3. Check the current output setting on the display (12). Fr. 1 = low cutting portion at high coagulation intensity
Fr. 9 = great cutting portion at low coagulation intensity
4. Optional: use buttons (15) or (18) to enable pulse or blend mode.
Blend mode:
5. Use the foot switch or the switch on the handpiece to activate The signal switches to a permanent signal from cutting to
the device. coagulation. The effect level determines the length of the
The indicator (14) will be lit during high frequency transition.
administration. bL. 1 = great cutting portion at low coagulation intensity
Monopolar coagulation bL. 9 = low cutting portion at high coagulation intensity
1. Optional: use program buttons (43/44/45) to open the setting.
Enabling/disabling effect mode
2. Optional: use buttons (21) and (23) to raise and lower the output,
Press effect button Pulse (36)/ Blend (37)/ Force (38).
respectively.
Mode is active when corresponding button is lit.
3. Check the current output setting on the display (19).
Adjusting effect levels
4. Optional: enable force mode using the button (22).
1. Press and hold effect button Pulse/Blend/Force for 2 seconds.
5. Use the foot switch or the switch on the handpiece to activate
The display switches to effect level mode (30/32/34). The figure
the device.
(31/33/35) indicates the effect level.
The indicator (20) will be lit during high frequency
2. Use buttons (39/40, 41/42) to increase/reduce the effect level.
administration.
Bipolar coagulation 3. To go back, press the effect button briefly.
1. Optional: use program buttons (43/44/45) to open the setting. Save
Programs
2. Optional: use buttons (27) and (29) to raise and lower the To save the current settings, press and hold the buttons (43) or
output, respectively. (44) for 2 seconds.
3. Check the current output setting on the display (24). Load
4. Optional: enable bipolar automatic control using the button 1. Press buttons (43) or (44). The button is lit.
(28). The stored settings are loaded.
The device comes on automatically as soon as the active 2. Press again to disable the program.
electrode comes into contact with the tissue.
Recall
5. If automatic control is disabled: use the foot switch or the When the device is switched off, the most recent settings will be
switch on the handpiece to activate the device. stored automatically in Recall memory. When the device is
The indicator (26) will be lit during high frequency switched back on, the default values will be restored. Press Recall
administration. (45) to load the settings you used last.
Effects Modes
Pulse mode:
Signal toggles between cutting and coagulating. The corresponding Söring GmbH Tel. +49 (0) 4106 6100 0
interval used for the signal portion is set via the effect level. Medizintechnik Fax +49 (0) 4106 6100 10
Justus-
Justus-von-
von-Liebig-
Liebig-Ring 2 E -Mail info@soering.com
Söring
Söring
PL. 1 = great cutting portion at low coagulation intensity Dok. Revision 03-
03-471R01
471R01 25451 Quickborn Web www.soering.com
PL. 9 = low cutting portion at high coagulation intensity 2008--11-17
2008