HEALTHCARE Chapter 5

Imaging Services Software Options


Document No: DD+DIS107.10E

NX 2.0.8300 / 3.0.8300
Type 4406/303

Clean Installation Procedure

► Purpose of this Document

This document describes the installation of an NX 2.0.8300 / 3.0.8300 system starting


from scratch.

► Document History

Edition. Release Changes


Revision Date compared to previous Version 1.0:
1.1 09-2010  Added remark to create partition D: on disk 0.
Refer to section 3.5
 Added step to select proper XRDI version if the
corresponding license is available.
Refer to section 5.2.

► Referenced Documents

Depending on the system, following documentation is needed on site:

Document Title
DD+DIS105.10E NX 2.0.8300 / 3.0.8300 Service Manual
DD+DIS116.10M NX 2.0.8300 / 3.0.8300 Spare Parts List
DD+DIS012.06E Licensing Service Documentation
DD+DIS124.09E ID Scanner Service Manual

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1
09-2010 printed in Germany Document Node ID: 29270600
sw_05_software-options_e_template_v05
Agfa Company Confidential Copyright © 2010 Agfa HealthCare N.V.
DD+DIS107.10E Software Options - Clean Installation Procedure

► Manufacturer
Agfa HealthCare N.V.

Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D – 81539 München
Germany
Copyright  2010 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.

WARNING:
Improper operation or service activities may cause damage or injuries.

INSTRUCTION:
1. Read the “Generic Safety Directions” document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
2. Strictly observe all safety directions within the “Generic Safety Directions” and on
the product.

NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type ‘Order List’ in the GSO library.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 2


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

LIST OF CONTENTS

1 OVERVIEW OF CLEAN INSTALL PROCEDURE....................................................................5


2 PREPARATION ........................................................................................................................6
2.1 Downloading / ordering the Clean Install Kit.............................................................................6

2.2 Obtaining additionally needed Tools, Documentation and Software ........................................6

2.2.1 Tools for a clean Installation Procedure ...................................................................................6

2.2.2 Documentation for a clean Installation Procedure ....................................................................7

2.3 Software for a clean Installation Procedure ..............................................................................7

2.3.1 TWAIN Digitizer Device Software (ASAP) ................................................................................7

2.3.2 Latest Hotfix ..............................................................................................................................7

2.3.3 Configuration File for the NX Workstation ................................................................................8

2.4 Obtaining Application License File (if needed) .........................................................................8

3 CLEAN INSTALL PROCEDURE ..............................................................................................8


3.1 Prerequisites .............................................................................................................................8

3.2 Reset a previous RAID Configuration.....................................................................................10

3.3 Adjust BIOS Settings ..............................................................................................................12

3.3.1 Adjusting BIOS Settings for DELL PCs ..................................................................................12

3.3.2 Adjusting BIOS Settings for HP PCs ......................................................................................13

3.4 Installing the Operating System (OS) .....................................................................................14

3.4.1 Starting the OS Installation .....................................................................................................14

3.4.2 Automatic Installation of Operating System............................................................................16

3.4.3 Finishing OS Installation .........................................................................................................17

3.5 Performing additional Partitioning ...........................................................................................18

3.6 Configuring Basic Operating System Settings........................................................................22

3.7 Installing the latest available Hotfixes.....................................................................................25

3.8 Configuring Network ...............................................................................................................26

3.8.1 Changing TCP/IP Settings of Local Area Connection (LAN)..................................................26

3.8.2 Configuring the wireless Network Connection for the wireless ICU Workflow .......................28

3.8.3 Changing the Hostname and/or Workgroup/Domain of the Operating System......................28

3.9 Installing NX Software and Documentation ............................................................................30

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 3


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

3.9.1 Installing NX Software.............................................................................................................31

3.9.2 Installing NX Software Update (SU) .......................................................................................32

3.9.3 Installing NX Online Help ........................................................................................................32

3.10 Installing an Uninterruptible Power Supply (UPS) ..................................................................33

3.11 Installing TWAIN Digitizer Software (ASAP)...........................................................................33

3.12 Installing FUSE Software ........................................................................................................34

4 OPERATING SYSTEM CONFIGURATION............................................................................35


5 LOADING LICENSES AND ACTIVATING THE INSTALLATION...........................................35
5.1 Loading Licenses (ALF) ..........................................................................................................35

5.2 Activation.................................................................................................................................36

6 INSTALLING ADDITIONAL DRIVERS ...................................................................................39


6.1 Installing Monitor Software and Calibration ............................................................................39

6.2 Upgrading ID Tablet Firmware................................................................................................40

6.3 Installing the ID Scanner Drivers ............................................................................................40

7 NX CONFIGURATION VIA CONFIGURATION TOOL...........................................................41


8 COMPLETING INSTALLATION .............................................................................................41
8.1 Activating RAID (Redundant Array of Independent Disks) .....................................................41

8.2 Adjusting Anti-Virus Settings...................................................................................................45

8.3 Backup of Configuration Data .................................................................................................45

8.4 Performing the NX Acceptance Test ......................................................................................46

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 4


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

1 Overview of Clean Install Procedure

Overview of required steps:


