ULTRAVIM
Application
Manual
Ver 5.5 1
ULTRAVIM APPLICATION MANUAL
1. Introduction
Describes the functionality and specifications of the UltraVim system.
2. Methodology
Explains the measurement definitions, methodology, and algorithms.
3. System Organization
Describes the operating system, the directories, and files.
4. UltraVim Operation
Gives an overview of the UltraVim software and the screens used.
5. Part Files
Shows how to create and maintain UltraVim part files.
7. Calibration
Describes how to verify the calibration of the system.
8. Managing Files
Shows how to copy and delete files using File Manager.
9. UltraMark Module
Describes the operation of the UltraMark module for mark inspection.
A. Adding a Printer
Explains how to connect a printer and install the drivers.
C. Custom Reports
Shows how to create a custom Lot Summary report.
D. Host Communications
Describes the host protocol required to communicate with the UltraVim.
Ver 5.5 2
ULTRAVIM APPLICATION MANUAL
1. Introduction
The UltraVim and UltraVim Plus inspection modules perform true three dimensional
inspections on many semiconductor devices such as Gull-Wing Parts, J-leaded parts, and
Ball Grid Arrays (BGA). They utilize a user-friendly interface and true 32-bit code for
fast and accurate results. A summary of the features is shown below.
Power Supply
Motherboard, Processor, and Memory
Floppy Disk and Hard Disk
Camera Power Supply
Digital IO Board
Framegrabber Board
Mouse and Keyboard
Ver 5.5 3
ULTRAVIM APPLICATION MANUAL
2. Methodology
Scanner Technologies has developed the UltraVim inspection system to provide the
fastest and most accurate three-dimensional inspections available in the industry. By
using an NIST traceable calibration standard, special optics, and a high-resolution
camera, the UltraVim system provides a solution that can meet your needs today, and in
the future.
System Calibration
Find Pixel Coordinates
Convert Pixel Coordinates into World Coordinates
Convert World Coordinates into Part Coordinates
2.1 Calibration
The secret to the accuracy of the UltraVim starts with the NIST traceable calibration
reticle. This standard is measured to an accuracy greater than 10 micro-inches. This
reticle utilizes 1,600 calibration references to define the World Coordinate Reference. By
using robust statistical techniques, the UltraVim is able to create and store this reference
with sub-micron accuracy.
The UltraVim has no moving parts. Once it has been calibrated, it stays calibrated. You
may want to calibrate the system once a month, or once a year to satisfy the verification
requirements or your company's internal process.
Ver 5.5 4
ULTRAVIM APPLICATION MANUAL
Each inspection reticle has its own calibration file. When you install a new inspection
reticle on the UltraVim system, make sure you load the correct file. This file contains
precise information about the geometry of the reticle.
There is no file associated with the calibration reticle. They are interchangeable. These
reticles are manufactured to specific measurements with a tolerance of less than 0.25
microns. Before the calibration reticle is certified for use with the system, it is tested at
the Scanner Technologies factory. If it does not meet the required tolerance, it is
destroyed.
Ver 5.5 5
ULTRAVIM APPLICATION MANUAL
The example below shows a sample image. This image will be used to demonstrate the
concepts used in this chapter. Note that this simple illustration does not include side
images, or ray tracing.
Assume that the software searched the image for the leads, and found the first lead at
386.2, 349.7 pixels. This is done for all the leads in the image, but for our example we
will just look at the first lead.
Pixel Measurements
X = 386.2 pixels
Y = 349.7 pixels
Ver 5.5 6
ULTRAVIM APPLICATION MANUAL
To show an example of how this is done we can use the formulas below. For this
example, assume the pixel size is 1.8 mils and the center pixel is 512. Note that this is a
very simplified version of what actually happens in the software.
X World = (X Found Pixel – X Center Pixel) * X Pixel Size
Y World = (Y Center Pixel – Y Found Pixel) * Y Pixel Size
X World = (386.2 – 512.0) * 1.8 = -226.4 mils
Y World = (512.0 – 349.7) * 1.8 = 292.1 mils
World Measurements
X = -226.4 mils
Y = 292.1 mils
Ver 5.5 7
ULTRAVIM APPLICATION MANUAL
After the Part Coordinate system has been calculated, the Part Measurements can be
determined. This is done by calculating the expected ideal position of each lead, based on
the Part Coordinate system. In the diagram below, you can see that the lead has a small
amount of Bent Lead and Tip Offset because the actual center of the lead is not the same
as the ideal center based on the Part Coordinate system.
Part Measurements
Ver 5.5 8
ULTRAVIM APPLICATION MANUAL
For absolute measurements, the tolerances will be positive numbers. For example, if the
Width should be 10.0, you may want to set a tolerance from 8.0 to 12.0 in the part file.
For relative measurements you will use both positive and negative numbers. For
example, for Bent Lead you may want to set a tolerance of -4.0 to +4.0 because a value of
0.0 shows that the lead is not bent at all.
Ver 5.5 9
ULTRAVIM APPLICATION MANUAL
3. System Organization
This chapter describes the operating system, the directories on the system, and the
important files required for operation.
The UltraVim system comes with the Microsoft Windows NT operating system. All of
the programs and drivers utilize true 32-bit code for maximum efficiency. When the
system starts, you will not find the standard Microsoft “Explorer” interface. Instead, a
simplified interface will be shown that has command buttons for the common programs
that you may need to run.
Computer Shows a dialog box that displays the contents of the computer as icons.
These icons include disk drives, the control panel, and printers. This icon
requires the user to enter the password for level 3 to run.
A:Setup Shows a dialog box that lists the executable files on the A: drive.
Use this button to run setup programs from the A: drive. This is
for installing upgrades to the software or new drivers.
UltraVim Starts the UltraVim software. There is no need to click on this button
when you power up the system. The UltraVim software will start
automatically without any user intervention.
Ver 5.5 10
ULTRAVIM APPLICATION MANUAL
Files Runs the File Manager program. This icon requires the user to enter
the password for level 3 to run.
Exit Exits Windows NT. This icon required the exit password before
it closes down Windows. If you need to power down the system,
or restart the system, it is best to do so by using this button.
3.2 Directories
The hard disk on the UltraVim is divided into three drives. The C: drive contains the
Windows NT operating system and drivers, the D: drive holds the UltraVim software, and
the E: drive is used to “Auto-Save” data.
C Drive
The C: drive contains the Windows NT operating system. It is very important that you do
not delete any files on this drive. A password level of 3 is required to delete files, so
make sure that only the system administrator knows this password and it is not given to
anyone else.
: If any files are deleted from the C: drive, the UltraVim system may no
longer function properly!
Ver 5.5 11
ULTRAVIM APPLICATION MANUAL
D Drive
The D: drive contains the UltraVim software. The following directories must be present
for the system to work correctly:
D:\UVIM All the executable files and dll’s are stored here.
D:\UVIM\Files This directory contains all the system files for the UltraVim.
D:\UVIM\Parts The part files are stored in this directory.
D:\UVIM\Data Any data that has been saved manually is stored in this directory.
: If any files are deleted from the D:\UVIM\Files directory, the UltraVim
system may no longer function properly!
; If you have multiple UltraVim systems, you can copy part files from the
Parts directory to other systems. The part files are not system
dependent.
E Drive
The E: drive contains the data from the “Auto-Save” function. It is important to archive
or delete the data on the E:\ when it is no longer needed to keep the drive from filling up.
When you select a part file, its Auto-Save directory is created. For example, if the Auto-
Save directory is specified as “QFP160” in the part file, then the “E:\QFP160.001”
directory is created. If the “E:\QFP160.001” directory already exists, then
“E:\QFP160.002” is created. The directory keeps incrementing up to “.999”.
The following files will be created in the “QFP160.001” directory (if the option to save
the file is checked in the Options dialog box, and Auto-Save is check in the Options
dialog box):
Ver 5.5 12
ULTRAVIM APPLICATION MANUAL
The ##### in the part files represent the number of the part that was measured such as
PRT00001, PRT00002, PRT00003, etc.
Note that if you have the Auto-Delete function turned on, the files on the E:\ drive will be
deleted after a certain number of days. This can be changed in the Options screen.
Ver 5.5 13
ULTRAVIM APPLICATION MANUAL
4. UltraVim Operation
This chapter gives an overview of the UltraVim software. Normally, the host system
controls inspections, but there may be times that you want to use the UltraVim software
for manual operations.
Ver 5.5 14
ULTRAVIM APPLICATION MANUAL
Select Part Use this command to manually select a part, or select the “Pick and Place
Setup” mode. Normally, the host system will automatically select a part.
Inspect Part Use this command to manually inspect a part. A part must be over the
inspection reticle first. Normally the host will automatically send the
inspect command.
Data Use this command to view and save data, or clear the data. If you do not
have the “Auto-Save” command on, all data is lost when the system is
shut down, the data is cleared, or a new part is selected.
Options This command lets you change options such as which language is shown,
or what measurement the SPC char displays. You need to enter password
level 2 to enable this command.
Maintenance This command lets you edit and create part files. If you are in password
level 1, you can only edit files. If you are in password level 2, you can
create and save files. If your password is level 3, you can delete files.
Passwords Use this command to change password level, change passwords using
the administration screen, or exit the program using the exit password.
Ver 5.5 15
ULTRAVIM APPLICATION MANUAL
The icons in the lower-left corner of the screen let the user select the part type. When the
part type is changed, the files listed will also change.
