Академический Документы
Профессиональный Документы
Культура Документы
mate
course and section:
on-tructor:
ExPERIMENT ac 2
Oscilloscope
OBJECTIVES
1. Become familiar with the construction, components, and fundamental operation ofan
oscilloscope.
EQUIPMENT
Instruments
Equipment issued, DMM (Fluke45_5035149), Oscilloscope (Tektronix_2245A), Function
Theory of Oscilloscope
The oscilloscope is an instrument that will display the variation of a voltage with time on a flat
screen monitor. The vertical axis is scaled off in volts while the horizontal axis is in units of
time. The number of vertical and horizontal divisions on the screen is not fixed, but the majority
have eight vertical divisions and 10 horizontal divisions The basic components of an
reduced in level (attenuated) or increased in level amplified. The horizontal input permits
applying another signal of any kind to interact with the vertical input to produce a waveform that
can often be quite informative. However, if you simply want to view the signal applied to the
vertical input versus time, then the sync input option is selected. The sweep oscillator will then
generate a saw tooth waveform, such as in Fig. 21, to move the applied signal across the screen,
If the frequency of the applied signal and that of the saw tooth waveform are the same, the
waveforms, are said to be "in sync" and the desired signal will sit stable on the screen. If the two
frequencies do not match, the waveform will appear to be continually moving horizontally.
Fortunately, all scopes have a sweep time control to adjust the frequency of the saw tooth
waveform so a steady image can be displayed. Note in Fig. 21 that the voltage of a saw tooth
waveform increases linearly with time. This is to ensure that the applied signal will appear across
the full width of the screen in an undistorted manner. For the case of an applied sinusoidal
voltage, the voltage between the two input terminals will increase and decrease in an oscillatory
manner. In the absence of the saw tooth wave format the horizontal input, the waveform on the
screen would simply be a vertical line with a high intensity spot moving up and down on the
screen with the same frequency as the applied signal. Applying the saw tooth voltage to the
horizontal input will move the waveform across the screen so the full sinusoidal pattern can be
displayed. Most of the controls for the proper operation of an oscilloscope are mounted on the
front panel of the instrument. Fig: 2.2 indicate the approximate locations of the controls found on
most general-purpose oscilloscopes. The locations of the controls shown vary according to
Control Purpose
Power For turning on power
Intensity To adjust the intensity of pattern on screen
Focus Focus electron beam to clearly define pattern
Triggering mode Determines the type of triggering for horizontal sweeping pattern
Triggering level Determine the level at which the triggering will occur
Vertical display mode Determines number of signals displayed on screen and the type of
Basic measurement
a. Voltage
DC levels
To use the scope to measure dc levels, first place the DC/AC/GND switch in the GND position
to establish the ground (0-V) level on the screen. Then switch DC/AC/GND switch to the dc
position to measure the dc level. In the ac mode, a capacitor blocks the dc from the screen. Next,
place the scope leads across the unknown dc level and use the equation 2.1.
𝑉
𝑑𝑐 𝑙𝑒𝑣𝑒𝑙 (𝑉) = 𝑑𝑒𝑓𝑒𝑙𝑒𝑐𝑡𝑖𝑜𝑛 (𝑑𝑖𝑣)𝑥 𝑣𝑒𝑟𝑡𝑖𝑐𝑎𝑙 𝑠𝑒𝑛𝑠𝑖𝑡𝑖𝑣𝑖𝑡𝑦 ( ) … (2.1)
𝑑𝑖𝑣
AC levels
After reestablishing the ground level, place the DC/AC/GND switch the ac position. Next, place
the scope leads across the unknown AC voltage. Peak to peak voltage can be measured fro
equation 2.2
𝑉
𝑉𝑝−𝑝 (𝑉) = 𝑑𝑒𝑓𝑙𝑒𝑐𝑡𝑖𝑜𝑛 𝑝𝑒𝑎𝑘 𝑡𝑜 𝑝𝑒𝑎𝑘 (𝑑𝑖𝑣)𝑥 𝑣𝑒𝑟𝑡𝑖𝑐𝑎𝑙 𝑠𝑒𝑛𝑠𝑖𝑡𝑖𝑣𝑖𝑡𝑦 ( ) … 2.2
𝑑𝑖𝑣
b. Frequency
The oscilloscope can be used to set the frequency of an audio oscillator or function generator
Determine the period of the desired waveform and then calculate the number of divisions
required to display the waveform on the horizontal axis using the provided us/div., ms/div.,
or s/div on the horizontal sensitivity control. Then adjust the audio oscillator or function
generator to provide the proper horizontal deflection for the desired frequency of course, the
reverse of the above procedure will determine the frequency of an unknown signal.
