US FDA QUALITY

SYSTEM REGULATION
VERSUS ISO 13485:2016
QUALITY MANAGEMENT
SYSTEM REQUIREMENTS
NSF INTERNATIONAL 21 CFR § 820 &
ISO 13485:2016 ALIGNMENT CHART
Author: Robert Ruff, Executive Director, NSF International
Researcher: Samuel Brown, Research Associate, NSF International

This tool clarifies the corresponding relationships between the US FDA Quality System Regulation and ISO 13485:2016
– Medical devices – Quality management systems – Requirements for regulatory purposes clauses. Use this tool to
ensure your quality management system meets applicable requirements of both US FDA and ISO 13485:2016

21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
820.1 Scope 1 Scope

Each manufacturer shall establish and


4.1.1 Quality
maintain a quality system that is appropriate
820.5 Quality management
for the specific medical device(s) designed No significant difference in requirements.
System system, General
or manufactured, and that meets the
requirements
requirements of this part.

Management with executive responsibility


shall establish its policy and objectives for,
5.3 Quality Policy ISO 13485:2016 specifies additional detail
and commitment to, quality. Management
820.20(a) relative to quality policy. For example: 5.3 b),
with executive responsibility shall ensure 5.4.1 Quality
Quality Policy 5.3 e). ISO 13485:2016 specifically requires
that the quality policy is understood, Objectives the quality objectives to be measurable.
implemented, and maintained at all levels of
the organization.

Each manufacturer shall establish and


maintain an adequate organizational structure 5.5.1
820.20(b)
to ensure that devices are designed and Responsibility and No significant difference in requirements.
Organization
produced in accordance with the requirements Authority
of this part.

Each manufacturer shall establish the


appropriate responsibility, authority, and
820.20(b)(1) 5.5.1
interrelation of all personnel who manage,
Responsibility Responsibility and No significant difference in requirements.
perform, and assess work affecting quality,
and Authority Authority
and provide the independence and authority
necessary to perform these tasks.

Each manufacturer shall provide adequate


resources, including the assignment of trained
820.20(b)(2) personnel, for management, performance 6.1 Provision of
No significant difference in requirements.
Resources of work, and assessment activities, including Resources
internal quality audits, to meet the
requirements of this part.

Management with executive responsibility


shall appoint, and document such
appointment of, a member of management
who, irrespective of other responsibilities,
shall have established authority over and ISO 13485:2016 specifies the management
responsibility for: representative has responsibility and authority
820.20(b)(3) 5.5.2
to ensure the promotion of awareness of
Management (i) Ensuring that quality system requirements Management
applicable regulatory requirements and
Representative are effectively established and effectively Representative
quality management system requirements
maintained in accordance with this part;
throughout the organization.
and
(ii) Reporting on the performance of the
quality system to management with
executive responsibility for review.
2
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
Management with executive responsibility
shall review the suitability and effectiveness 5.6 Management
of the quality system at defined intervals Review
ISO 13485:2016 specifies: (1) assessing
and with sufficient frequency according to 5.6.1 General
820.20(c) opportunities for improvement and the need
established procedures to ensure that the
Management 5.6.2 Review for changes to the QMS; and (2) specific
quality system satisfies the requirements of
Review Input requirements for management review input
this part and the manufacturer’s established
and output.
quality policy and objectives. The dates and 5.6.3 Review
results of quality system reviews shall be Output
documented.

Each manufacturer shall establish a quality ISO 13485:2016 specifies 5.4.2 Top
plan which defines the quality practices, management shall ensure that: b) The
5.4.2 Quality
820.20(d) resources, and activities relevant to devices integrity of the quality management
Management
Quality Planning that are designed and manufactured. system is maintained when changes to the
System Planning
The manufacturer shall establish how the quality management system are planned
requirements for quality will be met. and implemented.

Each manufacturer shall establish quality 4.2 Documentation


820.20(e) system procedures and instructions. An Requirements 21 CFR § 820 specifies an outline of the
structure of the documentation used in the
Quality System outline of the structure of the documentation 4.2.1 General
quality system; and ISO 13485:2016 specifies
Procedures used in the quality system shall be established
4.2.2 Quality a Quality Manual.
where appropriate.
Manual

Each manufacturer shall establish procedures


for quality audits and conduct such audits
to assure that the quality system is in 21 CFR § 820 specifies the date of the
compliance with the established quality audit be documented and the audit reports
system requirements and to determine the must be reviewed by management having
effectiveness of the quality system. Quality responsibility for the matters audited.
audits shall be conducted by individuals
ISO 13485:2016 specifies requirements for
who do not have direct responsibility for the
820.22 Quality 8.2.4 Internal the management responsible for the area
matters being audited. Corrective action(s),
Audit Audit being audited to ensure that corrections
including a reaudit of deficient matters, shall
and corrective actions are taken without
be taken when necessary. A report of the
undue delay;
results of each quality audit, and reaudit(s)
where taken, shall be made and such reports ISO 13485:2016 also specifies the
shall be reviewed by management having verification of the actions taken and
responsibility for the matters audited. The reports of the verification results.
dates and results of quality audits and
reaudits shall be documented.

Each manufacturer shall have sufficient 6.1 Provision of


820.25(a) personnel with the necessary education, Resources
Personnel, background, training, and experience to No significant difference in requirements.
General assure that all activities required by this part 6.2 Human
are correctly performed Resources

Each manufacturer shall establish procedures


for identifying training needs and ensure ISO 13485:2016 specifies more detail
that all personnel are trained to adequately than 21 CFR § 820 and addresses
perform their assigned responsibilities. “competence” as opposed to training
Training shall be documented. (e.g. competence via education, skills,
6.2 Human
820.25(b) (1) As part of their training, personnel experience).
Resources
Personnel, shall be made aware of device defects 21 CFR § 820 specifies requirements for:
Training which may occur from the improper 6.4.1(b) Work
(1) personnel performing verification and
performance of their specific jobs. Environment
validation activities, and (2) 21 CFR § 820
(2) Personnel who perform verification states that personnel shall be made aware
and validation activities shall be made of device defects which may occur from
aware of defects and errors that may be improper performance of their jobs.
encountered as part of their job functions.

