FIGURE 3-2: MAIN PC BOARD NEW BOARD

14 8 12 1
1 7 13 24

Power
Suppy
l

JP6
JP1

JP3
JP7
1 1 1 JP2
1

D. Load Cell Wiring


50514-4

TB1 Pin Number Description


1 - Excitation
2 + Excitation
3 + Sense
4 - Sense
5
6 Shield
7 + Signal
8 - Signal

See Figure 3-2 for location.

1. If standard load cell cable is over 25 feet long, then sense leads must be used.

2. If sense leads are used, then cut the jumpers at JP6 and JP7 located next to TB1.

E. Standard A/D Calibration

CAUTION: Recalibration of the Instrument will clear all Setpoint Programming.


Record all Setpoint Programming BEFORE Recalibration.

1. Press SW3 twice. After the first press, the display will show the program
number. After the second press the display will show “CAL 0.” Press 1 and
press the ENTER key once to enter the counts mode.

2. Remove all weight from the platform. When the display counts have
stabilized, press the ZERO key to store the new zero value or press ENTER
to retain the old zero reference.

NOTE: At any time during the calibration routine, up to the time the calibration
weight is entered, the process may be aborted by pressing the
GROSS/NET key, repeatedly, until the instrument returns to the weighmode.

12/06 41 50514 / SJ4819 Issue #4