# Step Section Duration*
PREPARATION
1. Download / order the clean installation kit. 2.1 -
2. Obtain additional tools, documentation and software. 2.2 -
3. Obtain the Application License File (ALF) if not available anymore. 2.3 -
CLEAN INSTALL PROCEDURE
4. Check prerequisites. 3.1 -
5. Reset a previous RAID Configuration. 3.2 10 min.
6. Adjust BIOS settings. 3.3 10 min.
7. Install operating system. 3.4 60 min.
8. Perform additional partitioning. 3.5 5 min.
9. Configure basic operating system settings. 3.6 5 min.
10. Install hotfixes (I). 3.7 50 min.
11. Configure network. 3.8 10 min.
12. Install application and documentation. 3.9 30 min.
13. Install UPS (optional). 3.10 5 min.
14. Install TWAIN digitizer software (if applicable) 3.11 10 min.
15. Install FUSE software. 3.12 10 min.
OPERATING SYSTEM CONFIGURATION
16. Edit security settings according to hospital guidelines. 4 10 min.
17. Create and configure users. 4 15 min.
LOADING LICENSES AND ACTIVATING THE INSTALLATION
18. Load Licenses 5.1 10 min.
19. Activate NX 5.2 15 min.
INSTALLING ADDITIONAL DRIVERS
20. Install monitor software and calibrate monitor. 6.1 20 min.
21. Upgrade ID Tablet firmware (if applicable) 6.2 10 min.
22. Install ID Scanner Driver (if applicable) 6.3 10 min.
NX CONFIGURATION VIA CONFIGURATION TOOL
23. Enable or disable active licenses if needed. 7 10 min.
24. Configure users. 7 10 min.
25. Import Configuration settings. 7 20 min.
26. Activate the configuration 7 2 min.
COMPLETING INSTALLATION
27. Activate RAID 8.1 130 min.
28. Adjust Anti-virus settings (if applicable) 8.2 20 min.
29. Create backup of configuration data. 8.3 10 min.
30. Perform Acceptance Test. 8.4 30 min.
Overall required time: Approximately 8 hours

* Reference value. Required time depends on the complexity of the situation on site.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 5


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

2 Preparation

2.1 Downloading / ordering the Clean Install Kit

NOTE:
Starting with NX 2.0 staged on DELL Optiplex 755 the Clean Install Material is
delivered with the system.

 The Clean Install material is available as spare part. See NX 2.0.8300 / 3.0.8300
Service Manual chapter 13 Spare Parts List DD+DIS116.10M (Intranet Link /
Extranet Link).
 The Clean Install material can also be downloaded from the FTP-Server:
(1) Open the FTP Access document on the GSO Library (Intranet Link /
Extranet Link).
(2) Connect to the FTP server according to the instructions given in the FTP
Access document.
(3) Download the following ISO files and burn it as disk image on DVD:
 NX_X.8300_DVD1_VX.Y.iso
 NX_X.8300_DVD2_VX.Y.iso

NOTE
For a specification of the contents of the Starterkit DVDs and the DVD versions that are
compatible with a specific NX profile refer to NX 2.0.8300 / 3.0.8300 Service Manual
chapter 1 SW/HW Compatibility Chart DD+DIS105.10E (Intranet Link / Extranet Link).

2.2 Obtaining additionally needed Tools, Documentation and Software

2.2.1 Tools for a clean Installation Procedure

Take the following tools on site:


 A standard serial cable to upgrade the firmware of the ID Tablet.
If the ID Tablet is connected via USB, make sure to take a serial cable on site.
 A portable medium (USB Stick or CD-R/RW).

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 6


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

2.2.2 Documentation for a clean Installation Procedure

Take the following documentation on site:


 NX 2.0.8300 / 3.0.8300 Service Manual, DD+DIS105.10E
(Intranet Link / Extranet Link)
 If the NX workstation is part of a Mammography or DR Solution: The corresponding
solution service manual.

2.3 Software for a clean Installation Procedure

2.3.1 TWAIN Digitizer Device Software (ASAP)

If the NX Workstation is connected to a TWAIN digitizer (e.g. DX-S, CR 30-X or


DX-G ASAP *), the compatible device software has to be downloaded from
Mednet GSO Library and installed on the workstation.
* DX-G is a DICOM digitizer as of SW version  NIM_2xxx

(1) To find the compatible version number(s) refer to NX 2.0.8300 / 3.0.8300 Service
Manual chapter 1 SW/HW Compatibility Chart DD+DIS105.10E
(Intranet Link / Extranet Link).
(2) Go to the Mednet GSO Library (Intranet Link / Extranet Link) to download the
device software packages:
GSO Library  Computed Radiography  CR Digitizers  <Digitizer name> 
Freeware

2.3.2 Latest Hotfix

During Clean Installation a Hotfix version is installed on the system that may not be the
latest available Hotfix version.
Run the FUSE updater on the Service PC to receive the latest available Hotfix version.
NOTE:
For instructions to obtain and install the Fast Update Service (FUSE)
software refer to the FUSE Service Manual FUSE - Fast Update Service
(Automatic Software Update) DD+DIS134.08E
(Intranet Link / Extranet Link).
For external partners: Please contact your local Agfa representative.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 7


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

2.3.3 Configuration File for the NX Workstation

If the configuration for the specific site is lost:


Organize the default configuration file of the NX workstation of the region or country
(to be setup by the country during the initial NX workstation installations).

2.4 Obtaining Application License File (if needed)

IMPORTANT:
When performing a clean installation, the Application License File (ALF) should
normally still be available and can be used for the same software version after the
clean install.
This is not the case if the ALF got lost after a system crash. In this case fetch it from
the Electronic License Management System (ELMS).

INSTRUCTION:
For a detailed description of the procedures to obtain a License File (ALF) refer to the
Licensing Service Manual (DD+DIS012.06E) on MedNet GSO Library
(Intranet Link / Extranet Link).

3 Clean Install Procedure

3.1 Prerequisites

ACTION:
Check the following items before starting the installation:

 Check hardware to see if it is supported for NX Operating System.


Install the NX only on computers that are on the supported Hardware list.
The list of supported hardware is specified in NX 2.0.8300 / 3.0.8300 Service
Manual, chapter 1 SW/HW Compatibility Chart DD+DIS105.10E
(Intranet Link / Extranet Link).

 Use the correct set of the Installation DVDs:


o NX Starterkit DVD 1
o NX Starterkit DVD 2

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 8


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

 Identify Windows License Key:

For all installations provide the Windows license key for the operating system. This
is often a ‘hologrammed’ sticker attached to the system itself by the manufacturer.
The key itself is referred to as the ‘Product ID’ and
has the form 5 characters in 5 blocks, for example:

HVYF3-T28PW-HWGKB-33J4W-B4VPM (Example)

 Make sure a network is connected:

When installing, the NX PC must have a network connection.


The minimum is a connection to a hub so the system can recognize a network
connection and another PC on the network.

