HAND SYSTEM
For fragment-specific fracture fixation with
variable angle locking and locking technology
SURGICAL TECHNIQUE
TABLE OF CONTENTS
AO Principles 5
Indications6
Featured Instruments 20
Insert Screw 50
Implant Removal 51
Instruments63
Graphic Cases 70
Set Lists 77
System Snapshot
• Extensive system of anatomically precontoured plates
• First to the market with 1.3 mm locking screws for
hand plating1
• Forceps that aid in fracture reduction and lag screw
application
• Forceps that aid in plate fixation
• Self-retaining screwdrivers
• Plates available in 316L stainless steel and titanium
• Color-coded instruments
1
DePuy Synthes Companies market analysis of leading
orthopaedic companies, conducted May 2015.
SYSTEM TECHNOLOGY
Variable Angle Locking Technology
1.5 mm and 2.0 mm plates feature variable angle locking
coaxial holes. Four columns of threads in the variable angle
locking hole provide four points for threaded locking
between the plate and the Variable Angle Locking Screw
to create a fixed-angle construct at the desired screw angle.
x 1.5 mm variable angle plate holes accept:
– 1.5 mm variable angle locking screws
– 1.5 mm cortex screws
– 1.5 mm locking screws. These screws can only
be used on-axis
x 2.0 mm variable angle plate holes accept:
– 2.0 mm variable angle locking screws
– 2.0 mm cortex screws
– 2.0 mm locking screws. These screws can only be
used on-axis
x Variable Angle Locking Screws can be angled anywhere
within a 30° cone
x 1.5 mm and 2.0 mm cortex screws can be used in the
plate positioning slots, if present, for traditional
compression and fixation
Locking Technology
• 1.3 mm plates feature locking, coaxial holes. Fully
threaded locking holes provide threaded locking
between the plate and the locking screw to create
a fixed-angle construct
• 1.3 mm locking plate holes accept:
– 1.3 mm locking screws
– 1.3 mm cortex screws
SYSTEM FEATURES
FEATURES BENEFITS
Extensive range of implant options Accommodates a variety of fractures and
patient sizes
Anatomically contoured plates Designed with contour to adapt to bony
anatomy
Plates designed for lateral and direct Accommodates plate placement to help
dorsal application avoid tendon insertion points
Low-profile, anatomically precontoured Designed to reduce soft tissue irritation
plates
Recessed screw heads Designed to sit flush with the plate surface
Variable Angle Locking Technology Offers screw placement options in a variety
of fragment patterns and around the joint
Provides fixed angle stability for
metaphyseal and osteopenic bone
Elongated plate holes Facilitate plate positioning
Modular, color-coded system Aids in clear identification of instruments
for Operating Room staff and Central
Processing
Facilitates proper instrumentation usage
by size
Comprehensive instrumentation Aids surgeon in all aspects of procedure:
plate adaption, provisional fixation, and
bone preparation
Implants offered in both stainless steel Allows options for surgeon preference and
and titanium patient need
T4 and T6 self-retaining screwdriver Aid in screw handling, measuring, and
shafts delivery
Hex coupling screwdriver handle Ergonomic connection mechanism in a
handle design that disassembles for
cleaning
Reduction Forceps Aids reduction and lag screw application
via forceps
Plate Holding Forceps Holds plate intraoperatively
Anatomicreduction
Anatomic reduction Stable
Stablefixation
fixation
Fracture reduction and fixation
Fracture reduction and fixation to
to Fracture
Fracturefixation
fixationproviding
providingabsolute
abso-
restore anatomical relationships.
restore anatomical relationships. or relative stability, as required
lute or relative stability, as by
the patient, the injury, and the
required by the patient, the injury,
1 2 personality of the fracture.
and the personality of the
fracture.
Early,
Early,active
active mobilization Preservationof ofblood
bloodsupply
supply
Early
Early and
and safe
mobilization
safe mobilization
mobilization andand 4 3 Preservation
Preservationofofthe
Preservation theblood
bloodsupply
supply
rehabilitation
rehabilitation of the injured part
of the injured part to soft tissues and bone by
to soft tissues and bone by gentle
and
andthe
the patient
patient as
as aa whole.
whole. gentle reduction
reduction techniques
techniques and and
careful handling.
careful handling.
