⑩Battery cabinet cover

INSTRUCTION MANUAL 1. SAFETY WARNINGS 2. Features  When battery capacity segment is less
than one, the user should replace the
This instruction manual contains warnings and safety rules which must be ●Wide Range 0.1∼199900 lux bat ter y. The bat ter y cabinet can be
observed by the user to ensure safe operation of the instrument and to ●MAX/MIN function opened with a screw driver and two AA
retain it in safe condition. ●Automatic Zero Adjust function type batteries should be installed.
 WARNING ●Large LCD Display with Backlight
●Read through and understand instructions contained in this manual
before starting using the instrument. 4. Specifications
●Save and keep the manual handy to enable quick reference whenever 3. Name of Parts and Position/Operation
necessary. MODEL KEW 5204
①Lens Protecting Cover
The symbol  indicated on the instrument means that the user must refer  This is used to protect the Light detector Silicon photo diode
to related parts of the manual for safe operation of the instrument. Be sure photocell from scratching and Measuring Range 0.0∼199,900 lx
to carefully read the instructions following each  symbol in this manual. create complete dark Range 199.9/1,999/19,990/199,900 lx
 DANGER :is reserved for conditions and actions that are likely to environment for the meter to Resolution 0.1 lx
cause serious or fatal injury. perform dark current Accuracy ±4%rdg±5dgt(23℃±2℃)
 WARNING:is reserved for conditions and actions that can cause calibration. 10° ±1.5%
serious or fatal injury. ②Light Sensing Probe/Photocell
Angle deviation from cosine 30° ±3%
 CAUTION :is reserved for conditions and actions that can cause ③LCD Display
minor injury or instrument damage. ④Power button characteristics 60° ±10%
 Push this button to turn on 80° ±30%
 WARNING the meter. Please push the Temperature characteristics ±5 % (at 23℃ reference and a range of 0 to 40℃)
●Do not install substitute parts or make any modification to the button again to turn it off. The Relative spectral sensitivity Deviation from spectral luminous
instrument. Return the instrument to Kyoritsu or your distributor for unit will automatically turn characteristics efficiency:9% or less
repair or re-calibration. itself off after 30 minutes if no Auto range:5s or less
key operation. To disable the Response time
 CAUTION Manual range:2s or less
automatic power off function,
●Do not expose the instrument to the direct sun, extreme temperatures Auto power off 30 min.
turn the meter off, push and
or dew fall. keep holding button, then push power button to turn on. Dimensions 169(L)x63(W)x37(D)mm
●This instrument isn't dust & water proofed. Keep away from dust and water. Weight 210g approx..
●Be sure to set the Function Selector Switch to the "OFF" position after ⑤Backlight button
use. When the instrument will not be used for a long period of time,  Push this button to turn on and off the LCD backlight. The backlight will Power source R6(AA)(1.5V) x 2
place it in storage after removing the batteries. be automatically off after 30 seconds to save battery power. Operation Temperature/ 0℃∼40℃,80%RH or less(without
⑥HOLD button
DIGITAL LIGHT METER ●Use a damp cloth and detergent for cleaning the instrument. Do not
use abrasives or solvents.  The user may hold the present and keep it on the LCD display by
Humidity condensation)
Storage Temperature/ -10℃∼60℃,70%RH or less(without
●The instrument should be calibrated once a year or so for the accurate pressing the button.
Humidity condensation)
measurement ⑦Max/Min button
LIGHT METER SERIES This symbol is marked where the user must refer to the
 Push this button to set the meter working under MAX/MIN mode. By Continuous measuring time Approx. 80 hours
pushing this button, the user may swap the LCD reading among current, JIS C 1609-1:2006
 instruction manual so as not to cause personal injury or maximum and minimum reading. Push and hold this key for 3 seconds
Applicable Standards
IEC61326

KEW 5204 instrument damage.


