ICAP-PF.0006368 Release_Bulletin_DICOM_Viewer_R3.

0_SP03 Authorized Version: 5

Philips Healthcare Quality System Management Page 1 of 9


Imaging System Platform Release Bulletin Template
ICAP

Release Bulletin
Philips DICOM Viewer R3.0L1-SP03
Printed copies are uncontrolled duplicates

Document id: ICAP-PF.0006368

Function Name Signature & Date


Binu Krishnan Available in Livelink
Printed copies are not controlled

Integration
Prepared by:
Manager
Kiran Ramaiah Available in Livelink
Prepared by: Test Engineer

Reviewed Sujaykumar Not applicable


Q&R Engineer Suvarna
by:
Rashmi Sharma Available in Livelink
Approved by: Project Manager

ICAP-T-030015.17 Corresponds to ICAP-W-030015.02 Confidential.


ICAP-PF.0006368 Release_Bulletin_DICOM_Viewer_R3.0_SP03 Authorized Version: 5

Philips Healthcare Quality System Management Page 2 of 9


Imaging System Platform Release Bulletin Template
ICAP

Table of Contents
SECTION 1 INTRODUCTION............................................................................................................................................3
1.1 PHILIPS DICOM VIEWER R3.0L1-SP03 .....................................................................................................................3
1.2 IDENTIFICATION ...........................................................................................................................................................3
1.2.1 Solved defects..................................................................................................................................................3
1.3 CONTENTS OF THE DELIVERY .......................................................................................................................................4
SECTION 2 INSTALLATION.............................................................................................................................................4
IHE COMPLIANT README FILE .................................................................................................................................................5
Printed copies are uncontrolled duplicates

SECTION 3 TEST PLAN....................................................................................................................................................5


3.1 T EST EQUIPMENT .........................................................................................................................................................5
3.2 SOLVED DEFECTS ........................................................................................................................................................5
3.3 IMPACTED TEST CASES ................................................................................................................................................5
Printed copies are not controlled

3.4 REGRESSION T EST CASES ...........................................................................................................................................5


3.5 HAZARD TESTS ............................................................................................................................................................5
3.6 SECURITY TESTS .........................................................................................................................................................6
3.7 VERIFICATION OF THE DELIVERY ..................................................................................................................................6
3.8 RELEASE OF DELIVERY ................................................................................................................................................6
SECTION 4 TEST REPORT .............................................................................................................................................7
4.1 CONCLUSION ...............................................................................................................................................................7
4.1.1 Actual environment used for testing............................................................................................................7
4.1.2 Verified Defects................................................................................................................................................7
4.1.3 Impacted test cases ........................................................................................................................................7
4.1.4 Hazard tests......................................................................................................................................................7
4.1.5 Security tests....................................................................................................................................................7
SECTION 5 APPLICABLE DOCUMENTS.......................................................................................................................8
5.1 INPUT DOCUMENTS .......................................................................................................................................................8
5.2 OUTPUT DOCUMENTS ....................................................................................................................................................8
SECTION 6 SCOPE FOR PHILIPS DICOM VIEWER R3.0L1-SP03.............................................................................8
SECTION 7 PREVIOUSLY RELEASED LEVELS/ SERVICE PACKS..........................................................................8
SECTION 8 REVISION HISTORY.....................................................................................................................................9

ICAP-T-030015.17 Corresponds to ICAP-W-030015.02 Confidential.


ICAP-PF.0006368 Release_Bulletin_DICOM_Viewer_R3.0_SP03 Authorized Version: 5

Philips Healthcare Quality System Management Page 3 of 9


Imaging System Platform Release Bulletin Template
ICAP

Section 1 Introduction
1.1 Philips DICOM Viewer R3.0L1-SP03

This is the release bulletin for the Philips DICOM Viewer R3.0L1-SP03. This release is based on Philips DICOM
Viewer R3.0L1-SP02.
The Philips DICOM Viewer R3.0L1-SP03 release includes the scope as specified in DV 3.0 bin in CQ

There are no functional changes (requirements) compared to R3.0 / R3.0 SP01/ R3.0 SP02.
Only defect solving is done.
Printed copies are uncontrolled duplicates

1.2 Identification
Printed copies are not controlled

Item Identification

Product Name DICOM Viewer R3.0L1-SP03


https://www.sc-
Delivery phlps.sfo.collab.net/sf/frs/do/viewRelease/projects.mip/frs.dicomviewer.r3_0_sp03
Location

