SERVICE MANUAL
Foreword . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1- Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2- Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
3- Nomenclature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Intelect/Genisys Ultrasound Therapy System . . . . . . . . . . . . .4
Intelect/Genisys Ultrasound Symbol Definitions . . . . . . . . . . .5
4- Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
5- Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-14
5.1 Intelect/Genisys Ultrasound Software Error Messages .7-9
5.2 Intelect/Genisys Ultrasound Diagnostics . . . . . . . . . . . .10
5.3 Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
5.4 Leakage Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
5.5 Unit Startop and Fan Testing . . . . . . . . . . . . . . . . . . . . .11
5.6 Ultrasound Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
5.7 Ultrasound Applicator Identification Test . . . . . . . . . . . . .12
5.8 Ultrasound Applicator Output Test . . . . . . . . . . . . . . . . .13
5.9 Ultrasound Duty Cycle Test . . . . . . . . . . . . . . . . . . . . . .14
6- Removal & Replacement . . . . . . . . . . . . . . . . . . . . . . . .15-20
6.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
6.2 Top & Bottom of Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
6.3 Fan Removal & Replacement . . . . . . . . . . . . . . . . . . . . .16
6.4 Power Supply Removal & Replacement . . . . . . . . . . . . .17
6.5 Ultrasound Board Removal & Replacement . . . . . . . . . .18
6.6 LCD Display Removal & Replacement . . . . . . . . . . . . . .19
6.7 Control Board Removal & Repalcement . . . . . . . . . . . . .20
7- Ultrasound Applicator Calibration . . . . . . . . . . . . . . . . . .21
8- Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-27
9- Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-33
10- Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
©2004 Encore Medical Corporation or its affiliates, Austin, Texas, USA. Any use of editorial, pictorial or layout composition of this publication without express written consent
from the Chattanooga Group of Encore Medical, L.P. is strictly prohibited. This publication was written, illustrated and prepared for print by the Chattanooga Group of Encore
Medical, L.P.
FOREWORD Intelect®/Genisys® Ultrasound Therapy Systems
Read, understand and follow the Safety Precautions and information contained in this manual.
This manual contains the necessary safety, and field service information for those Field Service Technicians,
approved by Chattanooga Group, to perform field service on the Intelect/Genisys Ultrasound Therapy Systems.
At the time of publication the information contained herein was current and up to date. However, due to
continual technological improvements and increased clinical knowledge in the field of electrotherapy, as well as
Chattanooga Group’s policy of continual improvement, Chattanooga Group reserves the right to make periodic
changes and improvements to their equipment and documentation without any obligation on the part of
Chattanooga Group.
It is the sole responsibility of field technicians to stay informed and trained in the latest technology utilized in
the Intelect/Genisys Ultrasound Therapy Systems by Chattanooga Group. From time to time, as significant
improvements are incorporated, Service Bulletins will be produced and made available on our web site
(www.chattgroup.com) in lieu of reprinting a complete manual prematurely. These Service Bulletins will provide
updated service information and technology improvements to the Intelect/Genisys Ultrasound Therapy Systems
for use by certified service technicians.
1. Level I- Those Field Service Technicians that have successfully completed the minimal training required by
Chattanooga Group in basic service techniques.
2. Level II- Those Field Service Technicians that have successfully completed Level I Training as well as
Level II training as required to perform specific troubleshooting and repair techniques and
procedures.
3. Level III- Those Field Service Technicians that have successfully completed Levels I & II Training as well as
Level III Advanced Training as required to perform all necessary Troubleshooting and Repair
techniques. The Technician having successfully completed the three levels of training and coupled
with experience should have the ability to train other technicians in Level I and Level II Training
with the necessary Training Materials from Chattanooga Group.
4. Temporary- Chattanooga Group, at its discretion and based on known experience of the technician, may
grant a “Temporary Certification” to a field technician for particular troubleshooting and repair
of a specific system requiring immediate attention. This “Temporary Certification” in no fashion
acknowledges the training level of a technician as defined above. This “Temporary
Certification” is utilized only in unique situations for a specific unit for a specific service
technique only and is documented as such.
