Вы находитесь на странице: 1из 64

MAMMOMAT NovationDR

Wiring Diagrams

© Sieme ns AB 2004
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.

English
Print No.: SPB7-250.844.01.03.02 Doc. Gen. Date: 08.04
Replaces: SPB7-250.844.01.02.02
0-2 Revision

Chapter Page Revision


all all 03

Document revision level


The document corresponds to the version/revision level effective at the time of system delivery. Revi-
sions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho-
rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.

Mammomat NovationDR SPB7-250.844.01 Page 2 of 4 Siemens AG


Wiring Diagrams Rev. 03 08.04 CS SD 24 Medical Solutions
Contents 0-3

Page
Document revision level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 - 2
Disclaimer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 - 2

1 _______System overview ______________________________________________ 1 - 1

System overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1

2 _______Stand Overview _______________________________________________ 2 - 1

PLACEMENT OF PC-BOARDS AND COMPONENTS, STAND . . . . . . . . . . .2-1


MAMMOMAT NovationDR - Front . . . . . . . . . . . . . . . . . . . . . . . .2- 1
MAMMOMAT NovationDR - Back (seen from the right side) . . . . . . . . . . .2-2
LIST OF BOARDS AND FUSES, STAND . . . . . . . . . . . . . . . . . . . . . .2-4
PCB Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2- 4
Fuses Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2- 4
LIST OF SWITCHES AND COMPONENTS, STAND . . . . . . . . . . . . . . . . .2-5
Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2- 5
Potentiometers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2- 6
Motors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2- 6
LIST OF SIGNALS AND TEST POINTS, STAND . . . . . . . . . . . . . . . . . . .2-8
LIST OF VOLTAGES AND TEST POINTS, STAND . . . . . . . . . . . . . . . . 2 - 12

3 _______Stand Functional Diagrams ______________________________________ 3 - 1

BLOCK DIAGRAM CABLE CONNECTIONS . . . . . . . . . . . . . . . . . . . . .3-1


GROUNDING CONNECTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . .3- 2
CPU BOARD INTERNAL, GENERATOR INTERFACE . . . . . . . . . . . . . . . .3-6
COLLIMATOR 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 15
COLLIMATOR 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 16
OBJECT TABLE 1 + 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 17

4 _______Generator Overview ____________________________________________ 4 - 1

MAINS VOLTAGE CONNECTION . . . . . . . . . . . . . . . . . . . . . . . . . .4-1


Electrical power to mains input . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Power line impedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4- 1
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
GENERATOR: LIST OF MODULES AND FUSES
PC-BOARDS: MASTER BOARD D750 . . . . . . . . . . . . . . . . . . . . . . . .4-2
Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4- 2
Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4- 2
PC-boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4- 2
GENERATOR: PLACEMENT OF MODULES, CONNECTORS,
LEDs AND POTENTIOMETERS . . . . . . . . . . . . . . . . . . . . . . . . . . .4- 4
Generator description of MAMMOMAT NovationDR . . . . . . . . . . . . . . .4-4
CONTROL PANEL FUNCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
D750 MASTER BOARD: LIST OF SWITCHES, TEST POINTS . . . . . . . . . . . .4-6
Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4- 6

Siemens AG SPB7-250.844.01 Page 3 of 4 MAMMOMAT NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
0-4 Contents

Page
Test points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
GENERATOR: LIST OF SWITCHES, POTENTIOMETERS, TEST POINTS, LEDs . 4-9
Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Potentiometers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Test points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 10

5 ______ General Functional Diagrams_____________________________________ 5 - 1

GENERATOR: BLOCK DIAGRAM . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1


GENERATOR: MAINS INPUT & MAINS INPUT CONVERTER,
PRINCIPLE DIAGRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
D750 MASTER BOARD: PRINCIPLE DIAGRAM . . . . . . . . . . . . . . . . . . 5-5
GENERATOR - TUBE POWER SUPPLY:
PRINCIPLE DIAGRAM, TUBE FILAMENT AND FOCUS SELECTION . . . . . . . 5-6
GENERATOR - TUBE POWER SUPPLY:
PRINCIPLE DIAGRAM TUBE ROTATION . . . . . . . . . . . . . . . . . . . . . . 5-7
GENERATOR - HV INVERTER PRINCIPLE DIAGRAM
CONTROL LOGIC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
GENERATOR - HV INVERTER AND HV TANK: PRINCIPLE DIAGRAM . . . . . . 5-9
D750 MASTER BOARD: PRINCIPLE DIAGRAM, AEC . . . . . . . . . . . . . . .5 - 11

6 ______ Generator PC Board Overview ____________________________________6 - 1

D750 MASTER BOARD: CONTACT LOCATION . . . . . . . . . . . . . . . . . . 6 - 1


GENERATOR: CONTACT DIAGRAM . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2

7 ______ Changes to previous version _____________________________________ 7 - 1

Changes to previous version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1

MAMMOMAT NovationDR SPB7-250.844.01 Page 4 of 4 Siemens AG


Wiring Diagrams Rev. 03 08.04 CS SD 24 Medical Solutions
X041E 1-1
SYSTEM OVERVIEW

This page/chapter is intentionally left blank and will be supplied with an update.

Siemens AG SPB7-250.844.01 Page 1 of 2 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
1-2 X041E
SYSTEM OVERVIEW 1

This page/chapter is intentionally left blank and will be supplied with an update.

Siemens AG SPB7-250.844.01 Page 2 of 2 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 2-1
PLACEMENT OF PC-BOARDS AND COMPONENTS, STAND 0

MAMMOMAT NovationDR - Front 0

Tube D805 M6 D810


M1 (Lift), X887

R871
K7-K9 R805
Object table
D807 Right side S1, S2, M
D807 Left side S890
D801
S867
M5 S880
M4:1 S896
M4:2
M4:3 S891
M4:4 X891
D814 S869
Lamp, X894 S897 R861 X874 R863
M2 (Compression) M3 (Rotation)
X888
Strain gauge D802
S894
D804 D808 Right side
D808 Left side
M, S1, S2
R803
S3 S860
S861
X898, R806
Detector array S862 S896 S895 X858
D750

D803
X801
F801-808

Incoming
T802 T801 mains
NOV00118

Fig. 1 Front of the stand

Siemens AG SPB7-250.844.01 Page 1 of 16 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
2-2 X041E
PLACEMENT OF PC-BOARDS AND COMPONENTS, STAND 0

MAMMOMAT NovationDR - Back (seen from the right side) 0

PXCM unit
(Brick unit)

S882
X601

X602

X603

PB0852
HV-tank
PB0851
HV inverter

S881

PB0854
Tube power

PB0855
AC inverter

PB0850
Mains input converter

S883

S884

T803

X885

X804
NOV00127

Z801 X803 PB0853


Mains input

Fig. 2 Back of the stand (seen from the right)

Siemens AG SPB7-250.844.01 Page 2 of 16 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 2-3
PLACEMENT OF PC-BOARDS AND COMPONENTS, STAND 0

18 x 24 object table Flat-panel detector (FD)


(used with Opdima) Object Table
(IRSD image receptor
holding device)
including
Detector array

Stand display

Foot switch

Fig. 3 Stand with rotating support arm

Stand with rotating support arm with one FD (Flat Detector) object table and one
18 x 24 object table (used with Opdima) and Mo/W X-ray tube.

Siemens AG SPB7-250.844.01 Page 3 of 16 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
2-4 X041E
LIST OF BOARDS AND FUSES, STAND 0

PCB Overview 0

D750 MASTER BOARD


D801 CPU BOARD
D802 MOTOR CONTROL BOARD
D803 DISPLAY BOARD
D804 FORCE AMPLIFIER BOARD
D805 WING BOARD
D807 BOARD FOR LIFT, ROTATION AND LAMP SWITCHES
D808 BOARD FOR LIFT AND ROTATION SWITCHES
D810 TILT SWITCH BOARD
D814 COLLIMATOR CONTROL BOARD

Fuses Overview 0

PCB FUSE FUSED VOLTAGE


D750 F1 0.5 A 10.5 VAC
D750 F2 0.5 A 10.5 VAC
D801 F1 2 AT +5V, +5V_REF, +5V_DSP, +7V
D801 F2 1 AT +5V_MPS
D802 F1 5 AT +5V, +5V_M, +14V, D805/+5V
D802 F2 3.15 AT +24VF, +15V
D802 F3 6.3 AT +24V
D802 F5 5AT 17VAC (LAMP)
D802 F4 1 AT 24VACF (FAN)
Chassis-mounted
fuses
F801 2 AT 11 VAC
F802 1 AT 11 V_MPS
F803 5 AT 14 VAC
F804 5 AT 17 VAC
F805 6.3 AT 24 VAC
F806 1AT 18 VAC
F807 2AT 400 VAC
F808 2AT 400 VAC
Brick unit-
mounted fuses
2 AT 230 VAC
2 AT 230 VAC

NOTE! See spare parts list for replacement fuses.

Siemens AG SPB7-250.844.01 Page 4 of 16 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 2-5
LIST OF SWITCHES AND COMPONENTS, STAND 0

Switches 0

Switch Page Description


S1/D801 3-6/1D Reset switch for stand CPU.
S1/Object table 1 3-17/2A Grid out switch located in the object table 1.
S1-1-4/D802 n.a. Current limit switches: 1=+3%, 2=+6%, 3=+12%, 4=+24%.
S1-5/D802 n.a. Tacho On/Off switch for rotation motor drive.
S1-S6/D807 n.a. Control switches for lamp, lift- and rotation movement. Left and right side.
S1-S4/D808 n.a. Control switches for lift- and rotation movement. Left and right side.
S2/Object table 1 3-17/2A Cassette inserted switch.
S860 3-17/3C Stereo lever switch.
S861 n.a. Compression limit switch, compression unit.
S862 n.a. Decompression limit switch, compression unit.
S867 3-16/3G Filter / mirror disk micro switch.
S869 n.a. Tacho optoswitch
S880 n.a. Emergency stop, switches off 24VAC and 14VAC, machine blocked.
S881 n.a. Lift up limit switch, rotating unit.
S882 n.a. Limit switch, distance tube head - floor <50 mm, switches off 24VAC,
machine blocked.
S883 n.a. Collision protection switch, risk of collision when tubehead is less than
120 mm from the floor.
S884 n.a. Lift down limit switch, rotating unit.
S890 3-17/3C AEC in position switch (Wing 1) (Not used).
S891 3-17/3C Object table locked in position (Wing1).
S894 3-17/4C Select wing 1 switch.
S895 3-17/4C Select wing 2 switch.
S896 n.a. Tilt opto-switch. 0° or ±100 used with stereo option.
S897 3-16/2G Diaphragm switch.
S3 n.a. Decompression button.
S1 3-17/1A Grid out switch located in object table 2.
S2 3-17/2A Grid front switch located in object table 2.
S3 3-17/2A Grid rear switch located in object table 2.

Siemens AG SPB7-250.844.01 Page 5 of 16 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
2-6 X041E
LIST OF SWITCHES AND COMPONENTS, STAND 0

Potentiometers 0

Potentiometer Page Description


R803 n.a. Tube angle, 2k.
R861 n.a. Preset force, 10k.
R863 n.a. Thickness, 2k.
R871 n.a. Preset angle, 10k.

Motors 0

Motor Page Description


M 3-17/2A Grid motor, placed in object table with grid.
M1 n.a. Lift motor.
M2 n.a. Compression motor.
M3 n.a. Rotation motor.
M4 :1, :2, :3, :4 3-16/5E Collimator stepping motor.
M5 3-16/4G Filter / mirror disk step motor.
M6 n.a. Fan, tube cooling.
M9 n.a. Rotation brake.
M10 n.a. Rotation brake.