Revision 47624 of https://project.sc-phlps.sfo.collab.net/svn/mip/branches/pii-


Baseline
sw112v1Lx
Build 47624_20130731_191509
32bits Build # 11.2.4960.1928
64bits Build # 11.2.4960.1935
https://project.sc-phlps.sfo.collab.net/svn/mip/tags/releases/DicomViewer/DV3.0-
Tag Location SP3

Tag Revision # 47627

1.2.1 Solved defects

ID headline Segment
MIP00113175 DV3SP2: “Alpha Blend” & "roll rotate" cursor Dicom
Viewer
MIP00112872 Image series could not be properly loaded and visualised on Philips DICOM Viewer Viewing
MIP00115483 The PMS DICOM viewer doesn’t show laterality in image information. Dicom
Viewer
Dicom
MIP00116282 Update the Philips brand name (so remove the Electronics part) in DICOM Viewer 3.0 Viewer

ICAP-T-030015.17 Corresponds to ICAP-W-030015.02 Confidential.


ICAP-PF.0006368 Release_Bulletin_DICOM_Viewer_R3.0_SP03 Authorized Version: 5

Philips Healthcare Quality System Management Page 4 of 9


Imaging System Platform Release Bulletin Template
ICAP

1.3 Contents of the delivery

Description Identification
Philips DICOM Viewer R3.0L1-SP03 software PmsDView.exe
(DICOM Viewer R3.0-SP03)
Philips DICOM Viewer R3.0L1-SP03 Release Bulletin (this ICAP-PF.0006368
document)
Product Safety Risk Overview Philips DICOM Viewer R3.0 ICAP-PF.0000995
DICOM Conformance Statement Philips DICOM Viewer R3.0 PIIOffc.0001192.
DICOM Conformance Approval Philips DICOM Viewer R3.0 PIIOffc.0001247
Printed copies are uncontrolled duplicates

IHE Integration Statement Philips DICOM Viewer R3.0 SP2 PIIOffc.0001240


Printed copies are not controlled

Section 2 Installation
No special installation steps are required, just copy the single-exe application (PmsDView.exe) to the location where
you want it and run it.

Note:
The Philips DICOM Viewer R3.0L1-SP03 will automatically detect the DICOMDIR if:

1. The DICOMDIR is located in the root of the drive the DICOM Viewer is started from.
2. The DICOMDIR is located in the current working folder, i.e. the folder where the DICOM Viewer is started from.

If the DICOMDIR cannot be located at any of these places, the DICOM Viewer will automatically detect DICOM files
with extension .dcm if:

1. The DICOM files are located in the root of the drive the DICOM Viewer is started from.
2. The DICOM files are located in the current working folder, i.e. the folder where the DICOM Viewer is
started from.

If no DICOMDIR or DICOM files can be located at any of these places, the DICOM Viewer will start and the user will
have to select the DICOMDIR or DICOM files manually.

A virus scanner or Internet security application may warn the Philips Dicom Viewer is trying to access the Windows
Registry. This is intended behavior and needed for the Philips Dicom Viewer to function properly.

Release Information:

Please read the safety risk Overview carefully, especially chapter 5 [INTENDED USE / INTENDED PURPOSE/
FORESEEABLE MISUSE]

The Philips DICOM Viewer R3.0L1-SP03 can be used to view DICOM image information available on CD/DVD/USB
or any DICOMDIR for non-diagnostic illustration purposes.
When DICOM image information will be distributed by means of CD/DVD or USB, the DICOM viewer can be
distributed together with the DICOM image information on the CD/DVD or USB.

The Philips DICOM Viewer R3.0L1-SP03 is not intended nor designed for any kind of
Clinical and/or diagnostic purposes

ICAP-T-030015.17 Corresponds to ICAP-W-030015.02 Confidential.


ICAP-PF.0006368 Release_Bulletin_DICOM_Viewer_R3.0_SP03 Authorized Version: 5

Philips Healthcare Quality System Management Page 5 of 9


Imaging System Platform Release Bulletin Template
ICAP

IHE compliant readme file


System integrators that intend to write the DICOM Viewer application to CD/DVD/USB while exporting image data to
CD/DVD/USB may want to include a IHE compliant readme file into the export. The following file can be used as a
template for such a readme file. Parts of this template have been filled with information on this Philips DICOM Viewer
R3.0L1-SP03 application:

README.txt
Printed copies are uncontrolled duplicates

Section 3 Test Plan


Printed copies are not controlled

The intention of this Dicom Viewer Release was to solve some deployment defects found by stakeholders
and in the field, based upon DV 3.0 SP02

The test approach chosen for the DICOM Viewer to verify the defects in scope and impacted areas if any.

