LEMBAR PERSETUJUAN

NAMA Jabatan Ttd Tgl

Disahkan Arif Suputra Yusup Direktur PT. Hospi


Medik Indonesia
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

I. DESKRIPSI PERALATAN DAN SPESIFIKASI SUMBER RADIASI YANG


DIGUNAKAN

1 NeuViz 16 Classic Hardware (International)

Gantry
Scan Control Panel
Controls and displays for gantry tilt, patient couch elevation and stroke are located on
both sides of the gantry.
Scan Control Box
Gantry and patient couch controls and displays are located conveniently at the
operator’s console. Additional functions include emergency stop, intercom, and scan
enable/pause buttons.
Gantry Aperture: 700 mm diameter
Adapt to the popular patient, convenient for intervention process.
Gantry Tilt: -30° to +30°; 0.5° increments. Adapt to all clinical examinations.
AutoVoice:
A standard set of commands for patient communication: before, during and after
scanning.
Multi-lingual AutoVoice:
Commands for patient communication in multiple languages including: English, French,
Spanish, Italian, German, Chinese, Japanese, Hebrew, Arabic, Russian, Turkish,
Georgian, Danish, Norwegian, Swedish, and Dutch. Also provides the ability to record
customized messages.
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

Patient Couch
Longitudinal motion
Moveable Range: 0-1550 mm
Scannable Range: 0-1450 mm
Speed: 1 to 100 mm/sec
Position accuracy: ±0.25 mm
Vertical motion
Range: 430 to 970 mm above floor; 1.0 increment
Table load capacity: 205 kg (452 lbs) with 0.25mm z-axis accuracy
Floating tabletop: Carbon-fiber table top with foot pedal and push buttons
On both sides for patient quick release.
Table Accessories
Table accessories include patient restraint straps, a standard head holder, a table pad,
cushions, and pads for optimal patient positioning and comfort.

23" LCD Monitor


Resolution 1920 x 1080

COMPUTER TABLE
This is computer table to put computer on.

Operation Manual
The user quick guidance for operations of NeuViz 16 Classic, the reference for
operators and radiologists.

Scan and image acquisition


Reliable, maximized system performance allows clinicians to remain focused on patient
care. The NeuViz 16 classic CT is perfectly balanced, combining power and flexibility
that maximizes image quality, speed and throughput while lowering patient dose.

Generator
The generator uses modern, low-voltage slip ring technology to provide a constant high
voltage to the CT x-ray tube assembly.
Output capacity: 28 kW
kV selections: 80, 100, 120, 140 kVp
mA selections: 10 to 233 mA

X-ray Tube
NeuViz 16 Classic has dynamic focal spots tube of Dunlee. Dynamic Focal Spot (DFS)
doubles the data sampling density from the detectors effectively doubling the number of
detectors and providing high spatial resolution in axial and spiral scanning.

Anode storage capacity: 4.0 MHU


Maximum cooling rate: 677 kHU/min
Focal spot (IEC): 0.4mm x 0.7 mm (small) | 0.6 mm x 1.3 mm (large)
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

Detector
Detector design is fundamental to the objective of acquiring high quality images while
minimizing patient dose.
NeuViz 16 Classic has brand-new ultra-efficentive detector system that minimize the
separation between elements to always provide the highest geometric detector
efficiency. Direct-to-digital signal conversion reduces dose and improves image quality.
Provide the stable guarantee to get high quality image.

 Modularization Design;
 Nanometer Connection;
 Super thin sensitization layer;
 Ultra high sensitivity;
 Ultra short afterglow time;
 Perfect conversion efficency;

Material: Solid State - Gadolinium Oxysulfide (GOS)


Slip Ring: Capacitive - 1.1 Gbps transfer rate
Data Sampling Rate: Up to 2320 views/revolution/element
Slice Collimations: 16x1.25mm, 16x0.625mm, 8x0.625mm, 4x0.625mm
Slice Increment: 0.1mm- 20.0mm

Image Quality
Spatial Resolution (High):
In-plane: 15lp/cm @ cutoff
Low Contrast Resolution: 4.0mm@ 0.3%

O-Dose
Lower dose is a philosophy, a set of principles and practices that ensure the best
possible outcomes with minimal risk to patients and staff. Neusoft focuses on system
design optimization, current (mA) optimization and increasing dosage awareness to
reduce the cumulative risk of radiation while obtaining high-quality images.
ACS (Automatic Current Selection)
Optimizes the dose for each patient based on the planned scan by suggesting the lowest
possible mAs settings to maintain constant image quality at low dose throughout the
exam.
Dose Modulation
Automatically controls the tube current (mA) longitudinally and rotationally, increasing
the signal over areas of higher attenuation to maintain constant image noise and
decreasing signal over area of less attenuation to save radiation dose.
Dose Displays
Volume CTDI (CTDIvol)
Dose Length Product (DLP)
Dedicated Pediatric Protocols
Age and weight-based infant and pediatric protocols help to balance the best clinical
results with minimal dose.
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

240° Exposure
Avoids direct exposure to operators while doing interventional or biopsy process and
decreases operation risk.
Organ Safe
Avoid direct exposure to some sensitive organs, for example lens, thyroid, galactophore,
gonad etc.
Brand-new Dectector Material
Compare with conventional detector, modularization detector design and new materials
will increase signal-to-noise ratio to get ideal aferglow.
Compacted design will reduce electron-noise and increase eometric efficiency.
Dose Check & Dose Report
Multiple checks realized by design of software ensure scanning dose not to exceed
presetted values.
All presetted dose could be evaluated before scanning, estimate CTDI and DLP.
Dose Report with DICOM conformity could be automatically created after scanning.
Comply of NEMA X25 Dose Check Standards.