3
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
(1) Each manufacturer of any class III or
class II device, and the class I devices
listed in paragraph (a)(2) of this section,
shall establish and maintain procedures
to control the design of the device in
order to ensure that specified design
requirements are met.
(2) The following class I devices are subject
to design controls:
(i) Devices automated with computer
software; and 21 CFR § 820 specifies exclusions from
820.30(a)
design control requirements based on the
Design Controls, (ii) The devices listed in the following 7.3.1 General
risk of the device (i.e. class I devices other
General chart. than those cited).
Section Device
868.6810 Catheter, Tracheobronchial
Suction.
878.4460 Glove, Surgeon’s.
880.6760 Restraint, Protective.
892.5650 System, Applicator, Radionuclide,
Manual.
892.5740 Source, Radionuclide Teletherapy.

Each manufacturer shall establish and maintain


plans that describe or reference the design and 21 CFR § 820 requires documented design
development activities and define responsibility plans.
820.30(b)
for implementation. The plans shall identify 7.3.2 Design and ISO 13485:2016 does not require a
Design and
and describe the interfaces with different development documented design and development plan,
Development
groups or activities that provide, or result in, planning but does require the documentation of
Planning
input to the design and development process. information specified in 7.3.2 a) - f) during
The plans shall be reviewed, updated, and design and development planning.
approved as design and development evolves.

Each manufacturer shall establish and


maintain procedures to ensure that the
design requirements relating to a device are
appropriate and address the intended use of
the device, including the needs of the user 21 CFR § 820 specifies the documentation
and patient. The procedures shall include of the date and signature of the individual(s)
7.3.3 Design and approving the design inputs.
820.30(c) a mechanism for addressing incomplete,
development
Design Input ambiguous, or conflicting requirements. ISO 13485:2016 specifies specific inputs;
inputs
The design input requirements shall be including applicable output(s) of risk
documented and shall be reviewed and management.
approved by a designated individual(s). The
approval, including the date and signature of
the individual(s) approving the requirements,
shall be documented.

Each manufacturer shall establish and maintain


procedures for defining and documenting
design output in terms that allow an adequate
evaluation of conformance to design input
requirements. Design output procedures shall
contain or make reference to acceptance
7.3.4 Design and ISO 13485:2016 specifies "b) provide
820.30(d) criteria and shall ensure that those design
development appropriate information for purchasing,
Design Output outputs that are essential for the proper
outputs production and service provision."
functioning of the device are identified.
Design output shall be documented, reviewed,
and approved before release. The approval,
including the date and signature of the
individual(s) approving the output, shall be
documented.
4
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
Each manufacturer shall establish and maintain
procedures to ensure that formal documented
reviews of the design results are planned
and conducted at appropriate stages of the
device's design development. The procedures
shall ensure that participants at each design 21 CFR § 820 specifies design reviews
review include representatives of all functions include "individual(s) who does not have
7.3.5 Design and direct responsibility for the design stage
820.30(e) concerned with the design stage being
development being reviewed".
Design Review reviewed and an individual(s) who does not
review
have direct responsibility for the design stage ISO 13485:2016 specifies that reviews will
being reviewed, as well as any specialists "identify and propose necessary actions".
needed. The results of a design review,
including identification of the design, the date,
and the individual(s) performing the review,
shall be documented in the design history file
(the DHF).

Each manufacturer shall establish and maintain


procedures for verifying the device design. ISO 13485:2016 specifies requirements for
Design verification shall confirm that the design verification of connectivity or interfaces with
7.3.6 Design and other medical devices as applicable.
820.30(f) Design output meets the design input requirements.
development
Verification The results of the design verification, including 21 CFR § 820 specifies that the individual(s)
verification
identification of the design, method(s), the performing the verification and the date of
date, and the individual(s) performing the the activity be documented.
verification, shall be documented in the DHF.

Each manufacturer shall establish and ISO 13485:2016 specifies requirements


maintain procedures for validating the device for validation of the device while connected
design. Design validation shall be performed to or interfaced with other medical devices
under defined operating conditions on initial as applicable.
production units, lots, or batches, or their
ISO 13485:2016 specifies “the organization
equivalents. Design validation shall ensure
shall perform clinical evaluations or
that devices conform to defined user needs
820.30(g) 7.3.7 Design and performance evaluations of the medical
and intended uses A1:D91 shall include
Design development device in accordance with applicable
testing of production units under actual or
Validation validation regulatory requirements” - which is not
simulated use conditions. Design validation shall
explicit to 21 CFR § 820 - but is expected as
include software validation and risk analysis,
discussed in the preamble comment 65.
where appropriate. The results of the design
validation, including identification of the design, 21 CFR § 820 specifies additional
method(s), the date, and the individual(s) documentation requirements such as the
performing the validation, shall be documented documentation (within the DHF) of the date
in the DHF. and individuals performing the validation.

ISO 13485:2016 specifies "that design and


development outputs are verified as suitable
for manufacturing before becoming final
production specifications and that production
capability can meet product requirements".
Each manufacturer shall establish and While this is not explicitly stated in 21
7.3.8 Design and
820.30(h) maintain procedures to ensure that the device CFR § 820, it is the expectation of design
development
Design Transfer design is correctly translated into production verification activities as well as design transfer
transfer
specifications. activities. Comment 86 of the QSR Preamble
states "The intent of the requirement was to
ensure that all design specifications released
to production have been approved, verified,
and validated before they are implemented
as part of the production process."
5
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
ISO 13485:2016 specifies additional
requirements. For example: ISO 13485:2016
specifies (1) "The review of design and
development changes shall include
Each manufacturer shall establish and evaluation of the effect of the changes on
7.3.9
maintain procedures for the identification, constituent parts and product in process
820.30(i) Design Control of design
documentation, validation or where appropriate or already delivered, inputs or outputs of
Changes and development
verification, review, and approval of design risk management and product realization
changes
changes before their implementation. processes" and (2) "The organization shall
determine the significance of the change to
function, performance, usability, safety and
applicable regulatory requirements for the
medical device."

Each manufacturer shall establish and maintain


a DHF for each type of device. The DHF shall
820.30(j) Design contain or reference the records necessary to 7.3.10 Design and
No significant difference in requirements.
History File demonstrate that the design was developed in development files
accordance with the approved design plan and
the requirements of this part.

