Вы находитесь на странице: 1из 10

review http://www.kidney-international.

org
& 2013 International Society of Nephrology

Achieving more frequent and longer dialysis for the


majority: wearable dialysis and implantable artificial
kidney devices
William H. Fissell1, Shuvo Roy2 and Andrew Davenport3
1
Nephrology and Hypertension, Vanderbilt University, Nashville, Tennessee, USA; 2Bioengineering and Therapeutic Sciences, University of
California, San Francisco, San Francisco, California, USA and 3UCL Center for Nephrology, Royal Free Hospital, University College London
Medical School, London, UK

The long-term survival for many chronic kidney failure Once hemodialysis had become established as a treatment for
patients who remain treated by dialysis in economically chronic kidney disease (CKD), the early pioneers realized the
advanced countries remains similar to that of those with limitations of the treatment, particularly in terms of the
solid-organ malignancy, despite a disproportionate amount impact the intermittent three times weekly hemodialysis had
of health-care expenditure. As such, the current paradigm of on patients’ quality of life—not only in terms of time spent
three times weekly in-center hemodialysis for 4 h or shorter on dialysis and the time traveling to and from treatment but
sessions needs to change to improve patient outcomes. also dietary and fluid restrictions. In addition, many patients
Although more frequent and longer dialysis sessions have lose employment as a direct consequence of starting dialysis,
been reported to improve cardiovascular risk surrogates and and not surprisingly, as with many chronic illnesses,
short-term outcomes, these options are only practically depression is well recognized among dialysis patients. This
available to a very small fraction of the total dialysis motivated a search for the holy grail—a wearable dialysis
population. As such, radically new approaches are required device that would allow patients to receive continuous
to improve patient outcomes and quality of life for the treatment while going on with the normal activities of daily
majority of dialysis patients. Currently, two different life. Such a device would not only provide adequate solute
approaches are being developed, wearable devices based on clearances and control both electrolyte and acid–base status,
current dialysis techniques and more futuristic implantable but also improve blood pressure control—all while allowing a
devices modeled on the natural nephron. liberal diet and fluid intake. Despite many attempts to
Kidney International (2013) 84, 256–264; doi:10.1038/ki.2012.466; develop such a wearable artificial kidney, it is only recently,
published online 13 February 2013 with the advent of micro-technologies, that it has been
KEYWORDS: continuous dialysis; daily dialysis sorbent; implantable dialysis; possible to construct a truly wearable device that can
middle molecule; protein bound toxins; wearable dialysis accurately regulate ultrafiltration and achieve adequate solute
clearances.1
Peritoneal dialysis (PD), as we know it today, can be
described as a portable or wearable mode of dialysis.
However, when first introduced in the 1960s, PD was limited
to patients with acute renal failure, through hard catheters
inserted percutaneusly using a sharp metal trochar and
dialyzed with sterile dialysate supplied in 1-liter glass bottles,
with treatments typically lasting for more than 24 h. Over the
next decade, PD started to be introduced as a treatment for
CKD, which was initially similar to that used for treating
acute renal failure but then progressed to patients being
admitted to hospital three times weekly for overnight PD
treatment sessions requiring a new catheter insertion for
each treatment. As such, in many centers PD was a temporary
Correspondence: William H. Fissell, Nephrology and Hypertension, Vander-
treatment modality until patients could be established on
bilt University, Nashville, Tennessee, USA. intermittent hemodialysis. However, a number of key
E-mail: william.fissell@vanderbilt.edu technical advances, including the introduction of indwelling
Received 24 August 2012; revised 19 October 2012; accepted 26 silicone catheters and commercially available sterile dialysate
October 2012; published online 13 February 2013 in plastic bags, have allowed the development of a wearable

