Академический Документы
Профессиональный Документы
Культура Документы
English
2039822-002 A (paper)
© 2009 General Electric Company.
All rights reserved.
B30 Patient Monitor
User's Guide
Related to software L-DICU08
0459
i
1
Safety precautions
These precautions refer to the entire system. Warnings and cautions specific to parts of the system can be found in the relevant section.
• Do not touch the patient, table, instruments, modules or the
Warnings monitor during defibrillation.
• If the integrity of the external protective earth conductor
A WARNING indicates a situation in which the user or arrangement is in doubt, use the monitor with battery operation.
the patient may be in danger of injury or death. • When detaching modules, be careful not to drop them. Always
support with one hand while pulling out with the other.
• Connect only one patient to the monitor at a time. • Use only approved accessories, batteries, mounts and
• Do not use the monitor without manufacturer approved defibrillator-proof cables and invasive pressure transducers. For
mounting attached. a list of approved supplies and accessories, see the "Supplies
• Use only hospital-grade grounded power outlets and power and Accessories" catalog delivered with the monitor. Other
cord. cables, batteries, transducers and accessories may cause a
• To avoid the risk of electric shock, this equipment must only be safety hazard, damage the equipment or the system, result in
connected to a supply mains with protective earth. increased emissions or decreased immunity of the equipment or
• Do not use an additional multiple socket outlet or extension cord. system or interfere with the measurement. Protection against
• After transferring or reinstalling the monitor, always check that it cardiac defibrillator discharge is due in part to the accessories
is properly connected and all parts are securely attached. Pay for pulse oximetry (SpO2), temperature (T) and invasive pressure
special attention to this in case of stacked mounting.
(P) measurement.
• If you accidentally drop the monitor or modules, have them
• Single-use accessories are not designed to be re-used. Re-use
checked by authorized service personnel prior to clinical use.
may cause a risk of contamination and affect the measurement
• Vibrations during intrahospital transport may disturb SpO2, ECG,
accuracy.
impedance respiration, NIBP and InvBP measurements. • Do not incinerate a battery or store at high temperatures, as it
• Some equipment malfunctions may not generate a monitor will explode.
alarm. Always keep the patient under close surveillance. • The monitor or its components should not be used adjacent to or
• To avoid explosion hazard, do not use the monitor in presence of stacked with other equipment. If adjacent or stacked use is
flammable anesthetics. necessary, the monitor and its components should be observed
• Do not use the monitor in high electromagnetic fields (for to verify normal operation in the configuration in which it will be
example, during MRI.) used.
• Do not connect any external devices to the system other than • Connecting electrical equipment together or using the same
those specified. extension cord for more than one device may cause their
leakage currents to exceed the limits specified in relevant safety
standards. Always make sure that the combination complies
with the international safety standard IEC 60601-1-1 for medical
Cautions
electrical systems and with the requirements of local authorities.
• Pins of connectors identified with the ESD warning symbol A CAUTION indicates a situation in which the unit or
should not be touched. Connections should not be made to devices connected to it may be damaged.
these connectors unless ESD precautionary procedures are
used. See "Safety precautions: ESD precautionary procedures" in • Before connecting the power cord to the power supply, check
the"User's Reference Manual" for details. that the local voltage and frequency correspond with the rating
• Other equipment may interfere with the system, even if that stated on the device plate.
other equipment complies with CISPR emission requirements. • Leave space for circulation of air to prevent the monitor from
• If liquid has accidentally entered the system or its parts, overheating.
disconnect the power cord from the power supply and have the • Refresh the batteries completely every six months (see "Cleaning
equipment serviced by authorized service personnel. and care").
• If the unit fails to respond as described, do not use the monitor • Do not store or use the monitor outside the temperature and
until tested and repaired by authorized service personnel. humidity ranges specified in the "Performance" section of this
• The system is intended for use by qualified medical personnel manual.
only. • After replacing a battery, always make sure that you close the
• Before cleaning, disconnect the monitor from the power supply. battery compartment by sliding the lid back to the right until it
• Use only an intact power cord. Replace the power cord if it is clicks.
cracked, frayed, broken or otherwise damaged. • Perform regular functional testing of each of the parameters and
• Do not apply tension to the power cord otherwise the cord may accessories
get damaged. • The monitor display is fragile. Ensure that it is not placed near a
• The power cord may only be connected to a three-wire, heat source or exposed to mechanical shocks, pressure,
grounded, hospital grade receptacle. moisture or direct sunlight.
• NOTE: The monitor is always internally powered when the • Do not immerse any part of the device in liquids or allow liquid to
batteries are connected. enter the interior.
• The B30 is always energized by the internal batteries. A short • Do not autoclave any part of the system with steam or sterilize
circuit may cause internal damage. Do not touch any exposed with ethylene oxide.
wiring or conductive surface inside, this may cause an electric • Do not apply pressurized air to any outlet or tubing connected to
shock. the monitor. Pressure may destroy sensitive elements.
• Make sure that the Pole Mount for PSMW is always used in • Before use, allow two minutes for warm-up and note any error
vertical position to prevent water from entering the E-PSM(P)W messages or deviations from normal operation.
module. • Do not short-circuit the battery terminals, this may produce a
very high current, which will damage the battery.
2
3
• Do not store or transport the monitor outside the specified − Class B equipment is suitable for use in domestic
temperature, pressure and humidity ranges. establishments and in establishments directly connected to
a low voltage power supply network which supplies buildings
Disposal used for domestic purposes.
• Dispose of the whole device, parts of it, its packing material and
this manual in accordance with local environmental and waste
disposal regulations.
Points to note
• This manual is intended for clinical professionals. Clinical
professionals are expected to have a working knowledge of
medical procedures, practices, and terminology, as required for
monitoring of critically ill patients.
• Medical electrical equipment needs special precautions
regarding electromagnetic compatibility and needs to be
installed and put into service according to the electromagnetic
compatibility information provided in the "Technical Reference
Manual" by qualified personnel.
• Portable and mobile RF communications equipment can affect
the medical electrical equipment.
• The allowed cables, transducers, accessories and mounts for the
system are listed in the "Supplies and Accessories" catalog.
• The equipment is suitable for use in the presence of
electrosurgery. Please notice the possible limitations in the
parameter sections and in the "Performance" section.
• Service and repairs are allowed for authorized service personnel
only.
• CISPR 11 classification: Group 1, Class B:
− Group 1 contains all ISM (Industrial, scientific and medical)
equipment in which there is intentionally generated and/or
used conductively coupled radio-frequency energy which is
necessary for the internal functioning of the equipment itself.
