High Voltage Power Supply 50KV max. for


DIY Electro-Spinning

MSK-ESPS-TE4020
KGC RESOURCES SDN. BHD.(Co. no. 223165-D)
Office: No. 2-2-3, Jalan Setia Prima E, U13/E, Seksyen U13
Setia Alam, 40170 Shah Alam, Selangor Darul Ehsan.
Tel: 03-33412880 Fax: 03-33439880
Email: kgcresources88@yahoo.com
Website: http://www.kgcscientific.com CIDB LICENSE: 1991124-SL054439 MOF LICENSE: 357-02067811

MSK-ESPS-TE4020 Function introduction

HVT MSK-ESPS-TE4020 Series


KGC Resources Sdn Bhd Electrospinning, electrostatic flocking,
digital voltage, current indication
Double closed loop control, high precision, digital voltage, current indication
 Maximum output power
 Maximum output voltage
 Adjustable output voltage
 Constant current control function
 Overvoltage and output short circuit
protection
 Remotely adjustable voltage and
current
 Safety interlock function

MSK-ESPS-TE4020 series high voltage power supply, high voltage power supply for
electrospinning equipment. The output voltage is adjustable, and the front panel of the power
supply has a voltage and current liquid crystal display instrument, which can achieve a linear
and steady rise of the high voltage output. The series of power supplies can also be connected
to a potentiometer for remote control of output voltage and current, and has external voltage
and current display. The high voltage output has overvoltage and short circuit protection, and
safety interlock.

Usual application:
KGC RESOURCES SDN. BHD.(Co. no. 223165-D)
Office: No. 2-2-3, Jalan Setia Prima E, U13/E, Seksyen U13
Setia Alam, 40170 Shah Alam, Selangor Darul Ehsan.
Tel: 03-33412880 Fax: 03-33439880
Email: kgcresources88@yahoo.com
Website: http://www.kgcscientific.com CIDB LICENSE: 1991124-SL054439 MOF LICENSE: 357-02067811

Electrospinning, electrostatic flocking


Capillary electrophoresis, electrostatic spraying, electrostatic precipitator
science experiment
Optional features:
SF output short circuit function

Specifications:
Input 220V ± 10% AC 50/60 HZ
Output The highest voltage output is 50KV
(optional), and the 0V to maximum voltage
is adjustable. Maximum power 100W
(optional)
Voltage control Power board surface: The output voltage can
be set between 0V and the highest voltage by
the potentiometer of the power supply.

External remote control: External remote


control signal (0-10V) adjusts the output
from 0V to the highest output current.
Current control Freely set the output current through the
power supply external multi-turn
potentiometer

Voltage regulation Relative load: 0.01% (no load to rated load)

Relative input: ±0.01% (input voltage


change is ±10%)
Current regulation Relative load: 0.01% (no load to rated load)
KGC RESOURCES SDN. BHD.(Co. no. 223165-D)
Office: No. 2-2-3, Jalan Setia Prima E, U13/E, Seksyen U13
Setia Alam, 40170 Shah Alam, Selangor Darul Ehsan.
Tel: 03-33412880 Fax: 03-33439880
Email: kgcresources88@yahoo.com
Website: http://www.kgcscientific.com CIDB LICENSE: 1991124-SL054439 MOF LICENSE: 357-02067811

Relative input: ±0.01% (input voltage


change is ±10%)
Ripple voltage Under the premise of outputting the rated
voltage, the peak value of the ripple voltage
is 0.1% of the maximum output voltage
(smaller ripple is optional)
Ambient temperature Working: -30°C - +50°C ; Saving: -20°C -
80°C
Temperature series Every degree Celsius 0.02%
Stability After turning on the machine for 30 minutes,
the stability is lesser than 0.1% every 8 hours
(better stability is optional)
Voltage and current indication 0-+10V, 1% accuracy under rated output
conditions
Remote control and indication Externally potentiometer and +10V
(optional) voltage reference inside the power
supply for remote control of the output
voltage DB15 includes 0-10V (optional)
voltage and current indication signals, which
can be connected to various numbers or
pointers
Dimensions Long 244mm, Wide 155mm, Height 118mm
Connectors and accessories Recessed epoxy insulated conduit, diameter
of 16 metal connector, high voltage cable for
a total length of 1 meter. DB15 interface
contains control and display signals
KGC RESOURCES SDN. BHD.(Co. no. 223165-D)
Office: No. 2-2-3, Jalan Setia Prima E, U13/E, Seksyen U13
Setia Alam, 40170 Shah Alam, Selangor Darul Ehsan.
Tel: 03-33412880 Fax: 03-33439880
Email: kgcresources88@yahoo.com
Website: http://www.kgcscientific.com CIDB LICENSE: 1991124-SL054439 MOF LICENSE: 357-02067811

