Вы находитесь на странице: 1из 150

CONTOURAâ

ACUTE CARE BEDS WITH


WEIGHING

MODEL
980B/985B/988B

SERVICE MANUAL

Document Ref. 746-270


Build Level 56
Issue Date 23.04.07

Huntleigh Healthcare
Nesbit Evans Products Division
Woden Road West
Wednesbury TEL:+44 (0)121 556 1511
West Midlands FAX:+44 (0)121 502 2092
WS10 7BL U.K. sales@huntne.co.uk
PRELIMINARIES
Preface
Read and understand this manual before attempting to service or repair
the equipment.

This manual is intended for use by Huntleigh Healthcare approved service


technicians. The manual may be provided to a customer in response to
customer requirements, but in no event will Huntleigh Healthcare Ltd be
responsible for any service or repair performed by customers or third parties.

Warnings, Cautions and Notes

WARNINGS given in this manual identify possible hazards in


procedures or conditions which, if not correctly followed, could result in
death, injury or other serious adverse reactions.
Cautions given in this manual identify procedures or conditions which,
if not correctly followed, could result in equipment failure or damage.
Notes given in this manual (indicated by +) are used to explain or
amplify a procedure or condition.

General Warnings

WARNING
Before starting any service or maintenance procedures, ensure
the equipment has been adequately decontaminated.
Electrical equipment can be hazardous if misused. Obey safety
instructions.
Do not use electrically-powered beds in the presence of
flammable gases such as anaesthetic agents.
The electrical power outlet at the foot end of the bed is only for
use with Huntleigh Healthcare approved pumps. Do not use it to
supply power to any other equipment.
Unless the UPS is turned off, the electrical power outlet at the
foot of the bed will remain powered at 230V a.c. even if the bed
is disconnected from the mains supply.
The bed and its sub-assemblies are very heavy and appropriate
precautions must be take to avoid injury when lifting or moving
them.

1 746-270-2
11/05
Contents

1 INTRODUCTION.............................................................................................4
1.1 About This Manual............................................................................................................ 4
1.2 Product Description........................................................................................................... 5
1.3 Operation........................................................................................................................... 7
1.4 Patient Weighing System................................................................................................. 15
1.5 Error Indications............................................................................................................. 21
1.6 Uninterruptable Power Supply (UPS)............................................................................. 23

2 OPERATIONAL MAINTENANCE.................................................................24
2.1 General ............................................................................................................................ 24
2.2 Cleaning........................................................................................................................... 24

3 PREVENTIVE MAINTENANCE ....................................................................26


3.1 General ............................................................................................................................ 26
3.2 Release Cables ................................................................................................................. 26
3.3 Castors, Brakes and Steering .......................................................................................... 26
3.4 Batteries........................................................................................................................... 27
3.5 Safety Sides...................................................................................................................... 27
3.6 Lifting Pole ...................................................................................................................... 27

4 TESTING ......................................................................................................28
4.1 Preliminary...................................................................................................................... 28
4.2 Power Operated Functions.............................................................................................. 28
4.3 Battery Operation............................................................................................................ 29
4.4 UPS Batteries................................................................................................................... 29
4.5 Manual Functions............................................................................................................ 29
4.6 Weighing System ............................................................................................................. 30

5 TROUBLESHOOTING..................................................................................31

6 SERVICING INSTRUCTIONS ......................................................................33


6.1. General ............................................................................................................................ 33
6.2. Special Tools and Equipment.......................................................................................... 34
6.3 Product Assembly Structure ........................................................................................... 35
6.4 Moulded Base Cover – Replacement............................................................................... 37
6.5 Brake Pedals – Replacement ........................................................................................... 37
6.6 Brake Adjustment ........................................................................................................... 37
6.7 Castors – Replacement .................................................................................................... 38
6.8 Height Actuator – Replacement ...................................................................................... 39
6.9 Mains Power Supply Cable – Replacement .................................................................... 41
6.10 Control Handsets – Replacement .................................................................................... 41
6.11 Control Box – Replacement............................................................................................. 42
6.12 Control Box Batteries – Replacement ............................................................................. 43
6.13 UPS Unit – Replacement ................................................................................................. 43
6.14 UPS Batteries – Replacement.......................................................................................... 45
6.15 Connection Box – Replacement....................................................................................... 46
6.16 Mains Filter Box – Replacement ..................................................................................... 47
6.17 Attendant Control Panel – Replacement ........................................................................ 48
6.18 ‘T’ Piece Cables – Replacement ...................................................................................... 48
6.19 User Interface Module – Replacement............................................................................ 49
6.20 QLC Extension Cable – Replacement............................................................................. 50
6.21 QLC – Replacement ........................................................................................................ 51
6.22 UPS Output Cable Assembly – Replacement.................................................................. 52

2 746-270-2
11/05
6.23 Backrest and Leg Actuators – Replacement ................................................................... 53
6.24 CPR Backrest Release / Adjustment Mechanism ........................................................... 54
6.25 CPR Release Cable – Replacement (beds with split safety sides) ................................... 55
6.26 Roller Buffer - Replacement (typical) ............................................................................. 56
6.27 Imaging Deck Laminate – Replacement ......................................................................... 57
6.28 SLS Switch Assembly – Replacement ............................................................................. 57
6.29 Dismantling the Bed ........................................................................................................ 60
6.30 Assembling the Bed ......................................................................................................... 66
6.31 Restoration of Paintwork ................................................................................................ 72

7 CALIBRATION .............................................................................................73
7.1 Calibration of Height/Tilt Actuators............................................................................... 73
7.2 Weighing System Calibration - Specific Gravitational Requirements. .......................... 74

8 TECHNICAL DATA ......................................................................................78

9 TRANSPORTATION & STORAGE ..............................................................80

APPENDIX A - ASSEMBLY SPARES MANUAL

APPENDIX B - DISPOSAL OF GAS SPRINGS

3 746-270-2
11/05
1 INTRODUCTION
1.1 About This Manual

This manual contains information on servicing and maintenance of the


Huntleigh Healthcare Contoura C980B series beds.

The manual comprises the following sections:


Section 1 - Introduction (this section): Includes a general description of
the equipment with an explanation of the function and operation of the
controls.
Section 2 - Operational Maintenance: Describes routine procedures and
checks to ensure correct operation during the service life of the
equipment.
Section 3 - Preventive Maintenance: Gives details of regular, periodic
maintenance actions to ensure correct operation of the equipment.
Section 4 - Testing: Defines a serviceability test to verify correct
operation of the equipment if function is suspect or following any
maintenance or servicing procedure.
Section 5 - Troubleshooting: Contains details of fault symptoms, their
possible cause and suggested actions to rectify the problem.
Section 6 - Servicing Instructions: Contains illustrated procedures for
removal and installation of replaceable parts and sub-assemblies.
Section 7 - Calibration: Gives calibration procedures for the electrical
controls and actuators and the weighing system.
Section 8 - Technical Data: Contains a list of technical data for the
equipment and components.
Appendix A - Assembly Spares Manual: Contains the Modification
Record, together with drawings and detailed parts lists for the
identification of replaceable parts.
Appendix B - Disposal of Gas Springs: Contains instructions for the safe
disposal of unserviceable gas springs.

4 746-270-2
11/05
1.2 Product Description

The Contouraâ C980B series are acute care beds designed to


accommodate patients weighing up to 250kg. The beds have multiple
patient positioning functions and other features to aid patient therapy.
A self-contained weighing system enables continuous monitoring of
patient weight and weight change.
The bed has a four-section profiling mattress platform with power driven
adjustment of bed height, two-way tilt angle, backrest angle and leg
elevation. Kneebreak angle is manually adjustable with power assistance
by gas spring.
The mattress platform incorporates a built-in bed extension and
removable (lift off) headboard and footboard. Revolving buffers are
provided on all four corners.
The bed may be supplied with single section (¾ length) safety sides or
split safety sides, the latter having two sections which may be raised or
lowered independently.
The bed is mounted on four 12.5cm castors with braking and steering
facility. The complete base of the bed is enclosed by a moulded cover.

Figure 1-1 Contoura C980B with split safety sides

5 746-270-2
11/05
On all models the power driven functions of height, backrest angle and
leg elevation are operated by push button controls. The Auto Contour
button simultaneously operates both backrest and leg elevation. The push
button handset(s) is accessible to either patient or attendant. It may be
pendant type or built-in to split safety sides.
An additional handset, for use by the attendant, is stored beneath the
Attendant Control Panel (ACP) at the foot of the bed. Tilt angle can be
controlled only from the attendant handset. The ACP enables the power
driven functions to be individually disabled. Operation of the bed by the
patient is thus at the discretion of the attendant.
The electrical actuators consist of a screw and nut assembly driven by a
24V d.c. motor and gearbox. Power for the actuators is derived from a
mains operated d.c. power supply with emergency battery back up.
A built-in power socket allows Huntleigh Healthcare air mattresses to be
connected directly to the bed. The socket is supplied from an
uninterruptable power supply (UPS) which will maintain operation of the
mattress system in the event of mains power failure or when the bed is
being moved. Some Models may be supplied without the power socket
and UPS.
The bed is equipped with an integral patient weighing system, which is
supplied with power from the UPS (if fitted).
Model C985B bed is equipped with an X-ray translucent backrest
incorporating a cassette carrier to provide the facility for chest X-ray
photography.
Model C988B bed is equipped with a full-length X-ray translucent
mattress platform, allowing head-to-toe X-ray photography to be carried
out.

6 746-270-2
11/05
1.3 Operation

1.3.1 Brakes and Steering

At the foot of the bed is a pair of linked brake pedals that operate on all
four castors.
When the pedal is pushed down, the brakes are applied on all four castors.
When the pedals are in the middle position, the brakes are free and all
four castors are able to rotate and swivel.
When the pedal is lifted up, one head end castor is engaged so that it
cannot swivel. Always engage the castor in line and trailing. The bed can
then be steered by pushing from the foot end.

Figure 1-2 Castors, Brakes and Steering

7 746-270-2
11/05
1.3.2 Safety Sides

1.3.2.1 ¾ Length sides

To lower the safety side: Press the release button (1) and lower the rails
toward the foot end (2).
To raise the safety side: Lift the top rail up and toward the head until the
latch engages at maximum height.

Figure 1-3 ¾ length safety side adjustment

1.3.2.2 Split Sides

To lower the safety side: Pull the release knob (1). Swing the side toward
the nearest end of the bed and down (2). Push the side in toward the
mattress platform (3).
To raise the safety side: Pull the side out away from the mattress
platform, lift and swing the side up until the latch engages at maximum
height.

Figure 1-4 Split safety side adjustment

8 746-270-2
11/05
1.3.3 Bed Extension

Before extending or shortening the bed, ensure that the leg sections are
lowered to the horizontal position.
To extend the bed: Apply the brakes. Lift the latch lever (1) and pull out
the foot end of the mattress platform (2) until the latch re-engages. Lift
the deck extension plate (3), pull the mattress retention rail (4) out toward
the foot of the bed and engage the plate in the outer location holes. Install
a mattress extension squab.

Figure 1-5 Extending or retracting the bed

To retract the bed: Remove the mattress extension squab. Lift the deck
extension plate (3), push in the mattress retention rail (4) and engage the
plate in the inner location holes. Lift the latch lever (1) and push in the
foot end of the mattress platform (2) until the latch re-engages.

Figure 1-6 Extending or retracting the mattress platform

9 746-270-2
11/05
1.3.4 Kneebreak Angle

When the legs are raised, the kneebreak angle can be manually adjusted
by lowering or raising the foot end of the calf section.
Hold the mattress retention rail on the foot end of the calf section, Lift the
release lever (1) and push the foot end down against the action of the gas
spring (2). Let go of the lever to lock the calf section at any angle.
To raise the foot end, lift the release lever and the gas spring will assist
the foot to rise.

Figure 1-7 Adjusting the kneebreak angle

1.3.5 Power Operation

The bed has power driven adjustment of tilt angle, height, backrest angle
and leg elevation.

1.3.5.1 Control Handsets

The patient handset, accessible to either patient or attendant, is connected


to the bed by an extendable, flexible cable. Alternatively the handset may
be incorporated in split safety sides (if fitted). An additional handset, for
use only by the attendant, is stored at the foot of the bed beneath the pull-
out tray.
When a handset push button is pressed the appropriate bed function will
operate until the button is released or the limit of travel is reached. The
Auto Contour button simultaneously operates both backrest and leg
elevation. Tilt angle can be controlled only from the attendant handset.
When reverting from a tilt condition, the bed will automatically pause in
the horizontal position. Then it will continue in the same direction unless
the push button is released.
10 746-270-2
11/05
Backrest

Leg
Elevation

Height

Auto
Contour

Figure 1-8 Patient Handset

Leg Backrest
Elevation

CPR Button

Tilt

Height

Leg
Elevation
Battery Tilt and
Backrest Indicator Height

Figure 1-9 Attendant handset and ACP

11 746-270-2
11/05
1.3.5.2 CPR Control Button

There is a CPR button on the attendant handset. Press and hold this button
to automatically return the bed to a flat, horizontal, low height condition.
Before using the CPR button, ensure that all lock out switches on the
ACP are in the vertical, unlocked position.

