regression line are
droxytriazolam, into the calibration and control samples. summarized in Table 2. The mean r2 (n ⫽ 5) was ⬎0.998
In addition, 2 peaks showed up on the reconstituted for all of the compounds except for 4-hydroxymidazolam
chromatogram of lorazepam: the first peak corresponded (r2 ⫽ 0.998).
to lorazepam and the second to the IS, flunitrazepam-d7 Limits of detection between 0.01 and 0.02 ␮g/L were
(transition m/z 321.13246.2). However, these 2 com- observed for all of the compounds except for bromaz-
pounds were perfectly separated, so that flunitraz- epam and ␣-hydroxyalprazolam (0.05 ␮g/L). The LLOQ
epam-d7 could be kept as IS. The SRM transitions, opti- and ULOQ values for the compounds are shown as the
mized collision energies, and RT are shown in Table 1. linear dynamic range in Table 2. The within-run impreci-
The deuterated IS selected for each compound is sion (CV) was ⬍17% in all cases for the LLOQ, and was
shown in Table 2. No cross-talk interference with the ⬎15% for only 3 drugs: loprazolam, bromazepam, and
deuterated IS was observed. The peak-area ratios of the ␣-hydroxytriazolam. For the other concentrations, only
drugs to their respective deuterium-labeled IS were plot- one CV was ⬎15% (16.4% for loprazolam at the 2 ␮g/L
ted against the corresponding enriched concentrations. concentration). The within-run deviation from target was
The linearity of the method was verified over the concen- satisfactory for all compounds under the noted criteria,
Clinical Chemistry 52, No. 7, 2006 1351

Table 2. Calibration data and limits of detection for the examined compounds.
Linear dynamic range, Correlation
Compound IS ␮g/L LOD,a ␮g/L coefficient (r2)
Zopiclone 7-Aminoclonazepam-d4 0.05–50 0.01 0.999
7-Aminoclonazepam 7-Aminoclonazepam-d4 0.02–50 0.01 0.999
Zolpidem 7-Aminoclonazepam-d4 0.02–50 0.01 0.999
7-Aminoflunitrazepam 7-Aminoflunitrazepam-d7 0.02–50 0.01 0.999
4-Hydroxymidazolam 7-Aminoclonazepam-d4 0.05–50 0.02 0.998
1-Hydroxymidazolam 7-Aminoflunitrazepam-d7 0.05–50 0.02 0.999
Loprazolam 7-Aminoflunitrazepam-d7 0.05–50 0.02 0.999
Midazolam 7-Aminoflunitrazepam-d7 0.05–50 0.02 0.999
Bromazepam 7-Aminoflunitrazepam-d4 0.1–50 0.05 0.999
␣-Hydroxytriazolam/4-Hydroxytriazolam ␣-Hydroxytriazolam-d4 0.05–50 0.02 0.999
␣-Hydroxyalprazolam ␣-Hydroxyalprazolam-d5 0.1–50 0.05 0.999
3-Hydroxyflunitrazepam Desmethylflunitrazepam-d4 0.05–50 0.02 0.999
Desmethylflunitrazepam Desmethylflunitrazepam-d4 0.02–50 0.01 0.999
Alprazolam Alprazolam-d5 0.05–50 0.02 0.999
Lorazepam Triazolam-d4 0.02–50 0.01 0.999
Triazolam Triazolam-d4 0.02–50 0.01 0.999
Clonazepam Triazolam-d4 0.05–50 0.02 0.999
Desalkylflurazepam Desalkylflurazepam-d4 0.05–50 0.02 0.999
Flurazepam 7-Aminoclonazepam-d4 0.05–50 0.02 0.999
Flunitrazepam Desalkylflurazepam-d4 0.05–50 0.02 0.999
Lormetazepam Desalkylflurazepam-d4 0.02–50 0.01 0.999
a
LOD, limit of detection; defined as the lowest concentration of the drug giving a signal-to-noise ratio of 3:1.