256 Kidney International (2013) 84, 256–264


WH Fissell et al.: Wearable and implantable dialysis devices review

dialysis system by the late 1970s,2 with patients being relies on using a fresh glucose-based dialysate each morning
dialyzed at home using simple mechanical devices, which and an overnight 7.5% icodextrin exchange to maintain
regulated dialysate inflow, dwell, and out-flow. However, it electrolyte homeostasis. In addition, there is no specific
was only with the advent of continuous ambulatory PD, ultrafiltration control, but ultrafiltration can be achieved by
pioneered by Montcrief and Popovich,3 that PD truly became adding additional glucose supplements to the regenerated
the major mode of renal replacement therapy we recognize dialysate during the last 2 h of daytime cycling therapy to
today, with now more than 230,000 patients being treated augment any ultrafiltration achieved from the nocturnal
worldwide. 7.5% icodextrin exchange.4
Spent peritoneal dialysate effluent contains proteins in low
CONTINUOUS WEARABLE PD DEVICES concentration, including fibrin, but with reuse this protein
PD is a wearable and portable dialysis therapy. However, content may well increase over time, and as such additional
continuous ambulatory PD typically requires three or four filters are required to prevent protein coating of the sorbents,
daily exchanges, which often limits the patient’s lifestyle thus reducing their efficiency.
flexibility. Although treatment with cycler PD allows daytime Currently, the ViWAK has not been formally tested in
freedom, the patient is connected to the machine overnight, animal or human clinical trials.
and treatment requires transport and storage of relatively The automated wearable artificial kidney (AWAK) is
large volumes of fresh dialysate. Currently, two new designs another continuous PD device designed for continued use,
have been proposed to increase patient freedom from which differs from the ViWAK in having single catheter
additional dialysate fluid exchanges. First, the Vicenza lumen access, and as such dialysate flow is discontinuous,
wearable artificial kidney (ViWAK)4 utilizes a dual lumen depending upon a tidal regimen requiring a reservoir for
catheter, allowing continuous flow PD powered by a small refreshed dialysate7 (Figure 1). Whereas the ViWAK is
lightweight battery-powered pump. After an initial 2-h dwell proposed to use two conventional peritoneal dialysate
with standard glucose-based dialysate, peritoneal dialysate is exchanges each day, the AWAK system is designed to
then continuously recycled by the passage of spent dialysate continuously regenerate dialysate, so that a single conven-
through a series of sorbents. Sorbents have traditionally tional glucose-based peritoneal dialysate solution may be
contained microporous carbon, zirconium, or polystyrene. continuously reused for up to a month or even longer. As
However, as neither of these readily absorbs urea, most such, this system has an additional chamber containing
sorbent designs have additionally incorporated urease to clear electrolytes, lactate, and glucose to refresh regenerated
urea, by metabolism to ammonium and carbon dioxide. As dialysate, and an ammonia sensor to monitor sorbent
ammonium is rapidly absorbed by zirconium phosphate, saturation. Around 750 ml of fresh dialysate is initially
ammonium generated by urease can be readily removed by infused into the peritoneal cavity, and then recirculated in a
placing zirconium phosphate adjacent to or in series with a tidal manner at 4 l/h using a battery-powered pump and any
urease sorbent. Even so, ammonia levels must be checked in ultrafiltrate generated over an 8–10-h period drained into a
the dialysate before return to the patient, for safety to ensure separate bag attached to the module. As recycled dialysate has
that the zirconium phosphate sorbent has not become lower glucose and changed electrolyte composition, it must
exhausted. Carbon dioxide micro- and macro-bubbles can be continuously regenerated, and as such the recycled
form in the extracorporeal circuit predominantly from the dialysate is reinfused with glucose and electrolyte solutions.
metabolism of urea, but also potentially from bicarbonate, Thus, the AWAK is designed to have both daily and monthly
which requires removal with a deaerating chamber, typically disposable sections, designed for ease of replacement
made of gas-permeable plastic polymers.5 Whereas (Figure 2). The rechargeable battery life is estimated to be
microporous carbons absorb a wide range of azotaemic around 18 h and requires recharging overnight.
toxins, including creatinine, uric acid, chloramines, oxidants, A 70-kg hemodialysis patient with a dietary protein intake
other organic compounds, heavy metals, middle molecules, of 1 g/kg could be expected to generate around 9–10 gm of
including beta 2 microglobulin, and protein-bound solutes, urea nitrogen per day.8 Although urease and 250 g of
other sorbents act largely as ion exchangers. Although zirconium can readily catalyze and absorb 2 g urea/h, this
zirconium phosphate absorbs ammonium, calcium, amount of urea clearance would typically exhaust the
magnesium, potassium, other cations, and metals, it currently available sorbent cartridges for wearable
releases hydrogen and, to a lesser extent, sodium ions, and peritoneal devices, potentially necessitating more than one
zirconium oxide and zirconium carbonate, which absorb daily cartridge exchange. To overcome this problem, the
phosphate, fluoride, and heavy metals, release sodium and AWAK device has produced two different sorbent cartridges,
bicarbonate, as well as, to a lesser extent, acetate.6 As such, one designed to extract 3.5 g of urea nitrogen and a larger
changes in dialysate electrolyte, bicarbonate, and glucose heavier cartridge to remove 10 g of urea nitrogen.9 During
composition occur over time. The proposed ViWAK system tidal PD, urea nitrogen clearance also depends upon
uses a combination of microporous carbon and polystyrene individual transporter status, and thus short-term 5-h
resin in a series of absorption columns, but does not have any studies simulating the AWAK tidal exchanges ranged from
specific system to correct electrolyte changes; however, it 12.9 to 17.1, which, based on the AWAK providing continual