For your notes:
4
5
Symbols
− Attention, consult accompanying documents. − On the rear panel this symbol indicates the
− On the modules or frames indicates that following warnings and cautions:
modules with identical measurements should −Electric shock hazard. Do not open the cover or
not be used in the same monitor. If such the back. Refer servicing to qualified
modules have been inserted, remove the personnel.
module that has been most recently connected. −For continued protection against fire hazard,
You can also remove both modules and re- replace the fuse only with one of the same
type and rating.
connect the new module after five second.
−Disconnect from the power supply before
− On the E-PSM(P)W module indicates that servicing.
protection against cardiac defibrillator −Do not touch the monitor during defibrillation.
discharge is due in part to the accessories for −Do not use the monitor without manufacturer
pulse oximetry (SpO2), temperature (T) and approved mounting attached.
invasive pressure (P) measurement.
Type BF (IEC 60601-1) protection against electric
− On the N-FC(REC) module indicates that airway
gases should be calibrated every six months in shock
normal use and every two months in continuous
use.
− On top of the monitor beside the battery cover: Type BF (IEC 60601-1) defibrillator-proof protection
Use manufacturer recommended batteries only. against electric shock
Follow the regional regulations for disposal.
A
B B Battery operation and remaining capacity. The SN,S/N Serial number
height of the green bar indicates the charging level.
Submenu. Selecting a menu item with this symbol
opens a new menu.
Battery (A) charging (white bar)
The monitor is connected to Network.
B Battery (A) failure A blinking heart next to the heart rate or pulse rate
value indicates the beats detected.
6
7
Date of manufacture
8
9
System introduction
(1) B30 monitor with modules
(2) Printer
(3) Other monitors in the network
1
NOTE: You cannot view other monitors on the 2
B30 monitor with L-DICU08 software.
WARNING: After transferring or reinstalling the WARNING: Do not use the monitor without
monitor, always check that it is properly connected manufacturer approved mounting attached.
and all parts are securely attached. Pay special
attention to this in case of stacked mounting. WARNING: Always make sure that the audio alarm
volume level is adequate in your care environment.
10
11
Monitor introduction
(1) Battery compartment, see below
1 2 3 (2) Transportation handle
(3) Alarm light, see page 39
4 (4) Alarm LEDs, see page 37
(5) Side panel keys, see page 14
(6) The ComWheel
(7) Command Board keys, see page 13
(8) Guide rail for GCX mounting
(9) Mains power and battery LEDs, see page 15
(10) ON/standby key
(11) Defibrillator & IABP synchronization connector (marked with X5)
5 (12) Measurement modules, see page 19
12
6
11
7
10 9 8
5 6
11
9 8 7
4
12
13
1 2 3 4 5 6
10 9 8 7
(1) ON/standby key (6) For returning the Normal Screen view to the screen
(2) Mains power ON (lit) or OFF (dark): indicates mains or external DC (7) For activating pulse oximetry, impedance respiration and
power temperature setup menus
(3) For admitting or discharging a patient; for selecting user modes (8) For printing and recording different trends and waveforms
(4) For viewing trends and alarm history (9) For setting up the monitor and for activating the Help menu
(5) For activating parameter specific menus. NOTE: All modules do (10) Battery operation LEDs, see page 15
not measure all of these parameters. For more information, see
page 19
Side panel keys
Invasive Pressur
NIBP
Start
Cancel 6 For starting or stopping the NIBP manual cycling, see page 68
Recorder
For starting or stopping local recording, see page 41
7 NOTE: Functional with the N-FREC and N-FCREC modules only.
14
15
Batteries
The B30 monitor allow to use two lithium-ion batteries at most, Battery indicators
located in the battery compartment. They can be charged
separately, and screen symbols and monitor frame LEDs indicate Screen Explanation Front panel battery LEDs
their charging level and possible failure, see table on the right. You Monitor is battery Green lit
can also check the battery status through Monitor Setup - Battery powered. Batteries are Orange dark
Setup. A
B B fully charged; the size of
the green bar indicates the
If you wish to have the battery charge visible at all times, select it in charging level.
one of the digit fields: Monitor Setup - Screen Setup - Digit Fields - Monitor is battery Green lit
Battery. You can now see how much charging time is left for each powered. Battery A is Orange dark
battery separately both in numbers and as symbols, and the total A B empty, battery B charge is
charging time in numbers. ok.
NOTE: Always use the B30 monitor with battery inserted. Otherwise Monitor is battery Green lit
all trend data and temporary settings are lost if the power cable is powered. Battery A failure, Orange flashing
detached from the mains. B battery B is full.
NOTE: When the monitor is battery powered, the green battery LED
is on. When the monitor is mains powered, the green mains LED is NOTE: If both batteries fail, the green battery LED is dark.
on. See also sections "Conditioning the batteries" and "Messages."
Monitor is battery Green lit
NOTE: When useable batteries are installed, if the monitor loses AC
powered. Battery A Orange flashing
power, it automatically switches to battery power without
missing, battery B is full.
interruption of operation.
16
17
A
B
(1) Open the lid of the battery compartment (2) Lift up the battery you want to change. (3) Push in the new battery. Make sure that
located behind the transportation handle Check the indicators and messages on the charging indicator is facing forward
by sliding it to the left. screen to make sure that you change the and push the battery down all the way.
battery with lower charge. Check the monitor indicators, see above.
For your notes:
18
19
4 7
6
Module keys Module versions
The Patient Side Modules have the following measurement
Auto capabilities:
On/Off Auto On/Off: for starting or stopping the
NIBP automatic cycling, see page 68 E-PSMW:
− Non-invasive blood pressure
− Temperatures
Start Start Cancel: for starting or stopping the
Cancel − Pulse oximetry
NIBP manual cycling, see page 68 − ECG and impedance respiration
20
21
4 Module versions
The Extension Modules have the following measurement capabilities
and features:
5
N-FREC:
3
− Built-in strip chart recorder for local recording
6
N-FCREC (in the drawing):
− Built-in strip chart recorder
7 − CO2 measurement
N-FC:
− CO2 measurement
Inserting a module Using two modules
1. Align the module with the insertion guides. E-PSM(P)W and N-Fx To install an E-PSM(P)W and an N-Fx module:
modules are all inserted the same way. 1. Insert the N-Fx module first as explained in "Inserting a module."
2. Push the module into the monitor frame until it clicks: 2. Attach the E-PSM(P)W to the N-Fx.
Removing a module
1. Pull the module outwards using the tab. Make sure not to drop it
when it comes out.
22
23
Monitoring basics
Navigating in menus Using modes
A menu is a list of functions or commands. To display a menu, press The B30 monitor has six user modes. These user modes are
one of the hard keys. Selections in the menus are made with the predefined combinations of settings. They determine, for example,
ComWheel. For example, to change the ECG display: what is displayed on the screen and in trends and what the alarm
• Press the ECG key to open the function menu. limits are. In other words, by choosing a specific mode you get
• Turn the ComWheel to select the desired function in the menu. suitable settings on the screen without having to choose all features
• Push the ComWheel to open a submenu or an adjustment one by one.
window.