except for loprazolam, for which the MRE value reached The clinical trial in healthy volunteers given a single
24% at the LLOQ, and bromazepam, for which it reached dose of bromazepam, clonazepam, flunitrazepam, loraz-
22.4% and 21.2% at the LLOQ and 2 ␮g/L, respectively. epam, zolpidem, or zopiclone (2 patients per drug) at the
The between-run CV was ⬍15% for all compounds at lowest dosage available showed that lorazepam, broma-
all concentrations of the calibration range, including the zepam, and 7-aminoflunitrazepam could be detected in
LLOQ, except for zolpidem, which had a slightly higher urine up to 144 h (1 week) post dose, whereas flunitraz-
result for the LLOQ (CV ⫽ 15%), although it was still epam was not detected after 132 h, zopiclone after 84 h,
acceptable according to the Food and Drug Administra- and zolpidem after 48 h in one volunteer and after 96 h in
tion recommendations. For all substances, between-run the other. These results demonstrate the suitability of this
MRE values were ⱕ14% at all concentrations, including method, which we have used successfully in our labora-
the LLOQ. tory for the investigation of actual DFSA cases.
The extraction recovery of the analytes at 3 concentra- The results obtained in a DFSA case are shown in Fig.
tions with the SPE procedure was 79.0%–99.3% at the 2. In this urine sample, the assay detected alprazolam at
LLOQ, 77.2%–99.6% at the intermediate concentration, 3.5 ␮g/L and its characteristic metabolite, ␣-hydroxy-
and 88.5%–110.0% at the ULOQ (Table 3). No significant alprazolam, at a low concentration of 0.17 ␮g/L.
ion suppression occurred during chromatographic runs,
except for alprazolam, for which the baseline decreased
almost 20% (see the Data Supplement that accompanies Discussion
the online version of this article at http:www.clinchem. Because higher concentrations of drugs and their metab-
org/content/vol52/issue7). The stability of the extracts in olites can usually be found in urine, it is the specimen of
the autosampler, studied with an enriched urine sample choice for toxicologic testing in DFSA cases (15 ). One of
extracted in quadruplicate (Table 4), showed that 4-hy- the most important issues in DFSA investigations is the
droxymidazolam may begin to deteriorate after 6 h and sensitivity of the screening and confirmatory techniques
midazolam and loprazolam after 12 h. This instability of because some of the compounds (e.g., BZPs) are typically
BZPs in solution, despite low temperatures (the auto- used in a single low dose (16, 17 ). Kintz et al. (18 )
sampler was cooled at 4 °C in this study) and protection demonstrated that the LC-MS and gas chromatography
from light, is well known and probably unavoidable. As a (GC)-negative chemical ionization-MS/MS techniques
consequence, when large batches of samples were to be are the most sensitive for detecting low quantities of the
analyzed, a portion of each extract was stored dry at 4 °C drug metabolite 7-aminoflunitrazepam several days after
before being reconstituted and placed in the autosampler administration of flunitrazepam. A highly sensitive GC-
a few hours before analysis. negative chemical ionization-MS method for the simulta-
1352 Quintela et al.: LC-MS/MS Assay for BZPs

Table 3. Extraction recovery of the 21 BZPs and metabolites.


Relative recovery, % (n ⴝ 3)

Lowa Mediumb Highc


Compound concentration concentration concentration
Zopiclone 83.5 91.2 96.6
7-Aminoclonazepam 84.1 94.1 94.0
Zolpidem 79.0 96.3 92.0
7-Aminoflunitrazepam 81.4 95.7 98.5
4-Hydroxymidazolam 95.5 96.5 99.3
1-Hydroxymidazolam 97.3 93.4 95.6
Loprazolam 92.4 91.5 100.0
Midazolam 97.0 94.2 91.8
Bromazepam 77.2 88.5
␣-Hydroxytriazolam/4-Hydroxytriazolam 92.6 99.3 94.6
␣-Hydroxyalprazolam 99.6 93.3
3-Hydroxyflunitrazepam 94.2 95.4 96.0
Desmethylflunitrazepam 83.3 96.8 95.8
Alprazolam 95.6 96.6 96.0
Lorazepam 90.0 97.6 91.6
Triazolam 83.1 95.5 91.3
Clonazepam 89.7 96.4 96.3
Desalkylflurazepam 91.7 96.4 110
Flurazepam 99.3 85.1 88.8
Flunitrazepam 95.7 95.9 90.4
Lormetazepam 79.4 86.8 91.5
a
LLOQ (0.02 or 0.05, depending on compound).
b
2 ␮g/L for all compounds.
c
ULOQ (50 ␮g/L for all compounds).