Miscellaneous 0

Misc. Page Description


COMP. UNIT n.a. Compression unit.
GROUND RAIL 3-2/1E System grounding terminal.
K7 n.a. Fan thermostat.
K8 n.a. Tube overtemperature thermostat.
K9 n.a. Tube housing overpressure switch.
L804 - L809 n.a. VHF choke (EMC)
LAMP 3-15/5G Field light lamp.
R805 n.a. Brake resistor, 6.2Ω.
R806 n.a. Brake varistor
R807 n.a. Lift motor varistor.
R860 n.a. Strain gauge (R=300-400 Ω)
T801 n.a. Mains transformer.
T802 n.a. Transformer for PXCM unit (Brick unit)
T803 n.a Transformer for Opdima system
Z801 n.a. Filter for Opdima system
C801 - C803 n.a. EMC capacitors

Siemens AG SPB7-250.844.01 Page 6 of 16 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 2-7
LIST OF SIGNALS AND TEST POINTS, STAND 0

This page is intentionally left blank.

Siemens AG SPB7-250.844.01 Page 7 of 16 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
2-8 X041E
LIST OF SIGNALS AND TEST POINTS,STAND 0

Page Signal Name Origin Via Destination Testpoint LED


n.a. ACT_SPEED TACHO D802 ACT_SPEED
3-17/3C AEC_POS S890 D805 D801
n.a. ANGLE_MEM D807 D805 D801
n.a. BR_OPEN D804 D804.X843 pin 8 V2
n.a. BR_RET D804 D804.X843 Pin 7
n.a. *BRAKE D801 D802
n.a. *BRAKE_RET D802 M9, M10 D802
n.a. *BRAKE_RELAY D802 not used
n.a. C_LIM D802 V11
n.a. C_PWM D802 M2 D802 C_PWM
3-17/3C CASS_LOADED OBJ. TABLE 1 D805 D801
3-16/2F COLL_MOT_INT D814 D805 D801
3-16/2D COLL_MOT_RX
3-16/2D COLL_MOT_TX
n.a. COMP_LIMIT D802 D805 COMP. UNIT
n.a. *COMP_DIR D801 D802 D802 COMP_D
n.a. COMP_FORCE D804 D801 D804.X843 pin 5
n.a. COMP_OK D802 D801 D802 V2
COMP_ERR
n.a. COMP_PROT D802 D801 D802 COMP_PROT
n.a. *COMP_SPEED D801 D802 D801, D802
COMP_SP
n.a. COMPRESS PEDAL D805 D801, D802
n.a. D_PWM D802 D802 D_PWM
n.a. DECOMP_ LIMIT D802 D805 COMP. UNIT
n.a. DECOMPRESS PEDAL / ** D801, D805
n.a. *DIR D801 D802 D802 DIR
n.a. DRIVE_OK D802 D801 D802
V24DRIVE_ERR
n.a. DRIVE_PROT D802 D801 D802 DRIVE_PROT
3-16/1G EXT_COLL S897 D805 D801
3-16/2F FILT_MOT_INT D814 D805 D801
3-16/2D FILT_MOT_RX
3-16/2C FILT _MOT_TX
3-16/4G FM OUT 1A D814 M5
3-16/4G FM OUT 1B D814 M5
3-16/4G FM OUT 2A D814 M5
3-16/4G FM OUT 2B D814 M5
“*SIGNAL” = “SIGNAL” active low
**plus decompression button on control console.

Siemens AG SPB7-250.844.01 Page 8 of 16 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 2-9
LIST OF SIGNALS AND TEST POINTS, STAND 0

Signal Name Description


ACT_SPEED Test point for frequency to voltage converted tacho signal.
AEC_POS AEC-detector in position near breast.
ANGLE_MEM Operator control signal for preset stop angle.
BR_OPEN Bridge open, compression force strain gauge not working correctly.
BR_RET Compression force strain guage return line (appr. 0V).
*BRAKE Control signal to brake reversing relay.
*BRAKE_RET Output to the brakes for max or min braking when active.

*BRAKE RELAY Output to brake reversing relay, max brake if *BRAKE_RET active.
C_LIM Rotation and fast-acting current limit.
C_PWM Pulse Width Modulated power output to compression motor.
CASS_LOADED Cassette inserted in object table 1 (not used).
COLL_MOT_INT Collimator motor interface signal.
COLL_MOT_RX Collimator motor receive signal.
COLL_MOT_TX Collimator motor transmit signal.
COMP_LIMIT Power output via compression limit switch to compression motor.

*COMP_DIR Controls compression motor to compress (moving down).


COMP_FORCE Analog signal indicating compression force.
COMP_OK Compression motor working correctly, no over current.
COMP_PROT Compression relay K4 is working correctly, checked 1s after pedal release.

*COMP_SPEED Pulse Width Modulated signal to control speed of compression motor (20kHz).
COMPRESS Operator control signal to move compression plate down.
D_PWM Pulse Width Modulated power output to rotation or lift motor.
DECOMP_LIMIT Power output via decompression limit switch to compression motor.
DECOMPRESS Operator control signal to move compression plate up.

*DIR Selecting direction of rotation or lift motor.


DRIVE_OK Rotation and lift motor are working correctly, no over current, approx. 1.5 s delay.
DRIVE_PROT DMG relay K6 is working correctly, checked 3s after rot/lift button is released.
EXT_COLL External diaphragm switch, diaphragm in place.
FILT_MOT_INT Filter motor interface signal.
FILT_MOT_RX Filter motor receive signal.
FILT _MOT_TX Filter motor transmitt signal.
FM OUT 1A Filter / Mirror disk step motor M5.
FM OUT 1B Filter / Mirror disk step motor M5.
FM OUT 2A Filter / Mirror disk step motor M5.
FM OUT 2B Filter / Mirror disk step motor M5.

Siemens AG SPB7-250.844.01 Page 9 of 16 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
2-10 X041E
LIST OF SIGNALS AND TEST POINTS, STAND 0

Page Signal Name Origin Via Destination Testpoint


3-6/3B GEN_PREP_RD (VH) D750 D801
3-17/2A GRID_FRONT OBJ.TABLE 2 D805 D801
3-17/3B GRID_MOTOR_N D802 D805 OBJ. TABLE D802 GRID_M
3-17/3A GRID_OUT OBJ. TABLE D805 D801 D801 GRID_O
3-6/2B GRID_RUNNING (AR) D801 D750 D801 AR
3-17/2A GRID_REAR OBJ.TABLE 2
3-17/3G *GRID_SPEED D801 D802 D801 GRID_SP
n.a. I1_OUT D804 D804.X843 pin 6
n.a. K8_TEMP TUBE/K8 D750
n.a. K9_PRESS TUBE/K9 D750
3-15/2A LAMP D807 D805 D801
3-15/4C *LAMP_CTRL D801 D802
3-15/5F LAMP_V D802 LAMP D802 LAMP
n.a. LED D802 TACHO switch
n.a. LIFT_DOWN D807, D808 D805 D801, D802
n.a. LIFT_UP D807, D808 D805 D801, D802
n.a. *MIRROR_MAGNET D802 (not used) D802 MIRROR
3-6/3B MPS D801 D750 D801 MPS
3-16/5C OUT 1A D814 M4:1, :2, :3, :4
3-16/5C OUT 1B D814 M4:1, :2, :3, :4
3-16/5C OUT 2A D814 M4:1, :2, :3, :4
3-16/5C OUT 2B D814 M4:1, :2, :3, :4
n.a. PGM_SPEED D802 D802 PGM_SPEED
n.a. POT_RETURN POT D801
n.a. PRES_FORCE R861 D801
n.a. PRESET_ANGLE R871 D805 D801
“*SIGNAL” = “SIGNAL” active low

Siemens AG SPB7-250.844.01 Page 10 of 16 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 2-11
LIST OF SIGNALS AND TEST POINTS, STAND 0

Signal Name Description


GEN_PREP_RDY (VH) The stand is informed that the generator has completed its preparation.
GRID_FRONT Active when the grid is in front position.
GRID_MOTOR_N Grid motor return line.
GRID_OUT Active when the grid is not in end position.
GRID_RUNNING (AR) Signal given to Master Board D750 as exposure request. If grid used, grid has
started.
GRID_REAR Active when the grid is in rear position.

*GRID_SPEED Pulse Width Modulated signal control speed of grid motor.


I1_OUT Output signal from compression force preamplifier.
K8_TEMP Signal from temperature switch K8 on the tube.
K9_PRESS Signal from the over pressure switch K9 on the tube.
LAMP Operator control signal for the field light lamp.

*LAMP_CTRL Signal to switch on the field light lamp.


LAMP_V Output to the field light lamp.
LED Output to the LED in the tacho opto switch in the rotation motor.
LIFT_DOWN Operator control signal to move compression plate down.
LIFT_UP Operator control signal to move compression plate up.

*MIRROR_MAGNET Output to mirror magnet. Enabled via *LAMP_CTRL. (not used)


MPS Multi processor serial communication for information exchange Master - Slaves.
OUT 1A Collimator stepping motor M4:1, M4:2, M4:3, M4:4.
OUT 1B Collimator stepping motor M4:1, M4:2, M4:3, M4:4.
OUT 2A Collimator stepping motor M4:1, M4:2, M4:3, M4:4.
OUT 2B Collimator stepping motor M4:1, M4:2, M4:3, M4:4.
PGM_SPEED Test point for duty cycle to voltage converted programmed speed signal.
POT_RETURN Return line for potentiometers connected to 0VA via a 20Ω resistor.
PRES_FORCE Analog signal as preset value for maximum compression force.
PRESET_ANGLE Analog signal as preset value for requested rotation angle.

Siemens AG SPB7-250.844.01 Page 11 of 16 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
2-12 X041E
LIST OF VOLTAGES AND TEST POINTS, STAND 0

Page Signal Name Origin Via Destination Testpoint LED


n.a. PU D802 D802 PU
3-16/5C QUAD A D814 M4:1, :2, :3, :4
3-16/5C QUAD A RET (0V) D814 M4:1, :2, :3, :4
3-16/5C QUAD B D814 M4:1, :2, :3, :4
3-16/5C QUAD B RET (0V) D814 M4:1, :2, :3, :4
n.a. *RELEASE D801 D802 D802 RELEASE
3-6/1E *RESET D801
3-16/2C RESET D801 D805
3-6/1E *RESET_HW D801

3-6/1E RESET_SW D801


n.a. ROT_CW D807, D808 D805 D801, D802
n.a. ROT_CCW D807, D808, D805 D801, D802
3-6/3D RXD D801 D801 RXD
n.a. *SEL_ROT D801 D802 D802 SEL_ROT
3-17/3C SELECT_WING1** S891, S894 D805 D801
3-17/4C SELECT_WING2 S895 D805 Object table 2
3-16/1D SMART_RX D805 SMART_RX
3-16/1F SMART_TX D805 SMART_TX
n.a. *SPEED D801 D802 D801, D802 SPEED
3-6/2B STAND_REQ_HV_OFF D801 D750 KVA
(*KVA)
3-17/4C STEREO_LEVER S860 D805 D801
3-17/1B TABLE_CONFIG OBJ.TABLE D805 D801
n.a. TACHO Tacho switch D802 D802 TACHO
n.a. THICKNESS R863 D801
3-11/5E THICKNESS_B THICKNESS D801
n.a. THICKNESS_STE- D801 X885
REO
n.a. TILT S896/D810 D805 D801 D810 V8
n.a. TUBE_ANGLE R803 D805 D801
3-6/4D TXD D801 D750 D801 TXD
n.a. ZERO D804 D804.X843 pin 4
“*SIGNAL” = “SIGNAL” active low.
** = Not used

Siemens AG SPB7-250.844.01 Page 12 of 16 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 2-13
LIST OF VOLTAGES AND TEST POINTS, STAND 0

Signal Name Description


PU Pull up signal for error comparators = +15V, +5V, +24V voltages OK.
QUAD A Collimator stepping motor M4:1, M4:2, M4:3, M4:4.
QUAD A RET (0 V) Collimator stepping motor M4:1, M4:2, M4:3, M4:4.
QUAD B Collimator stepping motor M4:1, M4:2, M4:3, M4:4.
QUAD B RET (0 V) Collimator stepping motor M4:1, M4:2, M4:3, M4:4.