1.3.5.3 Attendant Control Panel (ACP)

To gain access to the ACP, lift and pull out the sliding tray at the foot of
the bed. (This tray also carries the user interface module (UIM) for the
weighing system and storage for the attendant handset).

Figure 1-10 UIM and ACP with sliding tray open

To prevent operation by the patient, the individual bed functions can be


disabled by turning the appropriate lock-out switch so that the pointer is
horizontal. In this condition the selected function is disabled on both the
patient handset and the attendant handset.

1.3.5.4 Calibration Error

When operating the function buttons, if a bleeping sound is heard and the
function is inoperative this indicates an actuator calibration error. Refer to
Section 7, Calibration.

1.3.5.5 Batteries

The emergency back-up battery is automatically switched on if the bed is


disconnected from the mains power supply. The batteries will enable

12 746-270-2
11/05
emergency power operation for short periods, e.g. when the bed is being
moved.
To ensure that the batteries are kept fully charged, the bed must be
connected to the mains power supply at all times during normal use.
If, when the bed is operated using the batteries, a bleeping sound is heard,
this indicates a low charge condition. The bed should be re-connected to
the mains power supply, for a minimum of 24 hours, in order to recharge
the batteries.
The battery charge indicator on the ACP illuminates when the batteries
are being charged.

1.3.6 CPR Emergency

1.3.6.1 Single section ¾ length safety sides

On beds equipped with single section, ¾ length safety sides, a CPR


emergency release handle is provided on both sides of the bed on the
mattress deck.
To lower the backrest, lift the CPR release handle at the side of the bed
(1) and push the backrest down (2).

WARNING
The backrest will be free to fall quickly (2). Keep hands clear to
avoid trapping.

Figure 1-11 CPR emergency release handle

13 746-270-2
11/05
In addition to the CPR handle there is also a CPR button on the attendant
hand set Refer to paragraph 1.3.5. Power Operation.

1.3.6.2 Split safety sides

On beds equipped with split safety sides, a CPR emergency release


handle is provided on both sides of the bed at the top of the backrest.
To lower the backrest, lift the CPR release handle at the top of the
backrest (1) and push the backrest down. (2).

WARNING
The backrest will be free to fall quickly (2). Keep hands clear to
avoid trapping.

Figure 1-12 CPR emergency release handle

In addition to the CPR handle there is also a CPR button on the attendant
handset. Refer to paragraph 1.3.5. Power Operation.

14 746-270-2
11/05
1.3.7 X-ray Facilities (where fitted)

1.3.7.1 X-ray Backrest cassette tray (model 985B only)

The backrest x-ray cassette carrier provides the facility for chest x-ray
photography.
For information regarding use refer to the User Manual, publication
number 746-182.

1.3.7.2 Full length x-ray mattress platform (model 988B only)

To use the full-length x-ray facility the mattress platform must be in the
flat, horizontal position.

Caution
Operation of the bed profile while using the X-ray cassette tray
will cause damage to the product.

For information regarding use of the x-ray facility, refer to the User
Instructions, publication number 746-194 (for C888/ACC1 dual X-ray
cassette trays) or 746-211 (for C988B/ACC1 X-ray cassette with
positioning pole).

1.4 Patient Weighing System

1.4.1 General

Four load cells are installed in the bed frame. The electrical output signals
from the load cells are integrated in a load converter (QLC) which
provides a single output to the user interface. Separate displays are
provided for patient weight and patient weight change.

1.4.2 Accuracy

The weighing system is very sensitive and will be affected by a number


of outside influences. It is therefore important that the following criteria
are observed.
a) The bed must be installed on a flat and level floor.
b) The transport screws must be fully disengaged prior to use (Figure
1-13).
c) The bed should be positioned clear of obstructions such as walls,
furniture, trailing leads or curtains.
d) The weighing system should be used within the normal operating

15 746-270-2
11/05
ambient temperature range of 10°C to 40°C.
e) For best results, the patient should be positioned approximately
centrally on the bed.
f) The weighing system is capable of detecting relatively small
movements of the patient. It is therefore necessary for patient
movement to be minimised during the few seconds required to
obtain a stabilised reading and that the operator / nurse stands clear
of the bed during this time.
g) When weighing, the handsets must be replaced in position on the
platform.

Figure 1-13 Transport locking screw

1.4.3 User Interface

Lift and pull out the sliding tray beneath the foot end of the mattress
platform to gain access to the UIM.

16 746-270-2
11/05
1.4.4 Operation of the Weighing System

1.4.4.1 Setting up

Situate the bed on a flat and level floor and apply


! the brakes. Fully unscrew all four transport
screws.
Add the mattress and bed linen to the bed. The
mattress may be either foam type or an air
mattress system complete with pump. Do not
install any additional equipment at this stage.
Connect the bed to a suitable mains power supply.
When first switched on, the system will perform a
self test sequence during which the upper display
(E) will indicate CoLd. After a few minutes this
message will disappear and the system becomes
operational.
Zero the upper and lower displays.

17 746-270-2
11/05
Figure 1-14 User Interface Module (UIM) panel

Item Description Function


A Reset Push Button Resets upper display to show gross weight
Resets lower display to zero
B Enter Push Button Used to enter data on lower display
(patient weight change)
C Enter Push Button Used to enter data on upper display
(patient weight)
D Lower Display Shows patient weight change
E Upper Display Shows patient weight
F System Stable Indicator Confirms system is stable (OK to read display)
G Numerical Keypad Used for manual input of weight or
weight change
H Clear Push Button Clears last keypad entry
J Minus or Decimal Push Enters decimal point following entry of any
Button digit, OR enters minus (-) if no digit previously
entered.
K Set Zero Push Button Used to zero display
L Auto Compensation Push Allows addition or removal of objects from the
Button bed without affecting subsequent readings
M Display ON / OFF Turns display on or off

18 746-270-2
11/05
1.4.4.2 Zero Upper or Lower Display
Press the adjacent Enter Push Button.
The display will read zero.
Press the Zero Push Button.
The display will continue to read zero.
Press the Enter Push Button again.
The display will stabilise after a few seconds
and indicate 0.0.

+
If the displays do not zero correctly or if a series of horizontal
dashes appear in the middle of the display, refer to Error Indications
(1) Zeroing Error.

1.4.4.3 Patient Weight


Place the patient on the bed. When the system
stabilises, the indicator (F) will illuminate and
the upper and lower displays will indicate
patient weight.

+ If a series of horizontal dashes appear at the top of the display,


refer to Error Indications (2) Overload Error.

1.4.4.4 Patient Weight Change


Re-zero the lower display.
After this any change in patient weight will be
indicated on the lower display.
After recording a weight change, either re-zero
the lower display to monitor further change or
leave the display to monitor change
continuously.

1.4.4.5 Auto-Compensation

In order to add (or remove) additional equipment, pillows, bed linen or


other objects to the bed without affecting subsequent weight readings, the
auto-compensation feature must be used.

19 746-270-2
11/05
Press either Enter Push Button.
The adjacent display will read zero.
Press the Auto- Comp Push Button.
The adjacent display will continue to read zero,
the other display will continue to indicate its
original figures.
Press either Enter Push Button.
The upper display will read AUTO-COMP
The lower display will be blank.
Wait until the System Stable Indicator
illuminates.
The display is now in auto-compensation mode.
Add or remove objects as required.
Repeat the above sequence of push button
operations.
Both displays will revert to continuous weight
display.

+
If, after adding a number of heavy accessories, a series of
horizontal dashes appear at the bottom of the display refer to Error
Indications (3) Auto-Compensation Error.

1.4.4.6 Manual Weight Input

At any time the power supply to the bed may be lost e.g. due to power
failure or if the bed has to be moved. If the UPS is switched on and
functional this will not affect operation of the weighing system. If no UPS
is installed or it is not functional, when supply is restored the upper
display will indicate gross weight, which is the combined total for patient
weight and added equipment. The lower display will indicate 0.0. In this
circumstance, the last recorded patient weight can be entered manually.

Press the upper Enter Push Button.


The display will read zero.
Enter known patient weight on the keypad.
(e.g. 132.4). The display will read 132.4
Press the upper Enter Push Button again.
The display will stabilise after a few seconds
and indicate 132.4
After a manual input of weight, the symbol U will appear at the left-hand
side of the display. This symbol indicates that a manual input has been
made and it will not disappear until the weighing system is disconnected

20 746-270-2
11/05
from its power supply or the reset button is pressed.
If the reset button is pressed, the display will
revert to showing gross weight. (Refer to Error
Indications (3) Auto-Compensation Error).

1.4.4.7 Confidentiality Mode


The display can be turned off or on by pressing
the display On/Off button.

1.4.4.8 Cancelling an Error


If an error is made when entering data on the
keyboard, press the Clear Push Button to delete
the error, then enter correct data.
If an error is made when entering a command
sequence (e.g., Set Zero is pressed instead of
Auto-Comp), press the button again to de-
select the command then continue the sequence.

1.5 Error Indications

1.5.1 (1) Zeroing Error

A series of horizontal dashes along the centre of the display indicates a


zeroing error.
In order to maintain accuracy over its weighing range, the system can
only be set to zero within 50kg of the original calibrated zero. The dashes
indicate that an excessive amount of equipment has been added to the bed
prior to setting the initial zero. This is why it is recommended that only
the mattress and bed linen are added to the bed prior to setting zero for
the bed.
To clear this error: remove all equipment from the bed except mattress
and bed linen. Disconnect the weighing system from the power supply
(turn off the UPS) for a few seconds. Reconnect the power supply and
then re-zero the displays.
When installing additional equipment on the bed, use the auto-
compensation facility.

1.5.2 (2) Overload Error

A series of horizontal dashes along the top edge of the display indicates
21 746-270-2
11/05
that the safe working load of 250kg has been exceeded.
This means either the patient is too heavy for the bed or an excessive
amount of equipment has been added to the bed prior to placing the
patient on the bed.
If the latter is the case, remove equipment from the bed using the auto-
compensation facility.

1.5.3 (3) Auto-Compensation Error

A series of horizontal dashes along the lower edge of the display indicates
that the allowable limit for added equipment of 100kg has been exceeded.

To clear this error, press the Reset button. The


display will revert to showing gross weight.
To proceed, remove all additional equipment
from the bed, thereby returning to a display
showing patient weight. Then, using the auto-
compensation facility, replace all the additional
equipment previously fitted.
Alternatively, remove the last piece of
equipment added and manually input the patient
weight.

If, when the Reset button is pressed, a series of horizontal dashes appear
along the top edge of the display, this indicates an overload condition.
Refer to (2) above.

22 746-270-2
11/05
1.6 Uninterruptable Power Supply (UPS)

The UPS is located in an enclosure on the base of the bed at the head-end. It
provides power at 230V a.c. to the outlet socket at the foot-end of the bed as
well as to the patient weighing system. When the bed is disconnected from the
mains power supply the UPS will continue to provide this power derived from
its own internal batteries.
To turn the UPS off: Open the cover (1) on the patient’s left-hand side of the
enclosure and press the on/off button (2).

Figure 1-15 UPS on/off switch


To turn the UPS on: Open the cover (1), press and hold the on / off button (2)
for approximately 10 seconds. The button contains an LED that will be green in
normal circumstances.

An audible warning beep sounds when the mains power is


disconnected and the UPS is in operation. If this occurs, restore
+ mains power to the bed as soon as possible. If the unit continues to
beep when the mains power is restored, contact Huntleigh
Healthcare.

If the UPS has been in use and its internal batteries have become discharged,
reconnect to the mains power supply then turn the UPS off and back on again as
described above. This will reinitialise the UPS and, over an eight-hour period,
recharge the batteries to full capacity.

23 746-270-2
11/05
2 OPERATIONAL MAINTENANCE

2.1 General

The equipment has been designed to be virtually maintenance free


between service intervals. The extent of periodic cleaning and
maintenance necessary will be determined by use and condition.
Examine the bed for obvious signs of damage. If found take appropriate
remedial action.

2.2 Cleaning

WARNING
Disconnect the bed from the mains power supply and switch off
the UPS before cleaning.

Cleaning Wearing suitable protective clothing, clean all surfaces with


a disposable cloth soaked in a neutral detergent and hand hot
water.
The head and foot end panels can be removed from the bed
for cleaning. If the bed has detachable mattress platform
sheets these can also be removed for cleaning.
Start by cleaning the upper sections of the bed and work
along all horizontal surfaces. Work methodically towards the
lower sections of the bed and clean the wheels last. Take
extra care to clean areas that may trap dust or dirt.
Rinse with clean water and dry with disposable paper towels.
Allow the cleaned parts to dry before replacing the mattress.

24 746-270-2
11/05
Disinfecting After cleaning the bed as described above, wipe all surfaces
with Sodium Dichloroisocyanurate (NaDCC) at a
concentration of 1,000 parts per million (0.1%) of available
chlorine.
In the case of pooling body fluids, e.g. blood, the
concentration of NaDCC should be increased to 10,000 parts
per million (1%) of available chlorine.
Refer to Huntleigh Healthcare’s Customer Infection Control
Procedures.