Power requirements
-Power capacity: 50 KVA
-Voltage: 380VAC
-Voltage Variation: within 5% of the rating
-3-phase imbalance: less than 5%
-Frequency: 50Hz/60Hz ± 1Hz
-Power Ground Independent grounding system impedance≤4Ω, OR Integrated
grounding system impedance ≤1Ω

Environmental requirements
-Temperature of scanning room 18℃~24℃with variation less than 5℃/hour
-Humidity of scanning room 30%~60%(Non-condensing)
Temperature of operation room 18℃~28℃ with variation less than 5℃/hour
-Humidity of operation room 20%~80%(Non-condensing)
-Atmospheric pressure 70kPa~106kPa
-Temperature of transportation and storage -20℃~+55℃with variation less than
10℃/hour
-Humidity of transportation and storage10%~90% (Non-condensing)

2 NeuViz 16 Classic Software (International)

Daily Operations
The Daily Operations include Air Calibration, Tube Warm-up, Image Quality Valuation
and Disk Defragment.

Patient and Image Management


Patient Management
The Patient Management system includes Patient Registration, Schedule, and Patient
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

information archive/modify/delete/search.
Image Management and Archiving
Image archiving is organized according to the DICOM 3.0 hierarchical model and in a
DICOM 3.0 compliant image format. A lossless image compression/decompression
algorithm is used during image storage/retrieval to/from all local archives. Images can
be auto-archived to selected archive media.
Image Storage Capacity: 500 GB available
Number of uncompressed Images: 960,000 512x512 Matrix.

CD/DVD
CD/DVD is an archive solution for storing CT datasets. It provides an inexpensive,
reliable
method for high-speed recording. CD/DVD supports storing multiple patients in a single
writing session.
DVD Storage Capacity:
4.7 GB DVD: 512 X 512 Image Matrix = 15,000 typical number of compressed images
CD Storage Capacity:
700 MB CD: 512 X 512 Image Matrix = 1200 typical number of compressed images
CD Viewer
The CD Viewer application is included in each CD and can be automatically started on
the viewer’s PC. The common 2D image functions are supported in CD Viewer.

Scanning Modes
Spiral Scanning
Multiple contiguous slices acquired simultaneously with continuous table movement
during scans. Multiple, bi-directional acquisitions
Spiral exposure: Up to 100 sec. of uninterrupted spiral scanning.
Axial Scanning
Multiple-slice scan with up to 16 contiguous slices acquired simultaneously with
incremental table movement between scans.
Fused modes for reconstructing partial volume artifacts free thick slices from thin slice
acquisition.
Scan Times:
360° full scan: 0.78s, 1.0s, 1.5s, 2.0s
240° partial scan: 0.52s, 0.67s, 1.0s, 1.33s
Test Injection Bolus Timing(TIBT): This feature establishes the optimum delay time for
contrast injection. By using a test injection, a real-time graph of the enhancement in the
selected region of interest is displayed. The delay time is then selected to provide
optimal peak contrast enhancement and reduced contrast usage-ideal for CTA.
Bolus Tracking
Bolus Tracking is an automated injection planning technique that permits the user to
monitor actual contrast enhancement and initiate scanning at a pre-determined
enhancement level. Combine with the SAS option for full automation and efficacy.
Spiral Auto Start (SAS)
Spiral Auto Start integrates the injector with the scanner, allowing the technologist to
monitor the contrast injection to check for extravasation, and to initiate the scan (with
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

the predetermined delay) while in the scan room.


CT Angiography (CTA)
Specified protocols for different vessels, used with Vessel Analysis realize one click or
two click vessel extraction from contrast images. One button bone remove function
provided.

Reconstruction and Display


RapidView Reconstruction
RapidView reconstruction provides users with the images they desire at the time they
need them.
Reconstruction Rate: Up to 12 images per second (512x512 image matrix)
Cone Beam Reconstruction
Employing an Adaptive Spiral Tracking Reconstruction Algorithm (ASTR) enables
three-dimensional data acquisition and reconstruction in spiral scanning. This avoids
and/or corrects artifacts present in large collimation reconstructions by reducing pixel to
noise ratio, resulting in superior multi-slice image quality.
Reconstruction Modes
Concurrent: Axial and spiral modes - image reconstruction concurrent with acquisition.
Off-Line (batch): Background image reconstruction of user-defined groups of raw data
files with automatic image storage.
Add Reconstruction Enables quick and easy unplanned or modified reconstructions of
part or all of the images prospectively or retrospectively planned.
Reconstruction parameters
Any study can be set up to automatically reconstruct using various reconstruction
parameters. Exams can be tailored online while planning the scan, or during off-line
recon. Image reconstruction parameters include image matrix, filters, zoom and pan,
and archive.
Cine Display
Display a serise images as a movie to promote the clinical diagnositic efficiency.