Kidney International (2013) 84, 256–264 257


review WH Fissell et al.: Wearable and implantable dialysis devices

Cycler Cartridge
Non-disposables Daily disposables
Controller Gas
remover
Enrichment
solution
Monthly disposables
Sorbent
Pump Storage
cartridge
module

Ammonia
Pressure
sensor Pump sensor
Fibrin
filter Filter
UF bag

Catheter
Peritoneal
cavity
Connector

Figure 1 | The artificial wearable ambulatory kidney (AWAK) consists of two main components: a daily replaceable sorbent cartridge
containing electrolyte and glucose solution to regenerate spent dialysate, and a second assembly comprising ammonia detector,
protein, and bacterial filters designed to be changed monthly.

Cycler Cartridge
Non-disposables Daily disposables
Controller
Gas
remover
Enrichment
solution
Monthly disposables
Sorbent
Pump cartridge
Storage module
Ammonia
Pressure
sensor Pump sensor
Fibrin
filter Filter
UF bag

Catheter
Peritoneal
cavity
Connector

Cycler Cartridge
Non-disposables Daily disposables
Controller Gas
remover
Enrichment
solution
Monthly disposables
Storage Sorbent
Pump cartridge
module

Ammonia
Pump Pressure
sensor
sensor
Fibrin
Filter
filter
UF bag

Catheter
Peritoneal
cavity
Connector

Figure 2 | The artificial wearable ambulatory kidney (AWAK) uses a single lumen peritoneal dialysis access catheter, and therefore
peritoneal effluent is either pumped into the AWAK (upper panel) or returned to the patient (lower panel).

dialysis, extrapolates to a weekly Kt/Vurea of 3.4–6.5. terms of dietary protein intake, physical activity, and urea
However, the duration of effectiveness for any sorbent will generation rate, and the renewed interest in sorbent
depend markedly on individual patient characteristics in technology may well lead to the development of newer,