neous quantification of clonazepam and its major metab- LC-MS/MS) assays have been reported for forensic appli-
olite in urine detected the metabolite 28 days after a single cations, including use in DFSA cases (19 –23 ).
dose (3 mg) of clonazepam (19 ). Other LC-MS (and We evaluated several previous procedures for extrac-

Table 4. Stability in the autosampler of extracts of the 21 BZPs and metabolites at 10 ␮g/L.
Recovery (%), with elapsed time before analysis of:

0 6h 12 h 15 h
Zopiclone 100 80 85 113
7-Aminoclonazepam 100 97 101 104
Zolpidem 100 81 84 90
7-Aminoflunitrazepam 100 85 78 96
4-Hydroxymidazolam 100 77 71 61
1-Hydroxymidazolam 100 99 93 83
Loprazolam 100 85 80 76
Midazolam 100 99 86 69
Bromazepam 100 96 88 90
␣-Hydroxytriazolam 100 105 103 106
␣-Hydroxyalprazolam 100 103 99 105
3-Hydroxyflunitrazepam 100 97 89 91
Desmethylflunitrazepam 100 103 106 97
Alprazolam 100 108 107 106
Lorazepam 100 99 101 97
Triazolam 100 100 101 100
Clonazepam 100 108 112 105
Desalkylflurazepam 100 105 105 106
Flurazepam 100 92 91 99
Flunitrazepam 100 114 111 100
Lormetazepam 100 104 111 89
Clinical Chemistry 52, No. 7, 2006 1353

Fig. 2. Chromatograms of a forensic DFSA sample collected ⬍24 h after the assault.
The urine sample contained alprazolam and its metabolite ␣-hydroxyalprazolam at 3.5 and 0.17 ␮g/L, respectively.

tion of the BZPs and related compounds during develop- principal reason for our use of an SPE method. Variation
ment of the method we report here (24 ). The presence of in the pH of urine samples induced significant variations
compounds in a relatively wide polarity range was the in compound retention on the SPE column and in chro-
1354 Quintela et al.: LC-MS/MS Assay for BZPs