*RELEASE Control signal to release the brakes from min or max to permanent magnet braking.
*RESET Reset signal to the CPU.
RESET Reset signal to the smart point interface.
RESET_HW Reset signal from the CPU to the reset cicuit, caused by slave restart and CPU watchdog
reset.
RESET_SW Signal from the reset switch to the CPU via the reset circuit.
ROT_CW Operator control signal to rotate the x-ray system clockwise.
ROT_CCW Operator control signal to rotate the x-ray system counter clockwise.
RXD Input for receiving data via MPS.

*SEL_ROT Low level signal selects the rotation drive to be controlled.


SELECT_WING1 Signal from select wing 1 switch. Wing 1 selected (not used).
SELECT_WING2 Signal from select wing 2 switch. Wing 2 selected.
SMART_RX Smart point transmit signal.
SMART_TX Smart point receive signal.

*SPEED Pulse Width Modulated signal to control speed of rotation or lift motor (20 kHz).
STAND_REQ_HV_OFF Signal to terminate exposure. WIRED-OR line, see Master board D750.
(*KVA)
STEREO_LEVER Signal from stereo lever switch, indicating lever not in stereo position.
TABLE_CONFIG 4 inputs from object table, see table configuration .
TACHO Signal from rotation motor tachometer.
THICKNESS Analog signal indicating the compression thickness.
THICKNESS_STEREO Isolated analog signal indicating compression thickness for stereo evaluation unit.
TILT Indicates that the stereo angle is 0° or ±10°. Visible through vent opening behind tube.
TUBE ANGLE Analog signal indicating the rotation angle.
TXD Output for sending data via MPS.
ZERO Used for adjusting offset of the compression force amplifiers.

Siemens AG SPB7-250.844.01 Page 13 of 16 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
2-14 X041E
LIST OF VOLTAGES AND TEST POINTS, STAND 0

Page Voltage Origin Supplied from Supply for Test LED Fuse
n.a. +5V D801 11V Digital IC, 5V V7 D801/F1 F801
+5V_REF
n.a. +5V_REF D801 +5V Pots. 5V_REF D801/F1 F801
D804 X843 pin 2
n.a. +5V_DSP D801 11V D803, Display 5V_DSP V6 D801/F1 F801
n.a. +7V D801 11V Analog IC, 7V V5 D801/F1 F801
D804 X843 pin1
3-6/4D +5V_MPS D801 11V_MPS Generator 5V_MPS V26 D801/F2 F802
interface
n.a. 0VD GND D801/X811 Digital IC 0VD1,OVD2
n.a. 0VA 0VD D801 Analog IC 0VA1
D804 X843 pin 3
3-6/5D 0V_MPS D750/GND Generator inter- D801 MPS
face ground
n.a. 11V T801 230VAC +5V, +5V_DSP, D801/F1 F801
+7V
3-6/5B 11V_MPS T801 230VAC +5V_MPS D801/F2 F802
n.a. +5V D802 14VAC Digital IC 5V V57 D802/F1 F803
n.a. +5V_M D802 +5V D801-D802 V57 D802/F1 F803
opto interface
3-2/5C 0V_W D801 D805/X856 Optointerface
3-2/4C 0V_M D801 D802/X822 Optointerface
n.a. +14V D802 14VAC +5V, D805 14V D802/F1 F803
n.a. +15V D802 +24VF Transistor 15V V16 D802/F2
drivers M1,M3
n.a. +24V D802 24VAC +15V D802/F3 F805
n.a. +24VF D802 +24V Switches, 24VF D802/F2
relays, etc
n.a. 14VAC T801 230VAC +14V D802/F1 F803
3-15/5B 17VAC T801 230VAC LAMP 17VACF D802/F5 F804
3-15/4E 17VACF D802 17VAC LAMP 17VAC_RET D802/F5 F804
3-15/4D 0V_L D802 D802 Lamp control
3-15/5D -V_L D802 17VACF Lamp control
n.a. 24VAC T801 230VAC D802/F3 F805
n.a. 24VACF D802 24VAC Fan D802/F4
n.a. 0V GND D802/X821 Ground 0V, 0V1
n.a. +5V D805 D802/+14V Digital IC 5V V17 D802/F1 F803
n.a. +5V_R D805 +5V User switches D802/F1 F803
n.a. +12V D805 +14V V_MOT D814 12V D802/F1 F803

Siemens AG SPB7-250.844.01 Page 14 of 16 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 2-15
LIST OF VOLTAGES AND TEST POINTS, STAND 0

Voltage Description ( typical value)


+5V Regulated DC voltage, 4.75V < +5V < 5.25V
+5V_REF DC voltage, +5V -5% < +5V_REF < +5V
+5V_DSP Regulated DC voltage, 4.75V < +5V_DSP < 5.25V
+7V Regulated DC voltage, 6.60V<+7V<7.25V
+5V_MPS Regulated DC voltage, 4.75V < +5V_MPS < 5.25V
0VD Digital ground
0VA Analog ground
0V_MPS Generator interface ground.
11V AC supply voltage, 11VAC 1) 2)
11V_MPS AC supply voltage, 11VAC 1) 2)
+5V Regulated DC voltage, 4.75V < +5V < 5.25V
+5V_M Regulated DC voltage, 4.75V < +5V_M < 5.25V
0V_W Digital ground optointerface D801 - D805
0V_M Digital ground optointerface D801 - D802
+14V Unregulated DC voltage, 18V 1)
+15V Regulated DC voltage, 14.25V < +15V < 15.75V
+24V Unregulated DC voltage, 32V 1)
+24VF Unregulated DC voltage, 32V 1)
14VAC AC supply voltage, 14 VAC 1) 2)
17VAC AC supply voltage, 17 VAC 1) 2)
17VACF AC supply for lamp 1)2)
0V_L Floating ground for LAMP voltage regulation 1)2)
-V_L Unregulated supply voltage for lamp control circuit 1)2)
24VAC AC supply voltage, 24VAC 1) 2)
24VACF AC supply voltage for the fan 1) 2)
0V Circuit ground
+5V Regulated DC voltage, 4.75V < +5V < 5.25V
+5V_R Regulated DC voltage, 4.40V < +5V_R < 5.25V
+5V_SP Regulated DC voltage, 4.85V < +5V < 5.15V
8.2V_SP Regulated DC voltage, 7.8V < +8.2V < 8.6V
1) Standby typical value
2)
Floating voltage. Measure differentially.

Siemens AG SPB7-250.844.01 Page 15 of 16 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
2-16 X041E
LIST OF VOLTAGES AND TEST POINTS, STAND 0

NOTE For incoming mains voltage, see page 4-1.

Siemens AG SPB7-250.844.01 Page 16 of 16 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E
Medical Solutions
Siemens AG
A B C D E F G

OBJECT TABLE 1 COLLIMATOR DETECTOR ARRAY OBJECT TABLE 2 MAG BOARD


STEREO OPTION
CPL
X608

1
X885
X847 X845
X846

CONTROL CONSOLE X894 X832


Rev. 03 08.04
SPB7-250.844.01

X893 X847
X858 X849
X804

BLOCK DIAGRAM CABLE CONNECTIONS


X830 X869 X831
FOOT SWITCHES X871 X872
X839 X873

2
X857
WING BOARD X860
CS SD 24
Page 1 of 17

X858 D805 X855


X803 X859 X856
X848 COMPRESSION UNIT

TRANSFORMER X874
T801
X886

TUBE
TRANSFORMER
T802 X802

3
X895
X896
M2 M1 X897
TRANSFORMER
T803 DISPLAY
230 VAC ROTATION LIFT
Z801 D803 X826
T803 X888 X887
To Opdima
Filter X801

X803

X753
X755
D750

4
X756
Wiring Diagrams
Mammomat Novation

MASTER BOARD X757 X822 X812 X809


X761 X811
X768 X821 D801 X813
X823 D802 X814
X760 X752 X824 MOTOR CONTROL CPU BOARD X815
X825 X816

Incoming
DR

mains X601 X604


X707 X602
X701 X713 X721 X725 X603
PB0853 X726 X605
PB0850 PB0854 PB0852

5
MAINS INPUT X706 X710 PB0855 X711 X723 X720 X731 PB0851 X730 X6
MAINS INPUT CONVERTER AC INVERTER TUBE POWER HV INVERTER HV TANK PXCM
(BRICK)
X705 X700 X712 X722 X724 X732

3-1
0
Medical Solutions
Siemens AG
A B C D E F G

3-2
INCOMING MAINS

GROUNDING CONNECTIONS
P.E. STAND

1
GROUND RAIL
GENERATOR
GROUND RAIL
Door CONTROL PANEL
X803 CURTAIN
D741 8
HV D740
Common
Tank X32 X5 X30
Rev. 03
SPB7-250.844.01

1 2 6 8 10
X6 X5
3 3 6 3 UPPER X896
H3

X730 HV Inverter
1 2 6 8 10 D-subplate X897
08.04

17 1 5 6 7 12 18 19 TUBE REAR FRONT


0V_sec CHASSIS COVER COVER
LIFT
TUBE
2 4 8 10 16 20 14 D810 CARRIAGE COMPRESSION
X731 PLATE
X81

2
0V UNIT
Page 2 of 17
CS SD 24

TUBE
AC X713
17 1 5 PATIENT PATIENT HOUSING
Inverter X804 6 7 12 18 19
LOWER HANDLE HANDLE X895

X720 Tube K8
2 4 8 10 16 20 14 Power 3
Supply 5
X726
6 X883 K9
7 4
4
6 Ferrit sleeve
0V_sec
X725
0VA
X758 X755 X891 X1 X2 X893 X885
X721 3 4 8 1 22 22
8 3
10 2 23 23
X722 2 4 6 8 4 X756
24 24
5 X768

3
4 2 5 8 X860 8 X859 25 25
7
X752 X821
8
6 X811 6 7
X700 Mains Input X871
10 D802 D805
Conv. D803 D801 X812 X822 WING 1
X701 X760 1 1 MOTOR WING
2 4 6 8 DISPLAY CPU
6 6 2 2 BOARD BOARD
0V A 0V D
0V_sec 8 8 X801 X816 3 3 MAG TABLE
15 2
15 4 4 X823
10 10
0VA 0V A 16 16 5 5 8 5 WING 2 Frame left
12 12
17 17 6 6
D750 18 18
0VD 16
8 8 0V Detector
MASTER X855 Beam
X765 0V_M 0V
0V_RS232 BOARD 0V D 0V D 12
3 SHLD_1
6 X757 X815
9 1 1 X813 X856 Frame right
10 3 3 1 27

4
7 7 2 25 X846
Wiring Diagrams
Mammomat Novation

Service 0V D
X760 8 8 33 29
PC 16 34 30
0V_RS232 X761 X830
X602 X603 0V Filter/mirror
X762 2 14 E
COMPRESSION 2 2
8 0V_MPS
8 17
0V_W
UNIT 9 9 disk switch
5
5 10 18 PRES_FORCE
D804 X838
X805 GND_S_PC COMPRESSION 0V
12 19 X814 R861 R863 THICK FORCE
X764 2 S867
A 24 25 14 10k 2k NESS TUBE_ANGLE D814
X874 X841
B
GND_PTR
20 23 16 7 2 Collimator

0
0V A R803
DR

5 22 24 15 2 contr. board
0VI_RS232 17 PXCM X868 2k
5 PRESET_ANGLE
(BRICK) 6
X601 6 NOTE
X763 R871
5 X809 X855 DOTTED CONNECTIONS INDICATE
13 SHLD 10k
X769 3 SHIELDED CABLES WITHOUT
8 7
POT_RET 0V CONNECTION