The system can be connected to a larger network like the hospital or company
network. In this case, if other NX systems are installed at the same time, an IP
conflict can occur.

 Optionally, connect a QWERTY (US international) keyboard:

When logged on as administrator MS Windows will always be set to “English”.


Therefore it is easier working with a QWERTY keyboard (US international) attached
when typing something.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 9


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

3.2 Reset a previous RAID Configuration

REQUIRED TIME:
Approximately 10 minutes

NOTE:
These steps are only required if RAID was already activated before on the installed
hard disks.
If the installed hard disks have not been using RAID before, continue with section
3.3 "Adjust BIOS Settings".

(1) Press <CTRL-i> during system boot, when the screen is displaying the Intel
Matrix Storage Manager RAID overview.
On a HP PC it is necessary to first press <Enter> on the message:
Press any key for option ROM messages
(2) Choose option 3 to Reset Disks to Non-RAID in the Main Menu:

Figure 1

(3) Select both Member Disks with <Space> and press <Enter> to complete.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 10


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(4) Press <Y> to confirm the Warning dialog:

Figure 2

(5) Reboot and enter the BIOS-setup:


 On a Dell PC: Press <F2>.
 On a HP PC: Press <F10>.
(6) Set the drive access to IDE (Non-RAID):
 On a Dell PC: Set SATA operation to be to: RAID/Autodetect/ATA
 On a HP PC: Set SATA operation to: IDE
(7) Insert the NX Starterkit DVD 2.
(8) Reboot the PC and
 On a DELL PC: Press <F12> during the boot sequence (custom boot).
 On a HP PC: Press <F9> during the boot sequence (custom boot).
NOTE:
After changing the BIOS settings on HP desktop PCs, an additional
screen is displayed that informs about updated configuration options.
Confirm the changes by pressing <F1>.

(9) Select to boot from CD-ROM and press <Enter>.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 11


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

NOTE:
Be aware that the installation has to be cancelled in the next step!

(10) Cancel the installation by pressing <N>:


+-------------------------------------------------------------+
! AGFA NX System !
! Multi Language !
+-------------------------------------------------------------+
! !
! New Installation. !
! !
! (Y) Installation !
! (N) Exit !
+-------------------------------------------------------------+

Figure 3

(11) Access the CD-ROM drive by typing into the DOS prompt: K:\ or I:\
Remark: Depending on the SATA (Serial Advanced Technology Attachment)
emulation the CD-ROM will now be available on either K: or I:
(12) Type: Gdisk.exe 2 /i /DEL /ALL /Y
Remark: This will erase all parititons on the 2nd harddrive.
(13) Continue with section 3.4 Installing the Operating System (OS).

3.3 Adjust BIOS Settings

REQUIRED TIME:
Approximately 10 minutes

NOTE:
To setup the BIOS settings on DELL PCs refer to section 3.3.1.
For HP PCs refer to section 3.3.2.

3.3.1 Adjusting BIOS Settings for DELL PCs

(1) Enter the BIOS setup by pressing <F2> during booting.


(2) Make sure that the boot sequence is set as follows: First hard drive (HD), then
CD/DVD drive.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 12


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(3) Set ATA/SATA mode:


 For machines that use serial ATA drives find a BIOS option to set the ATA
mode. Normal or combined can be chosen here. Make sure to set this setting
to ‘combined’.
 For Dell 745: SATA operation to be set to: Legacy
 For Dell 755: SATA operation to be set to: RAID/Autodetect/ATA
 For Dell 960: SATA operation to be set to: RAID/Autodetect/ATA

(4) Activate SMART Reporting: Set Drives\SMART Reporting to <ON>.


(5) Deactivate Hyper-threading (not applicable for multiple core CPUs):
Select Performance\Hyper-Threading and set it to <Off>.

(6) Activate Multiple CPU core (only applicable for multiple core CPUs (e.g. Dell
D830, Dell Optiplex 755 and Dell Optiplex 960):
Set Multiple CPU core to <ON>.
NOTE:
Also check if a SCSI hard disk is installed. If yes, additional
configurations need to be performed later in section 3.6.

(7) Save changes and exit BIOS:


 On a Dell 960: Press <Apply> to confirm all changes and press <Exit> to
leave the BIOS settings
 On other Dell PCs: Exit with <ESC> and save the changes.

3.3.2 Adjusting BIOS Settings for HP PCs

(1) Enter the BIOS setup by pressing <F10> during booting.


(2) Make sure that the boot sequence is set as follows: First the hard drive (HD),
then the CD/DVD drive.
NOTE:
On HP 8530p and HP 8540p laptops the CD/DVD drive is referred to as
"Notebook Upgrade bay".

(3) Set Storage Options:


 For HP DC7800 and HP DC7900 and 8000 Elite set: SATA emulation to IDE
 For HP 8530p and HP 8540p laptops set: SATA device mode to ATA/IDE

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 13


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

NOTE:
Also check if a SCSI hard disk is installed. If yes, additional
configurations need to be performed later in section 3.6.

(4) Go to File  Save Changes & Exit, press <Enter> and accept with <F10>.

NOTE:
After changing the BIOS settings on HP desktop PCs, an additional
screen is displayed that informs about updated configuration options.
Confirm the changes by pressing <F1>.

3.4 Installing the Operating System (OS)

NOTE:
Recommendation: Have the Acceptance Test Protocol document at hand and fill it in at
every installation and configuration step instead of filling it in afterwards.
The document is available in NX 2.0.8300 / 3.0.8300 Service Manual chapter 6
Acceptance Test DD+DIS105.10E (Intranet Link / Extranet Link).

IMPORTANT:
Installing the Operating System leads to a removal of all available data on the complete
hard disk, including all existing partitions!

3.4.1 Starting the OS Installation

(1) Insert the NX Starterkit DVD 2.


(2) Reboot the PC and
 On a HP PC: Press <F9> during the boot sequence (custom boot).
 On a DELL PC: Press <F12> during the boot sequence (custom boot).
(3) Select to boot from CD-ROM and press <Enter>.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 14


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(4) Confirm the installation by pressing <Y>.