3.1 Test equipment


Philips DICOM Viewer R3.0L1-SP03 will be tested on Windows 7 (64-bits) and Windows 8 (32-bits)

3.2 Solved defects

Segment Defect Headline Hazard Securit Status


related y
related
Dicom Viewer MIP00113175 DV3SP2: “Alpha Blend” & "roll No No Verified
rotate" cursor
Viewing MIP00112872 Image series could not be No No Verified
properly loaded and visualised on
Philips DICOM Viewer
Dicom Viewer MIP00115483 The PMS DICOM viewer doesn’t No No Verified
show laterality in image
information.
Dicom Viewer MIP00116282 Update the Philips brand name No No Verified
(so remove the Electronics part)
in DICOM Viewer 3.0

3.3 Impacted test cases


No impacted tests were identified

3.4 Regression Test cases


No impact on regression is identified

3.5 Hazard tests


Defect Test case Description Tester

ICAP-T-030015.17 Corresponds to ICAP-W-030015.02 Confidential.


ICAP-PF.0006368 Release_Bulletin_DICOM_Viewer_R3.0_SP03 Authorized Version: 5

Philips Healthcare Quality System Management Page 6 of 9


Imaging System Platform Release Bulletin Template
ICAP

Defect Test case Description Tester


None

3.6 Security tests


Defect Test case Description Tester
None
Printed copies are uncontrolled duplicates

3.7 Verification of the delivery


 PR verification
Printed copies are not controlled

 All relevant files (see chapter Error! Reference source not found.) are included in the delivery of the
Service Pack.

3.8 Release of delivery


The software will be released to the customer when:
 The tests described in chapter Error! Reference source not found., 3.3, 3.5 and 3.6 have passed
 The verification of the delivery items (4.1) was successful.

ICAP-T-030015.17 Corresponds to ICAP-W-030015.02 Confidential.


ICAP-PF.0006368 Release_Bulletin_DICOM_Viewer_R3.0_SP03 Authorized Version: 5

Philips Healthcare Quality System Management Page 7 of 9


Imaging System Platform Release Bulletin Template
ICAP

Section 4 Test Report


4.1 Conclusion
The required tests which are described in Error! Reference source not found. are completed and passed.
Philips DICOM Viewer R3.0L1-SP03 can be released to the customer.
Printed copies are uncontrolled duplicates

4.1.1 Actual environment used for testing

The Philips DICOM Viewer does not have additional requirements on the hardware. The hardware must be
Printed copies are not controlled

capable of running the Windows version efficiently.


PmsDView.exe (Spoon version) was tested on the below OS flavours
Windows 7 (64-bits) and Windows 8 (32-bits)

4.1.2 Verified Defects

Segme Defect Headline Verified Result


nt By
Dicom MIP00113175 DV3SP2: “Alpha Blend” & "roll rotate" Kiran Verified
Viewer cursor Ramaiah
Viewing MIP00112872 Image series could not be properly Kiran Verified
loaded and visualised on Philips Ramaiah
DICOM Viewer
Dicom MIP00115483 The PMS DICOM viewer doesn’t Kiran Verified
Viewer show laterality in image information. Ramaiah
Dicom MIP00116282 Update the Philips brand name (so Kiran Verified
Viewer remove the Electronics part) in Ramaiah
DICOM Viewer 3.0

4.1.3 Impacted test cases


None

4.1.4 Hazard tests


Defect Test case Description Tester Result
None

4.1.5 Security tests


Defect Test case Description Tester
None

ICAP-T-030015.17 Corresponds to ICAP-W-030015.02 Confidential.


ICAP-PF.0006368 Release_Bulletin_DICOM_Viewer_R3.0_SP03 Authorized Version: 5

Philips Healthcare Quality System Management Page 8 of 9


Imaging System Platform Release Bulletin Template
ICAP

Section 5 Applicable Documents


5.1 Input documents
Ref. Document title Document ID
[PP] SIS_Project_Plan_for_ISPv5.0 PIICons.0000404
[DRS] DRS Dicom Viewer R3.0 ICAP-PF.0000780
[IOP] Interoperability Validation Scope and Planning PMS DICOM Viewer PIIOffc.0000162
R3.0
Printed copies are uncontrolled duplicates

5.2 Output documents


Document title Document ID
Philips DICOM Viewer R3.0L1-SP03 Release Bulletin (this document) ICAP-PF.0006368
Printed copies are not controlled