5
9 4 X863
20 17 5 1
13 0V_A
GND_OPD_WS 0V_RS232

X041E
X041E 3-3
MAINS CONNECTION, POWER SUPPLY CPU 0
1 2 3 4 5
This page/chapter is intentionally left blank and will be supplied with an update.
G
F
E
D
C
B
A DR
Siemens AG SPB7-250.844.01 Page 3 of 17 Mammomat Novation
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
3-4
X041E
POWER SUPPLY MOTOR CONTROL 0
1 2 3 4 5
This page/chapter is intentionally left blank and will be supplied with an update.
G
F
E
D
C
B
A DR
Siemens AG SPB7-250.844.01 Page 4 of 17 Mammomat Novation
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 3-5
DISPLAY 0
1 2 3 4 5
This page/chapter is intentionally left blank and will be supplied with an update.
G
F
E
D
C
B
A DR
Siemens AG SPB7-250.844.01 Page 5 of 17 Mammomat Novation
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
Medical Solutions
Siemens AG
A B C D E F G

3-6
D801

CPU BOARD INTERNAL, GENERATOR INTERFACE


1
RESET S1
COMP
*RESET SW
+5V <4.6V *RESET D801 TESTPOINTS AND LED’s
0VD
AR
*RESET_HW
Rev. 03
SPB7-250.844.01

KVA X812
COLL_ST
GRID_SP
COMP_SP X813
*RESET
V60 SPEED
D750
08.04

FILT_P
MASTER
*KVA STATUSLED’s FOCUS_P
J1 ROT_POS
COLL_P2

2
V1
X757 STAND_REQ_HV_OFF X815 COLL_P1
Page 6 of 17
CS SD 24

12MHz X1
6 ( KVA) 6 0VD2
* V2
PROM V60
X816
V3 V1
*AR V2
PLD
CPU & V4 V3
GEN_PREP_RDY PLD V4
5 (VH) 5 FUNC. GRID_0
GRID_RUNNING TXD ROM
DATA AND ADDRESS
4 (AR) 4 CPU J1
RXD
0VA1
1 1
7V
3 X814
3 0V_MPS 0VD1 V5

3
0V_MPS 0VA
7 7
PROM RAM EEPROM 5V_REF
8 8 0VD 128k x 8 32k x 8 8k x 8
+5V_MPS
RXD 5V_DSP
I9
GND 0V_MPS V6
X811
MPS 0VD1 F1
2 5V
2 MPS
COMMUNICATION
V7

TXD
ROM MPS
X815

4
0V_MPS V26 F2
Wiring Diagrams
Mammomat Novation

5V_MPS

S1
5V_MPS
RESET SWITCH
X811
1 F2
~ V26 0V_MPS
+5V_MPS

0
2 1AT
DR

= 0V_MPS
STATUS LED’s DURING STARTUP:
X886
PROM/RAM TEST V1, V4 LIT.
5 6 GENERATOR INTERFACE

5
FLOATING, GROUNDED BY GENERATOR V2, V3 LIT IF PROM CHECKSUM ERROR.
F802
1AT
V3, V4 LIT IF RAM ERROR.

X041E
X886A
11V_MPS

5 6
ALL TESTS PASSED, V1-V4 ARE DARK.
BN BU
V60 ~ 1HZ WHEN PLD IS PROGRAMMED
T801
X041E 3-7
TUBE, TEMPERATURE MONITORING 0
1 2 3 4 5
This page/chapter is intentionally left blank and will be supplied with an update.
G
F
E
D
C
B
A DR
Siemens AG SPB7-250.844.01 Page 7 of 17 Mammomat Novation
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
3-8
X041E
LIFT AND ROTATION 0
1 2 3 4 5
This page/chapter is intentionally left blank and will be supplied with an update.
G
F
E
D
C
B
A DR
Siemens AG SPB7-250.844.01 Page 8 of 17 Mammomat Novation
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 3-9
LIFT AND ROTATION MOTOR DRIVE 0
1 2 3 4 5
This page/chapter is intentionally left blank and will be supplied with an update.
G
F
E
D
C
B
A DR
Siemens AG SPB7-250.844.01 Page 9 of 17 Mammomat Novation
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
3-10
X041E
LIFT AND ROTATION SWITCHES 0
1 2 3 4 5
This page/chapter is intentionally left blank and will be supplied with an update.
G
F
E
D
C
B
A DR
Siemens AG SPB7-250.844.01 Page 10 of 17 Mammomat Novation
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 3-11
LIFT AND ROTATION SWITCHES 0
1 2 3 4 5
This page/chapter is intentionally left blank and will be supplied with an update.
G
F
E
D
C
B
A DR
Siemens AG SPB7-250.844.01 Page 11 of 17 Mammomat Novation
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
3-12
X041E
COMPRESSION MEASUREMENT SIGNALS 0
1 2 3 4 5
This page/chapter is intentionally left blank and will be supplied with an update.
G
F
E
D
C
B
A DR
Siemens AG SPB7-250.844.01 Page 12 of 17 Mammomat Novation
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 3-13
COMPRESSION MOTOR DRIVE 0
1 2 3 4 5
This page/chapter is intentionally left blank and will be supplied with an update.
G
F
E
D
C
B
A DR
Siemens AG SPB7-250.844.01 Page 13 of 17 Mammomat Novation
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
3-14
X041E
TUBE ANGLE MEASUREMENT, POT. RETURN, TILT SIGNALS 0
1 2 3 4 5
This page/chapter is intentionally left blank and will be supplied with an update.
G
F
E
D
C
B
A DR
Siemens AG SPB7-250.844.01 Page 14 of 17 Mammomat Novation
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E
Medical Solutions
Siemens AG
A B C D E F G

COLLIMATOR 1
D805 D801 D802

1
LEFT SIDE BOARD D805
X865

12V
D807 ** X812 X822

5V
0V
X807 X851 X851
1 1 COLL_ST
+5V_R X813
X839 MIRROR
Rev. 03 08.04
SPB7-250.844.01

S6 X856
FILT_P
LAMP 7 7 LAMP 7
FOCUS_P
LAMP ON X807 X865
D807 ** 1 1 COLL_P2
COLL_P1
S6

5V_SP
8.2V_SP

SMART_TX
5V_SP
LAMP 7 7 LAMP
X823

2
X831 SMARTPOINT
Page 15 of 17
CS SD 24

SMART
RIGHT SIDE BOARD SMARTPOINT
1 INTERFACE 17VACF
POINT
INTERFACE 2
MEM

SMART_RX
X860
4 TAG 17VAC_RET

8.2V_SP
+24V F5 X821
X856 X831

X869

X860

3
D801 D802 +24VF X823
X812 X822
1 12
2
3
X813
13
CPU & 4 MIRROR
LAMP 5
PLD
6
FUNC. 8

0V_M 0V
X812 X822 0V

4
*LAMP_CTRL
Wiring Diagrams
Mammomat Novation

12 12 0V

17VACF 0V
X821
V52
7 F5

5AT
COMPRESS 29
SET LAMP_V
R131
SEE PAGE 3-13
0V_L 0V_L

0
DR

COLLIMATOR

Phase
-V_L + angle LAMP
RMS X823 X894
control
/DC 16 1 1

5
URMS angle LAMP_V

3-15
-V_L L1 LAMP
X886A X886
YE 11 F804 11 0V_L 17VAC_RET

5AT 9 18 3
2
T801 17VAC
12 12
OG
3-16
Medical Solutions
Siemens AG
A B C D E F G

COLLIMATOR 2
X813
D801 D805

1
SMART RX

SMART TX
1 15 EXT COLL CPU & PLD
FUNCTION

2 26 COLL MOT INT X813 X856 X855 EXTERNAL


SMART TX 19 26 SMART RX SMART TX 7 6 COLLIMATOR
Rev. 03
SPB7-250.844.01

SMARTPOINT INTERFACE
15 1
3 6 FILT MOT INT S897
X855 2 X839
RESET 20 11 RESET
1
+5V
08.04

4 24 COLL MOT RX
X855
COLL MOT TX 21 21 COLL MOT RX COLL MOT INT 16 2

FILT MOT INT 14 3

2
5 12 FILT MOT RX
Page 16 of 17
CS SD 24

COLL MOT TX 13 4
FILT MOT TX 22 35 FILT MOT RX
FILT MOT TX 8 5
6 5 SMART RX
GRID FRONT X856/19 7

X869/9
7 32 GRID FRONT +5V 0V SEE PAGE 3-17

0V +12V

X830 1 2 4 7 6 3 8 5 9 10

X832

3
D814 +5V
+5V 0V 0V V MOT FILTER/MIRROR
DISK MICROSWITCH
X838 S867
1
STEP MOTOR CONTROL
2
0V

X837
* Step motor M4:1 drives the right collimator plate, 1 FM OUT 1A

4
STEP MOTOR DRIVE
Wiring Diagrams
Mammomat Novation

Step motor M4:2 drives the left collimator plate, 2 FM OUT 1B


Step motor M4:3 drives the rear collimator plate,
Step motor M4:3 drives the front collimator plate,
all as viewed from the patient position.

Pin position for each collimator motor connector: 3 FM OUT 2A


4 FM OUT 2B M5
1: OUT 1A
2: OUT 2B FILTER/MIRROR

0
3: QUAD B RET (0V) DISK STEP MOTOR
DR

4: QUAD B
5: QUAD A
6: QUAD A RET (0V) X834
X836 X835 X833
7: OUT 1B
8: OUT 2A * * * *

5
OUT denotes motor winding signals, M4:4 M4:3 M4:2 M4:1
QUAD denotes quadrature motor signals.

X041E
X041E
Medical Solutions
Siemens AG
A B C D E F G
X849
OBJECT TABLE 1
OBJECT TABLE 2 X869
+24VF 2 10 D805 CONNECTOR
D801
11
CPU &
REAR VIEW
PLD
S1 M

OBJECT TABLE 1 + 2
FUNC. 1 3
GRID_OUT

1
GRID_MOTOR_N 4 7
1 8 X856 X813
1 TABLE_CONFIG_1 10
6
3 3 TABLE_CONFIG_3 3 1 2 TABLE_CONFIG_2 9
7 9
X847
X845 TABLE CONFIG 6 3 TABLE_CONFIG_3 14
Rev. 03 08.04
SPB7-250.844.01

MAG TABLE (SEE BELOW)


2 2 TABLE_CONFIG_2 2 4 TABLE_CONFIG_0 7
1.8 X845
GRID_O
1 TABLE_CONFIG_1 1 5 CASS_LOADED 4
1
1.5 X846 0 4 TABLE_CONFIG_0 4 8 GRID_OUT 16 MAG TABLE

3 9 0VD2 (+24V) X845


GRID_FRONT RED
GRID_REAR
S3 X871
1 1.8 GREY X845
0VD

2
GRID_FRONT 19 GRID_FRONT 32 1.5 BLUE X846
Page 17 of 17
CS SD 24

S2
GREEN/YELLOW X846
OBJECT TABLE 1 +24VF 6 AEC_POS 3
X891 X857
25 1
S2 1 TABLE_CONFIG_1 1 1
CASS 1
TABLE 27 2
LOADED
CONFIG 2 2 TABLE_CONFIG_2 2 29 33 GRID_SP
(SEE 30 34
BELOW) 3 3 TABLE_CONFIG_3 3
4 TABLE_CONFIG_0 20KHz
0 4 0V
0V_W PWM
STEREO_LEVER 23
S1 5 CASS_LOADED 5
GRID_OUT
M 6 7

3
8 GRID_OUT 8 SELECT_WING_1 11
9
6 AEC_POS
S890 18
GRID_MOTOR_N 11
X812 19
S891 *GRID_SPEED
7 10 X860 GRID_MOTOR_N
9 X859 1 7