Due to the complex nature of the technology utilized by Chattanooga Group, the recommended troubleshooting
techniques are to determine “Bad Board” and board replacement only. No board component level
troubleshooting is recommended nor will information or parts be supplied by Chattanooga Group. Any board
component level troubleshooting performed will be at sole risk and liability of the Service Technician
performing such troubleshooting techniques.
This equipment is to be used only under the prescription and supervision of a licensed medical practitioner.
1
1- SAFETY PRECAUTIONS Intelect®/Genisys® Ultrasound Therapy Systems
! CAUTION
! CAUTION
Text with a “CAUTION” indicator will explain • Read, understand and practice the precautionary
possible safety infractions that have the potential to and operating instructions. Know the limitations
cause minor to moderate injury or damage to and hazards associated with using any ultrasound
equipment. device. Observe the precautionary and operational
B. WARNING decals placed on the unit.
• DO NOT operate the Intelect or Genisys when
! WARNING connected to any unit other than Chattanooga
Group devices. Do not operate the unit in an
Text with a “WARNING” indicator will explain environment of short-wave diathermy use.
possible safety infractions that will potentially cause • The Ultrasound modality should be routinely
serious injury and equipment damage. checked before each use to determine that all
C. DANGER controls function normally; especially that the
intensity control does properly adjust the intensity
! DANGER of the ultrasonic power output in a stable manner.
Also, determine that the treatment time control
does actually terminate ultrasonic power output
Text with a “DANGER” indicator will explain when the timer reaches zero.
possible safety infractions that are imminently
hazardous situations that would result in death or • Use of controls or adjustments or performance of
serious injury. procedures other than those specified herein may
result in hazardous exposure to ultrasonic energy.
D. EXPLOSION HAZARD
• DO NOT use sharp objects such as a pencil point
Do not use this equipment in the presence or ballpoint pen to operate the buttons on the
of flammable anesthetics. This symbol is control panel as damage may result.
also prominently displayed on the serial • Operate, transport and store this unit in
number plate of the unit. temperatures between 59 °F and 104 °F (15 °C
and 40 °C), with Relative Humidity ranging from
E. DANGEROUS VOLTAGE 30%- 60%.
• Inappropriate handling of, and subjecting the
Text with a “Dangerous Voltage” indicator serves to ultrasound applicator to physical abuse, may
inform the Service Technician of possible hazards adversely affect its characteristics.
resulting in the electrical charge retained by faulty
power supplies. • Inspect Sound Head for cracks, which may allow
the ingress of conductive fluid before each use.
F. CORROSIVE HAZARD (NiMH Battery) • Inspect all cables, leads and associated
connectors before each use.
Text with a “Corrosive Hazard” indicator
will explain possible safety infractions if the
chemical components of this product are
exposed to air, skin or other materials.
G. NOTE:
Throughout this manual “NOTE” may be found.
The Notes are helpful information to aid in the
particular area or function being described.
2
2- THEORY of OPERATION Intelect®/Genisys® Ultrasound Therapy Systems
2.1 Overview
The Intelect/Genisys Ultrasound Therapy Systems are comprised of several PC board assemblies housed within a
common enclosure. These assemblies each support a distinct function in the product. The basic elements are User
Interface, Control Board, Ultrasound Board, Ultrasound Applicator, and Power Supply Circuits.
3
3- NOMENCLATURE Intelect®/Genisys® Ultrasound Therapy Systems
1 3
2
16
15
4
17
14 5
13 6
12
10
11
FIGURE 3.1
1. On/Off Switch- Turns system On and Off. 11. Treatment Pause Button- Press to pause therapy session.
2. Fan Grill- Allow exhaust of air circulated by system fan. Press again to continue session.
3. LCD Contrast Knob-Adjusts contrast of LCD Display. 12. Treatment Stop Button- Press to completely stop therapy
4. Plynth- Allows for wall mounting or sitting on a flat surface. session.
5. Ultrasound Applicator- Used to administer Ultrasound 13. Multi-Control Button- Press the respective icon on this
Therapy to patient. button to change parameters and to scroll through Clinical
6. Applicator LED Indicator- Flashes 5 times upon connecting Resources Library
applicator to system. Illuminates constantly when applicator 14. Treatment Time Button- Press to adjust treatment time up
is coupled to patient treatment area. or down.