Image Processing
The interactive image viewer is designed for fast, efficient and simple image review and
filming purposes. Images can be handled individually or in user-selected groups.
Image viewer window: Displays a single image or a selection of images.
Features: Zoom, Pan, Scroll Bar, Cine, Invert Image, Image Parameters Display.
Post-Processing Analysis Tools
2-D SlabViewer
MPR – Multiplanar Reformation
Curved Multiplanar Reformation (CPR)
Maximum Intensity Projection (MIP)
Minimum Intensity Projection (MinIP)
Average Intensity Projection (AIP)
3-D SSD Reconstruction
Volume Rendering
Advanced 3D application package for the optimal display and differentiation of
different organs is through independent control of color, opacity. The doctor can modify
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

the protocol and preset the threshold for the different display mode. NeuViz 16 Classic
also provides the bone remove function to help display the contrast vessel and soft
tissue. The software supports the transparent 3D reconstruction image to display the
bronchus, colon and skeleton system.
Virtual Endoscopy
The VE Application is a review function that allows user to perform a flythrough of
structures filled with contrast or air, including general vessels, the bronchus, and the
colon.
The software support SSD/VR display model to show the anatomical structure inside
and outside.
Vessel Measurable Analysis
Vessel Analysis (VA) offers a set of tools for general vascular analysis. Following
functions is supported in vessel Analysis Application:One click bone remove; Vessel
centerline extraction; Various review modes may be used: Volume Rensering, MIP,
MPR,CPR; Measurements are provided for vessel assessment, including maximum and
minimum cross section diameters, lumen areas.
Tissue Segmentation
Cut the tissue to display the non-surface injure and disease to promote the clinical
diagnosis.
Bone Removal
One click bone removal function is supported in both Volume application and Vessel
Analysis application. All automatic workflow, fast and whole bone removed, MIP, VR
and Transparent display model, structure and narrow measurement, hemangioma and
support analysis, to give the optimum scheme for clinic.
Volume Calculation
Volume Measurements of segmented tissues, based on special algorithm for the ROI,
the doctor can measure the hemorrhagic volume and node. For node analysis, the doctor
can measure and observe the nodes through the MPR, VRT to promote the clinical
diagnosis.
AutoFilming
The filming function allows the user to set up and store desired filming parameters. Pre-
stored protocols can also include auto-filming. The operator can film immediately after
each image, at the end of a series, or film after the end of a study and review images
prior to print. The operator can also automatically film the study at three different
windows and incorporate Combine Images functionality to manage large datasets. Basic
monochrome and color DICOM Print capability are supported.

Networking/Connectivity
Network Requirements
Network connections should be located within 10 feet of the console. Supports
10/100/1000Mbps (10/100/1000BaseT) network speeds.
For optimal performance, Neusoft recommends a minimum of 100Mbps network speed
(1Gbps preferred) and for the CT network to be segmented from the rest of the hospital
network.
DICOM Connectivity
A full implementation of the DICOM 3.0 communications protocol allows connectivity
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

to DICOM 3.0 compliant scanners, workstations, and printers.


DICOM MPPS
Enhances system performance DICOM Modality Worklist service class interface to
hospital HIS/RIS system. Modality Worklist automatically imports patient
demographics
and study information from the HIS/RIS system thereby improving speed and
eliminating data entry errors. DICOM Modality PPS (Performed Procedure Step) can
record patient information during the scan, and send them to HIS/RIS archive.
DICOM Modality Worklist
Provides HIS/RIS interface through DICOM modality worklist service class; enhances
clinical workflow by importing patient demographics and study information from an
information management system.
Service Software
Build in Remote service software. Connecting to Neusoft remote service center for
Remote failure diagnose, Trouble shooting and software upgrade.

3 ClearView
ClearView is an advanced iterative algorithm for double spaces, one is for original raw
data space and another is for image space. With three levels of Slight, Standard and
Ultra we can get perfect image quality with lower dose.

4 Main AC Cable4
Cable used to connect the power distribution system with the gantry.

5 Tool Box
A set of tools for daily maintenance to make sure the system to work with best status.

6 NMS Head Holder Assembly


To support patients’ head during scan. Extend couch scannable range.

7 Head Holder Cushion


A set of cushions for better patient comfort while doing head scanning.

8 Knee Joint Cushion


Position cushion used for Knee Joint study.

9 Belt1
Fasten patient on the couch to avoid motion artifact.

10 Belt2
Fasten patient on the couch to avoid motion artifact.

11 Belt3
Fasten patient on the couch to avoid motion artifact.
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

12 Belt4
Fasten patient on the couch to avoid motion artifact.

13 Gantry Transportation Dolly (with package)


The easy-to-use transportation tools for gantry installation, location adjustment.

14 Couch Transportation Dolly (with package)


The easy-to-use transportation tools for couch installation, location adjustment.
15 Arm-Head Cushion
16 QA Phantom
Image Quality Assurance phantom
17 Continuous CT *
CCT is a scanning mode that allows the physician to perform extended, low-dose scans
while performing a biopsy. You can control the scan by pressing the foot-pedal switch in
the scanning room.
The resulting images display on a remote monitor in the scanning room,
providing near-real-time visual feedback during the biopsy.

18 Dual Screen*
Two monitors to display the console.

19 Dental Analysis*
The Dental Analysis is used to create true-size (life size) film images of the Mandible
and the Maxilla for assisting oral surgeons in planning implantation of prostheses.