MR SAFETY INFORMATION
These devices have not been evaluated for safety and
compatibility in the MR environment. The devices have not
been tested for heating or migration in the MR environment.
Right
Surgical Approach
Ulnar Radial
Right Hand
Left Hand
11 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Featured Plates & Technique Highlights
11 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Featured Plates & Technique Highlights
11 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Featured Plates & Technique Highlights
11 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Featured Plates & Technique Highlights
1.3 mm Screws
1.3 mm Locking Screws
• Only for on-axis insertion in the locking holes
• Threaded, conical head locks securely into the
locking hole
• Self-retaining,T4 STARDRIVE Recess
• Self-tapping tip facilitates insertion
1.5 mm Screws
1.5 mm Variable Angle Locking Screws
• Threaded, rounded head locks securely into the
variable angle locking hole
• Purple, color-coded head to easily distinguish as
Variable Angle Locking Screw
• Self-retaining, T4 STARDRIVE Recess
• Blunt self-tapping tip
Also Available
• 1.5 mm Locking Screws only used on-axis –
Compatible with 1.5 mm Variable Angle Locking Plates
11 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Screws in the System
2.0 mm Screws
2.0 mm Variable Angle Locking Screws
• Threaded, rounded conical head locks securely into
the locking hole
• Purple, color-coded head to easily distinguish as a
Variable Angle Locking Screw
• Self-retaining, T6 STARDRIVE Recess
• Blunt, self-tapping tip
Also Available
• 2.0 mm Locking Screws only used on-axis –
Compatible with 2.0 mm Variable Angle Locking Plates
22 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Featured Instruments
Alternative Instrument
22 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Featured Instruments
22 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Featured Instruments
Variable angle technique using the cone-shaped end. Nominal angle technique using the fixed-angle
end marked “coaxial.”
Variable angle technique using the variable Nominal angle technique using the fixed-angle
angle end, which allows freehand angulation end marked “coaxial.”
of the drill bit.
22 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
PREOPERATIVE PLANNING AND REDUCTION
1
Preparation
Complete preoperative radiographic assessment and
preoperative planning.
2
Reduce Fracture
Instruments
Reduction Forceps
319.39.96 Sharp Hook
Optional Instrument
22 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
LAG SCREW INSERTION (OPTIONAL)
Lag Screw Insertion – Reduction Forceps
1
Select a Drill Sleeve and Drill Bit
Lag screws may be inserted using the Reduction Forceps
(03.130.291). The associated drill sleeves and drill bits use
a color-coded stripe system. The colors on the drill sleeves
and drill bits correspond to the screw size. A single stripe
indicates that the drill sleeve or drill bit is used for a
threaded hole; a double stripe indicates that the drill
sleeve or drill bit is used for a gliding hole.
Screw Size Color Coding Drill Sleeve Drill Bit for Drill Sleeve Drill Bit for
for for Threaded Hole for Gliding Hole
Instruments Threaded Hole (One Color Stripe)* Gliding Hole (Two Color Stripes)*
1.3 mm Yellow
1.5 mm Red
2.0 mm Blue
*Drill bits are available with various couplings, including Mini Quick Coupling, K-wire,
and J-Latch.
2
Assemble Drill Sleeve to Forceps
Press the drill sleeve into the barrel of the Reduction
Forceps so that the sleeve is fully seated and a click
is felt.
3
Attach the Reduction Forceps to the Bone
Once in place, close the Reduction Forceps to the desired
tightness. The forceps is designed to remain in place after
the handles are released.
33 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Lag Screw Insertion (Optional)
Lag Screw Insertion—Reduction Forceps
4 Gliding Hole
Drill Threaded and Gliding Hole
Threaded Hole
Gliding Hole: Using the drill sleeve, drill the near cortex,
so that the thread of the screw does not obtain purchase.
5
Countersink (Optional)
Remove the Drill Sleeve from the forceps. Attach the
Countersink (03.130.215) to the Screwdriver Handle.
Countersink, if desired.
6
Measure
Insert the Depth Gauge (03.130.250) through the barrel
of the forceps, to determine the screw length needed.
The Depth Gauge is designed for one-handed use.