Product Safety Risk Overview Philips DICOM Viewer R3.0 ICAP-PF.0000995


NOR-D DICOM Viewer R3.0 L1 SP03 ICAP-PF.0006371
DICOM Conformance Statement Philips DICOM Viewer R3.0 PIIOffc.0001192
DICOM Conformance Approval Philips DICOM Viewer R3.0 PIIOffc.0000250
IHE Integration Statement Philips DICOM Viewer R3.0 PIIOffc.0001240

Section 6 Scope for Philips DICOM Viewer R3.0L1-SP03


Defect Headline Segment
MIP00113175 DV3SP2: “Alpha Blend” & "roll rotate" cursor Dicom Viewer
Image series could not be properly loaded and
MIP00112872 visualised on Philips DICOM Viewer Viewing
MIP00115483 The PMS DICOM viewer doesn’t show
laterality in image information. Dicom Viewer
Update the Philips brand name (so remove the
MIP00116282 Electronics part) in DICOM Viewer 3.0 Dicom Viewer

Section 7 Previously released Levels/ Service Packs


Philips DICOM Viewer R3.0

Item Identification
Delivery location https://www.sc-
(CollabNet) phlps.sfo.collab.net/sf/frs/do/listReleases/projects.mip/frs.dicomviewer
Baseline Revision 42339 of https://project.sc-
phlps.sfo.collab.net/svn/mip/branches/pii-sw112v1lx
Source location https://project.sc-
phlps.sfo.collab.net/svn/mip/tags/releases/DicomViewer/DV3.0
Tag revision # 42386

ICAP-T-030015.17 Corresponds to ICAP-W-030015.02 Confidential.


ICAP-PF.0006368 Release_Bulletin_DICOM_Viewer_R3.0_SP03 Authorized Version: 5

Philips Healthcare Quality System Management Page 9 of 9


Imaging System Platform Release Bulletin Template
ICAP

Philips DICOM Viewer R3.0L1-SP01.

Item Identification
Release location https://www.sc-phlps.sfo.collab.net/sf/go/projects.mip/frs.dicomviewer.r3_0_sp01
Baseline Revision 44373 of https://project.sc-phlps.sfo.collab.net/svn/mip/branches/pii-
sw112v1
Source location Revision 44484 of https://project.sc-
phlps.sfo.collab.net/svn/mip/tags/releases/DicomViewer/DV3.0-SP1
Build 44373_20120808_111502
Printed copies are uncontrolled duplicates

Philips DICOM Viewer R3.0L1-SP02.

Item Identification
Delivery https://www.sc-
Printed copies are not controlled

Location phlps.sfo.collab.net/sf/frs/do/viewRelease/projects.mip/frs.dicomviewer.r3_0_sp02
Baseline Revision 46981 of https://project.sc-phlps.sfo.collab.net/svn/mip/branches/pii-sw112v1Lx
Build 46981_20130415_083218
32bits Build # 11.2.4853.0845
64bits Build # 11.2.4853.0852
Tag Location https://project.sc-phlps.sfo.collab.net/svn/mip/tags/releases/DicomViewer/DV3.0-SP2
Tag Revision # 46984

Section 8 Revision History

Revisi Date Status Changed by: Reason/ Description of change. (Insert Technical
on Review record)
Avail 01- See Livelink Binu Krishnan Draft version
able Aug- or header of
in 2013 printed
Liveli 01- documents Binu Krishnan Incorporated review comments from Sujaykumar
nk Aug- Suvarna
2013

DefectLoggingExcelR
eport_DV3.0 SP03.xlsx

02-Aug- Binu Krishnan added comment from Ynse Hoornenborg in the defect
2013 logging sheet

End of Document

ICAP-T-030015.17 Corresponds to ICAP-W-030015.02 Confidential.


ICAP-PF.0006368 Release_Bulletin_DICOM_Viewer_R3.0_SP03 Authorized Version: 5

This is a representation of an electronic record that was signed electronically in our Regulated System.

This page is the manifestation of the electronic signatures used in compliance with the organizations
electronic signature policies and procedures.
Printed copies are not controlled

UserName: Binu Krishnan (ing03207)


Title: Software Engineer
Date: Friday, 02 August 2013, 02:14 PM India Standard Time
Meaning: This document has changed to Authorized status
================================================

UserName: Kiran Ramaiah (300227082)


Title: Software Engineer
Date: Friday, 02 August 2013, 02:14 PM India Standard Time
Meaning: This document has changed to Authorized status
================================================

UserName: Rashmi Sharma (ing08775)


Title: Project Manager
Date: Friday, 02 August 2013, 03:15 PM India Standard Time
Meaning: This document has changed to Authorized status
================================================