20 Nerve System DSA*


Realizes the DSA result by subtracting the bone tissue or some soft tissue with two
series image with different density and get the clear vessel display.

21 Brain Perfusion*
Brain Perfusion is a relatively new technique to diagnosis and evaluation of Brain
Stroke using the rapid qualitative and quantitative evaluation of cerebral perfusion by
generating maps of cerebral of cerebral blood flow (CBF), cerebral blood volume
(CBV), mean transit time (MTT), and Time to Peak (MTT).

22 Body Perfusion*
Give the evaluation for some organs like liver, calculating the perfusion result and
provides reference information for some abnormality as nodules or tumors.

23 Lung Density Evaluation*


Evaluates the density abnormality for lung, give reference information as inflammation,
calcification and reduced lung function degree.

24 Three-Dimensional Lung Nodule Analysis*


Automatic nodule detection and analysis, give the quantitative information about the
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

size, shape, and change over time. Doubling time (DT) helps doctor to identify benign
and malignant nodules.

25 Virtual Colonoscopy*
Designed to assist radiologists and other clinicians in the evaluation of the colon
disease. Automatic segmentation of colon, supplied with interactive edit tools.
High-quality VR view based on segmentation result. Along with the full-angle 3D view
technology, provides a better view for colonic contour and mucosal.
Automatic generation of fly-through view, supplied with user-defined observation
points and path.
• 2D and 3D fly-through visualization of prone and supine images
• Can provide the axial, sagittal, frontal,external view with fly-through view of
the colon images at the same time.
• Supply the real-time 3D Advanced (Panorama) expansion roaming view of
single/double series data, and this full-scale field will not fail any enteral details.
• Supply the VR image mark and corresponding position matching of MPR to the
suspected lesion tissues or ROI
• Supply the spatial 3D distance measurement and the CT value measurement of
the interesting point to the marked ROI.
• Selectively label the suspected lesion tissues according to the relative
information about the anatomy and pathology.

26 Tumor Evaluation*
Evaluate the tumor character, benign or malignant and some information as location,
quantity and size to give some reference for diagnose or surgery treatment. Tracing the
therapy for prognosis.

27 Workstation*
Post-Processing Analysis Tools
2-D SlabViewer
MPR – Multiplanar Reformation
Curved Multiplanar Reformation (CPR)
Maximum Intensity Projection (MIP)
Minimum Intensity Projection (MinIP)
Average Intensity Projection (AIP)
3-D SSD Reconstruction
Volume Rendering
Advanced 3D application package for the optimal display and differentiation of
different organs is through independent control of color, opacity. The doctor can modify
the protocol and preset the threshold for the different display mode. NeuViz 16 Classic
also provides the bone remove function to help display the contrast vessel and soft
tissue. The software supports the transparent 3D reconstruction image to display the
bronchus, colon and skeleton system.
Dental Analysis*

Nerve System DSA*


PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

Brain Perfusion*
.
Body Perfusion*

Lung Density Evaluation*

Three-Dimensional Lung Nodule Analysis*

Virtual Colonoscopy*

Tumor Evaluation*

“*”functions are same as the functions on the console.


28 Isolation Transformer (37KVA)*
Capacity: 37KVA
Input Voltage: 190/200/208/220/230/240/380/400/415/440/460/480V AC
Output Voltage: 380V/220V
Frequency: 50/60 Hz
Dimension: 600×450×850 (mm)
29 Power Stabilizer (50KVA)*

30 Arm Support*
To support patients' arm when transfusion.

31 Cervical Vertebra Cushion*


Position cushion used for Cervical vertebra study.

32 Cradle Handle*
For patient couch movement manually

33 APC UPS*
Capacity: 2.7KVA
Input Voltage: 151-302V
Output Voltage: 220V,230V or 240V
Frequency: 50/60 Hz

34 Coronal Head Holder*


This head holder is used for coronal head scan of patients lying on their backs. The
slope of the holder positions the patient with the neck extended and the head dropped
back.

35 Head Side Cushion-L*


Large size head side Cushion
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

36 Head Side Cushion-M*


Middle size had side cushion

37 Head Side Cushion-S*


Small size head side cushion

38 Couch Extension (Include Cushion)*


For the taller patient, sometime we need inserting another couch extension into the
bedplate end so as to the taller patient can put all body parts on the couch. It is required
for Europe and USA because of regulation needs. Include cushion for Couch Extension.

39 Flat Head Holder (Include Cushion)*

40 Flat Table Top (Include Belt)*


For patient’s lesions position location.

41 Infant Cradle*
For baby’s support and immobilization.

Note:“*” MEANS OPTIONAL CHOICE,


All parameters mentioned above subject to change without notice.
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

II. PRINSIP KERJA ALAT DAN SISTEM KESELAMATAN PERALATAN

A. INSTRUMEN PADA REVIEW


 Menu Umum
 Scroll pada mouse berfungsi untuk melihat image satu per satu pada satu / lebih
series secara cepat. Yaitu dengan melakukan klik pada image yang diinginkan lalu
putar scroll pada mouse ke atas / ke bawah.

 Cursor

Ini merupakan kursor secara umum.

 Pan

Klik icon Pan seperti pada gambar di atas, kemudian drag image yang diinginkan
(klik kiri mouse seraya digerakan), maka image akan bergeser sesuai dengan arah
mouse yg digerakkan.