MSK-ESPS-TE4020 Operating Instruction

1. Checking and Installation

1.1 Checking at Early Stage

Check the outer packaging of the power supply to find out if there are any signs of damage
caused during transportation. If you find it, please inform KGC and do not remove the materials
used for packaging. Check the panel and casing after unpacking to see if there are any signs of
damage.

1.2 Installation
The power supply can be installed in an electrical cabinet or placed on a workbench. The
following shows the dimension of the machine.
KGC RESOURCES SDN. BHD.(Co. no. 223165-D)
Office: No. 2-2-3, Jalan Setia Prima E, U13/E, Seksyen U13
Setia Alam, 40170 Shah Alam, Selangor Darul Ehsan.
Tel: 03-33412880 Fax: 03-33439880
Email: kgcresources88@yahoo.com
Website: http://www.kgcscientific.com CIDB LICENSE: 1991124-SL054439 MOF LICENSE: 357-02067811

High voltage (on)

High
voltage
Voltage Setting
(close)
adjustment

KGC Resources Sdn Bhd


On/Off
KGC RESOURCES SDN. BHD.(Co. no. 223165-D)
Office: No. 2-2-3, Jalan Setia Prima E, U13/E, Seksyen U13
Setia Alam, 40170 Shah Alam, Selangor Darul Ehsan.
Tel: 03-33412880 Fax: 03-33439880
Email: kgcresources88@yahoo.com
Website: http://www.kgcscientific.com CIDB LICENSE: 1991124-SL054439 MOF LICENSE: 357-02067811

Power input
JPI Control interface AC 220V

Caution

KGC Resources Sdn Bhd High voltage output

Behind part of the machine


KGC RESOURCES SDN. BHD.(Co. no. 223165-D)
Office: No. 2-2-3, Jalan Setia Prima E, U13/E, Seksyen U13
Setia Alam, 40170 Shah Alam, Selangor Darul Ehsan.
Tel: 03-33412880 Fax: 03-33439880
Email: kgcresources88@yahoo.com
Website: http://www.kgcscientific.com CIDB LICENSE: 1991124-SL054439 MOF LICENSE: 357-02067811

2. Safety Precautions

Safety precautions

Improper use of high voltage power supplies may threaten personal safety.

1. High voltage power supply must be grounded reliably.

2. Do not touch the high voltage connector unless the high voltage power supply is turned off
and the internal capacitance of the load and power supply is completely discharged.

3. Wait 5 minutes after turning off the high voltage power supply, so that the internal
capacitance of the power supply is fully discharged.

4. Do not operate the power supply in a humid environment, nor ground yourself.

5. Maintenance may require the removal of the power supply cover. The operation must be
carried out by qualified personnel to prevent electric shock.
KGC RESOURCES SDN. BHD.(Co. no. 223165-D)
Office: No. 2-2-3, Jalan Setia Prima E, U13/E, Seksyen U13
Setia Alam, 40170 Shah Alam, Selangor Darul Ehsan.
Tel: 03-33412880 Fax: 03-33439880
Email: kgcresources88@yahoo.com
Website: http://www.kgcscientific.com CIDB LICENSE: 1991124-SL054439 MOF LICENSE: 357-02067811

2.1 Steps

1. Check the label on the power supply to make sure the rating of the power supply is the same
as you requested.

2. Good grounding: The outer casing of the high-voltage power supply must be well grounded.
The grounding end of the power supply and the bottom wire can be directly connected by
copper wires. The return line of the high voltage load should be directly connected to the
ground of the power supply. It is not recommended to ground the return line of the high voltage
load and the ground end of the high voltage power supply separately.

3. Connect the high voltage output line and load.

4. If the user does not have remote control requirements, you can directly short the 1 pin and
15 pin of the DB15 terminal, and short the 8 pin and 2 pin; if the user has remote control
requirements, it can be connected at the 1 pin and 2 pin of the DB15 terminal. Adjust the
potentiometer to adjust the current and voltage.

