Instrumentation of GC
• Injection of the sample may be made manually of
using autosampler through a resealable rubber
septum
• The sample is evaporated in the heated injection
port area and condenses on the head of the column
• The column may either be a capillary or a packed
column. Mobile phase is a gas; nitrogen or helium
• The column is enclosed in an oven which mat be set
at any temperature between ambient and ~400
• The most commonly used detector is the flame
ionisation detector (FID)
TOWARDS EXCELLENCE
Schematic diagram of a
capillary GC system
TOWARDS EXCELLENCE
Gas Chromatography
Carrier Gas
TOWARDS EXCELLENCE
GC Syringes
TOWARDS EXCELLENCE
Split/Splitless Injection
• Split/Splitless Injection is used in conjunction with
capillary column GC
• Capillary column internal diameter between 0.2 and 0.5
mm X 12 and 50 in length. Injection takes place into a
heated glass or quartz liner rather than directly onto
column
• In split mode the sample is split into two unequal
portions the smaller of which goes onto the column
• Split ration range between 10:1 and 100:1 with large
portion being vented in the higher flow out of the split
vent
• This technique is used with concentrated samples
TOWARDS EXCELLENCE
Gas Chromatography
Sample Injection
TOWARDS EXCELLENCE
Gas Chromatography
Sample Injection
TOWARDS EXCELLENCE
Gas Chromatography
Sample Injection
TOWARDS EXCELLENCE
GC Oven
• GC ovens incorporate a fan, ensures uniform heat
distribution throughout the oven
• Oven can be programmed either produce a
constant temperature, isothermal conditions or
gradual increase in temperature
• Oven temperature programming rates from 1-
40C/min.
• Complex temperature programmes can be
produced involving a number of temperature ramps
interspersed with periods of isothermal conditions,
e.g. 60 C (1 min)/ 5 C/min to 100 C(5min)/ 10
C/min to 200 C (5 min)
TOWARDS EXCELLENCE
Cool on-column injection
• Direct on-column injection into the capillary column
may be carried out in a manner analogous to
injection into packed columm
• This technique requires a syringe with a very fine
fused silica needle
• The technique has the advantage of:
(1) Reduced discrimination between components in
mixture
(2) No sample degradation in a hot injector
(3) No backflash hence quantitative sample transfer
TOWARDS EXCELLENCE
Gas Chromatography
Sample Injection
TOWARDS EXCELLENCE
GC Oven
• GC ovens incorporate a fan, ensures uniform heat
distribution throughout the oven
• Oven can be programmed either produce a
constant temperature, isothermal conditions or
gradual increase in temperature
• Oven temperature programming rates from 1-
40C/min.
• Complex temperature programmes can be
produced involving a number of temperature ramps
interspersed with periods of isothermal conditions,
e.g. 60 C (1 min)/ 5 C/min to 100 C(5min)/ 10
C/min to 200 C (5 min)
TOWARDS EXCELLENCE
Gas Chromatography
Temperature and Pressure Programming
TOWARDS EXCELLENCE
Advantages of temperature
programming
• Materials of widely differing volatilities can be
separated in a reasonable time
• Injection of the sample can be carried out at low
temperature where it will be trapped at the head
of the column and then the temperature can be
raised until it eluted
TOWARDS EXCELLENCE
Type of column-Capillary Column
• Capillary columns are made from fused silica, usually
coated in the outside with polyamide to give the
column flexibility
• Coating on the outside with aluminium has also been
used for high temperature (>400 C ) work
• The internal diameter of the columns ranges between
0.15 and 0.5 mm.