Each manufacturer shall establish and maintain 4.2.4 Control of


820.40 documents
procedures to control all documents that are
Document No significant difference in requirements.
required by this part. The procedures shall 4.2.5 Control of
Controls
provide for the following: records

Each manufacturer shall designate an


individual(s) to review for adequacy and 21 CFR § 820 specifies the documentation
approve prior to issuance all documents of the date and signature of the individual(s)
established to meet the requirements of this approving the document.
part. The approval, including the date and 4.2.4 Control of
820.40(a) ISO 13485:2016 specifies documented
signature of the individual(s) approving the documents
Document procedures to “ensure that documents
document, shall be documented. Documents
approval and 4.2.5 Control of of external origin, determined by the
established to meet the requirements of this
distribution records organization to be necessary for the planning
part shall be available at all locations for
which they are designated, used, or otherwise and operation of the quality management
necessary, and all obsolete documents shall system, are identified and their distribution
be promptly removed from all points of use or controlled.”
otherwise prevented from unintended use.

Changes to documents shall be reviewed


and approved by an individual(s) in the same
function or organization that performed the 21 CFR § 820 specifies that approved
original review and approval, unless specifically changes be communicated to appropriate
designated otherwise. Approved changes shall personnel in a timely manner.
820.40(b) be communicated to the appropriate personnel 21 CFR § 820 specifies “Change records
4.2.4 Control of
Document in a timely manner. Each manufacturer shall shall include a description of the change,
documents
Changes maintain records of changes to documents. identification of the affected documents,
Change records shall include a description the signature of the approving individual(s),
of the change, identification of the affected the approval date, and when the change
documents, the signature of the approving becomes effective.”
individual(s), the approval date, and when the
change becomes effective.

6
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
ISO 13485:2016 specifies written quality
agreements between the organization and
Each manufacturer shall establish and 4.1.5 QMS General suppliers.
820.50 maintain procedures to ensure that all Requirements 21 CFR § 820 requires “... purchased or
Purchasing purchased or otherwise received product
7.4.1 Purchasing otherwise received product” to be included
Controls and services conform to specified
process in the Purchasing Controls procedures.
requirements.
ISO 13485:2016 limits its requirement to
purchased product.

Each manufacturer shall establish and


maintain the requirements, including
quality requirements, that must be met
by suppliers, contractors, and consultants.
Each manufacturer shall: ISO 13485:2016 specifies purchasing
criteria to be "d) proportionate to the
(1) Evaluate and select potential suppliers,
risk associated with the medical device."
contractors, and consultants on the
820.50(a) Although not explicitly stated in 21 CFR §
basis of their ability to meet specified
Evaluation 7.4.1 820, comment 115 of the QSR Preamble
requirements, including quality
of Suppliers, Purchasing discusses this requirement.
requirements. The evaluation shall be
Contractors, documented. process ISO 13485:2016 specifies requirements
and Consultants relative to the re-evaluation of suppliers.
(2) Define the type and extent of control to
Although not explicitly stated in 21 CFR
be exercised over the product, services,
§ 820, comment 105 of the preamble
suppliers, contractors, and consultants,
discusses this requirement.
based on the evaluation results.
(3) Establish and maintain records of
acceptable suppliers, contractors, and
consultants.

Each manufacturer shall establish and


maintain data that clearly describe or
reference the specified requirements, ISO 13485:2016 specifies "The
including quality requirements, for organization shall ensure the adequacy of
purchased or otherwise received product specified purchasing requirements prior to
and services. Purchasing documents shall their communication to the supplier."
7.4.2
820.50(b) include, where possible, an agreement that ISO 13485:2016 specifies "To the extent
Purchasing
Purchasing Data the suppliers, contractors, and consultants required for traceability given in 7.5.9,
information
agree to notify the manufacturer of the organization shall maintain relevant
changes in the product or service so that purchasing information in the form of
manufacturers may determine whether the documents (see 4.2.4) and records
changes may affect the quality of a finished (see 4.2.5)."
device. Purchasing data shall be approved in
accordance with 820.40.

Each manufacturer shall establish and ISO 13485:2016 specifies more detailed
maintain procedures for identifying product requirements including the documentation
820.60
during all stages of receipt, production, 7.5.8 Identification of procedures to ensure medical devices
Identification
distribution, and installation to prevent returned to the organization are identified
mix-ups. and distinguished from conforming product.

Each manufacturer of a device that is


intended for surgical implant into the body
or to support or sustain life and whose
failure to perform when properly used
in accordance with instructions for use
provided in the labeling can be reasonably ISO 13485:2016 requires procedures for
820.65 expected to result in a significant injury 7.5.9 Traceability traceability based on regulatory requirements.
Traceability to the user shall establish and maintain 7.5.9.1 General 21 CFR § 820 specifies the types of devices
procedures for identifying with a control that require traceability.
number each unit, lot, or batch of finished
devices and where appropriate components.
The procedures shall facilitate corrective
action. Such identification shall be
documented in the DHR.

7
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
Each manufacturer shall develop, conduct,
control, and monitor production processes
to ensure that a device conforms to its
specifications. Where deviations from device
specifications could occur as a result of the
manufacturing process, the manufacturer
shall establish and maintain process control
procedures that describe any process
controls necessary to ensure conformance
to specifications. Where process controls are
needed they shall include: In this clause, ISO 13485:2016 specifies
(1) Documented instructions, standard requirement that may align with other
820.70(a)
operating procedures (SOP’s), and 21 CFR § 820.70 requirements beyond
Production 7.5.1 Control of
methods that define and control the 21 CR 820.70(a). For example, in clause
and Process Production and
manner of production; 7.5.1, ISO 13485:2016 specifies "b)
Controls, Service Provision
qualification of infrastructure"; whereas,
General (2) Monitoring and control of process this requirement appears in "820.70(f)
parameters and component and device Buildings" of 21 CFR § 820.
characteristics during production;
(3) Compliance with specified reference
standards or codes;
(4) The approval of processes and process
equipment; and
(5) Criteria for workmanship which shall
be expressed in documented standards
or by means of identified and approved
representative samples.

Each manufacturer shall establish and


maintain procedures for changes to
a specification, method, process, or
820.70(b)
procedure. Such changes shall be verified 4.1.4 Quality
Production
or where appropriate validated according to Management No significant difference in requirements.
and Process
820.75, before implementation and these System
Changes
activities shall be documented. Changes
shall be approved in accordance with
820.40.