If the user has password level 3, the reticle can be changed. The reticle shown must
always match the reticle that is actually on the system when you are inspecting Gull Wing
and J-Leaded parts. For BGA parts, the user can ignore the reticle selection.
Ver 5.5 16
ULTRAVIM APPLICATION MANUAL
If you have a new reticle, insert the reticle setup disk in the A: drive and click on the
“Load” button. If you retire a reticle, or send it to the manufacturer for repair, you should
select that reticle in the list, then click on the “Delete” button to remove the file from the
system.
: If the wrong reticle file is selected, the inspection results will not be
valid for Gull Wing and J-Leaded parts. The reticle files have the
extension “RET” and are stored in the D:\UVIM\Files directory.
If you check the “Pick & Place Setup” checkbox, the system will perform only a two-
dimensional XY inspection and report the X, Y information in output channels C0-C7.
This should be used with the setup reticle to determine the proper placement of the
hander system over the center of the reticle.
If you want to test the system by measuring parts with the setup reticle, choose the reticle
called "SETUP", and the system can measure parts with the setup reticle, but note that the
coplanarity will not be checked because the setup reticle does not have prisms on it.
; Any part file can be loaded manually, but the host can only load part
files that contain 1 or 0 in the file name.
Ver 5.5 17
ULTRAVIM APPLICATION MANUAL
When a part is selected, the "Resume" button may become active. If the user clicks on
the "Resume" button, the lot summary information from the previous lot will be loaded
into the program. This includes the number of pass, rework, and fail parts as well as the
Lot Summary statistics.
To make sure the "Resume" button will become enabled for a part, the Auto-Save feature
must be used to save the Lot Summary report. When the user selects a part, the program
looks in the Auto-Save directory for the LOTSUMM.RPT file. If this file is found, the
"Resume" button will become enabled.
If there are multiple Auto-Save directories, the program will look in the last directory
found. In the example below, the "Resume" button will load the information from the
E:\QFP160.003 directory.
E:\QFP160.001
E:\QFP160.002
E:\QFP160.003
Note that the Lot Statistics information will not be loaded into the program when a lot is
resumed – only the Lot Summary. Also, the current user information that is entered will
be used.
Ver 5.5 18
ULTRAVIM APPLICATION MANUAL
When the user clicks on the “Inspect Part” button, or the host give the inspect command,
the UltraVim performs the following steps:
If a lead (or ball) was not found, the result is shown as “NF”. You should be able to look
at the image and see a red marker where the lead was expected. This can occur for the
following reasons:
Ver 5.5 19
ULTRAVIM APPLICATION MANUAL
4.5 Data
To view data, click on the “Data” button. The following dialog will be shown:
Note that you can change the User and Lot information on this screen. This will change
any of the files that you view, but it will not change the part files that have already been
saved with the “Auto-Save” function.
Ver 5.5 20
ULTRAVIM APPLICATION MANUAL
4.6 Options
Inspection Cycles The inspection cycles should be left at “NORMAL”. If you select
“Demo” the system will inspect the same part until you click on the
cancel button. Demo 1.0 inspects once per second, Demo 2.0
inspects once every two seconds, etc.
SPC Subgroup Size The SPC subgroup size shows how many values to average before
saving in the SPCSTAT.CSV file. It is best to set this number to 5
or 10. If you have large lot sizes, the SPC file can get very large.
Especially if this value is set to 1.
Ver 5.5 21
ULTRAVIM APPLICATION MANUAL
Auto-Delete If you turn on the Auto-Delete option, files on the E:\ drive
will automatically be deleted after a certain number of days.
Specify the number of days in the edit box to the right of
the Auto-Delete choice.
Graphics This determines if the graphics are shown. If you have a host
system that is very fast (above 10,000 uph) you should turn the
graphics off to speed up the system. If not, it is best to leave
graphics on.
Image You can use the save and load buttons to save or load TIF images.
If you have a part that looks strange, or think there is a problem
with your system, save the image and send it to customer support.
SPC Chart Select the measurements you want displayed on the SPC chart
on the main screen.
Ver 5.5 22
ULTRAVIM APPLICATION MANUAL
4.7 Maintenance
When you click on maintenance, the dialog below will be shown. From this dialog, you
can click on the icon of the type of part you want to edit or create. If your system only
supports one part type, this dialog box will not be shown.
4.8 Passwords
From the Passwords dialog box you can enter the password to change user level, exit the
program, or change passwords. The default passwords are:
Level 2 2
Level 3 3
Exit EX
Administration ADMIN
Ver 5.5 23
ULTRAVIM APPLICATION MANUAL
If the OK button remains gray, the old password is not correct, or the new password does
not match the “re-entered” new password.
Ver 5.5 24
ULTRAVIM APPLICATION MANUAL
5. Part Files
This chapter describes how to create and edit part files for the UltraVim system. It is very
important to make sure the values in the part files are correct for the system to operate
correctly. It is best to use the CAD drawing of the part to verify the values you enter in
the part file.
5.1 Maintenance
When you click on the “Maintenance” button from the UltraVim software, a dialog will
be shown that lets you select the part type. If your system is set up for only one type of
part, then this dialog will not be shown.
; If you cannot edit the part because all the fields are gray, go to the
Passwords screen and enter the password for level 2 or level 3.
Ver 5.5 25
ULTRAVIM APPLICATION MANUAL
Enter a short description of the part. This is especially important if the File Name
contained only 1’s and 0’s. This helps you identify the part file for manual operations,
and it will also show up on the reports.
The Auto-Save Directory can be up to eight characters long. Make sure that the Auto-
Save Directory for each part is unique. When Auto-Save is selected from the UltraVim
software, the Auto-Save Directory will be created on the E:\ drive with an extension from
001 to 999. For example, the first time you run the QFP160 part, the Auto-Save data will
be saved in the E:\QFP160.001 directory.
Ver 5.5 26
ULTRAVIM APPLICATION MANUAL
Select the Units and Lead Type, then enter the Lead Dimensions and Body Dimensions.
For a J-Leaded part, there will be no Foot Length or Foot Angle. The most common
orientation is Counter Clockwise. If you are creating a J-Leaded part, the Center
checkbox will become active. This lets you define lead 1 at the center of side 1 instead of
the corner.
If these conditions are not met you can create an “Irregular” part. To do this, first create a
normal part file that is as close to the part as you can make it. Next, select “Irregular” in
the Orientation section of the screen, then click on “Position”. The following screen will
be shown:
Ver 5.5 27
ULTRAVIM APPLICATION MANUAL
From this screen, you can enter the Side and X,Y position of each lead. Make sure that
you start with Side 1 and stop at Side 3 for a two-sided part or Side 4 for a four-sided
part.
If a part has a missing lead, you can delete the X and Y positions for that lead. This will
cause the lead to keep its lead number, but it will not be measured. Note that it will show
up as 0.0 on some of the graphs because a blank line is seen as a 0.0 value by the
graphing software.
5.3 Tolerances
The Tolerance screen defines how the values are calculated, and what values will cause
the part to be rejected or reworked. Some host systems may only have a reject output, but
you may still want to enter rework values to keep track of how many parts are “close” to
being rejected.
Ver 5.5 28
ULTRAVIM APPLICATION MANUAL
At the bottom of the tolerance screen there are several selections to change the
calculations used by the UltraVim software. A summary of these calculations is shown
below.
Ver 5.5 29
ULTRAVIM APPLICATION MANUAL
Ver 5.5 30
ULTRAVIM APPLICATION MANUAL
5.4 Measurements
Coplanarity
Coplanarity is the distance from the lead to a flat surface when
the part is placed on that surface.
Standoff
Standoff is the distance from the body to a flat surface when the
part is placed on that flat surface.
Tip Offset
Tip Offset is the distance from the tip of the lead to the ideal
position of the lead.
Tip Error
Tip Error uses a best fit line to calculate the tip offset relative to
the other leads on a side. The formula used is: Tip Error = Tip
Offset – (A*N + B) where A is the slope parameter in the
regression, N is the lead number, and B is the intercept.
Skew
Skew is the distance from the edge of the lead to the ideal center of
the lead. Skew = abs(Bent Lead) + Width/2
Ver 5.5 31
ULTRAVIM APPLICATION MANUAL
Bent Lead
Bent Lead is the distance from the measured center of the lead to the
ideal center of the lead.
Width
Width is the distance from one edge of the lead to the other edge.
This measurement is calculated 1/4 width from the lead tip.
Terminal Dimension
The terminal dimension is the tip to tip distance between two
opposite leads.
Pitch
The lead pitch is the distance between the measured centers of two
adjacent leads.
Burr
The burr measures any width of the lead that is larger than the
average width of all the leads. The foot length controls how far
towards the shoulder the software will search for a burr.
Ver 5.5 32
ULTRAVIM APPLICATION MANUAL
Enter a short description of the part. This is especially important if the File Name
contained only 1’s and 0’s. This helps you identify the part file for manual operations,
and it will also show up on the reports. The Auto-Save Directory can be up to eight
characters long. Make sure that the Auto-Save Directory for each part is unique. When
Auto-Save is selected from the UltraVim software, the Auto-Save Directory will be
created on the E:\ drive with an extension from 001 to 999. For example, the first time
you run the BGA208 part, the Auto-Save data will be saved in the E:\BGA208.001
directory.