Caution
Do not use abrasive compounds or pads as these may damage
the paint finish.
Do not use phenol-based disinfectant solutions.

25 746-270-2
11/05
3 PREVENTIVE MAINTENANCE

The following preventive maintenance checks and procedures should be


carried out every 12 months.
If the result of any check is unsatisfactory, take the appropriate remedial
action.

WARNING
Before commencing any maintenance activity, disconnect the
bed from the electrical power supply and switch off the UPS.
Avoid skin contact with lubricants. Gloves and protective
clothing should be worn when carrying out maintenance work.

3.1 General

Examine the bed for obvious signs of damage. Make sure that all nuts,
bolts and other fastenings are secure and not missing.
Examine the flexible mains and handset cables for cuts, abrasions, kinks
or other deterioration. Check the power connector plug for damage.
Check that the adapter fuse is of the correct type and rating (BS 1362, 5
Amp).
If the power connector plug or cable is damaged, they must be replaced as
a complete assembly.

3.2 Release Cables

After long use, release cables may stretch. This could prevent proper
operation of the CPR release.
To remedy this, adjust the two lock nuts on the inner cable at the release
lever end.
Refer to Section 6, sub-section 6.24 and 6.25.

3.3 Castors, Brakes and Steering

Check the brakes for efficient operation. Apply the brakes and push the
bed. If any of the castors rotate, the brake is not fully effective. Set the
brake pedal in the steering position and verify that one head end castor is
engaged so that it cannot swivel.

26 746-270-2
11/05
Brakes that fail to lock effectively indicate that the castor requires
adjustment or replacement. Refer to Section 6, sub-section 6.6 and 6.7.

3.4 Batteries

If, when the bed is operated using the batteries a bleeping sound is heard,
this indicates a low charge condition. The bed should be re-connected to
the mains power supply, for a minimum of 24 hours, in order to recharge
the batteries.
Check the condition of the batteries using the following procedure:
1. Ensure that the bed has been connected to the mains power supply for
24 hours.
2. Disconnect the bed from the mains power supply.
3. Apply a load of 80 kg (= 12 stone person) to the mattress platform.
4. Raise and lower the mattress platform, from minimum to maximum
height, three times.
5. If the bed does not operate satisfactorily in step 4, perform steps 1 to 4
again.
If the bed again fails to operate satisfactorily, the batteries are not holding
sufficient charge and they should be replaced. Refer to Section 6, sub-
section 6.12.

3.5 Safety Sides

Check the safety sides for obvious signs of damage. Make sure that all
nuts, bolts and other fastenings are secure and not missing. Make sure
that the mechanism that locks the safety side in the raised position
engages positively.

3.6 Lifting Pole

Examine the handle and strap for signs of damage, fraying or loose
stitching. Replace the complete strap and handle assembly if any defect is
found.

27 746-270-2
11/05
4 TESTING

The following serviceability tests should be performed before returning


the bed to use after service, or if any function is suspect.
These instructions should be read in conjunction with Section 1, sub-
section 1.3.
If the result of any test is unsatisfactory, investigate and take the
appropriate remedial action. Refer to Section 5, Troubleshooting and
Section 6, Servicing Instructions.

+ All references to left or right of the bed are when viewed from the
head end.

4.1 Preliminary

a) Verify that no furniture or other obstruction can impede the


movements of the bed.
b) Connect the bed to the mains power supply.
c) Verify that the battery charge indicator (on the attendant control
panel) is illuminated.
d) Verify that the rotary switches on the ACP are in the vertical
(unlocked) position.
e) Level the mattress platform (zero tilt).

4.2 Power Operated Functions

a) Using the patient handset, check the function of bed height,


backrest angle and leg elevation. Verify satisfactorily operation
over the full range of movement as specified in Section 8,
Technical Data. Verify that the actuators are de-energised when the
push button is released or the limit of travel is reached.
b) Repeat step a) using the second patient handset (if fitted).
c) Repeat step a) using the attendant handset. Verify that the lock-out
switches on the ACP disable the push buttons on both the patient
and attendant handsets.
d) Check the function of the tilt control using the attendant handset.
Verify satisfactory operation over the full range of movement as
specified in Section 8, Technical Data. Check that, on reaching the
horizontal condition, the mechanism will pause and then resume
movement until the push button is released.
e) Position the bed at near maximum height with both backrest and
legs raised and apply foot down tilt. Press and hold the CPR button.
Verify that the bed reverts to flat, horizontal, low height condition.
28 746-270-2
11/05
4.3 Battery Operation

a) Disconnect the bed from the mains power supply.


b) Check the functions of bed height, backrest angle, leg elevation and
two-way tilt using the attendant handset. Verify that the patient
handsets are also operative.
c) Re-connect the bed to the mains power supply. Verify that the
battery charge indicator on the ACP is illuminated.
d) Test the batteries as described in sub-section 3.4.

4.4 UPS Batteries

A battery test is carried out while the UPS unit is in start-up phase, then
automatically every seven days. If a battery fault is detected, the battery
indicator light on the front panel will flash and a warning beep will be
heard at one second intervals.
Recharge the battery (connect UPS to mains power supply) for 8 hours. If
the fault persists, replace the batteries. Refer to Section 6, sub-section
6.14.

4.5 Manual Functions

4.5.1 Brakes and Castors

a) Refer to paragraph 1.3.1. Check the brakes for efficient operation.


Apply the brakes and push the bed. None of the castors should
rotate.
b) Set the brake pedal in the steering position and verify that one head
end castor is engaged so that it cannot swivel.

4.5.2 Safety Sides

a) Refer to paragraph 1.3.2. Check operation of single section ¾


length safety sides (if fitted). Verify that the release button is
effective and the sides are firmly locked when raised.
b) Check operation of all four sections of split safety sides (if fitted).
Verify that the spring release latch mechanism is effective and that
the sides are firmly locked when raised.

4.5.3 Backrest CPR Release

a) Using any push button handset, raise the backrest to approximately


60°.
29 746-270-2
11/05
b) Refer to paragraph 1.3.6. Lift the release handle and push the
backrest down. Verify that the release mechanism is effective and
the backrest reclines in a smooth, controlled manner with minimum
force applied.

4.5.4 Kneebreak

a) Using any push button control panel, raise the legs to maximum
height.
b) Refer to paragraph 1.3.4. Verify that the foot end of the calf
section can be raised and lowered over the full range of movement
as specified in Section 8.
c) Lift the foot end to an intermediate position and check that the gas
spring is firmly locked. Lower the foot end again.
d) Using push button controls, lower the legs. Check that the
kneebreak angle returns to zero as the legs are lowered to
horizontal.

4.5.5 Bed Extension

Refer to paragraph 1.3.3. Check the operation of the bed extension.


Verify that the spring release and latching mechanism is effective.

4.6 Weighing System

a) Refer to sub-section 1.4 and zero both displays.


b) Place a ‘patient’ on the bed and make a note of the weight
indicated on the upper display.
c) Re-zero the lower display.
d) Add a known weight to the mattress platform. Verify that this
weight is shown on the lower display and the appropriate increase
is indicated on the upper display.
e) Select auto-compensation mode. Remove the additional weight and
revert to continuous weight display. Verify that the weight shown
on the top display is the same as was indicated in step d).
f) Remove ‘patient’ and weights from the bed and zero both displays.
g) Verify that manual data input is functioning by entering values on
both upper and lower displays.
h) If appropriate, calibrate the weighing system as described in
Section 7, Calibration.

30 746-270-2
11/05
5 TROUBLESHOOTING

The following table identifies some fault symptoms, their possible cause
and suggested remedial action(s).
Test
Symptom Possible cause Action
step
None of the actuators Power disconnected or Make sure power supply
will work batteries discharged lead is connected

Fuse blown Check fuse in adaptor


4.2
Line fuse blown Check the inline fuses
F1, F2 in the mains filter
box
One actuator does not work Actuator plug disconnected Check that actuator plug
4.2 is fully inserted in the
control box
Battery charge Power disconnected or Check and rectify
indicator not fuse blown
illuminated
ACP disconnected Check that plugs are fully
4.3
inserted in the control box

Check by substitution of
Battery charger fault control box
Brakes or steering Wear or damage Adjust or replace castor
4.5.1
lock not effective
Backrest CPR Release mechanism Check and re-adjust
release not effective not correctly adjusted
4.5.3
Movement too easy
not damped Faulty gas spring Replace gas spring
No weight display No supply to UPS Check line fuses in
connection box

4.6 UPS turned off Check and rectify

No supply to QLC Check fuse F5 in mains


filter box
Very erratic weighing Bed obstructed Check for obstructions or
or large weighing tight power cables
4.6 errors
Transport screws not Check screws are fully
fully unlocked unlocked
Slightly erratic or ‘Patient’ moving Get ‘patient’ to lie
small weighing errors down and keep still

Bed touching wall, Check and rectify


4.6
curtains etc.

System due for Refer to Section 7


recalibration

+ Refer to Section 6 for instructions on how to access the control


box, mains filter box and mains connection box.

31 746-270-2
11/05
5.1 Weighing System Error Messages

If the display on the User Interface Module shows an error message


Err1 to Err4, this indicates that there is no communication between the
load cell and the UIM.
Check the wiring between the load cell(s), the QLC and the UIM. If no
fault found, contact Huntleigh Healthcare.

32 746-270-2
11/05
6 SERVICING INSTRUCTIONS

WARNING
Before starting any service or maintenance procedures, ensure
that the equipment has been adequately decontaminated.
Avoid skin contact with lubricants. Rubber gloves and protective
clothing should be worn when carrying out maintenance work.
Heavy equipment hazard. The bed and its sub-assemblies are
very heavy and could cause serious injury or death by crushing.
When working beneath the platform, the equipment must be
properly supported with suitable lifting devices, blocks or
stands.
Unless the UPS is turned off, the electrical power outlet at the
foot of the bed and wiring in the mains filter box will remain live
at 230V ac even if the bed is disconnected from the mains power
supply.
Components of the bed can move under battery power even if
the bed is disconnected from the mains power supply.
Do not connect any cable (except the SLS switch, if fitted) to
Port 5 of the control box as this may cause the mattress
platform to lower unexpectedly. Do not remove the blanking
plug (if fitted).

6.1. General

The servicing procedures given in this section include instructions for:


! Replacement of specific components and sub-assemblies
! Dismantling and assembly of the main structure
! Restoration of paint finish
Procedures which may be considered self-evident have not been included.
Do not disassemble more than is necessary to replace a defective item.
Remove old Loctite screw adhesive and other contamination from
components before re-assembly. Apply Loctite screw adhesive to screw
threads on assembly of items as specified in Appendix A. Before
applying Loctite, clean and degrease both internal and external screw
threads using a suitable volatile solvent e.g. methylated spirits.
Apply Rocol Sapphire grease to bushes and joints on assembly of items
as specified in Appendix A.
33 746-270-2
11/05
Throughout this section, numbers given in parenthesis thus (02-14) refer
to the Assembly Spares Manual drawing sheet and item number for the
part.

+ All references to left or right of the bed are when viewed from the
head end.

Before the bed is returned to use following service or repair:


! Carry out a visual inspection for signs of damage
! Perform a serviceability test as given in Section 4
! Clean the equipment using the method described in Section 2

6.2. Special Tools and Equipment

In addition to normal workshop tools, the special tools and equipment


listed below will be required for servicing.

Item Description Part No. Use


1 Hoist SWL 250kg L6J/01/250JDS General
2 Sling strap (2) 745.08 General
3 Brake cam rotating tool 745.04 Fitting castors
4 Clip rotating tool 745.03 Removing / installing
radius arms
5 Clip removal tool 745.01 Removing / installing
radius arms
6 Dummy load adapter (set of 4) 791.267 Installing load cells

6.2.1 Using the Hoist

When using the hoist to lift any part of the bed, always ensure that the
sling straps are applied to main frame sections and do not bear upon
components or minor fabrications.
When lifting the entire bed or the deck assembly, first remove all deck
sheets and lift off the head end and the foot end. (See note). Pass both
sling straps around the main longitudinal members of the deck frame.
Apply the straps inboard of the deck hinges, one between the back and
seat sections and the second between the thigh and calf sections.

On C988B beds with full-length x-ray facility, the deck sheets are
not removable. Ensure that the deck is in the flat, horizontal
+ condition and pass the sling straps around the deck assembly
ensuring that they pass beneath the main longitudinal members of
the deck frame. Use suitable stretcher bars if necessary.

34 746-270-2
11/05
6.3 Product Assembly Structure

Each bed comprises a number of sub-assemblies and components as


defined by the Assembly Spares Manual and Modification Record
(ASMMR) Appendix A.
The Model number (Ref No) and Serial number are marked on a label
attached to the base of the bed.
The Model number e.g. C980BS/E etc defines the configuration of the
bed. The first group of four digits is the basic Model number :
C980B - Standard
C985B - X-ray backrest
C988B - Full length x-ray mattress platform
The next letter defines the safety sides fitted:
N - No safety sides
S - Split safety sides
L - ¾ length safety sides
The following groups of letters define various available options that were
incorporated when the bed was built.
The Serial number and the ASMMR will identify the applicable
modification / build standard of the bed.

