Section 1……………………………Overview
Section 2…………………………..User Interface
Section 3…………………………..Mythic™18 Fluidic
Section 4…………………………..Mythic™18 Electronic
Section 5…………………………..Mythic™18 Mechanic
Power Switch
Numerical
LCD Display Keyboard
touchscreen
Status Indicator
LED
Sampling module
Counting module
Syringes module
USB ports
TCP/IP port
Barcode
reader
Waste
connector RS232 port
Diluent
connector Centronic port
24 V Power
supply
External connector
grounding
Lyse solution
Straw
Cleaner Straw
Physical dimensions
Dimension Instrument
Height 350 mm
Width 250 mm
Depth 340 mm
Weight 11 kg
• Learning Objectives.
• Operator Screens.
• Technician Screen.
Section 2 will explain the user and service interface. After reviewing
this section of the workbook you should be able to do the following:
M ain M en u
C A L IB R A T IO N
A u t o m a t o n c o e f. a d j u s t .
Q U A L IT Y C O N T R O L
QC
R E SU LT S
V ie w c h a r t r e s u lt s
RES V IE W
V ie w R E S n u m b e r V ie w fu ll s c r e e n
s e le c t e d h is t o r y
SET U P
A DVA NCED
( B io lo g is t p a s s w o r d )
P R IN T E R C O M M U N IC A T IO N A N L Y S IS LA B . C A L IB R A T IO N OT HERS
O P T IO N S PA RA MET ERS FA C T O R S S E T T IN G S
L IM IT S
T H R E SH OLD S
A LA R M S
BLOO D T Y P E
CHA NGE
L IM IT S
T H R ESH OLD S
A LA R M S
S E R V IC E C O R R E C T . FA C T O R S
C ONT R OL C Y C LE M O T O R IN IT H Y D R A U L IC S M E C H A N IC S R E P E A T A B IL IT Y LOG E R R OR S T E C H N IC IA N
G e ne ra l C o ntro l M e c h a n ic C o n t r o l R u n s c le a n in g R u n s m e c h a n ic a l R u n s fo r A u to m a to n s ta tu s T e c h n i c ia n
c y c le c y c le c y c le s c o n t r o l c y c le s r e p e a t a b ilit y c o n t r o l r e po r ting fu n c t io n a lit ie s
D R A IN A L L C H EC K V A LV ES
R ESET A LL
R IN S E A L L
C H EC K NEED LE
C L E A N IN G
CHECK ROCKER
B A C K FLU SH
C H E C K S Y R IN G E
B LEA C H
C HECK SENSOR S
OP ER A T OR LOG OU T IN IT P R E S S
Q u it a n d la u n c h n e w u s e r H G B LE D O FF
N EED LE
REA GENT S D IS M A N T L IN G
R e a g e n t fu n c t io n s
L U B R IC A T IO N
P OS.
P R IM E P R IM E
P R IM E L Y S E
D IL UENT C L EA NER
LOG S
A u t o m a t o n d ia r y
V ie w R e s u lts
RUN N E X T ID
1- Password:
2- Language:
The USB key must be formatted before any use. The format should be done
on the Mythic 18.
How to proceed?
o One of the two USB port for the Mythic 18 equipped with normal USB port.
o The USB port on the lower position for the Mythic 18 equipped with small
USB port (like camera USB port).
o If the USB key is not compatible, a prompt will appear to inform you. In
this way, try another one.
NOTE: The format is done in FAT16. For this reason it is not necessary to
equip the Mythic 18 with a USB key with a capacity higher than 2 Gb.
2 Gb capacity = Around 60.000 patients.
128 Mb capacity = Around 10.000 patients.
Once your USB key is formatted, you can start to work with it.
1/ Common datas
RESULTS\ 0047.CSV Result analysis One sub folder per Year/month (YYYY- ASCII + UTF-8
(sequence number. CSV) MM), current unit
YYYY-MM\
One sub folder per day (DD), YES / YES
DD One file per result
Limited to 510 files
SETTINGS TECH.SET Private analyzer settings One file ASCII + UTF-8
YES / YES
REAGENTS REAGLOGS.CSV Private reagents log One file ASCII
YES / YES
Once installed, you can use the USB key how you want.
Two cases:
All the patient results will be saved on the internal memory of the instrument.
NOTE: The capacity of the internal memory is 1.500 patient results instead of 500 in the
previous version.
When the key will be locked on the instrument, the Mythic 18 will ask you if you want to
transfer the data on the USB. In this way, these data will be erased from the internal
memory.