matogram background noise. The best results were ob- document drug-facilitated crimes: four cases involving bromaz-
tained by dilution of urine samples with Sorensen buffer epam. J Anal Toxicol 2004;28:516 –9.
(pH 7.4). Because of the high sensitivity of our method, 9. Drummer OH, Odell M, eds. The Forensic Pharmacology of Drugs
of Abuse. London: Arnold, 2001:103– 66.
which allowed detection of the parent drug and phase I
10. Kraemer T, Maurer H. Therapeutic drugs: sedatives and hypnotics.
metabolite at very low concentrations, and the instability In: Bogusz MJ, ed. Handbook of Analytical Separations. Amster-
of several analytes during incubation (mainly the amino dam: Elsevier Science BV, 2000:197–223.
metabolites), we processed the urine samples without 11. Salamone SJ, ed. Benzodiazepines and GHB: Detection and
enzymatic hydrolysis. Pharmacology. Totowa, NJ: Humana Press, 2001:160pp.
Our results demonstrate that this technique meets 12. US Department of Health and Human Services, Food and Drug
method validation criteria for measuring analytes of in- Administration. Guidance for industry: bioanalytical method val-
idation. http: // www.fda.gov/ cder/ guidance/4252fnl.pdf (ac-
terest in urine. In the case of bromazepam and loprazo-
cessed March 2006).
lam, certain values were not in the expected range, but 13. Shah VP, Midha KK, Findlay JW, Hill HM, Hulse JD, McGilveray IJ,
this should not present a problem for applying this et al. Bioanalytical method validation-a revisit with a decade of
technique to urine samples in DFSA cases, where precise progress. Pharm Res 2000;17:1551–7.
quantification is not the most important goal. 14. International Conference on Harmonization. Validation of analyti-
The use of HPLC to separate the compounds before cal procedures: text and methodology. Q2(R1). http://www.
detection by MS avoids the temperature-induced degra- ich.org/ (accessed March 2006).
15. LeBeau M, Andollo W, Hearn WL, Baselt R, Cone E, Finkle B, et al.
dation of certain BZPs that can occur when GC methods
Recommendations for toxicological investigations of drug-facili-
are used (10 ). Another major advantage of HPLC for tated sexual assaults. J Forensic Sci 1999;44:227–30.
analyzing BZPs is that no derivatization of the metabolites 16. Negrusz A, Gaensslen RE. Analytical developments in toxicological
is required (11 ). investigation of drug-facilitated sexual assault. Anal Bioanal Chem
Compared with other reports on analysis of several 2003;376:1192–7.
BZPs (25–35 ), in the screening method described in this 17. Couper FJ, Logan BK. Determination of ␥-hydroxybutyrate (GHB) in
study, the LLOQs for 19 BZPs or their major metabolites biological specimens by gas chromatography-mass spectrometry.
J Anal Toxicol 2000;24:1–7.
and 2 related compounds (zopliclone and zolpidem) are
18. Kintz P, Villain M, Cirimele V, Goulle JP, Ludes B. Crime under the
lower and seem to be appropriate for the detection and influence of psychoactive drugs: the problem of the duration of
quantification of these compounds in urine samples from detection. Acta Clin Belg 2001;1(Suppl):24 –30.
DFSA cases. 19. Negrusz A, Bowen AM, Moore CM, Dowd SM, Strong MJ, Janicak
PG. Elimination of 7-aminoclonazepam in urine after a single dose
of clonazepam. Anal Bioanal Chem 2003;376:1198 –204.
20. Marquet P, Lachâtre G. Liquid chromatography-mass spectrome-
We thank Bea López and Anna Chasnoff for helpful
try: potential in forensic and clinical toxicology. J Chromatogr B
assistance in the preparation of this manuscript. This Biomed Sci Appl 1999;733:93–118.
work was supported by a grant from the Ministry of 21. Kintz P, Villain M, Cirimele V, Pepin G, Ludes B. Windows of
Education and Science of Spain under the Formación de detection of lorazepam in urine, oral fluid, and hair, with a special
Profesorado Universitario Program. focus on drug-facilitated crimes. Forensic Sci Int 2004;145:
131–5.
22. Kintz P, Villain M, Dumestre-Toulet V, Ludes B. Drug-facilitated
References sexual assault and analytical toxicology: the role of LC-MS/MS: a
1. Stewart CH. Hypnotics and sedatives. In: Goodman Gilman A, case involving zolpidem. J Clin Forensic Med 2005;12:36 – 41.
Goodman LS, Rall TW, Murad F, eds. Gilman’s The Pharmacolog- 23. Quintela O, Cruz A, Castro A, Concheiro M, Lopez-Rivadulla M.
ical Basis of Therapeutics, 7th ed. New York: Macmillan Publish- Liquid chromatography-electrospray ionization mass spectrometry
ing Company, 1985:339 –71. for the determination of nine selected benzodiazepines in human
2. Currie D, Hashemi K, Fothergill J, Findlay A, Harris A, Hindmarch I. plasma and oral fluid. J Chromatogr B Analyt Technol Biomed Life
The use of anti-depressants and benzodiazepines in the perpetra- Sci 2005;825:63–71.
tors and victims of accidents. Occup Med 1995;45:323–5. 24. Drummer OH. Methods for the measurement of benzodiazepines
3. Ray WA, Fought RL, Decker MD. Psychoactive drugs and the risk of in biological samples. J Chromatogr B Biomed Sci Appl 1998;713:
injurious motor vehicle crashes in elderly drivers. Am J Epidemiol 201–25.
1992;136:873– 83. 25. Pirnay S, Ricordel I, Libong D, Bouchonnet S. Sensitive method for
4. Neutel CI. Benzodiazepine-related traffic accidents in young and the detection of 22 benzodiazepines by gas chromatography-ion
elderly drivers. Hum Psychopharmacol 1998;13(Suppl 2):115–23. trap tandem mass spectrometry. J Chromatogr A 2002;954:235–
45.
5. Barbone F, McMahon AD, Davey PG, Morris AD, Reid IC, McDevitt
26. Bishop SC, Lerch M, McCord BR. Micellar electrokinetic chromato-
DG, et al. Association of road-traffic accidents with benzodiaz-
graphic screening method for common sexual assault drugs
epine use. Lancet 1998;352:1331– 6.
administered in beverages. Forensic Sci Int 2004;141:7–15.
6. Logan BK, Couper FJ. Zolpidem and driving impairment. J Forensic 27. Jinno K, Taniguchi M, Hayashida M. Solid phase micro extraction
Sci 2001;46:105–10. coupled with semi-microcolumn high performance liquid chroma-
7. Dowd SM, Strong MJ, Janicak PG, Negrusz A. The behavioral and tography for the analysis of benzodiazepines in human urine.
cognitive effects of two benzodiazepines associated with drug- J Pharm Biomed Anal 1998;17:1081–91.
facilitated sexual assault. J Forensic Sci 2002;47:1101–7. 28. McClean S, O’Kane E, Hillis J, Smyth WF. Determination of
8. Villain M, Cheze M, Dumestre V, Ludes B, Kintz P. Hair to 1,4-benzodiazepines and their metabolites by capillary electro-
Clinical Chemistry 52, No. 7, 2006 1355