33 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Lag Screw Insertion (Optional)
Lag Screw Insertion—Reduction Forceps
7
Insert Cortex Screw
Instruments
33 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
LAG SCREW INSERTION — DOUBLE DRILL SLEEVE
(OPTIONAL)
Instruments
For 1.3 mm, 1.5 mm, and 2.0 mm cortex screws used as
independent lag screws, use a double-ended drill sleeve
and corresponding drill bit. Drill bits and drill guides are
color-coded. Drill bits are available with several different
couplings (see item listing). Etch lines on drill bits are
spaced 5 mm apart.
Screw Size Color Coding Double Drill Sleeve Drill Bit For Drill Bit for
for Threaded Hole Gliding Hole
Instruments (One Color Stripe)* (Two Color Stripes)*
1.3 mm Yellow
03.130.125 1.0 mm 1.3 mm
1.5 mm Red
03.130.225 1.1 mm 1.5 mm
2.0 mm Blue
03.130.325 1.5 mm 2.0 mm
*Drill bits are available with various couplings, including Mini Quick Coupling,
K-wire, and J-Latch.
Instruments
33 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
PREPARE AND INSERT PLATE
1
Trim Plate
Instrument
Optional Instrument
Cut portion
03.130.270 Plate Cutter, In-line for 1.3 mm Locking, of plate
1.5 mm VA Locking, and 2.0 mm
VA Locking Plates
33 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Prepare and Insert Plate
2
Contour Plate (Optional)
If necessary, contour the plate to fit the patient’s anatomy.
Instruments
44 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Prepare and Insert Plate
3
Position Plate
Instruments
Optional Instrument
44 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Prepare and Insert Plate
4
Determine Screw Type
Depending on the individual case, cortex screws or
Variable Angle Locking/locking Screws may be inserted.
Locking or cortex screws can be used in any of the screw
holes, with the exception of the nonthreaded elongated
hole, if present.
5
Drill
Instruments
Select a drill bit based on the size of the screw. Drill bits Screw Size Drill Bit Size/Color Code
are color-coded to the size of the associated screw.
1.3 mm 1.0 mm - Yellow
Note: The fracture pattern will dictate the optimal
screw placement.
1.5 mm 1.1 mm - Red
Fracture pattern will determine the order of screw insertion.
In all cases, make sure the plate is in the correct position
and flush with the bone. If placing a plate on the condyle, 2.0 mm 1.5 mm - Blue
be cognizant of reducing and restoring the articular surface.
44 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Prepare and Insert Plate
Variable Angle Locking Double Drill Guide—with If necessary, drill at a different angle and verify again under
Cone (03.130.220, 03.130.320) imaging.
Variable Angle Locking Screws can be inserted using Precaution: Avoid redrilling, especially in
2 different techniques, depending on which end of osteopenic bone.
the drill guide is used:
Variable angle technique using the cone-shaped end. Nominal angle technique using the fixed-angle
end marked “coaxial.”
Fully insert the variable angle end of the drill guide into Fully insert the coaxial end of the drill guide into the
the Variable Angle Locking hole. When drilling, the tip of Variable Angle Locking hole and drill.
the drill guide should remain fully seated in the hole
Drill the hole with the drill bit at the desired angle within
the cone by positioning the drill bit.
Drill for Variable Angle Screws—Freehand Verify the drill bit angle and depth under radiographic
imaging to ensure the desired angle has been achieved.
Variable Angle Locking Double Drill Guide—
Freehand (03.130.221, 03.130.321)
If necessary, drill at a different angle and verify again under
imaging. Avoid redrilling, especially in osteopenic bone.
Variable Angle Locking Screws can be inserted using
2 different techniques, depending on which end of the
drill guide is used:
Variable angle technique using the variable angle. Nominal angle technique using the fixed-angle
end marked “coaxial.”
Fully insert the variable angle end of the drill guide into Fully insert the coaxial end of the drill guide into the
the Variable Angle Locking hole. Variable Angle Locking hole and drill.
Drill the hole with the drill bit at the desired angle by
positioning the drill guide in the hole.
44 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Prepare and Insert Plate
7
Determine Screw Length
Use the Depth Gauge (03.130.250) to determine screw
length. Screw length is indicated by the black line.