Ver 5.5 33
ULTRAVIM APPLICATION MANUAL
If you click on the “Setup” button, the screen below will be shown. From this screen you
can remove missing balls from your part file. To add or remove a ball, click on that ball
with the mouse. As you move the mouse over a ball, the ball label will be shown on the
left. The Part Reference should be set to Upper Right. If you click on Full the software
will automatically set all the balls to become active. If you click on staggered, the
software will set every other ball to become active.
Ver 5.5 34
ULTRAVIM APPLICATION MANUAL
In Figure 6A below, you can see that the balls are rotated on the board, and the X and Y
datum are also rotated to match the rotation of the balls. In this example, the X Error and
Y Error of the balls would be small because the positions of the balls are close to the ideal
positions define by the X and Y datum.
Figure 6B shows the same part with the datum define with respect to the board. The
board is not rotated and the X and Y datum are not rotated. This will create X and Y
Errors that are larger than in Figure 6A because the ideal positions of the balls defined by
the datum will not be close to the actual position of the balls.
Ver 5.5 35
ULTRAVIM APPLICATION MANUAL
Figure 6A Figure 6B
In Figure 6C below, you can see that the balls are offset on the board, and the X and Y
datum are also offset to match the balls. In this example, the X Error and Y Error of the
balls would be small because the positions of the balls are close to the ideal positions
define by the X and Y datum.
Figure 6D shows the same part with the datum define with respect to the board. This will
create X and Y Errors that are larger than in Figure 6C because the ideal positions of the
balls defined by the datum will not be close to the actual position of the balls.
Figure 6C Figure 6D
Ver 5.5 36
ULTRAVIM APPLICATION MANUAL
3.0
2.0
1.0
0.0
-1.0 Hysteresis
Range
-2.0
-3.0
3.0
2.0
1.0
0.0
-1.0
-2.0
-3.0
Ver 5.5 37
ULTRAVIM APPLICATION MANUAL
5.8 Measurements
X, Y Diameter
This is the distance from the left edge of the ball to the right edge
or the distance from the top edge of the ball to the bottom edge.
X Error
This is the distance from the X position of the ball to the ideal X
position defined by the X datum.
Y Error
This is the distance from the Y position of the ball to the ideal Y
position defined by the Y datum.
R Error
This is the distance from the X,Y center of the ball to the ideal X,Y
center defined by the X and Y datum.
Coplanarity
This is the Z distance of the ball to the average Z measurement as
defined by the LMS regression plane or Hysteresis plane.
Ver 5.5 38
ULTRAVIM APPLICATION MANUAL
Standoff
This is the Z height of the ball relative to the board. This
measurement is calculated by added the nominal standoff with
the relative ball height.
Warpage
This is the average Z distance from the center of the package to the
corners of the package.
Ball Quality
This is a measurement of the roundness of the ball. It is calculated
by finding the minimum diameter and maximum diameter of the
ball and using the following formula:
Ball Quality = min / max * 100%
X, Y Grid Offset
This is the X and Y offset of the center of the ball array grid to the
center of the package.
Ver 5.5 39
ULTRAVIM APPLICATION MANUAL
Ver 5.5 40
ULTRAVIM APPLICATION MANUAL
; The following types of reports use a file extension of RPT. You can load
any of these types of reports from this screen.
BGA Part Summary
BGA Lot Summary
Gull Wing & J-Lead Part Summary
Gull Wing & J-Lead Lot Summary
Ver 5.5 41
ULTRAVIM APPLICATION MANUAL
The second section of the report shows a summary of the minimum values for each part.
The Min and Max show the “Minimum of the Minimums” and the “Maximum of the
Minimums” for the lot. For example, the minimum Bent Lead will always be a negative
number so the Min and Max will both be negative numbers as well. The average and
standard deviation are calculated for the entire lot as well.
The next section of the report shows a summary of the maximum values for each part.
The Min and Max show the “Minimum of the Maximums” and the “Maximum of the
Maximums” for the lot. For example, the maximum Bent Lead will always be a positive
number so the Min and Max will both be positive numbers as well.
Ver 5.5 42
ULTRAVIM APPLICATION MANUAL
The second section of the report shows a summary of the minimum values for each part.
The Min and Max show the “Minimum of the Minimums” and the “Maximum of the
Minimums” for the lot. For example, the minimum X Error will always be a negative
number so the Min and Max will both be negative numbers as well. The average and
standard deviation are calculated for the entire lot as well.
The next section of the report shows a summary of the maximum values for each part.
The Min and Max show the “Minimum of the Maximums” and the “Maximum of the
Maximums” for the lot. For example, the maximum X Error will always be a positive
number so the Min and Max will both be positive numbers as well.
Ver 5.5 43
ULTRAVIM APPLICATION MANUAL
By using the command buttons at the top of the screen, you can load a file, save a file,
delete a file, or print the charts or data. To delete a file you must have password level 3
before you enter this screen.
;Note: You can load any UltraVim report with a CSV extention from
this screen such as part measurements or lot statistics.
Ver 5.5 44
ULTRAVIM APPLICATION MANUAL
If you click on the "Stdev" selection on the left of the screen, you will see an expanded
view of just the standard deviation. If you click on the "3D" button, you will see a 3D
chart of the coplanarity.
Ver 5.5 45
ULTRAVIM APPLICATION MANUAL
6.7 SPC
The SPC data shows the minimum, maximum, average, standard deviation, and CPK for
each measurement. The number of data points in the SPC chart depends on the SPC
Subgroup Size that is set in the Options dialog. For example, if you had 1,000 parts in
your lot and the Subgroup Size is set to 10, you will have 100 data points in the SPC
chart. Each data point will be the average of the previous 10 parts measured.
The CPK number is equal to the difference between the maximum and minimum
tolerance divided by six standard deviations. Generally, a CPK number greater than one
shows that the process is stable.
If the SPC file is very large, a dialog will be shown that shows the percent of the file that
is loading. If you click on the cancel button, the file will stop loading and you will be
able to see the data that was loaded. For example, if you clicked cancel on the dialog
below, you would still be able to see 25% of the SPC data.
Ver 5.5 46
ULTRAVIM APPLICATION MANUAL
6.8 Printing
When you click on the Print button, the screen below will be shown. From this screen
you can select to print the chart, or the data. If you select “Data”, you can enter the pages
you want to print, and select the report you want to print. The “Report” limits the
columns of data that will print out.
When you click on the “Reports” button, the screen below will be shown. From this
screen you can name a report and select the columns that you want to print. To select a
column just click on it in the listbox. To de-select it, click on it again. Make sure you
click on the “Save” button to save your changes. To delete the report, click on the
“Delete” button while the report is active.
Ver 5.5 47
ULTRAVIM APPLICATION MANUAL
6.9 3D Charts
If you click on the “3D” selection, a three dimensional chart of Coplanarity will be
shown. This option is available for all Gull Wing and J-Leaded parts, but it is only
available for BGA parts if the UltraVim has the additional 3D-BGA option. To change
the view of the chart, click on the chart, hold the left mouse button down, then drag the
box to the desired angle you would like to see.
Ver 5.5 48
ULTRAVIM APPLICATION MANUAL
To run the Golden Accuracy report, first place the device on the UltraVim. Next, select
the correct part file and click on the Inspect button several times to inspect the device. A
sample size of 30 inspections is normally used. After you have completed the
inspections, click on the Password button and enter password “G”. The dialog below will
be shown. Select the correct device serial number, then click on the OK button.
Ver 5.5 49
ULTRAVIM APPLICATION MANUAL
If the correct device is not listed in the dropdown selection, copy the correct *.CRB or
*.CRQ file to the D:\UVIM\Library. For example, if you are using device GB16-123, you
will need to copy the GB16123.CRB file. This file contains the values from the
certificate file and is used to compare with the inspection results to generate the accuracy
report.
An example accuracy report is shown below. To calculate the results, the following
formula is used:
Delta[i] = abs(C[i] – M[i])
C is certificate value.
M is measured value.
From the calculated “Delta” values, the maximum delta and the 3-Sigma accuracy values
are calculated and displayed:
3-Sigma = (Average + 3*Stdev)
Accuracy(+/-) = +/- (3-Sigma)*0.5
Ver 5.5 50
ULTRAVIM APPLICATION MANUAL
To run the Gage R&R Report, first click on the Options button and set the SPC Subgroup
Size to 1. Next, select the correct part file and inspect the parts. The most common way
to run the test is to use one tray of parts and cycle the tray 9 times through the system.
The first three runs are called Group A, the next three are called Group B, and the last
three are called Group C.
If you had a tray with 5 parts that was cycled 3 times you would have the following
inspections for Group A:
#1..... Part 1, Run 1 #6..... Part 1, Run 2 #11..... Part 1, Run 3
#2..... Part 2, Run 1 #7..... Part 2, Run 2 #12..... Part 2, Run 3
#3..... Part 3, Run 1 #8..... Part 3, Run 2 #13..... Part 3, Run 3
#4..... Part 4, Run 1 #9..... Part 4, Run 2 #14..... Part 4, Run 3
#5..... Part 5, Run 1 #10..... Part 5, Run 2 #15..... Part 5, Run 3
Another method that is sometimes used is to inspect each part 9 times in a row. This can
be used when it is difficult to cycle the tray of parts through the system. With this
method, if you had a tray with 5 parts and measured each one 3 times, you would have the
following inspections for Group A:
Ver 5.5 51
ULTRAVIM APPLICATION MANUAL
After you have completed the inspections, click on the Passwords button and enter
password “G” to display the Gage R&R Dialog. In the Gage R&R section of the dialog,
the SPC file will already be selected. Note that you can change the file in case you have
saved the data from an old test. Enter the tolerance value for the measurement that you
want a report on. Select the part type that you are testing and the measurement that you
want the report generated for.