phoresis and high-performance liquid chromatography using ultra- metabolites, and benzodiazepine-like substances in whole blood
violet and electrospray ionization mass spectrometry. J Chro- by liquid chromatography-(tandem) mass spectrometry. J Chro-
matogr A 1999;838:273–91. matogr B Analyt Technol Biomed Life Sci 2004;811:13–20.
29. Rittner M, Pragst F, Bork WR, Neumann J. Screening method for 33. Yuan H, Mester Z, Lord H, Pawliszyn J. Automated in-tube solid-
seventy psychoactive drugs or drug metabolites in serum based phase microextraction coupled with liquid chromatography-electro-
on high-performance liquid chromatography-electrospray ioniza- spray ionization mass spectrometry for the determination of
tion mass spectrometry. J Anal Toxicol 2001;25:115–24. selected benzodiazepines. J Anal Toxicol 2000;24:718 –25.
30. Miki A, Tatsuno M, Katagi M, Nishikawa M, Tsuchihashi H. 34. Klette KL, Wiegand RF, Horn CK, Stout PR, Magluilo Jr J. Urine
Simultaneous determination of eleven benzodiazepine hypnotics benzodiazepine screening using Roche Online KIMS immuno-
and eleven relevant metabolites in urine by column-switching assay with ␤-glucuronidase hydrolysis and confirmation by gas
liquid chromatography-mass spectrometry. J Anal Toxicol 2002; chromatography-mass spectrometry. J Anal Toxicol 2005;29:
26:87–93. 193–200.
31. Zweigenbaum J, Heinig K, Steinborner S, Wachs T, Henion J. 35. Gunnar T, Ariniemi K, Lillsunde P. Determination of 14 benzodi-
High-throughput bioanalytical LC/MS/MS determination of benzo- azepines and hydroxy metabolites, zaleplon, and zolpidem as
diazepines in human urine: 1000 samples per 12 hours. Anal tert-butyldimethylsilyl derivatives compared with other common
Chem 1999;71:2294 –300. silylating reagents in whole blood by gas chromatography-mass
32. Smink BE, Brandsma JE, Dijkhuizen A, Lusthof KJ, de Gier JJ, spectrometry. J Chromatogr B Analyt Technol Biomed Life Sci
Egberts AC, et al. Quantitative analysis of 33 benzodiazepines, 2005;818:175– 89.

Вам также может понравиться