Service Bulletin

Carestream Health, Rochester, NY 14608, USA

2017-05-11

DRYVIEW 5800, 5850 and TRIMAX TX50 Laser Imagers


Service Codes: 1115, 7153, 8835, 2833, 2832

EC 21-624 after Software Upgrade or Software Reload

IMPORTANT: When doing the procedures in this document, you must use safe work
practices and wear the correct Personal Protective Equipment (for example, Safety
Eyewear) according to your Company's Standard Operating Procedures.

Problem:

After performing a software upgrade (V2.7.b1) or software reload (reimage), the


5800/5850/TX50 printer may fail to calibrate small-size films e.g. 25 x 30 cm (10 x 12
in.) and 20 x 25 cm (8 x 10 in.) film sizes, and declares the error code EC 21-624.
However, the printer is able to successfully calibrate the 35x43 cm (14 x 17 in.) size.

This is often seen if a backup cannot be created or restored.

NOTE: The application log shows all calibration steps values to be read as 0 (zero):

Root cause:

The calibration pattern on 25 x 30 cm (10 x 12 in.) and 20 x 25 cm (8 x 10 in.) films is


shifted by around 2 cm so the densitometer cannot read the 21 calibration steps; i.e. the
calibration step wedges do not fall under the densitometer reading zone.
This happens because the 58x0 ghost image has a bad default for the Optics Parameters and
the FE did not restore the backup configuration so the original Optics Parameters are lost.

Page 1 Carestream Health, Confidential Pub No. SBAF3716


Solution:

For a DV5800 the Start Index Delay of 900 is good for 5800 as starting point (but it is no
good for 5850).

For DV5850 adjust the Start Index Delay for 10 in. film trays from 900 (wrong value) to
around 1400.

This is a temporary adjustment that will allow the printer to calibrate the small film sizes.

Important!
The FE should now perform the optics offset and delay adjustments as written and
described in the Service Manual Adjustments Section to set the values “correctly”.

Page 2 Carestream Health, Confidential Pub No. SBAF3716


For more information, please contact:

Carestream Health
Center of Excellence (COE)
1049 West Ridge Rd
Rochester, NY 14615
USA
USA Only: 1-800-328-2910
Outside the USA: 585-627-1864
Canada: 1-866-927-1017
Outside the USA and Canada: Contact your
local Shared Service Center (SSC)

CARESTREAM and DRYVIEW are trademarks of Carestream Health.

Created in the USA


© Carestream Health, Inc., 2017
Rochester, NY USA 14608

Page 3 Carestream Health, Confidential Pub No. SBAF3716

Вам также может понравиться