Modes can be hospital specific. The monitor starts in start-up mode,
• Push the ComWheel to confirm the selection.
which is one of the user modes chosen during configuration. The
default modes are STEP-DOWN, ED, PACU, CCU, NEURO and
PEDIATRIC. For more information about the default user modes, see
the "Default Configuration Worksheet" delivered with the monitor.
24
25
Battery setup
Through this menu, you can check the battery status:
1. Press the Monitor Setup key.
2. Select Battery Setup. Battery information is now available.
Changing the monitor installation settings
The monitor installation settings are the same in all user modes. The
changes are preserved until changed again. Press Monitor Setup,
enter the password and select Install/Service - Installation.
Units Printer
Change units for height, weight and blood pressure. ECG Printout Type: Select 2x6-25, 2x6-50 or 3x4-25.
Printer Connection: Select printer connection (default: None).
You can change temperature units through Others - Temp Setup Paper Size: Select A4 or Letter.
and CO2 units through Airway Gas - CO2 Setup. The changes are
permanent. NOTE: Network printer only.
26
27
28
29
30
31
32
33
34
35
Screen setup
Modifying the screen temporarily Changing the split screen contents
• Press the Monitor Setup key and select Screen Setup. You can split the Normal Screen page into two parts. The other half
• Change the waveform and digit field measurements, split screen of the split screen shows trend data.
option, minitrend length and sweep speeds. 1. Press the Monitor Setup key.
• To make other setup changes, such as scale changes, press a 2. Select Screen Setup.
parameter key and select its setup menu. For example, press the 3. Select Split Screen and choose from the options.
ECG key and select ECG Setup. • Trend shows minitrends beside waveforms.
• None shows no split screen.
Changes are valid until the monitor is turned off (+15 minutes) or
until you discharge the patient from the monitor. Only time and date
settings are stored permanently.
36
37
Alarms
Enabling the alarms Alarm indications
To enable the alarms, connect patient cables. If the alarm source is When the monitor is turned on, you will hear a beep: this tells you
selected, the alarms are operative also when the measurement is that the alarm audio signal is working. Also, the alarm LED indicators
not displayed (except the impedance respiration measurement light up for a few seconds. To check them, see "Cleaning and care:
alarms). Functioning of the alarms." You can also check the functioning of the
audio signal and alarm light through Alarms Setup - Alarm
Volume or Alarm Light.
WARNING: Always make sure that necessary alarm
limits are active and set according to the patient's When an alarm becomes active, messages appear in the order of
priority. The alarming measurement value flashes (except for low
clinical condition when you start monitoring. A priority alarms) and the color indicates the alarm category; see the
potential hazard can exist if different default alarm table below. In some cases, there may be a message on the screen
settings are used for the same or similar equipment giving more detailed information. An audible alarm is also triggered,
in any single area. and the alarm LEDs on the monitor side panel indicate the alarm
level. If enabled, also the alarm light flashes red or yellow according
to alarm levels, see below.
CAUTION: Setting alarm limits to extreme values may
NOTE: If the monitor is connected to the network, it also sends
render the alarm system useless
alarms to the Central.
Alarm categories
The priority depends primarily on the cause and alarm duration. The priority increasing with the duration and according to the physiological
significance.
Visual Meaning Tone pattern (selected Side panel LED Alarm light (if
when the system is indicators enabled)
configured)
Red For life threatening situations: HIGH Triple + double beep every 5 red LED flashing flashing red
PRIORITY ALARM seconds or continuous beep
--- -- 5 --- --/ -----
Yellow For serious but not life threatening Triple beep every 19 seconds or yellow LED flashing flashing yellow
problems: MEDIUM PRIORITY double beep every 5 seconds
ALARM --- 19 --- / -- 5 -- 5 --
White Advisory: LOW PRIORITY ALARM Single beep - yellow LED lit dark
38
39
40
41
42
43
44
45
Trends
Symbols
Trend bar: the gap shows the blood pressure mean
value
Trends view
(1) Measurement trend field
(2) Real time ECG
(3) Numerical value of a measurement at the trend cursor point
(4) Time and marker field
(5) Trend page number
Viewing graphical trends Viewing numerical trends
1. Press the Pt.Data & Trends key. 1. Press the Pt.Data & Trends key.
2. Select Trends - Graphical. 2. Select Trends - Numerical.
• To see more parameters, select Scroll Pages and scroll with • To see more pages, select Scroll Pages and scroll with the
the ComWheel. ComWheel.
• To see more data, select Cursor and scroll the page left and • To see more data, select Cursor and scroll the page up and
right with the ComWheel. down with the ComWheel.
Graphical trends contain four trend pages each having up to six Numerical trends contain four pages with the maximum of 72 hours
preconfigured fields with different parameters. Five fields can be of trend data. Real-time ECG waveform is displayed at the top of
displayed, and six fields can be printed. Real-time ECG waveform is each page.
always displayed at the top of each page.
The graphical trend time scale varies from 20 minutes to 72 hours Erasing trends
and the resolution from 10 seconds to one minute. With the 20 1. Press the Admit/Discharge key.
minute trend length, the displayed time period is 30 minutes and the 2. Select Discharge.
resolution 10 seconds. With trend lengths from one to 72 hours, the
displayed time period is 72 hours and the resolution is one minute. If the monitor has been turned off from the ON/standby key but the
For HR, ST, PVC, SpO2 and temperature measurements you can patient has not been discharged, the trend data will be stored in the
memory for 15 minutes.
select the scale in the Trend Scales menu.
46
47
48
49
Water trap in the N-FCREC and N-FC modules Conditioning the batteries
• Empty the container whenever half full. Condition batteries regularly to maintain their useful life. This is best
• Change the water trap every two months and when the text done on an external charger. Condition a battery every six months
'Replace D-Fend' appears. or when the message 'Replace Battery A' or 'Replace Battery B'
• The water trap cartridge is disposable. Do not wash or reuse the appears on the screen. Always observe the messages and symbols
cartridge. on the screen to see the battery status. You can also check the
status through Monitor Setup - Battery Setup. For more
WARNING: After cleaning, ensure that every part of information, see "Replacing the batteries", "Symbols" and
the system is dry before reconnecting it to the power "Messages."
supply. Detailed instructions for refreshing the batteries can be found in the
“User's Reference Manual.”
If the monitor is turned off, trend data and the latest user-made
settings remain in the monitor memory for 15 minutes even if the
mains power is interrupted. If not, contact service personnel. After 15
minutes, trend data is lost and the monitor returns to the user
default settings (startup mode).
Changing fuses
1. Remove the power cord if used.
2. Remove the fuse holder by pushing the locking pin and pulling
the holder gently out.