5. Connect a switch to the 6-pin and 11-pin of DB5 to control the remote high-voltage switch.
The wiring diagram is shown in Figure 2.1.

6. Connect the 22V AC power input cable.

7. At this point, you can see that there is a high voltage output; adjusting the external
potentiometer can change the output voltage. With a slow-start high-voltage power supply, the
output will rise slowly and the rise time is about 2-4 seconds.
KGC RESOURCES SDN. BHD.(Co. no. 223165-D)
Office: No. 2-2-3, Jalan Setia Prima E, U13/E, Seksyen U13
Setia Alam, 40170 Shah Alam, Selangor Darul Ehsan.
Tel: 03-33412880 Fax: 03-33439880
Email: kgcresources88@yahoo.com
Website: http://www.kgcscientific.com CIDB LICENSE: 1991124-SL054439 MOF LICENSE: 357-02067811

8. Turn off the voltage power switch or cut off the 220V power input line of the high voltage
power supply to turn off the high voltage power supply completely.

Warning

1. After cutting off the power, do not approach and disconnect the load until the internal power
of the high voltage power supply is completed.

2. When the power supply of the high-voltage power supply is cut off, the voltage indication
of the high-voltage power supply panel does not represent the actual voltage output. At this
time, it is likely that high-voltage static electricity still exists in the high-voltage output and the
load.

3. When operating the high-voltage power supply, be sure to cover the power supply. Do not
operate or repair the internal circuit of the high-voltage power supply. Dangerous or even fatal
voltage may be generated inside the power supply.

2.2 Power Supply Precautions


1. Attention! The high-voltage output of the high-voltage power supply is strictly limited under
any circumstances. Under normal circumstances, the rated output voltage is not allowed to
exceed. Although the power supply itself limits the high-voltage output, it is only a steady-state
limit and cannot guarantee the high-voltage output. The transient does not exceed the rated
output voltage. If the high voltage output exceeds the rated voltage of the high voltage power
supply (steady state or transient), it may cause permanent damage or even damage to the high
voltage power supply and its load (such as X tube).
KGC RESOURCES SDN. BHD.(Co. no. 223165-D)
Office: No. 2-2-3, Jalan Setia Prima E, U13/E, Seksyen U13
Setia Alam, 40170 Shah Alam, Selangor Darul Ehsan.
Tel: 03-33412880 Fax: 03-33439880
Email: kgcresources88@yahoo.com
Website: http://www.kgcscientific.com CIDB LICENSE: 1991124-SL054439 MOF LICENSE: 357-02067811

2. If the above damage is not avoided, the rise time of the high voltage output must be strictly
controlled to ensure that the high voltage rises slowly. Under normal circumstances, the high
voltage rise time of the high voltage power supply is not less than 2-4 seconds. In order to
achieve a slow rise of the high voltage, high voltage output is required. The control signal is a
ramp-up signal from the starting voltage to the target voltage. The rise time of the signal is not
less than 2 seconds. It is strictly forbidden to use the step-down signal instead! The step-down
signal or the ramp signal that rises too fast will cause overshoot of the high-voltage output, that
is, the high-voltage output exceeds the set value of the high-voltage output at the moment of
high-voltage rise.

2.3 Standard function

1. Remote interface notes:


If possible, isolate the high voltage power interface signal from the remote circuitry. Try to
simulate the voltage and current given signals, and the analogue voltage and current sense
signals are isolated by an analogue isolation amplifier. All interface wires and cables must be
well shielded. Remotely given signal, the output voltage and current of the high voltage power
supply can be remotely adjusted by an external potentiometer.