35 746-270-2
11/05
16

Mains
230V ac

1 Transformer 6 QLC 11 Height Actuator


2 Mains Filter Box 7 UIM 12 Battery Box
3 Connection Box 8 ACP 13 UPS Outlet Socket
4 Control Box 9 Backrest Actuator 14 Load Cells
5 UPS 10 Leg Actuator 15 Control Handsets
16 SLS Switch

Figure 6-1 Block wiring diagram

36 746-270-2
11/05
6.4 Moulded Base Cover – Replacement

If the moulded base cover is to be removed for the purpose of access to


base assembly components, it can be lifted and temporarily secured to the
deck assembly with cords or wire hooks.
6.4.1 Removal

a) Raise the deck to maximum height.


b) Remove the four button head socket screws (01-34). The cover
(01-6) can now be lifted clear of the base. Secure it to the deck
with cords or wire hooks.
c) To completely remove the base cover from the bed, it is first
necessary to separate the deck assembly from the base assembly
and remove the leg fabrications as described in 6.28.8 and 6.28.9.

6.4.2 Installation

a) Installation of the base cover (01-6) is by the reversal of the above


procedure. Refit the screws (01-34).

6.5 Brake Pedals – Replacement

6.5.1 Removal

a) Support the brake pedal (01-16) (01-17) with a suitable anvil block,
drive out and discard the spirol pin (01-21). Pull the brake pedal
(01-16) (01-17) off the brake bar (01-48). Remove the spacer (01-
41).

6.5.2 Installation

a) Slide the spacer (01-41) on to the brake bar (01-48). Install the
brake pedal (01-16) (01-17) on the brake bar (01-48) ensuring
correct orientation. Align the holes in the pedal and brake bar.
Support the brake pedal with a suitable anvil block and drive in a
new spirol pin (01-21).

6.6 Brake Adjustment

a) Remove the base cover and tie it to the deck as described in 6.4.

37 746-270-2
11/05
b) Bend back the shakeproof pressing (01-56) and remove two
hexagon head screws (01-55) and the pressing.
c) On the brake/free castors (01-11) turn the bearing housing ‘A’
clockwise in 180° steps until the brake is more effective.
Alternatively, on the tracking castor (01-54), turn the bearing
housing anti-clockwise until the steering lock is more effective.
d) Do not over-adjust by turning the housing too many times.
e) Fit a new or serviceable shakeproof pressing (01-56) and refit two
hexagon head screws (01-55). Bend up the pressing to lock the
screws.

(01-42)
(01-48)

(01-55)

(01-56)

(01-11)
(01-54)

Figure 6-2 Castor installation

f) Refit the base cover as described in 6.4.


g) If satisfactory adjustment cannot be achieved, replace the castor.

6.7 Castors – Replacement


6.7.1 Removal
a) At the foot end, remove the brake pedals (01-16)(01-17) as
described in 6.5. At the head end, Remove two button head screws
(01-20) and washers (01-19) from the ends of the brake bar (01-
42). Raise the base cover as described in 6.4.
b) Use the hoist (Tool 1) to raise the bed off the floor by

38 746-270-2
11/05
approximately 150mm. Support the outer base frame on suitable
blocks.
c) Remove two nuts (01-45) and washers (01-46) and release the
castor link tubes (01-7) from the castor levers (01-44). Do not lose
the nylon washers (01-47).
d) Slide the brake bar (01-48) or (01-42) through just far enough to
clear the castor tube. Bend back the shakeproof pressing (01-56),
remove two hexagon head screws (01-55) and withdraw the castor
from the base fabrication. The castor bush (01-12) should remain in
the base fabrication.

6.7.2 Installation

a) Insert the stem of the castor into the castor bush (01-12) within the
base fabrication and align the holes. Use the brake cam rotating
tool (Tool 3) to verify that the brake is actuated by rotation in the
appropriate direction. If not, turn the stem of the castor through
180°. Slide the brake bar (01-48) or (01-42) through the castor
stem.
b) Install a new or serviceable shakeproof pressing (01-56) and two
hexagon head screws (01-55). Bend up the pressing to lock the
screws.
c) Install the nylon washers (01-47) and the castor link tubes (01-7)
on the castor levers (01-44) and secure with washers (01-46) and
new nuts (01-45).
d) Install the base cover as described in 6.4. Install the brake pedals as
described in 6.5.

6.8 Height Actuator – Replacement


6.8.1 Removal

a) Remove the head bow and the foot bow from the bed. Remove the
deck sheets from the deck assembly. Remove the base cover as
described in 6.4.
b) If the height actuators are functional: Raise the deck to maximum
height and use the hoist to support it. Disconnect the bed from the
mains power supply.
c) If the height actuators are not functional and bed is at low height:
Disconnect the bed from the mains power supply. Use the hoist to
39 746-270-2
11/05
support the deck and tilt the bed on one side far enough to gain
access to the actuator(s). Remove the ‘E’ ring (01-10), washer (01-
9) and clevis pin (01-13) to release the actuator ram from the radius
arm. Set the bed back on its castors and use the hoist to raise the
deck to maximum height.
d) Refer to the Electrical Assembly Schematic, Sheet 04 and
disconnect the actuator connector from the control box. Spring off
the covers from the rigid conduit on the deck frame and pull out the
actuator cable.
e) If not already done in c) above, remove the ‘E’ ring (01-10),
washer (01-9) and clevis pin (01-13) to release the actuator ram
from the radius arm.
f) Remove the ‘E’ ring (01-10), washer (01-9) and clevis pin (01-8) to
release the actuator (01-15) from the base frame.

6.8.2 Installation

a) If fitting a new actuator, refer to Electrical Assembly Schematic,


Sheet 04 and temporarily connect the actuator to the control box.
Use the control hand set to extend the actuator to near maximum
length.
b) Position the actuator, ensuring correct orientation and attach the
body to the base frame with the clevis pin (01-8), washer (01-9)
and ‘E’ ring (01-10).
c) Use the hoist to lower or raise the deck to align the hole in the ram
with the hole in the radius arm.
d) Attach the ram to the radius arm with clevis pin (01-13), washer
(01-9) and ‘E’ ring (01-10).
e) Lightly lubricate the ‘O’ ring on the actuator connector with
petroleum jelly (Vaseline). Refer to the Electrical Assembly
Schematic, Sheet 04 and connect the actuator to the control box.
Lay the cable into the rigid conduit on the deck and replace the
covers.
f) Install the base cover as described in 6.4. Remove the hoist.
Replace the deck sheets, head bow and foot bow.
g) Calibrate the actuators as described in Section 7.

40 746-270-2
11/05
6.9 Mains Power Supply Cable – Replacement

6.9.1 Removal

a) Raise the deck to maximum height. Raise the base cover as


described in 6.4. Disconnect bed from mains power supply.
b) Release the mains cable and strain relief bush from the bracket at
the head end of the base frame. Remove the cable clamp securing
the cable to the base frame. Spring off the covers from the rigid
conduit on the base frame and pull out the cable.
c) Remove four screws (05-8) and washers (05-4) from the top of the
transformer, remove the lid of the transformer case and disconnect
the mains cable from the transformer input terminals.

6.9.2 Installation

a) Connect the new mains power supply cable assembly to the


transformer input terminals. Refit the lid of the transformer case
and attach with four screws (05-8) and washers (05-4).
b) Position the cable in the rigid conduit and refit the cover.
c) Install the cable and strain relief bush in the bracket. Make sure
that the cable is not strained and secure it to the base frame with the
cable clamp.
d) Install the base cover as described in 6.4.

6.10 Control Handsets – Replacement


6.10.1 Removal

a) Roving patient handsets are removed by simply disconnecting the


flexible cable from the socket on the deck frame.
b) To remove the attendant handset, cut and remove cable ties as
necessary to release the flexible cable and disconnect it from the
socket beneath the seat section on the deck frame.
c) Patient handsets embedded in split safety sides are factory fitted
items and can only be renewed by replacing the complete safety
side assembly. Remove the safety side assembly as described in
6.28.4.

41 746-270-2
11/05
6.10.2 Installation

a) Connect the patient handset to the appropriate socket on the deck


frame.
b) Connect the attendant handset to the appropriate socket beneath the
seat section on the deck frame. Arrange the flexible cable tidily,
ensuring that it cannot snag moving parts and secure with cable
ties.
c) For patient handsets embedded in split safety sides, install a new
safety side assembly as described in 6.29.9.

6.11 Control Box – Replacement


6.11.1 Removal

a) Raise the bed to maximum height and raise the base cover as
described in 6.4.
b) Refer to Electrical Assembly Schematic, Sheet 4 and disconnect all
electrical cables from the control box.
c) Remove four nuts (22-5) and screws (22-11) and remove the
control box (22-16) from the mounting plate (22-17).

6.11.2 Installation

a) Position the control box (22-16) on the mounting plate (22-17) and
attach with four screws (22-11) and new nuts (22-5).
b) Refer to Electrical Assembly Schematic, Sheet 4 and reconnect all
cable plugs to the control box. To assist insertion of the actuator
connectors, lightly lubricate the ‘O’ rings with petroleum jelly
(Vaseline).
c) Refit the base cover as described in 6.4.
d) Calibrate the actuators as described in Section 7.

42 746-270-2
11/05
6.12 Control Box Batteries – Replacement
6.12.1 Removal

WARNING
To avoid short-circuiting the battery, remove rings, watches and
other metal objects. Use tools with insulated handles.
Lead acid batteries are a potential hazard to health and the
environment. Store batteries in accordance with the
manufacturer’s instructions. Dispose of unserviceable batteries
safely in accordance with local regulations.

a) Raise the bed to maximum height and raise the base cover as
described in 6.4. Disconnect bed from mains power supply. Switch
off the UPS.
b) Remove four screws and remove the lid of the battery compartment
on the control box (22-16).
c) Disconnect the batteries and remove them from the battery
compartment.

6.12.2 Installation

a) Position the new batteries in the compartment in the control box,


ensuring correct orientation. Reconnect the electrical wires in
accordance with the instruction label in the lid.
b) Refit the lid and attach with four screws.
c) Refit the base cover as described in 6.4.

6.13 UPS Unit – Replacement


6.13.1 Removal

a) Raise the bed to maximum height. Disconnect the bed from the
mains power supply. Switch off the UPS.

43 746-270-2
11/05
C

A B

Figure 6-3 UPS Power Connections

b) Remove four screws (22-39) and remove the lid (22-1) from the
control box and UPS assembly (01-57).
c) Refer to Figure 6-3. Disconnect the mains input cable connector
(A) from the UPS. Disconnect both output cable connectors (B)
from the UPS.
d) Lift the UPS unit (22-13) off the UPS brackets (22-24) and remove
it from the control box and UPS assembly.

6.13.2 Installation

a) Position the UPS unit (22-13) on the UPS brackets (22-24) inside
the control box and UPS assembly (01-57).
b) Check that the UPS is switched off and connect both the output
cables (B).
c) Connect the UPS mains input cable (A). Switch on the UPS and
verify output power is being supplied to the pump outlet and the
weighing system.
d) Replace the lid (22-1) on the control box and UPS assembly (01-
57) and attach with four screws (22-39).

44 746-270-2
11/05
6.14 UPS Batteries – Replacement
6.14.1 Removal

WARNING
To avoid short-circuiting the battery, remove rings, watches
and other metal objects. Use tools with insulated handles.
Store batteries in accordance with the manufacturer’s
instructions.
Dispose of unserviceable batteries safely in accordance with
local authority regulations.

a) Remove the UPS from the bed as described in 6.13.


b) Refer to Figure 6-3. Use a small flat-bladed screwdriver to lever
off the rear right hand UPS foot (C). Swing open the battery
compartment door.
c) Pull out the two batteries. Unplug the black (-) and red (+) UPS
cables from the batteries. Disconnect and retain the short jumper
cable that connects the two batteries.
d) Remove the unserviceable batteries for safe disposal or recycling.

6.14.2 Installation

a) Connect the red (+) UPS cable to red (+) terminal on one battery.
Connect the black (-) UPS cable to the black (-) terminal of the
other battery.
b) Using the short jumper cable retained in 6.14.1 c), link the
unconnected black (-) terminal of the first battery to the
unconnected red (+) terminal of the second battery.

+ A small spark may occur when connecting the UPS batteries.


This is normal.
c) Push the two connected batteries back into the UPS enclosure
ensuring that wires are not trapped.
d) Close the battery compartment door and refit the UPS foot (C).
e) Install the UPS on the bed as described in 6.13.

45 746-270-2
11/05
6.15 Connection Box – Replacement
6.15.1 Removal

a) Remove the UPS as described in 6.13.1.


b) Remove four screws securing the lid of the connection box (22-18)
and remove the lid. Disconnect all the wires from within the
connection box. Slacken the cable glands and pull the three cables
out of the connection box.
c) Disconnect the earth bonding strap from the side of the connection
box. Remove the four screws (22-11) securing the connection box
to the base of the electrical enclosure (22-32).
d) Lift the connection box out of the electrical enclosure and remove
it from the bed.

