Вы находитесь на странице: 1из 2

50.1.41 Table  2011.19A.

  ICP-MS parameters
AOAC Official Method 2011.19 Typical operating conditions
Chromium, Selenium, and Molybdenum RF power, W 1600
in Infant Formula and Adult Nutritional Products RF matching, V 1.8
Inductively Coupled Plasma-Mass Spectrometry
Sampling depth, mm 9
First Action 2011
Extract 1 lens, V 0
A.  Principle Carrier gas, L/min 0.9
Test portion is heated with nitric acid in a closed vessel microwave Make-up gas, L/min 0.2
digestion system at 200°C. Digested test solution, or an appropriate Nebulizer (glass concentric) MicroMist
dilution, is presented to the inductively coupled plasma-mass Spray chamber temperature, °C 2
spectrometer (ICP-MS) instrument standardized with acid matched Interface cones Ni
standard calibrant solutions. An ionization buffer (potassium) is He cell gas flow rate, mL/min 4.5
used to minimize easily ionizable element (EIE) effects, methanol H2 cell gas flow rate, mL/min 4.2
is added to normalize the carbon content, and nickel and tellurium
Nebulizer pump rate, rps 0.1 (0.5 mL/min)
are used as internal standards.
Analyte/internal standard/ 52
Cr,95Mo/ 60Ni,/He 78Se/130Te/H2
B.  Apparatus   gas mode
(a)  Microwave.—Commercial microwave designed for
laboratory use at 0–300°C, with closed vessel system and and 0.4, 2.0, and 10 ng/mL Se, plus blank, with both Ni and Te
controlled temperature ramping capability. It is recommended that internal standard, in HNO3. Ni is used as the internal standard for
the vessel design be selected that will withstand the maximum both Cr and Mo, and Te must be used for Se.
possible pressure, since organic material, and also carbonates if (b)  Sample preparation.—Prepare powder samples by
not given sufficient time to predigest, will generate significant reconstituting approximately 25 g sample in 200 mL warm
pressure during digestion. (Vessels can reach 700 psi or more on laboratory water (60°C). Accurately weigh approximately 1.8 g
occasion.) Vessels must be designed to operate with only 6 mL reconstituted test portion into the digestion vessel. This represents
solution volume, or the volume must be adjusted accordingly. approximately 0.2 g original powder sample. Fluid samples may
Vent according to manufacturer’s recommendation. (Caution: be prepared by accurately weighing approximately 1 g test portion
Microwave operation involves hot pressurized acid solution. Use weighed directly into the digestion vessel after mixing. For a
appropriate face protection and laboratory clothing.) Additional 1-step digestion (two stages in microwave program), add 0.5 mL
instrument parameters are summarized in Table 2011.19A.
5000 ng/mL Ni and Te internal standard solution and 5 mL trace
(b)  ICP-mass spectrometer.—With collision reaction cells (CRCs).
metal-grade HNO3 followed by 2 mL H2O2 to the microwave
(c)  Various plasticware and pipets.
digestion vessels. Seal vessels according to manufacturer’s
C.  Chemicals and Reagents directions and place in microwave. Ramp temperature from
[Caution: Use normal laboratory safety precautions (laboratory ambient to 180°C in 20  min, and hold for 20  min in stage 1. In
coats and safety glasses with side shields) when handling stage 2, the microwave will automatically ramp to 200°C in 20 min,
concentrated acids, bases, and organic solvents. Additional and hold for 20 min (see Table 2011.19B).
protections such as face shields, neoprene gloves, and aprons For microwave ovens without the 2-stage program and where
should be used where splashing may occur. Avoid breathing vapors it is more convenient, use the 2-step digestion. Add 0.5  mL
by working in approved hoods.] 5000  ng/mL Ni and Te internal standard solution and 5 mL trace
(a)  Laboratory water.—Use 18 MΩ water throughout for dilution. metal-grade HNO3. With power settings appropriate to microwave
(b)  Concentrated nitric acid (HNO3).—65–70% trace metal- model and number of vessels, ramp temperature from ambient to
grade HNO3 throughout. 200°C in 20 min. Hold at 200°C for 20 min. Cool vessels according
(c)  Hydrogen peroxide.—30% ACS reagent grade. to manufacturer’s directions, approximately 20 min. Slowly open
(d)  Methanol.—99.99% analytical reagent grade for matrix the microwave vessels, venting the brownish nitrogen dioxide
matching. gases. (Caution: Venting must be performed in a hood because NO2
(e)  Potassium.—10 000 mg/L in nitric acid for matrix matching. is very toxic.) Add 1 mL H2O2 and redigest samples by ramping the
temperature from ambient to 180°C in 15 min. Hold at 180°C for
D.  Standards
15 min and cool for 20 min.
(a)  2 mg/L Cr and Mo and 1 mg/L Se multielement stock standard (c)  Preparation of test solution.—Add approximately 20 mL
solution in nitric acid.—High-Purity Standards (Charleston, SC), laboratory water to the contents of the vessel with the digested
or equivalent. samples and transfer to a 50 mL sample vial. Rinse the vessel and
(b)  5 mg/L Ni and Te multielement stock standard solution in transfer the rinsate into the sample vial. Add 0.5 mL methanol to
nitric acid.—High-Purity Standards, or equivalent. the sample vial and dilute to 50 mL with laboratory water.
E.  Procedure F.  Determination