7. Ultrasound Intensity Button- Used to increase and 15. Clinical Resources Library Button- Press to access the
decrease ultrasound output intensity. Clinical Resources Library.
8. Ultrasound Applicator Hook- Used to store applicator when 16. LCD Display- Displays necessary information for operation
not in use. and maintenance.
9. Ultrasound Applicator Connector- Used to connect the 17. Mains Power Cord- Connects system to required electrical
Ultrasound Applicators to the System outlet.
10. Treatment Start Button- Press to start therapy session. Not Illustrated- Battery Access Cover located beneath the
plynth.
4
3- NOMENCLATURE Intelect®/Genisys® Ultrasound Therapy Systems
On/Off Switch- Flashes when plugged into Frequency Press to change frequency to 3.3 MHz or
power outlet. Illuminates steady when system 1MHz
is On
Stop Treatment- Press to stop therapy session. Intensity Display- Press to display “Watts” or
“Watts per cm2“ for the output of the ultrasound.
5
4- SPECIFICATIONS Intelect®/Genisys® Ultrasound Therapy Systems
6
5- TROUBLESHOOTING Intelect®/Genisys® Ultrasound Therapy Systems
5.1 Intelect/Genisys Ultrasound Software Error Messages primarily user definable and remedied by following
A. The information provided below is intended to aid in the instructions given by the unit. Error meesages
defining the Software Error Messages of the Ultrasound in the ranges of 200- 299 and 300-399, require the
Therapy System. Once a particular Error Message is attention of a field technician certified by
defined the information will also list probable causes Chattanooga Group in order to perform the
and possible remedies. Once the problem area is necessary diagnostics and properly replace any
determined subsequent diagnostics for verification necessary components.
will be necessary to determine “Bad Board”. All
diagnostic tests will be to validate a “Bad B. Once a particular PCB has been determined as
Board” only. No component level troubleshooting bad, refer to the appropriate Removal and
information is or will be provided by Chattanooga Replacement Section for the board affected for
Group for field troubleshooting of board components. proper instructions for replacement of the board.
Error messages in the range of 100 to 199 are
CODE No. TYPE MESSAGE PROBABLE CAUSES POSSIBLE REMEDIES
Icon- Sound Head 1. Sound Head has overheated to a A. Turn Off Head Warming feature.
overtemp temperature that could cause damage B. Allow Sound Head to cool before
to the crystal. continuing therapy.
C. Connect a known good Applicator to
the system and try again. If system
works properly, re-calibrate Applicator
or replace with new Applicator.
Icon- No applicator 1. Applicator not plugged into system A. Plug Applicator into system.
plugged into connector.
system 2. Applicator connector not completely A. Make certain Applicator connector is
seated in system connector. completely seated in system
connector.
3. Applicator Cable disconnected from A. Connect Applicator Cable to
Applicator PCB. Applicator PCB
4. Applicator cable damaged. A. Replace with new Applicator.
5. Applicator not calibrated A. Calibrate Applicator.
6. Bad Ultrasound Board A. Replace Ultrasound Board
100 WARN_US_UNPLUGGED 1. Applicator not plugged into system A. Plug Applicator into system.
Technician attempted to connector.
calibrate an ultrasound 2. Applicator connector not completely A. Make certain Applicator connector is
applicator but none is seated in system connector. completely seated in system
plugged into system. connector.
3. Applicator Cable disconnected from A. Connect Applicator Cable to
Applicator PCB. Applicator PCB
4. Applicator cable damaged. A. Replace with new Applicator.
101 WARN_US_APPL_ 1. Applicator not plugged into system A. Plug Applicator into system.
BECAME_UNPLUGGED connector.
Applicator came unplugged 2. Applicator connector not completely A. Make certain Applicator connector is
from system while seated in system connector. completely seated in system
administering therapy. connector.
