Академический Документы
Профессиональный Документы
Культура Документы
Purpose
This technical note demonstrates how a synchronized network of three or more Gocator sensors
mounted side by side to scan a wide and generally flat target surface can be system calibrated in order
for all sensors to report data in a common coordinate frame.
Equipment Used
Any Gocator model sensor (but must be same model within system)
Master 400 (or higher)
Table of Contents
1 Introduction ....................................................................................................................................................... 2
2 Gocator Coordinate Definition ........................................................................................................................... 3
3 Calibration Target ............................................................................................................................................. 3
4 Calibration Processing Flow .............................................................................................................................. 4
4.1 Input to Calibration Process ........................................................................................................................ 4
4.2 Sensor Configuration .................................................................................................................................. 5
4.3 Processing Steps ....................................................................................................................................... 5
5 Implementation using the Gocator SDK ............................................................................................................. 6
Gocator-2000-2300_Appnote_Multi-Sensor-Alignment-Calibration.doc
Application Note: Gocator – Multi-sensor Alignment Calibration Page 2 ( 6)
1 Introduction
This technical note explains how several Gocator sensors mounted side by side to scan a wide and generally flat
target surface can be system calibrated in order for all sensors to report data in a common coordinate frame. A
calibration target in the form of a bar with reference holes is positioned in the sensor system allowing for a calibration
procedure to calculate translation parameters for each sensor. In this case, each sensor will independently establish
three parameters, a rotational tilt compensation parameter, an X position offset based on which hole a particular
sensor is calibrated against, and a Z offset based on the distance to the calibration bar. During run-time these
parameters are then applied to the data points to create a single 3D model of the scanned surface.
The Gocator’s built-in Buddy concept can handle calibration of dual sensor configurations, but when creating systems
with three or more sensors, the user has to work with the Gocator Software Development Kit (SDK) to build a custom
solution. A common requirement with multi sensor systems is to precisely control the timing of each sensor to avoid
interference in areas where the measurement zones overlap, often referred to as “cross talk” between sensors. In the
type of sensor mounting layout pictured in Figure 1, there is typically interference between neighbouring sensors so
time multiplexing is necessary.
This technical note is bundled together with an open source example code project that the user can study, compile
and run in parallel to reading it.
This document assumes that the reader is already familiar with the Gocator SDK and how to multiplex timing in a
multi-sensor system. It is strongly recommended to first study the example code provided with the Gocator SDK
(called the Go2Samples), and then read the technical note on multi-sensor operation, which can be downloaded at:
http://www.lmi3d.com/product/gocator-2000-family/support/files/all
Gocator-2000-2300_Appnote_Multi-Sensor-Alignment-Calibration.doc
Application Note: Gocator – Multi-sensor Alignment Calibration Page 3 ( 6)
Sensor Coordinates
Prior to any custom system calibration applied by the user, individual
sensors use the coordinate system shown here. The Z-axis represents the
sensor’s measurement range (MR), with the values increasing towards the
sensor. The X-axis represents the sensor’s field of view (FOV). The origin is
at the center of the MR and FOV. The positive direction of the X-axis is to the
right when viewing the connector side of the sensor.
System Coordinates
In the general case, any arbitrary coordinate system definition can be defined
and implemented by the user. However, the Gocator has built-in support to
transform each data point with an X offset, Z offset and Angle (rotation in the X-Z
plane). Taking advantage of this support, the system coordinates are aligned
such that the system X-axis is parallel to the calibration target surface. The
system Z-origin is set to the base of the calibration target object. The system
X-origin is user defined and given by the X-positions of the reference holes.
3 Calibration Target
For wide, multi-sensor systems, a calibration bar is required to match the length of the system by following the
guidelines listed below (LMI Technologies does not manufacture or sell calibration bars).
Gocator-2000-2300_Appnote_Multi-Sensor-Alignment-Calibration.doc
Application Note: Gocator – Multi-sensor Alignment Calibration Page 4 ( 6)
Note that the position of the X-origin is arbitrary and user defined. However, to be consistent with the Gocator’s
internal coordinate definition, it is recommended to define the positive X-direction to be from left to right when viewing
the system from the connector side of the sensors. In the example illustrated in Figure 2, the X-origin is somewhere in
the centre of the system, meaning that reference holes to the left of the origin have negative values.
Gocator-2000-2300_Appnote_Multi-Sensor-Alignment-Calibration.doc
Application Note: Gocator – Multi-sensor Alignment Calibration Page 5 ( 6)
Active Area – This is a fundamental configuration that refers to a region of interest within the sensor’s
maximum field of view that defines the area used for laser profiling. A smaller Active Area may allow the
sensor to run faster. Changing the Active Area affects the sensors internal coordinate frame. Therefore it is
not possible to modify this setting after the system calibration has been established.
Exposure – The sensor should be configured for Single Exposure, which is static over time. Do not use
Dynamic Exposure, since the first few frames may not have good data. Although the exposure time can be
changed after the system calibration, it is still strongly recommended to establish this setting prior. It is critical
for the calibration routine that the data collected from the calibration target is not under- or over-exposed. If
the measurement of the target suffers from data drop-out, this can be mistaken for the reference hole. Any
noise or outliers due to over-exposure or reflections may skew the angle correction and cause errors in the Z
offset.
Gocator-2000-2300_Appnote_Multi-Sensor-Alignment-Calibration.doc
Application Note: Gocator – Multi-sensor Alignment Calibration Page 6 ( 6)
In an effort to keep this example code short and easy to review, a number of assumptions and simplifications have
been made.
The sensors all have to be pre-configured using the web UI using the following rules:
o All sensors have to run at the same speed (use the same frame period)
o All sensors have to use the same exposure time
o The exposure time has to be shorter than half the frame period (for multiplexing to work)
The calibration parameters and the sensor layout information is hardcoded in the header file App.h. This is to
avoid having to add code to parse a file format like xml. Before attempting to run the code, the user will have
to edit the following defines and declarations to reflect their particular setup.
Gocator-2000-2300_Appnote_Multi-Sensor-Alignment-Calibration.doc