See pages 32–33 for additional technique information for the Depth Gauge.
44 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Prepare and Insert Plate
8
Insert Screw
Instruments
Postoperative Treatment
Postoperative treatment with locking plates or variable
angle locking plates does not differ from conventional
internal fixation procedures.
55 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
IMPLANT REMOVAL
1
Preoperative Planning
To ensure that the appropriate screw removal instruments
are obtained, the surgeon should have the following
information before implant removal:
x Implant type
x Time of implantation
x Material (steel, titanium)
x Recess geometry and dimension of screws
x Any visible damage to the implant (eg, broken
screw shaft)
2
Clean Recess
Instrument
3
Implant Removal
Instruments
Optional Set
55 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
SUMMARY TABLE INSTRUMENTS
Screw Size
Gliding Hole
Gliding Hole
(Two Stripes) 1.0 mm 1.3 mm 1.5 mm 2.0 mm
Threaded Hole
Threaded Hole
Drill Bit (One Stripe) 0.76 mm 1.0 mm 1.1 mm 1.5 mm
Double Drill
Guide for 03.130.125 03.130.225 03.130.325
n/a
Drill Guide Cortex Screws (1.3 mm/1.0 mm) (1.5 mm/1.1 mm) (2.0 mm/1.5 mm)
1.0 mm Cortex 1.3 mm Locking or 1.5 mm Variable Angle 2.0 mm Variable Angle
Cortex Screws Locking or Cortex Screws Locking or Cortex Screws
n/a
Variable Angle
Double Drill
Guide— 03.130.220 03.130.320
with Cone n/a (1.1 mm) (1.5 mm)
Variable Angle
Drill Guide Double Drill
Guide— 03.130.221 03.130.321
Freehand (1.1 mm) (1.5 mm)
n/a n/a
Threaded
Drill Guide 03.130.120
Cruciform
Screwdriver 314.482 T4 STARDRIVE - 03.130.010 T6 STARDRIVE - 03.130.020
55 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Implants
Stainless
Steel Titanium
02.130.262 04.130.262 1.5 mm Variable Angle Locking
Box Plate, 4 Holes
02.130.263 04.130.263
1.5 mm Variable Angle Locking
First Metacarpal Plate,
Dorsal
02.130.264 04.130.264
1.5 mm Variable Angle Locking
First Metacarpal Plate,
Lateral, Right
02.130.265 04.130.265
1.5 mm Variable Angle Locking
First Metacarpal Plate,
Lateral, Left
02.130.266 04.130.266
1.5 mm Variable Angle Locking
Rotation Correction Plate,
6 Holes
02.130.267 04.130.267
1.5 mm Variable Angle Locking
Fracture Plate, 6 Holes
02.130.268 04.130.268
1.5 mm Variable Angle Locking
Metacarpal Neck Plate
55 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Implants
55 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Screws
*Ti-6Al-7Nb
*Ti-6Al-7Nb
66 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Screws
*Ti-6Al-7Nb
66 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
INSTRUMENTS
66 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Instruments
66 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Instruments
In Stainless Sets
292.06 0.6 mm Kirschner Wire w/Trocar Point
70 mm, 10 each
292.08 0.8 mm Kirschner Wire w/Trocar Point
70 mm, 10 each
292.10 1.0 mm Kirschner Wire w/Trocar Point
150 mm, 10 each
292.12 1.25 mm Kirschner Wire w/Trocar Point
150 mm, 10 each
In Titanium Sets
492.08 0.8 mm Ti Kirschner Wire w/Trocar Point
70 mm, 10 each
492.10 1.0 mm Ti Kirschner Wire w/Trocar Point
150 mm, 10 each
492.12 1.25 mm Ti Kirschner Wire w/Trocar Point
150 mm, 10 each
66 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Instruments
Graphic Cases*
60.130.000 1/3 Instrument Tray, Standard Instruments,
for the Variable Angle Locking Hand System
77 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Variable Angle Locking Hand System
Graphic Cases
60.130.013
60.130.213
60.130.313
60.130.015
60.130.215
60.130.315
77 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Variable Angle Locking Hand System
Graphic Cases
60.130.020
60.130.220
60.130.320
60.130.005 Drill Insert Module for Reduction Forceps
03.130.291
77 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Also Available
77 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
VA LOCKING HAND INSTRUMENT AND IMPLANT SET—STANDARD
STAINLESS STEEL (01.131.100) AND TITANIUM (01.131.101)
VA LOCKING HAND INSTRUMENT AND IMPLANT SET—COMPACT
STAINLESS STEEL (01.130.225) AND TITANIUM (01.130.235)
Graphic Cases
60.130.000 1/3 Instrument Tray, Standard Instruments,
for the Variable Angle Locking Hand System
Item Number For detailed cleaning and sterilization instructions, please refer to