3. Applicator Cable disconnected from A. Connect Applicator Cable to
Applicator PCB. Applicator PCB
4. Applicator cable damaged. A. Replace with new Applicator.
7
5- TROUBLESHOOTING Intelect®/Genisys® Ultrasound Therapy Systems
8
5- TROUBLESHOOTING Intelect®/Genisys® Ultrasound Therapy Systems
9
5- TROUBLESHOOTING Intelect®/Genisys® Ultrasound Therapy Systems
! WARNING
UNIT FAILING DIELECTRIC WITHSTAND AND/OR LEAKAGE
TESTS COULD INDICATE SERIOUS INTERNAL SYSTEM
PROBLEMS.
DO NOT PLACE UNIT BACK INTO SERVICE! SEND UNIT TO
FACTORY FOR REPAIR! DO NOT ATTEMPT TO REPAIR IN
THE FIELD!
FIGURE 5.5
12
5- TROUBLESHOOTING Intelect®/Genisys® Ultrasound Therapy Systems
! WARNING
USE ONLY DEGASSED WATER IN POWER METER FOR
TESTING ULTRASOUND APPLICATORS. USE OF OTHER
TYPES OF WATER WILL CAUSE FALSE TEST RESULTS.
SEE PAGE 10 FOR DEGASSED WATER RECIPES.
DO NOT AERATE WATER WHEN FILLING POWER METER.
2.Place an Applicator into the Power Meter retainer.
As per Ohmic Manual, make certain the Applicator’s
aluminum face is submerged in the degassed water
and centered directly over the S.S. cone.
3. “Zero” the Power Meter.
4. Turn the Ultrasound Therapy System on.
5. Press the button until “Continuous” is APPLICATOR OUTPUT SPECIFICATIONS
displayed in the Duty Cycle area of the display.
POWER SETTING
APPLICATOR SIZE OUTPUT RANGE
6. Press the button until “Watts” appears in the (WATTS)
intensity area of the display. 1 0.8 - 1.2
1 cm2
7. Increase Intensity for the applicator being tested 2 1.6 - 2.4
according to Figure 5.6 1 0.8 - 1.2
2 cm2 2 1.6 - 2.4
8. Press “Start”.
4 3.2 - 4.8
9. Compare Power Meter readings, displayed on 1 0.8 - 1.2
power meter for all the settings for each respective
2 1.6 - 2.4
applicator being tested to the values in Figure 5.6. 5 cm2
5 4.0 - 6.0
10. Press the button until “3.3 MHz” is diplayed in 10 8.0 - 12.0
the Frequency area of the display. Repeat test and 1 0.8 - 1.2
compare readings to Figure 5.6.
5 4.0 - 6.0
B. Ultrasound Applicator Output Test Results 10 cm2 10 8.0 - 12.0
1. Output ranges fall within the specified ranges as 15 12.0 - 18.0
listed in Figure 5.6. 20 16.0 - 24.0
a) Unit passed test. FIGURE 5.6
2. Readings fall outside specified ranges of Figure 5.6.
a)Possible bad Degassed Water in Power Meter.
b)Possible use of Power Meter other than Ohmic
Instruments UPM DT 10 or UPM DT 100
Ultrasound Power Meter.
c)Possible bad or out of calibration Applicator.
d)Possible bad Ultrasound Board.
13
5- TROUBLESHOOTING Intelect®/Genisys® Ultrasound Therapy Systems
! WARNING
USE ONLY DEGASSED WATER IN POWER METER FOR
TESTING ULTRASOUND APPLICATORS. USE OF OTHER
TYPES OF WATER WILL CAUSE FALSE TEST RESULTS. SEE
PAGE 10 FOR DEGASSED WATER RECIPES.
DO NOT AERATE WATER WHEN FILLING POWER METER.
14
6- REMOVAL & REPLACEMENT Intelect®/Genisys® Ultrasound Therapy Systems
! WARNING
UNPLUG THE UNIT FROM THE POWER SOURCE
AND/OR REMOVE BATTERY BEFORE ATTEMPTING
ANY REMOVAL OR REPLACEMENT PROCEDURES
TO PREVENT ELECTRICAL SHOCK.