 Zoom

Klik icon Zoom seperti pada gambar di atas, kemudian drag image yang
diinginkan (klik kiri mouse seraya digerakan) ke atas / ke bawah, maka image
akan membesar / mengecil.

 Enhance

Klik icon Enhance seperti pada gambar di atas, kemudian drag image yang
diinginkan (klik kiri mouse seraya digerakan) ke atas / ke bawah, maka image
akan semakin tajam / semakin halus.

 Contrast

Klik icon Contrast seperti pada gambar di atas, kemudian drag image yang
diinginkan (klik kiri mouse seraya digerakan), maka densitas image akan berubah
sesuai pergerakan mouse.

 Rotate

Klik icon Rotate seperti pada gambar di atas, kemudian drag image 3D (klik kiri
mouse seraya digerakan), maka image akan berputar sesuai dengan pergerakan
mouse. Menu ini digunakan pada menu 3D dan VE.

 Show Information

Klik icon ini, maka akan muncul informasi mengenai parameter pada image
tersebut. Parameter ini pun dapat di print (cetak) dengan klik tombol film.
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

 Expand / Reset Image Window

Icon ini digunakan untuk merubah tampilan image menjadi 1 x 1, jika icon ini di
klik lagi maka tampilan image akan kembali seperti pengaturan sebelumnya.

 Show / Hide Patient Panel

Icon ini digunakan untuk menampilkan / menghilangkan tampilan daftar


microfilm / series yang ada di bawah tampilan utama image, klik icon ini untuk
menampilkan / menghilangkannya.

 Save

Icon ini digunakan untuk menyimpan gambar pada local disk.

 Report

Icon ini digunakan untuk mengirimkan image ke tools Report.

 Print

Icon ini pada menu Review digunakan untuk mengirimkan gambar ke Filming,
sedangkan pada menu Filming, icon ini digunakan untuk melakukan printing /
mencetak image ke film.

 Polygon

Icon ini digunakan untuk membuat area pengukuran dalam bentuk yang
diinginkan, klik icon ini dan klik melingkari area yang akan diukur dengan
bentuk apa saja, kemudian klik dua kali setelah selesai.

 Rectangle

Icon ini digunakan untuk membuat area pengukuran dalam bentuk persegi dengan
berbagai ukuran sesuai yang diinginkan. Klik icon ini kemudian klik dan drag
mouse sesuai dengan area yang diinginkan.
 Ellipse

Icon ini digunakan untuk membuat area pengukuran dalam bentuk lingkaran /
elips dengan berbagai ukuran sesuai yang diinginkan. Klik icon ini kemudian klik
dan drag mouse sesuai dengan area yang diinginkan.

 Length

Icon ini digunakan untuk membuat garis pada image dan melakukan pengukuran
terhadapnya.
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

 Ruler

Icon ini digunakan untuk membuat garis pada image dan melakukan pengukuran
terhadapnya.

 Angle

Icon ini digunakan untuk membuat sudut pada image dan melakukan pengukuran
terhadapnya.

 Arrow

Icon ini digunakan untuk membuat panah dan textbox yang dapat dimasukkan
teks yang diperlukan di dalamnya.

 Text

Icon ini digunakan untuk membuat textbox yang dapat dimasukkan teks yang
diperlukan di dalamnya.

 Pixel Value

Icon ini digunakan untuk melakukan pengukuran HU, klik icon ini kemudian klik
dibagian yg diinginkan pada image, maka akan muncul nilai HU pada titik
tersebut.

 Profile

Icon ini digunakan untuk melakukan pengukuran ct value (HU) pada image yang
dilalui garis ini.

 Histogram

Icon ini digunakan untuk melakukan pengukuran ct value (HU) pada image yang
dilalui area gambar ini.

 Grid

Icon ini digunakan untuk menampilkan gridline / garis kisi-kisi pada image.

Icon modified ini digunakan untuk memilih kondisi windowing gambar, yaitu
kondisi gambar dengan nilai window width/level dengan standard yang sudah
disediakan.

PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

Jika icon ini dipilih maka label / teks judul pada gambar akan tidak ditampilkan,
klik kembali untuk menampilkannya.

Icon ini digunakan untuk mengembalikan kondisi gambar seperti saat sebelum di-
edit (original / kondisi awal).

 2D

 , Setelah gambar diinput dari Home


ke Review, maka pada bagian bawah layar Review, berupa layar kecil akan
muncul dengan berbagai image yang diinput seperti pada gambar, ditampilkan
pada layar utama maka klik dua kali (double click) pada image yang diinginkan
pada layar kecil tersebut.

 Compare

Menu ini digunakan untuk menggandakan tampilan / layar utama sehingga kita
dapat memasukkan series image (kumpulan image) yang berbeda dan saling
membandingkannya.

 , menu ini merupakan bagian dari menu compare dan berfungsi untuk
membandingkan 2 (dua) buah series image (kumpulan image) dengan batas
posisi vertikal.