+-------------------------------------------------------------+
! AGFA NX System !
! Multi Language !
+-------------------------------------------------------------+
! !
! New Installation. !
! !
! (Y) Installation !
! (N) Exit !
+-------------------------------------------------------------+

Figure 4

(5) Confirm the installation once more by pressing <Y>. Be aware that as soon as
<Y> is pressed, the (re-)installation is starting, and all data will be erased.
+-------------------------------------------------------------+
! AGFA NX on XP !
! Multi Language !
+-------------------------------------------------------------+
! !
! Installation will erase all current systems !
! !
! (Y) This machine will be totally reinstalled !
! (N) Exit to Operating system !
+-------------------------------------------------------------+

Figure 5

(6) After approximately 5 minutes, the RAID BIOS setting instruction is displayed.
Remove the DVD from the drive and press <CTRL+ALT+DEL> to reboot.

Figure 6

IMPORTANT:
Perform the next four steps only on an HP DC7800, HP DC7900, HP 8000 and
HP8540p system. On DELL systems continue with step (10).
(7) Enter the BIOS setup by pressing <F10> during booting.
(8)  On HP8540p: Set “SATA device mode” to: AHCI
 On HP 8530p: Do not activate the storage option RAID.
 On all other HPs: Set Storage Options/SATA emulation to: RAID
(9) Save changes and exit the BIOS.
NOTE:
After changing the BIOS settings on HP desktop PCs, an additional
screen is displayed that informs about updated configuration options.
Confirm the changes by pressing <F1>.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 15


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

IMPORTANT:
Perform the next steps only on a DELL 755 or DELL 960 system. Also perform
them if installing a DELL 755 or DELL 960 system without RAID!
(10) Enter the BIOS setup by pressing <F2> during booting.
(11) Set ATA/SATA mode to: RAID ON
(12) Save changes and exit the BIOS.
NOTE:
Directly after reboot an extra screen is displayed showing automatically
updated configurations. Save the changes by pressing <F1>.

NOTE:
The RAID setting is applied for consistency between all hardware types and has no
effect on the functionality of a system without RAID hardware.

Result From now on the installation is automatic.

3.4.2 Automatic Installation of Operating System

REQUIRED TIME:
Approximately 60 minutes

IMPORTANT:
Now the Operating System installs automatically. On RAID systems, the installation
may be interrupted by a failure of the system to boot after an automatic reboot.
This can be solved by changing the boot order of the hard disks:
On an HP PC:
(1) Press <F10> during booting to enter the BIOS setup.
(2) Go to Storage and select Boot Order.
(3) Set the other hard disk as primary boot media.
(4) Press <F10> to accept changes.
(5) Select Save Changes and Exit and press <F10> to accept.
On a DELL PC:
(1) During booting press <F12> to enter the BIOS setup.
(2) Select the other hard disk as primary boot media.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 16


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

Result After changing the boot order, the installation will continue.

IMPORTANT:
Don’t click anything until the log on screen described in the next section appears.
It could interrupt the installation.
The automatic installation consists of the Norton Ghost Restore Procedure, the
Windows Setup and post-installation customization.
 Norton Ghost Restore Procedure:
A progress bar indicates the progress of the restore procedure. The PC will
automatically reboot afterwards.
 Windows Setup:
During the installation a lot of screens of the Windows Setup are displayed.
The installation runs in the background.
 Post-installation customization:
Automatic scripts run once the Windows Setup process has completed. Wait until
the log on screen appears. Do not interrupt these scripts.

3.4.3 Finishing OS Installation

NOTE:
It is possible that after the Operating System installation the screen resolution is set
very low. In this case set the resolution to minimum 1280 x 1024 pixels. See below for
resolutions of laptop PCs:
 On a Dell D830 and HP 8530p laptop PC: Set it to 1680 x 1050 pixels.
 On a HP 8540p laptop PC: Set it to 1600 x 900 pixels.

NOTE:
Attach the Windows License sticker to the upgraded workstation next to the Windows
Server License sticker (if not yet done).

Result The Operating System installation is finished. It is now necessary to carry out system
configuration steps.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 17


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

3.5 Performing additional Partitioning

REQUIRED TIME:
Approximately 5 minutes

The operating system has now been installed on the System & Configuration partition
(drive letter C:).
For NX one more partition (Data & Backup partition) has to be defined with
the following parameters:

Drive letter: D:
Volume Label: DATA
Size: Complete Remaining rest of disk-space

IMPORTANT:
RAID Sytems contain a second hard disk for RAID mirroring. On RAID systems create
both partitions on one disk!

NOTE:
The screenshots below are made on a two hard disk system. If the platform contains
only one harddisk, the previous OS installation step has allocated an amount of
Gbyte of the first hard disk found for the C partition. The D partition has to be made
manually.

To create a new partition, perform the following steps:

(1) Right-click My Computer and select Manage from the menu:

Figure 7

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 18


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(2) In the Computer Management window, select Disk Management:

Figure 8: Example – The disk sizes may differ.

(3) Right-click an unallocated disk partition on disk 0.


(4) Select ”New Partition” from the menu.
(5) Click <Next> in the welcome window:

Figure 9

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 19


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(6) Select Primary Partition and click <Next>:

Figure 10

(7) Leave the default maximum diskspace as partition size for the new partition click
<Next>.

212869

Figure 11 – Example

(8) Choose the drive letter D and click <Next>.

Figure 12

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 20


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(9) Enter the following settings:


 Choose the partition formatting.
 Fill in the appropriate Volume Label DATA.
 Check: Perform a quick format
 Uncheck: Enable file and folder compression

Figure 13

(10) Click <Next>.


(11) Click <Finish> in the summary window to create the partition:

Figure 14

Result: The new partition and the remaining unpartitioned space is displayed in
the computer management window.

(12) Close the window Computer Management.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 21


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

3.6 Configuring Basic Operating System Settings

REQUIRED TIME:
Approximately 5 minutes

IMPORTANT:
Whenever the language or unicode settings are changed it is advised to use the
Administrator account.
Reason: The service language should always be English. In case service is performed
remotely, the service engineer might not be able to understand the local language.

