Medical Gases Systems

Presented By: Mohamed Salah

THE FIFTH LECTURE.


In an effort to continuously improve our products, 2005628

mAIR MEDICAL
the right is reserved to change the specification of the

AIR SYSTEMS
03/05/12
items described herein at any time. Please contact us Page 1 of 9
for further information and up to date specifications. Issue 3

mAIR Medical Air, cAIR Combined Air & sAIR Surgical Air Systems
– EN ISO 7396-1/HTM 02-01 and HTM2022 EurPh
400V 50Hz, 4 Bar, 7 Bar & 11 Bar Outlet
SPECIFICATION
Air Plant System Compressors
The Medical Air system shall conform to EN ISO 7396-1 and NHS Compressors shall be Atlas Copco GA-MED oil injected rotary screw
Health Technical Memorandum HTM02-01. Medical quality air to the compressors suitable for both continuous and frequent start/stop
European Pharmacopoeia monograph shall be delivered at pressures of operation at a nominal outlet pressure of 750 kPa (7.5 bar), 950 kPa (9.5
400 kPa (4 bar), 700kPa (7 bar) or 1100 kPa (11 bar) gauge for supply of bar) or 13000 kPa (13 bar) gauge. Compressors shall be supplied with
the hospital medical or surgical air systems. The entire system shall be a block and fin style after cooler with a dedicated quiet running fan to
‘duplexed’ such that any single functional component failure will not maximise cooling and efficiency. A multistage oil separator capable of
affect the integrity of the medical compressed air supply. achieving 2ppm oil carry over shall be fitted to minimise contamination
and maintenance. EFF1 (CEMEP) rated TEFC, IP55 class F electric motors
Surgical air systems shall have a ‘duplexed’ dryer system and a simplex shall be used and incorporate maintenance-free greased for life bearings.
compressor. Additional compressors shall be available to fully ‘duplex’ Motors with lower efficiency ratings are not acceptable.
the surgical air system, such that any single functional component
failure will not affect the integrity of the air supply. Variable Speed Drive - Optional
Variable Speed Drive (VSD) shall be available on compressor(s). By
Sources Of Supply - HTM02-01/EN ISO 7396-1 including an AC-DC converter, along with associated control hardware
Triplex or quadruplex compressor configurations will produce the and software it will enable the compressor to match it’s running speed
primary supply with two compressors in standby. Each compressor on a 1:1 ratio with the flow demand required by hospital. By using
will be capable of supplying the specified volumetric flow for duplex such technology energy of savings of up to 35% shall be achievable as
and triplex plant, and half flow for quadruplex. For duplex plant the well as prolonged machine life. Start currents will be reduced and the
secondary supply shall be from an automatic manifold capable of motor will run from 0-100% speed increasing or decreasing it’s speed as
supplying at average demand for 4 hours. For triplex and quadruplex necessary to continuously match the required demand. The compressor
plant the standby compressors will form the secondary supply. shall operate from 400-1300kPa (4-13 bar) gauge.

Sources Of Supply - HTM2022 EurPh ‘Full VSD’ air plant shall incorporate VSD controllers on all compressors,
Duplex, triplex or quadruplex compressor configurations will produce cycling the lead compressor to ensure even wear as per HTM02-01
the primary supply with one compressor on standby. A secondary requirements.
supply shall be from an automatic manifold capable of supplying at
average demand for 4 hours. ‘Mixed VSD’ air plant shall incorporate a VSD controller on the lead
compressor and the remaining compressors will be fixed speed.
Control System
The central control system shall provide an intelligent human machine Dryer/Filter/Regulator System
interface incorporating on board flash memory and real-time clock The duplexed filter and dryer module shall incorporate high efficiency
for recording operational parameters in the in built event log. The water separators, oil filters, heatless regenerative desiccant dryer,
central control system shall operate at low voltage and include BMS dust/activated carbon filters, hopcolite filters and bacterial filters with
connection for plant fault, plant emergency, reserve fault and pressure autoclavable element. Electrical contacts shall be installed on the filters
fault. Visualisation of plant inputs, outputs and status through a web to provide warning alarms on the dryer controller in the event of high
browser, using a simple Ethernet connection shall be available. The pressure drop (ie blockage) and shall also include connections for BMS.
central control unit shall incorporate a user friendly 5.7” high-definition Contaminants in the delivered air downstream of the bacterial filters
colour display with clear pictograms and LED indicators, providing easy shall be maintained at levels below those shown in the following table:
access to system operational information.
Contaminant Threshold
A mechanical back-up facility shall ensure continued operation in the H2O 67 ppm v/v
event of a control system malfunction. The control system shall normally
employ automatic rotation of the lead compressor to maximise life and Free from visible particulates in a 75 litre
Dry particulates
ensure even wear. sample
Oil (droplet or mist) 0.1 mg/m³
Optional Control Equipment
CO 5 ppm v/v
An advanced monitoring system shall be available to give immediate
access to valuable information such as system status, trends, historical CO2 500 ppm v/v
data and system performance. Data collected from all pumps shall be SO2 1 ppm v/v
made available in real-time visualisation pages and shall be accessed
through the hospital’s LAN, such that total data security is assured. NO 2 ppm v/v
NO2 2 ppm v/v
The Airconnect™ monitoring system shall also include :-

• Logging and trending for an accurate performance status of the system.

• Desktop event notification to avoid constant status checking.

• E-mail and SMS event notification for additional convenience.

BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
In an effort to continuously improve our products, 2005628
the right is reserved to change the specification of the 03/05/12
items described herein at any time. Please contact us Page 2 of 9
for further information and up to date specifications. Issue 3

Dryer Purge Control - Optional


The dryer control system shall incorporate a Purge Saver Energy Oil-free Compressors - Optional
Management system that freezes the regeneration of the desiccant once Medical air plant, surgical air plant 7 bar and combined medical and
adequate dew point is reached in the inactive tower. Only when the surgical air plant 7 bar can be supplied with oil-free compressors.
dewpoint level in the active tower deteriorates to an unacceptable level Incorporating either Atlas Copco ZT Claw compressors or SF Scroll
will the intelligent controller switch towers. This shall be achieved by compressors - all available upon special request.
including an additional dew point sensor and associated software in the
dryer controller to effectively manage the system as well as providing on CE Marking
screen measurements of purge savings. The standard range of BeaconMedæs Medical Air plant systems are ‘CE’
marked under the Medical Devices Directive 93/42/EEC with approval
Dew Point Monitoring from notified body no. 0088 (Lloyd’s Register Quality Assurance). Under
The dryer shall incorporate a ceramic dew point hygrometer with an this directive, the specified products are classified as Class IIa Medical
accuracy of ±10C in the range -20 to -800C atmospheric dew point and Devices.
4-20mA analogue output. Aluminium oxide or palladium wire sensors
are not acceptable. An alarm condition shall trigger on the dryer control Typical Layout
panel if the dew point exceeds a -460C atmospheric set point. The plant
control unit shall incorporate a multifunctional LCD displaying, amongst
other things, the dew point of the delivered air to enable monitoring of Receivers
Compressors
the air quality by the hospitals estates department. Volt free contacts
shall be included to enable the dew point alarm signal to be connected
to a central medical gas alarm system and/or building management
system (BMS). To enable periodic calibration of the dew point sensor
element, the hygrometer shall be remotely connected downstream of
the dryer via a micro-bore tube. It is not acceptable to install the sensor
directly into the medical air supply pipeline.

