How to adjust the calibration for the DI-2000

We would like to inform the procedure of calibration for the accuracy.


The accuracy of the infusion pump is depending on the quanty of i.v set manufacturer, production lot, thickness, hardness,
diameter and position of i.v set.
So, we recommend to test the accuracy every one unit at least 3 times.

Please try to check out the accuracy error with analyzer or measuring cylinder according to the below.
First of all, you have to set "USER MODE" and please check out the accuracy of the pump 3 times.
If you can see "U001" on third display, press the button "MODE" during 3 seconds.
* How to enter the "USER MODE" of the DI-2000 *
If you can see "U001" on third display, press the button "MODE" during 3 seconds.
1. Please press button “Stop” and "Sel" at the same time during about 3 seconds.
2. If you can see "I V.St" on first display, press button “MODE” 2 times and then you can see "Set 3" on second display.
3. Please press button “BOULS” 1 time and check out the accuracy against below procedure.

* Procedure *

1. Please check out the measure flow rate(50ml/h or 100ml/h) during 1 hour by using analyzer or measuring cylinder.
2. Please repeat 3 times every one pump.
3. Please calculate the everage of accuracy error.
4. Please adjust the calibration value of USER MODE in the pump.
5. Please check out the measure flow rate(50ml/h or 100ml/h) during 1 hour by using the pump analyzer or measuring cylinder.

* How to adjust the calibration value for USER MODE *

1. Please press the buton "Stop" and "Sel" at the same time during 3 seconds.
2. If you can see "U001" on third display, press the button "MODE" during 3 seconds.
3. Please press the button "BOLUS" after adjusting the calibration value.

(Example)

- Error rate = (Set the Flow rate – Measure Flow rate) / Set Flow rate * 100

Flow rate Test

Error rate: (Set Flow rate - Measure Flow rate) / Set Flow rate * 100

Before Measure Flow After


Flow
NO Calibration(CLAC) rate/volume(m Error rate Calibration(CLAC)
rate(ml/h)
Value l/h,ml) Value
1 0.0 100 113 13%
2 0.0 100 114 14%
3 0.0 100 115 15%
Average Error + 14% 0.0 -> -14.0

If you will get the result like above, you will be able to adjust the calibration value in the pump.

Flow rate Test

Error rate: (Set Flow rate - Measure Flow rate) / Set Flow rate * 100

Measure Flow
After Calibration Flow
NO rat/volume(ml/ Error rate
(CLAC)Value rate(ml/h)
h,ml)
1 -14.0 100 100 0.00%
Average 0.00%

And you should check out the accuracy of DI-2000 once again.