(1) Log in using the Administrator account (don't use crservice account here).
(2) Go to Start  Settings  Control Panel. If the Control Panel is in
Category View click Switch to classic view.
(3) Double-click Date & Time and adjust the Date and Time Properties.

Figure 15

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 22


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(4) Open the tab Time Zone.

Figure 16

(5) Select the correct time zone and check the checkbox: Automatically adjust for
daylight saving changes
(6) Click <OK>.
(7) Go back to the Control Panel and choose Regional & Language Settings.
(8) Set the appropriate setting in the tabs Regional Options and Languages.
(9) Go to the tab Advanced and select the language to match the language version of
non-unicode programs:

Figure 17

(10) Check the box Apply all settings to the current user account and to the default
user profile.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 23


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(11) Click <OK> on following message:

Figure 18

NOTE:
This has only effect when users are not in a domain. For instructions to:
 Change the language of single users
 Change to non-standard string formats (e.g. Date & Time, figure
formats)
 Configure the default language for new NX users
refer to NX 2.0.8300 / 3.0.8300 Service Manual chapter 4 Installation and
Configuration Appendices DD+DIS105.10E section Changing the
regional settings of the NX (Intranet Link / Extranet Link).

(12) Go to the tab Languages and click <Details...>.


(13) If the correct keyboard is not listed under Installed services, click <Add...>,
select the keyboard layout and click <OK>.
(14) Select in the Default Input Language dropdown list, the correct input language.
(15) Click <OK>.
(16) Click <OK> if this screenshot appears:

Figure 19

(17) Confirm that the files already installed on the hard disk can be used
(Only if a dialog is displayed asking for a decision).
(18) Reboot the system if it is requested by the system.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 24


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

NOTE:
The following steps are only applicable to PCs with SCSI disk, not for SATA disks:

(19) On the desktop, click right on My Computer and select Properties.


(20) Go to tab Hardware and click Device Manager.
(21) Select the disk drive, click right and select Properties.
(22) In the tab Policies check Enable write caching on the disk.

Figure 20

(23) Click <OK>.


(24) Repeat the steps from step (19) if the system contains more than one SCSI disk.

3.7 Installing the latest available Hotfixes

REQUIRED TIME:
Approximately 50 minutes

Check if the latest Hotfix version is installed on the clean installed NX:
(1) To find out the installed version refer to NX 2.0.8300 / 3.0.8300 Service Manual
chapter 1 SW/HW Compatibility Chart DD+DIS105.10E
(Intranet Link / Extranet Link).
(2) If a higher Hotfix version has been downloaded by FUSE (see section 2.2.2),
install the latest Hotfix version on the system.

NOTE:
Some of the Hotfix patches are skipped during installation since the NX application is
not installed yet. This is why the Hotfix installation will be repeated after installing the
application (section Fehler! Verweisquelle konnte nicht gefunden werden.).

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 25


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

3.8 Configuring Network

REQUIRED TIME:
Approximately 10 minutes

NOTE:
The following actions need to be performed to configure the Windows network settings.

Checkpoints: See section:


1. Changing TCP/IP Settings 3.8.1

Configuring the wireless Network Connection for the 3.8.2


2.
wireless ICU Workflow

Changing the Hostname and/or Workgroup/Domain of the 3.8.3


3.
Operating System

3.8.1 Changing TCP/IP Settings of Local Area Connection (LAN)

IMPORTANT:
If the wireless Intensive Care Unit (ICU) feature will be used on this NX workstation,
do not perform the steps described here (in section 0). Continue with section 3.8.2.

(1) Log in using crservice account.


(2) Go to Start  Settings  Control Panel. If Control Panel is in Category View,
switch to Classic View and select Network Connections.

Figure 21

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 26


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(3) Right-click Local Area Connection and go to Properties. The window Local Area
Connection Properties opens.
(4) Select Internet Protocol (TCP/IP):

Figure 22

(5) Click <Properties>.


The window Internet Protocol (TCP/IP) Properties opens.
(6) Change the values according to the customers needs:

Figure 23

(7) Click <OK> (on the Internet Protocol (TCP/IP) Properties dialog).
(8) Click <OK> (on the Local Area Connection Properties dialog).

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 27


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

3.8.2 Configuring the wireless Network Connection for the wireless ICU Workflow

IMPORTANT:
 This section is only applicable if the wireless Intensive Care Unit feature (technical
license: NX_ICU_CONFIGURATION) is used on this NX workstation.
 The wireless NX must not be plugged into the Ethernet network: Only one operating
mode is possible, either wireless or fixed network.
 The wireless NX must only be used for GenRad applications.
 The images on the wireless NX are not available for Central Monitoring.

(1) Go to: Start  Settings  Control Panel  Network Connections


IMPORTANT:
Do not disable the Local Area Connection.

(2) Right-click Wireless Network Connection and choose Properties.


(3) Configure the wireless network connection according to the settings of the
wireless network on site.

3.8.3 Changing the Hostname and/or Workgroup/Domain of the Operating System

NOTE:
Gather information about network environment in the hospital (Domain or Workgroup)
or refer to the information gathered in chapter 3 Installation Planning
(DD+DIS106.10E).

To change the hostname and or the workgroup of the NX PC, perform following steps:

(1) Right-click the My Computer icon on the desktop and select Properties.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 28


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(2) Select the tab Computer Name:

Figure 24

(3) Click <Change…>. The window Computer Name Changes opens.


(4) Enter the new parameters:
 Computer name (Host Name)
 And/or the Workgroup
 And/or the Domain

Figure 25

(5) Click <OK>.


(6) If the Workgroup or Domain has been changed the following welcome message
is displayed:

Figure 26

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 29


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(7) Click <OK>. A message appears asking to reboot.


(8) Click <OK> (on the Computer Name Changes dialog box).
(9) Click <OK> (on the System Properties dialog box).
(10) Reboot the PC.