2. Remote monitoring
The corresponding output of pin 7 and pin 4 of connector DB15 on the power panel can be
used as the test point of the current of the voltage respectively. The output signal is always 0
to 10V (optional) regardless of whether the polarity of the output is positive or negative.
Positive signal. 0 to 10V represents a nominal output of 0 to 100%. The output resistance of
this output signal is less than 100 ohms.
KGC RESOURCES SDN. BHD.(Co. no. 223165-D)
Office: No. 2-2-3, Jalan Setia Prima E, U13/E, Seksyen U13
Setia Alam, 40170 Shah Alam, Selangor Darul Ehsan.
Tel: 03-33412880 Fax: 03-33439880
Email: kgcresources88@yahoo.com
Website: http://www.kgcscientific.com CIDB LICENSE: 1991124-SL054439 MOF LICENSE: 357-02067811

3. High pressure on and off:


The 6 pin of DB15 can be used as the input of the external high voltage control signal. When
it is grounded (11 or 12 feet), the high voltage of the power supply can work. When the high
voltage of the power supply is working normally, turning on the switch (floating) can put the
power supply into the high voltage off state. The wiring diagram is shown in Figure 2.1.

4. Remote voltage and current regulation:


If the user does not have remote control requirements, you can directly short the 1 pin and 15
pin of the DB15 terminal, and short the 8 pin and the 2 pin. If the user has remote control
requirements, the DB15 terminal 2 pin can be connected to the tap end of the potentiometer.
The other two ends are connected to 10 feet and ground (11 feet or 12 feet) respectively. The
potentiometer can be adjusted to change the output voltage. Also connect a potentiometer from
the 1 pin and adjust the potentiometer to change the output current. The 10 pin and 12 (or 11
pin) pins of DB15 are 10V reference voltage and ground respectively. The wiring diagram is
shown in Figure 2.1.
KGC RESOURCES SDN. BHD.(Co. no. 223165-D)
Office: No. 2-2-3, Jalan Setia Prima E, U13/E, Seksyen U13
Setia Alam, 40170 Shah Alam, Selangor Darul Ehsan.
Tel: 03-33412880 Fax: 03-33439880
Email: kgcresources88@yahoo.com
Website: http://www.kgcscientific.com CIDB LICENSE: 1991124-SL054439 MOF LICENSE: 357-02067811

Voltage control signal output


Current control signal output

Voltage output indication signal

Chain output

Chain circuit (optional)

Current output indication signal

Chain circuit (optional)

Voltage control signal input

electricity failure
Current control signal
input

Figure 2.1 Wiring Diagram


KGC RESOURCES SDN. BHD.(Co. no. 223165-D)
Office: No. 2-2-3, Jalan Setia Prima E, U13/E, Seksyen U13
Setia Alam, 40170 Shah Alam, Selangor Darul Ehsan.
Tel: 03-33412880 Fax: 03-33439880
Email: kgcresources88@yahoo.com
Website: http://www.kgcscientific.com CIDB LICENSE: 1991124-SL054439 MOF LICENSE: 357-02067811

OTHERS
1. Maintenance and testing guide

1.1 Regular maintenance: This product does not require regular maintenance.

1.2 Testing:
WARNING: High voltage hazard, testing high voltage power supplies requires professionally
qualified personnel.

1.3 High voltage divider


The high-voltage divider recommended for high-voltage measurement is not less than 1%, and
its range is greater than the highest output voltage of the high-voltage power supply.

2. Order and replace parts

2.1 Replacement parts


Please contact KGC directly to order and replace related accessories. The company can provide
customers with the necessary types of accessories and components. It is recommended that
only qualified personnel can repair and replace the accessories and components. High pressure
is very dangerous, even a small mistake in maintenance can have serious consequences.

2.2 Ordering accessories


Each power supply has an identification label with the model number and serial number of the
power supply. Please indicate the power supply model number and serial number when the
user needs other related engineering and application information.
KGC RESOURCES SDN. BHD.(Co. no. 223165-D)
Office: No. 2-2-3, Jalan Setia Prima E, U13/E, Seksyen U13
Setia Alam, 40170 Shah Alam, Selangor Darul Ehsan.
Tel: 03-33412880 Fax: 03-33439880
Email: kgcresources88@yahoo.com
Website: http://www.kgcscientific.com CIDB LICENSE: 1991124-SL054439 MOF LICENSE: 357-02067811

3. Warranty terms
3.1 KGC provides a free warranty service for the power products it produces during the
warranty period. The warranty covers products with defects in materials and workmanship or
products that fail due to product defects under normal use.

3.2 The company is not responsible for warranty services other than this
3.3 The company's warranty liability and the purchaser's compensation will be limited to the
purchase price of the product. The repair or replacement of the product during the warranty
period depends on the company's inspection results for the returned product. The customer will
bear the cost of returning the product and returning the product.