258 Kidney International (2013) 84, 256–264


WH Fissell et al.: Wearable and implantable dialysis devices review

more effective sorbents. The AWAK device is currently also used in two proof-of-concept human clinical trials, first
undergoing clinical trials, and the results are eagerly awaited. as an ultrafiltration device, and second for hemodialysis.12
These trials were of relatively short duration, and as such the
CONTINUOUS WEARABLE HEMODIALYSIS DEVICES sorbents were not tested to exhaustion. Laboratory studies
The challenges to design and produce a wearable continuous have suggested that they would be expected to last for at least
hemodialysis treatment are similar to those for PD, but are 24 h, allowing a simple daily exchange. Similarly, ex vivo
complicated by the requirement for blood access. Although experiments suggested a battery life in excess of 24 h. Clotting
continuous renal replacement therapy (CRRT) can deliver of the extracorporeal circuit did occur in some patients when
significantly higher solute clearances than conventional three heparin anticoagulation was withdrawn; however, the circuit
times weekly standard hemodialysis, simple miniaturization used contained many additional temporary ‘T’ connectors to
of the CRRT circuit is not without technical challenges. allow for detailed blood sampling and pressure monitoring.1
Standard hemodialysis machines use roller pumps to As such, these additional connectors added areas of blood
generate countercurrent blood and dialysate flows. To be stagnation and increased the presence of air–blood interfaces,
truly portable, an energy-efficient dual-chamber pump, and as such circuit clotting would be expected to be reduced
capable of being powered by long-life lightweight batteries by careful design of the circuit tubing to minimize pressure
without generating excess heat is required. As such, several drops and areas of stagnation,13 and this device has recently
different pump designs have been suggested to provide been fast-tracked by the US FDA for further clinical trials
adequate flows, but also to reduce protein fouling of the (Figure 3). Other devices designed to use an external blood
dialyzer membrane and mechanical activation of platelets access and to be worn externally on a belt, again using
and leukocytes. As with the AWAK and ViWAK, the system is nanosorbent technology using a mixture of clay and
primed with a small volume of sterile dialysate, which is then biopolymer, nanoporous carbon, and a catalytic adsorbent
regenerated by sorbents.10 After passage through the ion- with ultrafiltration control, are also being developed at an
exchange sorbents, electrolytes, including bicarbonate, have experimental stage (Nanodialysis, Oirschot, Netherlands).
to be returned to ‘refresh’ the regenerated dialysate. As with
standard extracorporeal circuits, anticoagulation is required THE IMPLANTABLE ARTIFICIAL KIDNEY
to prevent clotting, and ultrafiltration needs to be regulated. Although wearable options for both PD and hemodialysis are
The control and monitoring of electrolytes, anticoagulation, now potentially therapeutic options on the horizon for
and ultrafiltration volumes all add complexity to the circuit, patients with CKD, the question arises as to whether these
but can be overcome using small lightweight minipumps, treatments will be suitable for all patients or just a minority,
with servo-controlled safety feedback loops, supported with as with current home hemodialysis programs. Previous
point-of-care testing just as a diabetic would check blood studies looking at the barriers to self-care dialysis have
glucose. The traditional hemodialysis circuit encompasses a shown that the major reasons given by patients were for the
venous blood chamber, which acts as a safety device to most part related to achieving competence with the
prevent air emboli returning to the patient. As blood–air technology used for the treatments,14 and concerns over
interfaces increase the risk of clot formation in the vascular access ranging from needle phobia to undetected
extracorporeal circuit, this standard design has to be changed venous needle dislodgement. Even new dialysis machine
for wearable devices, by utilizing developments in gas- developments designed for the home market, such
permeable, water-impermeable plastics to remove micro- as the NxStage System 1 (NxStage Medical, Lawrence,
bubbles, typically carbon dioxide, which can develop in the Massachusetts), have not managed to recruit large numbers
blood circuit depending on the pressure and temperature. of in-center patients to adopt home therapies, and similarly
Similarly, gas-permeable plastics are also required in the the more frequent hemodialysis trials had difficulty recruiting
dialysate circuit again to remove carbon dioxide generated by sufficient patients willing to accept more frequent and/or
the metabolism of urea by urease in the sorbents. As a longer dialysis sessions.15
wearable hemodialysis device requires blood access, addi- Home-based therapies certainly have advantages over
tional safety features are required to prevent air emboli and center-based dialysis; however, they also tend to place a large
blood loss, using a protective bubble detector sensor placed burden on the caregivers of the dialysis patient, in terms of
after the blood pump, designed to stop blood flow if gas adopting responsibility for the treatment and maintaining
bubbles are detected in the blood circuit, and a second vigilance for complications of therapy. In addition, home-
servomechanism to halt the ultrafiltration pump if the blood based therapies require space not only for dialysis equipment
flow stopped for any reason. In addition, by using a pulsatile but also storage for consumables, increased energy costs, and
blood pump with self-limited capacity to generate negative water usage for home hemodialysis. As dialysis typically
pressure for suction from the arterial side of the vascular remains to be a disease of the disadvantaged, many of these
access, any disconnection on the arterial side would stop the patients lack the financial and housing requirements nece-
blood pump.11 ssary for successful home hemodialysis. Thus, a truly implan-
Currently, a wearable hemodialysis device has been table artificial kidney could have both greater patient appeal
extensively trialed in pig models of acute kidney injury, and and applicability, in addition to reduced complications from

Kidney International (2013) 84, 256–264 259


review WH Fissell et al.: Wearable and implantable dialysis devices

Blood-leak/bubble detector
and pump power-up and
alarm/shut-off system
Dialysate
Shuttle pump Battery Mini regenerating
pump system
Mini
pump

Mini
Mini pump
pump

External flowmeter
Probe measures Dialyzer Blood-leak-detecting probe
blood and dialysate flow rate Bubble-detecting probe

Ultrafiltrate

Figure 3 | Circuit diagram of version 1.0 of the wearable artificial kidney (WAK) showing micro-shuttle pump for countercurrent blood
and dialysate flows, a series of sorbent cartridges, and minipumps for refreshing electrolyte solution, bicarbonate, heparin, and
ultrafiltration control.