A. Tools & Equipment Required
1. #1 Phillips Screwdriver
2. Flat Blade Screwdriver FIGURE 6.1
6.2. Top & Bottom of Unit
A. Removing Top from Bottom
REMOVE
1. Place unit face down on a grounded work surface. FAN GRILL
NOTE:
Use a clean, soft cloth or cushion underneath the display lens
to prevent scratching or damage to the lens.
2. Remove the plynth from the bottom of the system.
See Figure 6.1. REMOVE BATTERY
COVER AND BATTERY
3. Remove the Mains Power Cord from the system.
See Figure 6.1.
FIGURE 6.2
4. Remove Battery Cover with the Flat Blade
Screwdriver and remove Battery from the system if
equipped. See Figure 6.2.
5. Remove the Rear Fan Grill from the system to gain
access to the two rear mounting screws. See Figure
6.2.
REMOVE
6. Remove the two front feet from the system to gain FOUR SCREWS
access to the two front screws. See Figure 6.3.
7. Remove the four mounting screws from the bottom
housing using the #1 Phillips Screwdriver. See REMOVE
Figure 6.3. FRONT FEET
8. Carefully separate the upper and lower housings of
the system. It may be necessary to pry the housings FIGURE 6.3
apart at the Applicator Hook. See Figure 6.4.
9. Disconnect the Power Supply and Fan Harnesses CAREFULLY PRY APART
from the Control PCB of the system. See Figure 6.4. HERE IF NECESSARY
! WARNING
UNPLUG THE UNIT FROM THE POWER SOURCE
BEFORE ATTEMPTING ANY REMOVAL OR
REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK.
A. Fan Removal
1. Separate Top from Bottom. Refer to 6.2, part A. FIGURE 6.5
2. With #1 Phillips screwdriver, remove the 2 screws
securing the Fan. See Figure 6.5.
FAN
3. Remove Fan from the unit. BAFFLE
4. Remove Fan Baffle from fan. See Figure 6.5.
B. Installing Fan
1. Route Fan Harness underneath fan as shown in
Figure 6.6.
2. Install Fan Baffle onto fan as shown in Figure 6.6.
ROUTE
3. Place new Fan, Part Number 27266, into system HARNESS
base. AS SHOWN
4. Make certain the Fan Baffle sits onto System FIGURE 6.6
Housing Baffle Tabs as shown in Figure 6.7.
5. Secure fan to housing with the 2 mounting screws.
See Figure 6.8. BAFFLE
TABS
NOTE:
Do not overtighten the screws. Overtightening will damage the
threads of the brass inserts.
FIGURE 6.7
SECURELY TIGHTEN
MOUNTING SCREWS
FIGURE 6.8
16
6- REMOVAL & REPLACEMENT Intelect®/Genisys® Ultrasound Therapy Systems
! WARNING
UNPLUG THE UNIT FROM THE POWER SOURCE
BEFORE ATTEMPTING ANY REMOVAL OR
REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK.
REMOVE SCREWS
FIGURE 6.12
17
6- REMOVAL & REPLACEMENT Intelect®/Genisys® Ultrasound Therapy Systems
! WARNING
UNPLUG THE UNIT FROM THE POWER SOURCE
BEFORE ATTEMPTING ANY REMOVAL OR
REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK.
FIGURE 6.16
18
6- REMOVAL & REPLACEMENT Intelect®/Genisys® Ultrasound Therapy Systems
6.6 LCD Display Removal & Replacement RELEASE RETAINING
2 SCREWS BRACKET TABS
FIGURE 6.19
19
6- REMOVAL & REPLACEMENT Intelect®/Genisys® Ultrasound Therapy Systems
! WARNING
UNPLUG THE UNIT FROM THE POWER SOURCE
BEFORE ATTEMPTING ANY REMOVAL OR
REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK.