If you have activated the option Real time runs save in the Storage Options, all the
results from the patient will be directly saved on the USB key.
NOTE: To avoid the problem of memory full, we recommend activating the option FIFO
Mode (First In, First Out) in the storage options. When the memory is almost full the first
result will be erased and replaced by the last one.
With the upgrade 2.0.0-009 or higher, we provide you with a USB driver; this driver is a
PCL3 driver, common to many USB printers.
In the Advanced Set-up you can select 3 new options in the Printer menu:
Archive of patients:
In the previous version, all the patients were archived in the same table from 1 to 500
(max.).
With this new version, each patient result is archived in list of the day with the following
architecture:
- YY/MM1
o dd1
o dd2
o etc …
- YY/MM2
o dd1
o dd2
o etc …
- etc …
Step 1:
To display one result for one patient, you just need to load
the list of the day while pressing the DATE button in the
ARCHIVE menu.
Step 2:
Select either “INTERNAL MEMORY” or “MASS
STORAGE”
INTERNAL MEMORY = Mythic 18 memory
MASS STORAGE = USB Drive.
Step 3:
From the archive menu, it is possible to save some results from the internal memory to
the USB drive.
To proceed, press on TOOLS from the main screen of the archive menu, select the days
you want to save, press SAVE.
With the update version 2.0.0-009 or higher, it is now possible to connect the Mythic 18
either with RS232 cable or using TCP/IP protocol.
This new option is available in Set-up, Advanced, and Communications.
Due to the new functionalities on the Mythic 18, please refer to the document
“Communication_protocol_v2.0.0_M18.pdf” to modify the communication protocol for the
existing connection.
Mythic S/N
Password
SW version
CPU Board n°
Input
- The lot diluent
- Expiry date
- Volume of
container in
ml.
- Press Change
Reagent to
confirm.
Note: Either you change one reagent and you prime this reagent or you
change all the reagents and you prime all.
1. Choose one lot 2. Input lot, targets and limits or press load
to load a new batch from USB key
RESTORE is used to restore a lot and all of the results from USB key to Mythic memory.
Only available with technician
password.
Default parameters :
Mythic 18 PC
DB9 male DB9 Female
1 --------------------------------------------------------------------- 1 CD
2 --------------------------------------------------------------------- 2 RXD
3 --------------------------------------------------------------------- 3 TXD
4 --------------------------------------------------------------------- 4 DTR
5 --------------------------------------------------------------------- 5 GND
6 --------------------------------------------------------------------- 6 DSR
7 --------------------------------------------------------------------- 7 RTS
8 --------------------------------------------------------------------- 8 CTS
9 --------------------------------------------------------------------- 9 RI
Technician Menu
Pressure:
- In the stand by mode, the pressure is 0.
- To check a leak of pressure:
o Run a cycle (Run Analysis)
o Stop the cycle by pressing the
ON/OFF button, when pistons are
in medium position.
o Go to Check Sensors.
o Check the value for Pressure.
o This value must be unchanged
along the time.
o If the pressure decreases, check
the tubing and the syringe block.
HGB sensor:
- If the LED is OFF, the value must be
around 0.
- If the LED is ON (turn ON in the previous
menu), the value must be between 850
and 950.
WBC and RBC sensors:
- In Stand by mode, the values are less
than 30. These values are measured
directly on the preamplifier board.
- If one or both values are higher (higher
than 100), check for interferences inside
the machine or check for an interference
source outside the machine.
Note: This part is only available with the technician password: 8SUP
CO=Aperture Current.
To check the current inside the counting
bath.
1. Turn 0N by pressing the button.
2. Open the fluidic door.
3. Put a pin of the voltmeter inside the
courting bath, a pin to the ground.
4. Measure the voltage.
1. When you receive a new upgrade, you receive a zip file like the following:
Mythic-2.0.1-003.zip
3. Copy the file with the extension “.b” in the USB key in the folder VERSION.
5. Go to Set-Up/Advanced/Version Release.
6. Select the upgrade file and press load. The upgrade will be automatically
done.
• Learning Objectives.
• Fluidic diagram.
• The different modules.
o Sampling module
o Counting module
o Syringe module
• Troubleshooting on the fluidic part.