 , menu ini merupakan bagian dari menu compare dan berfungsi untuk
membandingkan 2 (dua) buah series image (kumpulan image) dengan batas
posisi horizontal.

 menu ini merupakan bagian dari menu compare dan berfungsi untuk
membandingkan 3 (tiga) buah series image (kumpulan image) dengan batas
posisi vertical.

 menu ini merupakan bagian dari menu compare dan berfungsi untuk
membandingkan 4 (empat) buah series image (kumpulan image) dengan batas
posisi vertical dan horizontal.
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

 menu ini merupakan bagian dari menu compare dan berfungsi untuk
menampilkan per 1 (satu) buah image dalam setiap series image (kumpulan
image) yang dibandingkan.

 menu ini merupakan bagian dari menu compare dan berfungsi untuk
menampilkan per 4 (empat) buah image dalam setiap series image (kumpulan
image) yang dibandingkan.

 menu ini merupakan bagian dari menu compare dan berfungsi untuk
menampilkan per 9 (sembilan) buah image dalam setiap series image (kumpulan
image) yang dibandingkan.

 menu ini merupakan bagian dari menu compare dan berfungsi untuk
menampilkan per 16 (enam belas) buah image dalam setiap series image
(kumpulan image) yang dibandingkan.

 , menu ini digunakan untuk menentukan jumlah


tampilan per image dalam setiap series image (kumpulan image) yang akan
dibandingkan, hanya tinggal memasukkan nilai Row(baris) dan Col(kolom)
sesuai yang dibutuhkan, rentang nilai yang dapat dimasukkan adalah 1-10.

 menu ini digunakan untuk membalik atau menukar posisi image bagian
atas dengan bagian bawah.

 menu ini digunakan untuk membalik atau menukar posisi image bagian
kanan dengan bagian kiri.

 menu ini digunakan untuk memutar image dengan sudut sebesar 90o
searah jarum jam.

 menu ini digunakan untuk memutar image dengan sudut sebesar 90o
berlawanan dengan arah jarum jam.

 menu ini digunakan untuk memilih / menunjuk image satuan (hanya 1


image).

 menu ini digunakan untuk memilih / menunjuk series images (1 kumpulan


images). Kemudian akan aktif dua buah menu berikut
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

, kedua menu ini digunakan untuk mengurutkan tampilan image dan


mengembalikan urutannya.

 menu ini digunakan untuk memilih / menunjuk seluruh (all) images yang
ada pada layar.

 MPR (Multi Planar Reconstruction)


 Pada aplikasi ini, layar utama akan dibagi menjadi 4 bagian, dimana layar kiri atas
merupakan tampilan secara axial, layar kanan atas merupakan tampilan coronal,
layar kiri bawah merupakan tampilan dari menu oblique / curved (jika digunakan,
dan layar kanan bawah merupakan tampilan sagital.

 Curved , menu ini digunakan untuk membuat gambar curved dari gambar
axial / coronal / sagital hanya dengan klik dan membentangkan garis di bagian

yang akan diiris dengan menu ini, dan gunakan menu ini jika diperlukan
untuk mengedit image curved yang telah didapat.

 , menu ini digunakan untuk menampilkan image originalnya.

 , menu ini digunakan untuk menampilkan image dengan intensitas yang


rendah.

 , menu ini digunakan untuk menampilkan image dengan intensitas yang


menengah.

 , menu ini digunakan untuk menampilkan image dengan intensitas yang


tinggi, biasanya menu ini digunakan untuk menampilkan High Resolution CT,
dengan menambahkan ketebalan irisan pada kolom thickness seperti berikut

 Batch , menu menu ini digunakan untuk


mendapatkan sejumlah potongan-potongan gambar yang berurut seperti saat
scanning namun secara offline. Dengan meletakan garis horizontal / vertical yang
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

ada pada layar (axial / coronal / sagital) kemudian klik from , kemudian tarik

garis hingga batas yang dikehendaki dan klik to , kemudian klik play
maka image akan diiris dari batas awal hingga batas akhir.

 All , menu ini untuk menampilkan batch secara keseluruhan tanpa batas
yang ditentukan seperti menu from dan to.

 Clear , menu ini digunakan untuk menghapus batas batch yang ditentukan
melalui menu from dan to.

 , menu ini digunakan untuk menentukan spasi pemotongan dari


image pada batch.

 , menu ini digunakan untuk menentukan jumlah image yang akan


dihasilkan pada batch.

 Save , menu ini digunakan untuk menyimpan batch yang telah dibuat ke
dalam computer.

 Report , menu ini digunakan untuk mengirim batch yang telah dibuat ke
report.

 Film , menu ini digunakan untuk mengirim batch yang telah dibuat ke
filming.

 , menu ini berfungsi untuk menampilkan batch yang telah dibuat dari batas
awal ke batas akhir.

 , menu ini berfungsi untuk menghentikan tampilan batch (pause).

 menu ini berfungsi untuk menampilkan batch yang telah dibuat dari batas
akhir ke batas awal.
 , menu ini digunakan untuk mengatur kecepatan dan
resolusi tampilan batch yang telah dibuat, semakin tinggi resolusi yang digunakan
(gambar halus) maka semakin lambat perjalanan batch.

 3D

Bentuk tampilan 3D yaitu :


PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

 menu ini berfungsi untuk menampilkan volume image secara axial


(tampilan dari atas).

 menu ini berfungsi untuk menampilkan volume image secara coronal


(tampilan dari depan).

 menu ini berfungsi untuk menampilkan volume image secara sagital


(tampilan dari samping).

 menu ini berfungsi untuk menampilkan volume image.

 menu ini berfungsi untuk menampilkan shaded surface display, dimana


menampilkan lapisan permukaan.

 menu-menu ini digunakan untuk menampilkan image 3D


dengan intensitas yang berbeda (minimum, rata-rata, atau maksimum).