Next, you must select the correct test type and the number of runs and parts that were
used for the test. A test type of R1(P1,P2..Pn) R2(P1,P2,Pn) indicates that you cycled the
tray for each run. A test type of P1(R1,R2..Rn) P2(R1,R2,Rn) indicates that you cycled
the tray of parts one time, but measured each part more than once.
Ver 5.5 52
ULTRAVIM APPLICATION MANUAL
An example report is shown below. The top section of the report shows the values that
were used to calculate the Repeatability & Reproducibility.
These are the average ranges for each of the three groups. For example, Range
Group A (RA) is the average range of Run 1, Run 2, and Run 3 for all 10 parts.
Range Group B is the average range of Run 4, Run 5, and Run 6 for all 10 parts.
Range Group C is the average range of Run 7, Run 8, and Run 9 for all 10 parts.
Ver 5.5 53
ULTRAVIM APPLICATION MANUAL
RX = (RA + RB + RC)/3
Average A = 3.60
Average B = 3.69
Average C = 3.65
RY = Max(A,B,C) – MIN(A,B,C)
= 3.69 – 3.60
= 0.09
Repeatability (EV)
Repeatability shows the variation when measuring the same part several times.
It is calculated by multiplying the statistical constant K1 times the average range.
EV = K1*RX
= 3.55*RX
Ver 5.5 54
ULTRAVIM APPLICATION MANUAL
Reproducibility (AV)
Reproducibility shows the variation between groups A, B, and C. Often a
different operator is used for each group to check the reproducibility. Also,
systems are sometimes calibrated between groups to check the reproducibility
of the calibration. The formula for reproducibility is shown below.
AV = sqrt(K2*K2*RY*RY – EV*EV/N)
RR = sqrt(EV*EV + AV*AV)
Ver 5.5 55
ULTRAVIM APPLICATION MANUAL
7. Calibration
This chapter describes how to calibrate the UltraVim system. Because the UltraVim has
no moving parts, you do not have to calibrate often. It is recommended that you calibrate
the first time you setup the machine, any time you move the machine, and one time each
year. You may also want to calibrate the system if you think there are any problems. It is
a good way to verify the basic operations of the system.
When you first start the calibration software, the Calibrate and Options buttons will be
grayed out. If you enter password level 2, the Calibrate button will become active. If you
enter password level 3, the Options button will become active. The spreadsheet on the
left side of the screen shows the calibration log. The result, values, units, and date are
shown. The system must calibrate to a value 0.06 mils or less to pass. Any value greater
than this number will cause a “Fail” result.
Ver 5.5 56
ULTRAVIM APPLICATION MANUAL
If you click on the “?” button, the system will perform a short calibration to calculate the
field of view. If you click on the image, the system will turn on the lights for a short
period of time.
Before you can calibrate the system, you must install the calibration reticle and the
calibration diffuser that came in the UltraVim toolkit. Make sure the reticle is clean and is
seated properly. Assemble the calibration diffuser as shown in the diagram above, and
place it on the calibration reticle. After you have installed the reticle and diffuser, click
on the image to test the installation. You should be able to see the black dots clearly, and
the background should be a smooth gray color.
After you have verified that the image is correct, click on the Calibrate button. The
software will acquire an image, then find all the dots on the screen. The blue bar at the
top of the screen indicates the progress.
As the system calibrates, the X Error and Y Error will become smaller. After the
software has acquired 16 images, the calibration will complete. If both the X Error and Y
Error are less than 0.06 mils, the calibration will pass and the results will be saved.
Ver 5.5 57
ULTRAVIM APPLICATION MANUAL
An example of a map of the non-linear in the system is shown below. All lenses have
these distortions to some degree. The calibration using a NIST traceable calibration
reticle with hundreds of calibration points is critical to the accuracy of the system.
Ver 5.5 58
ULTRAVIM APPLICATION MANUAL
Note that the errors reported in step 4 show the 3 Sigma distances from the measured
centers of the dots to the known NIST centers of the dots. The average 3 Sigma error
value can be up to 0.06 mils. If the value is greater than 0.06 mils, the calibration will not
pass.
7.3 Options
If you click on the Options button, the screen below will be shown. The screen should
only be used the very first time the UltraVim is installed in the host. It shows the
relationship of the part to the inspection reticle.
The default setup is A, and you should try to install the UltraVim so this is the way the
parts are oriented on the system. If you cannot do this, you may changed the orientation
to match the orientation of the actual part as it is inspected on the system.
Ver 5.5 59
ULTRAVIM APPLICATION MANUAL
8. Managing Files
To manage your UltraVim files, exit the software by entering
the exit password. Then click on the Files icon on the main
command bar to start File Manager. You will need to enter
password level 3 to start the program. For on-line help,
press the F1 key.
In File Manager, your files and directories are displayed in a directory window. The
directory window is split: the left half displays the directory tree; the right half displays
the contents of the current directory. When you select a different directory in the directory
tree, the contents of that directory are displayed in the right half of the window.
In the directory window, each filename has an icon next to it, graphically indicating what
kind of file it is.
Directories.
Shared directories. These are directories users can connect to on the network.
Program files, PIFs, and batch files. Their default filename extensions are
.EXE, .COM, .PIF, and .BAT. These files start applications.
Ver 5.5 60
ULTRAVIM APPLICATION MANUAL
Document files. These files are associated with applications. When you choose
a document file, the application starts and opens the file.
File Manager displays a toolbar and drivebar below the menu bar. The toolbar provides
quick access to File Manager commands through buttons. The drivebar displays icons that
represent available drives, including hard disk drives, floppy disk drives, network drives,
and CD-ROM drives. You can change to a different drive by selecting its drive icon.
Ver 5.5 61
ULTRAVIM APPLICATION MANUAL
To change drives, click on a different drive. For example, if you click on the D: drive, the
screen will change to the screen below. At the bottom of the screen you can see that the
D: drive has 653 MB free.
Ver 5.5 62
ULTRAVIM APPLICATION MANUAL
Now click on the “File” menu, then click on “Copy”. Enter A: in the “To” edit box.
Then insert a floppy disk in the A: drive and click OK. The program will copy the
contents of this directory to the A: drive. You may want to back up your D:\UVIM\Files
and D:\UVIM\Parts directories in this way.
To delete a file or directory, simply select it in the directory or file list then press the
delete key. You can also choose delete from the File menu. In the screen below, pressing
the delete key would delete the D:\UVIM\Data directory. Without this directory, the
system will not run correctly. Any time you use File Manager, you must be very careful
not to delete any required files or directories.
Ver 5.5 63
ULTRAVIM APPLICATION MANUAL
9. UltraMark Module
This chapter describes how to install and operate the UltraMark inspection module. This
module requires additional software, framegrabber board, and a CCD camera.
9.1 Hardware
The list below describes all the major hardware components of the UltraMark system and
notes on installation.
Camera Cable The split end of the camera cable should be connected to the
power plug on the computer, and the BNC plug on the
framegrabber board. The single end of the cable should be
connected to the camera.
9.2 Startup
After the UltraMark module has been installed, a command button will appear at the
bottom of the UltraVim main screen. You can click on this button at any time to display
the UltraMark main screen.
To get back to the UltraVim main screen, you can click on the UltraVim command button
at the bottom of the UltraMark screen.
Ver 5.5 64
ULTRAVIM APPLICATION MANUAL
Select Part Shows a list of predefined part files that can be selected.
Learn Mark Finds the characters in the image and saves them in a template.
Inspect Mark Compares the characters in the image to the template and reports
Pass or Fail.
Passwords Lets the user exit the program, or change password level.
Ver 5.5 65
ULTRAVIM APPLICATION MANUAL
Camera Icon This button shows if the camera is in live mode. You
can toggle between modes by clicking on this button.
Lot Summary The Lot Summary window shows the results of the current lot.
Pass The number of parts that have passed.
Fail The number of parts that have failed.
Total The total number of parts inspected.
Part Summary The Part Summary window shows the results of the last part.
Result The result of the last inspection.
Pin1 The result of the last pin1 inspection.
Contrast The average contrast of the mark.
Min Corr The worst correlation value on the mark.
Ave Corr The average correlation value on the mark.
Splatter The number of excess white pixels in the image.
X Pos The X position of the mark in pixels.
Y Pos The Y position of the mark in pixels.
Angle The angle of the mark.
Minimize Icon This button minimizes the screen to just a small image. To enlarge
the image again, double click on it.
Ver 5.5 66
ULTRAVIM APPLICATION MANUAL
After you have selected a part file, a default Region of Interest will be selected. You may
want to change the region of interest to closely outline the characters on the part. To do
this, follow the steps below:
Ver 5.5 67
ULTRAVIM APPLICATION MANUAL
2. Click once in the upper-left corner of the mark and hold the mouse button down.
For the best results, click as close to the characters as you can.
3. Hold the mouse down and move to the lower-right corner of the mark. Let go
of the mouse button when the dotted rectangle is in the correct position.
4. Click on the “Learn Mark” button to find the characters in the region of interest.
In the example below, the software found twelve characters.
