D 1652 – 04
6. Hazards 9.2 Tetraethylammonium Bromide Solution in Glacial Ace-
6.1 Hydrogen bromide and glacial acetic acid are corrosive. tic Acid (Warning—See Section 6):
/
2
D 1652 – 04
EEW Approximate Specimen Size, g 15.2 Tetramethyl Ammonium Bromide (TEAB), solution in
170–375 0.4 glacial acetic acid.
/
375–600 0.6
15.3 Methylene Chloride (MECL).
Licensed by WEX to TAIWAN UNIVERSITY. Downloaded: 11/8/20071:21:44PM single-user license only, copying and networking prohibited.
600–1000 0.8
1000–1500 1.3 15.4 Potassium Hydrogen Phthalate (KHP).
1500–2000 1.8
2000–2500 2.3
2500–5000 2.8 16. Reagent Preparation
11.2 Add 10 to 15 mL of methylene chloride to the 16.1 Perchloric Acid (0.1 N Solution in Glacial Acetic
specimen. Insert a clean stirring bar and adjust the magnetic Acid) (Warning—See Section 6.)—Prepare in the following
stirrer to effect solution. Continue agitation through the titra- manner and sequence in order to avoid an excessive rise in
tion procedure. temperature.
11.3 Add 10 mL of tetraethylammonium bromide reagent 16.1.1 Place approximately 250 mL of glacial acetic acid
and 6 to 8 drops of crystal violet indicator solution. Titrate with into a 1 L volumetric flask. Add 13 mL of 60 % perchloric acid
0.1 N perchloric acid reagent to a sharp blue to green end point and mix. Add 50 mL of acetic anhydride, dilute to the mark
which is stable for at least 30 s. Record the volume of with glacial acetic acid, and mix thoroughly.
perchloric acid reagent used to titrate the specimen. 16.1.2 Allow to stand at least 8 h for completion of reaction
between the acetic anhydride and water. A shorter time period
12. Calculation may be used if completion of the reaction is analytically
12.1 If 10.1 (Standardization of Potassium Acid Phthalate) verified.
is used for standardization, calculate weight percent epoxide, 16.2 Tetraethylammonium Bromide Solution in Glacial Ace-
E, as follows: tic Acid (Warning—See Section 6):
E 5 4.3 3 V 3 N/W (3)
16.2.1 Dissolve, with agitation at room temperature, 100 g
of tetraethylammonium bromide in 400 mL of glacial acetic
12.2 If 10.2 (Standardization of Potassium Acid Phthalate) acid.
is used for standardization, calculate the weight percent ep-
oxide, E, as follows: 17. Standardization of 0.1 N Perchloric Acid Reagent
E 5 F 3 V/We (4) 17.1 Standardization with Potassium Acid Phthalate:
where: 17.1.1 Dissolve 0.25 g of potassium acid phthalate, weighed
We = weight of epoxy resin specimen used, g. accurately to the nearest milligram, in 50 mL of glacial acetic
12.3 Calculate the epoxy equivalent weight, WEEW, as acid, and 15 mL 20 % TEAB solution. Titrate with perchloric
follows: acid reagent solution.
17.1.2 Calculate and record the perchloric acid reagent
WEEW 5 43 3 100/E (5)
normality as follows:
where 43 = mol weight of the epoxy ring.
N 5 ~W 3 1000!/~204.2 3 V! (7)
12.4 Calculate weight percent of oxirane oxygen, O, as
follows: where:
O 5 16/43 3 E 5 1.6 3 V 3 N/W (6) N = normality of perchloric acid reagent,
W = potassium acid phthalate used, g, and
13. Precision V = volume of perchloric acid reagent required to titrate
13.1 A liquid epoxy resin sample with approximately the standard, mL.
24.1 % epoxide was tested by seven laboratories where ten
analysts obtained the following results: 18. Procedure
13.1.1 Repeatability—The difference between two results 18.1 Fill burette with 0.1 N perchloric acid solution in
obtained by the same analyst should not vary by more than glacial acetic acid.
1.22 % relative. 18.2 Flush all titrator lines to assure no air bubbles are
13.1.2 Reproducibility—The difference between two re- present.
sults, each the mean of two determinations obtained by 18.3 Add the appropriate amount of sample to a sample cup.
analysts in different laboratories should not vary by more than 18.4 Add 30 mL of MECL and 15 mL of TEAB to the
2.97 % relative. sample.
18.5 Select the correct method on the titrator and run the
Automatic Titration Method sample. The method should prompt when to enter the sample
14. Apparatus weight.
18.6 The titrator will automatically print out the results
14.1 Automatic Titrator, equipped with a 10 mL burette and using the formula in Section 19.
a pH electrode.
14.2 A Four Place Analytical Balance.
19. Titration Calculation
15. Reagents and Materials 19.1 Calculate:
15.1 Perchloric Acid, 0.1N in glacial acetic acid. R2 5 R 3 C 3 0.1000/M (8)
3
D 1652 – 04
SUMMARY OF CHANGES
Committee D01 has identified the location of selected changes to this standard since the last issue
(D 1652 - 97) that may impact the use of this standard.
(1) The range of this standard was entered into the Scope (4) Manual Titration Method – corrected the Potassium Hy-
section. drogen Phthalate Standard number.
(2) Test Method A was deleted. (5) Automatic Titration Method was added.
(3) Test Method B was renamed Manual Titration Method.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.