 Cut , menu ini berfungsi untuk memotong bagian dari volume image yang
tidak dibutuhkan.

 , menu ini digunakan untuk undo / mengembalikan image pada keadaan


sebelumnya.

 menu ini digunakan untuk menghilangkan tulang (bone removal).

 menu ini digunakan untuk undo dari menu bone removal.


 , menu ini untuk menampilkan / menghilangkan garis
horizontal, vertikal serta kotak batas image pada layar.
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

 , menu ini untuk menampilkan / menghilangkan kotak batas image


pada layar.
 , menu ini untuk menunjukkan letak / lokasi titik point yang
ditunjuk.
 : Berfungsi untuk menampilkan atau tidak daftar protocol di
sebelah kiri layar. Yang dimaksud daftar protocol adalah sebagai berikut :

Klik , untuk membuat protocol baru yang kemudian akan diatur


kondisinya pada tampilan berikut :

Klik , untuk mengedit protocol yang sudah ada. Dan klik , untuk
menghapus protocol yang sudah dibuat.
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

 VIRTUAL ENDOSCOPY
 , menu ini sebagai standar menu pada virtual endoscopy.

 , menu ini berfungsi untuk menentukan lokasi awal endoscopy.


 , menu ini berfungsi untuk mengatur arah endoscopy yang telah ditentukan.
 , menu ini berfungsi untuk menentukan titik-titik perjalanan endoscopy.
 , menu ini digunakan untuk konfirmasi bahwa endoscopy akan berlangsung
sesuai dengan titik-titik yang telah ditentukan.
 , menu ini digunakan untuk menentukan arah pengambilan kamera.

 , menu ini digunakan untuk menentukan resolusi dan kecepatan


tampilan endoscopy dimana semakin tinggi resolusi maka semakin lambat
tampilan endoscopy.
 , menu ini berfungsi untuk menjalankan maju endoscopy dari awal secara
otomatis.
 , menu ini berfungsi untuk menjalankan maju endoscopy dari awal secara
manual step by step.
 , menu ini berfungsi untuk menjalankan mundur endoscopy secara otomatis.
 , menu ini berfungsi untuk menjalankan mundur endoscopy dari secara manual
step by step.
 , menu ini berfungsi untuk menghentikan endoscopy ataupun rekaman
endoscopy.
 , menu ini berfungsi untuk menyimpan navigasi endoscopy.
 , menu ini berfungsi untuk menyimpan endoscopy dengan format video.
 , menu ini berfungsi untuk menyimpan endoscopy ke dalam USB / CD.

B. FILMING
 , menu ini digunakan untuk menampilkan 1x1 gambar pada selembar film
yang akan dicetak.
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

 , menu ini digunakan untuk menampilkan 4x5 gambar pada selembar film
yang akan dicetak.

 , menu ini digunakan untuk menampilkan 5x7 gambar pada selembar film
yang akan dicetak.

 , menu ini digunakan untuk menentukan jumlah tampilan gambar pada


selembar film yang akan dicetak.

 , menu ini digunakan untuk menampilkan sub window / pada sebuah bagian
dari tampilan dapat dibagi lagi untuk menampilkan beberapa images, dengan format

tampilan , , , , dan menu untuk mengembalikan


sub window menjadi 1 layar kembali.

 , menu ini digunakan untuk menampilkan lokasi irisan scanning dari awal
hingga akhir.

 , menu ini digunakan untuk menampilkan lokasi irisan scanning awal dan
akhir.

 , menu ini digunakan untuk menghilangkan tampilan irisan.

 , menu ini digunakan untuk membalik atau menukar posisi image bagian atas
dengan bagian bawah.

 , menu ini digunakan untuk membalik atau menukar posisi image bagian
kanan dengan bagian kiri.

 , menu ini digunakan untuk memutar image sebesar 90o searah jarum jam.

 , menu ini digunakan untuk mencetak dengan printer film x-ray.

 , menu ini digunakan untuk mengatur format pencetakan.

 , menu ini digunakan untuk mencetak dengan film kertas.

 , menu ini berfungsi untuk menghapus image.


PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

SISTEM KESELAMATAN PERALATAN

Radiology Safety
Sinar X-Ray dan sinar gamma berbahaya bagi operator dan orang lain di sekitar kecuali jika
prosedur pemaparan aman yang benar benar-benar diperhatikan.

Radiasi pengion yang berguna dan tersebar dapat menghasilkan luka tubuh yang serius atau
fatal pada pasien dan orang di daerah sekitarnya jika digunakan oleh operator tidak terampil.
Tindakan pencegahan yang memadai harus selalu dilakukan untuk menghindari paparan sinar
yang berguna, serta untuk menghilangkan kebocoran radiasi dari tempat sumber atau radiasi
yang tersebar akibat pelepasan radiasi melalui materi.