X868 8 X822 D802


MOTOR AND GRIDOUT SWITCH SEE PAGE 10
+24VF 19
OMITTED IF NO GRID 3-11 GRID_M
S860 +24VF
X893 1 25 1 STEREO_LEVER
TABLE CONFIGURATION
S895
TABLE CONFIG TABLE TYPE 3 SELECT_WING_2 K1
X885 X824

4
X856
Wiring Diagrams
Mammomat Novation

3 2 1 0 1 12
S894 11 +24VF
NC NC NC NC NO TABLE LOCKED IN PLACE 2 SELECT_WING_1 0V GRID_MOTOR_N GRID_SP
STEREO
NC NC NC 24V AN: 18X24 WITH GRID (OPTION)
25
NC NC 24V NC AN: 24X30 WITH GRID +24VF
NC NC 24V 24V AN: 18X24 WITHOUT GRID
D801 D802
NC 24V NC NC AN: 24X30 WITHOUT GRID K8

NC 24V NC 24V AN: MAGNIFICATION 1.5 D805


X822

0
NC 24V 24V NC AN: STEREOTAXI (18x24) GRID_SP
DR

X812
NC 24V 24V 24V AN: MAGNIFICATION 1.8 0V 0V
GRID_SP X824
24V NC NC NC DI: NORMAL (NO MAG) X813
X868
24V NC NC 24V DI: INVALID (TRANSITIONAL)
X855 1
GRID_M A 0V
X875 X876 X869

5
24V NC 24V NC DI: INVALID (TRANSITIONAL) 0V

0V

-1.25
X856 0V
24V NC 24V 24V DI: MAGNIFICATION 1.5
0V X857 X872
24V 24V NC NC DI: INVALID (TRANSITIONAL)

UREG
0VD2 0V
X871

3-17
24V 24V NC 24V DI: MAGNIFICATION 1.8

U
X860 X859
24V 24V 24V NC DI: INVALID GRID_0 X873
0V

I
24V 24V 24V 24V DI: INVALID

0V
X041E 4-1
MAINS VOLTAGE CONNECTION 0

The mains input can be either 1-phase or 2-phase, 50 or 60 Hz.

Electrical power to mains input: 0

Mains voltage 1-phase: 208, 230, 240 and 277 V ±10%


or
2-phase: 208, 230, 240, 277 and 400 V ±10%
Mains Frequency: 50 or 60 Hz
Maximum load: 10kVA (momentary) and 0.8 kVA (long time load).
Fuse: 25 A (external) 208 -> 400 V.

Power line impedance: 0

The generator section of the mammomat requires a line impedance not exceeding the
maximum values according to the table below:

Nominal input voltage [VAC] Rimax[Ω]


208 0.45
230 0.50
240 0.60
277 0.65
400 0.85

Connections: 0

Power system Type of Power voltage Power


connection connections
Singel phase Phase to neutral 208 V-277 V Connect to 3 and 4
Two phase Phase to phase 208 V-400 V Connect to 3 and 4.
phase to phase. Use two phases (of three) in a
Note: Max 277 V normal power distribution sys-
phase to neutral/ tem.The left over wires, neutral
ground allowed. and the third phase may be con-
nected to terminal 1 and 2. See
Fig. 1.

X707

PE 1 2 3 4
NOV00123

Note 1. PE = Protective Earth


Note 2. Terminals 1 and 2 are not connected internally. Can be used to secure wires when four or five
wire power system is used.
Fig. 1

Siemens AG SPB7-250.844.01 Page 1 of 10 MAMMOMAT NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
4-2 GENERATOR: LIST OF MODULES AND FUSES X041E
PC-BOARDS: MASTER BOARD D750 0

Modules: 0

PB0853 Mains input (Filter unit)


PB0850 Mains input converter
PB0855 AC inverter
PB0854 Tube power supply
PB0851 HV inverter
PB0852 HV Tank

Fuses: 0

• F2 8A/660VAC

Placed inside PB0850 - Mains input converter


Type: Ferraz Shawmut - 660GRB 10-08 / A070GRB08T13 or equivalent.
• F1 5A/660VAC

Placed inside PB0850 - Mains input converter


Type: Ferraz Shawmut - This fuse is used to protect the auxiliary board at unwanted
failure, it does not protect internal components and is not intended to be a spare part.

PC-boards: 0

Master board D750

Siemens AG SPB7-250.844.01 Page 2 of 10 MAMMOMAT NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 4-3
LEFT BLANK 0

This page is intentionally left blank. 0

Siemens AG SPB7-250.844.01 Page 3 of 10 MAMMOMAT NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
4-4 GENERATOR: PLACEMENT OF MODULES, X041E
CONNECTORS, LEDs AND POTENTIOMETERS 0

Generator description of MAMMOMAT NovationDR 0

PXCM unit
(Brick unit)

PB0852
HV Tank

X5, X6, X7, X8 X1, X2, X3

X740 mAs-jumper
PB0851
X730 HV inverter

LED:
HV_ON

Potentiometers:
I_OUT X732
HV_READ X725
HV_WISH X726
0VA

X731 PB0854
X720 Tube power
Potentiometers:
Bias Mo Potentiometer:
Bias W Anode drive frequency
Testpoints:
0_VA X724
FIL_CURR_READ X713
MA_READ
PB0855
X721 AC inverter
X722
X723 X712
X705
LED:
AC_ON Fuse F2: Inside the module
X706
X710
X711 PB0850
Mains input converter
LEDs:
28V_prim PB0853
5V_RS232 Mains input
-15VA (Filter unit)
+15VA
28V_sec
5VD
NOV00122

X700 LEDs X701 S700 X707 LEDs


DC_BUS_OK H3
DC_BUS_FAIL H4

Fig. 2 Back of the stand (seen from the left)

Siemens AG SPB7-250.844.01 Page 4 of 10 MAMMOMAT NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 4-5
CONTROL PANEL FUNCTIONS 0

15 16 17 18 19 20 21 22

Mo Mo W
kV mAs / mGy AEC Mo Rh Rh Program Auto Lim.
WARNING

NOV00124
1 2 3 4 5 6 7 8 9 10 11 12 1 13 14
Fig. 3

1 Exposure release buttons with green lights. Both buttons must be pressed during the entire exposure. This is
to ensure that the operator is standing behind the radiation shield during the exposure.
2 +/- kV selection and display
3 MAMMOMAT NovationDR ON/OFF
4 +/- mAs selection and display. In AEC mode (Automatic Exposure Control), the display remains dark before
exposure. After the exposure, the mAs value reached during radiography is indicated.
If the dose calculation option is enabled, the dose value is shown on the same display as the mAs value.
5 “dr” (digital radiography) is displayed on the film density display to indicate that the MAMMOMAT NovationDR
is ready to be used. If the display is dark the communication with the DR-system is not working. The buttons
are not used.
6 AEC. These buttons are used to activate AEC mode (Automatic Exposure Control). H stands for low dose
exposure and D for high dose exposure.
7 Buttons for selection of anode/filter combination in manual mode.
8 Buttons for selection of one of four pre-programmed exposure parameter settings.
9 Decompression button.
10 Button for storing exposure parameter (see pos 8).
11 Button to select/deselect auto mode.
12 Button to select/deselect automatic decompression after exposure. When off, use foot switch or the knobs for
manual decompression or use the decompression button (see pos 9).
13 Limit button (acknowledgement). Serves to reset the system after errors, indicated by an error code, or faulty
exposures (limit errors), indicated also by the limit lamp and acoustic signal (e.g limit mAs, dose monitoring,
interrupted exposure).
14 Limit lamp. Flashing: Tube temperature is above normal. Lit: Limit error (see pos 13).
15 Not used for digital mammography.
16 Indicates an unknown table configuration. e.g. magnification table not fully inserted.
17 Risk of collision between tube head and floor (limit switch is activated).
18 Not used for digital mammography.
19 Collimator not in correct position or wrong combination of object table and compression plate.
20 Focal spot indication. Indicates focus in operation. Small focus is automatically selected, when a magnifica-
tion table is used.
21 Green light indicates that the unit is ready for exposure. This indicator has the same function as the green
lights on the stand display and the exposure release buttons.
22 Yellow light and an acoustic signal indicate radiation ON.

Siemens AG SPB7-250.844.01 Page 5 of 10 MAMMOMAT NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
4-6 D750 MASTER BOARD: X041E
LIST OF SWITCHES, TEST POINTS 0

Switches 0

Switch Page Description


S2 6-1 HV disable; HV = High Voltage
Upper position = Normal mode
Lower position = HV disable mode
S3 6-1 Manual activation of Power aid unit
S4 6-1 Manual reset of AEC processor

Test points
Test point Page Description
5VD_1 6-1 Logic supply voltage: 5 volt digital
0VD_1,2,3,4,5 6-1 Logic supply voltage: 0 volt digital
P15VA 6-1 Logic supply voltage: Positive 15 volt analog
0VA_1,2,3 6-1 Logic supply voltage: 0 volt analog
N15VA 6-1 Logic supply voltage: Negative 15 volt analog
5V_RS232 6-1 Logic supply voltage: 5 volt; for communication with PXCM unit
(Brick unit)
0V_RS232 6-1 Logic supply voltage: 0 volt; for communication with PXCM unit
(Brick unit)
5VI_RS232 6-1 Logic supply voltage: 5 volt; for communication with Service-PC,
Opdima WS and Printer
0VI_RS232 6-1 Logic supply voltage: 0 volt; for communication with Service-PC,
Opdima WS and Printer
M_WD_RESET_ 6-1 Signal that resets the master CPU and thus causes the entire Mam-
momat system to restart.
GEN_POWER_ON 6-1 Signal to the generator that turns on the generator and thus the
Mammomat system.
DC_BUS_OK 6-1 Signal from the generator, indicating the DC-bus voltage is within its
allowed voltage window.
HV_OFF_INT_ 6-1 Indicates that the signal M_PLD_HV_OFF_ from the PLD circuit
has detected an event that shall terminate or prevent high voltage
generation.
It causes the signal HV_ON from the Master board to go inactive,
which immediately prohibits high voltage generation.
REQ_HV_ON 6-1 Signal from PLD and CPU for generating high voltage and thus start
the exposure.
It will together with the signal GRID_O_HV_ON set the signal
HV_ON.
HV_ON 6-1 Signal set by the signals REQ_HV_ON and GRID_O_HV_ON out-
putted to the generator. One of two necessary signals, together with
the signal EN_HV_ON, to generate high voltage and thus start the
exposure.
This signal goes low at grid returns, due to GRID_O_HV_ON, thus
causing the intermittent loading mode during exposure.
EN_HV_ON 6-1 Signal from the exposure release buttons at the Control panel to the
generator. One of two necessary signals, together with the signal
HV_ON, to generate high voltage and thus start the exposure.