Ver 5.5 68
ULTRAVIM APPLICATION MANUAL
In the example below, part of the “S” in the word “SAMPLE” is missing. The software
can only find a match of 27% and the character fails. A typical correlation tolerance used
is 70%.
The Lot Summary shows the results of the current lot. This
includes the total number of parts that have passed and the
number of parts that have failed.
The Part Summary shows the inspection results for the last
part inspected. This summary only shows the
measurements that have been selected in the part file. Note
that if none of the searches or characters are defined as
Pin1, the result for Pin1 will always show Pass.
Ver 5.5 69
ULTRAVIM APPLICATION MANUAL
After the inspection is complete, the user can click on the Data button to view a Mark
Summary report.
This report shows the number of parts that have passed and failed the inspection
tolerances. The report also shows the yeild for each inspection item and the minimum,
average, and maximum values. Results for each individual search area are also shown
Ver 5.5 70
ULTRAVIM APPLICATION MANUAL
Report As The reporting can be set to MARK or PIN1. If this is set to PIN1,
the software will set the B line low if the character fails.
Updatable This lets the software know if a search can be updated after the
initial Learn Mark is pressed by the user.
Search Range This tells the software how far to search for the character.
Search Color This tells the software the color of the mark and background.
Threshold This is the threshold used when the software learns the mark.
Setting this to LOW can help when characters run into each other.
Setting this to HIGH can help find faint characters.
Ver 5.5 71
ULTRAVIM APPLICATION MANUAL
9.8 Passwords
Use the Passwords dialog box to change the user level, change passwords, or exit the
program.
Default Passwords
2 Password level 2 enables the “Options” button.
Users can create files, but not delete them.
To change a password, enter the old password in the first text box. Then enter the new
password in the next two text boxes. When this is done correctly, the “OK” button will
become enabled. For example, to change the supervisor password from “2” to “SUPER”
enter the following:
Ver 5.5 72
ULTRAVIM APPLICATION MANUAL
Old Password: 2
New Password: SUPER
Re-Enter New: SUPER
9.9 Options
Use the Options dialog box to set the default settings for the program, or to save and
image. From this screen you can change the default language to English, Japanese, or
Chinese. If you are in password level 4, you can select the camera and output lines that
are used by the software. The normal output lines used by the mark inspection prgoram
are REQ2. This means that the software will respond to input request line A1, and will
acknowledge the request be setting output line A2. The result lines set will be B2 and
B3. This is shown in the dialog as REQ2 (A1 -–A2B2B3).
Ver 5.5 73
ULTRAVIM APPLICATION MANUAL
9.10 Maintenance
The maintenance screen lets the user edit, create, or delete part files. The part file is
described below.
File
Enter the file name for the part. The part files are saved in the D:\UVIM\Parts directory
and have a *.MRK file extension.
Description
The user can enter additional information in the description field. This text is shown
when the Select Part screen is displayed.
Ver 5.5 74
ULTRAVIM APPLICATION MANUAL
Setup – References
The references can be set up for GLOBAL SEARCH or LOCAL SEARCH. This tells the
software how to use the references. For LOCAL SEARCH the software will search for
each reference as if it were independant from the other references. This could be used if
the software is looking at multiple parts, for example.
Setup – Calculation
This parameter tells the software how to calculate the Average Correlation. If
AVERAGE CORR is selected the software will calculate the average of all items in the
search box. If MIN OF 5 is selected, for example, the software will calculate the average
from the worst five correlations.
Setup – Rotation
With this parameter, you can turn on the test for rotated parts. If TEST180 is selected the
software will test the image rotated at 180 degrees if it fails at 0 degrees. If the image
passes after it has been rotated, the Pin1 inspection will be recorded as a fail.
Ver 5.5 75
ULTRAVIM APPLICATION MANUAL
Tolerances – Contrast
This parameter sets the tolerance for the contrast of the part. If the contrast falls below
this level, the part will fail. The calculation for contrast is the minimum of 100% or 5.0
times the standard deviation of the pixels in the search area.
Tolerances – Correlation
The correlation tolerance is the most important setting. This is the primary method of
passing or failing the mark. If minimum correlation value for any character is below this
setting, the part will fail.
Tolerances – Average
The average correlation tolerance sets the minimum acceptable value for average
correlation. There are several methods used to calculate the average value. The
calculation method is set in the Setup – Calculation section.
Tolerances – Splatter
The Splatter is the number of white ink spots that are found on the part. The software
looks for ink splatter in any area that does not contain a character. The Search – Splatter
parameter must also be turned on.
Tolerances – X Position
This parameter sets the allowable number of pixels that a mark can be from the trained
position. If the found position is greater than this value the part will fail.
Ver 5.5 76
ULTRAVIM APPLICATION MANUAL
Tolerances – Y Position
This parameter sets the allowable number of pixels that a mark can be from the trained
position. If the found position is greater than this value the part will fail.
Tolerances – Rotation
The rotation sets the allowable rotation of the mark in degrees. If the angle of the mark is
greater than this setting, it will fail.
Search – Enabled
For a search to affect the results it must first be set to enabled. The software will skip any
search that is not enabled. Drawing a new search on the image with this mouse will
automatically set that search to be enabled.
Search – X1..Y2
These parameters set the Region of Interest for the search. This region is only used when
the user learns the mark with the Learn button, or updates the mark with the Update
button. The software can find characters that are outside the search if they are in the
allowable range and time allocated.
Search – Part
The part sets what part the search is used for. If the software is set up to test multiple
parts, the user must set this parameter. If the user is only testing one part at a time, this
parameter will be set to PART 1.
Ver 5.5 77
ULTRAVIM APPLICATION MANUAL
Search – Type
This attribute tells the software how to learn the mark. If it is set to CHARACTERS the
software will look for any character that is completely inside the search area, is larger
than the minimum character size allowed, and is smaller than the maximum character
size allowed. If this attribute is set to LOGO/PIN1 the software will use the entire search
area as one character.
Search – Fail
This parameter tells the software what type of failure the search will report. If it is set to
MARK, the A line will be set on the digital IO. If it is set to Pin1, the B line will be set
on the digital IO.
Search – Updateable
This attribute lets the software know if a search can be updated by the user during a lot
inspection. If there are any updateable search areas, the Learn button will display Update
after the user clicks on it the first time. If there are no areas that can be updated, the
Learn button will become disabled.
Search – Reference
The references can speed up a search because the help the software determine the general
position of the mark. The reference can be set to FIRST, LAST, or BOTH. A user can
manually change the references by clicking on a character in the image after the mark has
been learned.
Search – Splatter
The splatter attribute defaults to NO TEST because this inspection can add over 100ms to
the total inspection time. If it is set to AUTOMATIC the software will calculate a
threshold value and search the entire area for ink splatter. A manual threshold can also be
set.
Ver 5.5 78
ULTRAVIM APPLICATION MANUAL
Search – Range
The Search Range tells the software how many pixels it can look for a character from the
expected position. Setting a higher number here will find characters that are misplaced,
but will slow down the search.
Search – Timout
This parameter tells the software how long it can search for failing characters. If the time
is greater than this setting, the software will break out of the search and report the best
match that it could find.
Search – Speed
The Search Speed is how fast the software looks for the object. To speed up the search,
larger search steps are taken. For a large object such as a logo, it may help to set the
speed to VERY FAST.
Search – Color
This attribute describes the color of the mark and the background. The default is WHITE
ON BLACK. If this is not set correctly, the Learn function will not find the characters
correctly.
Search – Threshold
The Threshold is used by the software to find the characters when the user clicks on the
Learn button. Setting this to LOW can help find characters that run into each other.
Setting this to HIGH can help the software find faint characters. The AUTOMATIC
setting will work for most parts.
Search – Sensitivity
This parameter adjust the sensitivity of the correlation. Setting it to LOW will give a
higher correlation result. You may want to set a logo to LOW, for example.
Ver 5.5 79
ULTRAVIM APPLICATION MANUAL
1 3
INSPECT
2 5
BUSY
4
RESULT
ORIENTATION
2. The software sets the BUSY line high and acquires an image.
3. The host acknowledges the busy line by setting the INSPECT line low.
4. The software completes the inspection and sets the result and orientation lines.
5. The software sets the busy line low, letting the host know the results are ready.
If the INSPECT line is not low at this point, the software will time out after
3 seconds, and an error will be shown on the screen.
Ver 5.5 80
ULTRAVIM APPLICATION MANUAL
B1B0 = 00 Invalid
B1B0 = 10 Fail
B1B0 = 01 Rework
B1B0 = 11 Pass
B3B2 = 00 Pass
B1B0 = 10 Fail
Ver 5.5 81
ULTRAVIM APPLICATION MANUAL
9.12 Troubleshooting
The software responds to the host inspect command, but then it says “Timed Out”.
1. The host must set the INSPECT line low after it detects that the busy line
is high. If the INSPECT line is not low within 3 seconds, the inspection
will time out.
Ver 5.5 82
ULTRAVIM APPLICATION MANUAL
Ver 5.5 83
ULTRAVIM APPLICATION MANUAL
After the part file has been loaded, the search area must be selected by the user. To do
this, the user can use the mouse to draw the search box on the image. The search box
should be close to all the tips of the leads on the part. Below are some examples of
search boxes that are drawn correctly and some that are not correct.