6.15.2 Installation

a) Position the connection box (22-18) on the bed and secure it from
below the electrical enclosure base (22-32) with four screws (22-
11).
b) Connect the earth bonding strap to the earth terminal on the side of
the connection box. Pass the three cables through their respective
cable glands and reconnect all the cables within the connection box
– refer to Figure 6-4 for wiring details.
c) Check that the cables and wires are not trapped or strained at any
point and tighten the cable gland nuts to secure the three cables.
d) Refit the connection box lid and secure with four screws.
e) Install the UPS unit (22-13) as described in 6.13.2.

Blue Blue
Power in
Brown Brown Power out to
from isolating
control box
transformer Gn/yel
Gn/yel
Blue Power out to
Brown
UPS

Figure 6-4 Connection box wiring

46 746-270-2
11/05
6.16 Mains Filter Box – Replacement
6.16.1 Removal

a) Remove the UPS unit as described in 6.13.1.


b) Remove four screws and remove the lid from the mains filter box
(22-19). Slacken off the four cable glands, disconnect the wires and
pull the four cables out of the mains filter box.
c) Disconnect the earth bonding strap from the earth terminal on the
side of the mains filter box.
d) Remove the four screws (22-11) securing the mains filter box to
the electrical enclosure base (22-32) and remove the mains filter
box from the bed.

6.16.2 Installation

a) Refer to the connection diagram Figure 6-5. Thread the cables


through the cable glands into the mains filter box and connect the
wires to the correct terminals.

To UPS To QLC

Brown
Blue

To isolating Brown
transformer E N L
Black

Red

Blue
F1 4AT

F2 4AT

F5 2AT

To mains Brown
connection
box Blue

0V
A
+VE

Figure 6-5 Mains filter box - connections

b) Make sure that the cables and wires are not strained and tighten the
cable gland nuts. Connect the earth bonding strap to the earth stud
(A) on the mains filter box and secure with nut and washer.
c) Fit the lid on the mains filter box and attach with four screws. Place
the box in position on the electrical enclosure base (22-32) and
attach with four screws (22-11).
d) Install the UPS unit (22-13) as described in 6.13.2.
47 746-270-2
11/05
6.17 Attendant Control Panel – Replacement
6.17.1 Removal

a) Raise the bed to maximum height and raise the base cover as
described in 6.4. Disconnect the bed from the mains power supply.
Remove the deck panels.
b) Refer to Electrical Assembly Schematic. Sheet 4 and disconnect
the ACP connecting plug from the control box (22-16). Spring off
the covers from the rigid conduit on the base frame and pull out the
cable.
c) Tie a length of cord to the plug and pull the cable out of the
flexible conduit (01-5). Untie the cord from the cable leaving the
cord in the conduit. Cut and remove cable ties as necessary to
release the cable from the deck frame.
d) Remove two nuts (03-9) and screws (03-8) and remove the ACP
(03-28) from the foot end assembly.

6.17.2 Installation

a) Position the ACP in the tray (03-28) of the foot end assembly and
attach with two screws (03-8) and washers (03-9).
b) Secure the cable to the deck frame ensuring that it is not strained
and cannot snag moving parts. Use the cord to pull the cable
through the flexible conduit (01-5). Remove the cord.
c) Refer to Electrical Assembly Schematic, sheet 4 and connect the
ACP cable to the control box (22-16). Lay the cable in the rigid
conduit on the base frame and replace the conduit cover.
d) Refit the base cover as described in (6.4).

6.18 ‘T’ Piece Cables – Replacement


6.18.1 Removal

a) Raise the bed to maximum height and raise the base cover as
described in 6.4. Disconnect the bed from the mains power supply.
Remove the deck panels.
b) Refer to Electrical Assembly Schematic, sheet 4 and disconnect the
long ‘T’ piece cable (04-2) from the control box (22-16). Spring off
the covers from the rigid conduit on the base frame and pull out the
cable.
48 746-270-2
11/05
c) Tie a cord to the cable and pull it out of the flexible conduit (01-5)
into the deck area, leaving the cord in the conduit.
d) Disconnect the handset and the short ‘T’ piece cable. Remove four
screws (04-4), nuts (04-5) and washers (04-6) to release the cable
sockets from the bed.
e) Cut and remove cable ties as necessary to release the short ‘T’
piece cable from the deck frame.
f) Disconnect the patient handset and the attendant handset. Remove
four screws (04-4), nuts (04-5) and washers (04-6) to release the
cable sockets from the bed.

6.18.2 Installation

a) Attach the fixed sockets on the short ‘T’ piece cable (04-3) to the
deck frame and secure with four screws (04-4), nuts (04-5) and
washers (04-6).
b) Attach the fixed sockets on the long ‘T’ piece cable (04-2) to the
deck frame and secure with four screws (04-4), nuts (04-5) and
washers (04-6).
c) Secure the cables to the deck frame ensuring that they are not
strained and cannot snag moving parts. Use the cord to pull the
long cable through the flexible conduit (01-5). Remove the cord.

Figure 6-6 Attach cord to pull in the cable (actuator connector shown)

d) Refer to Electrical Assembly Schematic, sheet 4 and connect the


long cable to the control box (22-16). Arrange the cable in the rigid
conduit on the base frame and replace the conduit cover.
e) Refit the base cover as described in (6.4).

6.19 User Interface Module – Replacement


6.19.1 Removal

a) Disconnect the UIM extension cable connector from the fixed


49 746-270-2
11/05
socket on the QLC extension cable (04-1). Cut and remove cable
ties to release the UIM cable.
b) Remove four pan head screws (03-10) to release the UIM from the
tray (03-32) on the foot end assembly.

6.19.2 Installation

a) Position the UIM in the tray (03-32) and attach with four pan head
screws (03-10).
b) Connect the UIM extension cable connector to the fixed socket on
the QLC extension cable (04-1). Secure the UIM extension cable
with cable ties.

6.20 QLC Extension Cable – Replacement


6.20.1 Removal

a) Raise the bed to maximum height and raise the base cover as
described in 6.4. Disconnect the bed from the mains power supply
and switch off the UPS. Remove the deck panels.
b) Remove four screws (22-39) and remove the lid (22-1) from the
control box and UPS assembly (01-57). Remove four screws and
the lid of the QLC (22-10).
c) Refer to the wiring diagram Figure 6-1, identify the QLC extension
cable (UIM cable) and disconnect the electrical wires from the
terminals. Loosen the cable gland and pull the cable out of the box.
d) Spring off the covers from the rigid conduit on the base frame. Tie
a length of cord to the cable and pull the cable out of the flexible
conduit (22-15) and the rigid conduit. Pull the cable out of the
flexible conduit (01-5) up to the deck frame. Untie the cord from
the cable leaving the cord in the conduit. Cut and remove cable ties
as necessary to release the cable from the deck frame.
e) Remove four screws (04-4), nuts (04-5) and washers (04-6) to
release the cable assembly from the bed.

6.20.2 Installation

a) Install the QLC extension cable on the deck frame and attach the
fixed socket with four screws (04-4), nuts (04-5) and washers (04-
6).

50 746-270-2
11/05
b) Secure the cable to the deck frame with cable ties, ensuring that it
is not strained and cannot snag moving parts. Use the cord to pull
the cable through the flexible conduit (01-5) and (22-15) into the
control box and UPS assembly enclosure. Remove the cord.
c) Arrange the cable in the rigid conduit on the base frame and
replace the conduit cover. Pass the cable through the gland into the
QLC enclosure.
d) Refer to the wiring diagram Figure 6-1 and connect the electrical
wires to the terminals. Tighten the cable gland, replace the lid on
the QLC and secure with four screws.
e) Calibrate the weighing system as given in Section 7.

6.21 QLC – Replacement


6.21.1 Removal

a) Remove the UPS unit (22-13) from the bed as described in 6.13.1.
b) Remove four screws and remove the lid from the QLC (22-10).
c) Disconnect all the electrical wires from the terminals. Loosen the
cable glands and pull the cables out of the box.
d) Remove the two screws (22-11) and Nyloc nuts (22-5) attaching
the QLC bracket (22-12) to the control box and UPS assembly (01-
57). Lift out the QLC, with the bracket attached, and remove it
from the bed.
e) Remove two screws (22-4) and Nyloc nuts (22-5) to detach the
QLC bracket (22-12) from the QLC.

6.21.2 Installation

a) Attach the QLC (22-10) to the QLC bracket (22-12) with two
screws (22-4) and Nyloc nuts (22-5). Attach the QLC and bracket
assembly to the control box and UPS assembly (01-57) with two
screws (22-11) and nuts (22-5).
b) Refer to the Electrical Assembly Schematic, Sheet 4 and thread the
cables through the cable glands into the QLC box. Refer to the
connection diagram Figure 6-7 and connect the wires to the
terminals. Ensure that the wires and cables are not strained and
tighten the cable glands.
c) Refit the lid on the QLC and secure with four screws. Refit the

51 746-270-2
11/05
UPS unit (22-13) as described in 6.13.2.
d) Calibrate the weighing system as described in Section 7.

Priority Load Cell Vex 10VAC Vin 30mV I ex 200mA

Figure 6-7 QLC Mk. IV - connections

6.22 UPS Output Cable Assembly – Replacement


6.22.1 Removal

a) Raise the bed to maximum height. Disconnect the bed from the
mains power supply. Turn off the UPS. Raise the base cover as
described in 6.4.
b) Remove four screws (22-39) and the lid (22-1) from the control
box and UPS assembly. Disconnect the output cable connector
from the UPS.
c) Remove the lid (24-2) of the plug box at the foot of the bed,
disconnect the cable wires from the terminals and remove the
socket (24-3) together with the lid (24-2). Loosen the cable gland
(24-13) and pull the cable out of the box.
d) If it is necessary to replace the box, remove two screws (24-5), and
remove the plug box base from the deck frame.
e) Pull the cable out of the tube (24-12). Spring off the cover from the
rigid conduit on the base frame. Tie a cord to the cable and pull the
cable out of the flexible conduit (01-5), the rigid conduit and the
flexible conduit (22-15). Leave the cord in the conduit.

6.22.2 Installation

a) If removed, install the plug box base (24-1) on the deck frame and
attach with two screws (24-5). If removed, assemble the socket

52 746-270-2
11/05
(24-3) onto the plug box lid (24-2).
b) Use the cord to pull in the new cable through the flexible conduit
(22-15) and (10-5) up to the deck frame. Untie the cord and thread
the cable through the tube (24-12). Thread the cable through the
cable gland (24-13) into the box and connect the wires to the socket
as shown on sheet 24. Check that the gland is tight and the wires
are not strained and re-fit the lid (24-2) on the box base (24-1).
c) Arrange the cable on the deck assembly ensuring that it is not
strained and cannot snag moving parts. Lay the cable in the rigid
conduit on the base frame and replace the conduit cover.
d) Check that the UPS is switched off and connect the output cable to
the UPS.
e) Replace the UPS bracket lid (13-2) and attach with four pan head
screws (13-9) and two self-tapping screws (13-10).

6.23 Backrest and Leg Actuators – Replacement


6.23.1 Removal

a) Raise the backrest and leg sections. Raise the bed to maximum
height and raise the base cover as described in 6.4. Disconnect the
bed from the mains power supply and switch off the UPS. Remove
the deck panels.
b) Refer to Electrical Assembly Schematic, sheet 4 and disconnect the
actuator cable from the control box. Spring off the covers from the
rigid conduit on the base frame and pull out the actuator cable.
c) Tie a length of cord to the plug and pull the cable out of the
flexible conduit (01-5). Untie the cord from the cable leaving the
cord in the conduit. Cut and remove cable ties as necessary to
release the cable from the deck frame.
d) Use the hoist (Tool 1) to support the backrest or leg sections as
appropriate.
e) If removing the backrest actuator, disconnect the CPR release
mechanism as described in 6.24.
f) Remove ‘E’ ring (02-18), two washers (02-17) and clevis pin (02-
16) to release the actuator ram from the moving component. Do not
allow the actuator to drop.

53 746-270-2
11/05
Some beds may be fitted with an additional damper. In this case the

+ backrest actuator ram and the damper are both secured by clevis
pin (02-82), washers (02-78) and ‘E’ ring (02-79), reference
drawing 804.76.
g) Support the actuator and remove ‘E’ ring (02-18), two washers (02-
17) (12-23) and clevis pin (02-16) to release the actuator from the
bed.

6.23.2 Installation

a) Position the actuator, ensuring correct orientation and attach the


body to the deck frame with clevis pin (02-16), two washers (01-
17) and ‘E’ ring (02-18).
b) Use the hoist to raise or lower the section and align the hole in the
ram with the hole in the moving component. Attach the ram with
clevis pin (02-16), two washers (01-17) and ‘E’ ring (02-18).
c) On the backrest actuator, connect the CPR release mechanism as
described in 6.24.
d) Secure the cable to the deck frame ensuring that it is not strained
and cannot snag moving parts. Use the cord to pull the cable
through the flexible conduit (01-5). Remove the cord.
e) Lightly lubricate the ‘O’ ring on the actuator connector with
petroleum jelly (Vaseline). Refer to Electrical Assembly
Schematic, sheet 4 and connect the cable to the control box (22-
16). Arrange the cable in the rigid conduit on the base frame and
replace the conduit cover.
f) Refit the base cover as described in 6.4.
g) If the backrest actuator has been replaced, adjust the CPR release
mechanism as described in 6.24.