3. If a fuse is blown, replace it with a fuse of the correct type and
rating.
Regular checks
When you start monitoring, check that: NIBP
• The module is firmly in place. • Ensure that you are using correct cuff size and have selected
• Accessories are intact and properly connected. correct inflation limits. For children and when using hoses
• You have selected desired parameters to be displayed in the without identification, the inflation limit must be set manually.
digit and waveform fields. • Check that the cuff hose detection works properly.
• Check that the pressure values are displayed.
ECG and impedance respiration
• Start the Venous Stasis mode and check that the pump is not
• Check that the message 'Leads off' disappears and waveforms restarting during the measurement. If it does, the cuff may be
are displayed when the cable is connected to the patient. leaking.
Pulse oximetry Airway gas (CO2)
• Check that the red light is lit in the sensor.
• Check that the water trap is empty.
• Check that the SpO2 value is displayed and the message
• Occlude the sampling line and check that the 'Sample line
'SpO2 probe off' disappears when the sensor is connected to the blocked' message appears within 30 seconds and gas
patient. waveforms are showing zero at the same time.
Temperature
• Check that the temperature value is displayed when the probe is
connected to the patient.
InvBP
• Check that the monitor recognizes cable connections (activates
the display) for all the pressure channels used and the pressure
values are shown.
• Make sure that all transducers are zeroed correctly.
50
51
52
53
ECG and ST
You need NOTE: For a comprehensive list of accessories, see the "Supplies and
(1) E-PSMW or E-PSMPW module with ECG measurement capability Accessories" catalog.
(2) Multi-Link 5-lead ECG trunk cable, or 3-lead ECG cable with
NOTE: Keep the ECG cable, lead set and module connectors dry.
integrated leadwires
Avoid excessive use of liquids when cleaning cables and connectors.
(3) 3 or 5 leadwire set
NOTE: In 5-lead ECG, place the 5th electrode (C/V) in one of the six
ECG electrodes (pre-gelled electrodes are recommended). Check the
places indicated, and select the corresponding V lead label.
expiration date.
R=RED (IEC) L=YELLOW (IEC) R=RED (IEC) L=YELLOW (IEC)
RA=WHITE (AAMI) LA=BLACK (AAMI) RA=WHITE (AAMI) LA=BLACK (AAMI)
LEAD I
R/ RA L/ LA
LEAD III
LE
AD C=WHITE (IEC)
II V=BROWN (AAMI)
F/LL N=BLACK (IEC)
F=GREEN (IEC)
1 RL=GREEN (AAMI)
LL=RED (AAMI)
F=GREEN (IEC)
LL=RED (AAMI)
RA
R L
3
R L
R A L A
I I
I
L L
L
A
III
WARNING: Vibrations during intrahospital transport Preparing the patient and applying the electrodes
1. Prepare the skin properly to ensure optimal signal quality:
may disturb ECG measurement.
• Shave any hair from the electrode sites.
• Gently rub the skin surface to increase capillary blood flow
Connecting ECG leadwire sets to ECG trunk cables and remove dead skin cells and oil.
• Clean the skin using a mild soap and water solution.
• For 3-lead ECG, use the Multi-Link 3-lead ECG cable with
• Dry the skin completely before applying the electrodes.
integrated leadwires or connect a 3 leadwire set to the Multi-Link
2. Apply the electrodes (see figures and instructions above). Avoid
3- or 5-lead ECG trunk cable.
bones close to the skin, obvious layers of fat and major muscles.
• For 5-lead ECG, connect a 5 leadwire set to the Multi-Link 5-lead
ECG cable.
Selecting the ECG filter
WARNING: Make sure that the leadwire set clips or 1. Press the ECG key.
snaps do not touch any electrically conductive 2. Select ECG Setup - Filter:
STfilt filters high-frequency artifacts but catches slow ST
material including earth. changes.
Monit filters high-frequency artifacts and slow ST changes.
Diagn catches high-frequency changes and slow ST changes.
54
55
Selecting the number of electrodes for 5-lead ECG Selecting how to view ECG waveforms
1. Press the ECG key. • To set the number of ECG waveforms in Normal Screen, press
2. Select ECG Setup. Monitor Setup and select Screen Setup - Waveform Fields.
3. Select 5-lead Cable - 3elect or 5elect. With 3-lead ECG, one lead, and with 5-lead ECG, up to three
leads can be viewed at the same time.
Selecting the user leads • To cascade a lead, press ECG and select ECG2 Lead/ECG3 Lead
1. Press the ECG key. - Casc.
2. Select a lead for ECG1 Lead, ECG2 Lead or ECG3 Lead. • To increase ECG amplitude, press ECG and select ECG Size.
• To change the waveform sweep speed: Monitor Setup - Sweep
With 3-lead ECG, you can select only one user lead (ECG1 Lead). Speeds. Select Hemodynamics and adjust the value.
With 5-lead ECG, you can select three user leads.
NOTE: The module input circuits are protected against the effects of
Selecting a label for V Lead electrosurgery and defibrillation. However, the ECG waveform on the
With 5-lead ECG, one V lead is measured according to the placement monitor screen may be disturbed during electrosurgery.
of the V lead electrode.
Displaying ECG grid
To select a label for the lead:
To view the ECG waveforms over gridlines on the screen:
1. Press the ECG key. 1. Press the ECG key and select ECG Setup.
2. Select ECG Setup - V Lead. 2. Select Grid - ON. To view without gridlines, select OFF.
Selecting what to display with HR WARNING: When using the electrosurgery unit,
You can select what is displayed with heart rate: ensure proper contact of the ESU return electrode to
1. Press the ECG key. the patient to avoid possible burns at monitor
2. Select ECG Setup. measurement sites.
3. Select Display with HR - PR/PVC/None.
Monitoring pacemaker patients WARNING: Do not rely entirely upon rate meter
1. Press the ECG key. alarms when monitoring patients with pacemakers.
2. Select ECG Setup - Pacemaker and select one of the following: The monitor may count the pacemaker pulses as
• Show = Pacemaker spike is displayed on ECG. heartbeats. In this case, asystole and ventricular
• Sensit = Sensitive pacemaker detection; spike displayed on fibrillation may go undetected. Always keep these
ECG.
patients under close surveillance and monitor their
• Hide = Pacemaker spike is not displayed on ECG.
vital signs carefully.
NOTE: If the patient has an atrial pacer, ST calculations can be
performed if the pacer does not coincide with the ISO point's
adjustment range.
NOTE: Pacemaker detector may not operate correctly during the use
of high-frequency (HF) surgical equipment. The disturbance of HF
surgical equipment typically causes false positive pacer detection.
56
57
Arrhythmia monitoring
NOTE: With the L-DICU08 software only severe analysis is available.
