Академический Документы
Профессиональный Документы
Культура Документы
Operation Manual
NDT X-Ray equipment belongs to precision instrument, for your successful operation and safety,
1. Operating in outdoor field, 2mm thick lead plate can be employed as protective shield. If no any
suitable radiation protection, please make sure that there will be nobody within 20m around area from
2. X-Ray machine should be at vertical direction while carrying and transportation because x ray tube
3. Operators must be trained and qualified , operation should be strict accordance with the manual.
Manufacturer don't take any responsibility for abuse use or force work caused damage.
4. X-Ray machine is safe work at 0.35-0.5 Mpa, prohibit operating less than 0.35 Mpa.
5. X-Ray tube inside should be treat most careful because pressure inside. Wrong method to open the
generator will damage x ray tube or HV coil, even make explosion of x ray tube.
6. Unauthorized installation, removal, and replacement of accessory is forbidden. For safety, operator
should contact manufacturer before repair and disassemble x ray generator , so that no damage or
7. Problem not caused by the quality of the equipment itself (such as drop, crash, drop into the water,
I、 General……………………………………………… 1
VI、 Maintenance………………………………………… 8
VIII、Drawings…………………………………………… 9
ZHONG YI NDT CO., LTD
I. General
The frequency-converted and gas-insulated portable X-ray NDT equipment with ZY–3000 controller, is one of
upgrading generation of the products by ZHONG YI NDT CO.,LTD (Dandong Zhongyi Electronic Co., Ltd).
Besides of its advantages of small size, light weight, easy to use, simple operation and self-troubleshooting, this new
type of controller has the excellent performance of strongly interference-free capability, good working stability and
operating reliability. With the popularization and spreading application of this new type NDT X-ray equipment all
over the country, it has been recognized and welcomed by the customers from different industries
This portable X-ray NDT equipment can be used in the fields of ship-building, mechanical manufacturing, aviation,
railroad, pressure vessel, boiler, petroleum, chemical and national defense, performing the non-destructive testing,
The X-ray NDT equipment is used to show the internal discontinuity or faults of materials, products or welding from
a radiographic image on a film, according to the principle of the different absorption of the X-ray for different
material depending on the different thickness of the materials. We can evaluate the quality of materials and products
based upon the characters, size and position of the discontinuity or faults.
In this series of products, the X-ray generator is furnished with SF6 gas as an insulation, electric fan for forced
cooling, the controller is equipped with the advanced microprocessor controlling system, SCR modules for voltage
regulation, converting modules to control the primary and secondary side of the SCR as well as the advanced
electronic circuits of voltage & current stabilization and interference-free .Thus, the performance of the complete
1
ZHONG YI NDT CO., LTD
Ⅱ. Technical Data
Classify Model Tube voltage Capacity Focus Size Max. penetrate
(KV) (KW) (mm) A3 steel (mm)
2
ZHONG YI NDT CO., LTD
1. Controller
The controller is shaped like a suitcase. All operative buttons are furnished on the front panel while the cable socket,
main switch and earthing terminal are located inside the control box on the right side of it.
A. Control board;
B. Main board;
C. Display board;
D. LC filter;
G. Cooling fan;
2. X-ray Generator
The X-ray generator is the module type. The X-ray tube, HV transformer and insulated gas (SF6) are encapsulated in
an aluminum shell of sylindrical shape. The fan and heat radiator are mounted on one end of the X-ray generator.
A. X-ray tube;
B. HV transformer;
C. Temperature relay;
3
ZHONG YI NDT CO., LTD
voltage turns to the smooth dc voltage by means of the LC filter. This voltage converts to the
frequency-adjustable square-wave pulse with single flow direction, becoming the electrical supply of the
The mA stabilizing unit will change the frequency of the single-phase pulse against the voltage increase or
decrease of the X-ray tube filament, to ensure the stable current in the X-ray tube.
The kV regulating unit can continuously control the voltage up and down, to meet with the different
The interference-free circuit improves the shock- resistance capability to protect the control system from
2. The X-ray generator is equipped with the electric shock absorber. The anode of X-ray tube is connected to
the ground and an over temperature protection device is furnished.. Once the over-temperature occurs, the
controller is able to shut off the HV automatically to prevent the X-ray tube being damaged from the over
temperature.