Mereka yang diberi wewenang untuk mengoperasikan, berpartisipasi atau mengawasi operasi
peralatan harus benar-benar menguasai dan sepenuhnya mematuhi faktor pemaparan aman
yang ada saat ini dan prosedur yang dijelaskan dalam publikasi, seperti "Sistem X-Ray
Diagnostik dan komponen utamanya", bagian Dari sub-bab J dari Kode 21 Peraturan
Peraturan Federal, dan Dewan Nasional Perlindungan Radiasi (NCRP) No 102,
"Perlindungan sinar-X Medis dan sinar gamma untuk energi hingga 10 desain dan
penggunaan peralatan MEV", seperti yang direvisi atau diganti di kemudian nanti.

Selain itu, para operator sangat mendesak untuk memperkenalkan rekomendasi Komisi
Internasional Perlindungan Radiologi Internasional saat ini, dan di Amerika Serikat, dengan
Dewan Nasional untuk Perlindungan Radiologi AS.

ICRP, Pergamon Press, Oxford, New York, Beijing, Frankfurt, Sao Paul, Sydney,Tokyo,
Toronto

NCRP, Suite 800, 7910 Woodmont Avenue, Bethesda, Maryland 20814, Amerika Serikat

Mereka yang bertanggung jawab atas perencanaan instalasi peralatan X-Ray dan gamma ray
harus benar-benar menguasai dan sepenuhnya mematuhi NCRP No. 49, "Perancangan dan
evaluasi perisai struktural untuk Medical of X-Rays dan sinar gamma energi sampai 10
MEV", Seperti direvisi dan diganti di masa depan.

Kegagalan untuk mengamati peringatan ini dapat menyebabkan cedera fisik yang serius atau
fatal pada operator atau di area tersebut.
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

III. PROSEDUR PENGOPERASIAN ALAT

STANDAR PENGOPERASIAN
CT-SCAN NEUVIZ 16 CLASSIC, NEUSOFT

MENGHIDUPKAN (MENGAKTIFKAN)
No. Langkah-langkah Gambar
1. Nyalakan panel dengan memutar kenop hitam pada panel listrik
ke kanan / searah jarum jam (lampu hijau menyala).

2. Nyalakan gantry, Atur switch gantry pada posisi On, merujuk ke

gambar .

3. Nyalakan UPS pada bagian kontrol konsol dengan menekan


tombol ”test”.

4. Nyalakan CPU komputer konsol dengan menekan tombol On.

5. Nyalakan printer (Tekan tombol power [on/off] pada printer).


PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

6. Nyalakan injektor (bila perlu)


Atur switch on/off pada (bagian belakang) power box injektor
ke posisi On.
7. Alat didiamkan sekitar 30 menit untuk mendapatkan kestabilan.
8. Alat siap dioperasikan.

\
STANDAR PENGOPERASIAN
CT-SCAN NEUVIZ 16 CLASSIC, NEUSOFT

MEMATIKAN (MENON-AKTIFKAN)
No. Langkah-langkah Gambar
1. Matikan injector dengan mengatur switch pada (belakang)
power box injektor ke posisi off.

2. Matikan printer dengan menekan tombol power on/off pada


printer ke posisi off.

3. Pastikan suhu tube (tube storage) yang ditampilkan di bagian


kanan bawah monitor sudah bernilai kurang dari 20%.
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

4. Keluar dari software dengan klik Home lalu klik Log off di
bagian kanan bawah, kemudian matikan komputer console
(klik Start kemudian klik Power Off / Shut Down).

5. Matikan UPS dengan menekan tombol off.

6. Matikan Gantry (pindahkan posisi switch gantry dari posisi

On ke posisi Off, merujuk ke gambar .

7. Matian panel dengan memutar kenop hitam kembali ke kiri


posisi awal (lampu hijau mati).

8. Alat sudah dalam keadaan off.

STANDAR PENGOPERASIAN
CT-SCAN NEUVIZ 16 CLASSIC, NEUSOFT

PENGOPERASIAN

1. Klik Start – Neuviz 16 Host ,

2. Isi User Name dan Password, klik OK


PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

,
3. Lakukan Warm-up tube jika diperlukan, dengan prosedur berikut :
 Klik menu Service

 Klik Tube Warm Up ,

 Klik Start ,
 Tunggu beberapa menit, jika warm-up selesai (tulisan warm-up finished /

complete ditampilkan) maka klik Exit,

4. Klik menu ”Scanning>>” pada pojok kanan atas, seperti pada gambar berikut:

a) Klik Start Study

b) Isi data pasien, terutama yang bertanda bintang berwarna merah (wajib diisi) ,
kemudian klik Exam Protocol Groups,
c) Pilih protokol yang akan digunakan dengan memilih nama protokol yang akan
digunakan, kemudian klik jenis pemeriksaan yang akan dilakukan, selanjutnya
klik Exam Protocols.
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

d) Lakukan pengaturan yang diperlukan.


e) Klik Go,
PT. HOSPI MEDIK INDONESIA
Medical Equipment & Supplier

f) Tekan dan tahan tombol Enable (hingga tampilan perintah untuk


menekan menghilang) pada CT-Box (jika monitor komputer menampilkan perintahnya),

g) Tekan tombol Scan pada CT-Box jika monitor komputer telah


menampilkan perintahnya.
h) Jika scan surview / topogram telah selesai, maka lakukan pengaturan area
scanning, kemudian klik GO, dan lakukan seperti langkah f dan g.
i) Jika telah selesai secara keseluruhan maka klik End Study,
j) Lalu muncul prompt dengan kalimat, ”End Study?”, maka klik YES.