Siemens AG SPB7-250.844.01 Page 6 of 10 MAMMOMAT NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E D750 MASTER BOARD: 4-7
LIST OF SWITCHES, TEST POINTS 0

Test point Page Description


GEN_POWERED_ON 6-1 Signal to the PXCM unit (Brick unit), that the power to the
Mammomat system has been switched on by the operator.
PREP_DR_DET_ 6-1 Signal to the PXCM unit (Brick unit).
The positive edge indicates to the Brick that the generator preparation
phase for making an exposure has started and requests that the DR detec-
tor shall prepare for image acquisition.
The negative edge of the signal comes 5 ms after the negative edge of the
signal GEN_RDY_FOR_DR_EXP at normal exposure termination, in case
the hardware detects an error the negative edge of the signals will coincide.
If a tube arc occur the signal will not go inactive.
GEN_RDY_FOR_DR_EXP 6-1 Signal to the PXCM unit (Brick unit).
The generator preparation phase has finished and the generator is ready for
generating high voltage to the tube (i.e. x-ray).
The inactivation of the signal will coincide with negative edge of the signal
X_RAY_ON at normal termination of the exposure. In case of an abnormal
termination of the exposure, the inactivation of the signal coincides with the
inactivation of the signal PREP_DR_DET.
If a tube arc occur the signal will not go inactive.
X_RAY_ON 6-1 Signal to the PXCM unit (Brick unit). The exposure is in progress.
Note: The signal stays active
- at the grid turning points when x-ray actually is off.
- at the first two tube arcs when x-ray actually is off.
DR_DET_RDY_FOR_EXP 6-1 The DR detector is making image data acquisition.
DR_SYST_RDY 6-1 The DROC of the DR system indicates that it is ready for an
exposure.
DR_DET_ERR_ 6-1 The DR system indicates to the generator that it has detected an error.
HV_SET 6-1 Analog output signal to the generator for setting the magnitude of the tube
voltage (high voltage).
The relation between the tube voltage and the analog control signal is 5kV/V.
HV_ACT 6-1 Analog input signal HV_ACT, from the generator, that is proportional to the
true tube voltage.
The relation between the tube voltage and the analog signal is 5kV/V.
MA_SET 6-1 Analog output signal to the generator for setting the magnitude of the tube
current.
The relation between the tube current and the analog control signal is
40mA/V.
MA_ACT 6-1 Analog input signal MA_ACT, from the generator, that is proportional to the
true tube current.
The relation between the tube current and the analog signal is 40mA/V.
FIL_CURR_SET 6-1 Analog output signal to the generator for setting the magnitude of the fila-
ment current.
The relation between the filament current and the analog control signal is
1A/V.
FIL_CURR_ACT 6-1 Analog input signal FIL_CURR_ACT, from the generator, that is proportional
to the true filament current.
The relation between the filament current and the analog signal is 1A/V.

Siemens AG SPB7-250.844.01 Page 7 of 10 MAMMOMAT NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
4-8 D750 MASTER BOARD: X041E
LIST OF SWITCHES, TEST POINTS 0

Test point Page Description


ACC_ANODE 6-1 Signal to the generator to start the anode rotation by accelerating
the stator of the tube. The acceleration power (AC current) to the
tube continues as long as the bit is set.

Note, the generator stops the acceleration power to the tube if this
signal is set longer than about 2.9 sec.
BRAKE_ANODE 6-1 Signal to the generator to stop the anode rotation by braking the
stator of the tube. The braking power (DC current) to the tube con-
tinues as long as the bit is set.

Note, the generator stops the braking power to the tube if this signal
is set longer than about 7.4 sec.
ANODE_FREQ 6-1 Signal from the generator for measurement of the anode rotation
speed. The frequency of the signal is equal to the anode rotation
frequency.

Note, PLD measurement of this signal starts 200 ms after the signal
ACC_ANODE goes low, mean value of 10 pulses is calculated.

Siemens AG SPB7-250.844.01 Page 8 of 10 MAMMOMAT NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E GENERATOR: LIST OF SWITCHES, 4-9
POTENTIOMETERS, TEST POINTS, LEDs 0

Switches: 0

Switch Page Module Description


S700 4-5 Mains input On/Off circuit breaker for mains input, inside the covers. The
(Filter unit) breaker is intended for service/maintenance as an extra possibility
to turn off the mains for safety at work with the unit.
It shall be in On position when the covers are mounted. The breaker
also protects the generator, it trips at overcurrent (30A) plus has a
trip coil used for DC-bus overvoltage protection.

Potentiometers 0

Potentiometer Page Module Description


Bias Mo 5-6 Tube power Adjustment of bias voltage for Molybden (Mo).
supply
Bias W 5-6 Tube power Adjustment of bias voltage for Tungsten (W).
supply
Anode drive freq. 5-7 Tube power Adjustment of drive freq. 3 - 5 Hz from its nominal value, for
supply decreasing noise that can occur due to mechanically vibrations
from the tube assembly.

Test points 0

Test point Page Module Description


0VA 5-6 Tube power Logic supply voltage: 0 volt analog
supply
Note: 0VA = 0V_sec
FIL_CURR_READ 5-6 Tube power Analog signal proportional to the true filament current.
supply The relation between the filament current and the analog signal is
1A/V.
MA_READ 5-6 Tube power Analog signal proportional to the true tube current.
supply The relation between the tube current and the analog signal is
40mA/V.
I_OUT 5-8 HV inverter Analog signal proportional to the output inverter current fed to the
HV tank.
HV_READ 5-8 HV inverter Analog signal proportional to the true tube voltage.
The relation between the tube voltage and the analog signal is
5kV/V.
HV_WISH 5-8 HV inverter Analog control signal for setting the magnitude of the tube voltage
(high voltage).
The relation between the tube voltage and the analog control signal
is 5kV/V.
0VA 5-6 HV inverter Logic supply voltage: 0 volt analog

Note: 0VA = 0V_sec

Siemens AG SPB7-250.844.01 Page 9 of 10 MAMMOMAT NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
4-10 GENERATOR: LIST OF SWITCHES, X041E
POTENTIOMETERS, TEST POINTS, LEDs

LEDs: 0

LED Page Module Description


28V_prim 4-5 Mains input Logic supply voltage: 28-30 volt, only to the generator modules.
converter
Note: The voltage is hazardous.
5V_RS232 4-5 Mains input Logic supply voltage: 5 volt; for communication with PXCM unit
converter (Brick unit)
-15VA 4-5 Mains input Logic supply voltage: Negative 15 volt analog
converter
+15VA 4-5 Mains input Logic supply voltage: Positive 15 volt analog
converter
28V_sec 4-5 Mains input Logic supply voltage: 28-30 volt, only to the generator modules.
converter
+5VD 4-5 Mains input Logic supply voltage: 5 volt digital
converter
DC_BUS_OK 4-5 Mains input The DC-bus voltage is OK, that is, inside its upper and lower limits.
converter
Note: The LED can only be lit when the D750 Master board gives a
GEN_POWER_ON signal.
DC_BUS_FAIL 4-5 Mains input The DC-bus voltage is outside its upper or lower limits.
converter
Note: The LED can only be lit when the D750 Master board gives a
GEN_POWER_ON signal.
H3 4-5 Mains input The DC-bus voltage is present (>15 volt) to the AC inverter and to
converter the Tube power supply.

Note: If LED H3 not is lit but H4 is lit, indicates the fuse F2 inside
the Mains input converter has blown and need to be replaced.
H4 4-5 Mains input The DC-bus voltage is present (>15 volt) to the HV inverter.
converter
AC_ON 4-5 AC inverter The AC output voltage is present.
HV_ON 4-5 HV inverter The EN_HV_ON signal is set and thus x-ray exposure is on.

Comment: Before making any service/maintenance on the generator modules


both LEDs H3 and H4 shall be off, thus making sure the DC-bus
voltage not is present.
The DC-bus voltage discharges below 60 volt within 3 minutes.

Siemens AG SPB7-250.844.01 Page 10 of 10 MAMMOMAT NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
A B C D E F G
INCOMING
MAINS X707
VARISTOR MAINS INPUT

Siemens AG
X041E

PE
V CONVERTER
H V INVERTER MODULE FILAMENT

Medical Solutions
1 INPUT CIRCUIT MAINS MODULE CURRENT
1

BREAKER INPUT FULL-BRIDGE


S700 CONVERTER X-RAY TUBE
2 MODULE HALF SINE HV TANK
WAVE
385VDC HIGH VOLTAGE
BOOST-
=
3 ~ FLYBACK + F2 +
FILTER CONVERTER FILTER
4
FILTER ~
=

Rev. 03
BIAS VOLTAGE

POWER ON

08.04
SPB7-250.844.01
OVERVOLT. 0
COIL TRIP PROTECTION
H V-SET I
2

F1
II
F808 F807 28V_PRIM H V_ACT
0V_PRIM
T802
28V_SEC

CS SD 24
FLYBACK

Page 1 of 2
BRICK CONVERTER
FLYBACK AUXILIARY
PXCM
CONVERTER POWER SUPPLY
BOARD 385VDC
0V_SEC
=
FILTER +
~
FOCUS SELECTOR
28V_SEC 5V_RS232 5VD +15VA -15VA FORWARD
0V_SEC 0V_RS232 0VD 0VA CONVERTER
H V_SET
+
3

GND

MA_ACT
AC INVERTER MODULE
T801 MA_SET
= DC POWER 385VDC
GENERATOR: BLOCK DIAGRAM

FIL_CURR_SET
230VAC
50 HZ
MA_ACT
~
FIL_CURR_ACT
TUBE POWER
H V_ACT MODULE
SINE WAVE
GENERATOR

A A
D µP D
4

INPUT

OUTPUT
GEN_POWER_ON
RS232
SERVICE PC
Z801
T803 µP
FILTER
TO OPDIMA
MASTER

Wiring Diagrams
STAND

Mammomat NovationDR
µP
CONTROL
PANEL BOARD
5
5-1

0
5-2 X041E
GROUND, 0V AND PROTECTION GROUND DISTRIBUTION
1 2 3 4 5

This page/chapter is intentionally left blank and will be supplied with an update.
G
F
E
D
C
B
A

Siemens AG SPB7-250.844.01 Page 2 of 2 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
A B C D E F G
TERMINALS
INCOMING RECTIFIED LINE
MAINS INPUT CONVERTER MODULE
MAINS X705 X732

Siemens AG
X041E

X707 BOOST-FLYBACK
PE CONVERTER 5 3

Medical Solutions
+DC-BUS
MAINS INPUT MODULE
1

X703 DIODE H V INVERTER


6 4
1 CIRCUIT BRIDGE
1 V1 V2 MODULE
BREAKER COMMON MODE 3 1
2 S700 FILTERS INPUT FILTER
2 L2 L3 VARISTOR
L4
4 2
3 3
L1
+ + F2
4
X702

Rev. 03
X705 X724
1 BLEADER
7 1 TUBE POWER
2 -DC-BUS
SUPPLY MODULE
Y Y

08.04
SPB7-250.844.01
3 1 4

8 1 AC INVERTER
2

X9 MODULE
X704 SENSE_CURRENT 7 3

1 TRIP COIL DC_VOLTAGE 385V


X700 X722

CS SD 24
Page 1 of 2
2 28V_SEC
TUBE

F808
1,3

F807
1,3
GENERATOR:

OVERVOLTAGE 0V_SEC POWER


PROTECTION F1 2,4 2,4 SUPPLY
X802B 4 1 28V_SEC
+ +15VA BOARD
COMP 0V_SEC 5 5
0VA
PRINCIPLE DIAGRAM

U REG +15VA 6,8 6,8


-15VA
+ +15 V
3

X802B 6 3 + 7 7
DC_POWER_OK 0VA
+ GEN. CONNECT
X802C 6 3 U REG -15VA 9 9
COMP
-15 V
X603 U REG 5VD
+ +5 V
0VD X701 X760
BRICK GEN. CONNECT
PXCM 14 14
U REG X1
28V_PRIM U REG 5V_RS232
+24V + + +5 V 13 13 MASTER
0V_PRIM 0V_RS232 +15VA +15VA BOARD
5 5
X706
MAINS INPUT & MAINS INPUT CONVERTER

0VA 0VA
28V_PRIM
4

6,8 6,8
1 -15VA -15VA
POWER_ON 5VD 7 7
2
5VD 5VD
9,11 9,11
+ DC_POWER_OK 0VD 0VD
SENSE_CURRENT POWER_ON 10,12 10,12

Wiring Diagrams
- DC_POWER_OK

5V_RS232 5V_RS232
DC_VOLTAGE 385V 15 15

Mammomat NovationDR
V1_GATE
0V_RS232 0V_RS232
RECTIFIED LINE V1_EMITTER AUXILIARY POWER SUPPLY BOARD 16 16

GEN_POWER_ON_P GEN_POWER_ON_P
5

UC 3854A 3 3
GEN_POWER_ON_N GEN_POWER_ON_N
4 4
0

DC_BUS_OK_P DC_BUS_OK_P
V2_GATE
1 1
V2_EMITTER DC_BUS_OK_N DC_BUS_OK_N
2 2
5-3
5-4 X041E
LEFT BLANK 0

1 2 3 4 5

This page is intentionally left blank.