CONTRAST KNOB
FIGURE 6.22
20
7- ULTRASOUND APPLICATOR CALIBRATION Intelect®/Genisys® Ultrasound Therapy Systems
7.1 General
A. Tools and Equipment Required
1. Ultrasound Therapy System being serviced.
2. All Ultrasound Applicators for the unit being serviced.
3. Ohmic Instruments UPM DT 10 or UPM DT 100
Ultrasound Power Meter, set to “watts”.
4. Degassed Water. Refer to page 10 for Degassed
Water Recipes.
! WARNING
USE ONLY DEGASSED WATER IN POWER METER FOR
CALIBRATING ULTRASOUND APPLICATORS.
USE OF OTHER TYPES OF WATER WILL CAUSE FALSE FIGURE 7.1
READINGS AND BAD TEST RESULTS.
SEE PAGE 10 FOR DEGASSED WATER RECIPES.
USE OF OTHER BRANDS OR TYPES OF TOOLS,
EQUIPMENT, FIXTURES, MATERIALS AND SUPPLIES
OTHER THAN THOSE SPECIFICALLY LISTED IN “A. Tools
and Equipment Required” ABOVE WILL GIVE BAD TEST
AND CALIBRATION RESULTS.
IF PROPER EQUIPMENT IS NOT AVAILABLE OR CAN NOT
BE OBTAINED, SEND THE ULTRASOUND APPLICATORS
TO THE FACTORY FOR CALIBRATION.
SET METER
TO WATTS
FIGURE 7.3
21
8- PARTS Intelect®/Genisys® Ultrasound Therapy Systems
Fan Connector
3&4
12
11
(See Pages26 - 27) Power Supply Connector
5 5
10
Battery Connector
6
7 (Optional)
22
8- PARTS Intelect®/Genisys® Ultrasound Therapy Systems
Ultrasound Assembly
ITEM PART DESCRIPTION QTY
NUMBER NUMBER REQ’D
1 27402 Top Assembly (Grey)- See page 24 for breakdown 1
2 27247 Top Assembly (Blue)- See Page 24 for breakdown 1
3 27361 Rear Vent (Grey) 1
4 27363 Rear Vent (Blue) 1
5 27142 M3 x 6mm S.S. Pan Head Screw 4
6 27365 Right Foot 1
7 27267 Battery Pack (Optional) 1
8 27410 Battery Compartment Cover Assembly 1
9 27253 Plynth 1
10 27363 Left Foot 1
11 27409 Bottom Assembly- See Pages 26-27 for breakdown 1
12 27373 Baffle 1
23
8- PARTS Intelect®/Genisys® Ultrasound Therapy Systems
1 & 1A
2 & 2A
3 & 3A
4
(See Page 25)
24
8- PARTS Intelect®/Genisys® Ultrasound Therapy Systems
25
8- PARTS Intelect®/Genisys® Ultrasound Therapy Systems
26
8- PARTS Intelect®/Genisys® Ultrasound Therapy Systems
27
9- SCHEMATICS Intelect®/Genisys® Ultrasound Therapy Systems
SCL_APPL
SDA_APPL
MCLR
1
1
1
C2 C1 R2
R6 R1 10nF 10nF
1K
10K 1K R4
42.2K 1%
F1
2
2
U1
VCC_APPL 1 8
1
1
1 2
VDD VSS
1
2 7
GP5/T1CKI/OSC1/CLKIN GP0/AN0/CIN+/ICSPDAT
3 6
GP4/AN3/T1G/OSC2/CLKOUT GP1/AN1/CIN-//VREF/ICSPCLK
4 5 C5
C3 C6 GP3/MCLR/VPP GP2/AN2/T0CKI/INT/COUT RT1
0.1uF Next to Pin 1
0.1uF 1.0uF/16V 100K Thermal connect
2
PIC12F675-II /SN L2
2
2
2
DGND_APPL
NOT Case
Connected 220uH
3
LED1
2
R5 R7
1 2 2 1 1 2 1
Q4
220
1
10K
MMBT3904
2
R3
C4 10K
10nF
2
1
D1
BAS16_NL
1 2
1- N/C K1
3
2,4- US+ 8
3- MCLR 1
5- +5V 5 1 50PAD
6,8- US-
1
6 PAD4
7- I C2 DATA 7 1 50PAD
1
