Shimadzu Detector Splitting Software

User Manual

NOTICE
(1) All rights are reserved, including those to reproduce this
publication or parts there of in any form without permission in
writing from Shimadzu Corporation.
(2) Information in this publication is subject to change without
notice and does not represent a commitment on the part of the
vendor.
(3) Any errors or omissions which may have occurred in this
publication despite the utmost care taken in its production will
be corrected as soon as possible, but not necessarily immediately
upon detection.
(4) Repair and maintenance parts of this model are supplied until
seven years pass from stopping production of this model. After
that, those parts cannot be supplied. Please be forewarned.
However, except incorrupt parts of Shimadzu Corporation, other
corporation produces those parts and decides the length of
provision.
(5) Copyright © 2009 Shimadzu Corporation
Notice
The equipment may be different from the parts or figures of this manual for
production improvement. And the parts in this manual are not all maintenance
and repair parts.
This manual contains technical know-how about this equipment. Bring this
manual back necessarily after installing or fixing the equipment.
System Requirements
GCsolution (Ver.2.32 or later) or GCMSsolution (Ver.2.53 or later) is required for
loading GCsolution or GCMSsolution method files and data files on this
software.

1. Overview

This hardware is designed to allow a single injection to be split into two capillary
columns for separation and analysis by multiple detectors. This hardware is
optimized for use with capillary columns.

2. Overview of Software

Shimadzu detector splitting software is part of the Advanced Flow Technology


series.
This system provides column splitting for up to three channels.
Column flow, linear velocity, split ratio is based on the dimensions of the
restrictors connected to the detectors.
GCsolution or GCMSsotution by itself, can not calculate the correct values.
This software calculates the correct values and downloads to GCsolution or
GCMSsolution.
2.3 Launching of the software

Double-click on the file to open it Shimadzu AFT.

Program launcher appears, and double-clicks the launcher.


2.4 Construction of the software

This software consists of two program windows.

2.4.1 Setting window

Parameter of flow and


pressure

Flow rate of detectors Parameter for column


2.4.2 Property Window

Property window is opened by pushing configuration button.

Configuration

Select GCsolution or GCMSsolution

Number of the detectors

Use AUX-APC or not

Set the length of the


restrictor.

Set the inner diameter of the


restrictor.
3. Operation
(Assuming GC solution or GCMS solution is previously installed on the PC.)

1) Configuration
Detector splitting analysis can be performed with the following process.
At first select GCsolution or GCMSsolution in the property window.
And select Yes or No for APC.
And set the dimension of the restrictors.

3.1.3 Open the method file

Open the method file of GC/GCMS solution. Click the Open button in menu bar.

Open button
3) Check the method

Check and confirm the method, split ratio, column flow, and column velocity and
so on.

4) Save the method


Finally, click “Save” button on the tool bar to save parameters.

The characters “-2DET”or “-3DET” are automatically added to the original file
name in save dialog. When checkbox “overwrite save” is checked, the file is
overwritten. (“-2DET” or “3DET”is not added) To save a file with other name is
also available.

Save button
5) Analysis
Run the analysis using the method as developed. It is not necessary to use this
detector splitting software for this analysis. Please close this software and start
GC/GCMS solution.

4. Operation
(With the PC without GCsolution、GCMSsolution)

1) Configuration
Detector splitting analysis can be performed in the following process.
At first select GCsolution or MSsolution in the property window.
And select Yes or No for APC.
And set the dimension of the restrictors.

3) Check the method

Check and confirm the method, split ratio, column flow, and column velocity and
so on.
5) Analysis
Set the parameters of the method which was made. Set to GC directly or other software to
control the GC and analysis the sample.

4.Error message

The AUX-APC pressure is set too low.


Set the AUX-APC pressure to higher pressure value.

When the APC pressure is higher than the inlet pressure, this message is displayed.
Set the APC pressure to a lower value or set the inlet pressure to the higher value.

When the APC pressure is too high, this message is displayed.


This state may cause FID flame instability or Mass spectrometer vacuum errors.
Set the APC pressure to a lower value or use thin and long restrictors.