G
F
E
D
C
B
A

Siemens AG SPB7-250.844.01 Page 2 of 2 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
5-5 D750 MASTER BOARD: X041E
PRINCIPLE DIAGRAM 0

1 2 3 4 5
X858
GENERATOR 1 MA_ACT_P 1 D750 MASTER BOARD 1 P15V_10W 1 1 5 DOUBLE DETECTOR
2 MA_ACT_N 2 2 N15V_10W 2 2 3
3 HV_ACT_P 3 3 RDLI 3 3 1
4 HV_ACT_N 4 X768 4 0V_RDL 4 4 4 X883
5 FIL_CURR_ACT_P 5 5 TRAI 5 5 2
6 FIL_CURR_ACT_N 6 6 SHIELD 6
G

7 HV_SET_P 7
8 HV_SET_N 8 1 DOOR_SWITCH_P DOOR
9 FIL_CURR_SET_P 9 2 DOOR_SWITCH_N SWITCH
10 FIL_CURR_SET_N 10 X754 3 GEN_POWER_AID_P POWER X-RAY TUBE
11 MA_SET_P 11 4 GEN_POWER_AID_N AID
12 MA_SET_N 12
0VD (K8)
X895
1 3 3 K8
13 HV_REGUL_ERR_P 13
2 X858
0VD (K9) 5 5
14 1 MA_ACT_P
HV_REGUL_ERR_N
GENERATOR 14 1 D750 MASTER BOARD DOUBLE DETECTOR K9
2 MA_ACT_N 2
X7551 P15V_10W
3 K8_TEMP
1 1
4
5
4
2 N15V_10W 2 2 3
15 MA_WERCURR_P 15
3 HV_ACT_P 3 RDLI 4 K9_PRESS 6
3 3 3 1 6
16 MA_WERCURR_N
4 HV_ACT_N 16 4 X768 4 0V_RDL 4 4 4 X883
17 GEN_CONNECT_P
5 FIL_CURR_ACT_P 17 5 5 TRAI 1 5 5 2
X804
PB_POWER_ON_ 15 15 1
6 FIL_CURR_ACT_N 6
18 GEN_CONNECT_N
7 HV_SET_P
18
7
6 SHIELD
2
6
PB_POWER_OFF_ 16 16 2 D741
19 GEN_IN_SPARE_1_P
8 HV_SET_N 19 8 1 DOOR_SWITCH_P
3 0VD DOOR 17 17 3
X32
F

2 DOOR_SWITCH_N SWITCH
20 GEN_IN_SPARE_1_N
9 FIL_CURR_SET_P 20 9
4 0VD 1 1 3
FIL_CURR_SET_N 10 X754 3 GEN_POWER_AID_P POWER
21
10
ACC_FREQ_P 21 X756
4 5 5VD
GEN_POWER_AID_N AID 2
X-RAY TUBE
2 1
D740
11 MA_SET_P 11
22 ACC_FREQ_N
12 MA_SET_N 22 12 6 PB_EXP_ON_
X895 3 3 2
23 STATUS_HV_ON_SIGNALS_P
13 HV_REGUL_ERR_P 23 13
1 0VD (K8) 3 3 K8 X5
2 7
0VD (K9) MPS
5 5 4 4 4
14 HV_REGUL_ERR_N 14 K9
24 STATUS_HV_ON_SIGNALS_N 24 X752 X755 K8_TEMP 0VD4
3 8 4 5 5 6
X721 25
15 MA_WERCURR_P
BIAS_VOLT_DETECTED_P 25
15
4 K9_PRESS 6 6
16 MA_WERCURR_N 16
BIAS_VOLT_DETECTED_N 1GRID_OUT_PX804 16
26 17 GEN_CONNECT_P 26 17 1 PB_POWER_ON_ 15 15 1 D801
27 SPARE_P18 GEN_CONNECT_N
27 18 X753
2 2 GRID_OUT_W
PB_POWER_OFF_ 16 16 1 2 X813 D741
19 GEN_IN_SPARE_1_P 19 3 0VD 17 17 3
X32
28 SPARE_N20 GEN_IN_SPARE_1_N
28 20 4 0VD 2 MPS 1 1 2 3
29 ACC_ANODE_P
21 ACC_FREQ_P 29 21 X756 5 5VD 2 2 1
D740
22 ACC_FREQ_N 22
4 GRID_RUNNING 4
30 ACC_ANODE_N 30 6 PB_EXP_ON_ 3 3 2
X5
23 STATUS_HV_ON_SIGNALS_P 23 X757 5 GEN_PREP_RDY 5
31 BRAKE_ANODE_P
24
31
STATUS_HV_ON_SIGNALS_N 24
7 MPS 4 4 4 X815
X721 X752 8 0VD 6 STAND_REQ_HV_OFF_
5 5 6 6
32 BRAKE_ANODE_N
25 BIAS_VOLT_DETECTED_P 32 25
E

BIAS_VOLT_DETECTED_N33 1 GRID_OUT_P 16
33 HV_ON_P26 26
1 D801
27 SPARE_P 27 X753 2 GRID_OUT_WEXT_POWER_ON_
1 X813
34 HV_ON_N 34 2 EXT_POWER_OFF_
28
FIL_CURR_ON_P
SPARE_N
35
28
2 MPS 2 PRODUCTION TEST
35 29 ACC_ANODE_P 29 6 REQ_HV_ON_
36 30 ACC_ANODE_N
FIL_CURR_ON_N 36 30
4
X758 GRID_RUNNING 4 EQUIPMENT
X757 5 7 +15VA
GEN_PREP_RDY 5
X815
31 BRAKE_ANODE_P 31 RL2
37 MA_REGUL_ON_P 37 6 STAND_REQ_HV_OFF_
8 +15VA 6
32 BRAKE_ANODE_N 32
38 MA_REGUL_ON_N
33 HV_ON_P 38 33 1 EXT_POWER_ON_
39 34 HV_ON_N
SEL_LARGE_FOCUS_P 39 34 X759
2 1 EXT_EXP_LAMP_P
EXT_POWER_OFF_
PRODUCTION TEST
40
35 FIL_CURR_ON_P
SEL_LARGE_FOCUS_N 40
35 6 2 EXT_EXP_LAMP_N
REQ_HV_ON_ EXTERN EXPOSURE LAMP
36 FIL_CURR_ON_N 36 X758 EQUIPMENT
7 +15VA
41 SEL_MO_ANODE_P 41 RL2
37 MA_REGUL_ON_P 37 8 +15VA
1 5V_RS232 1
42 MA_REGUL_ON_N
SEL_MO_ANODE_N
38 42 38 BRICK
39 SEL_LARGE_FOCUS_P 39 X759 1 2 0V_RS232
EXT_EXP_LAMP_P 14
43 EN_HU_ON_P 43 2 EXT_EXP_LAMP_N EXTERN EXPOSURE LAMP
40 SEL_LARGE_FOCUS_N 40 3 PREP_DR_DET_P 2
44 EN_HU_ON_N
41 SEL_MO_ANODE_P 44 41
4 PREP_DR_DET_N
45 GEN_OUT_SPARE_1_P
42 SEL_MO_ANODE_N 45 42
1 5V_RS232 1
BRICK15
0V_RS232
D

2 5 14
GEN_RDY_FOR_DR_EXP_P
43 EN_HU_ON_P 43 3
46 GEN_OUT_SPARE_1_N 46 3 PREP_DR_DET_P 2
44 EN_HU_ON_N 44 6 GEN_RDY_FOR_DR_EXP_N
PREP_DR_DET_N 16
47 GEN_OUT_SPARE_2_P 47 4 15
45 GEN_OUT_SPARE_1_P 45
5 7 DR_DET_RDY_FOR_EXP_P
GEN_RDY_FOR_DR_EXP_P 3 4
48 GEN_OUT_SPARE_2_N
46 GEN_OUT_SPARE_1_N 48 46
GEN_RDY_FOR_DR_EXP_N
6 8 16
DR_DET_RDY_FOR_EXP_N 17
47 GEN_OUT_SPARE_2_P 47
7 DR_DET_RDY_FOR_EXP_P 4
1 48 GEN_OUT_SPARE_2_N
DC_BUS_OK_P 1 48 9 DR_SYST_RDY_P 5
8 DR_DET_RDY_FOR_EXP_N 17
2 DC_BUS_OK_N
1 DC_BUS_OK_P 2 1 9 10 DR_SYST_RDY_N
DR_SYST_RDY_P 5 18
2 DC_BUS_OK_N 2 10 DR_SYST_RDY_N
11 18
DR_DET_ERR_P 6
3 GEN_POWER_ON_P 3
3 GEN_POWER_ON_P 3 11 DR_DET_ERR_P 6
4 GEN_POWER_ON_N 4 12 DR_DET_ERR_N 19
4 GEN_POWER_ON_N 4 12 DR_DET_ERR_N 19
5 +15VA 5 +15VA 5 5 X761X761
13 13 GEN_POWERED_ON_P
GEN_POWERED_ON_P 7 X602 7 X602
6 0VA 6 14 GEN_POWERED_ON_N
14 20
GEN_POWERED_ON_N 20
6 0VA 6
7 -15VA 7 15 X_RAY_ON_P 8
7 -15VA 7 15 X_RAY_ON_P 8
8 0VA 8 16 X_RAY_ON_N 21
8
X701
0VA 8 16 X_RAY_ON_N 21
X701 9 5VD 9 X760 17 SHIELD

9 5VD 10 0VD 9 10 X760 18 17 SHIELD 22


11 5VD 11 19 TXD_DR 10
C

10 0VD 10 18 22
12 0VD 12 20 0V_RS232 23
11 5VD 13 11
GEN_CONNECTION_LOOP 13 21 19
PXD_DR TXD_DR 11 10
12 0VD 14 12
GEN_CONNECTION_LOOP 14 22 20
0V_RS232 0V_RS232 24 23
15 5V_RS232 15 23 5V_RS232 12
13 GEN_CONNECTION_LOOP 13 21 PXD_DR 11
16 0V_RS232 16 24 0V_RS232 25
14 GEN_CONNECTION_LOOP 14 22 0V_RS232 24
HV_ACT 1 2 RXD_S_PC 2
15 5V_RS232 15 23 5V_RS232 12
X805

PRODUCTION MA_AT 2 3 TXD_S_PC 3


X762 24 0V_RS232
SERVICE - PC
TEST
16 0V_RS232 0VA 16 3 5 0VI_RS232 5 25
EQUIPMENT HV_SET 4
2 RXD_OPD_WS
HV_ACT MA_SET 1 5 2 RXD_S_PC 2
X805

TXD_OPD_WS
PRODUCTION MA_AT 0VA 2 6 X765 3
3 TXD_S_PC OPDIMA WORKSTATION
3
RL1 X762
4 DTR_OPD_WS SERVICE - PC
TEST 0VA 3
7
X763 5 5
0VI_RS232 0VI_RS232 5
8
EQUIPMENT HV_SET 0VA 4 9
+15VA 6 DSR_OPD_WS
8 2 RXD_OPD_WS
GND_OPD_WS
MA_SET 0VA 5 10
3 TXD_OPD_WS
0VA 6 X765 2 RXD_PTR OPDIMA WORKSTATION
B