Where environmental conditions could


reasonably be expected to have an adverse
effect on product quality, the manufacturer
shall establish and maintain procedures to
820.70(c) adequately control these environmental
6.4.1 Work
Environmental conditions. Environmental control system(s) No significant difference in requirements.
environment
Control shall be periodically inspected to verify that
the system, including necessary equipment,
is adequate and functioning properly.
These activities shall be documented and
reviewed.

Each manufacturer shall establish and


maintain requirements for the health,
cleanliness, personal practices, and clothing
of personnel if contact between such
personnel and product or environment
820.70(d) could reasonably be expected to have 6.4.1 Work
No significant difference in requirements.
Personnel an adverse effect on product quality. environment
The manufacturer shall ensure that
maintenance and other personnel who are
required to work temporarily under special
environmental conditions are appropriately
trained or supervised by a trained individual.

8
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
ISO 13485:2016's specified requirements
are "as appropriate" unless relative to sterile
medical devices. However, as stated in ISO
13485:2016 0.2 Clarification of concepts,
when a requirement is qualified by the
Each manufacturer shall establish and maintain phrase "as appropriate", it is deemed to
820.70(e) procedures to prevent contamination of 6.4.2 be appropriate unless the organization can
Contamination equipment or product by substances that could Contamination justify otherwise.
Control reasonably be expected to have an adverse control For sterile devices, ISO 13485:2016 specifies
effect on product quality. "... the organization shall document
requirements for control of contamination
with microorganisms or particulate matter
and maintain the required cleanliness during
assembly or packaging processes." 21 CFR §
820 is not this specific.

Buildings shall be of suitable design and


820.70(f) contain sufficient space to perform necessary
6.3 Infrastructure No significant difference in requirements.
Buildings operations, prevent mix-ups, and assure orderly
handling.

Each manufacturer shall ensure that all


equipment used in the manufacturing
process meets specified requirements and is
appropriately designed, constructed, placed,
and installed to facilitate maintenance,
adjustment, cleaning, and use.
(1) Maintenance schedule. Each manufacturer
shall establish and maintain schedules
for the adjustment, cleaning, and other
maintenance of equipment to ensure 21 CFR § 820 specifies specific activities
that manufacturing specifications are relative to the maintenance schedules,
met. Maintenance activities, including periodic inspections to assure adherence
6.3 Infrastructure
the date and individual(s) performing the to equipment maintenance schedules, and
820.70(g) maintenance activities, shall be documented. 7.5.1 Control of posting of inherent limitations or allowable
Equipment production and tolerances.
(2) Inspection. Each manufacturer shall conduct
service provision
periodic inspections in accordance with 21 CFR § 820 also requires the
established procedures to ensure adherence documentation of dates and individuals
to applicable equipment maintenance conducting specified activities.
schedules. The inspections, including the
date and individual(s) conducting the
inspections, shall be documented.
(3) Adjustment. Each manufacturer shall ensure
that any inherent limitations or allowable
tolerances are visibly posted on or near
equipment requiring periodic adjustments or
are readily available to personnel performing
these adjustments.

Where a manufacturing material could


reasonably be expected to have an adverse
effect on product quality, the manufacturer
shall establish and maintain procedures for
820.70(h) ISO 13485:2016 specifies conditions that
the use and removal of such manufacturing 7.5.2 Cleanliness
Manufacturing would require documentation of cleanliness
material to ensure that it is removed or limited of Product
Material requirements.
to an amount that does not adversely affect
the device's quality. The removal or reduction
of such manufacturing material shall be
documented.
9
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
4.1.6 QMS General
When computers or automated data processing Requirements
systems are used as part of production or 7.5.6 Validation
the quality system, the manufacturer shall of processes for
820.70(i)
validate computer software for its intended production and
Automated No significant difference in requirements.
use according to an established protocol. All service provision
Processes
software changes shall be validated before
7.6 Control
approval and issuance. These validation
of monitoring
activities and results shall be documented.
and measuring
equipment

Each manufacturer shall ensure that all


inspection, measuring, and test equipment,
including mechanical, automated, or electronic
inspection and test equipment, is suitable for its
820.72(a) intended purposes and is capable of producing
7.6 Control
Control of valid results. Each manufacturer shall establish
of monitoring
Inspection, and maintain procedures to ensure that No significant difference in requirements.
and measuring
Measuring, and equipment is routinely calibrated, inspected,
equipment
Test Equipment checked, and maintained. The procedures shall
include provisions for handling, preservation,
and storage of equipment, so that its accuracy
and fitness for use are maintained. These
activities shall be documented.

Calibration procedures shall include specific


directions and limits for accuracy and precision.
When accuracy and precision limits are not met,
there shall be provisions for remedial action to
reestablish the limits and to evaluate whether
there was any adverse effect on the device’s
quality. These activities shall be documented.
(1) Calibration standards. Calibration standards
used for inspection, measuring, and test
equipment shall be traceable to national
or international standards. If national or
21 CFR § 820 specifies specific
international standards are not practical 7.6 Control
documentation requirements including: "The
820.72(b) or available, the manufacturer shall use an of monitoring
equipment identification, calibration dates,
Calibration independent reproducible standard. If no and measuring
the individual performing each calibration,
applicable standard exists, the manufacturer equipment
and the next calibration date..."
shall establish and maintain an in-house
standard.
(2) Calibration records. The equipment
identification, calibration dates, the individual
performing each calibration, and the next
calibration date shall be documented.
These records shall be displayed on or
near each piece of equipment or shall be
readily available to the personnel using such
equipment and to the individuals responsible
for calibrating the equipment.

7.5.6 Validation
Where the results of a process cannot be fully of processes for
verified by subsequent inspection and test, the production and
process shall be validated with a high degree service provision
820.75(a) of assurance and approved according to 7.5.7 Particular 21 CFR § 820 specifies the documentation
Process established procedures. The validation activities requirements of the date and signature of the individual(s)
Validation and results, including the date and signature for validation approving the validation.
of the individual(s) approving the validation of processes for
and where appropriate the major equipment sterilization and
validated, shall be documented. sterile barrier
systems
10
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
Each manufacturer shall establish and maintain
procedures for monitoring and control of
process parameters for validated processes to
ensure that the specified requirements continue ISO 13485:2016 clause 8.2.5 specifies
to be met. the expectations for monitoring and
measurement of processes.
820.75(b) (1) Each manufacturer shall ensure that 8.2.5 Monitoring
Process validated processes are performed by and measurement 21 CFR § 820 specifies that the monitoring
Validation qualified individual(s). of processes and control methods and data, the date
performed, and individual(s) performing
(2) For validated processes, the monitoring the process or the major equipment be
and control methods and data, the date documented.
performed, and, where appropriate, the
individual(s) performing the process or the
major equipment used shall be documented.