6.24 CPR Backrest Release / Adjustment Mechanism


6.24.1 Removal

a) Loosen two locknuts (02-26)(23-23) on the adjustment pin (02-


24)(23-22). Remove the ‘E’ ring (02-30)(23-27) to free the
assembly from the CPR release fabrication (02-31)(23-28). The
adjustment pin can now be disconnected from the actuator.

54 746-270-2
11/05
6.24.2 Installation

a) Assemble the lever screw (02-28) (23-25) and nut (02-29)(23-26)


on the adjustment rod fabrication (02-27)(23-21). Loosely
assemble adjustment pin (02-24)(23-22) and locknuts (02-26)(23-
23).
b) Connect the assembly to the actuator and attach to the CPR release
fabrication (02-31)(23-28) with ‘E’ Ring (02-30)(23-27) and
washer if fitted.

6.24.3 Adjustment

a) Adjust the position of the pin (02-24)(23-22) by means of the


locknuts (02-26)(23-23) to achieve satisfactory operation. Adjust
the stop bolt (02-83)(23-94) and locknut (02-26)(23-23) to prevent
straining of the actuator release.

6.25 CPR Release Cable – Replacement (beds with split


safety sides)
6.25.1 Removal

a) Remove two locknuts (23-71) and the spherical washer (23-78) to


release the cable inner from the CPR handle (23-73) or (23-76).
Remove the nut (23-81) and release the cable outer from the
backrest frame.
b) Remove nut (23-71) to release cable inner from the pivot rod (23-
82). Remove the nut (23-81) and release the cable outer from the
deck frame.

6.25.2 Installation

a) Attach the cable outer to the deck frame with nuts (23-81). Set the
outer cable with minimum protrusion past the nut. Attach the cable
inner to the CPR release fabrication (23-28) with pivot rod (23-82)
and two nuts (23-71).
b) Attach the cable outer to the backrest with cable spacer (23-79) and
nuts (23-81). Attach the cable inner to the CPR handle (23-73) or
(23-76) with spherical washer (23-78) and two locknuts (23-71).

55 746-270-2
11/05
6.25.3 Adjustment

a) First adjust the actuator release mechanism as described in 6.24.


b) Adjust the nuts (17-5) or (18-32) on the cable inner or the nuts
(17-6) or (18-30) on the cable outer to give about 2 mm of free
movement at the cable inner attachment on the CPR handle.
c) Check that the actuator release operates correctly and tighten the
nuts.

6.26 Roller Buffer - Replacement (typical)


6.26.1 Removal

a) Use a large, flat bladed screwdriver to release the flanges on the


roller buffer retainer (02-7) and pull out the retaining pin (02-8).
Remove the roller buffer retainer (02-7), roller buffer (02-6) and
roller buffer cover (02-5) from the deck assembly or foot end
assembly.

(02-5)

(02-6)

(02-8)

(02-7)

Figure 6-8 Roller buffer assembly

6.26.2 Installation

a) Assemble the roller buffer (02-6), cover (02-5) and retainer (02-7)
and install them on the deck assembly or foot end assembly so the
hole in the retainer is aligned with the hole in the deck frame.
b) Install the retaining pin (02-8), ensuring that the groove engages
between the flanges on the moulding.

56 746-270-2
11/05
6.27 Imaging Deck Laminate – Replacement
6.27.1 Removal

a) Use a 5mm diameter drill, to drill out the pop rivets (02B-83)
securing the laminate to the bed section.
b) On seat section laminate (02B-41) remove two screws (02B-84),
nuts (02B-29) and nylon spacers (02B-85).
c) On thigh section laminate (02B-42), remove two screws (02B-91)
and threaded nylon spacers (02B-90).

6.27.2 Installation

a) The seat section laminate must be fixed first. Refer to the


Assembly Spares Manual, sheet 02B and ensure that the dimension
of 35mm as shown is achieved. The remaining laminates are fitted
outwards from the seat section. Ensure that there is a nominal 2-
3mm gap between edges of adjoining laminates.
b) Secure the laminate to the appropriate bed section with pop rivets
(02B-83).
c) On seat section laminate (02B-41), attach two nylon spacers (02B-
85) with screws (02B-84) and nuts (02B-29).
d) On thigh section laminate (02B-42), attach two threaded nylon
spacers (02B-90) with two screws (02B-91).

6.28 SLS Switch Assembly – Replacement

+ The SLS switch was fitted to new beds from November 2004
onwards, and was also retrofitted to some older beds.

6.28.1 Removal

a) Use the control handset to raise the deck to maximum height.


Disconnect the bed from the mains power supply and raise the base
cover as described in Paragraph 6.4.
b) Remove four self-tapping screws (22-39) and remove the electrical
enclosure lid (22-1) from the bed. Remove two screws (22-37) and
nuts (22-38) to release the SLS assembly (22-35). Collect the O-
ring (22-36).
c) Cut the SLS cable adjacent to the switch body. Unplug the SLS
cable from the control box (22-16) and pull the cut SLS through the
flexible conduit (22-15) and onto the control box mounting plate
(22-17). Remove and discard the SLS cable and switch.
57 746-270-2
11/05
6.28.2 Installation

a) Refer to Figures 6-9 and 6-10. Use a T10 Torx driver to remove
two screws from the SLS switch body and collect the long spring.
Take out the cable assembly and desolder the two wires (A) from
the microswitch (B), making a note of their colours and the
terminals to which they are attached.

+ Take care not to damage the microswitch body or the wires’


plastic insulation when using the soldering iron.

E C

B
D
H

Figure 6-9 SLS cable and microswitch assembly

G F

Figure 6-10 Complete SLS assembly

58 746-270-2
11/05
b) Cut and remove the cable tie (C) and pull off the strain relief sleeve
(D). Thread the SLS cable through the flexible conduit (22-15) into
the electrical enclosure. If necessary apply a suitable temporary
lubricant (e.g. IPC P-80) to the SLS cable.
c) Refit the strain relief sleeve (D) to the SLS cable and fit a new
2mm wide cable tie (C) to retain it in position. Solder the two wires
(A) to the appropriate microswitch terminals.
d) Apply a thin bead of clear silicone sealant all around the locating
groove (E) of the strain relief sleeve. Fit the spring (F) and install
the microswitch over the two lugs (G) with its actuator “pip” (H)
adjacent to the spring. Reassemble the SLS housing, fit the two
Torx screws and verify that the switch operates freely.
e) Plug the SLS switch cable into Port 5 on the control box (22-16)
and pull the excess cable through the flexible conduit into the
electrical enclosure.
f) Fit the O-ring (22-36) to the SLS assembly (22-35), and secure it to
the inside top face of the electrical enclosure lid (22-1) using two
screws (22-37) and nuts (22-38). Make sure that the O-ring is
secured between the lid and the SLS switch housing.

59 746-270-2
11/05
6.29 Dismantling the Bed

This sub-section gives procedures for dismantling the main structure of


the bed. It must be read in conjunction with the preceding sub-sections
which describe the removal of specific components and sub-assemblies.
Dismantling of the bed is not a routine procedure and should only be
carried out to the extent necessary.
The information given is for guidance since procedures will be varied
according to the purpose for which they are undertaken. For example,
it is not necessary to take off the deck sections in order to remove the
deck from the base.
Minor procedures which may be deemed self-evident have not been
included.

6.29.1 Foot End Assembly – Removal

a) Lift off the foot panel. Remove the thigh, and calf deck sheets (06-
3)(06-4)(06-5). Remove the attendant handset (03-26) from the
holder.
b) Remove four nuts (03-7) and shoulder bolts (03-6) and release the
ACP tray (03-32), together with the ACP and UIM, from the foot
end assembly. Alternatively, disconnect and remove the ACP and
UIM as described in 6.17 and 6.19.
c) Remove Nyloc nut (03-18) from the latch fabrication (03-19).
Remove two pan head screws (03-10) which retain the extension
bushes (03-15).
d) Lift the latch handle (03-11) and pull the foot end assembly off the
bed.

6.29.2 Safety Sides - Removal (3/4 length safety sides)

The following instructions refer specifically to L/H Safety Side


+ Assembly sheet 11. The same procedures apply for all ¾ length
safety sides.

a) Raise the safety side and use the hoist to support it.
b) Remove two countersunk socket head screws (11-15) and pivot cap
(11-16) and pull the side rail assembly off the mounting

60 746-270-2
11/05
fabrications (11-17)(11-18).
c) If necessary, remove four Nyloc nuts (11-24) and bolts (11-23) and
remove the mounting fabrications (11-17)(11-18) from the deck
frame.

6.29.3 Safety Sides - Removal (split safety sides)

+ The following instructions refer specifically to Head End Safety


Sides L/H sheet 13. The same procedures apply for all other split
safety sides.

a) Cut and remove the cable tie securing the CPR release cable to the
mounting bracket (13-19). Lower the safety side and push in,
toward the deck, to approximately mid travel.
b) Use the hoist to support the safety side assembly and remove four
button head screws (13-16) and washers (13-20).
c) Remove the complete safety side assembly from the bed. If
appropriate, return the complete assembly to manufacturer for
service or replacement.

6.29.4 Safety sides - Removal (sides with embedded control handset)

+ Control handsets embedded in split safety sides are a factory fitted


item and can only be replaced by replacing the complete safety
side assembly.

a) Disconnect the handset cable assembly (26-1)(27-1) from the


socket on the deck. Remove two self-tapping screws (26-2)(27-2)
to release the cable guide (26-3)(27-3) from the backrest.
b) Remove the safety side assembly from the bed as described in
6.28.3 above.

6.29.5 Backrest – Removal

a) Remove the deck sheet from the backrest. Raise the backrest and
use the hoist to support it.
b) Disconnect the backrest actuator ram by removing E’ ring (02-18),
two washers (02-17) and clevis pin (02-16).

61 746-270-2
11/05
+ Some beds may be fitted with an additional damper. In this case
the backrest actuator ram and the damper are both secured by
clevis pin (02-82), washers (02-78) and ‘E’ ring (02-79), reference
drawing 804.76.

c) Disconnect the CPR release cable if appropriate - reference 6.25.


Remove the head end safety sides if fitted - reference 6.28.
d) Remove two shoulder screws (02-12) to release the backrest from
the stabiliser fabrication (02-10). Remove self-tapping screw (02-
72) and shakeproof washer (02-74) to release the bonding strap
(02-71) from the backrest fabrication (02-9).
e) Drive out two spring pins (02-13) and withdraw the guide bars (02-
14). Remove the backrest from the bed.

6.29.6 Calf Section – Removal

a) Lift off the foot panel. Remove the calf deck sheet (06-4). Raise the
legs and use the hoist to support the calf section.
b) If you require to remove the extension deck sheet (06-5), pull or
drive out two spring pins (02-41) and pull the calf extension bow
(02-40) out of the bushes (02-42).
c) Remove ‘E’ ring (02-50), washer (02-49), clevis pin (02-46), plain
bushes (02-47) and headed bushes (02-48) to release the gas spring
from the deck frame. Unscrew the gas spring (02-51) from the gas
spring mounting (02-63). Remove the spacer (02-61) and rubber
washer (02-62) from the gas spring stem.
d) If necessary, remove two screws (02-65) to release the gas spring
mounting (02-63) from the calf section. Collect two headed bushes
(02-64).
e) Use a suitable flat bladed tool to lever off four nut covers (02-37).
Remove two Benton nuts (02-36), nylon washers (02-81) and pivot
bolts (02-35). Remove the calf section from the thigh section.

6.29.7 Thigh Section Removal

a) Remove the calf section as described in 6.28.6.


b) Use the hoist to support the thigh section and disconnect the leg
section actuator ram as given in 6.23.
c) Use a suitable flat bladed tool to lever off four nut covers (02-37).
Remove two Benton nuts (02-36), nylon washers (02-81) and pivot

62 746-270-2
11/05
bolts (02-35). Remove the thigh section from the bed.

6.29.8 Deck Assembly – Removal

a) Raise the deck to maximum height and use the hoist to support the
deck as described in 6.2. Disconnect the bed from the mains power
supply. Switch off the UPS.
b) Raise the base cover as described in 6.4. If not already removed,
disconnect the QLC extension cable and pull it out of the flexible
conduit (01-5) as described in 6.20. Leave the cord in the conduit.
c) Disconnect the UPS output cable and release it from the deck
frame as described in 6.22. Tie a cord to the cable and pull it out of
the flexible conduit (01-5). Leave the cord in the conduit.
d) At the control box, disconnect the ACP, backrest actuator, leg
actuator and ‘T’ piece cables and pull them out of the flexible
conduit (01-5) as described in 6.17, 6.18 and 6.23. Leave the cords
in the conduit.
e) Release the long bonding strap from the deck earthing point on the
seat section. Tie a cord to the bonding strap and pull it out of the
flexible conduit (01-5). Leave the cord in the conduit.
f) Cut and remove the cable tie securing the flexible conduit and free
the conduit from the deck assembly.