Receiver Assembly
Air receivers shall comply with DIRECTIVE 97/23/EC, supplied with
relevant test certificates. Each air receiver shall be hot dip galvanised
Dryer
inside and out and fitted with a zero loss electronic drain valve. Float
type drain valves are not acceptable. The receiver assembly shall be
fitted with a pressure safety valve capable of passing the maximum flow
Note:-
output of the compressor at 10% receiver overpressure. The receiver
Inter connecting pipework (blue illustration) between components to
shall be further protected by a safety pressure relief valve and include a
be made on site and provided by the installer. Controller CAN cables
pressure gauge.
are provided as a 10m assembly with each compressor which can be
shortened on site if required.
Optional Items
There shall be the followings options available for enhanced operation
of the air plant system:-
Combined Air Plant Sizing Guide
• Phase sequence relays that prevent unintentional reverse operation of In HTM02-01, the relative size of receiver capacity and compressor
the compressors capacity on surgical air or combined medical/surgical air systems
changes according to the design flow rate. In order to correctly calculate
• OCS electronic water/oil drains for the air plant system the receiver capacity and compressor capacity, both the medical and
surgical design flow-rates (DF’s) are required. It should be noted that
• EWD zero loss electronic water drains for the dMED dryer including for all combined air systems, an additional duplex regulating station
(ordered separately) is needed to supply the medical air pipeline.
secure mounting to the dryer base

• VSD compressors and optional harmonics filter for installation close to Design Flow Value ‘A’
highly sensitive equipment (l/min) FAD (l)
Surgical Air <500 0.33 x DF
• Synthetic oil for increased compressor life Compressors
500-3500 0.66 x DF
• QDT saturation indicators to give clear visual indication of oil carry >3500 0.5 x DF
over to the activated carbon tower
Table 1: Surgical Air Plant Flow Rate Multiplier Value ‘A’.
• Tropical thermostatic sensors for countries with high humidity
Design Flow Value ‘B’ Receiver
• Heavy duty inlet filters for compressors installed in areas of highly (l/min) water capacity (l)
concentrated dust levels
<500 1 x 200% x DF
Surgical Air
• Painted hot dipped galvanised vessels Receivers 500-2000 2 x 66.6% x DF
2001-3500 2 x 50% x DF
• CO and CO2 monitors including full integration into the ES-Med
central controller giving alarm warnings when unacceptable CO and >3500 3 x 33.3% x DF
CO2 levels are present Table 2: Surgical Air Receiver Multiplier Value ‘B’.

BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
In an effort to continuously improve our products, 2005628
the right is reserved to change the specification of the 03/05/12
items described herein at any time. Please contact us Page 3 of 9
for further information and up to date specifications. Issue3

Example 1 - Small Day Treatment Centre (Upgrade) Example 2 - Large District Hospital

Flow Rate and Dryer Sizing Flow Rate and Dryer Sizing
Medical Air DF = 555 l/min (FAD) (4 Bar) Medical Air DF = 4920 l/min (FAD) (4 Bar)
Surgical Air DF = 1138 l/min (FAD) (7 Bar) Surgical Air DF =2888 l/min (FAD) (11 Bar)
Combined/total DF = 1693 l/min (FAD) (11 Bar high pressure system) Combined/total DF = 7808 l/min (FAD) (11 Bar high pressure system)

A dryer greater than 1693 l/min outlet flow should be selected A dryer greater than 7808 l/min should be selected
(outlet flow is the inlet flow minus purge losses) (outlet flow is the inlet flow minus purge losses)
= dMED25 inlet flow 2025 l/min, outlet flow 1755 l/min = dMED145 inlet flow 11745 l/min, outlet flow 10092 l/min

Flow Rate and Compressor Sizing Flow Rate and Compressor Sizing
From Table 1 surgical air DF is between 500-3500 l/min, so the From Table 1 surgical air DF is between 500-3500 l/min, so the
multiplying factor ‘A’ = 0.66 multiplying factor ‘A’ = 0.66

Compressor flow rate = Med. DF + (Surg. DF x A) Plant flow rate = Med. DF + (Surg. DF x Value ‘A’)
= 555 + (1138 x 0.66) = 4920 + (2888 x 0.66)
= 555 + 751 = 4920 + 1907
= 1306 l/min = 6827 l/min

We also need to add the purge losses to the compressor output. For We also need to add the purge losses to the compressor output. For
additional purge consumption use:- additional purge consumption use:-

dMED inlet - dMED outlet = purge losses lpm dMED inlet - dMED outlet = purge losses lpm
= 2025 - 1755 = 270 l/min = 11745 - 10092 = 1653 l/min

Compressors should be selected with a flow rate greater than Compressors should be selected with a flow rate greater than
1306 l/min + 270 l/min = 1576 l/min 6827 l/min + 1653 l/min = 8480 l/min

Receiver Sizing Receiver Sizing


From Table 2 surgical air DF is between 500-2000 l/min, so the From Table 2 surgical air DF is between 2001-3500 l/min, so the
multiplying factor ‘B’ = 2 x 2/3 multiplying factor ‘B’ = 2 x 1/2

Capacity = (Med. DF x 0.5) + (Surg. DF x B) Capacity = (Med. DF x 0.5) + (Surg. DF x B)


= (555 x 0.5) + (1138 x 2 x 2/3) = (4920 x 0.5) + (2888 x 2 x 1/2)
= 278 + 1518 = 2460 + 2888
= 1796 litres = 5348 litres

A combination of receivers with a minimum number of 2 should be A combination of receivers with a minimum number of 2 should be
selected selected

Selected receiver capacity = 2000 litres (2 x 1000 litre) Selected receiver capacity = 6000 litres (3 x 2000 litre)

Plant System Selection Plant System Selection


Selected plant part number = cAIR-1755-TGA11 Selected plant part number = n/a

If no standard model is available for selection from the standard range a If no standard model is available for selection from the standard range a
bespoke configuration of dryer, compressors and receivers are available bespoke configuration of dryer, compressors and receivers are available
and can be quoted by our sales and sales support teams. and can be quoted by our sales and sales support teams.

Receiver Selection Table

Receiver Capacity (litres) 250 500 1000 1500 2000


Maximum working pressure (bar) 16 16 16 16 16
Individual Receiver Dimensions
500/1950 600/2350 800/2550 1000/2525 1150/2605
(diameter, height, mm)
Receiver Weight (kg) 80 160 304 445 557
Receiver pipe size (mm) 22 22 28 28 35
Receiver Part Number 8101 0211 97 8101 0212 62 8101 0213 12 8102 0213 38 8101 0213 53
Accessory Kit Part Number * Specific to plant Specific to plant Specific to plant Specific to plant Specific to plant
Drawing Number tbc tbc tbc tbc tbc

* Accessory kit for medical air receiver complete with plant data plate, test certificate, pressure safety valve, zero-loss electronic drain
valve (with isolation and bypass valve), pressure gauge (with isolation valve), safety pressure relief valve and inlet and copper outlet
union connection pipes (each with isolation valve).

BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
In an effort to continuously improve our products, 2005628
the right is reserved to change the specification of the 03/05/12
items described herein at any time. Please contact us Page 4 of 9
for further information and up to date specifications. Issue 3

Compressor Selection Table - Fixed Speed

Model Name GA-MED 5 GA-MED 7 GA-MED 11 GA-MED 15 GA-MED 18 GA-MED 22

Output flow (litres/minute) *


900 1308 1842 2580 3150 3612
7.5 bar variant

Output flow (litres/minute) *


702 1032 1560 2178 2610 3102
10 bar variant

Output flow (litres/minute) *


504 852 1320 1806 2232 2700
13 bar variant

1140 x 700 x 1140 x 700 x 1140 x 700 x 1285 x 680 x 1285 x 680 x 1285 x 680 x
Footprint L x W x H (mm)
1240 1240 1240 922 922 922

Compressor weight (kg) 270 284 310 375 395 410

Service connection (mm) 22 22 22 22 22 22

Noise level/pump (dB[A]) 60 61 62 72 73 74

Maximum ambient temperature (0C) 46 46 46 46 46 46

Supply voltage (v) 400 400 400 400 400 400

Supply frequency (Hz) 50 50 50 50 50 50

Nominal motor rating (kW) 5 7 11 15 18 22

Full load current per compressor (A) 17 22 32 36 43 54

Starting current (A) 98.6 131 188 252 302 383

Cooling air flow per compressor


0.8 0.8 1 1.1 1.15 1.2
(m3/s)

Part Number - 7.5 bar 8153 0344 86 8153 0344 94 8153 0345 02 8153 0350 04 8153 0350 38 8153 0350 61

Part Number - 10 bar 8152 0341 89 8153 0342 05 8153 0342 21 8153 0350 12 8153 0350 46 8153 0350 79

Part Number - 13 bar 8153 0341 97 8153 0342 13 8153 0342 39 8153 0350 20 8153 0350 53 8153 0350 87

Drawing Number 9828 4969 26 9828 4969 26 9828 4969 26 9820 6060 17 9820 6060 17 9820 6060 17

* Output flow stated at reference conditions

Compressor Selection Table - Variable Speed Drive

Model Name GA-MED VSD 5 GA-MED VSD 7 GA-MED VSD 11 GA-MED VSD 15

Output flow (litres/minute)


900/ 792 / 600 1218 / 1008 / 828 1842/ 1446 / 1242 2226 / 1854 / 1488
7.5 bar , 10 bar and 13 bar variants *

Footprint L x W x H (mm) 1345 x 700 x 1240 1345 x 700 x 1240 1345 x 700 x 1240 1345 x 700 x 1240

Compressor weight (kg) 290 295 308 315

Service connection (mm) 22 22 22 22

Noise level/pump (dB[A]) 62 64 66 69

Maximum ambient temperature (0C) 46 46 46 46

Supply voltage (v) 400 400 400 400

Supply frequency (Hz) 50 50 50 50

Nominal motor rating (kW) 5 7 11 15

Full load current per compressor (A) 20.9 24.3 36.3 45.3

Cooling air flow per compressor (m3/s) 0.8 0.8 1 1

Part Number 8153 0343 61 8153 0343 79 8153 0343 87 8153 0343 95

Drawing Number 9828 4969 27 9828 4969 27 9828 4969 27 9828 4969 27

* Variable speed drive compressor operate from 4-13 bar at 0-100% speed - start currents nominal for VSD

BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
In an effort to continuously improve our products, 2005628
the right is reserved to change the specification of the 03/05/12
items described herein at any time. Please contact us Page 5 of 9
for further information and up to date specifications. Issue 3

Dryer Selection Table

Model Name dMED7 dMED13 dMED25 dMED35 dMED50 dMED70 dMED80 dMED100 dMED145

Inlet flow at 7.5 bar (litres/minute) 420 780 1500 2100 3000 4200 4800 6000 8700

Output flow (litres/minute)


340 632 1230 1722 2430 3444 3936 4860 7047
at 4 bar line pressure *

Inlet flow at 10 bar (litres/minute) 504 936 1800 2520 3600 5040 5760 7200 10440

Output flow (litres/minute)


424 788 1530 2142 3030 4284 4896 6060 8787
at 7 bar line pressure *

Inlet flow at 13 bar (litres/minute) 567 1053 2025 2835 4050 5670 6480 8100 11745

Output flow (litres/minute)


487 905 1755 2457 3480 4916 5616 6960 10092
at 10 bar line pressure *

1920 x 1920 x 1920 x 1920 x 1920 x 2160 x 2160 x 2160 x 2160 x


Footprint L x W x H (mm) 600 x 600 x 600 x 600 x 600 x 1250 x 1250 x 1250 x 1250 x
1590 1590 1590 1590 1870 1850 1850 1850 1870

Dryer weight (kg) 255 275 325 345 415 615 700 705 895

Inlet and outlet connections (mm) 15 15 15 28 28 28 28 28 28

Supply voltage (v) 230 230 230 230 230 230 230 230 230

Supply frequency (Hz) 50 50 50 50 50 50 50 50 50

Central control supply 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5
- single phase (mm2/Amps) (2) (2) (2) (2) (2) (2) (2) (2) (2)
8102 1431 72
0000 0203 58

8102 1431 80
0000 0203 58

8102 1431 98
0000 0203 71

8102 1432 06
0000 0203 71

8102 1432 14
0000 0203 72

8102 1432 22
0000 0203 73

8102 1432 30
0000 0203 73

8102 1432 48
0000 0203 74

8102 1432 55
0000 0203 75
Part Number - dryer at 4 bar outlet +
QDT hopcolite filter for EurPh
8102 1434 04
8102 1433 54
0000 0203 58

8102 1433 62
0000 0203 58

8102 1433 70
0000 0203 71

8102 1433 88
0000 0203 71

8102 1433 96
0000 0203 72

0000 0203 73

8102 1434 12
0000 0203 73

8102 1434 20
0000 0203 74

8102 1434 38
0000 0203 75
Part Number - dryer at 7 bar outlet +
QDT hopcolite filter for EurPh
8102 1434 46
0000 0203 58

8102 1434 53
0000 0203 58

8102 1434 61
0000 0203 71

8102 1434 79
0000 0203 71

8102 1434 87
0000 0203 72

8102 1434 95
0000 0203 73

8102 1435 03
0000 0203 73

8102 1435 11
0000 0203 74

8102 1435 29
0000 0203 75
Part Number - dryer at 10 bar outlet +
QDT hopcolite filter for EurPh
9827 7000 08

9827 7000 09

9827 7000 10

9827 7000 11

9827 7000 12

9827 7000 13

9827 7000 14

9827 7000 15

9827 7000 16

Drawing Number

* Output flow stated includes calculated purge lost during the regenration process of between 15-19% depending on model and inlet pressure.

Notes on plant

• Design flow in terms of free air delivered after losses at working pressure with the reserve compressor(s) on standby. Tolerance ±5%.
• Component dimensions supplied do not include maintenance access space, and are provided to allow customer to arrange plant
components within plant room. Complete installation drawings are available on request. Quote the drawing number required.
• Duplex systems must be installed with a manifold as the third source of supply for HTM02-01 complaince.
• Mean sound level in accordance with ISO 2151.
• Electrical details are provided for guidance only. Site conditions may impose a larger cable size. For exact cable sizing, and fuse / MCB
ratings, consult a qualified electrical engineer.

BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
In an effort to continuously improve our products, 2005628
the right is reserved to change the specification of the 03/05/12
items described herein at any time. Please contact us Page 6 of 9
for further information and up to date specifications. Issue 3

Standard Models

HTM02-01 Medical Air 4 bar

mAIR-3445-QGF4

mAIR-3935-QGF4

mAIR-4860-QGF4

mAIR-5570-QGF4

mAIR-7045-PGF4
mAIR-6000-PGF4
mAIR-1000-TGF4

mAIR-1230-TGF4

mAIR-1465-TGF4

mAIR-1720-TGF4

mAIR-2000-TGF4

mAIR-2410-TGF4

mAIR-2855-TGF4
mAIR-340-TGF4

mAIR-500-TGF4

mAIR-630-TGF4
Model
Number

Design flow
340 500 630 1000 1230 1465 1720 2000 *2430 2855 3445 3935 4860 5570 6000 7045
(litres/minute)
Number of
3 3 3 3 3 3 3 3 3 3 4 4 4 4 5 5
compressors
GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA-
Type of
MED MED MED MED MED MED MED MED MED MED MED MED MED MED MED MED
compressor
5 5 5 7 11 11 15 15 18 22 15 15 18 22 15 18
dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED
Type of dryer
7 13 13 25 25 35 35 50 50 70 70 80 100 145 145 145
Number of receivers 1 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2
Receiver
volume (l) 250 250 250 250 500 500 500 500 1000 1000 1000 1000 1500 1500 1500 2000
(each)
Drawing Number

9827 8413 00

8153 0353 19
9827 8413 00

8153 0353 27
9827 8414 00

8153 0353 35
9827 8414 00

8153 0353 43
9827 8415 00

8153 0353 50
9827 8415 00

8153 0353 68
9827 8420 00

8153 0353 76
9827 8420 00

8153 0353 84
9827 8421 00

8153 0353 92
9827 8421 00

8153 0354 00
9827 8421 00

8153 0354 18
9827 8421 00

8153 0354 26
9827 8422 00

8153 0354 34
9827 8422 00

8153 0354 42
9827 8422 00

8153 0354 59
9827 8424 00
8153 0353 01
Part Number

HTM02-01 Combined Medical & Surgical Air 7 bar


cAIR-3030-QGF7

cAIR-4285-QGF7

cAIR-4895-QGF7

cAIR-6000-QGF7

cAIR-8000-PGF7

cAIR-8790-PGF7
cAIR-1000-TGF7

cAIR-1530-TGF7

cAIR-2140-TGF7

cAIR-2450-TGF7
cAIR-420-TGF7

cAIR-790-TGF7

Model
Number

Design flow (litres/minute) 420 790 998 1530 2140 2450 3030 4285 4895 6000 8000 8790
Required compressor output * 282 524 664 1269 1776 2035 2515 3556 4064 4980 6000 6593
Required receiver volume * 530 985 1248 1402 1962 2245 2777 3213 3672 4500 6000 6593
Number of compressors 3 3 3 3 3 3 4 4 4 4 5 5
GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA-
Type of compressor MED MED MED MED MED MED MED MED MED MED MED MED
5 5 7 11 15 18 11 15 18 22 18 18
dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED
Type of dryer
7 13 25 25 35 50 50 70 80 100 145 145
Number of receivers 1 2 1 2 2 2 2 2 2 3 3 4
Receiver volume (l) (each) 1000 500 1500 1000 1000 1500 1500 2000 2000 1500 2000 2000
Drawing Number

8153 0354 67
9827 8416 00

8153 0354 75
9827 8415 00

8153 0354 83
9827 8418 00

8153 0354 91
9827 8417 00

8153 0355 09
9827 8421 00

8153 0355 17
9827 8422 00

8153 0355 25
9827 8419 00

8153 0355 33
9827 8424 00

8153 0355 41
9827 8424 00

8153 0355 58
9827 8423 00

8153 0355 66
9827 8425 00

8153 0355 74
9827 8426 00
Part Number

* Flow output based on a 50/50 split of medical and surgical air design flow (D.F)
For examples on how to calculate the required design flow, compressor flow and receiver volume refer to earlier page,
in line with HTM02-01 requirements and tables.

BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
In an effort to continuously improve our products, 2005628
the right is reserved to change the specification of the 03/05/12
items described herein at any time. Please contact us Page 7 of 9
for further information and up to date specifications. Issue 3

HTM02-01 Combined Medical & Surgical Air 11 bar

cAIR-3480-QGF11

cAIR-4470-QGF11

cAIR-4915-QGF11

cAIR-5460-QGF11

cAIR-8760-QGF11
cAIR-6000-PGF11

cAIR-6955-PGF11

cAIR-8000-PGF11
cAIR-1260-TGF11

cAIR-1755-TGF11

cAIR-2180-TGF11

cAIR-2460-TGF11
cAIR-485-TGF11

cAIR-800-TGF11

cAIR-905-TGF11
Model
Number

Design flow
485 800 905 1260 1755 2180 2460 3480 4470 4915 5460 6000 6955 8000 8760
(litres/minute)
Required compressor
324 532 601 1046 1457 1809 2039 2888 3710 4079 4532 4980 5773 6000 6570
output *
Required receiver
609 1000 1130 1155 1609 1998 2252 3190 3353 3686 4095 4500 5217 6000 6570
volume *
Number of
3 3 3 3 3 3 3 4 4 4 4 5 5 5 4
compressors
GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA-
Type of compressor MED MED MED MED MED MED MED MED MED MED MED MED MED MED MED
5 7 7 11 15 18 22 15 18 22 22 18 22 22 30
dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED
Type of dryer
7 13 13 25 25 35 35 50 70 70 80 100 100 145 145
Number of receivers 1 1 2 2 2 2 2 2 2 2 3 3 3 3 4
Receiver volume (l)
1000 1000 1000 1000 1000 1000 1500 2000 2000 2000 1500 1500 2000 2000 2000
(each)
Drawing Number

9827 8424 00
8153 0355 82
9827 8416 00

8153 0355 90
9827 8416 00

8153 0356 08
9827 8417 00

8153 0356 16
9827 8417 00

8153 0356 24
9827 8421 00

8153 0356 32
9827 8421 00

8153 0356 40
9827 8422 00

8153 0356 57

8153 0356 65
9827 8424 00

8153 0356 73
9827 8424 00

8153 0356 81
9827 8423 00

8153 0356 99
9827 8423 00

9827 8425 00

8153 0357 15
9827 8425 00
8153 0357 07
Part Number

n/a special
* Flow output based on a 50/50 split of medical and surgical air design flow (D.F).
For examples on how to calculate the required design flow, compressor flow and receiver volume refer to earlier page,
in line with HTM02-01 requirements and tables.

HTM02-01 Surgical Air 11 bar


sAIR-1445-SGF11

sAIR-1755-SGF11

sAIR-2460-SGF11

sAIR-3225-SGF11

sAIR-3890-SGF11

sAIR-4915-SGF11

sAIR-5615-SGF11
sAIR-485-SGF11

sAIR-750-SGF11

sAIR-905-SGF11

Model
Number

Design flow (litres/minute) 485 750 905 1445 1755 2455 3225 3890 4915 5615
Required compressor output * 161 495 597 954 1158 1642 2129 1945 2458 2808
Required receiver volume * 974 1000 1207 1927 2340 2460 3225 3890 4915 56105
Number of compressors 1 1 1 1 1 1 1 1 1 1
GA-MED GA-MED GA-MED GA-MED GA-MED GA-MED GA-MED GA-MED GA-MED GA-MED
Type of compressor
5 7 7 11 15 18 22 22 30 30
dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED
Type of dryer
7 13 13 25 25 35 50 70 70 80
Number of receivers 1 2 2 2 2 2 2 3 3 3
Receiver volume (l) (each) 1000 500 1000 1000 1500 1500 2000 1500 2000 2000
Drawing Number

8153 0357 23
9827 8416 00

8153 0357 31
9827 8415 00

8153 0357 49
9827 8417 00

8153 0357 56
9827 8417 00

8153 0357 64
9827 8422 00

8153 0357 72
9827 8422 00

8153 0357 80
9827 8424 00

8153 0357 98
9827 8423 00
Part Number

n/a special

n/a special

* For examples on how to calculate the required design flow, compressor flow and receiver volume refer to earlier page,
in line with HTM02-01 requirements and tables.

BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
In an effort to continuously improve our products, 2005628
the right is reserved to change the specification of the 03/05/12
items described herein at any time. Please contact us Page 8 of 9
for further information and up to date specifications. Issue 3

HTM02-01 Combined Medical & Surgical Air 11 bar - Varible Speed Drive

Available upon request

HTM2022 EurPh Medical Air 4 bar

mAIR-7045-QGF4
mAIR-6000-QGF4
mAIR-1000-DGF4

mAIR-1230-DGF4

mAIR-1465-DGF4

mAIR-1720-DGF4

mAIR-2000-DGF4

mAIR-2430-DGF4

mAIR-2855-DGF4

mAIR-3445-TGF4

mAIR-3935-TGF4

mAIR-4860-TGF4

mAIR-5570-TGF4
mAIR-340-DGF4

mAIR-500-DGF4

mAIR-630-DGF4
Model
Number

Design flow
340 500 630 1000 1230 1465 1720 2000 2430 2855 3445 3935 4860 5570 6000 7045
(litres/minute)
Number of
2 2 2 2 2 2 2 2 2 2 3 3 3 3 4 4
compressors
GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA-
Type of
MED MED MED MED MED MED MED MED MED MED MED MED MED MED MED MED
compressor
5 5 5 7 11 11 15 15 18 22 15 15 18 22 15 18
dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED
Type of dryer
7 13 13 25 25 35 35 50 50 70 70 80 100 145 145 145
Number of receivers 1 1 1 1 1 1 1 1 1 1 1 1 2 2 2 2
Receiver volume (l)
250 250 500 500 1000 1000 1000 1000 1500 1500 2000 2000 1500 1500 1500 2000
(each)
Drawing Number

8153 0358 06
9827 8427 00

8153 0358 14
9827 8427 00

8153 0358 22
9827 8428 00

8153 0358 30
9827 8428 00

8153 0358 48
9827 8429 00

8153 0358 55
9827 8429 00

8153 0358 63
9827 8430 00

8153 0358 71
9827 8430 00

8153 0358 89
9827 8431 00

8153 0358 97
9827 8431 00

8153 0359 05
9827 8434 00

8153 0359 13
9827 8434 00

8153 0359 21
9827 8432 00

8153 0359 39
9827 8432 00

8153 0359 47
9827 8432 00

8153 0359 54
9827 8435 00
Part Number

HTM2022 EurPh Combined Medical & Surgical Air 7 bar

cAIR-6000-QGF7

cAIR-7655-QGF7
cAIR-1000-DGF7

cAIR-1290-DGF7

cAIR-1530-DGF7

cAIR-1800-DGF7

cAIR-2140-DGF7

cAIR-2535-DGF7

cAIR-8790-PGF7
cAIR-3000-TGF7

cAIR-3600-TGF7

cAIR-4285-TGF7

cAIR-4895-TGF7
cAIR-425-DGF7

cAIR-555-DGF7

cAIR-785-DGF7

Model
Number

Design flow
425 555 785 1000 1290 1530 1800 2140 2535 3000 3600 4285 4895 6000 7655 8790
(litres/minute)
Number of
2 2 2 2 2 2 2 2 2 3 3 3 3 4 4 5
compressors
GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA-
Type of compressor MED MED MED MED MED MED MED MED MED MED MED MED MED MED MED MED
5 5 7 11 11 15 15 18 22 15 15 18 22 18 22 18
dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED
Type of dryer
7 13 13 25 25 25 35 35 50 50 70 70 80 100 145 145
Number of receivers 1 1 1 1 1 1 1 1 1 1 1 2 2 2 2 3
Receiver volume (l)
250 500 500 500 1000 1000 1000 1500 1500 1500 2000 1500 1500 1500 2000 1500
(each)
Drawing Number

8153 0359 62
9827 8427 00

8153 0359 70
9827 8428 00

8153 0359 88
9827 8428 00

8153 0359 96
9827 8428 00

8153 0360 02
9827 8429 00

8153 0360 10
9827 8430 00

8153 0360 28
9827 8430 00

8153 0360 36
9827 8431 00

8153 0360 44
9827 8431 00

8153 0360 51
9827 8431 00

8153 0360 69
9827 8434 00

8153 0360 77
9827 8432 00

8153 0360 85
9827 8432 00

8153 0360 93
9827 8432 00

9827 8435 00

8153 0361 19
9827 8433 00
8153 0361 01
Part Number

BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
In an effort to continuously improve our products, 2005628
the right is reserved to change the specification of the 03/05/12
items described herein at any time. Please contact us Page 9 of 9
for further information and up to date specifications. Issue 3

HTM2022 Combined Medical & Surgical Air 11 bar

cAIR-4915-QGF11

cAIR-5615-QGF11

cAIR-6960-QGF11
cAIR-1000-DGF11

cAIR-1535-DGF11

cAIR-1755-DGF11

cAIR-2325-DGF11

cAIR-7270-PGF11

cAIR-8790-PGF11
cAIR-3000-TGF11

cAIR-3710-TGF11

cAIR-4645-TGF11
cAIR-425-DGF11

cAIR-500-DGF11

cAIR-705-DGF11

cAIR-905-DGF11
Model
Number

Design flow
425 500 705 905 1000 1535 1755 2325 3000 3710 4645 4915 5615 6960 7270 8790
(litres/minute)
Number of
2 2 2 2 2 2 2 2 3 3 3 4 4 4 5 5
compressors
GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA- GA-
Type of compressor MED MED MED MED MED MED MED MED MED MED MED MED MED MED MED MED
5 7 7 11 11 15 18 22 15 18 22 18 18 22 18 22
dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED dMED
Type of dryer
7 13 13 13 25 25 25 35 50 70 70 70 80 100 145 145
Number of receivers 1 1 1 1 1 1 1 1 1 1 2 2 2 2 2 3
Receiver volume (l)
250 250 500 500 500 1000 1000 1500 1500 2000 1500 1500 1500 2000 2000 2000
(each)
Drawing Number

8153 0361 27
9827 8427 00

8153 0361 35
9827 8427 00

8153 0361 43
9827 8428 00

8153 0361 50
9827 8428 00

8153 0361 68
9827 8428 00

8153 0361 76
9827 8430 00

8153 0361 84
9827 8430 00

8153 0361 92
9827 8431 00

8153 0362 00
9827 8431 00

8153 0362 18
9827 8434 00

8153 0362 26
9827 8432 00

8153 0362 34
9827 8432 00

8153 0362 42
9827 8432 00

8153 0362 59
9827 8435 00

8153 0362 67
9827 8435 00

8153 0362 75
9827 8436 00
Part Number

• Design flow in terms of free air delivered after losses at working pressure with the reserve compressor(s) on standby. Tolerance ±5%.
• Component dimensions supplied do not include maintenance access space, and are provided to allow customer to arrange plant
components within plant room. Complete installation drawings are available on request. Quote the drawing number required.
• Duplex systems must be installed with a manifold as the third source of supply for HTM02-01 complaince.
• Mean sound level in accordance with ISO 2151.
• Electrical details are provided for guidance only. Site conditions may impose a larger cable size. For exact cable sizing, and fuse / MCB
ratings, consult a qualified electrical engineer.