4.1.4 QMS General


Requirements
When changes or process deviations occur, 7.5.6 Validation
820.75(c) the manufacturer shall review and evaluate of processes for
Process the process and perform revalidation production and No significant difference in requirements.
Validation where appropriate. These activities shall be service provision
documented.
8.2.5 Monitoring
and measurement
of processes

820.80(a) 7.1 Planning


Receiving, Each manufacturer shall establish and of product
in-process, and maintain procedures for acceptance activities. realization
No significant difference in requirements.
finished device Acceptance activities include inspections, tests, 8.2.6 Monitoring
acceptance, or other verification activities. and measurement
General of product

Each manufacturer shall establish and maintain


820.80(b) procedures for acceptance of incoming product.
7.4.3 Verification
Receiving Incoming product shall be inspected, tested, or
of purchased No significant difference in requirements.
Acceptance otherwise verified as conforming to specified
product
Activities requirements. Acceptance or rejection shall be
documented.

Each manufacturer shall establish and maintain


acceptance procedures, where appropriate, to 7.5.10 Customer
820.80(c) ensure that specified requirements for in-process property
In-Process product are met. Such procedures shall ensure
8.2.6 Monitoring No significant difference in requirements.
Acceptance that in-process product is controlled until the
Activities required inspection and tests or other verification and measurement
activities have been completed, or necessary of product
approvals are received, and are documented.

Each manufacturer shall establish and maintain


procedures for finished device acceptance to
ensure that each production run, lot, or batch
of finished devices meets acceptance criteria.
Finished devices shall be held in quarantine or
otherwise adequately controlled until released.
Finished devices shall not be released for 7.5.10 Customer
820.80(d) Final distribution until: property 21 CFR § 820 specifies “Finished devices
Acceptance 8.2.6 Monitoring shall be held in quarantine or otherwise
(1) The activities required in the DMR are
Activities and measurement adequately controlled until released”.
completed;
of product
(2) the associated data and documentation is
reviewed;
(3) the release is authorized by the signature of
a designated individual(s); and
(4) the authorization is dated.
11
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
Each manufacturer shall document acceptance
activities required by this part. These records shall
include: 4.2.5 Control of
records
(1) The acceptance activities performed;
7.1 Planning
820.80(e) (2) the dates acceptance activities are performed; of product Although both require records, 21 CFR
Acceptance § 820 specifies specific record content
(3) the results; realization
Records requirements.
(4) the signature of the individual(s) conducting 8.2.6 Monitoring
the acceptance activities; and and measurement
of product
(5) where appropriate the equipment used.
These records shall be part of the DHR.

Each manufacturer shall identify by suitable


means the acceptance status of product, to
indicate the conformance or nonconformance
of product with acceptance criteria. The
820.86
identification of acceptance status shall be
Acceptance 7.5.8 Identification No significant difference in requirements.
maintained throughout manufacturing,
status
packaging, labeling, installation, and servicing of
the product to ensure that only product which
has passed the required acceptance activities is
distributed, used, or installed.

Each manufacturer shall establish and maintain


procedures to control product that does
not conform to specified requirements. The
procedures shall address the identification,
820.90(a) documentation, evaluation, segregation, and 8.3 Control of
nonconforming Standard Clause 8.3.3 specifies requirements
Control of disposition of nonconforming product. The
product for nonconforming product detected after
Nonconforming evaluation of nonconformance shall include a
delivery.
Product determination of the need for an investigation 8.3.1 General
and notification of the persons or organizations
responsible for the nonconformance. The
evaluation and any investigation shall be
documented.

1) Each manufacturer shall establish and


maintain procedures that define the
responsibility for review and the authority
for the disposition of nonconforming
product. The procedures shall set forth the
review and disposition process. Disposition 8.3.2 Actions
of nonconforming product shall be in response to
documented. Documentation shall include nonconforming
the justification for use of nonconforming product detected
820.90(b) product and the signature of the individual(s) before delivery
Nonconformity authorizing the use. 8.3.3 Actions No significant difference in requirements.
Review and
(2) Each manufacturer shall establish and in response to
Disposition
maintain procedures for rework, to nonconforming
include retesting and reevaluation of the product detected
nonconforming product after rework, after delivery
to ensure that the product meets its 8.3.4 Rework
current approved specifications. Rework
and reevaluation activities, including a
determination of any adverse effect from
the rework upon the product, shall be
documented in the DHR.
12
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
Each manufacturer shall establish and
maintain procedures for implementing
corrective and preventive action. The
procedures shall include requirements for:
(1) Analyzing processes, work operations,
concessions, quality audit reports, quality
records, service records, complaints,
returned product, and other sources
of quality data to identify existing and
potential causes of nonconforming
product, or other quality problems.
Appropriate statistical methodology shall 21 CFR § 820 specifies “(4) Verifying or
be employed where necessary to detect validating the corrective and preventive
recurring quality problems; action to ensure that such action is effective
(2) Investigating the cause of nonconformities and does not adversely affect the finished
relating to product, processes, and the 8.4 Analysis of device.“
quality system; data 21 CFR § 820 specifies “(6) Ensuring that
820.100(a) (3) Identifying the action(s) needed to correct 8.5 Improvement information related to quality problems or
and prevent recurrence of nonconforming nonconforming product is disseminated to
Corrective 8.5.1 General
product and other quality problems; those directly responsible for assuring the
and Preventive 8.5.2 Corrective quality of such product or the prevention of
Action (4) Verifying or validating the corrective and Action such problems.”
preventive action to ensure that such
action is effective and does not adversely 8.5.3 Preventive ISO 13485:2016 contains specific clauses
affect the finished device; Action relative to “corrective action” and
“preventive action”.
(5) Implementing and recording changes
in methods and procedures needed to ISO 13485:2016 requires corrective action to
correct and prevent identified quality be taken “without undue delay.”
problems;
(6) Ensuring that information related to
quality problems or nonconforming
product is disseminated to those directly
responsible for assuring the quality of
such product or the prevention of such
problems; and
(7) Submitting relevant information on
identified quality problems, as well as
corrective and preventive actions, for
management review.