3. The voltage fluctuation of power supply can be regulated automatically by the controller itself, ensuring to
get a stable X-ray. The exposure time and operation status of the complete equipment is controlled by the
4
ZHONG YI NDT CO., LTD
1st After unpacking, please check all items in the package according to the packing list, to see whether there is any
damage, shortage or any incorrect parts packed by mistake.
2nd Confirm that the gas pressure meter on the X-ray generator points the range of 0.40 ~ 0.45 Mpa (Very small
pressure variation caused by different environmental temperature ).
3rd Get the controller on a working desk, switch off the controller power, connect the power cable. Make sure that
the main power supply should be in line with the specifications.
4th Connect one end of the grounding wire to the earthing terminal of controller while another end of the wire links
to a humid ground or specialized earthing point.
5th Link the controller with X-ray generator by the connecting cable and ensure this connection perfect.
6th The protective shield should be equipped when operating this equipment. In case of operating in outdoor field,
2mm thick lead plate can be employed as protective shield. If there is no any suitable radiation protection, please
make sure that there will be no any body within 20m around area from the focal spot of the x-ray tube when
radiography exposure.
2. Operation
1st Turn on the power-supply switch, the Digital screen shows different models:
Buzzer will make sound with green lights on when equipment is ready, meanwhile the controller shows origin kv
value according to different models machine. Controller will show last exposure setting parameters next use.
Origin Parameter Value:
Model KV shows Time-delay shows
1005 040 05.0
1605 060 05.0
2005 100 05.0
2505 150 05.0
3005 200 05.0
3505 250 05.0
5
ZHONG YI NDT CO., LTD
2nd When the machine ready, system will automatic release the key of the button. Now setting the
different parameter .
3rd KV setting :
KV setting should be done in the standby mode.
Press the SET button once, KV light flashing and adjust the KV value as requirement with the decrease and increase
button, each stepping rate is 2KV.
After setting KV value, press the SET button , meanwhile TIME light start flashing, press SET button again for stop
the flashing.
Start-up HV and machine training can be proceed after setting.
6
ZHONG YI NDT CO., LTD
No need human operation during training,different models machine have different training times.
Model Time
1005 1
1605 3
2005 3
2505 3
3005 3
3505 3
The machine can automatic training to specific kv value. Eg: 2505 machine , setting kv is 184V, the machine will
automatic stop when training value reached 184V.
10th Notice
A. Frequently check the gas pressure in the X-ray generator. If it is lower than 0.35Mpa, the operation is prohibited.
B. For safely operation, verify the earth terminal on the controller to be connected with the ground reliably
C. Ensure the cable is under good contact condition and electric socket is clean.
D. The ratio between working and resting time of the equipment should strictly follow by 1:1. The main power
should not be cut off during resting time, so the cooling fan keeping rotation continuously.
E. The circuit may get protection when large voltage fluctuation takes place. This is not malfunction. The machine
can rework when voltage gets stable.
7
ZHONG YI NDT CO., LTD
VI. Maintenance
When failure occurs, the operation should be stopped and the failure reason should be found out.
1. When powering on, the cooling fan does not rotate and no indication on the Digitron:
2. The displayer shows the failure code with the meaning as below:
A. “LA” No mA warning;
B. “HA” over mA warning;
C. “PU” over voltage warning;
D. “PT” over temperature warning;
3. When pressing the ON button, power switch cut off or sometimes sounds unusual:
8
ZHONG YI NDT CO., LTD
Ⅷ. Drawings
Picture 1 Picture 2
Picture 3 Picture 4
9
ZHONG YI NDT CO., LTD
10
ZHONG YI NDT CO., LTD
11
ZHONG YI NDT CO., LTD
12
ZHONG YI NDT CO., LTD
13
ZHONG YI NDT CO., LTD
Quality Warranty
1. 1 year free guarantee. If the product fails or became inoperative without evidence of misuse, abuse, improper
installation, or unauthorized repair, we will provide the spares for repair free .
2. Engineers will online assist for the repairing work , according to fault description and picture sent by users.
3. For difficult fault, users can send back the machine back to China for repair. Please pack the machine properly so
that it will not be damaged during transportation, meanwhile accompany by a fully completed Field Failure Report.
Engineer Provide Services in Working Site for training Product application method, maintenance work.
NOTE: For the users who visit factory for learning the technical:
F. All the accommodation and transportation expenses in China should bear by visitors .Factory will provide
This manual is for reference only, if there is any change, without prior notice.
14