(a)  Standard preparation.—Prepare intermediate standards Table 2011.19A summarizes typical instrument parameters
from commercial stock standards at 40 ng/mL Cr and Mo, and for analysis. Analyze test solutions using an ICP-MS instrument
20  ng/mL Se. Custom-blended multielement stock standard in standardized with standard solutions. Ni is used as the internal
HNO3 is acceptable. Prepare a minimum of three multielement standard for both Cr and Mo (helium mode), and Te must be used for
working standards containing 0.8, 4.0, and 20 ng/mL Cr and Mo Se (hydrogen mode). Analyze a 4 ng/mL Cr and Mo, and 2 ng/mL

© 2012 AOAC INTERNATIONAL


Table  2011.19B.  Operating parameters SD of 10 digested blanks by a factor of 10, and the instrument
Stage 1 sample digestion
LOQ by multiplying the instrument LOD by 3 (1). However, in this
method the useful LOQ, or practical LOQ (PLOQ), was determined
1 Power 100% (1600 W)
to be the lower linear limit value of the calibration curve because
2 Ramp to temperature 20 min the accuracy and precision of sample measurements below that
3 Hold time 20 min value would be uncertain. Almost all mineral-fortified nutritional
4 Temperature 180°C products can be prepared with a DF such that Cr, Se, and Mo will
5 Cool down 20 min be present in the analytical solution above the PLOQ.
Stage 2 sample digestion (c)  Matrix matching with methanol.—The presence of carbon
(organic compounds) in analytical solutions causes signal enhancement
1 Power 100% (1600 W)
of Se during analysis by ICP-MS (2–4). To determine the optimum
2 Ramp to temperature 20 min
concentration of methanol (source of carbon) needed to compensate
3 Hold time 20 min for Se signal enhancement, various concentrations of methanol were
4 Temperature 200°C added to both calibration standards and digested samples.
5 Cool down 20 min (d)  Effects of EIEs.—Many nutritional products contain
significant levels of EIEs, such as Ca, Na, K, and Mg. Therefore,
Se working standard or other suitable quality control solution every blank solutions and solutions containing 4 ng/mL Cr and Mo and
10 test portions to monitor for instrument drift and linearity (result 2 ng/mL Se were analyzed both with and without EIEs to determine
100 ± within 5% of nominal). The inclusion of a method blank (run any changes in concentrations of the analytes.
as a sample), a duplicate sample [relative percent difference (RPD) (e)  Specificity.—Specificity of the method is its ability to
≤ within 10%], and known reference materials serving as control accurately measure the analyte in the presence of other components
samples (recovery check within control limits) are considered in the sample matrix that might cause spectral interferences.
mandatory for good method performance. If any of these QC To demonstrate the specificity of the method, undigested blank
checks fails, results should be considered invalid. solutions were spiked with multielement solutions at concentrations
G.  Calculations that are representative of nutritional products in samples for ICP-
Sample concentrations were automatically calculated by the MS analysis. The typical H2 gas mode for Se, and He gas mode for
ChemStation software using a nonweighted least-squares linear Cr and Mo, were used.
regression calibration analysis to produce a best-fit line: (f)  Accuracy.—Accuracy was demonstrated by analyzing three
National Institute of Standards and Technology (NIST) standard
Y = ax + blank reference materials (SRMs) on 2 independent days, measuring
spike recoveries in 10 nutritional products on 3 different days,
The analyte concentration in the sample was then calculated: and comparing results for 10 nutritional products obtained by this
method to results obtained by other in-house validated ICP-AES
and atomic fluorescence spectrometry (AFS) methods. The spike
where x = analyte concentration (ng/g); y = sample response ratio levels of the analytes added to the products were between 50 and
(ng/mL), which is the measured count of each analyte’s standard 200% of the analyte concentrations in each product.
solution data point in the calibration curve divided by the ratio of (g)  Precision.—Both within- and between-day RSD values were
the counts/concentration of the internal standard at the same level; determined by analyzing two in-house laboratory control samples.
blank = blank standard solution (ng/mL), which is the measured Within-day precision was determined by analyzing the laboratory
count of the blank standard solution data point in the calibration control samples in duplicate on each day, and between-day
curve divided by the ratio of the counts/concentration of the precision was measured by using the mean results of the duplicate
internal standard at the same level as the blank standard solution; samples analyzed on each day on 10 different days.
a = slope of the calibration curve; and DF = dilution factor of the
(h)  Ruggedness and robustness.—To determine the ruggedness of
sample solution divided by sample weight (mL/g).
the method, laboratory control samples were analyzed by two analysts
H.  Method Validation on 10 different days. Also, NIST SRM 1849 was analyzed in triplicate
(a)  Linearity.—All calibration curves were prepared using with varying sample weights and with different internal standards.
a nonweighted least-squares linear regression analysis, and References: (1)  IUPAC (1996) Pure Appl. Chem. 68, 1167–1193
correlation coefficient (r) values were calculated with each
(2)  Kralj, P., & Verber, M. (2003) Acta Chim. Slov.
calibration curve. Each calibration curve was prepared with four
50, 633–644
multielement standard solutions, including the blank standard
(3)  Gammelgaard, B., & Jons, O. (1999) J. Anal.
solution. It should be noted that all analyte concentrations in
samples were within linear range of the calibration curve and above Atom. Spectrom. 14, 867–874
the established lower linearity limit. (4)  Feldmann, I., Jakubowski, N., Thomas, C., & Stuewer,
(b)  LOQ.—The LOQ is the lowest concentration of the analyte D. (1999) Fresenius’ J. Anal. Chem. 365, 415–421
in the sample that can be reliably quantitated by the instrument. The J. AOAC Int. 95, 588(2012); AOAC SMPR 2011.009,
method LOQ is typically determined by multiplying the average J. AOAC Int. 95, 297(2012)

© 2012 AOAC INTERNATIONAL

Вам также может понравиться