BeaconMedæs
Telford Crescent,
Staveley, Derbyshire
S43 3PF, England
Tel: +44 (0) 1246 474 242
Fax: +44 (0) 1246 472 982
www.beaconmedaes.com 0088
E-mail: gbn.info@beaconmedaes.com
2005161

VACUUM SYSTEM
In an effort to continuously improve our products,

mVAC MEDICAL
the right is reserved to change the specification of the 08/11/11
items described herein at any time. Please contact us Page 1 of 4
for further information and up to date specifications. Issue 6

mVAC Medical Vacuum System 400v 50Hz


– EN ISO 7396-1 & HTM 02-01
SPECIFICATION

mVAC Medical Vacuum Optional Control Equipment


The mVAC Medical Vacuum System shall conform to EN ISO 7396-1 An advanced monitoring system shall be available to give immediate
and NHS Health Technical Memorandum No. 02-01 (HTM 02-01). The access to valuable information such as system status, trends, historical
Medical Vacuum System shall ensure the minimum pipeline vacuum data and system performance. Data collected from all pumps shall be
level of 450mmHg is maintained at the plant service connection point at made available in real-time visualisation pages and shall be accessed
the rated volumetric ‘free air’ flow rate with two pumps in standby. The through the hospital’s LAN, such that total data security is assured.
bacteria filtration system shall be ‘duplexed’ such that each filter can be
isolated for replacement of the filter cartridge. The Airconnect™ monitoring system shall also include :-

Vacuum Pumps • Logging and trending for an accurate performance status of your
Vacuum pumps shall be air-cooled, oil lubricated rotary vane type system.
suitable for both continuous and frequent start/stop operation at
nominal inlet vacuum levels of between 578mmHg and 728mmHg. • Desktop event notification to avoid constant status checking.
Composite carbon fibre rotor blades shall be fitted to minimise the
cost of maintenance. Rotors shall be driven by directly coupled TEFV, • E-mail and SMS event notification for additional convenience.
IE2 efficiency electric motors. Pump inlets shall include a wire mesh
filter and integral non-return valve to prevent oil suck back and pressure Optional Items
increases in the vacuum system. Each vacuum pump shall have an There shall be the followings options available for enhanced operation
integral separator filter to ensure a virtually oil-free exhaust. Each pump of the vacuum plant system:-
shall be fitted with anti-vibration pads between the pump foot and
mounting frame. • Synthetic oil for increased pump life.

Bacteria Filters • Painted hot dipped galvanised vessels.


The duplex bacteria filter system shall incorporate high efficiency filter
elements. A differential vacuum indicator shall be installed across • Oil level switch and integrated ES-VAC alarm for notification of low
the filter to indicate blockage. Additional pressure sensors shall be oil level in the pumps.
installed at the inlet and outlet of the filter to measure the pressure
drop across the filters. Each filter shall be designed and sized to carry Vacuum Receiver(s)
the full plant design flow capacity with a pressure drop not exceeding Vacuum receiver(s) shall be supplied with relevant test certificates and
33mbar (25mmHg). Bacteria Filter elements shall have penetration have a total volume of at least 100% of the plant output in 1 minute
levels not exceeding 0.005% when tested by the sodium flame method in terms of free air aspired at normal working pressure. Each vacuum
in accordance with BS 3928:1969 and utilising particles in the 0.02 to receiver shall be hot dip galvanised inside and out.
2 micron size range. Drain flasks shall be connected to each filter. Drain
flasks shall be manufactured from transparent Pyrex® with a polymer Pyrex® is a registered trademark of Corning Glass.
coating on the inner and outer surfaces in order to maintain a seal in the
event of inadvertent breakage of the Pyrex® flask. All drain flasks shall be CE Marking
suitable for sterilisation and be connected via a manual isolating valve. The standard range of BeaconMedæs Medical Vacuum plant systems
are ‘CE’ marked under the Medical Devices Directive 93/42/EEC
Control System with approval from notified body no. 0088 (Lloyd’s Register Quality
The central control system shall provide an intelligent human machine Assurance). Under this directive, the specified products are classified as
interface incorporating on board flash memory and real-time clock for Class IIa Medical Devices.
recording operational parameters in the in built event log. The central
control system shall operate at low voltage and include BMS connection
for commont fault. Visualisation of plant inputs, outputs and status
through a web browser, using a simple Ethernet connection shall be
available. The central control unit shall incorporate a user friendly 5.7”
high-definition colour display with clear pictograms and LED indicators,
providing easy access to system operational information.

Cascading of vacuum pumps shall be achieved by measuring the vacuum


level at the plant inlet with a pressure transducer. A mechanical back-up
facility shall ensure continued operation in the event of a control system
malfunction. The control system shall normally employ automatic
rotation of the lead pump to maximise pump life and ensure even wear.

BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
In an effort to continuously improve our products, 2005161
the right is reserved to change the specification of the 08/11/11
items described herein at any time. Please contact us Page 2 of 4
for further information and up to date specifications. Issue 6

Typical Layout

Model L (ength) W (idth) H (eight) LMA WMA F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø

mVAC250 T 2300 980 1650 3600 1980 500 800 54 54 n/a

mVAC330 T 2300 980 1650 3600 1980 500 800 54 54 n/a

mVAC500 T 2400 980 1650 3700 1980 500 800 54 54 n/a

Model L (ength) W (idth) H (eight) ØD LMA WMA F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø

mVAC660 Q 1910 980 2100 660 3210 3740 500 800 54 54 54 outl. unit - 54 inl. vessel

mVAC1000 Q 1910 980 2100 660 3210 3740 500 800 54 54 54 outl. unit - 54 inl. vessel

mVAC2560 Q 2600 1200 2400 1060 3900 4360 500 800 76 54 76 outl.unit - 54 inl. vessel

mVAC3300 Q 2600 1200 2500 1230 3900 4530 500 800 76 54 76 outl.unit - 76 inl. vessel

mVAC3840 P 4100 1250 2500 1230 5990 4580 500 800 76 54 76 outl.unit - 76 inl. vessel

BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
In an effort to continuously improve our products, 2005161
the right is reserved to change the specification of the 08/11/11
items described herein at any time. Please contact us Page 3 of 4
for further information and up to date specifications. Issue 6

Model L (ength) W (idth) H (eight) ØD LMA WMA F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø

mVAC1280 T 2200 1100 2100 660 4280 3860 500 800 76 54 54 outl. unit - 54 inl. Vessel

mVAC4950 P 4100 1250 2500 1230 5990 4580 500 800 108 54 108 outl. unit - 76 inl. Vessel

mVAC6000 H 4100 1250 2500 1230 7220 4580 500 800 108 54 108 outl.unit - 76 inl. Vessel

Model L (ength) W (idth) H (eight) ØD LMA WMA F G I (nlet) Ø E (xhaust) Ø V (essel con.) Ø

mVAC6600 H 4100 1250 2500 1230 7220 4580 500 800 108 54 108 outl.unit - 76 inl. Vessel

Note!

mVAC-6600-H includes 4 x vessels

Drawing shows other models not used in this range and are for
illustration purposes only.

BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
In an effort to continuously improve our products, 2005161
the right is reserved to change the specification of the 08/11/11
items described herein at any time. Please contact us Page 4 of 4
for further information and up to date specifications. Issue 6

mVAC-1000-Q

mVAC-3300-Q
mVAC-2560-Q

mVAC-6000-H

mVAC-6600-H
mVAC-500-TH
mVAC-250-TH

mVAC-330-TH

mVAC-3840-P

mVAC-4950-P
mVAC-1280-T
mVAC-660-Q
mVAC-500-T
mVAC-250-T

mVAC-330-T
Model Number

Plant Output
250 250 330 330 500 500 660 1000 1280 2560 3300 3840 4950 6000 6600
(litres/minute)
System Flow 40 40 52 52 79 79 105 159 203 406 524 609 786 952 1047
(m3/h - l/m) 665 665 865 865 1315 1315 1750 2650 3380 6765 8730 10150 13100 15865 17450
Number of pumps 3 3 3 3 3 3 4 4 3 4 4 5 5 6 6
Number of receivers 1(H) 1 1(H) 1 1(H) 1 2 2 3 2 2 2 3 3 4
Total receiver
500 500 500 500 500 500 1000 1000 1500 3000 4000 4000 6000 6000 8000
volume (litres)
Receiver
n/a 54 n/a 54 n/a 54 54 54 54 54 76 76 76 76 76
connection(s) (mm)
Inlet/service
54 54 54 54 54 54 54 54 54 76 76 76 108 108 108
connection (mm)
Exhaust connection
54 54 54 54 54 54 54 54 54 54 54 54 54 54 54
(mm, per pump)
Maximum exhaust
60 60 60 60 60 60 60 60 60 60 60 60 60 60 60
back pressure (mbar)
Noise level/pump
67 67 69 69 70 70 69 70 72 72 75 72 75 75 75
(dB[A])
Weight (kg) 650 - 690 - 750 - 960 1040 1475 2215 2335 2660 3190 3390 3820
Motor rating (kW) 1.1 1.1 1.5 1.5 2.2 2.2 1.5 2.2 5.5 5.5 7.5 5.5 7.5 7.5 7.5
Motor cable size 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 2.5 1.5 2.5 2.5 2.5
(mm2/Amps) (13) (13) (13) (13) (13) (13) (13) (13) (13) (13) (17) (13) (17) (17) (17)
Motor rated supply
10 10 10 10 10 10 10 10 20 20 20 20 20 20 20
per pump (A)
FLC per pump (A) 2.8 2.8 3.6 3.6 5.1 5.1 3.6 5.1 12 12 14.9 12 14.9 14.9 14.9
Starting current (A) 14 14 18 18 26 26 18 26 60 60 75 60 75 75 75
Central control supply
1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5
- single phase
(5) (5) (5) (5) (5) (5) (5) (5) (5) (5) (5) (5) (5) (5) (5)
(mm2/Amps)
Maximum Inlet
40 40 40 40 40 40 40 40 40 40 40 40 40 40 40
Temperature (0C)
Cooling air flow
0.1 0.1 0.1 0.1 0.2 0.2 0.1 0.2 0.3 0.3 0.4 0.3 0.4 0.4 0.4
per pump (m3/s)
Installation Proposal

8152 1306 04
9820 6365 00
8152 1306 00

8152 1306 48

9820 6365 00
81521 306 02

81521 306 49

9820 6365 00

8152 1306 50

9820 6368 00
8152 1306 06

9820 6368 00
8152 1306 08

9820 6369 00
8152 1306 10

9820 6368 00
8152 1306 12

9820 6368 00
8152 1306 14

9820 6368 00
8152 1306 16

9820 6369 00
8152 1306 18

9820 6369 00
8152 1306 20

9820 6370 00
8152 1306 22
Part Number

• Plant Output in terms of free air aspired at a vacuum of 450 mmHg at the inlet connection with two pumps on standby and with a
tolerance of ±10%.
• System Flow at atmospheric pressure at the inlet connection with two pump on standby and with a tolerance of ±10%.
• Plant dimensions include the required space around the plant for maintenance access.
• Mean sound level measured at a distance of 1m as measured to ISO 2151 / DIN 45635.
• Electrical details are provided for guidance only and are referenced at 400C ambient temperature. Site conditions may impose a larger
cable size. For exact cable sizing, and fuse / MCB ratings, consult a qualified electrical engineer.
• Plant weight includes packaging for shipping purposes.

BeaconMedæs
Telford Crescent,
Staveley, Derbyshire
S43 3PF, England
Tel: +44 (0) 1246 474 242
Fax: +44 (0) 1246 472 982
www.beaconmedaes.com
E-mail: gbn.info@beaconmedaes.com
GAS SCAVENGING
In an effort to continuously improve our products, 2005208

ANAESTHETIC
the right is reserved to change the specification of the 16/03/12
items described herein at any time. Please contact us Page 1 of 3
for further information and up to date specifications. Issue 7

Anaesthetic Gas Scavenging System


HTM02-01/HTM2022/ISO7396-2
SPECIFICATION

Anaesthetic Gas Scavenging System Simplex starter/isolator panels c/w alarm pressure switch and duplex
The Anaesthetic Gas Scavenging (AGS) System shall comply with HTM units incorporate line pressure switch. This line pressure switch
02-01, HTM2022 and either EN ISO 7396-2 or BS 6834. The AGS monitors vacuum levels and provides an additional control of the
system shall be a dedicated, specifically designed active extraction remote start switch and starter/isolator panel green ‘RUNNING’
and disposal system for waste anaesthetic gas. It shall provide a indicators. The pressure switch shall also include a digital display
maximum flow rate of 80 l/min (EN ISO 7396-2) or 130 l/min (BS providing an accurate readout of the vacuum level in the pipeline in
6834) with a 1 kPa resistance to flow, and a minimum of 50 l/min (EN order to assist with installation/commissioning and annual
ISO 7396-2) or 80 l/min (BS 6834) with a 2 kPa (EN ISO 7396-2) or 4 re-commissioning.
kPa (BS 6834) resistance to flow at each terminal unit,
irrespective of the number of terminal units in use. The AGS system Simplex installations shall use remote start switches that include a red
shall use dedicated radial blowers in a simplex or duplex ‘PLANT EMERGENCY’ indicator. This indicator shall illuminate on all
configuration. The AGS pump assemblies shall be skid mounted and remote start switch panels if the vacuum level falls below the
included on the skid shall be the simplex or duplex pump(s), motor pressure switch set point level when the pump has been called, or if
control unit(s) with starter/isolator, moisture drain flask and flexible the overload trips. The on/off rocker switch shall include a green
connector(s) to connect the plant to the pipeline. Each pump shall illuminated surround to indicate ‘mains on’.
include an electric motor and directly coupled impeller assembly.
Impeller bearings in the pump(s) shall not require lubrication. The Duplex installations shall use remote start switches that include an
pump(s) shall be air cooled and rated for continuous operation. amber ‘PLANT FAULT’ indicator. This shall illuminate, if either pump is
set to ‘HAND’, or if one of the overloads trip. A red ‘PLANT
Vacuum/Flow Regulating Valve EMERGENCY’ indicator shall also be provided and shall illuminate on
A vacuum/flow regulating valve shall be provided and positioned all remote start switch panels if the vacuum level falls below the
at the pump, comprised of a spring-loaded plate valve and inlet pressure switch set point level when the pump has been called. The
silencer. The valve should be changeable with the pipeline inlet in on/off rocker switch shall include a green illuminated surround to
order to provide flexibility on site. The plate shall control air ingress indicate ‘mains on’.
into the pipeline system, thereby controlling the vacuum level within.
An optional air inlet filter shall be available should the air quality Where a duplex system is installed each pump shall be controlled by
be poor/dusty offering further protection against dirt ingress into a separate motor control panel to enable servicing of either pump or
the pump. Additional in line vacuum/flow regulating valves may be control gear whilst maintaining system operation.
installed if required and shall be determined by the pipeline designer.
The vacuum/flow regulating valve shall ensure a maximum vacuum Volt free relay kits for replicating alarm conditions to BMS shall be
of 200mb below atmospheric pressure is not exceeded and shall be available as an optional extra. To be either installed either at factory or
factory preset at 150mb. as a retro-fit kit for onsite installation.