60.116.451 Blank www.synthes.com/cleaning-sterilization or sterilization instructions, if provided.
Standard Sets
01.131.100 VA Locking Hand Instrument and
Implant Set—Standard
01.131.101 VA Titanium Locking Hand Instrument
and Implant Set—Standard
77 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
VA Locking Hand Instrument and Implant Set—Compact
Stainless Steel (01.130.225) and Titanium (01.130.235)
Compact Sets
01.130.225 VA Locking Hand Instrument and
Implant Set—Compact
01.130.235 VA Titanium Locking Hand Instrument
and Implant Set—Compact
88 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
VA Locking Hand Instrument and Implant Set—Standard
Stainless Steel (01.131.100) and Titanium (01.131.101)
VA Locking Hand Instrument and Implant Set—Compact
Stainless Steel (01.130.225) and Titanium (01.130.235)
In Titanium Sets
492.08 0.8 mm Ti Kirschner Wire w/Trocar Point
70 mm, 10 each
492.10 1.0 mm Ti Kirschner Wire w/Trocar Point
150 mm, 10 each
492.12 1.25 mm Ti Kirschner Wire w/Trocar Point
150 mm, 10 each
Implants
1.3 mm Locking Plates – 316 Stainless Steel or 1.5 mm Variable Angle Locking Plates –
TiCP Titanium 316L Stainless Steel or TiCP Titanium
Stainless Stainless
Steel Titanium Steel Titanium
02.130.150 04.130.150 1.3 mm Locking Straight 02.130.250 04.130.250 1.5 mm Variable Angle Locking
Plate, 6 Holes* Straight Plate, 6 Holes*
02.130.151 04.130.151 1.3 mm Locking Straight 02.130.251 04.130.251 1.5 mm Variable Angle Locking
Plate,12 Holes Straight Plate, 12 Holes
02.130.152 04.130.152 1.3 mm Locking T-Plate, 02.130.252 04.130.252 1.5 mm Variable Angle Locking
3 Holes Head, 5 Holes Shaft T-Plate, 3 Holes Head,
7 Holes Shaft
02.130.153 04.130.153 1.3 mm Locking Y-Plate,
3 Holes Head, 5 Holes Shaft 02.130.253 04.130.253 1.5 mm Variable Angle Locking
Y-Plate, 3 Holes Head,
02.130.154 04.130.154 1.3 mm Locking
7 Holes Shaft
Phalangeal Base Plate,
2 Holes Head, 5 Holes Shaft 02.130.254 04.130.254 1.5 mm Variable Angle Locking
Phalangeal Base Plate,
02.130.156 04.130.156 1.3 mm Locking Phalangeal
2 Holes Head, 6 Holes Shaft
Head Plate, Right
02.130.255 04.130.255 1.5 mm Variable Angle Locking
02.130.157 04.130.157 1.3 mm Locking Phalangeal
Condylar Plate, 2 Holes Head,
Head Plate, Left
6 Holes Shaft
02.130.158 04.130.158 1.3 mm Locking Strut Plate,
02.130.256 04.130.256 1.5 mm Variable Angle Locking
8 Holes, Oblique, Right
Phalangeal Head Plate, Right
02.130.159 04.130.159 1.3 mm Locking Strut Plate,
02.130.257 04.130.257 1.5 mm Variable Angle Locking
8 Holes, Oblique, Left
Phalangeal Head Plate, Left
02.130.161 04.130.161 1.3 mm Locking Web Plate,
02.130.258 04.130.258 1.5 mm Variable Angle Locking
14 Holes*
Strut Plate, 8 Holes,
Oblique, Right
88 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
VA Locking Hand Instrument and Implant Set—Standard
Stainless Steel (01.131.100) and Titanium (01.131.101)
VA Locking Hand Instrument and Implant Set—Compact
Stainless Steel (01.130.225) and Titanium (01.130.235)
1.5 mm Variable Angle Locking Plates – 2.0 mm Variable Angle Locking Plates, 316L