9- I C2 CLOCK PAD3
pplicator
10- Applicator Body 4 1 50PAD
C8 1
3 PAD2
2 1 50PAD
1
C7 PAD1
10nF JP1
T1 G6K-2
2F-YY5VDC
US+ 1 8
1 2
10nF
3 4 PIN1
C9
5 6
330pF/200V 2 7 1
7 8 1
9 10
3 6
US- 4 5
1
1
PIN2 PIN3
Case Connected
FB1
HZ1206 L3
100uH
1
1
1 2 GND
VCC
8
7
6
J1
C11 C12 KMDG-8
8S-BBS
7
4.7uF/25V 0.1uF
8
6
HEAD_CLOCK
2
4
4 R156 R151
2
5 3 0 0
1
5 3
GND
R157
1
1
R152
2 1
S1
S2
S3
2 1
C1 GND
0 0
2
2
T1
1
100pF/100V US_GND
2
4 5 R7 R8
S1
S2
S3
HEAD_DATA
1
21
21
0.1uF/50V TX2-2
24V 11.5uH/2A 3300pF/250V TX2-2
24V
1
1
1
3 6 3 2 1 1 2 3
1
DRAIN
1
2
2 7 4 4
C4 C98 5 5 C13 C14
2
2
2
47 0.1uF/50V 10 1000pF/250V 10
1
2
2
1 8 9 2 1 1 2 9
2
C2 8 8
100pF/100V GND 1 L2 1
1
VIN IRF
ENG00787 12 2.8uH/2A 12 GND
2
FB9
1
VCC L4
R5 100uH_NL
1
D3 0.56 TP1 1 2
1
1
2
1
C15 4.7uF/25V FB10 US Ground
1 2
U2 0.1uF C16 U3 VCC C20
R19
1
3
2
1 5 4.7uF/25V 5 1 0.1uF_NL
VDD VCC VCC VDD 1.5K_NL
2
R112
2
C18
2N7002
2
2
0.1uF HZ1206 0
Q8
4
4
R22
12
2 GND Q4 Q7 2 1 GND 1 2
GND GND HEAD_DETECT
1
1_3MHZ
2
2
IRLR110 IRLR110
1
1
1.5K_NL
2
2
GND R9 GND GND
3 4 1 2 1 1 1 2 4 3 GND US_GND R20
1
1
IN OUT OUT IN 300K_NL R23
R13 C19 C23
0_NL
3
3
TPS2819DBV 2.2 R12 TPS2819DBV 10K 0.1uF_NL
0.1uF C22
1
2
2.2
2 2
2
1
R10 R11 2 1
VCC
1.0M 1.0M
R17
U4
7
2
GND GND 0.1uF GND
14.0K
3 +
12
GND 6
VRF
2
-
R18
LMH6642MA
1.00K C21
4
2 1
-5
5V
1
GND 0.1uF
GND 1 2
R21
1.00K
DUTY_CYCLE
CLK_REF U1A
C17
13
12 2 1 2
1
1
1
1
620
2
R25 R26
U1B
2.2K 2.2K U5 3
4 6
2
1 16 5
I2C_CLOCK SCL CLKN
2 15 2 1
I2C_DATA SDA CLKP 74HC27D
3 13
ADDR0 FBK
7 12 R28
1
ADDR1 REF
5 10 620
XIN EXTLF
6 9
XOUT LOCK/IPRG
8 14 C28
VDD VDD
4 11
VSS VSS
2
GND 47pF
FS6131-0
01
1
1
GND GND GND VCC
C29
TP14
C25 C27 1 2
0.1uF 0.1uF TESTPOINT
14
2
2
0.1uF
1
1
11
VCC
U1C
HZ1206 C26
74HC27D
7
4.7uF/25V
GND
2
GND
Ultrasound Board
Schematic 1 of 3 (AMP)
31
9- SCHEMATICS Intelect®/Genisys® Ultrasound Therapy Systems
Ultrasound Board
Schematic 2 of 3 (Ports)
32
9- SCHEMATICS Intelect®/Genisys® Ultrasound Therapy Systems
Ultrasound Board
Schematic 3 of 3 (V Supply)
33
10- WARRANTY Intelect®/Genisys® Ultrasound Therapy Systems
Chattanooga Group ("Company") warrants that the Ultrasound Therapy System("Product") is free of defects in material and
workmanship. This warranty shall remain in effect for two years (24 months) from the date of original consumer purchase. If this
Product fails to function during the two years warranty period due to a defect in material or workmanship, Company or the selling
dealer will repair or replace this Product without charge within a period of thirty (30) days from the date on which the Product is
returned to the Company or the dealer.