820.100(b) 8.5.2 Corrective


Corrective All activities required under this section, and Action
No significant difference in requirements.
and Preventive their results, shall be documented. 8.5.3 Preventive
Action Action

Each manufacturer shall establish and 7.5.1 Control of


820.120 Device
maintain procedures to control labeling production and No significant difference in requirements.
Labeling
activities. service provision

ISO 13485:2016 does not prescribe specific


requirements for labeling other than stating
that production controls shall include
“implementation of defined operations for
7.5.1 Control of
Labels shall be printed and applied so labelling and packaging.”
production and
as to remain legible and affixed during service provision 21 CFR § 820 provides specific detail
820.120(a)
the customary conditions of processing, relative to label integrity. However, ISO
Label Integrity 7.5.11
storage, handling, distribution, and where 13485:2016 requires preservation of product
appropriate use. Preservation of
as well as design verification and design
product
validation which would include confirming
label integrity throughout the product’s
production, packaging, distribution and
storage channels; and environments.

13
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
Labeling shall not be released for storage ISO 13485:2016 does not prescribe
or use until a designated individual(s) has 7.4.3 Verification specific requirements for labeling other
examined the labeling for accuracy including, of purchased than stating that production controls
where applicable, the correct unique device product shall include “implementation of defined
820.120(b) identifier (UDI) or universal product code 7.5.1 Control of operations for labelling and packaging”
Labeling (UPC), expiration date, control number, production and and does not specifically address the
Inspection storage instructions, handling instructions, service provision inspection of labels. However, 8.2.6
and any additional processing instructions. Monitoring and measurement of product
8.2.6 Monitoring
The release, including the date and and 7.4.3 Verification of purchased
and measurement
signature of the individual(s) performing the products should account for the additional
of product
examination, shall be documented in the DHR. activities cited in 21 CFR § 820.120(b).

ISO 13485:2016 does not prescribe


6.4.1 Work specific requirements for labeling other
environment than stating that production controls
820.120(c) Each manufacturer shall store labeling in a
shall include "implementation of defined
Labeling manner that provides proper identification 7.5.1 Control of operations for labelling and packaging"
Storage and is designed to prevent mix-ups. production and and does not specifically address the
service provision labeling storage requirements cited in 21
CFR § 820.

Each manufacturer shall control labeling and 7.5.1 Control of


production and ISO 13485:2016 does not prescribe specific
820.120(d) packaging operations to prevent labeling
service provision requirements for documenting labels and
Labeling mix-ups. The label and labeling used for
labeling used for each production unit, lot,
Operations each production unit, lot, or batch shall be 4.2.3 Medical or batch in a specific file/record.
documented in the DHR. device file

7.5.8 Identification ISO 13485:2016 does not require control


Where a control number is required by
7.5.9.2 Particular numbers, although it does specify
820.120(e) 820.65, that control number shall be on
requirements requirements for UDI (when required)
Control Number or shall accompany the device through
for implantable and traceability for implantable medical
distribution.
medical devices devices.

7.3 Design and


Each manufacturer shall ensure that device development
packaging and shipping containers are 7.5.1 Control of
820.130 Device designed and constructed to protect the production and
No significant difference in requirements.
Packaging device from alteration or damage during the service provision
customary conditions of processing, storage,
7.5.11
handling, and distribution.
Preservation of
product

Each manufacturer shall establish and


maintain procedures to ensure that mix-ups, 7.5.11 21 CFR § 820 specifies procedures must
820.140
damage, deterioration, contamination, or Preservation of include provisions to assure "mix-ups" do
Handling
other adverse effects to product do not occur product not occur.
during handling.

Each manufacturer shall establish and


maintain procedures for the control of
storage areas and stock rooms for product
to prevent mix-ups, damage, deterioration,
contamination, or other adverse effects
7.5.11
820.150(a) pending use or distribution and to ensure ISO 13485:2016 does not specify a provision
Preservation of
Storage that no obsolete, rejected, or deteriorated for stock rotation.
product
product is used or distributed. When the
quality of product deteriorates over time,
it shall be stored in a manner to facilitate
proper stock rotation, and its condition shall
be assessed as appropriate.

Each manufacturer shall establish and


820.150(b) maintain procedures that describe the ISO 13485:2016 does not specify a
Storage methods for authorizing receipt from and comparable requirement.
dispatch to storage areas and stock rooms.
14
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
Each manufacturer shall establish and maintain 7.2.1
procedures for control and distribution of Determination
finished devices to ensure that only those of requirements
devices approved for release are distributed and related to product
that purchase orders are reviewed to ensure 21 CFR § 820 specifies explicit provisions
820.160(a) 7.2.2 Review of
that ambiguities and errors are resolved before relative to the prohibition of the distribution
Distribution requirements
devices are released for distribution. Where a of expired or deteriorated devices.
related to product
device’s fitness for use or quality deteriorates
over time, the procedures shall ensure that 7.5.1 Control of
expired devices or devices deteriorated beyond production and
acceptable fitness for use are not distributed. service provision

Each manufacturer shall maintain distribution


records which include or refer to the location of: ISO 13485:2016 specifies provisions
for maintaining distribution records for
(1) The name and address of the initial implantable medical devices.
7.5.9.2 Particular
consignee;
820.160(b) requirements 21 CFR § 820 specifies distribution records
Distribution (2) The identification and quantity of devices for implantable for all devices; and specifies specific
shipped; medical devices additional information to be included
(3) The date shipped; and (or referred to the location of) in the
distribution records.
(4) Any control number(s) used.

Each manufacturer of a device requiring


installation shall establish and maintain
adequate installation and inspection
instructions, and where appropriate test
procedures. Instructions and procedures
820.170(a) 7.5.3 Installation
shall include directions for ensuring proper No significant difference in requirements.
Installation activities
installation so that the device will perform as
intended after installation. The manufacturer
shall distribute the instructions and procedures
with the device or otherwise make them
available to the person(s) installing the device.

The person installing the device shall


ensure that the installation, inspection,
and any required testing are performed
820.170(b) 7.5.3 Installation
in accordance with the manufacturer's No significant difference in requirements.
Installation activities
instructions and procedures and shall
document the inspection and any test results
to demonstrate proper installation.