It detects asystole, bradycardia, tachycardia, ventricular fibrillation
Selecting leads for the arrhythmia analysis
and ventricular tachycardia. When measuring 5-lead ECG, you can affect the selection of the two
ECG leads used for detecting beats and ventricular fibrillation. The
Setting arrhythmia alarm priority selection of user leads (ECG1, ECG2, ECG3) on the monitor affects the
leads used for detection. The first lead used for detection is lead I or
1. Press the ECG key.
II. The algorithm uses the lead appearing first in user leads. The
2. Select ECG Alarms - Arrh. Alarms - Adjust.
second lead used for detection is one of the precordial leads (V1 -
3. Select Red (high priority), Yellow (medium priority) or White (low
V6). The algorithm uses the precordial lead appearing first in the user
priority) with the ComWheel. NOTE: You can only set the priority
leads.
for V Tachy alarm.
4. Select Exit o confirm the changes and return to menu. To change the user lead:
NOTE: Alarm priorities can also be set using the Central if the Central 1. Press the ECG key.
configuration allows it. 2. Select a lead for ECG1 Lead, ECG2 Lead, ECG3 Lead.
Alarm Criteria
Asystole Cardiac arrest, no QRS complexes for five seconds
Bradycardia HR below the HR alarm limit
Tachycardia HR over the HR alarm limit
Ventricular fibrillation Fibrillatory waveform caused by ventricular fibrillation
Ventricular tachycardia Five or more consecutive PVCs and rate of successive beats over 100 bpm
58
59
Impedance respiration
You need
Improving waveform readability
Use the same setup as in the ECG measurement, see "ECG and ST."
1. Press the Others key and select Resp Setup.
2. Select Size and adjust the waveform size.
Starting
Correcting the respiration number
1. Select respiration to a waveform or digit field, otherwise
respiration data is not included in trends and no alarms are
activated. Normally, the AUTO detection limit is recommended. However, if the
2. Turn on the measurement: respirations are particularly weak or affected by artifacts, they may
not be included in the respiration rate. To ensure the correct
• Press the Others key and select Resp Setup.
respiration number, adjust detection limits closer to each other:
• Select Measurement - ON.
1. Press the Others key.
NOTE: Impedance respiration measurement is intended for patients
over three years old. 2. Select Resp Setup - Detection Limit.
3. Adjust the limits.
WARNING: Vibrations during intrahospital transport WARNING: The impedance respiration measurement
may disturb impedance respiration measurement. may cause rate changes in Minute Ventilation Rate
Responsive Pacemakers. Set the pacemaker rate
WARNING: Make sure that the leadwire set clips or responsive mode off or turn off the impedance
snaps do not touch any electrically conductive respiration measurement on the monitor.
material including earth.
WARNING: The impedance respiration measurement
WARNING: This device is not an apnea monitor is inherently very sensitive as it measures very small
system intended to alarm primarily upon the physiologic signals (changes of impedance of the
cessation of breathing. In central apnea it indicates patient's chest area). Conducted RF current above 1
an alarm after a pre-determined time since the last Vrms may cause erroneous measurements at various
breath detection. Do not attempt to use it for frequencies, for example interference with the signal/
detecting obstructive or mixed apneas, since waveform leading to respiration rate readings
respiration movements and impedance variations inconsistent with the patient's true respiration rate. If
may continue in these cases. you notice this, use another form of respiration
monitoring. For further information, see the
WARNING: When using the electrosurgery unit, "Technical Reference Manual."
ensure proper contact of the ESU return electrode to
the patient to avoid burns at measurement sites.
60
61
Temperature
You need NOTE: Use only GE Healthcare temperature probes or defibrillator-
(1) E-PSMW or E-PSMPW module proof YSI 400 series probes.
(2) Adapter cable for temperature probes NOTE: For a comprehensive list of accessories, see the "Supplies and
(3) Reusable temperature probe Accessories" catalog.
(4) Adapter cable for disposable temperature probes
(5) Disposable temperature probe
3
2
4
Changing temperature label Combining different temperatures
1. Press the Others key. The monitor displays the difference between different temperatures
2. Select Temp Setup - T1 Label or T2 Label. if they are displayed in the same digit field.
62
63
Pulse oximetry
You need NOTE: For a comprehensive list of accessories, see the "Supplies and
(1) E-PSMW or E-PSMPW module Accessories" catalog.
(2) OxyTip+ interconnect cable
NOTE: For each SpO2 accessory, refer to the instructions for use in
the accessory package for patient weight limits.
OXY-E-UN
OXY-SE-3
OXY-F-UN
1
2 OXY-W-UN
OXY-AP OXY-AF
Applying sensors Adjusting SpO2 settings
• Use dry and clean sensors only. You can adjust the volume of the beat sound, the waveform scaling
• Clean the application site. Remove nail polish, artificial and response averaging time:
fingernails, earrings etc. 1. Press the Others key.
• Clip long fingernails. 2. Select SpO2 Setup.
• Attach the sensor cable to the wrist or bedclothes to prevent the 3. Select Beat Sound Volume, Pleth Scale or SpO2 Response.
cable and sensor from moving.
64
65
WARNING: Allow sensor and cable to dry completely WARNING: Inaccurate SpO2 data can result if a sensor
after cleaning. Moisture and dirt on the connector is past its useful life. Therefore, re-evaluate the
can affect the measurement accuracy. measurement periodically by performing additional
assessment of the patient and equipment, including
WARNING: Vibrations during intrahospital transport consideration of use of alternate monitoring methods
may disturb SpO2 measurement. such as direct measurement of arterial
oxyhemoglobin saturation (SaO2).
66
67
3 4 5
1
2
To produce a single measurement
Starting
• Press the Start Cancel side panel or module key, or
Note that the measurement unit may be mmHg or kPa. The unit is
press the NIBP key and select Start Manual.
selected during configuration through Monitor Setup - Install/
Service - Installation - Units.
To measure automatically after set intervals
The monitor automatically sets inflation limit for adults to Adult
according to the hose used. For children and when using hoses • Press the Auto On/Off side panel or module key, or
without identification, the inflation limit must be set manually. To do press the NIBP key and select Start Cycling.
this: The possible intervals for autocycling are 1, 2.5, 3, 5, 10, 15, 30 or 60
minutes, 2 or 4 hours.
1. Connect the cuff hose to the NIBP connector.
• Press the NIBP key and select Cycle Time
2. Press the NIBP key.
• Select teh alternative with the ComWheel.
3. Select NIBP Setup - Inflation Limits.
NOTE: When using hoses without identification, the monitor goes
to this selection automatically when you try to start the NIBP To measure continuously for five minutes
measurement using one of the procedures described in the • Press the NIBP key and select Start STAT.
following. With these hoses, AUTO option is not available.
4. Select the limit with the ComWheel. For children, select Child.
and for infants, select Infant.