• The wall of the column is coated with the liquid
stationary phase, which may have a thickness between
0.1 and 5 m
• The most common type of coating is based on organo
silicone polymer, which are chemically bonded to the
silanol groups on the wall of the column and the chains
of the polymers are further cross-linked
TOWARDS EXCELLENCE
Capillary Column
• These types of phases have more or less replaced the
wall-coated open tubular (WCOT) and support coated
open tubular (SCOT) columns
• Which are reported in earlier literature, for most
routine applications
• SCOT columns are sometimes encountered in very high
temperature work
• The wall-bonded phases are stable to at least 325C
and some types of coating will withstand temperature
370C
• The non-silicone based polymers, e.g. Carbowax,
cannot be bonded onto the wall of the column in the
same way and columns with these coatings are less
temperature stable
TOWARDS EXCELLENCE
Capillary Column
• The temperature limit for a carbowax capillary
column is ~240C
• Common carrier gas is helium and flow rate 0.5 – 2
ml/min
• Flow rate from the end of the capillary column is
low compared to the internal space of some
detectors, ‘make up’ gas often has to be added to
the gas flow post column in order to sweep the
sample through the internal volume of the detector
at a reasonable rate
• ~100 ng per component is loaded onto capillary
column
TOWARDS EXCELLENCE
Gas Chromatography
Instrumentation
➢ Increasing Resolution
- Narrow columns Increase resolution
TOWARDS EXCELLENCE
Gas Chromatography
Instrumentation
TOWARDS EXCELLENCE
Gas Chromatography
Instrumentation
Also, increase in
capacity factor and
reduce peak tailing
TOWARDS EXCELLENCE
Gas Chromatography
Instrumentation
➢ To reduce “bleeding” of
stationary phase:
- bond (covalently
attached) to silica
- Covalently cross-link to
itself
TOWARDS EXCELLENCE
Types of column: Packed columns
• The column are usually made from glass which is
silanised to remove polar silanol Si-OH groups from its
surface that can controbute to the peak tailing af the
peaks of polar analytes
• These columns have internal diameters 2-5 mm
• The columns are packed with particles of a solid
support is a diatomaceeous earth (mainly calcium
silicate)
• This material is usually acid washed to removed mineral
impurities and then silanised as shown in the following
figure 11.4 to remove the polar Si-OH groups on the
surface of the support, which can lead to tailing of the
analyte peak
TOWARDS EXCELLENCE
Types of column: Packed columns
• Mechanically coated with a variety if liquid
stationary phase
• Mobile phase in packed column: nitrogen with flow
rate of ~20mL/min
• Resolution of packed column is lower than capillary
– 4000-6000 plates for 2 m packed column
– >100 000 plated for 25 m capillary column
• Upper temperature limit of packed column is 280C
• Beyond 280C, the liquid stationary phase
evaporates and creates a large background signal
• Pack column are adequate for routine quality
control (QC) operations
TOWARDS EXCELLENCE
Gas Chromatography
Instrumentation
TOWARDS EXCELLENCE
Gas Chromatography
Instrumentation
500 L chromatography
3.) Packed Columns column
➢ The major advantage and use is for large-scale or
preparative purification
TOWARDS EXCELLENCE
Selectivity of liquid phases-
Kovat indices and column polarity
• Kovats indices (I-values) are based on the retention
times of the series of n-alkanes
• For particular GC phase, I-values are very reproducible
from one column or from one GC to another
• However, they are slightly affected by GC programming
conditions
• n-Alkanes have most affinity for non-polar phases and