The audio oscillators designed to provide a sinusoidal waveform in the frequency range
audible by the humaneur. A function generator typically expands on the capabilities of the
audio oscillator by providing a square wave and triangular waveform with an increased
frequency range, Either instrument is suitable for this experiment, since we will be dealing
only with sinusoidal waveforms in the audio range. Most oscillators and generators require
that the magnitude of the output signal be set by an oscilloscope or DMM. That is, the
amplitude dial of the oscillator is not graduated and the peak or peak-to-peak value is set by
connecting the output of the oscillator to a scope or meter and adjusting the amplitude dial
Procedure:
Part 1
(a) Instructor provided information about the basic operation of the oscilloscope and function
generator.
(b) The oscilloscope was turned on and a horizontal line centered on the face of the screen
was established. There were no connections to the vertical input sections of the scope for this
part
(c) The controls listed in table 2.2 were set and effects were observed
a. Set the DC/ACGND switch to the GND position and adjust the Y-position control until the
b. Once the 0-V level is established, move the DC/ACGND switch to the de position and set
the vertical sensitivity to 1 V/div and connect one channel of the scope across the 1.5-V
Multisim result
Determine the dc voltage that established the shift by multiplying by the vertical sensitivity. That
Dc voltage= (1.4)(1)
Change the sensitivity to 0.5 V/div. and note the effect on the vertical shift Recalculate the do
Multisim result
(e) Disconnect the 1.5-V battery and re-establish the 0-v reference line. Then connect the vertical
input section of the scope as shown in Fig. 2.4 with the vertical sensitivity set 1 V/div.
Fig. 2.4
Multisim results
In this part of the experiment, we will learn how to set the magnitude of a sinusoidal signal using
Oscilloscope:
Fig. 2.5
b. Frequency of oscilloscope/generator was set to 500 Hz using the dial and appropriate
Multism result
c. Vertical sensitivity of the Scope was set to I V/div and the Horizontal sensitivity of the Scope
to 0.5ms/div and turn on both the scope and the oscillator or generator.
Multism result
d. DC/AC/GND switch was set to the GND position to establish the 0V reference level and then
Multism result
e. amplitude control of the function generator was so adjusted that signal reached to 6V(peak to
peak) swing. The resulting waveform has the following mathematical equation
Multism result
f. Switch to the dc position and comment below about the any change.
Lab result
g. make the necessary arrangements to display the following waveforms on the screen.
Multisim result
Fig 2.6
1. v=8 sin 2π500t
Lab result
Fig 2.6
DMM was directly connected across the oscillator in the ac rms mode and the oscillator was
adjusted so that output reaches until V = 2.121 v. Then the output of the oscillator was connected
directly to the scope for noting the total peak-to-peak swing. Is the waveform the same as that
Same.
Lab result
(i) Use the DMM to set the following sinusoidal output from the oscillator v = 0,5 sin 2π500t
Veff= 0 365V. Set Veff with the DMM by adjusting the output of the oscillator, and place the
Same.
This section will demonstrate how the oscilloscope can be used to set the frequency output of an
oscillator. In other words, the scope can be used to make fine adjustments on the frequency set
For the signal such as 2 sin 2π500t, the frequency is 500 Hz and the period is 1/500 Hz= 2ms.
With a horizontal sensitivity of 0.5 ms/div., the waveform should appear in exactly four
horizontal divisions. If it does not, fine adjust control on the frequency of the oscillator or
generator can be adjusted until it is exactly four divisions. The scope has then set output
frequency of the oscillator. Make the necessary adjustment to place the following waveforms on
the scope. Sketch the waveforms, indicating the number of vertical and horizontal deflections
F= 10000Hz, T= 0.0001 s
Horizontal sensitivity=20ms
2. V=5 sin 377t
F= 60Hz, T= 0.0167 s
Vertical sensitivity= 2V
Horizontal sensitivity=5ms
Part 5 sinusoidal waveforms on a dc level
a. Set the oscillator to an output of 1sin 500t using the vertical sensitivity of 1V/div. on the
b. Measure the dc voltage of one of the D cells and insert in Fig. 2.10.
E=0.51 V almost
c. Construct the series combination of supplies as shown in fig. 2.10 and connect the scope as
indicated.
Fig. 2.10
Lab result
d. The input signal has a dc level equal to the dc voltage of the D cell. Set the DC/AC/GND
switch to GND position and adjusts line to the center of the screen.
f. Now switch to the DC mode and sketch the waveform on the same scope pattern
Lab result
as part 5(e) (g) what was the effect of switchingfrom the Actoth* DC mode?
Up 1. 5V
No
How does the vertical shift compare to the de level of the battery?
Up 1. 5V
(h) Switch to the GND mode and describe what happened to the waveform. In general, what is
the effect of switching to the GND position, no matter where the leads of the scope are
connected?
Goes to zero