Caution
Do not allow the leg section fabrication (01-4) to fall when
released.

g) Remove four pivot bolts (23-35), washers (23-17) and pivot bushes
(23-32) and remove the deck assembly from the base assembly.

6.28.9 Radius Arms – Removal

a) Remove the deck assembly from the base assembly as described in


6.28.8.
b) Use a suitable flat bladed tool to lever off eight plastic covers (01-
63). Remove four Benton nuts (01-61), four bolts (01-64), eight
washers (01-60) and four pivot bushes (01-62) and remove the
stabilisers.

63 746-270-2
11/05
Caution
Do not allow the leg section fabrication (01-4) to fall when
released.

c) Use a suitable flat bladed tool to lever off eight plastic covers (01-
63). Remove four Benton nuts (01-61), four bolts (01-64), eight
washers (01-60) and four pivot bushes (01-62) and remove the leg
fabrications (01-4).
d) Remove the base cover (01-6) from the base assembly, together
with flexible conduit (01-5).
e) Remove the ‘E’ ring (01-10), washer (01-9) and clevis pin (01-13)
to release the actuator ram from the radius arm fabrication (01-2).

WARNING
Plastic clips may fracture and fly off at high velocity when
removed. Use a cloth or other suitable shield to prevent possible
injury.

f) Use special tool (Tool 4) to rotate the plastic clip (01-23) until the
slot is uppermost. Use the special tool (Tool 5) to push off the clip.
Discard the clip - do not re-use.
g) Remove the radius arm fabrication (01-2) by sliding to one side
and lifting it clear.

6.28.10 Transformer – Removal

a) Remove four screws (05-8) and washers (05-4) from the top of the
transformer, remove the lid of the transformer case and disconnect
the transformer output cable (04-8) and the input cable assembly
from the terminals.
b) If necessary, release the mains cable and strain relief bush from the
bracket at the head end of the bed. Remove the cable clamp
securing the cable to the base frame. Spring off the covers from the
rigid conduit on the base frame and pull out the mains cable.
c) If required, remove four screws (05-3), washers (05-4) and pillars
(05-2). Remove the transformer and mains cable assembly from the
bracket.

64 746-270-2
11/05
Caution
Do not remove the transformer bracket. Do not loosen or remove
the transformer bracket attaching parts as this could cause
misalignment of the weighing system load cells.

6.28.11 Load Cells – Removal

a) Remove four screws (22-39) and remove the lid (22-1) from the
UPS and control box assembly. Remove the lid from the QLC (22-
10) and disconnect the load cell cables Loosen the cable glands and
pull the cables out of the box. Spring off the cover from the fixed
conduit and pull out the load cell cables.
b) Release the mains cable and strain relief bush from the bracket at
the head end of the bed. Remove the cable clamp securing the
cable to the base frame.
c) Support the UPS and control box assembly on suitable wooden
blocks. Remove two nuts (01-77), bolts (01-75) and washers (01-
76). Remove two hex head screws (01-55) and washers (01-46) and
release the UPS and control box assembly from the inner frame
sub-assembly.
d) Remove four flange headed bolts (01-40) and nuts (01-38). Use the
hoist to raise the inner frame sub-assembly and remove four load
adapters (01-26) together with ‘O’ rings (01-27).
e) Remove two high tensile screws (01-30), nuts (01-32) and four
washers (01-31) and remove the load cell (01-28) and the hardened
plate (01-29) from the inner frame sub-assembly.

(01-1) (01-29) (01-28)

(01-27)

(01-26)

(01-30)
(01-31)
(01-58) (01-27) (01-32)

Figure 6-11 Load cell assembly

65 746-270-2
11/05
6.30 Assembling the Bed

6.29.1 Load Cells – Installation

+ Load cells are supplied in matched sets of four. They must never be
replaced individually.

a) With the inner sub-frame assembly supported by the hoist, attach


the load cell(s) (01-28) and hardened plate (01-29) to the pressing
fabrication (01-1)(01-71) with two high tensile screws (01-30),
four washers (01-31) and two nuts (01-32). Do not fully tighten the
nuts at this step.
b) Insert four dummy load adapters (Tool number 6) in the receptacles
in the outer base frame fabrication (01-58). Carefully lower the
inner sub-frame assembly, aligning the load cells on the dummy
load adapters. The body of the load cells should be aligned parallel
to the axis of the base frame.
c) Fully tighten the high tensile bolts (01-30) and nuts (01-32) to a
torque setting of 136Nm.
d) Raise the inner sub-frame assembly again and remove the dummy
load cell adapters. Lightly grease the adapter (01-26) and ‘O’ ring
(01-27) assemblies and install them in the receptacles in the outer
base frame fabrication (01-58).
e) Lower the inner sub-frame assembly on to the load cell adapters
and remove the hoist. Ensure that the jackscrews (01-37) are
backed off. Screw in four flange headed bolts (01-40), Adjust the
under head gap to 3mm and lock the bolts in position with nuts
(01-38). Tighten the jack screws.
f) Attach the UPS and control box assembly to the inner sub-frame
assembly with two hex head screws (01-55) and washers (01-46),
two bolts (01-75), washers (01-76) and nuts (01-77).
g) Arrange the load cell cables in the rigid conduit, thread the cables
through the flexible conduit (22-15) and into the QLC. Refer to
connection diagram Figure 6-6 and connect the wires to the
terminals. Ensure that the cables are not strained and tighten the
cable glands.

6.29.2 Transformer – Installation

a) If necessary, remove the screws and nuts attaching the transformer


66 746-270-2
11/05
case lid.
b) Position the transformer in the transformer bracket (05-1) and
attach four pillars (05-2) to the bracket with screws (05-3) and
washers (05-4).
c) Connect the input and output cables to the transformer terminals.
Fit the lid on the transformer case and secure transformer and lid
with four screws (05-8) and washers (05-4).
d) If removed, arrange the mains input cable assembly in the rigid
conduit and refit the conduit cover. Install the strain relief bush on
the bracket at the head of the bed and secure the cable with the
cable clamp.

6.29.3 Radius Arms – Installation

a) Refer to the drawing to ensure correct orientation. Install the radius


arm on the longer pivot on the base frame then slide it to one side
to engage on the shorter pivot.

+ Installation of plastic clips will be easier if they are first warmed by


immersion in hot water for a few minutes.

b) Install a new radius arm clip (01-23) with a firm straight push then
rotate the clip using the special tool (Tool item 4) until the internal
spigot locates in the hole in the pivot.
c) Install the stabilisers (01-3) and attach with pivot bushes (01-62),
bolts (01-64), washers (01-60), and Benton nuts (01-61). Fit plastic
covers (01-63).
d) Attach the height actuator ram to the radius arm fabrication with
clevis pin (01-13), washer (01-9) and ‘E’ ring (01-10).
e) Install the moulded base cover (01-6), together with the flexible
conduit (01-5).
f) Attach the leg fabrications (01-4) with pivot bushes (01-62), bolts
(01-64), washers (01-60), and Benton nuts (01-61). Fit plastic
covers (01-63).

6.29.4 Deck Assembly – Installation

a) Use the hoist to position the deck assembly and attach it to the leg
fabrication (01-4) with four pivot bushes (23-32), washers (23-17)
and pivot bolts (23-35).

67 746-270-2
11/05
b) Attach the flexible conduit (01-5) to the deck frame with a cable
tie.
c) Use the cords in the flexible conduit to pull in the UPS outlet cable,
earth bonding strap, deck actuator and ‘T’ piece cables and QLC
extension cable.
d) Refer to the wiring diagram Figure 6-6 and reconnect the QLC
extension cable (04-1). Reconnect the earth bonding strap (04-13)
to the deck earthing point. Reconnect the UPS output cable to the
pump outlet socket. Reconnect the deck actuator, ‘T’ piece and
ACP cables to the control box (22-16).
e) Lay the cables into the rigid conduit on the base and replace the
covers. Secure cables to the deck assembly as appropriate.
f) Replace the UPS bracket lid (06-2) (13-2) and attach with four pan
head screws (06-9) (13-9) and two self tapping screws (06-10) (13-
10).
g) Install the base cover as described in 6.4.

6.29.5 Thigh section – Installation

a) Use the hoist to position the thigh section and attach it to the deck
frame with pivot bolts (02-35), nylon washers (02-81) and Benton
nuts (02-36). Install four nut covers (02-37).
b) Use the hoist to lower or raise the section and align the hole in the
ram with the hole in the thigh section fabrication. Attach the ram
with ‘E’ ring (02-18), washers (02-17) and clevis pin (02-16).

6.29.6 Calf Section – Installation

a) Use the hoist to position the calf section and attach it to the thigh
section with pivot bolts (02-35), nylon washers (02-81) and Benton
nuts (02-36). Install four nut covers (02-37).

68 746-270-2
11/05
(02-39)
(02-65)

(02-64)
(02-51)
(02-61)
(02-63) (02-64) (02-65)

(02-62)

Figure 6-12 Assembly of gas spring to calf section

b) If removed, install the gas spring mounting (02-63) with two


screws (02-65) and headed bushes (02-64). Install the spacer (02-
61) and rubber washer (02-62) on the gas spring (02-51) and screw
the stem into the counter-bored side of the gas spring mounting
(02-63).

(02-46)
(02-1)
(02-49)

(02-48)

(02-47)

(02-49) (02-50)
(02-51)

(02-47) (02-48)

Figure 6-13 Assembly of gas spring to deck frame

c) Attach the gas spring to the deck frame with clevis pin (02-46),
plain bushes (02-47), headed bushes (02-48), washers (02-49) and
‘E’ ring (02-50) – refer to Figure 6-11.
d) If removed, assemble the release moulding (02-66) and release stop
(02-68) to the calf section with bolt (02-67), washers (02-70) and
self locking nut (02-69) as shown in the figure. Make sure that the
release fabrication (02-52) rests on the stop (02-68) as shown in
Figure 6-12.

69 746-270-2
11/05
(02-39)

(02-52)

(02-66) (02-68)

(02-67) (02-69)

(02-70) (02-70) (02-70)

Figure 6-14 Assembly of release moulding

e) If removed, assemble the calf deck sheet (06-4) onto the extension
bow (02-40), install the extension bow on the calf section
fabrication and secure with two spring pins (02-41).

6.29.7 Backrest – Installation

a) Use the hoist to position the backrest and slide the guide bars (02-
14) through the deck fabrication (02-1) and the backrest fabrication
(02-9). Drive in two spring pins (02-13) to retain the guide bars.
b) Attach the stabiliser fabrication (02-10) with two shoulder screws
(02-12). Reconnect the earth bonding strap (02-71) with self
tapping screw (02-72) and shakeproof washer (02-74).
c) Adjust the height of the backrest to align the hole in the ram with
the hole in the backrest fabrication. Attach the ram with ‘E’ ring
(02-18), washers (02-17) and clevis pin (02-16).

Some beds may be fitted with an additional damper. In this case the

+ backrest actuator ram and the damper are both secured by clevis
pin (02-82), washers (02-78) and ‘E’ ring (02-79), reference
drawing 804.76.

d) Attach the CPR backrest release cable and adjust as described in


6.25.

6.29.8 Safety Sides - Installation (split safety sides)

The following instructions refer specifically to Head End Safety


+ Side L/H, sheet 13. The same procedures apply for all other safety
sides.

70 746-270-2
11/05
a) Use the hoist to support the safety side assembly and attach it to the
backrest fabrication with four button head screws (13-16) and
washers (02-20). Adjust the lateral position of the assembly in the
mounting bracket to gain access to all four fixing holes.
b) Attach the CPR backrest release cable to the mounting bracket (13-
19) with a cable tie.

6.29.9 Safety sides - Installation (sides with embedded control handset)

a) Install the safety side as described in 6.29.8.


b) Attach the cable guide (26-3)(27-3) to the backrest with two self
tapping screws (26-2)(27-2). Connect the handset cable assembly
(26-1)(27-1) to the socket on the deck.

6.29.10 Safety sides - Installation (3/4 length safety sides)

+ The following instructions refer specifically to L/H Safety Side


Assembly sheet 11. The same procedures apply for all ¾ length
safety sides.

a) If removed, install the mounting fabrications (11-17) and (11-18)


and attach with four bolts (11-23) and Nyloc nuts (11-24).
b) Use the hoist to support the safety side in position and attach with
pivot caps (11-16) and countersunk socket head screws (11-15).
Apply Loctite 270 to the screws on assembly.

6.29.11 Foot End Assembly – Installation

a) Align the foot end assembly with the guides in the deck frame,
noting orientation of the bushes (03-15). Lift the latch handle (03-
11) and slide the foot end assembly on to the bed.
b) Install self-locking nut (03-18) on the latch fabrication (03-19). Fit
two pan head screws (03-10) to retain the extension bushes (03-
15).
c) Install the ACP tray (03-32) (together with the ACP and UIM) and
attach with four shoulder bolts (03-6) and nuts (03-7). Replace the
attendant handset (03-26) in the holder.