316L Stainless Steel or TiCP Titanium (continued) Stainless Steel or TiCP Titanium
Stainless Stainless
Steel Titanium Steel Titanium
02.130.263 04.130.263
1.5 mm Variable Angle Locking 02.130.350 04.130.350 2.0 mm Variable Angle Locking
First Metacarpal Plate, Dorsal Straight Plate, 6 Holes*
02.130.264 04.130.264
1.5 mm Variable Angle Locking 02.130.351 04.130.351 2.0 mm Variable Angle
First Metacarpal Plate, Locking Straight Plate,
Lateral, Right* 12 Holes
02.130.265 04.130.265
1.5 mm Variable Angle Locking 02.130.352 04.130.352 2.0 mm Variable Angle
First Metacarpal Plate, Locking T-Plate,
Lateral, Left* 3 Holes Head, 7 Holes Shaft
02.130.266 04.130.266
1.5 mm Variable Angle Locking 02.130.353 04.130.353 2.0 mm Variable Angle
Rotation Correction Plate, Locking Y-Plate,
6 Holes* 3 Holes Head, 7 Holes Shaft
02.130.268 04.130.268
1.5 mm Variable Angle Locking 02.130.354 04.130.354 2.0 mm Variable Angle
Metacarpal Neck Plate Locking Phalangeal Base Plate,
2 Holes Head, 6 Holes Shaft
1.0 mm Cortex Screws, Self-Tapping, Cruciform, 2 each 1.3 mm Locking Screws, Self-Tapping, T4 STARDRIVE Recess,
316L Stainless Steel or TiCP Titanium 4 each
316L Stainless Steel or Titanium*
Stainless Steel Titanium Length (mm)
200.526 400.526 6 Stainless Steel Titanium Length (mm)
200.527 400.527 7 02.130.104 04.130.104 4
200.528 400.528 8 02.130.105 04.130.105 5
200.529 400.529 9 02.130.106 04.130.106 6
200.530 400.530 10 02.130.107 04.130.107 7
200.531 400.531 11 02.130.108 04.130.108 8
200.532 400.532 12 02.130.109 04.130.109 9
200.533 400.533 13 02.130.110 04.130.110 10
200.534 400.534 14 02.130.111 04.130.111 11
02.130.112 04.130.112 12
1.3 mm Cortex Screws, Self-Tapping, T4 STARDRIVE Recess, 02.130.113 04.130.113 13
4 each 02.130.114 04.130.114 14
316L Stainless Steel or Titanium* 02.130.115 04.130.115 15
Stainless Steel Titanium Length (mm) 02.130.116 04.130.116 16
02.130.004 04.130.004 4 02.130.118 04.130.118 18
02.130.005 04.130.005 5
02.130.006 04.130.006 6 1.5 mm Cortex Screws, Self-Tapping, T4 STARDRIVE Recess,
316L Stainless Steel or Titanium*
02.130.007 04.130.007 7
02.130.008 04.130.008 8 Stainless Steel Titanium Length (mm) Qty
02.130.009 04.130.009 9 02.214.104 04.214.104 4 2
02.130.010 04.130.010 10 02.214.105 04.214.105 5 2
02.130.011 04.130.011 11 02.214.106 04.214.106 6 4
02.130.012 04.130.012 12 02.214.107 04.214.107 7 4
02.130.013 04.130.013 13 02.214.108 04.214.108 8 6
02.130.014 04.130.014 14 02.214.109 04.214.109 9 6
02.130.015 04.130.015 15 02.214.110 04.214.110 10 6
02.130.016 04.130.016 16 02.214.111 04.214.111 11 6
02.130.018 04.130.018 18 02.214.112 04.214.112 12 6
02.214.113 04.214.113 13 6
02.214.114 04.214.114 14 6
02.214.115 04.214.115 15 4
02.214.116 04.214.116 16 4
02.214.118 04.214.118 18 4
02.214.120 04.214.120 20 4
02.214.122 04.214.122 22 4
02.214.124 04.214.124 24 4
*Ti-6Al-7Nb
88 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
VA Locking Hand Instrument and Implant Set—Standard
Stainless Steel (01.131.100) and Titanium (01.131.101)
VA Locking Hand Instrument and Implant Set—Compact
