Вы находитесь на странице: 1из 4

RISH Max -10

Digital multimeters

Data Sheet
RISH Max - 10
Digital Multimeter

www.rishabh.co.in Page 1 of 4 Version:1/05/VIII/1211/B


Application Overload warning
RISH Max-10 digital multimeters are suited for universal, general An acoustic signal occurs when measuring AC voltage>750V, DC
applications in the electrical and electronics fields, as well as in Voltage>1000V, AC/DC mA current>400.0mA, AC/DC
radio and television service, training and education. current>10.00A.
They are of especially flat design, and thus fit into any bag. The Energy saving circuit
protective cover, which is provided as standard equipment, can be The instrument is switched off automatically, if none of the operating
opened at an angle for convenient reading from the workbench, and elements have been activated for about 15 minutes.
provides for easy transport.
Protective cover for rough operating conditions
Product Features A protective cover of ABS with a built-in stand protects the
instrument against jolts and falls. It also secures the test probe for
Hold
one-hand operation, and allows for winding of the measurement
By pressing the HOLD/ON key, the currently displayed cable which provides protection during transport.
measurement value can be held and “HOLD” is simultaneously
displayed. Calibration
Rish Max -10 multimeters are calibrated using precision calibrators
Relative measurement (REL) having accuracy better than at least 5 to 10 times depends upon the
By pressing the REL key, the zero correction is made and Relative functions and ranges. These sources are calibrated at regular
Value is measured. All functions can measure Relative Value intervals.
except Hz/Duty, Diode, Continuity and 0C functions.
Theft protection
Automatic/manual measuring range selection Company name and name of the user can be entered into the field
The measurement function are chosen with the rotary selector next to the display with an indelible etching needle for identification
switch. The measuring range is automatically adjusted to the of the owner.
measurement value. The measuring range can also be manually
selected with the AUTO/MAN button.
n
Direct and alternating voltages from 100m V ... 1000V
0
Note: For Temperature ( C ), Frequency (Hz), Duty cycle (%) and
n
Direct and alternating currents from 10m A ... 10.00A
Capacitance (F) measuring range is AUTO. No Manual range n
Resistances from 100mW ... 40.00MW with zero correction
selection is possible. n
Capacitance from 1pF ... 200.00m F with zero correction .
n
Frequencies from 10.00Hz ... 500.0kHz
Temperature Measurement n
Diode measurement and continuity testing
Rish Max - 10 allows you to measure temperature with “K” type n
Hold measurement .
0
Thermocouple (NiCr - Ni) sensor in the range from 0C0 to +1300 C. n
Relative measurement
n
Duty cycle (%) measurement
Diode and continuity testing n
Temperature measurement with K type Thermocouple
This provides for the testing of the polarity of diodes, as well as
inspection for short -circuits and circuit interruptions. In addition to
the display, resistance of less than 100 W (approx.) are indicated
with an acoustic signal.

Characteristic values for Rish Max - 10


1)
Meas. Measuring Input Digital display inherent deviation at Overload capacity
Resolution
Function Range Impedance reference condition
V(AC)/V(DC) +(...%rdg + ...digits) Overload value Overload Duration
400.0mV 100m
V >20MW 0.75+2
4.000V 1mV 11MW
V(DC) 40.00V 10mV 10MW 0.5+2 1050V(DC) Continuous
400.0V 100mV 10MW
1000.0V 1V 10MW
400.0mV 100m
V 11MW 1.5+5
4.000V 1mV 11MW 1050V(AC) Continuous
10mV 10MW 1+5 rms
40.00V
V(AC)
400.0V 100mV 10MW
1000V 1V 10MW
Approx. voltage drop 1+10
at max. meas. current

40.00mA 10m
A 450mV
0.8+2 480mA Continuous
A(DC) 400.0mA 100m
A 4.2V
4)
10.00A 10mA 750mV 1.5+5 4) 4)

40.0mA 10m
A 450mV 480mA Continuous
1+5
A(AC) 400.0mA 100m
A 4.2V
4)
4)
10.00A 10mA 750mV 2+5 4)

www.rishabh.co.in Page 2 of 4 Version:1/05/VIII/1211/B


1)
Meas. Measuring Input Digital display inherent deviation at Overload capacity
Resolution
Function Range Impedance reference condition
V(AC)/V(DC) +(...%rdg + ...digits) Overload value Overload Duration
Open - circuit voltage

400.0W 100mW 0.8+5


W 4.000KW1W
40.00KW10W 0.8+2 500V
approx. 0.45V
DC/AC 10 min
400.0KW100W
rms
4.000MW1KW 1+5
40.00MW10KW 2+5

BUZZER 400.0W 100mW 0.8+5


Acoustic signal for 0...<100W
approx
DIODE 1.000V 1mV approx. 1V 2+10
2)
5.000nF 1pF 3+40
2)
50.00nF 10pF 2+10
2) 500V
500.0nF 100pF 0.5+3
F 2)
DC/AC 10 min
5.000m
F 1nF 1+2 rms
2)
50.00m
F 10nF 1.5+2
200.0m
F 100nF 3)
fmin 5+10