68
69
Inflation limit Maximum venous stasis inflation limit Venous stasis time
Infant 145±5 mmHg (19.3 ±0.7 kPa) 40±5 mmHg (5.3 ±0.7 kPa) 1 minute
Child 200±10 mmHg (26.7 ±1.3 kPa) 60±5 mmHg (8.0 ±0.7 kPa) 2 minutes
Adult 280±10 mmHg (37.3 ±1.3 kPa) 80±5 mmHg (10.7 ±0.7 kPa) 2 minutes
Venous stasis pressure may be lower than the values above if the WARNING: If Luer lock connectors are used in the
patient has low blood pressure. The venous stasis pressure adapts construction of tubing, there is a possibility that they
to the measured mean pressure being the same as mean pressure might be inadvertently connectd to intravascular
but always at least the following:
fluid systems, allowing air to be pumped in to a blood
Infant 20 ± 5 mmHg (2.7 ±0.7 kPa)
Child 30 ± 5 mmHg (4.0 ±0.7 kPa) vessel.
Adult 40 ± 5 mmHg (5.3 ±0.7 kPa)
70
71
1
7
6
Starting Labeling channels
1. For the setup, prepare the transducer kit according to the The label of the pressure channel sets its display scale, color, filter,
manufacturer's instructions. alarm source and alarm limits. The label descriptions are
2. Ensure that there is no air in the line. preconfigured.
3. Zero the transducer by opening it to air, pressing the Zero ALL
To change the label:
side panel key or the Invasive Pressures monitor key and
selecting Zero Pressures - Zero ALL. You can zero one channel 1. Press the Invasive Pressures key.
at a time with the module keys Zero P1 and Zero P2. Zero each 2. Select P1 Setup - Label. For factory default descriptions, see
channel. next page.
NOTE: Selecting Zero ALL does not zero ICP. Zero it separately.
NOTE: The transducer is always leveled to the mid right atrium.
Zero the transducer and pressures whenever the patient’s
position is changed.
4. Open the line to the patient.
72
73
74
75
76
77
78
79
Troubleshooting
NOTE: Always check the patient’s condition first in problematic situations or if an alarm is triggered. See also “Messages.” Also note that if the
measurement or function does not appear on the screen, check module connections.
Airway gases ECG
Values are too low: ECG signal is noisy or no QRS is detected:
• Check the sampling line and connectors for leakage. • Ensure that the patient is not shivering.
• Select the correct ECG filter through ECG - ECG Setup -
Filter.
Arrhythmia • Check the electrode quality and positioning. Do not place
Extra arrhythmias are detected. them on body hair, bones close to skin, layers of fat and
• Start relearning manually through ECG - Relearn. major muscles. Pre-gelled electrodes are recommended.
Extra Ventricular Fibrillations are detected: • Change the ECG lead.
• Check the patient. • Remove the ECG cable from the module and reinsert it.
• Check that the amplitude signals are sufficient:
Leads I and II: Select the one with the largest amplitude to
Impedance respiration
ECG 1.
Measurement fails:
After selecting the leads, start relearning manually.
• Check the electrode quality and positioning.
• Adjust the detection limits. During ventilator supported
Batteries breathing, the respiration calculation may count only
Battery operation time is markedly shortened: ventilator-produced inspirations and expirations.
• Condition the batteries, see “Conditioning the batteries” and • Remove other electrical devices from the vicinity of the
the “User's Reference Manual.” respiration measurement as they may cause interference.
Invasive pressures
Readings seem unstable:
• Make sure that there are no air bubbles in the transducer
system: Flush and zero.
• Place the transducer on the patient’s mid-heart level and
zero.
Monitor Non-invasive blood pressure
The monitor does not start: Measurement does not work or values seem unstable:
• Check that the batteries are inserted and sufficiently • Check that cuff tubings are not bent, stretched, compressed
charged, see page 17. or loose.
• Check that the power cord is properly connected. • When using hoses without identification, make sure that you
• Check the fuses and replace them if necessary, see have selected the inflation limits in the NIBP Setup menu, see
“Cleaning and care.” “Non-invasive blood pressure (NIBP).”
• Prevent motion artifacts.
• Use cuffs of correct size.
Measurements
The measured values are not displayed:
• Check that you have selected the desired parameter to a Pulse oximetry
waveform or digit field, see “Screen setup.” SpO2 signal is poor:
You cannot perform a measurement or a function. • Check the sensor and sensor positioning.
• Check that the measurement module is properly installed. • Change the SpO2 Response (averaging time) to Normal.
• Remove the module and reinstall it. • Note that skin pigment causes differences.
• Make sure that the patient is not moving.
Printing Temperature
Printing is not possible: Measurement fails:
• Check the printer setting through Print/Record - Printer • Check that you are using a correct probe.
Connection. • Try another probe.
• Check that the printer is connected to the network.
• Check the network cable.
Recording
Recording is not possible:
• Connect an N-FREC or N-FCREC module.
• Check the Central recorder if you are recording through
network.
80
81
Messages
Always check the patient first. If any problem or message persists, contact qualified service personnel.
Messages are listed here in alphabetical order.
• Check NIBP
− Check the patient status.
− Check the measurement setup.
− Check the cuff.
• Check SpO2 probe • Identical modules
− SpO2: Check the sensor and connections. − You are trying to use two or more E-PSM(P)W modules or two
or more N-Fx modules at the same time. You can only use
• Check sample gas out one E-PSM(P)W and/or one N-Fx module at a time. Remove
− Gases: Remove blockage from the sample gas outlet. extra modules.
82
83
• Zero error
− Gases: Repeat the calibration procedure.