Ver 5.5 84
ULTRAVIM APPLICATION MANUAL
Ver 5.5 85
ULTRAVIM APPLICATION MANUAL
10.4 Options
Inspection Cycles The inspection cycles should be left at “NORMAL”. If you select
“Demo” the system will inspect the same part until you click on the
cancel button. Demo 1.0 inspects once per second, Demo 2.0
inspects once every two seconds, etc.
SPC Subgroup Size The SPC subgroup size shows how many values to average before
saving in the SPCSTAT.CSV file. It is best to set this number to 5
or 10. If you have large lot sizes, the SPC file can get very large.
Especially if this value is set to 1.
Ver 5.5 86
ULTRAVIM APPLICATION MANUAL
Auto-Delete If you turn on the Auto-Delete option, files on the E:\ drive
will automatically be deleted after a certain number of days.
Specify the number of days in the edit box to the right of
the Auto-Delete choice.
Graphics This determines if the graphics are shown. If you have a host
system that is very fast (above 10,000 uph) you should turn the
graphics off to speed up the system. If not, it is best to leave
graphics on.
Image You can use the save and load buttons to save or load TIF images.
If you have a part that looks strange, or think there is a problem
with your system, save the image and send it to customer support.
Setup The setup area of the options dialog is used to set the camera
and digital IO options. You must have password level 4 to
change these options.
Ver 5.5 87
ULTRAVIM APPLICATION MANUAL
A. Adding a Printer
This chapter shows how to add a printer to the UltraVim system. Before you start the
steps in this chapter, make sure the printer has paper in it, is connected to the PRINTER
port on the UltraVim, and is plugged in.
From the “My Computer” screen, double click on the “Printers” icon.
Ver 5.5 88
ULTRAVIM APPLICATION MANUAL
From the Printer Wizard screen, select “My Computer”, then click on Next.
Ver 5.5 89
ULTRAVIM APPLICATION MANUAL
Click on the printer manufacturer in the list on the left of the screen, then select the
printer in the right list on the screen. If your printer is not shown, you need a floppy disk
with drivers on it. Insert the floppy disk in the A: drive, then click on “Have Disk” to
select the drivers. When you have selected the printer, click on Next.
Ver 5.5 90
ULTRAVIM APPLICATION MANUAL
Now you can click on the Finish button. If you have installed the printer correctly, a test
page will print out.
Before you buy a printer for the UltraVim system, you should make sure it is compatible
with Windows NT. Windows NT has many default drivers for printers from Cannon,
Fujitsu, Hewlett Packard, NEC, and Okidata. Check with the printer manufacturer for
more details.
Ver 5.5 91
ULTRAVIM APPLICATION MANUAL
B.1 Hardware
In this example a 3COM Etherlink XL PCI 3C900-Combo NIC is used. When you
select a network card, make sure it has Window NT 4.0 drivers that are available on a
floppy disk.
First, power down the UltraVim system and unplug the power cord. Then, remove the
cover and carefully pull the two 50-pin ribbon cables off the digital I/O board so you have
access to the network PCI slot.
If you have the small STC computer chassis, bend the mounting bracket on the Network
Interface Card as shown below. Remove the existing bracket (be careful to not drop the
mounting screw or nylon washer in the computer).
Install the network card in the slot labeled "Network Card". Make sure it is fully seated
in the PCI slot. Put the mounting screw and washer back in place to secure the card.
Finally, put the cover back on the UltraVim computer module.
Ver 5.5 92
ULTRAVIM APPLICATION MANUAL
From the “My Computer” screen, double click on the “Control Panel” icon.
Ver 5.5 93
ULTRAVIM APPLICATION MANUAL
When you click on "Network", you will be asked if you want to install Windows NT
Networking. Click "Yes" to start the Network Wizard.
Click "Next" on the first screen of the Setup Wizard to connect the computer to a
Network Adapter.
The next screen will let you search for a Network Adapter, or select from a list. You
should have a floppy disk with network drivers on it for your network card. Insert this
disk in the floppy drive now. If you purchased the 3COM 3C900 card, insert Setup Disk
2 in the floppy drive.
Ver 5.5 94
ULTRAVIM APPLICATION MANUAL
Click on "Select From List..." to get to the "Select Network Adapter" screen.
From the "Select Network Adapter" screen click on the "Have Disk..." button.
Make sure your driver disk is in the floppy drive, and click "OK" to load the driver list
from the A: drive.
Ver 5.5 95
ULTRAVIM APPLICATION MANUAL
The Network Adapter should now be displayed in the list shown below. Click "Next" to
go to the next screen.
Select the correct protocols for your network. If you are not sure, click on all three
selections as shown in the screen below. When you have selected the protocols, click on
the "Next" button.
Ver 5.5 96
ULTRAVIM APPLICATION MANUAL
The network services shown below should be selected by default. Click on the "Next"
button to move to the next screen.
Windows NT is now ready to install the components you selected. Click "Next" to start
the installation.
Windows NT will now copy some files from both your A: and C: drives. When you see
the screen below, click Continue. If you see a message saying that a file could not be
found, change the A:\ to C:\ and try again.
Ver 5.5 97
ULTRAVIM APPLICATION MANUAL
If you installed the TCP/IP protocol you will see the screen shown below. If you have a
DHCP server on your network click "Yes". If you are not sure, click "No".
If you installed the TCP/IP protocol you will be asked for an IP address for the computer.
Ask your System Administrator to give you this address. If you are not sure, you can
make up an address and change it later.
Ver 5.5 98
ULTRAVIM APPLICATION MANUAL
Now the Setup Wizard will install the network bindings. Click "Next" to continue. If the
Wizard cannot find a file, change the A:\ to C:\ and try again.
After the bindings are installed, you can click on the "Next" button to finish the
installation.
Ver 5.5 99
ULTRAVIM APPLICATION MANUAL
Enter the computer name and the workgroup. You may wish to enter the serial number of
your UltraVim system for the computer name.
Click on "Finish" to restart your computer and initialize the network. Make sure you
have connected the network cable to your network card before you restart the system.
After the computer restarts, you should not see any errors. If you see a message that says
"At least on service failed to start" your Network Card has not been installed correctly.
Start File Manager and select "Connect Network Drive" from the "Disk" directory. You
should now see a list of computers that are connected to your network. Use this screen to
map a drive to the computer. To map a printer to your computer, follow the directions in
Chapter 9, but select "Network Printer Server" instead of "My Computer.
While you are in the UltraVim software, press CTRL-ALT-F7 to access the custom report
editor. The custom report must is saved in the D:\UVIM\Files\REPORT32.INI file. This
file is updated automatically as you make changes in the Custom Reports dialog.
To enable the custom report, click on "Enable Custom Report" in the dialog. Note that if
you click on "Leaded Parts" or "BGA Parts" the screen will change as it loads the data
from the REPORT32.INI file.
Up to 45 lines can be defined in your custom report. The following example shows a
custom report definition:
[CustomQFP]
Enabled=1
01=
02=
03=HEADR_RES
04=
05=TOTAL_RES
06=
07=
08=
09=HEADR_AVE
10=
11=COPLQ_AVE
12=TPOFF_AVE
13=LSKEW_AVE
14=BLEAD_AVE
15=WIDTH_AVE
16=STOFF_AVE
17=PITCH_AVE
18=TTT13_AVE
19=TTT24_AVE
20=
21=
22=
23=Scanner Technologies
24=Minneapolis, MN 55447
25=
Note that it is good to use blank lines to separate sections of the report. When you do
this, you can also click on "Enable Line Graphics" to have the software automatically
draw lines around the different sections you create.
You can add text to any line, but be aware that not all text will get automatically
translated into other languages. Make sure you use commas to separate your columns if
you add custom text.
Note that some measurements, such as pitch, may not have a result or a tolerance value.
These line types must be combined with a variable type to create a custom line.
D. Host Communications
This chapter describes the protocol required to communicate with the UltraVim system
using the digital I/O cables. When the UltraVim software starts, it is automatically
checking the digital I/O lines for commands from the host. This lets the user inspect part
in manual mode, or in host mode without changing any of the settings.
There are two OPTO22 50-pin connectors on the UltraVim PC. They are labeled "Input"
and "Output". The input connector used 16 lines for commands and data input, and the
output connector uses 24 lines for results, and other outputs.
The figure below shows the layout of the connectors. Each pair of pins carries a signal
and is labeled A0 – C7. Note that C0 – C7 are not used on the input connector.
Ground 50 z z 49 +5V
Ground 48 z z 47 A0
Ground 46 z z 45 A1
Ground 44 z z 43 A2
Ground 42 z z 41 A3
Ground 40 z z 39 A4
Ground 38 z z 37 A5
Ground 36 z z 35 A6
Ground 34 z z 33 A7
Ground 32 z z 31 B0
Ground 30 z z 29 B1
Ground 28 z z 27 B2
Ground 26 z z 25 B3
Ground 24 z z 23 B4
Ground 22 z z 21 B5
Ground 20 z z 19 B6
Ground 18 z z 17 B7
Ground 16 z z 15 C0
Ground 14 z z 13 C1
Ground 12 z z 11 C2
Ground 10 z z 9 C3
Ground 8 z z 7 C4
Ground 6 z z 5 C5
Ground 4 z z 3 C6
Ground 2 z z 1 C7
From the UltraVim software, press the following keys at the same time to display the
Digital I/O Test dialog:
CTRL – ALT – F4
This will show the status of the 16 input lines and the 24 output lines. Note that the
UltraVim will not process any host commands while this dialog is being displayed. You
can test individual output lines by clicking on the “0/1” button next to the line you wish
to test. Also, you can invert the input or output lines by clicking on the “Invert Signals”
checkbox.