Learning Objectives
Sampling cycle
3- Rocker move above the WBC bath and the syringes up to drain the
waste (valve 7 ON)
5- Syringe down
WBC bath drain (valve 1 ON)
6- Syringe up
Second outside needle cleaning (valve 4 ON)
Waste syringe drain (valve 7 ON)
7- Syringe down
WBC bath drain (valve 1 ON)
8- Syringe up
Waste syringe drain (valve 7 ON)
Sample dilution
WBC Dilution
9- Needle down to the dilution position and the syringes down to prime
the diluent (valve 8 ON)
14- Syringe up
Second outside needle cleaning (valve 4+8 ON)
Lysis in the WBC bath (0,3 to 0,5ml) (valve 9 ON)
Bubble to mix the dilution (valve 1 ON)
Measurement:
20- Syringe up
25- Syringe up
RBC bath rinse diluent (valve 4 ON)
Waste syringe drain (valve 7 ON)
26- Needle up and syringes up for back flush (valve 4+6 ON)
28- Syringe up
WBC bath rinse diluent (valve 4 ON)
Waste syringe drain (valve 7 ON)
31- Syringe up
Waste syringe drain (valve 7 ON)
Note1: For replacement of the O-rings on the syringe block, please refer to
the Mythic™18 service manual and the technical note.
Note2: The O-rings on the seal plate of the syringes module are greased in
the factory, as well as the pistons. It is very important that the greasing of
these parts is maintained all the time. We recommend greasing the
pistons every 6 months and the O-rings once per year. The pistons can
be greased by the customer.
WBC: --- *
RBC: --- *
PLT: --- *
Or
In any case, the flags “W_CL” or “R_CL” appear in the flag box below the
results.
Service Menu
Hydraulic
Cleaning
End of
OK Run Sample
service
NO
Backflush
OK
Run Sample
NO
OK Run Sample
NO
It could happen that you observe a drop of blood falling from the needle after
the blood aspiration.
Note: As we explain before, none of the valves is used during this part of the
cycle. So, it is not a problem of valve.
This problem appears when a clog is located in the rinsing head of the probe.
Rinsing
head
Tubing
n°10 :
Tubing n°21 :
Aspiration
Diluent tubing
tubing
NO
- HCT result rejected or too low (not proportional with Hb result). Impossible
to recalibrate.
- Repeatability problem on the results.
1- Diagnosis
Login as Technician
Run analysis
1/ Service Menu
2/ Mechanics
3/ Check sensors
PRESSURE Check the value of the pressure. Should be stable along the time.
Check leak of liquid
STABLE If the pressure is decreasing, leak of pressure
PRESSURE UNSTABLE
NO
End of service
- Check if mixing by air inside RBC chamber seems correct: check if tubing
14 is clamped.
- Clean RBC/PLT aperture with a bleach solution and perform a backflush.
• Learning Objectives.
• Electronic diagram.
• Boards description.
• Wiring description.
• Valves replacement.
• Electronic gain adjustment.
Learning Objectives
Section 4 will explain the Mythic™18 Electronic part. After reviewing this
section of the workbook you should be able to do the following:
+12V DC
CMD Rocker
CMD Needle
+3.3V
CMD Syringe
EV
+ 5V CMD
Counting EV
Pulse
processing
DRA
logic
+3.3 V DC
+ 5 V DC
+60 V DC
+2.5 V DC
NOTE:
• Press in the same time on the numerical keyboard the key “.” and
“DEL”.
• A cross appears on the display.
• Press in the middle of the cross.
• Proceed as previously with the next cross which appears.
• When the calibration is done, the display comes back to its initial
status.
Door switch
Counting EV
Rocker Optical
Sensor
Needle motor
Motor
syringe Rocker motor
Syringe Optical
Sensor
NOTE: For the wiring of the different parts (sensors, motors, valves …),
please refer to the Mythic™18 service manual.
Note:
WARNING
EV 3/2 Ø 1.6mm
Note1:
• Main board
• Preamplifier board
• Counting chamber
• Counting heads
Note2:
GAIN
- Wait the end of the cycle and check that values are inside the following range:
- Wait the end of the cycle and check that values are inside the following range:
• Learning Objectives.
• Syringe motor sub-assembly and greasing
• Sampling module and greasing
Learning Objectives
Section 5 will explain the Mythic™18 mechanic part. After reviewing this
section of the workbook you should be able to do the following:
- The endless screw with the grease “Lubriplate DS-ES” ref. 410-0502-01.
- The guide of the pistons holder with the grease “Plastic grease SPG35”
ref. 410-0503-01
- Axis of the rocker c with the grease “Plastic grease SPG35” ref. 410-
0503-01
- The rail of the rocker d with the grease “Plastic grease SPG35” ref. 410-
0503-01
Refer to the technical note for
greasing recommendations.