84
85
Abbreviations
/min beats per minute, breaths per minute BAEP brainstem auditory evoked potential
°C Celsius degree Bal balance gas
°F Fahrenheit degree bar 1 atmosphere
µg microgram Beta, BE beta frequency band
Bigem. bigeminy
A alveolar BIS bispectral index
A arm (describing location) Blad bladder temperature
a arterial Blood blood temperature (C.O. measurement)
a/AO2 arterio-alveolar PO2 ratio Body body temperature
AA anesthetic agent BP blood pressure
AaDO2 alveolo-arterial oxygen difference Brady bradycardia
AAMI Association for the Advancement of Medical BSA body surface area
Instrumentation BSR burst suppression ratio
ABG arterial blood gases B-TO-B beat-to-beat
ABP arterial pressure BTPS body temperature and pressure, saturated gas
ADU Anesthesia Delivery Unit
AEP auditory evoked potential
AirW airway temperature c calculated/derived value
Alpha, Al alpha frequency band C chest
AM Anesthesia Monitor C(a-v)O2 arteriovenous oxygen content difference
Amp amplitude C.I. cardiac index
Ant. anterior C.O. cardiac output
APN apnea cal. calibration
Arrh. arrhythmia Calc calculated/derived value
Art arterial pressure Calcs calculations
ASY asystole CAM Compact Anesthesia Monitor
ATMP atmospheric pressure CaO2 arterial oxygen content
ATPD atmospheric/ambient temperature and pressure, Casc. cascaded (ECG)
dry gas cc cubic centimeter
ATPS ambient temperature and pressure, saturated CCCM Compact Critical Care Monitor
gas CCM Critical Care Monitor
aw airway CCO continuous cardiac output
AV atrioventricular CcO2 capillary oxygen content
aVF left foot augmented lead CCU cardiac (coronary) care unit
Avg. average
aVL left arm augmented lead CEL Celsius degree
aVR right arm augmented lead CISPR International Special Committee on Radio
Axil axillatory temperature Interference
CFI Cardiac Function Index
cmH2O centimeter of water
CMRR common mode rejection ratio EE energy expenditure (kcal/24h)
CO carbon monoxide EEG electroencephalogram
CO2 carbon dioxide EEG1 first EEG waveform
COHb carboxyhemoglobin EEG2 second EEG waveform
Compl compliance EEG3 third EEG waveform
Cont. continuous EEG4 fourth EEG waveform
Contrl controlled ventilation EEMG evoked electromyogram
Core core temperature EEtot total energy expenditure
Count count of responses elect electrode
CPB cardiopulmonary bypass elev. elevation
CPP cerebral perfusion pressure EMC electromagnetic compatibility
CSA compressed spectral array EMG electromyogram
CT computer tomography Enf enflurane
CvO2 (mixed) venous oxygen content Entr entropy
CVP central venous pressure EP evoked potential
ESD electrostatic discharge
d day Eso esophageal temperature
dB decibel ESV end-systolic volume
DBS double burst stimulation (NMT) ESVI end-systolic volume index
DEL delete ET, Et end-tidal concentration
Delta, De delta frequency band EtAA end-tidal anesthetic agent
depr. depression EtBal end-tidal balance gas
Des desflurane EtCO2 end-tidal carbon dioxide
Dia diastolic pressure EtN2O end-tidal nitrous oxide
Diagn diagnostic (ECG filter) EtO2 end-tidal oxygen
DIFF difference ET-tube, ETT endotracheal tube
DIS S/5 Device Interfacing Solution EVLW extravascular lung water
DO2 oxygen delivery EVLWI extravascular lung water index
DO2I oxygen delivery index exp expiratory
DSC digital signal converter
Dyn. dynamic F foot (describing location)
FAH Fahrenheit degree
e estimated FEMG frontal electromyogram
ECG electrocardiogram FFT fast Fourier transform
ECG1 first ECG waveform (top) FI Fi fraction of inspired gas
ECG1/r real-time ECG FiAA fraction of inspired anesthetic agent
ECG2 second ECG waveform Fib fibrillation
ECG3 third ECG waveform FiBal fraction of inspired balance gas
ED emergency department FiCO2 fraction of inspired carbon dioxide
EDV end-diastolic volume FiN2 fraction of inspired N2
EDVI end-diastolic volume index FiN2O fraction of inspired nitrous oxide
86
87
88
89
90
91
Performance
Alarm behavior
WARNING: Operation of the monitor outside the The maximum alarm delay
specified values may cause inaccurate results. of the alarm at the monitor
signal output to network: <5 seconds
NOTE: Information in this section may be especially useful to
clinicians. If the alarm mode is latched, the technical alarms are latched as
well. This does not comply with the NIBP (IEC 60601-2-30) and
B30 Patient Monitor invasive pressure (IEC 60601-2-34) standard requirements.
Any fluctuations within the specified limits do not affect the Silencing alarms for 5 minutes does not comply with the SpO2 (ISO
performance. 9919) standard requirements.
Power supply
Rated voltages and N-FREC, N-FCREC
frequencies: 100 to 240 V 50/60 Hz Power consumption: 3W
Allowed voltage fluctuations: ±10% Recorder type: Thermal array
Max. power consumption: 150 VA Print resolution:
Battery operation Vertical 8 dots/mm (200 dots/inch)
Batteries: Exchangeable lithium-ion, Horizontal 24 dots/mm (600 dots/inch) at a
2 pcs max. speed of 25 mm/s and slower
Charging time: 2 hours per battery pack Paper width: 50 mm, printing width 48 mm
Operation time: Up to 4.5 hours Waveforms: Selectable 1, 2, or 3 waveforms
Environmental conditions Print speed: 1, 6.25, 12.5, 25 mm/s
Operating temperature:
normal operation: +5 to +40°C (41 to 104°F)
while charging batteries: +5 to +35°C (41 to 95°F)
Storage and transport
temperature: -20 to +60°C (-4 to 140°F)
Relative humidity: 10 to 90% noncondensing,
in airway 0 to 100% condensing
Atmospheric pressure: 670 to 1060 mbar
(500 to 800 mmHg)
Defibrillator & IABP synchronization connector Hemodynamic modules E-PSMW, E-PSMPW
Analog output ECG1
ECG (Pin 8): Filter modes:
From first user lead (ECG1) With 50 Hz power supply frequency:
Gain: 1 V/mV ±10% monitoring filter 0.5 to 30 Hz
Delay: < 15 ms ST filter 0.05 to 30Hz
DC offset: ±100 mV max. diagnostic filter 0.05 to 150 Hz
Frequency response: 0.05 Hz to 40 Hz With 60 Hz power supply frequency:
Invasive blood pressure (Pin 7): monitoring filter 0.5 to 40 Hz
From pressure labeled ‘Art’ or P1 ST filter 0.05 to 40 Hz
Gain: 10 mV/mmHg ±2% diagnostic filter 0.05 to 150 Hz
Delay: < 35 ms‘ QRS minimum detection level:
DC offset: ±20 mV max. Minimum level 0.5 mV with duration between 40 and 120 ms.
Frequency response: DC to 30 Hz
The pacemaker pulses have been replaced with 2 ms ± 20% fixed Defibrillation protection: 5000 V, 360 J
digital pulses at the ECG analog output. Recovery time: <5 s according to
A device that fulfils the requirements of the IEC 60601-1 standard Heart rate:
can be connected to the defibrillator & IABP synchronization Measurement range: 30 to 250 bpm
connector. There are no other limitations, because the signals of the Measurement accuracy: ±5 % or ±5 bpm, whichever is greater
connector are galvanically isolated from patient applied part of the Displays average of 10-second median values2
ECG and invasive blood pressre measurements. Display update time: 1s
Maximum response time of heart rate meter to change in heart rate:
Synchronization pulse (Pin 1)
Response time 80 to 120 bpm: 6.9 s
Pulse width: 10 ms positive pulse Response time 80 to 40 bpm: 8.2 s
Delay: < 35 ms (R-wave peak to leading edge Maximum Tall T wave amplitude that does not disturb the heart rate
of pulse) calculation time: >1.4 mV
Amplitude: CMOS compatible Input Impedance: > 2.5 mΩ
3.5 V min. at 1 mA sourcing
0.5 V max. at 5 mA sinking
Output impedance: 50 ohm
Current limit: 10 mA
1 The isolation barrier capacitance in the module has been minimized to reduce the hazard of burns in the event of a defect in the ESU return electrode
connection.