All repairs to the Product must be performed by a service center certified by the Company. Any modifications or repairs performed
by unauthorized centers or groups will void this warranty.
The warranty period for applicators is one year (12 months).
This Warranty Does Not Cover:
Replacement parts or labor furnished by anyone other than the Company, the selling dealer or a certified Company service
technician.
Defects or damage caused by labor furnished by someone other than Company, the selling dealer or a certified Company service
technician.
Any malfunction or failure in the Product caused by product misuse, including, but not limited to, the failure to provide reasonable
and necessary maintenance or any use that is inconsistent with the Product User's Manual.
COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES.
Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may
not apply to you.
To Obtain Service From Company or the selling dealer under this warranty:
1. A written claim must be made within the warranty period to the Company or the selling dealer. Written claims made to the
Company should be
sent to:
4717 Adams Road
Hixson, TN 37343 USA
Telephone: +1 423 870-7200
FAX: +1 423 870-2046
and
2. The Product must be returned to the Company or the selling dealer by the owner.
This warranty gives you specific legal rights and you may also have other rights which vary from state to state or location to
location.
The Company does not authorize any person or representative to create for it any other obligation or liability in connection with the
sale of the Product. Any representation or agreement not contained in the warranty shall be void and of no effect.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING ANY WARRANTY OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
34
More Trusted Products from Chattanooga Group
®
Adapta Intelect ®
Massage Tables & Chairs Ultrasound and Electrotherapy Products
™
A.E.R. Boot Magneciser™
Auto Edema Reduction Boot Exercise Equipment
™
A.E.R. Compression System Measurement Instruments
Hot and Cold Compression Therapy Dynamometers, Goniometers, ect.
boo-boo pac™ Myossage®
Child Size Bear-Shaped Cold Pack Massage Lotion
®
Cambion Nylatex®
Shock Dampening Foot Care Products Elastic Wraps
®
Can-do OptiFlex®
Exercise Bands, Tubes and Balls Continuous Passive Motion
Carpal-Trac™ Opti-Ice™
Carpal Traction Accessory Cold Therapy System
Cervical Traction System Para-Care®
Clinical Cervical Traction Paraffin Wax Unit
®
ColPaC Pillow Perfect™
Chilling Units and Reusable Cold Therapy Products Cervical Pillow Line
Conductor Gel™ Pivotal Therapy System™
Highly Conductive Ultrasound Gel Orthotics for the Spine
®
Contracture Products PresSsion
Contracture Management Orthotic Products Intermittent Compression
™
DURA-STICK® Electrodes Pron Pillo
Self-Adhesive Electrodes Positioning Pillow
EMG Retrainer® and EMG Retrainer® IR SensaFlex™
Dual Channel Surface EMG Hot/Cold Pack
Ergostyle® SPORT-PAC™
Chiropractic Tables Soccer Ball Shaped Cold Pack
Flexi-PAC® I and II Therma-Wrap™
Reusable Hot
®
and Cold Compresses Hot and Cold Compression
Fluido DHT Triton®
Dry Heat Whirlpool Therapy Units Treatment and Traction Equipment
Gel Medex™ Tru-Trac®
Gel Mattress Overlay Traction Equipment
®
Hydrocollator Wellness 1st™
Heating Units and HotPacs™ Back Support