ID A7 A6 A5 A4 Command
0 0 0 0 0 Load Part
1 0 0 0 1 Inspect Part
2 0 0 1 0 Inspect Part
3 0 0 1 1 Inspect Part
4 0 1 0 0 Inspect Mark
5 0 1 0 1 Inspect Mark
6 0 1 1 0 Inspect Dambar
7 0 1 1 1 Inspect Lead Frame
8 1 0 0 0
9 1 0 0 1
10 1 0 1 0 Set Autosave On
11 1 0 1 1 Set Autosave Off
12 1 1 0 0 Get Part Angle
13 1 1 0 1
14 1 1 1 0 Turn Lights On
15 1 1 1 1 Clear Line
When the "Load Part" command is sent to the UltraVim, the file name will be created
from the input data lines B7 – B0 (B7B6B5B4B3B2B1B0). For example, to load the file
"0000001.PAR" you should set the data lines B7 – B1 to 0 and data line B0 to 1. Note
that you cannot have a file named "0000001.PAR" and a file named "00000001.BGA" on
the same system since the UltraVim has no way of telling the file extension from the eight
data lines.
Use the C code below as an example of how to load a part file from the host. The code
will set the data bits for part file "00000001", then send the command to load the file.
Assume the following:
///////////////////////////////////////////////////////////////////
//
// Example C code to load part “00000001”
//
///////////////////////////////////////////////////////////////////
BOOL LoadPartExample()
{
// Check if the MODE bit is on
if (GetBit(A6)==0) return FALSE;
The output channels B1 and B0 show the results of the inspection. The four possibilities
are shown below:
Inspection Results
B1 B0 Result
0 0 Invalid
1 0 Failure
0 1 Rework
1 1 Pass
The following diagram shows the general timing of the host communications to inspect a
part with the UltraVim. Assume that a part file has already been loaded.
1 3
Host REQ
2
UltraVim ACK
4 5
UltraVim BUSY
1. The host must hold a part over the reticle, set the command bits (A4-A7)
to the Inspect command, and send the REQ bit high to start the inspection.
2. As soon as the UltraVim sees the REQ bit high, it sets the ACK bit high
and starts the image acquisition.
3. The host acknowledges the start of inspection by setting the REQ bit low.
The REQ line must be set low as soon as the ACK line goes high.
4. When the image has been acquired the UltraVim sets the BUSY line high.
The host may move the part away from the reticle when this line goes high.
5. When the results have been computed, the UltraVim sets the result bits,
then sets the ACK and BUSY lines low. This tells the host that the
results are ready.
Use the C code below as an example of how to inspect a part from the host. Assume the
following:
return TRUE;
}
///////////////////////////////////////////////////////////////////
//
// Example C code to inspect a part
//
///////////////////////////////////////////////////////////////////
BOOL InspectPartExample()
{
// Check if the MODE bit is on
if (GetBit(A6)==0) return FALSE;
After an inspection is complete, the UltraVim returns the part results and its position over
the center of the reticle. The channels B2 – C7 contain the X,Y position information.
The following example code shows how to calculate the position in mils.
///////////////////////////////////////////////////////////////////
//
// Example C code to read part position from reticle center
//
///////////////////////////////////////////////////////////////////
int GetPositionX()
{
int nPos = 0;
return nPos;
}
///////////////////////////////////////////////////////////////////
//
// Example C code to read part position from reticle center
//
///////////////////////////////////////////////////////////////////
int GetPositionY()
{
int nPos = 0;
return nPos;
}
The following diagram shows the coordinate system of the UltraVim in relation to the
reticle:
+Y
-X +X
-Y
The diagram below shows the general move sequence for the handler.
When the inspect command is given to the UltraVim software, the lights are turned on
and the image is acquired. If you have a very high speed system, you can save about
50ms by sending the UltraVim software the command to turn on the lights (ID=14) before
the part is in place.
A. Send command to turn lights on, then move the handler above the reticle.
B. Move the part down to position.
C. When part is in place, send the inspect command.
A
B
After an inspection is complete, the UltraVim returns can report the part angle with
respect to the center of the reticle. To get this information, you must send the command
to get the angle by setting A7=1, A6=1, A5=0, and A4=0 (ID=12). After the command
has been set, set the request line high to report the angle in output channels C0 – C7.
This will report the angle up to +/- 12.7 degrees. The following example code shows how
to calculate the angle.
///////////////////////////////////////////////////////////////////
//
// Example C code to read part angle from reticle center
//
///////////////////////////////////////////////////////////////////
double GetAngle()
{
double dAngle = 0.0;
return dAngle;
}
The UVIM uses opto-couplers on it’s digital I/O signals to electricaly isolate the
UVIM and the customers equipment. This isolation requires a D.C. voltage source
to power the signals. If the customer does not have a voltage source available,
then the UVIM can be equipped with a DC to DC convertor whose +5vdc is
available on pin 49 and return (ground) on any even pin of the connectors. The
customer can use the voltage from these pins to power the signals that are wired
between the UVIM and their equipment. If the customers equipment already has a
voltage source of 5vdc to 32vdc @ ½ amp then the UVIM can be configured with
out the DC to DC convertor and the customer can connect their positive voltage to
pin 49 and their ground to any even pin of the connectors. There are three basic
configurations in which a customer can wire their signals to the optically isolated
signals on the UVIM digital I/O connector.
The customer can simply connect a 50 pin ribbon cable between the UVIM digital
I/O connectors and two 24 position OPTO-22 boards. The ribbon cable should
have standard 0.1” pitch keyed connectors at each end to ensure pin #1 of the
UVIM connector gets connect to pin #1 of the OPTO-22 board. The customer
should then follow the directions that came with the OPTO-22 board to wire their
signals to the appropriate UVIM signal. To power the OPTO-22 modules the
UVIM needs to be configured with the DC to DC convertor.
U V IM In p u t O u tp u t
5 0 C o n d u c to r 5 0 C o n d u c to r
R i b b o n C a b le O P T O - 2 2 S t y le R i b b o n C a b le
B o a rd s
W ir in g f r o m W ir in g f r o m
C u s to m er O u tp u t C u s to m er In p u t
Customer wires Directly to UVIM Digital I/O Connector, but They Have No
Isolation Power Supply
If the customer wishes to wire directly to the UVIM, but has no isolating power
supply available, they can order the UVIM with a DC to DC convertor. The DC to
UVIM Customers
Equipment
UVIM Input
Connector
CURRENT
LIMITER
+5VDC
50 49 ISOLATED
UVIM Input
Ch l0
Logic
47 A0
48
A1
46 45
Optical
I l ti
+5VDC
UVIM DC to DC
POWER CONVERTOR
Ground
OPTIONA
LPULL-UP +5VDC
RESISTO 50 49 ISOLATED
R
48 47 A0
UVIM
OChannel
t t
0 Logic 46 45
A1
Optical
I l ti
UVIM Output Connector
Customer wires Directly to UVIM Digital I/O Connector and They Have an
Isolation Power Supply
If the customer wishes to wire directly to the UVIM and they have a source of
5vdc to 32vdc @ ½ amp, they can order the UVIM without a DC to DC convertor.
The customer should wire the positive side of their supply to pin 49 of each
connector and the return (ground) side to any even pin on both connectors.
Illustrated below is the internal configuration of the UVIM to aide the customer in
determining how they will connect their signals to the UVIM.
UVIM Customers
Equipment
UVIM
IConnecto
t
CURREN
TLIMITE
R
50 49
UVIM Input
Channel 0
Logic
48 47
A0
46 45
A1
Optical
I l ti
+5 -
32VDC Customer
Powe
Suppl
Groun
d
OPTIONAL
PULL-UP
RESISTOR 50 49
48 47
A0
UVIM Output
Channel 0
Logic
46 45
A1
Optical
I l ti
UVIM Output
C t
WARNING:
There is no current limiting on the UVIM output transistor, therefore it is
the customers responsibility to ensure that the load they connect they to the
UVIM output does not exceed the transistor ratings listed in the table below.
The asynchronous communications protocol differs slightly from the industry standard
communications described in Appendix D, but it offers the added flexibility of sending
commands to different cameras and software packages at the same time.
To set the protocol, make sure you select "Option 2" at the bottom of the digital IO dialog
box.
There are two OPTO22 50-pin connectors on the UltraVim PC. They are labeled "Input"
and "Output". The input connector used 16 lines for commands and data input, and the
output connector uses 24 lines for results, and other outputs. The figure below shows the
layout of the connectors. Each pair of pins carries a signal and is labeled A0 – C7. Note
that C0 – C7 are not used on the input connector.
