Академический Документы
Профессиональный Документы
Культура Документы
01057
Test preparation
Instrument-specific information
Table 1 shows all of the instruments that have the program for this test. The table also
shows sample cell and orientation requirements for reagent addition tests, such as
powder pillow or bulk reagent tests.
To use the table, select an instrument, then read across to find the applicable information
for this test.
Table 1 Instrument-specific information
Instrument Sample cell orientation Sample cell
DR 6000 The fill line is to the right. 2495402
DR 3800
DR 2800
DR 2700
DR 1900
DR 5000 The fill line is toward the user.
DR 3900
DR 900 The orientation mark is toward the user. 2401906
Before starting
Install the instrument cap on the DR 900 cell holder before ZERO or READ is pushed.
To make sure that all forms of the metal are measured, digest the sample with heat and acid. Use the mild or vigorous
digestion. Refer to the Water Analysis Guide for more information.
Rinse all glassware with a 1:1 (50%) nitric acid solution. Rinse again with deionized water.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective
equipment.
The alkaline cyanide solution contains cyanide. Make sure to read the Safety Data Sheets and obey the safety precautions.
Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal
information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.
1
Items to collect
Description Quantity
Start
1. Start Program 2. Prepare the blank: Fill 3. Prepare the sample: Fill 4. Add the contents of one
290 Manganese, LR PAN. the sample cell with 10 mL a second sample cell with Ascorbic Acid Powder Pillow
For information about of deionized water. 10 mL of sample. to each sample cell.
sample cells, adapters or Digestion is necessary for
light shields, refer to total manganese
Instrument-specific determination.
information on page 1.
9. Swirl to mix. 10. Start the instrument 11. When the timer expires, 12. Insert the blank into the
The sample will show an timer. A 2‑minute reaction clean the blank sample cell. cell holder.
orange color if manganese time starts.
is present.
Zero Read
13. Push ZERO. The 14. Clean the prepared 15. Insert the prepared 16. Push READ. Results
display shows 0.000 mg/L sample cell. sample into the cell holder. show in mg/L Mn.
Mn.
Interferences
Interfering substance Interference level
Aluminum 20 mg/L
Cadmium 10 mg/L
Calcium 1000 mg/L as CaCO3
Cobalt 20 mg/L
Copper 50 mg/L
Hardness For samples that contain more than 300 mg/L hardness as CaCO3, add 4 drops of Rochelle Salt
Solution to the sample after the Ascorbic Acid Powder Pillow is added.
Iron 25 mg/L (If the sample contains more than 5 mg/L iron, increase the reaction time to 10 minutes.)
Accuracy check
Standard additions method (sample spike)
Use the standard additions method (for applicable instruments) to validate the test
procedure, reagents and instrument and to find if there is an interference in the sample.
Items to collect:
• Manganese Standard Solution, 10 mg/L Mn, PourRite Ampule
• Ampule breaker
• Pipet, TenSette, 0.1–1.0 mL and tips
1. Use the test procedure to measure the concentration of the sample, then keep the
(unspiked) sample in the instrument.
2. Go to the Standard Additions option in the instrument menu.
3. Select the values for standard concentration, sample volume and spike volumes.
4. Open the standard solution.
5. Prepare three spiked samples: use the TenSette pipet to add 0.1 mL, 0.2 mL and
0.3 mL of the standard solution, respectively, to three 10-mL portions of fresh sample.
Mix well.
6. Use the test procedure to measure the concentration of each of the spiked samples.
Start with the smallest sample spike. Measure each of the spiked samples in the
instrument.
7. Select Graph to compare the expected results to the actual results.
Note: If the actual results are significantly different from the expected results, make sure that
the sample volumes and sample spikes are measured accurately. The sample volumes and
sample spikes that are used should agree with the selections in the standard additions menu. If
the results are not within acceptable limits, the sample may contain an interference.
Summary of method
The PAN method is a highly sensitive and rapid procedure to measure low levels of
manganese. The ascorbic acid reagent reduces all oxidized forms of manganese to Mn2+.
The alkaline-cyanide reagent masks potential interferences. PAN Indicator then forms an
orange-colored complex with Mn2+. The measurement wavelength is 560 nm.
Consumables and replacement items
Required reagents
© Hach Company/Hach Lange GmbH, 1989–2014, 2017. All rights reserved. 12/2017, Edition 10