71 746-270-2
11/05
6.31 Restoration of Paintwork

Damage to the paint finish on the frame or other components can be


repaired by brush application of cellulose-based paint obtainable from
specialist paint suppliers or local distributors.

To ensure colour match, contact Huntleigh Nesbit Evans Ltd quoting the
product code and serial number, for confirmation of finish colour.

WARNING
Always use painting materials as directed by the manufacturer
and follow the safety instructions.

6.30.1 Procedure

a) Clean and degrease the damaged area by wiping with a lint-free


cloth moistened with a suitable volatile solvent e.g. methylated
spirits.
b) Make sure there are no loose chips of paint and use fine silicon
carbide abrasive paper to abrade the area where paint is to be
applied and blend to a smooth surface. Clean and degrease the area
again.
c) Use a soft bristle brush and apply the primer paint to the damaged
area. Allow to dry and apply the colour finish paint. Use both in
accordance with the manufacturer’s instructions.

72 746-270-2
11/05
7 CALIBRATION

7.1 Calibration of Height/Tilt Actuators

Minor accumulated errors in service may result in lost position (non-


parallelism) of the height actuators, as a result of which the bed may not
return to a true horizontal position. This may be automatically corrected
by driving the bed to its lowest height. If this is not effective, the
condition can be rectified by the re-calibration procedure described
below.

7.1.1 Preliminary

Ensure that the bed is connected to the mains power supply. Set the rotary
‘lock-out’ switches on the attendant control panel to enable all the bed
functions. Raise the leg section to about 5°.

7.1.2 Procedure

a) On any of the patient control handsets, press any three of the top
four buttons simultaneously and keep them pressed.
b) A continuous beep sound should be heard from the control box. If
not, repeat step a).
c) After approximately 5 seconds, the continuous sound should
change to an intermittent beep. (This indicates that the control box
is in calibration mode.) Release the push buttons.
d) Press the Height Down button for a few seconds. (This will drive
one of the base actuators inwards.)
e) Press the Height Up button for a few seconds. (This will drive the
other base actuator inwards.)
f) Repeat steps d) and e) until the deck is level at bottom height.
g) At this stage the system will recognise that calibration is complete.
The intermittent beep will stop and the system will revert back to
normal operation.

If at any time during calibration there is a period of more than 5


seconds before a button is operated, the system will revert to
+ normal operation and calibration data will not be stored.
Calibration data will only be stored after completion of all steps a)
to f) listed above i.e., when both base actuators are fully retracted.

73 746-270-2
11/05
7.2 Weighing System Calibration - Specific Gravitational
Requirements.

7.2.1 Equipment Required

Calibrated weight, 250kg.

7.2.2 Weighing System Calibration Procedure

a) Site the bed on a flat, level floor. Raise the bed to maximum height
and level the platform (zero tilt).
b) Fully disengage the transport jackscrews. Remove the lid from the
UPS and control box assembly to gain access to the Quad Load
Converter (QLC).
c) Disconnect the bed from the mains power supply and remove the
cover from the (QLC).
d) Connect the bed to the mains power supply and immediately, while
the system is going through its self test sequence, press the ‘setup’
switch inside the QLC. The upper display on the UIM will now
read ConFIg.
e) The following parameter headings form a circular Menu, which can
be scrolled up or down using the Change Enter Button (3) and the
Reset Button (4). (See Figure 7-1)

Figure 7-1 Calibration parameter menu

74 746-270-2
11/05
f) Assign values to the following parameters as shown:

SCALE 250000
FrAC 1
dEL 0.0001
d 100 or 200 #
Lb on or Lb oFF lb off
CoLd 120
grA 9.8130*

# Refer to approval label on UIM for scale division setting.


* This value will vary depending on the destination for the bed.

This is achieved as follows:

g) Scroll through the Menu as described in (e) to show the required


parameter on the Upper Display (5). Press the Weight Enter Button
(2). The parameter value or status will now be shown on the
display.
h) If the value shown for SCALE, FrAC or dEL is incorrect, press the
Weight Enter Button (2) again and the display will show zero. Key
in the required value on the Keypad (1). When the required value is
shown, press the Weight Enter Button (2) again. The value has now
been accepted and the display will show the parameter heading
again.
i) If the value shown for d or CoLd is incorrect, use the Change Enter
Button (3) and the Reset Button (4) to scroll through a series of
preset values. When the required value is shown, press the Weight
Enter Button (2) again. The value has now been accepted and the
display will show the parameter heading again.
j) If the value shown for grA is incorrect, press the Weight Enter
Button (2) and the display will show zero. Key in the required
figure for local gravity (e.g. Wednesbury, England 9.8130) and
press the Weight Enter Button (2) again. The value has now been
accepted and the display will show the parameter headings again.
Note that this figure needs to be set initially for the location at
which the bed is being calibrated. After calibration and testing it
may need to be reset to suit the final destination of the bed (e.g.
Lyon, France 9.8061). This revised figure needs to be input prior to
sealing the QLC and shipping the bed.
k) If the status of Lb on/ Lb oFF is incorrect, use the Weight Enter

75 746-270-2
11/05
Button (2) to toggle between the two settings.
l) Perform steps g) to k) until all six parameters have the correct
values.
m) Scroll through the Menu to select CAL on the Upper Display (5).
Press the Weight Enter Button (2), grA will now be shown on the
display. Press the Weight Enter Button (2) again and the stored
figure for local gravity will be displayed. Note that this local
gravity figure is for the location at which the bed is being
calibrated and may differ from the local gravity figure for the bed’s
final destination. To change the figure, press the Weight Enter
Button (2) again and input the required value. Pressing the Weight
Enter Button once more will store this new figure. LoAd O will
now be shown on the Upper Display (5).
n) Make sure that nothing is on or touching the bed platform, then
press the Weight Enter Button (2). The Upper Display (5) will now
show _ _ _ _ _ _
o) Do not touch or move the bed whilst this is displayed.
p) Wait until a stable reading has been obtained and the system has
determined the calibration data, LoAd I will be shown on the
Upper Display (5).
q) Place the calibrated weight (250kg) on the bed platform above
Load Cell 1 (see Figure 7-2), press the Load Enter Button (2) and
again wait until LoAd 2 is shown on the Upper Display (5).
r) Repeat step p) in sequence for load cells 2, 3 and 4, each time
placing the weight over the appropriate load cell. When this has
been completed, CAL will again be shown on the Upper Display (5).
Remove the weight(s) from the bed.
s) Scroll through the Menu to select ConFI9 on the Upper Display (5).
Press the Weight Enter Button (2) again.
t) Calibration is now complete but the weighing scale must be set up
as follows prior to use.

Figure 7-2 Load cell positions

76 746-270-2
11/05
7.2.3 Weighing Scale Set-up Procedure

a) Turn the mains power supply off then on again. When CoLd
disappears from the upper display, use the Keypad (1) to enter
3.141593 (π). If done successfully, the Upper Display (5) will show
End. Setup mode is implemented as a circular menu as follows:

Figure 7-3 Weighing scale setup menu

b) Assign value or status to the parameters as follows:


FIL C 3
Fon / FoFF Fon
LC1 - LC4 Not used

c) When the above parameters have been entered, scroll through the
Menu to select End on the Upper Display (5), then press the Weight
Enter Button (2).
d) The UIM will now perform its power-up routine, after which
normal mode is resumed. Both displays should indicate CoLd.
If not, then zero both displays as described in Section 1.
e) Disconnect the bed from the mains power supply. Replace the
cover on the QLC and the UPS and control box assembly. Do not
fit the QLC sealing label yet.

At this point the bed must be tested to EN45501 requirements by an


approved body (e.g. Local Trading Standards). After approval the local
gravity figure may need to be altered to suit the bed’s final destination
(see paragraph 7.3.2.j). The QLC can then be sealed for shipment.

77 746-270-2
11/05
8 TECHNICAL DATA

Huntleigh Healthcare Contoura 980B acute care bed with weighing


system.

Dimensions
Overall Length (normal) 226cm
(extended) 243cm
Overall width (max) 101cm
Overall height (max) 126cm
Height of mattress deck
(with 12.5cm castors) 48 to 82cm
50 to 84cm
(with 15.0cm castors)
In-bed length (normal) 204cm
(extended) 222cm
Tilt 13° foot up to 16°
foot down
Deck angles (max.)

Mattress 202 x 88 x 12.5cm


Castors (3 brake + 1 12.5cm or 15.0cm
brake/steer) diameter.

Product weight (approx.) 207kg


with Breeze mattress 221kg
Safe Working Load (patient plus accessories) 250kg
with Breeze mattress 140kg

Electrical Data
Power in 4A max at 230V a.c., 50/60Hz.
Duty rating 10% (2 min/20 min)
Electrical Safety Standards Complies with BS EN 60601-1
Electric Shock Protection Class I Type B

EMC Complies with BS EN 60601-1-2

78 746-270-2
11/05
Equipotential Terminal Complies with IEC 601-1

Liquid ingress protection IP54


Batteries 2 x 12V 1.2Ah series connected.
Sealed, rechargeable lead/acid gel
types.
UPS Batteries 2 x 12V 5Ah series connected.
Sealed, rechargeable lead/acid gel
types.
Weighing System
Minimum verification level (scale divisions) 100g or 200g #
Maximum capacity 250kg
Minimum capacity 2kg or 4kg #
Operating Temperature Range +10 to +40°C

# Refer to the UIM Approval Label for appropriate setting

For details of the Breeze low air loss mattress system, refer to the
product user manual (publication no. 624901).

79 746-270-2
11/05
9 TRANSPORTATION & STORAGE

Caution
The transport locking screws must always be firmly locked when
the bed is moved or in storage.

Handle with care. Do not drop. Avoid shock or violent impact.


The equipment should be stored in a clean, dry and well-ventilated area.

The following limits apply during transport and for a storage period of up
to 15 weeks:
Ambient temperature -25°C t o +70°C
Relative humidity 10% to 75%
Air pressure 50kPa to 106kPa

The following limits apply to operating conditions or longer periods of


storage:
Ambient temperature +10°C to +40°C
Relative humidity 30% to 75%
Air pressure 70kPa to 106kPa

80 746-270-2
11/05
APPENDIX A

ASSEMBLY SPARES MANUAL

A-1 General

The Assembly Spares Manual is for the identification and requisition of


replaceable parts. It contains assembly drawings of the equipment and
sub-assemblies. Each drawing sheet is accompanied by its related parts
list.
The Assembly / Spares Manual Modification Record (ASMMR) is the
configuration control document for the equipment.
The ASMMR identifies the sheet number and issue number of all
drawings relevant to a particular equipment. A separate ASMMR is
included for each model number covered by this manual.

A-2 How to Use the Assembly Spares Manual

Refer to the relevant ASMMR for the equipment model number.


By reference to the equipment serial number and the ASMMR, identify
the applicable drawing. Find the part on the drawing and note the item
number, which is given in a balloon. Refer to the related parts list for
information regarding the part.
Sometimes a number of identical parts may appear in different locations
but only one will be identified with a balloon. The number in the “Qty”
column of the parts list indicates the total number of times that the part
appears in the drawing.
When a part has optional variations, the last column of the parts list refers
to a separate drawing sheet defining the choices.

81 746-270-2
11/05
Form 130/2

Service Manual Amendments Sheet

Document Ref. 746-270

Product – Contoura® 980B Weighing Beds

Build
Mod Note Date Serial No. Sheet No. Drawing No. Issue
Level
52 4539 01.06.06 828448 22 804.153 5
53 4567 01.08.06 839164 30 804.61 12
54 4602 05.10.06 853832 03 804.46 4
55 4626 31.10.06 896053 01 804.44 10
11B 796.891 3
56 4726 28.04.07 893582
12B 796.892 3

Form 130-2 Service Manual Amendments Sheet_C980B.doc 23/04/2007


APPENDIX B

DISPOSAL OF GAS SPRINGS


The gas spring contains air and oil at high pressure and must be vented in
accordance with the following instructions before being discarded.
Under no circumstances should any attempt be made to open the device.

WARNING
Danger of explosion. Do not heat or incinerate.
High pressure gas. The sudden release of gas at high pressure
could cause serious injury or death. Wear suitable protective
clothing, eye protection and/or a face shield.

1. Operate the valve at the end of the piston rod and allow the piston rod
to extend fully.
2. Clamp the gas spring in a vice and drill a hole approximately 3mm
diameter at position 1 as shown in Figure B-1. Screen off the drilling
point because metal chips and oil may be ejected due to the high
internal pressure. Then drill a second hole at position 2 as shown. The
holes should be drilled to a depth of approximately 10mm.

Figure B-1 Disposal of gas springs

3. Operate the valve at the end of the piston rod and push in the piston
rod.
4. Drain oil from the gas spring by pumping the piston rod in and out
several times.
5. Dispose of the gas spring and oil through raw materials trade or
special refuse points. Do not dispose of gas springs or oil in household
refuse.
If correct disposal in accordance with these instructions is not possible,
the unit should be returned to the supplier.

82 746-270-2
11/05

Вам также может понравиться