Ground 50 z z 49 +5V
Ground 48 z z 47 A0
Ground 46 z z 45 A1
Ground 44 z z 43 A2
Ground 42 z z 41 A3
Ground 40 z z 39 A4
Ground 38 z z 37 A5
Ground 36 z z 35 A6
Ground 34 z z 33 A7
Ground 32 z z 31 B0
Ground 30 z z 29 B1
Ground 28 z z 27 B2
Ground 26 z z 25 B3
Ground 24 z z 23 B4
Ground 22 z z 21 B5
Ground 20 z z 19 B6
Ground 18 z z 17 B7
Ground 16 z z 15 C0
Ground 14 z z 13 C1
Ground 12 z z 11 C2
Ground 10 z z 9 C3
Ground 8 z z 7 C4
Ground 6 z z 5 C5
Ground 4 z z 3 C6
Ground 2 z z 1 C7
From the UltraVim software, press the following keys at the same time to display the
Digital I/O Test dialog:
CTRL – ALT – F4
Make sure you select "Option 2" at the bottom of the dialog box to set the protocol. This
dialog will show the status of the 16 input lines and the 24 output lines. Note that the
UltraVim will not process any host commands while this dialog is being displayed. You
can test individual output lines by clicking on the “0/1” button next to the line you wish
to test. Also, you can invert the input or output lines by clicking on the “Invert Signals”
checkbox.
Note that the REQ line is used to send a command to the main UltraVim software. The
command is determined by CMD1, CMD2, CMD3, and CMD4. The RQ2 line is used to
request a command from device number two, usually the Mark Inspection program. The
acknowledgement line AK2 responds to request line RQ2, and the results are stored in the
RS2A and RS2B lines.
The following input commands are used only for the REQ (A0) line. The RQ2, RQ3, and
RQ4 lines are only used to request and inspection of that device.
ID A7 A6 A5 A4 Command
0 0 0 0 0 Load Part
1 0 0 0 1 Inspect Part
2 0 0 1 0
3 0 0 1 1
4 0 1 0 0
5 0 1 0 1
6 0 1 1 0
7 0 1 1 1
8 1 0 0 0
9 1 0 0 1
10 1 0 1 0 Set Autosave On
11 1 0 1 1 Set Autosave Off
12 1 1 0 0 Get Part Angle
13 1 1 0 1 Get Y Position
14 1 1 1 0 Turn Lights On
15 1 1 1 1 Clear Line
When the "Load Part" command is sent to the UltraVim, the file name will be created
from the input data lines B7 – B0 (B7B6B5B4B3B2B1B0). For example, to load the file
"0000001.PAR" you should set the data lines B7 – B1 to 0 and data line B0 to 1. Note
that you cannot have a file named "0000001.PAR" and a file named "00000001.BGA" on
the same system since the UltraVim has no way of telling the file extension from the eight
data lines.
The following error codes are used only for the REQ (A0) line. The RQ2, RQ3, and RQ4
lines only use the AK2, AK3, and AK4 lines to determine the response of the intended
device.
The output channels RS1A (B0) and RS1B (B1) show the results of the inspection from
the UltraVim software. This is the main UltraVim system that performs an inspection of
a leaded or BGA device. The four possibilities are shown below:
Inspection Results
B1 B0 Result
0 0 Invalid
1 0 Failure
0 1 Rework
1 1 Pass
For Mark Inspection, the RS2A (B2) and RS2B (B3) lines are used. The RS2A line
indicates if the mark passed, and the RS2B line shows if the orientation passed.
The following diagram shows the general timing of the host communications to inspect a
part with the UltraVim. Assume that a part file has already been loaded.
1 3
Host REQ
2
UltraVim ACK
4 5
UltraVim BUSY
1. The host must hold a part over the reticle, set the command bits (A4-A7)
to the Inspect command, and send the REQ bit high to start the inspection.
2. As soon as the UltraVim sees the REQ bit high, it sets the ACK bit high
and starts the image acquisition.
3. The host acknowledges the start of inspection by setting the REQ bit low.
The REQ line must be set low as soon as the ACK line goes high.
4. When the image has been acquired the UltraVim sets the BUSY line high.
The host may move the part away from the reticle when this line goes high.
5. When the results have been computed, the UltraVim sets the result bits,
then sets the ACK and BUSY lines low. This tells the host that the
results are ready.
The general sequence for all Mark Inspection commands is shown below:
1 3
Host RQ2
2 4
Mark AK2
1. The host must make sure the part is under the mark inspection camera.
Then the host sets the RQ2 line high.
2. The Mark Inspection program sees the RQ2 line high, and sets the AK2
line after the image is acquired. The host may move the part as soon as AK2 goes
high.
3. The host acknowledges the start of inspection by setting the RQ2 line low.
The RQ2 line must be set low as soon as the ACK2 line goes high.
4. When the image has been processed and the RS2A and RS2B results are
ready, the mark inspection system sets the AK2 line low again.
Note that this timing diagram applies to the RQ3 and RQ4 lines as well. See the table
below:
Use the C code below as an example of how to inspect a part from the host. Assume the
following:
///////////////////////////////////////////////////////////////////
//
// Example C code to turn lights on
//
///////////////////////////////////////////////////////////////////
BOOL TurnLightsOn()
{
// Check if ACK bit is low
if (GetBit(A1)==1) return FALSE;
return TRUE;
}
///////////////////////////////////////////////////////////////////
//
// Example C code to inspect a part
//
///////////////////////////////////////////////////////////////////
BOOL InspectPartExample()
{
// Check if ACK bit is low
if (GetBit(A1)==1) return FALSE;
After an inspection is complete, the UltraVim returns the part results and its position over
the center of the reticle. The channels C0 – C7 contain the X position information. The
following example code shows how to calculate the position in mils. To get the Y
position information, send command 13 and the C0 – C7 lines will show the Y
information instead of the X information.
///////////////////////////////////////////////////////////////////
//
// Example C code to read part position from reticle center
// The position could be X, or Y if the command 13 was sent
// after the inspection.
//
///////////////////////////////////////////////////////////////////
int GetPosition()
{
int nPos = 0;
return nPos;
}
The following diagram shows the coordinate system of the UltraVim in relation to the
reticle:
+Y
-X +X
-Y
The diagram below shows the general move sequence for the handler.
When the inspect command is given to the UltraVim software, the lights are turned on
and the image is acquireE. If you have a very high speed system, you can save about
50ms by sending the UltraVim software the command to turn on the lights (ID=14) before
the part is in place.
A. Send command to turn lights on, then move the handler above the reticle.
B. Move the part down to position.
C. When part is in place, send the inspect command.
A
B
After an inspection is complete, the UltraVim returns can report the part angle with
respect to the center of the reticle. To get this information, you must send the command
to get the angle by setting A7=1, A6=1, A5=0, and A4=0 (ID=12). After the command
has been set, set the request line high to report the angle in output channels C0 – C7.
This will report the angle up to +/- 12.7 degrees. The following example code shows how
to calculate the angle.
///////////////////////////////////////////////////////////////////
//
// Example C code to read part angle from reticle center
//
///////////////////////////////////////////////////////////////////
double GetAngle()
{
double dAngle = 0.0;
return dAngle;
}
The diagram below shows the proper centering of the parts of the inspection reticle.
Note that each part must be the same distance from the center of the reticle. For BGA
parts, the datum lines of the reticle still apply, even though the reticle is not required for
the inspection.
Pitch Pitch
The Digital I/O Setup can be set up many ways, but to maintain asynchronous
communication with both the UltraVim and Mark Inspection software, the following
method is recommended. The steps for this setup are:
INI Files
The recommended Digital I/O setup for multiple parts is shown below. Note that the
commands in the HWARE32.ini file are repeated starting with ID=08. Also, in the
MARK32.ini file, the command changes to INSMZ starting with ID=08.
ID A7 A6 A5 A4 HWARE32.ini MARK32.ini
0 0 0 0 0 00=LOADP 00=INSPM
1 0 0 0 1 01=INSPP 01=INSPM
2 0 0 1 0 02=INSPZ 02=INSPM
3 0 0 1 1 03=ANGLE 03=INSPM
4 0 1 0 0 04=GETYP 04=INSPM
5 0 1 0 1 05=LIGHT 05=INSPM
6 0 1 1 0 06=NOCMD 06=INSPM
7 0 1 1 1 07=CLEAR 07=INSPM
8 1 0 0 0 08=LOADP 08=INSMZ
9 1 0 0 1 09=INSPP 09=INSMZ
10 1 0 1 0 10=INSPZ 10=INSMZ
11 1 0 1 1 11=ANGLE 11=INSMZ
12 1 1 0 0 12=GETYP 12=INSMZ
13 1 1 0 1 13=LIGHT 13=INSMZ
14 1 1 1 0 14=NOCMD 14=INSMZ
15 1 1 1 1 15=CLEAR 15=INSMZ
///////////////////////////////////////////////////////////////////
// Example C code to Inspect Part #1 on UltraVim
///////////////////////////////////////////////////////////////////
BOOL UltraVimInspectPart1()
{
// Set the command bits to “001” (INSPP)
SetBit(A6,0);
SetBit(A5,0);
SetBit(A4,1);
///////////////////////////////////////////////////////////////////
// Example C code to Inspect Part #2 on UltraVim
///////////////////////////////////////////////////////////////////
BOOL UltraVimInspectPart2()
{
// Set the command bits to “010” (INSPZ – No Grab)
SetBit(A6,0);
SetBit(A5,1);
SetBit(A4,0);
///////////////////////////////////////////////////////////////////
// Example C code to Inspect Part #1 on UltraMark
///////////////////////////////////////////////////////////////////
BOOL UltraMarkInspectPart1()
{
// Set the command bit to “0” (INSPM)
SetBit(A7,0);
///////////////////////////////////////////////////////////////////
// Example C code to Inspect Part #2 on UltraMark
///////////////////////////////////////////////////////////////////
BOOL UltraMarkInspectPart2()
{
// Set the command bit to “1” (INSMZ - No Grab)
SetBit(A7,1);