3.9 Installing NX Software and Documentation

REQUIRED TIME:
Approximately 30 minutes (without additional Software Update)

Prerequisites:
Successful OS installation & partitioning (C and D)
The crservice account language setting is 'English'.

NOTE:
The following actions need to be performed to install Application and Documentation.

Checkpoints: See section:


1. Installing Application 3.9.1

2. Installing NX Software Update (if available) 3.9.2

3. Installing NX Online Help 3.9.3

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 30


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

3.9.1 Installing NX Software

IMPORTANT:
 Before continuing, make sure the partition D (DATA) was created on disk 0 as
described in section 3.5.
 Do not disconnect the network during installation of the NX software, otherwise the
installation will fail!

(1) Log in using crservice account.


(2) Insert the NX Starterkit DVD 1.
(3) Browse in the CD-drive to the folder:
Service Software  NX Application NX X.0.8300.
(4) Double-click the file: NX X.0.8300 (Build.6.X.XXXX).exe
(5) The installation is guided by a setup wizard. Read carefully the messages and
instructions displayed in the wizard screens.
NOTE:
 The wizard will ask if a single partition system must be installed.
Do not check the box Install a single-partition system.

Figure 27

 This dialog decides on which partition the examination data (Pixel Files)
will be stored. The examination data must always be stored on
partition D on disk 0. Make sure the DATA partition D has been created
as described in section 3.5. If it hasn’t been created yet, cancel the
installation and create partition D.

(6) Reboot the workstation after finishing the installation wizard.


(7) Log in using the crservice account.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 31


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

NOTE:
The logging information of the process application installation can be found in the log
file: C:\Agfa\Healthcare\Log\NXMain.log

NOTE:
After first reboot of the system, Windows may detect the USB dongle as 'new
hardware' and the installation wizard pops up. Follow the instructions of the wizard to
install the WIBU drivers.

3.9.2 Installing NX Software Update (SU)

NOTE:
This is the moment when available NX software updates are installed. Check the NX
2.0.8300 / 3.0.8300 Service Bulletins to see if there is a SU released for NX 2.0.8300 /
3.0.8300. The most recent SU should be installed. The documentation and software is
available on NX Starterkit DVD 1.

(1) Insert the NX Starterkit DVD 1.


(2) Browse to the CD-drive to the folder Service Documentation to find the respective
installation instructions.
(3) Browse to the CD-drive to the folder Service Software to find the respective
installation package.

IMPORTANT:
Make sure that the ALF file to be activated, matches the version of the
software which is installed! How to pre-create an ALF file is described in section
2.3 "Obtaining Application License File (ALF)".

3.9.3 Installing NX Online Help

REQUIRED TIME:
Approximately 5 minutes

(1) Insert the NX Starterkit DVD 1.


Internet Explorer is automatically started, showing the Documentation CD
Navigation screen. If the Internet Explorer is not automatically started, browse to
the CD drive and double click index.htm.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 32


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(2) Click Install Online Help.


Three messages can pop up:
 Internet Explorer message “Active content in html …”: Click <Yes>.
 File download security warning: Click <Run>.
 Internet Explorer security warning “Publisher could not be verified …”:
Click <Run>.
(3) Click <Next>.
(4) The Online Help installer will guide through the Online Help installation
on the system. Choose one of the options.
(5) Click <Install>.
(6) Click <Finish> to end the installation procedure.
(7) Remove the NX Starterkit DVD 1.
(8) After finishing the upgrade inform the user representative about the usage of the
NX Online Help.

NOTE:
 Installing the Online Help (any language) will also install a special Thai font on the
NX Workstation that is required to read the Thai Online Help. This implies that
browsing Thai Online Help directly from the CD will not work before the Online Help
is actually installed on the Workstation.
 If required, the font can be manually installed by copying this file from the CD into
the Windows font folder: C:\…\Nx Online Help\Windows\Fonts\Angsd___.ttf

3.10 Installing an Uninterruptible Power Supply (UPS)

REQUIRED TIME:
Approximately 45 minutes
For the UPS installation and configuration procedure refer to NX 2.0.8300 / 3.0.8300
Service Manual chapter 4 Installation and Configuration Appendices DD+DIS105.10E
(Intranet Link / Extranet Link).

3.11 Installing TWAIN Digitizer Software (ASAP)

REQUIRED TIME:
Approximately 30 minutes

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 33


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

NOTE:
This section is only applicable if a TWAIN digitizer is connected to the NX workstation:

(1) Install the digitizer software of the TWAIN digitizer (CR 30-X or DX-S or DX-G with
SW < NIM_2XXX) as described in the installation instructions enclosed to the
software.

IMPORTANT:
When the configuration of the NX workstation is finished, and the digitizer is switched
on, it is possible, that the following error messages are displayed on the NX
workstation:

Symptom 1 Error indicating, that the registration of the serial number failed.
Reason Restore not performed: NX workstation does not recognize the serial
number of the connected device.
Solution Step 1: Switch off digitizer. Step 2: Restore the latest digitizer backup.
Step 3: Re-boot control PC. Step 4: Switch on digitizer.
Alternative solution:
Fill in site specific data (Digitizer service client menu Configuration – Site
specific data). When done, select buttons <update site specific
data> and <configure site with new EMS serial number>.
Symptom 2 Error indicating, that software versions loaded onto machine and
installed on PC are not identical.
Reason Software mismatch between digitizer software loaded on digitizer and
installed on NX workstation.
Solution Perform software upload to the digitizer as described in the instructions
in the digitizer device software package. If the software version on the
digitizer is higher than on the workstation, install the corresponding
digitizer software version on the workstation.

3.12 Installing FUSE Software

REQUIRED TIME:
Approximately 10 minutes

For instructions to obtain and install the Fast Update Service (FUSE) software refer to
the FUSE Service Manual FUSE - Fast Update Service (Automatic Software Update)
DD+DIS134.08E (Intranet Link / Extranet Link).
For external partners: Please contact your local Agfa representative.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 34


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

4 Operating System Configuration

Perform following steps as described in NX 2.0.8300 / 3.0.8300 Service Manual


chapter 4 Installation and Configuration DD+DIS105.10E (Intranet Link / Extranet Link):

 Editing Security Settings according to Hospital Guidelines


 Creating and Configuring Users on Operating System and Domain

5 Loading Licenses and Activating the Installation

5.1 Loading Licenses (ALF)

CAUTION:
Due to wrong date & time and time zone settings license may become invalid.
Before activating a license key file, make sure the correct time zone, date and time
have been configured!
For details refer to section 3.6.