This instrument satisfies the marking requirement defined in the
WEEE Directive. This symbol indicates separate collection for
to exit MAX/MIN mode.
⑧Range selector button.
 The user may use this button to select the preferred measuring or
Accessories
Carrying case
Instruction Manual
R6(AA)(1.5V)×2
electrical and electronic equipment. automatically selected by the meter. The available ranges are: AUTO
200, 2000, 20000, 200000 lx
⑨0 ADJ button (Please refer to Measurement)

DISTRIBUTOR 5. Measurement ADEQUATE LIGHT LEVELS FOR YOUR WORKING OR AT YOUR WORK AREAS. (JIS Z9110-2010)
5-1 Zero adjust LUXES (lx) 3,000 2,000 1,500 1,000 750 500 300 200 150 100 75 50 30 20 15 10 7
 Please perform zero adjustment before measurement.
Entrance hall (day-time) Inquiry office, Dining
 Press 0 ADJ button with the Lens Protecting Cover attached to the Light Director room room Stair
Night-duty room
Sensing Probe. Indoor emergency
OFFICE Council room staircase
 "CAP-" will auto-appear on the LCD display. After "CAL-" appears, the Electronic computer room Library, Tearoom
Reception room Dressing room Warehouse, Lift, Passage
meter enters the automatic zero adjustment. When automatic zero Guard station Toilet, lavatory
adjustment is completed, "CAL-" disappears and "0.0"appears. After Stair
confirming the "0.0" display, remove the cap to start measuring. Design room Control room, Indoor emergency
 If“ ERR1”appears, check the cap to make sure it is on properly. FACTORY Extra fine visual work Electric room Entrance
staircase
Drawing room Air conditioning machine Passage
Warehouse
Path
 After the measurement has been completed, press the power button room, lavatory, Toilet
again to turn off the power, and cover the light-detecting surface with Classroom, Faculty room,
Drawing room Dining room, Gymnasium Stair
the cap for protecting against stray light. Night-duty room
SCHOOL Clothing classroom,
Electronic computer room Auditorium, Assembly
5-2 Measurement Experimental laboratory room Locker room Passage
Reading room
 Place the light sensing probe parallel to the measuring object and Healthcare room lavatory, Toilet
measure it. At this time, please be careful that your own shadow does Consultation room Nursery room
not hit the sensing probe. Nurse station Waiting room
Operating room Pharmacist office Visiting room
 The light sensing probe can be fixed on the meter parallel or Pharmacist's office for outpatients Ward Emergency staircase
General examination room Passage
perpendicular to the LCD Display. Or it may be detached from the unit to Emergency room Physiological examination Changing room, lavatory
fit the measuring requirement. room, Radioisotope room Hospital charts room
HOSPITAL Funeral chapel Toilet
X-ray room
Kyoritsu reserves the rights to change specifications or designs described Physiotherapy room,
Exercise machine room
in this manual without notice and without obligations. Night-duty room Ophthalmic dark room
Endoscopy room, X-ray
fluoroscopy room
Lift lobby, Escalator Reception room Passage, Resting room
Most important lavatory, Toilet
STORE part of display Important part of display, Negotiation room
Register, Packaging stand Stair
Sample case Register Stair
RESTAURANT Cookroom
Dining table
Waiting room
Guest room Porch, Passage
Counter lavatory, Toilet
Spectator seats, Lobby Entrance
Ticket office Electric room Resting room
Mechanical room Projection room
THEATRE lavatory, Toilet Passage Monitoring room Projection room
CONCERT HALL Stair
(during performance) (during performance)
Stall, Green room Theatre basement work
Monitoring room room
Front desk Carriage porch Stair Important parts of garden
Counter, Office Dining room

REF. LEVEL INN, HOTEL Cuisine Banqueting hall, Salle


Game room, Guest room,
Entrance, Passage,
Guest room desk Lobby, lavatory, Toilet Bathroom
Undressing room
BEAUTY SALON Hair dressing, Hair dyeing Hair cut, Shaving
Dressing
Stair
BARBAR SHOP Hair set, Make-up
Shampooing, Register Toilet inside the shop Passage

05-18 92-2349