10.000Hz 0.001Hz 10Hz

100.00Hz 0.01Hz 10Hz <1KHz : 1000V


5)
1.0000KHz 0.1Hz 10Hz
Hz <10KHz : 400V
10.000KHz 1Hz 10Hz 0.2+2 Continuous

100.00KHz 10Hz 10Hz <500KHz : 40V


10Hz...1KHz: +5D except 400mV
500.0KHz 100Hz 10Hz
1KHz...10KHz: +5D/KHz
% 2.0...98.0% 0.1% -
Sensor
0 0
0...+1300 C 1C K 500V DC/AC
2+3 10 min
C NiCr-Ni rms

0 0
n
At C ... + 40 C
n
With zero adjustment „REL”;
n
Time requirement for measurement approximately 60 seconds. Ambient Conditions
n
max. 10 A/30 min Operating
12 A/5 min temperature range
0
-10 C ... + 50 C
0

16 A/30 s Storage temperature range


0 0
- 25 C ... + 70 C
n
Indication of the frequency measurement expanded to up to (without batteries)
9999 digit.
Relative humidity 45 ... 75 %
Elevation up to 2000 m
Reference Conditions
Ambient temperature
0
+ 23 C + 2 K Display
Relative humidity 45 % ... 55 %
LCD display field (50 mm x 30 mm) with digital display, and with
Frequency of
display of measurement unit, type of current and various special
meas. quantity Sine 50 Hz
functions.
Operating voltage 3V + 0.1V

Power Supply Digital


Display 7 segment
Battery 2 numbers of 1.5V mignon cell
Character height 10 mm
Zinc- carbon cell as per IEC R6
Number of digits 3 3/4 digit @3999 steps
Alkaline manganese dry cell as
Overflow display OL
per IEC LR 6
Polarity display „–” sign is displayed when plus
Service life Zinc-carbon cell: approx. 300
pole at „ ”
hours
Alkaline manganese dry cell: Measurement rate 3 measurement/s for V, I, W ,
appox. 600 hrs Capacitance,
Battery test Automatic display of „ ” symbol Frequency and Duty cycle
when battery voltage falls measurements
below following value:
approx. 2.4 V

Fuse
Fuse for ranges 1.6 A / 600V; 6.3 mm x 32 mm
up to 400 mA
Fuse for 16 A / 600V; 6.3 mm x 32 mm
10 A range

www.rishabh.co.in Page 3 of 4 Version:1/05/VIII/1211/B


Influence Interference Meas. Quantity /
variable Magnitude Meas. Range Attenuation

1000V DC/AC 50Hz All V DC >100 dB


sinusoidal
Common 1000 VDC All V DC >100 dB
Mode 400 mV / 4 V AC >80 dB
Interference
Voltage 1000V AC 50Hz 40 V AC >63 dB
sinus
400 V AC >43 dB
1000 V AC >23 dB
Series - Mode MAX. 1000 V AC 50/60Hz V DC >43 dB
Interference
voltage MAX. 1000 V DC V AC >55 dB

Aux. Voltage Influence


(without display) all ranges except Cap : + 8 D
Cap range : + 20 D
RISHmax display
1 Digital display with dot and polarity
2 Low Battery Indication
3 Display for REL and HOLD Applicable regulations and standards
4 Continuity test display:
speaker symbol appears when acoustic signal is switched on DIN 43751 Digital measuring instruments
5 Display for diode measurement DIN EN 60529 Test instruments and test procedures
-Degree of protection provided by enclosures
6 Measurement unit display DIN VDE 0470 part 1 (IP code)
7 Display for automatic measuring range selection IS 13875 Digital measuring instruments
8 Display for selected type of Voltage/Current (AC or DC)

Influence variable and effects Mechanical Design


Protection Instruments: IP 50
Influence Influence Meas. Quantity / Influence
variable range Meas. Range Connector sockets: IP 20
Effect
Dimensions W x H x D:
V ... 92 mm x 154 mm x 25 mm
V~
0
0 C ... +21 C
0
mA / A ... 0.1 x Weight Approx. 0.25 Kg with battery
intrinsic
Temperature 0 and 0
mA / A ~ error / K
+25 C ... +50 C W Standard Scope Of Supply
F
Hz 1 Cable set
Duty (%) 1 Multimeter
C 1 Copy Operating Instructions
1 Protective Case with tilt stand
Influence Influence range Frequency Inherrent Error at
variable (max. resolution) Ref. Designation Type Order Code
+(...%rdg. + ...digits)

400mV, 1000V 20Hz ... <50Hz 33061


2+3 Digital multimeter RISHmax 10
Frequency >50Hz ... 500Hz
VAC 20Hz ... <50Hz RISHmax Probe Set 42199
4V, 40V, 400V 750Hz ... 1KHz 2+3
RISHmax Fuse 1.6A 1.6 A / 600 V AC
RISHmax Fuse 16A 16 A / 600 V AC
Influence Influence Meas. Quantity / Influence
variable range Meas. Range Effect Safety cover RISHmax 10 42200
V~
mA / A ~ Subject to change without notice
Relative 55 ... 75% W 1 x Inherent
humidity error
F
Hz
Duty (%)
0
C

RISHABH INSTRUMENTS PVT.LTD.


F-31, MIDC, Satpur, Nashik-422 007,India.
Tel.: +91 253 2202160, 2202202 Fax : +91 253 2351064
E-mail : India :- marketing@rishabh.co.in
International :- exp.marketing@rishabh.co.in
www.rishabh.co.in
www.rishabh.co.in Page 4 of 4 Version:1/05/VIII/1211/B