REQUIRED TIME:
Approximately 10 minutes

NOTE:
If the installation wizard for the license dongle has not appeared yet, unplug the dongle
and reconnect it. The wizard will pop up. Click through it using the default settings.

Figure 28

To load the licenses the Application License File (*.alf) is needed (see section 0):

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 35


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(1) Copy the ALF to: C:\AGFA\Healthcare\NX\Configuration\Licensing


(2) Start up the license manager tool from:
Start  AGFA  Service  License manager
(3) Click the <Add License…> button.
(4) Browse to the folder: C:\AGFA\Healthcare\NX\Configuration\Licensing
(5) Select the ALF and click <Open>.
(6) Reboot the PC.
Result The license file has been loaded successfully.

NOTE:
The logging information of the Licensing process can be found in the log file:
C:\Agfa\Healthcare\Log\LicensingService.log

5.2 Activation

REQUIRED TIME:
Approximately 15 minutes

The NX workstation is setup to be configured as one of these profiles, based on the


information retrieved from the loaded ALF file:
In-room mixed use NX workstation: Depending on the available licenses, the NX
workstation will be able to perform:
o Only GenRad examinations
o Only Mammography examinations
o Both GenRad and mammography examinations
For more information about the license configurations refer to
NX 2.0.8300 / 3.0.8300 Service Manual chapter 1 Product Description
DD+DIS105.10E (Intranet Link / Extranet Link).
In-room veterinary NX workstation: Allows only configuring veterinary examinations.
Central monitoring system: Allows monitoring a set of in-room NX workstations.

IMPORTANT:
Activation will fail if the network is unplugged during activation of the NX Software.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 36


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(1) Log out and in again using the crservice account.

(2) Start the NX Activation Utility using the shortcut on the desktop, or
start it via
Start Menu  Agfa  NX  Service  Install Tools.
Figure 29

(3) Click <Next> on the message that a fresh NX will be activated.

(4) Choose the values for dose calculation: Logarithm Median (LGM) values or
Exposure Index (EI)
Decision should be based on these rules:
 If a CMS is installed, all connected NX in-room workstations must have the
same setting as the CMS.
 For information about which digitizer type is compatible with which dose
calculation, refer to NX 2.0.8300 / 3.0.8300 Service Manual chapter 1
SW/HW Compatibility Chart DD+DIS105.10E (Intranet Link / Extranet Link). For
more generic information about dose calculation refer to chapter 1 Product
Description, section Dose Monitoring (Intranet Link / Extranet Link) and the
NX User Manual.

(5) Click <Next>.

NOTE:
Depending on the licenses, following components are installed now:
 Micro Calcification Enhancement (MCE)
 X-Ray Generator and DR Detector Interfacing Software (XRDI)
 Sedecal Software Console (DX-D 300 U-arm)
The software for each component is already available in folder:
c:\Agfa\Healthcare\NX\Installers\

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 37


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(6) If one or more of the above mentioned


components are installed, select the
component and press next. See
example in Figure 30.
If none of the above mentioned
components are installed, continue
with next step.

Figure 30

IMPORTANT:
If more than one component is displayed (like in the example of Figure 30),
select the required component. For compatibility information refer to the
solution specific information.
Example for XRDI:

Solution GSO Library Path Document


CR Mammo 1C Computed Radiography
=> CR Mammography Systems
=> CR Mammo Phase 1C
CR Mammo (DX-M) Computed Radiography Software / Hardware
=> CR Mammography Systems Compatibility Chart
=> CR Mammo Solution (DX-M) and/or specific Service
DX-D 300 Direct Radiography => DR Equipment Bulletin
=> DX-D 300 (Soft Console)
DX-D 500 Direct Radiography => DR Equipment
=> DX-D 500n

(7) Select the default language for the NX workstation and click <Next>.

NOTE:
After Activation, the NX workstation will be configured to use the default
string formats (e.g. date and time) for the selected language, and the
selected language will be the default language for new users.
For instructions how to tune the local string formats (e.g. date and time) or
how to change a user's language refer to NX 2.0.8300 / 3.0.8300 Service
Manual chapter 4 Installation and Configuration Appendices
DD+DIS105.10E (Intranet Link / Extranet Link).

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 38


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(8) Check the overview of specified selections: To change options, click <Back>. To
cancel the activation, click <Cancel>. If all selections are correct, click
<Activate>. The activation starts.
(9) Click <Next> after the activation has completed.
(10) Choose Reboot now and click <Finish>.

NOTE:
The logging information of the process NX activation can be found in the log file:
C:\Agfa\Healthcare\Log\Nx\NXactivate.log

6 Installing additional Drivers

6.1 Installing Monitor Software and Calibration

REQUIRED TIME:
Approximately 5 minutes

(1) To identify and retrieve the required monitor software refer to


NX 2.0.8300 / 3.0.8300 Service Manual chapter 1 SW/HW Compatibility Chart
DD+DIS105.10E (Intranet Link / Extranet Link).
(2) For instructions to install the software and calibrating the monitors, refer to
NX 2.0.8300 / 3.0.8300 Service Manual chapter 5 Monitor Options
DD+DIS108.10E (Intranet Link / Extranet Link).

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 39


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

6.2 Upgrading ID Tablet Firmware

REQUIRED TIME:
Approximately 10 minutes

IMPORTANT:
 If the ID tablet firmware version is 1.12.07 or higher this step is not required.
 If a mobile ID Scanner is used with this NX workstation, updating the ID Tablet
firmware is not required.
For instructions refer to NX 2.0.8300 / 3.0.8300 Service Manual chapter 4 Installation
and Configuration Appendices DD+DIS105.10E section ‘Upgrading ID tablet firmware’
(Intranet Link / Extranet Link). This section also describes how to check the ID tablet
firmware version.