tend to elute more quickly from polar phases
• In contrast, polar analyte will elute more slowly from a
polar phase and thus relative to the n-alkanes, its
retention and thus its I-value will increase as the
polarity of the GC phase increase
TOWARDS EXCELLENCE
Selectivity of liquid phases-
Column polarity
• A measure of the polarity of a stationary phase is given by its
McReynold’s constant
TOWARDS EXCELLENCE
Commonly used GC detectors
TOWARDS EXCELLENCE
Commonly used GC detectors
TOWARDS EXCELLENCE
Commonly used GC detectors
TOWARDS EXCELLENCE
Examples of the separation
of mixtures by GC
• Analysis of peppermint oil on two GC phases
• Structures of some component in peppermint oil
• Figure indicates approximate I-values for some of the
components in peppermint oil on a BPX-5 column
(Molecular weight and shape)
• Shape of -pinine has the same MW as limonene but
more compact => lower I-value
TOWARDS EXCELLENCE
Analysis of peppermint oil on
two GC phases
• Methyl acetate has a higher I-
value tha menthol because of its
MW
• Methyl acetate elutes later than
menthol on BPX-5 non polar
column
• Methyl acetate elutes earlier
than menthol on Carbowax
column
• Polar alcohol is masked by
acetate ester group thus lower
polarity then menthol
TOWARDS EXCELLENCE
Analysis of peppermint oil on
two GC phases
• Carbowax is highly selective for
polar compounds
• Polar compounds including
menthol and methone is
resolved more extensively on a
carbowax column with alcohol
menthol and a number of other
minor alcohols eluting at around
12 minutes
• Less polar ketone menthone and
a number of ketones elute at
around 10 minutes
TOWARDS EXCELLENCE
Examples 2 of the separation
of mixtures by GC
• Analysis of the fatty acid composition of a fixed oil
by GC
• Very polar carbowax is used with high degree of
polar discrimination and fatty acid with differing
degrees of unsaturation
• Non-polar BPX-5 cannot discriminate 18 carbons
fatty acid (Stearic, oleic, linoleic and linolenic with
0, 1, 2 and 3 double bonds)
• BP monograph uses GC for quality control to the
content of fatty acids coposing triglycerides
TOWARDS EXCELLENCE
Separation of mixtures by GC
– Monograph for almond oil
• Composition of the fatty acids making up the
triglyceride should be:
• first number: number of carbon
• second number refers to number of double
bonds
• The percentage of each component is
determined in relation to the sum of the areas
of the chromatographic peaks
• Method: triglycerides
have to be hydrolysed
to liberate fatty acids TOWARDS EXCELLENCE
• BPX column-incomplete
resolution – C20 and C22
acids are not separated
• Carbowax column-complete
separation of oleic (18:1),
linoleic (18:2), stearic acid
(18:0) and small amount of
linolenic acid (18:3)
• Percentage of area as
follows: 16:0 (7.0%), 16:1
(0.4%), 17:0 (0.12%), 18:0
(1.5%), 18:1 (62.8%), 18:2
(28.4%), 18:3 (0.16%), 20:0
(0.09), 22:0 (0.09%)
• Almond oil within BP
specification
TOWARDS EXCELLENCE
Application of GC in quantitative
analysis
• HPLC is more important than GC method for
quantifying drugs in pharmaceutical preparation
• Quantitative GC is the method of choice in compound
that lacks chromophores eg. Amino acids, fatty acids
and sugars
• The BP assay format is often:
– Solution 1: A calibration containing more or less equal
amount of pure standard and internal standard
– Solution 2: An extract of sample containing no internal
standard to check for interference from the formulation
matrix
– Solution 3: An extract from the samplr containing the same
amount of internal standard as Solution 1
TOWARDS EXCELLENCE
Chromatogram of Solution 1, 2 and 3 prepared
for the analysis of methyltestosterone tables.
