DiST 3
• When high accuracy is required. values change over time. Fresh solution should be used for each calibration.
• At least once a month. • The probe should be rinsed with purified water each time before placing in 1413 µS/cm calibration solution,
HI70031P
calibration solution or sample to be tested. 20 mL sachets (25 pcs.)
Measurement 1413 µS/cm calibration solution,
HI98303
Press the ON/OFF button to turn the meter ON. Place the probe to the desired solution. HI7031M
230 mL bottle
The conductivity value, automatically compensated for temperature, will be shown Battery Replacement 1413 µS/cm calibration solution,
on the LCD. HI7031L
The measured temperature will be shown on the secondary LCD.
Note: Before taking any measurement make sure the meter has been calibrated. 0
500 mL bottle
EC Tester
0
IST98303N 01/17
Operation Specifications Operational Guide
Thank You To turn the meter ON and check the battery status
Thank you for choosing a Hanna Instruments product. Please read Range
0 to 2000 µS/cm
0.0 to 50.0 ˚C (32.0 to 122.0 ˚F) Press the ON/OFF button to turn the meter on. At start-up, all the LCD segments are
this instruction manual carefully before using the instrument. displayed for 1 second, then the percent indication of the remaining battery life is
1 µS/cm
For more information about Hanna Instruments and our products, Resolution
0.1 ˚C (0.1 ˚F)
displayed for another second. The meter then enters the normal measuring mode.
visit www.hannainst.com or e-mail us at sales@hannainst.com. Note: Keeping the ON/OFF button pressed while turning the meter on will
Accuracy ±2% F.S.
For technical support, contact your local Hanna Instruments display all LCD segments as long as the button is pressed.
µS/cm (@25 °C/77 °F) ±0.5 °C (±1.0 °F)
Office or e-mail us at tech@hannainst.com Calibration HI70031
Liquid Crystal Solutions (1413 µS/cm) To enter calibration mode
Find your local Hanna Instruments Office on www.hannainst.com Press the CAL button. “CAL“ message is displayed.
Display Calibration automatic, single point (1413 µS/cm)
CAL
Temperature Compensation automatic, 0.0 to 50.0 ºC (32.0 to 122.0 ºF) To enter setup mode
Remove the battery cover and press the Setup button located on the side of the battery.
Preliminary Examination Battery type CR2032 Li-Ion 3V (1 pc.)
Remove the meter from the packing material and examine it carefully to make sure that CAL button Battery life approximately 250 hours of continuous use Meter Setup
no damage has occurred during shipment. If noticeable damage is evident, contact your 0 to 50 ºC (32 to 122 ºF); While in measurement mode, remove the battery cover. Press the Setup button
local Hanna Instruments Office. Environment
RH 100% max located on the side of the battery in the battery compartment. The meter will enter
Each meter is supplied with: Weight in setup mode. Press the ON/OFF button to move through setup parameters. Change
• CR2032 battery (1 pc.) ON/OFF button (without battery)
68 g (2.4 oz.)
option by pressing CAL button. The default settings are: “Set t” measure unit - °C,
• Storage / Protection sleeve “AOFF” - 8 min.
• Instruction manual
DiST 3
• Quality Certificate
• 1413 µS/cm calibration solution sachet (4 pcs.)
Setup
Note: Save all packing material until you are sure that the instrument functions
correctly. All defective items must be returned in the original packaging with the
Button
supplied accessories.