2 When the heart rate (HR changes rapidly, the averaging is restarted.
92
93
1 The isolation barrier capacitance in the module has been minimized to reduce the hazard of burns in the event of a defect in the ESU return electrode
connection
NIBP Temperature2
Blood pressure measurement range: Measurement range: 10 to 45°C (50 to 113°F)
adult 25 to 260 mmHg (3.3 to 34.7 kPa) Measurement accuracy: ±0.1°C
child 25 to 190 mmHg (3.3 to 25.3 kPa) Digital display resolution: 0.1°C at 25 to 45°C with resuable
infant 15 to 140 mmHg (2.0 to 18.7 kPa) probles
Pulse rate range accepted: 30 to 250 bpm Digital display averaging: median of 5 samples
Cuff pressure Probe type: Use only GE Healthcare temperature
measurement range: -15 to + 350 mmHg (-2.0 to 46.7 kPa) probes or defibrillator-proof YSI 400
Typical measuring time: series probes.
adult Less than 30 s Temperature self-check: at start-up and then every 10 minutes
infant Less than 25 s
Overall system accuracy: Meets or exceeds SP10-2002 AAMI Time constant of temperature probes:
standards.1 Reusable skin temperature probe: 3s
Reusable adult central temperature probe: 6s
NOTE: The cuff pressure measurement range is equal to cuff Reusable pediatric central temperature probe: 4s
nominal and cuff indication ranges. Disposable skin temperature probe: 3 to 6 s
Disposable central temperature probe, 12F: 5 to 8 s
The ESU does not cause a burn hazard through the NIBP cuff, Disposable central temperature probe, 9F: 5 to 8 s
because there is no electrical connection between the cuff and the
NIBP measuring electronics. Esophageal stethoscope with temperature probe, 9F: 15 s
Esophageal stethoscope with temperature probe, 12F: 16s
NOTE: NIBP measurement is intended for patients weighing over Esophageal stethoscope with temperature probe, 18F3: 23 s
5 kg (11 lb). Esophageal stethoscope with temperature probe, 24F3 : 32 s
1 According to SP10-2002 AAMI 4.4.5.2.B, Intra-arterial method as the reference standard, mean difference of the test system and the comparison system shall
be ± 5 mmHg or less with standard deviation of 8 mmHg or less
2 The isolation barrier capacitance in the module has been minimized to reduce the hazard of burns in the event of a defect in the ESU return electrode
connection.
3 Response time of the probe exceeds 150 seconds.
94
95
1 Alarm limits and their adjustment range may vary depending on the mode used.
96
97
ElectroMagnetic Compatibility
Changes or modifications to this system not expressly approved by GE can cause EMC issues with this or other equipment. This system is
designed and tested to comply with applicable regulation regarding EMC and must be installed and put into service according to the EMC
information stated in this section.
98
99
Voltage dips, short <5% UT <5% UT Mains power quality should be that of a
interruptions and (>95% dip in UT) (>95% dip in UT) typical commercial or hospital
voltage variations on for 0.5 cycle for 0.5 cycle environment. If user of the equipment
power supply lines requires continued operation during
IEC 61000-4-11 40% UT 40% UT power mains interruptions, it is
(60% dip in UT) (60% dip in UT) recommended that the equipment be
powered from an uninterruptible power
for 5 cycles for 5 cycles
supply or a battery.
70% UT 70% UT
(30% dip in UT) (30% dip in UT)
for 25 cycles for 25 cycles
<5% UT <5% UT
(>95% dip in UT) (>95% dip in UT)
for 5 sec for 5 sec
Power frequency (50/ 3 A/m 3 A/m Power frequency magnetic field should be
60 Hz) magnetic field at levels characteristic of a typical location
IEC 61000-4-8 in a typical commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
100
101
d = 2.3 P
102
103
The B30 monitor is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the equipment can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the equipment as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum Separation distance according to frequency of transmitter
output power of m
transmitter
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
W
d = 1.2 P d = 1.2 P d = 2.3 P
d = 3.5 P (1
104
105
106
107
Warranty
This Product is sold by GE Medical Systems (China) Co., Ltd. under the GE's sole and exclusive obligation and Buyer's sole and exclusive
warranty set forth in the following paragraphs. Such warranty is remedy under the above warranty is limited to repairing or
extended only with respect to the purchase of this Product directly replacing, free of charge, at GE's option, a Product, which is
from GE or GE's Authorized Dealers as new merchandise and is telephonically reported to the nearest GE office or GE's Authorized
extended to the Buyer thereof, other than for the purpose of resale. Dealers office and which, if so advised by GE, is thereafter returned
For a period of twelve (12) months from the date of original delivery with a statement of observed deficiency, not later than seven (7)
to Buyer, this Product, other than expandable parts, is warranted days after the expiration date of the applicable warranty, to the GE
against functional defects in materials and workmanship and to office or GE's Authorized Dealers office during normal business
conform to the description of the Product contained in this manual hours, transportation charges prepaid, and which, upon GE's
and accompanying labels and/or inserts, provided that the same is examination, is found not to conform to the above warranty. GE
properly operated under the conditions of normal use, that regular shall not be otherwise liable for any damages including but not
periodic maintenance and service is performed and that the limited to incidental damages, consequential damages, or special
replacements and repairs are made in accordance with the damages.
instructions provided, using genuine parts and performed by a There are no express or implied warranties, which extend beyond
trained person. The foregoing warranty shall not apply if the Product the warranty hereinabove set forth. GE makes no warranty of
has been repaired by anyone other than GE or otherwise than in merchantability or fitness for particular purpose with respect to the
accordance with written instructions provided by GE, or altered by product or parts thereof.
anyone other than GE, or if the Product has been subject to abuse,
misuse, negligence, or accident.
For your notes:
108
World Headquarters European Representative Asian Headquarters
GE Medical Systems GE Medical Systems GE Medical Systems
Information Technologies, Inc. Information Technologies GmbH Information Technologies Asia
8200 West Tower Avenue Munzinger Straße 3-5 GE China Technology Park
Milwaukee, WI 53223 USA D-79111 Freiburg 1 Huatuo Road
Tel:+ 1 414 355 5000 Germany Shanghai 201203, P.R.China
1 800 558 5120 (US only) Tel: + 86 21 3877 7888
Fax:+ 1 414 355 3790 Fax: + 86 21 3877 7451
0459