6.3 Installing the ID Scanner Drivers

REQUIRED TIME:
Approximately 10 minutes

IMPORTANT:
This section is only applicable if the ID Scanner is used in combination with this NX
workstation.

INSTRUCTION:
For detailed instructions refer to the ID Scanner Service Manual chapter 1 Controls,
Connections, and Setup Procedures DD+DIS124.09E (Intranet Link / Extranet Link).
A short overview of the instructions is:
(1) Install the ID Scanner drivers on the NX Workstation.
(2) Adjust the COM Port Properties.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 40


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

7 NX Configuration via Configuration Tool

REQUIRED TIME:
Approximately 5 – 15 minutes depending on the amount of configuration tasks

(1) Perform the "Configuration via Configuration Tool" as described in NX 2.0.8300 /


3.0.8300 Service Manual chapter 4 Installation and Configuration
DD+DIS105.10E (Intranet Link / Extranet Link).
(2) Perform following settings in case the NX workstation is running on a laptop PC:
 Power settings as described in NX 2.0.8300 / 3.0.8300 Service Manual
chapter 4 Installation and Configuration Appendices DD+DIS105.10E section
Laptop Power Settings (Intranet Link / Extranet Link).
These settings also have to be applied on the automatically created
accounts: ‘crservice’ and ‘Administrator’
 On laptop PCs configure the screen resolution manually on each individual
user account:
o On the Dell D830 and HP 8530p laptop PCs: Change the screen
resolution to 1680 x 1050 pixels.
o On the HP 8540p laptop PC: Change the screen resolution to 1600 x
900 pixels
.
(3) Perform additional configuration steps if required for this specific system, e.g.
additional users, PCAnyWhere configuration, etc.

8 Completing Installation

8.1 Activating RAID (Redundant Array of Independent Disks)

IMPORTANT:
RAID is supported ONLY on hardware that has been ORIGINALLY delivered as RAID
system and equipped with a second harddisk.
For specification of hardware types supporting RAID refer to NX 2.0.8300 / 3.0.8300
Service Manual chapter 1 SW/HW Compatibility Chart DD+DIS105.10E
(Intranet Link / Extranet Link).
It is not allowed to install RAID on systems not originally delivered as RAID system,
even if the hardware is compatible with RAID.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 41


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

REQUIRED TIME:
Approximately 130 minutes (waiting time included)

(1) Insert the NX Starterkit DVD 2.


(2) Refer to NX 2.0.8300 / 3.0.8300 Service Manual chapter 1 SW/HW Compatibility
Chart DD+DIS105.10E to find the folder of the respective RAID tool installer,
depending on the NX PC that is used.
(3) Double-click the respective installer file: E.g. R208364.exe or sp44098.exe
(4) A security warning may be displayed. Click <Run>:

Figure 31

(5) Click <Continue> on the next dialog:

Figure 32

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 42


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(6) A location is requested where to unzip the files to:


Enter C:\DRV\<Name of the RAID installer file> and then click
<Ok>.

R208364

Figure 33: Example screenshot

(7) Create the location by clicking <Yes>.

Figure 34

(8) Click <Ok> on the next dialog:

Figure 35

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 43


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(9) Confirm all screens of the Intel Matrix Storage Manager setup wizard.
Result: The PC will be automatically rebooted:

Figure 36: First dialog of Setup Wizard Figure 37: Last dialog of Setup Wizard

(10) Log in using crservice account.


(11) Click the Start button. Go to All Programs  Intel Matrix Storage Manager  Intel
Matrix Storage Console to launch the Intel Storage Console.
(12) Select: Protect data from a hard drive failure with RAID 1:

Figure 38

(13) Click <Create a RAID 1 volume>.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 44


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(14) Click <Yes> on following message to start migrating:

Figure 39

Result The progress of migration is displayed:

Figure 40

IMPORTANT:
Do not perform a reboot of the system until the migration process has finished
completely! Apart from that the the migration process does not interfere with other
actions to be carried out on the PC.

8.2 Adjusting Anti-Virus Settings

To find the procedures on how to configure the supported Anti-virus software refer to
NX 2.0.8300 / 3.0.8300 Service Manual chapter 4 Installation and Configuration
Appendices DD+DIS105.10E (Intranet Link / Extranet Link).

8.3 Backup of Configuration Data

REQUIRED TIME:
Approximately 10 minutes

After installation and after any change to the configuration of NX, it is important to
make a backup. Follow this procedure:
(1) Open the Configuration Tool and load the active configuration.
(2) In the Configuration Tool menu select: File  Save as…

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 1 NX 2.0.8300 / 3.0.8300 Chapter 5 / 45


09-2010 Type 4406/303 Agfa Company Confidential
DD+DIS107.10E Software Options - Clean Installation Procedure

(3) Choose a portable medium to save the configuration file on and store this file on a
save place. Name the file that it can be identified later on: e.g.
<Site>_initial_configuration_<date>.xml
The file should be kept outside the NX hard disk to have it available in case the
NX hard disk crashes.
(4) Run the script ‘Backup all data’:
Agfa  NX  Service  Backup Restore  Backup all data

For more information about backup scripts, see NX 2.0.8300 / 3.0.8300 Service Manual
chapter 4 Installation and Configuration Appendices DD+DIS105.10E
(Intranet Link / Extranet Link).

8.4 Performing the NX Acceptance Test

REQUIRED TIME:
Approximately 30 minutes

IMPORTANT:
Check if all required fields of the Acceptance Test have been filled out, complete it and
hand it over to the customer according to the instructions given in the Acceptance Test
document.
The document is available in the NX 2.0.8300 / 3.0.8300 Service Manual chapter 6
Acceptance Test DD+DIS105.10E (Intranet Link / Extranet Link).

IMPORTANT:
Refer to the respective Solution Manual (e.g. CR Mammography Solution Manual) for
additional Acceptance Test on Solution level.