All records required by this part shall be


maintained at the manufacturing establishment
or other location that is reasonably accessible
to responsible officials of the manufacturer and 21 CFR § 820 requires records to be
to employees of FDA designated to perform readily available for review and copying
820.180 inspections. Such records, including those by FDA employees designated to perform
4.2.5 Control of inspections.
General not stored at the inspected establishment,
records
Requirements shall be made readily available for review and 21 CFR § 820 specifies records stored in
copying by FDA employee(s). Such records automated data processing systems must be
shall be legible and shall be stored to minimize backed up.
deterioration and to prevent loss. Those records
stored in automated data processing systems
shall be backed up.
15
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
21 CFR § 820 specifies the explicit
requirements allowing the manufacturer
to mark records as confidential to aid
FDA in determining publicly releasable
Records deemed confidential by the information under provisions of the
manufacturer may be marked to aid FDA Freedom of Information Act (FOIA).
820.180(a) 4.2.5 Control of
in determining whether information may
Confidentiality records ISO 13485:2016 specifies that the
be disclosed under the public information
regulation in part 20 of this chapter. organization shall document procedures
to define the controls needed for
(amongst other things) the security of
records; and the protection of confidential
health information.

All records required by this part shall be


retained for a period of time equivalent to 4.2.4 Control of
820.180(b) documents
the design and expected life of the device,
Record No significant difference in requirements.
but in no case less than 2 years from the 4.2.5 Control of
Retention Period
date of release for commercial distribution records
by the manufacturer.

This section does not apply to the reports


required by 820.20(c) Management
review, 820.22 Quality audits, and
supplier audit reports used to meet the
requirements of 820.50(a) Evaluation of
suppliers, contractors, and consultants,
but does apply to procedures established
under these provisions. Upon request
820.180(c) ISO 13485:2016 does not specify a
of a designated employee of FDA, an
Exceptions comparable requirement.
employee in management with executive
responsibility shall certify in writing that the
management reviews and quality audits
required under this part, and supplier audits
where applicable, have been performed and
documented, the dates on which they were
performed, and that any required corrective
action has been undertaken.

Each manufacturer shall maintain device


master records (DMR’s). Each manufacturer
shall ensure that each DMR is prepared and
approved in accordance with 820.40. The DMR
for each type of device shall include, or refer to
the location of, the following information:
(a) Device specifications including
appropriate drawings, composition,
formulation, component specifications,
and software specifications;
(b) Production process specifications
820.181 Device including the appropriate equipment 4.2.3 Medical
specifications, production methods, No significant difference in requirements.
Master Record device file
production procedures, and production
environment specifications;
(c) Quality assurance procedures and
specifications including acceptance
criteria and the quality assurance
equipment to be used;
(d) Packaging and labeling specifications,
including methods and processes
used; and
(e) Installation, maintenance, and
servicing procedures and methods.

16
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
Each manufacturer shall maintain device
history records (DHR’s). Each manufacturer
shall establish and maintain procedures to
ensure that DHR’s for each batch, lot, or unit
are maintained to demonstrate that the device 21 CFR § 820 specifies additional DHR
is manufactured in accordance with the DMR content beyond that specified in ISO
and the requirements of this part. The DHR 13485:2016. (e.g. (a) The dates of
shall include, or refer to the location of, the manufacture;...
following information:
(d) The acceptance records which
(a) The dates of manufacture; demonstrate the device is manufactured
7.5.1 Control of in accordance with the DMR;
820.184 Device (b) The quantity manufactured;
production and
History Record (e) The primary identification label and
(c) The quantity released for distribution; service provision
labeling used for each production unit;
(d) The acceptance records which demonstrate and
the device is manufactured in accordance
with the DMR; (f) Any unique device identifier (UDI) or
universal product code (UPC), and any
(e) The primary identification label and labeling other device identification(s) and control
used for each production unit; and number(s) used.)
(f) Any unique device identifier (UDI) or
universal product code (UPC), and any
other device identification(s) and control
number(s) used.

Each manufacturer shall maintain a quality


system record (QSR). The QSR shall include, or
refer to the location of, procedures and the While ISO 13485:2016 does not specify
documentation of activities required by this a requirement for a QSR, there are
820.186 Quality 4.2.5 Control of
part that are not specific to a particular type requirements for the various documents and
System Record records
of device(s), including, but not limited to, the procedures that are included (or referred to
records required by 820.20. Each manufacturer the location of) within the QSR.
shall ensure that the QSR is prepared and
approved in accordance with 820.40.

No significant difference in requirements.


However, the definition of “complaint”
Each manufacturer shall maintain complaint in 3.4 of ISO 13485:2016 includes the
files. Each manufacturer shall establish and term “usability” – not included in the
maintain procedures for receiving, reviewing, definition of the term contained in 21 CFR
and evaluating complaints by a formally § 820.3(c).
designated unit. Such procedures shall ensure The definition of “complaint” in 3.4 of
that: ISO 13485:2016 also includes concerns
(1) All complaints are processed in a uniform “..related to a service that affects the
820.198(a) and timely manner; 8.2.2 Complaint performance of such medical devices” as a
Complaint Files handling potential source of complaints.
(2) Oral complaints are documented upon
receipt; and ISO 13485:2016 defines the term
“complaint” as “Written, electronic or oral
(3) Complaints are evaluated to determine
communication that alleges deficiencies
whether the complaint represents an event
related to the identity, quality, durability,
which is required to be reported to FDA
reliability, usability, safety or performance
under part 803 of this chapter, Medical
of a medical device that has been released
Device Reporting.
from the organization’s control or related
to a service that affects the performance of
such medical devices.”

Each manufacturer shall review and evaluate


all complaints to determine whether
21 CFR § 820 specifies that when no
an investigation is necessary. When no
complaint investigation is made, the name
820.198(b) investigation is made, the manufacturer shall 8.2.2 Complaint
of the individual responsible for the decision
Complaint Files maintain a record that includes the reason no handling
not to investigate must be documented and
investigation was made and the name of the
maintained.
individual responsible for the decision not to
investigate.