Stainless Steel (01.130.225) and Titanium (01.130.235)
1.5 mm Variable Angle Locking Screws, Self-Tapping, 2.0 mm Cortex Screws, Self-Tapping, T6 STARDRIVE Recess
T4 STARDRIVE Recess 316L Stainless Steel or Titanium*
316L Stainless Steel or Titanium*
Stainless Steel Titanium Length (mm) Qty
Stainless Steel Titanium Length (mm) Qty 201.356.97 401.356.97 6 4
02.130.204 04.130.204 4 2 201.357.97 401.357.97 7 4
02.130.205 04.130.205 5 2 201.358.97 401.358.97 8 4
02.130.206 04.130.206 6 4 201.359.97 401.359.97 9 4
02.130.207 04.130.207 7 4 201.360.97 401.360.97 10 6
02.130.208 04.130.208 8 6 201.361.97 401.361.97 11 6
02.130.209 04.130.209 9 6 201.362.97 401.362.97 12 6
02.130.210 04.130.210 10 6 201.363.97 401.363.97 13 6
02.130.211 04.130.211 11 6 201.364.97 401.364.97 14 6
02.130.212 04.130.212 12 6 201.365.97 401.365.97 15 6
02.130.213 04.130.213 13 6 201.366.97 401.366.97 16 4
02.130.214 04.130.214 14 6 201.368.97 401.368.97 18 4
02.130.215 04.130.215 15 4 201.370.97 401.370.97 20 4
02.130.216 04.130.216 16 4 201.372.97 401.372.97 22 4
02.130.218 04.130.218 18 4 201.374.97 401.374.97 24 4
02.130.220 04.130.220 20 4
02.130.222 04.130.222 22 4 2.0 mm Variable Angle Locking Screws, Self-Tapping,
02.130.224 04.130.224 24 4 T6 STARDRIVE Recess
316L Stainless Steel and Titanium*
Stainless Steel Titanium Length (mm) Qty
02.130.306 04.130.306 6 4
02.130.307 04.130.307 7 4
02.130.308 04.130.308 8 4
02.130.309 04.130.309 9 4
02.130.310 04.130.310 10 6
02.130.311 04.130.311 11 6
02.130.312 04.130.312 12 6
02.130.313 04.130.313 13 6
02.130.314 04.130.314 14 6
02.130.315 04.130.315 15 6
02.130.316 04.130.316 16 4
02.130.318 04.130.318 18 4
02.130.320 04.130.320 20 4
02.130.322 04.130.322 22 4
02.130.324 04.130.324 24 4
*Ti-6Al-7Nb
Sets
01.131.013 1.3 mm Locking Hand Instrument and Implant
Module Set—Standard
01.131.113 1.3 mm Titanium Locking Hand Instrument
and Implant Module Set—Standard
01.131.015 1.5 mm VA Locking Hand Instrument and
Implant Module Set—Standard
01.131.115 1.5 mm Titanium VA Locking Hand
Instrument and Implant Module Set—Standard
01.131.020 2.0 mm VA Locking Hand Instrument and
Implant Module Set—Standard
01.131.120 2.0 mm Titanium VA Locking Hand Instrument
and Implant Module Set—Standard
88 DePuy Synthes Companies Variable Angle Locking Hand System Surgical Technique
Limited Warranty and Disclaimer: DePuy Synthes Trauma products are sold with a limited warranty to the original purchaser against defects in
workmanship and materials. Any other express or implied warranties, including warranties of merchantability or fitness, are hereby disclaimed.
WARNING: In the USA, this product has labeling limitations. See package insert for complete information.
CAUTION: Federal Law restricts these devices to sale by or on the order of a physician.
CAUTION:Some devices listed in this technique guide may not have been licensed in accordance with Canadian law and may not be for sale in Canada.
Please contact your sales consultant for items approved for sale in Canada.
www.depuysynthes.com