X769 RL1 3 4
TXD_PTR DTR_OPD_WS PRINTER/ID-FLASHER
7
T801 10V5_AC2 1 X764X763
5 0V_RS232
5 0VI_RS232
10V5_AC1 8 2 X772 A GND_PTR
+15VA 6
GND_PTR
DSR_OPD_WS
B
0VA 9
8 GND_OPD_WS
0VA 10
2 RXD_PTR
X769 3 TXD_PTR PRINTER/ID-FLASHER
T801 10V5_AC2 1 5 0V_RS232
X764
10V5_AC1 2 X772 A GND_PTR
B GND_PTR
A

Siemens AG SPB7-250.844.01 Page 1 of 1 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
A B C D E F G

Siemens AG
X041E

Filament focusing converter

Medical Solutions
28V_sec V_Bias
1

0V_sec
Bias_W Bias_Mo
PWM R111 R109 X721
385VDC Filament converter IC det. Bias_volt_detected_p 25
T201
logic
26

Rev. 03
on on
off off
V_Bias_ret
+15V_prim

08.04
X726

SPB7-250.844.01
Bias_voltage 1 PK G
2

N201
Inv YE h1M
V204
UCC3804
Sel_Mo_anode_p Small_focus_W 3
0V_prim h2M
BU

CS SD 24
Small_focus_Mo 2

Page 1 of 1
h1W
PRINCIPLE DIAGRAM

GY
Large_focus_W 5
L201 L202 L203
Large_focus_Mo 4
+ + WH h2W

385VDC_ret GND 6 BN h3

+15VA 7 BL
3

Generator
door
GENERATOR - TUBE POWER SUPPLY:

Voltage limit
TUBE FILAMENT AND FOCUS SELECTION

Current limit

X721 Sel_large_focus_p* Sel_Mo_anode_p*


35 Fil_curr_on
36 K1 K2

9 Fil_curr_set_p Fil_curr_set
4

10

MA_regul_on_p TP
37
FIL_CURR_READ
38 X721
Fil_curr_act_p 5
11 MA_set_p
MA_set 6
12 MA_set_n

Wiring Diagrams
MA_act

MA_overcurr_p 15

Mammomat NovationDR
Logic 16
circuit TP
Fil_curr_act
MA_READ
X720
5 MA_sense MA_act_p 1
5

6 MA_sens_ret 2
TP
X721 0VA
39 Sel_large_focus_p*

40

41 Sel_Mo_anode_p*
* = active low
5-6

42
A B C D E F G
X722 X721

Siemens AG
X041E

9 Gen_connect_p 17

Medical Solutions
1

1, 3 28V_sec
2, 4 0V_sec

5 +15VA
6, 8 0VA

Rev. 03
385 VDC
7 -15VA
385 VDC_ret
To Filament Converter

08.04
SPB7-250.844.01
Rotation Inverter
2

X724
DC-bus Tube
1 V6 V5
X725
High High

CS SD 24
+ Motor_II 2 WH II

Page 1 of 1
220mF & &
4 Low Low 3.3mF 2.2mF
side side L2
DC-bus_ret V13 V14
drive drive Motor_I YE I
4

Motor_0 1 BN 0
3

X723
28V_prim 15V_prim
1 15V_reg

To Filament Converter
2
V6+V14
28V_prim_ret 0V_prim PWM Switch
Circuit V13+V5 logic Square
Wave
circuit
GENERATOR -TUBE POWER SUPPLY:

Enable pin
PRINCIPLE DIAGRAM, TUBE ROTATION

Internal
freq_adj.
X720
Anode 28V_sec 1,3
4

drive freq.
0V_sec 2,4
X721
3 HV_act_p 11
28V_sec 4 HV_act_n 12
7 HV_set_p 13

Wiring Diagrams
8 HV_set_n 14
X721 Time-out 13 HV_regul_err_p 15
To To
29 Acc_anode_p acc. 14 16
D750 HV

Mammomat NovationDR
logic
30 Master 23 Status_hv_on_signals_p 18 inverter
board 24 20
Time-out
31 Brake_anode_p brake 33 HV_on_p 7
logic
5

32 34 8
43 En_hv_on_p 9
44 10
Rotation frequency
0

18 Gen_connect_n 19
21 Anode_freq_p

22
5-7
A B C D E F G
X730: 3 TP 5.1V_ref +24V

Siemens AG
I_OUT
X041E

X731:5 MA_SENSE TR1

Medical Solutions
X4
1

2 Overcurrent
X730:4 Comp
HV inv.
MA_SENSE_RET current
X731:6 1 ref
TP
HV_WISH
N4 UCC3806 OSC. 1 G1
Current
X731:13 HV_SET_P transformer

Rev. 03
Limit
+ A-out
7.5V
_

B-out

08.04
SPB7-250.844.01
X730:5 HV_act_+
E1
2

TP B-out
TP 0VA
HV_READ

X731:11 HV_ACT_P HV_READ

CS SD 24
+24V

Page 1 of 1
SHDWN CURSNS

X4
4
Interlock
RT CT

Common 3 Ramp OSC. 2 G2


GENERATOR - HV INVERTER:

X731:7 HV_ON_P
3

E2
X731:9 EN_HV_ON_P HV_ON A-out
PRINCIPLE DIAGRAM, CONTROL LOGIC

+24V

X731:19 GEN_CONNECT
0

X730:9
OSC. 3 G3
4

HV_READ A_out
X731:18 STATUS_HV_ON_SIGNALS_P
E3
HV A-out
X731:15 HV_REGUL_ERR_P Time out
circuit

Wiring Diagrams
OSC. 1

X731:1 28V_sec OSC. 2


OSC. +24V

Mammomat NovationDR
+24V
V-reg circuit OSC. 3
X731:3 Comp +24V

ref OSC. 4
X731: 2
0V_sec +12V
5

V-reg
X731: 4 OSC. 4 G4
0V_sec=0VA
X730: 1, 2, 6, 8, 10

E4
X731: 8, 10, 12, 14, 16, 20 B-out
5-8
A B C D E F G

Siemens AG
X041E

Medical Solutions
1

MAINS INPUT MAINS INPUT CONVERTER HV INVERTER

INCOMING MAINS CIRCUIT BREAKER S700 V1 V2


X707 X705 DC-BUS
FILTER FILTER
F2 X705
+ X732 6x220mF

G1 G3

Rev. 03
E1 E3

08.04
SPB7-250.844.01
AC INVERTER
2

+ G2 G4
TUBE POWER SUPPLY
E2 E4

CS SD 24
Page 1 of 2
M1

TR1 CURRENT TRANSFORMER

M2
PRINCIPLE DIAGRAM

OVERCURRENT
2 COMP
X4
1
REF

HV TANK
T1 6.6k X3 HV CABLE

C1 94M
X6 5 X730 HV INVERTER
HV_ACT
X-RAY
C2 18.8k TUBE
10c 1, 2, 6, 8, 10
GENERATOR - HV INVERTER AND HV TANK:

4 MA_SENSE_RET 0V_SEC

3 MA_SENSE

X5 X731

Wiring Diagrams
2, 4, 8, 10, 16, 20

X720
MAS CONNECTOR

Mammomat NovationDR
TUBE
POWER
0V_SEC X726
SUPPLY
0VA 6
5

GENERATOR DOOR
5-9
5-10 X041E

PRINCIPLE DIAGRAM, CONTROL PANEL


1 2 3 4 5

This page/chapter is intentionally left blank and will be supplied with an update.
G
F
E
D
C
B
A

Siemens AG SPB7-250.844.01 Page 2 of 2 Mammomat NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
A B C D E F G
X752
3 HV_ACT
D750

Siemens AG
X041E

AEC µP-BUS

Medical Solutions
MPS
1

HV_DETECTED_ EPROM EEPROM


RX RAM
COMP (FLASH)
1
X>Y TX
2
3
X771
1-2 NORMAL

Rev. 03
2-3 NO DOSE KVA_ABORT_EXP
MONITOR
CPU 1
CPU KVA_EXP_DONE

08.04
SPB7-250.844.01
OVD RESET
1
2

0V_RDL

PLD PROM

CS SD 24
PLD

Page 1 of 1
P15V_ION

PLD CONTROL
N15V_ION VOSCTL_
EOSM_

REQ_HV_ON
& PLD
STATUS LED’S
D750 MASTER BOARD:

V2 V8
3

V1 FACTORY
STATUS V9
PRINCIPLE DIAGRAM, AEC

USE ONLY GAIN


LED’S
V4 V7 1 V10

PLD V3 V6 2 V11

V5 4 V24
MUX
RDL1
OFS_RDL

RDL2 SEL_RDL A3_RDL AD_RDL


A4_RDL
A=4 HUM FILTER A=1-128
S&H

Wiring Diagrams
Mammomat NovationDR
4

U
-100 µV + F
OFFSET MUX VOSCTL1
TEST TEST FRDL
-0.9V VOSCTL1_ 0VA
VOSCTL1_
0VA1
EOSM_ -
&
VOSCTL1
J3 VOSCTL1

VOSCTL_ 1
& & VOSCTL1_
J1 J2
VOSCTL1
TRAI (NOT USED)
0

A_PLD_REQ_HV_OFF

0VA

0VD=GND, 0VA IS CONNECTED TO 0VD


JUMPERS J1, J2, J3 NOT MOUNTED
5-11

6 5 4 1 2 3 11
X768
X041E D750 MASTER BOARD: 6-1
CONTACT LOCATION 0
1 2 3 4 5
D750 MASTER BOARD
S4
G

5VD_1
0VD_1
0VA_1
F

X768
0VA_3
0VD_2
M_WD_RESET_
FIL_CURR_ACT
HV_ACT
E

X765
0VD_3 MA_ACT
HV_SET
N15VA
FIL_CURR_SET
P15VA
MA_SET
0VA_2
D

ANODE_FREQ
X752
DC_BUS_OK HV_OFF_INT_
EN_HV_ON
C

ACC_ANODE
HV_ON
BRAKE_ANODE S3
0VD_5 X758 GEN_POWER_ON
S2
X754
X755 REQ_HV_ON
X757 0VD_4
X760
X753
B

X756
DR_SYST_RDY
DR_DET_RDY_FOR_EXP
DR_DET_ERR_ X759
5V_RS232
0V_RS232
GEN_POWERED_ON_ 5VI_RS232
PREP_DR_DET_ X772
GEN_RDY_FOR_DR_EXP_
A

X_RAY_ON_
0VI_RS232
X761
X769 X764 X762
X763
Siemens AG SPB7-250.844.01 Page 1 of 2 Mammomat Novation
DR
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
GENERATOR:
6-2 CONTACT DIAGRAM 0
X041E
1 2 3 4 5
X5
X3
G

X6
X8
X1
X7
X2
F
E
D

Master board T801


F801
D750
S1
F802
230V X803 X886 S2
50Hz X713 P1 F803
X752 S3
F804
S5
F805
S4
F806
C

S6
S7
X706 To
X760 Master
board
F1 F2
F2 X705 Z801
0.5AT 0.5AT Boost-Flyback T803
To
+ DC-bus Filter Opdima
X772
B

From
T801: S7
Brick PXCM
X9 F807 T802 X802c 2AT
X603
F808 X802b X802b 2AT
A

Incoming mains
Siemens AG SPB7-250.844.01 Page 2 of 2 Mammomat Novation
DR
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 7-1
Changes to previous version 0

Chapter Page Change

2 2 Two new components, T803 and Z801, added in Fig 2.

2 3 New Fig 3.

2 6 Two new components, T803 and Z801, in the list Miscellaneous

3 1 New information because of T803 in the block diagram cable con-


nections.

5 1 New information in Generator: Block Diagram, because of T803


and Z801.

6 2 New information in Generator: Contact Diagram, because of T803


and Z801.

Siemens AG SPB7-250.844.01 Page 1 of 2 MAMMOMAT NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
7-2 X041E
Changes to previous version 0

This page intentionally left blank.

Siemens AG SPB7-250.844.01 Page 2 of 2 MAMMOMAT NovationDR


Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams

Вам также может понравиться