17
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
Any complaint involving the possible failure of
ISO 13485:2016 does not specify that an
a device, labeling, or packaging to meet any of
investigation is not necessary if such an
820.198(c) its specifications shall be reviewed, evaluated, 8.2.2 Complaint
investigation has already been performed for
Complaint Files and investigated, unless such investigation has handling
a similar complaint and another investigation
already been performed for a similar complaint
is not necessary.
and another investigation is not necessary.

Any complaint that represents an event which


must be reported to FDA under part 803 of this
chapter shall be promptly reviewed, evaluated,
and investigated by a designated individual(s)
and shall be maintained in a separate portion of ISO 13485:2016 specifies regulatory
the complaint files or otherwise clearly identified. reporting requirements must be met, but
In addition to the information required by does not include the specific details of 21
820.198(e), records of investigation under this 8.2.3 Reporting
820.198(d) CFR § 820 relative to the information to
paragraph shall include a determination of: to regulatory
Complaint Files be documented and maintained; nor that
authorities
(1) Whether the device failed to meet complaints representing events that must be
specifications; reported under 21 CFR 803 be maintained in
a separate portion of the complaint file.
(2) Whether the device was being used for
treatment or diagnosis; and
(3) The relationship, if any, of the device to the
reported incident or adverse event.

When an investigation is made under this


section, a record of the investigation shall be
maintained by the formally designated unit
identified in paragraph (a) of this section. The
record of investigation shall include:
(1) The name of the device;
(2) The date the complaint was received;
8.2.2 Complaint Although ISO 13485:2016 requires complaint
(3) Any unique device identifier (UDI) or
handling handling records to be maintained it does not
820.198(e) universal product code (UPC), and any
provide the specific detail relative to records
Complaint Files other device identification(s) and control 7.2.3 of complaint investigations that is specified in
number(s) used; Communication 21 CFR § 820.
(4) The name, address, and phone number of
the complainant;
(5) The nature and details of the complaint;
(6) The dates and results of the investigation;
(7) Any corrective action taken; and
(8) Any reply to the complainant.

ISO 13485:2016 does not specify a


comparable requirement. Although ISO
When the manufacturer's formally designated 13485:2016 does not discuss where records
complaint unit is located at a site separate may be located, clause 4.2.5 requires "The
820.198(f) from the manufacturing establishment, the organization shall document procedures
Complaint Files investigated complaint(s) and the record(s) of to define the controls needed for the
investigation shall be reasonably accessible to the identification, storage, security and integrity,
manufacturing establishment. retrieval, retention time and disposition
of records ... Records shall remain legible,
readily identifiable, and retrievable."

ISO 13485:2016 does not specify a


If a manufacturer’s formally designated
comparable requirement. Although ISO
complaint unit is located outside of the United
13485:2016 does not discuss where
States, records required by this section shall be
records may be located, 4.2.5 requires “The
reasonably accessible in the United States at
820.198(g) organization shall document procedures
either:
Complaint Files to define the controls needed for the
(1) A location in the United States where the identification, storage, security and integrity,
manufacturer’s records are regularly kept; or retrieval, retention time and disposition
of records ... Records shall remain legible,
(2) The location of the initial distributor.
readily identifiable, and retrievable.”
18
21 CFR § 820 US FDA QUALITY SYSTEM REGULATION ISO 13485:2016 SPECIFIC DIFFERENCES
Where servicing is a specified requirement,
each manufacturer shall establish and
820.200(a) 7.5.4 Servicing
maintain instructions and procedures for No significant difference in requirements.
Servicing activities
performing and verifying that the servicing
meets the specified requirements.

7.5.4 Servicing
activities
Each manufacturer shall analyze service reports ISO 13485:2016’s requirement for use of
820.200(b) 8.1 Measurement,
with appropriate statistical methodology in appropriate statistical techniques is found
Servicing analysis, and
accordance with 820.100. in 8.1.
improvement,
General

ISO 13485:2016 requires analysis to


Each manufacturer who receives a service determine if service activities carried
report that represents an event which must out by the organization or its supplier
820.200(d) be reported to FDA under part 803 of this 7.5.4 Servicing represent a complaint.
Servicing chapter shall automatically consider the report activities 21 CFR § 820 provides a specific type of
a complaint and shall process it in accordance service event that must automatically be
with the requirements of 820.198. considered a complaint (i.e. service reports
relative to a 21 CFR 803 reportable events).

Service reports shall be documented and shall


include:
(1) The name of the device serviced;
(2) Any unique device identifier (UDI) or
universal product code (UPC), and any
820.200(e) other device identification(s) and control 7.5.4 Servicing 21 CFR § 820 specifies specific content
Servicing number(s) used; activities requirements relative to service reports.
(3) The date of service;
(4) The individual(s) servicing the device;
(5) The service performed; and
(6) The test and inspection data.

Where appropriate, each manufacturer 21 CFR § 820 requires documented


shall establish and maintain procedures for 8.1 Measurement, procedures for identifying valid statistical
820.250(a) techniques.
identifying valid statistical techniques required analysis and
Statistical
for establishing, controlling, and verifying improvement, ISO 13485:2016 requires the organization to
Techniques
the acceptability of process capability and General plan for and determine appropriate statistical
product characteristics. techniques and the extent of their use.

Sampling plans, when used, shall be written 7.5.6 Validation


of processes for 21 CFR § 820 specifies specific detail
and based on a valid statistical rationale. Each
production and with respect to sampling plans including
manufacturer shall establish and maintain
820.250(b) service provision establishing and maintaining procedures
procedures to ensure that sampling methods
Statistical to ensure sampling methods are adequate
are adequate for their intended use and 8.1 Measurement,
Techniques for their intended use and to ensure when
to ensure that when changes occur the analysis and changes occur the sampling plans are
sampling plans are reviewed. These activities improvement, reviewed.
shall be documented. General
19
NSF International
Partner locations

NSF International is a global public health organization


that operates in more than 170 countries, with worldwide
laboratory testing facilities, and expert resources across a wide
range of professional fields including health sciences, software,
food and beverages, sustainability and agriculture.

For additional information, please contact medicaldevices@nsf.org


or visit www.nsfmedicaldevices.org.

NSF INTERNATIONAL | MEDICAL DEVICES


2001 Pennsylvania Avenue NW, Suite 950 Washington, DC 20006 USA
T +1 202-822-1850 | E medicaldevices@nsf.org
www.nsf.org
LMD-776-0418