Control System Terminal Units


Each motor control panel shall incorporate an emergency panel Terminal unit shall be provided with an adjustable orifice to allow
isolation switch facility, which controls all electrical power to the balancing of the terminal unit flows during commissioning. Venturi
exhauster unit, remote start switch panels and system indication style terminal units are not acceptable. Terminal units shall not be
lights. All control and status indication circuitry shall be limited to connected to the medical vacuum system.
24V a.c. A green ‘POWER ON’ indicator shall be fitted to the starter/
isolator panel, and shall illuminate whenever power is available to the CE Marking
24V control and indication circuit. A ‘HAND/OFF/AUTO’ switch shall The standard range of BeaconMedæs Anaesthetic Gas Scavenging
be provided to control operation of the pump, running the pump System are ‘CE’ marked under the Medical Devices Directive 93/42/EEC
continuously when selected to ‘HAND’. When selected to ‘AUTO’, with approval from notified body no. 0088 (Lloyd’s Register Quality
control of the pump shall be passed to the remote start switch panels. Assurance). Under this directive, the specified products are classified as
Operation of any of the remote start switches shall activate the pump. Class IIb Medical Devices.
The pump shall continue to run until all remote switches are selected
‘OFF’.

The starter/isolator panel shall incorporate a thermal protection


overload device. The thermal protection overload device shall also
monitor the electrical power supply and phase input. In the event of a
fault, the overload device shall break the circuit to the pump,
preventing operation until the system is manually re-set. Operation
of the overload device shall also break the circuit to the remote start
switch panels, extinguishing the green running indicator.

BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
In an effort to continuously improve our products, 2005208
the right is reserved to change the specification of the 16/03/12
items described herein at any time. Please contact us Page 2 of 3
for further information and up to date specifications. Issue 7

AGS Pump Assembly Details

Motor start current (A)

Min. cable size (mm2)


Supply frequency (Hz)

Noise level per pump


Motor rated fused

Connection (mm)
supply/pump (A)

Motor plate (kW)


FLC/pump (A)

Length (mm)

Height (mm)
Width (mm)
Part number

@ 1m dB(A)
Weight (kg)
Voltage (V)

Phase ~
Format
Model

AGS-
4153 1254 00 400 3 1.8 9 2.5 16 900 525 1160 96 0.75 60 54
520-D/3
Duplex

AGS-
4153 1255 00 400 3 3.3 23 2.5 16 900 525 1160 121 1.5 64 54
1560-D/3
AGS-
4153 1256 00 400 3 7.2 48 4 16 900 525 1160 139 3 69 54
2860-D/3
50 Hz

AGS-
4153 1250 00 230 1 5 20 2.5 16 525 525 1160 56 0.75 60 54
520-S/1
AGS-
4153 1251 00 400 3 1.8 9 2.5 16 525 525 1160 56 0.75 60 54
Simplex

520-S/3
AGS-
4153 1252 00 400 3 3.3 23 2.5 16 525 525 1160 68 1.5 64 54
1560-S/3
AGS-
4153 1253 00 400 3 7.2 48 2.5 16 525 525 1160 77 3 69 54
2860-S/3
AGS-
4153 1261 00 380 3 1.8 10 2.5 16 900 525 1160 96 0.9 64 54
650-D/3
Duplex

AGS-
4153 1262 00 380 3 3.3 23 2.5 16 900 525 1160 121 1.8 69 54
2210-D/3
AGS-
4153 1263 00 380 3 7.2 48 4 16 900 525 1160 139 3.6 70 54
3770-D/3
60 Hz

AGS-
4153 1257 00 220 1 5.5 18 2.5 16 525 525 1160 56 0.9 64 54
650-S/1
AGS-
4153 1258 00 380 3 1.8 10 2.5 16 525 525 1160 56 0.9 64 54
Simplex

650-S/3
AGS-
4153 1259 00 380 3 3.3 23 2.5 16 525 525 1160 68 1.8 69 54
2210-S/3
AGS-
4153 1260 00 380 3 7.2 48 2.5 16 525 525 1160 77 3.6 70 54
3770-S/3

Notes…

1. Noise levels are averages, measured in accordance with ISO 517:1996 Free Air Aspiration and ISO 2151 Mean Sound Level
2. Service connections are copper to BS EN 13348 and relate to the external pipe diameter

BeaconMedæs ● Telford Crescent, Staveley, Derbyshire S43 3PF ● Tel: +44 (0) 1246 474 242 ● Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) ● Company VAT Reg No. 598612590 ● www.beaconmedaes.com ● gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 ● CHAS accredited contractor ● SAFEcontractor Reg No. N00184450
In an effort to continuously improve our products, 2005208
the right is reserved to change the specification of the 16/03/12
items described herein at any time. Please contact us Page 3 of 3
for further information and up to date specifications. Issue 7

Typical AGS Pump Assembly Layouts

Duplex Pump Assembly

10 7

1 1 3
7
1160

8
Ø54 10 Ø54
6

5 ALT. 2 ALT. 1
INLET INLET
8

9 2
40 445 15 870
525 900 CABLE MOUNTING BY
ENTRANCE 4x M8 BOLT

Simplex Pump Assembly

2 Ø54 10

10
7 ALT. 2 ALT. 1
3 INLET INLET

1 9
7 9 2
1160

8 CABLE MOUNTING BY
6 ENTRANCE 4x M8 BOLT

REF. DESCRIPTION
1 FLEXIBLE HOSE
2 FLOW/REGULATING VALVE
40 445 15 495 3 PRESSURE SWITCH
525 525 4 NON-RETURN VALVE
5 DRAIN FLASK
6 ISOLATION VALVE
7 MOTOR CONTROL PANELS
8 PUMPS
9 INLET CONNECTION
10 EXHAUST CONNECTIONS

Note: Items 2 & 9, Flow regulating valve and inlet connection, can be re positioned on site if required

BeaconMedæs
Part of the Atlas Copco Group
Telford Crescent, Staveley,
Derbyshire, S43 3PF, England
Tel: +44 (0) 1246 474 242 0088
www.beaconmedaes.com