• RTX-1 column 15 m x 0.25 mm i.d. x 0.25 m film
• Programme 150C (1 min) then 10C /min to 320C (5 min)
• Analysis of methytestosterone in a tablet formulation using
testosterone as a internal standard
TOWARDS EXCELLENCE
Analysis of methyltestosterone
in tablets
• A calibration solution containing ca 0.04% w/v of
methyltestosterone and 0.04% w/v testosterone in
ethanol is prepared (Solution 1)
• A weight of tablet powder containing ca 20 mg of
methyltestosterone is extracted with 50 mL of
ethanol to prepare Solution 2
• Solution 3 prepared by dissolving tablet powder
contaning ca 20 mg of methyltestosterone in exactly
50 mL ethanol containing exactly the same
concentration of testosterone as Solution 1
• 0.5 µL amounts of the solutions were injected into
the GC in the splitless mode TOWARDS EXCELLENCE
Analysis of methyltestosterone
in tablets
TOWARDS EXCELLENCE
Data from analysis of
methyltestosterone tablets
TOWARDS EXCELLENCE
Quantification of ethanol in a
formulation
• GC is a useful method for quantifying very volatile
materials
• In this analysis, columns that strongly retain volatile
compounds
• Ethanol is used in the preparation or tincture and in
disinfectant solutions
• Typically ethanol quantified against a related alcohol
• BP assay of chloroxylenol solution ethanol is quantified
against a propan-1-ol internal standard
• Column: Porapak Q-porous polymeric stationary phase
which retains low molecular weight compounds
strongly
• Porapak is effective in separating gases such as CO2,
ammonia and acetylene TOWARDS EXCELLENCE
Determination of manufacturing
and degradation residues by GC
• Determination of pivalic acid in dipivefrin eye drops
• GC provides a useful technique for estimate volatile degradation
products
• Pivalic acid release from the hydrolysis of dipuvefrin in an eyedrop
preparation used from treating glaucoma may be estimated by GC
• Isovaleric acid, which is an isomer of pivalic acid, is suitable
internal standard
• Breakdown products of esters are more likely to occur in aqueous
formulations such as eyedrops or injection
TOWARDS EXCELLENCE
Determination of manufacturing
and degradation residues by GC
• Dimethylaniline in bupivacaine injection
• Dimethylaniline is both a manufacturing impurity in
bupivacaine and since it is formulated in injections a
possible breakdown product
• Hydrolysis of amide is much slower than hydrolysis of
esters
• BP uses a spectrophotometric method to assay for this
impurity but GC provides a more sensitive and specific
method for this determination
TOWARDS EXCELLENCE
Determination of manufacturing
and degradation residues by GC
• The GC trace obtained from injection of a 10% w/v solution of
bupivacaine free base extracted from an injection gave the
trace shown below
• Comparison with a standard from dimethyaniline that there is
≤0.1% of the impurity present although a number of other
peaks due to excipients or impurities can be seen in the GC
chromatogram
TOWARDS EXCELLENCE
Determination of manufacturing
and degradation residues by GC
• Determination of residual solvents
• Current BP methods for determination of sulvent
residues in pharmaceuticals remaining from the
manufacturing process rely on direct injection of
the sample dissolved in a suitable solvent (often
water) and based on packed column GC
TOWARDS EXCELLENCE
Determination of manufacturing
and degradation residues by GC
• Determination of residual solvents and volatile impurities by head
space analysis
• A more refined method for determining residual solvents and
volatile impurities in based on head space analysis
• The simplest method of sampling is to put the sample into sealed
vial and heat it
• Sample is placed in a sealed vial fitted with rubber septum, heated
• Fixed volume of head space (1 mL) is
withdrawn
• The sample is then injected into a GC in
the usual way
• If capillary column GC is used a split
injection has to be used to facilitate
sample injection
TOWARDS EXCELLENCE
Determination of residual solvents and
volatile impurities by head space analysis
• Several points are important to note:
TOWARDS EXCELLENCE
Purge trap analysis
• Another form of head space analysis uses purge
trapping device to trap volatile impurities
• In this technique a gas e.g. Helium is bubbled through
the sample which is dissolved in suitable solvent
(usually water)
• The volatile impurities are thus ‘stripped’ from the
solution and passed in the stream of gas through a
polymeric adsorbent where they become trapped and
thus concentrated
• The stream of gas is then switched so it passes in
reverse direction through the polymeric trap, which is
heated to desorb the trapped volatiles and the gas
stream is then diverted into the GC
• Used in environmental analysis to concentrate volatiles
in water which are present at low levels
TOWARDS EXCELLENCE
Gas Chromatography
Purge and Trap
➢ Removes volatile analytes from liquids or solids,
concentrates sample and transfer to GC
Connect port to GC
TOWARDS EXCELLENCE