an extracorporeal blood loop, which must be repeatedly range from decades (dental implants) to months (implanted
opened and reclosed, thus reducing the risks of access- continuous glucose sensors). The risk, cost, and effort
associated infection and accidental disconnection. associated with medical devices that have either failed or
worn out vary considerably; generator exchanges for pace-
DESIGN SPECIFICATION FOR AN IMPLANTABLE ARTIFICIAL makers are generally low-risk uneventful outpatient proce-
KIDNEY dures, whereas on the other hand an left-ventricular assist
The operational requirements for an implanted renal device device replacement for driveline fracture, hemolysis, or
are shown in Table 1. Medications and dietary modification increased power consumption from bearing wear may not
can help compensate for the loss of some of the physiological even be feasible for the patient. The lifetime of any
functions of the mammalian kidney, and several studies of implantable artificial kidney device anastomosed to the iliac
elderly patients starting dialysis have reported that symptom vessels and connected to the bladder would have to be long
control, survival, and quality of life are similar for those enough to overcome the disadvantage of subsequent removal
treated by dialysis and those managed conservatively with or replacement of the whole device or parts of the device.
drugs, including erythropoietin, activated vitamin D analogs, Vascular access is the first hurdle to overcome in terms of
phosphate binders, bicarbonate, and dietary modifica- device lifetime, whereas synthetic arterial grafts have been
tion.16,17 As such, any implantable kidney would clinically successful; those used as arterio–venous dialysis
predominantly have to achieve two key goals: waste access grafts have often suffered venous clotting problems.
elimination and homeostasis of extracellular fluid volume. Animal studies have shown that, whereas noticeable adherent
Whereas typical vascular access devices for hemodialysis, clot and mural thrombus develops over time using custom-
including catheters, needles, and tubing sets, have significant engineered silicon darts inserted into the femoral vein,
flow resistance such that peristaltic pumps are required not surface modification of the darts with polyethylene glycol
just to circulate blood through the hollow-fiber dialyzer, but appeared to prevent both clot adherence (Figure 4) and
also to overcome viscous flow resistance in the access device, mural thrombi (Figure 5).18
an implantable device anastomosed to the iliac vessels would Clotting of dialyzer fibers often occurs during a standard
potentially have very low flow resistance, obviating the need 4-h hemodialysis treatment session because of fiber protein
for energy-intensive blood pumps. deposition and clot formation due to the activation of
The three dominant considerations in the design en- inflammatory cells and the extrinsic coagulation pathway.19
gineering of an implantable device for treating CKD The hydrostatic pressure drop from blood inlet to blood
(CKD5d) are device lifetime, power requirements, and waste outlet along the length of a hollow-fiber dialyzer leads to
removal. Several other classes of implantable devices are now significant internal filtration within the proximal portion of
in common clinical use, including orthopedic and dental the hollow fiber, and these blood–membrane interactions
implants, pacemakers, implantable defibrillators, and left- actually dominate solute clearance, rather than the prescribed
ventricular assist devices. Service lifetimes for these devices modality.20,21 The concentration polarization of plasma

260 Kidney International (2013) 84, 256–264


WH Fissell et al.: Wearable and implantable dialysis devices review

Table 1 | Replacement of native renal function by dialysis versus miniaturized renal assist device
Physiologic function Native kidney Conventional dialysis Miniaturized RAD
Toxin clearance High-volume convective ultrafiltration, Predominantly diffusive clearance of small High-volume convective ultrafiltration by
followed by active tubular electrolyte, water-soluble solutes from plasma water silicon nanoporous membrane, followed
water, organic anion, and small protein to dialysate, with some convective by selective reabsorption of electrolytes
reabsorption from the ultrafiltrate with clearance of middle-sized molecules, and water by tubule cell bioreactor or
fine tuning by subsequent cellular depending on dialyzer flux and design synthetic ‘tubular’-like membrane
secretion
Volume homeostasis Active distant and intra-renal neuro- Ultrafiltration volume determined Computerized synthesis of physiologic
humoral feedback systems controlling clinically by assessment of dry weight and data and patient-entered weight data to
glomerular blood flow, perfusion pres- ultrafiltration rate by dialysis session time control balance between ultrafiltration
sure, filtration, and tubular reabsorption and reabsorption rates
Calcium–phosphorus Calcium and phosphorus are filtered by As dialysis does not adequately remove Calcium and phosphorus are filtered by
metabolism the glomerulus and differentially reab- sufficient phosphate, patients are the silicon nanoporous membrane, some
sorbed by the tubule under hormonal required to avoid phosphate-rich food- calcium and phosphate is reabsorbed,
and phosphatonin control and tubular stuffs, and require phosphate-binding and any deficiencies will be supplemen-
flow rate medications. Additional active vitamin D ted by oral medications
sterols are required. Dialysate calcium is
adjusted to maintain serum calcium
concentration
Potassium regulation Potassium is filtered by the glomerulus, Patients require a potassium-restricted Patient will self-monitor blood potassium
then reabsorbed by the tubule, and then diet, and then dialysate potassium is at home by finger stick blood sampling,
secreted under hormone control and adjusted to maintain a target pre-dialysis and adjust diet, and if necessary adjust
exchanged to maintain acid–base home- concentration ion exchange resin dose, accordingly
ostasis
Red blood cell mass Renal tubular cells in renal interstitium Require exogenous erythropoiesis-stimu- Require exogenous erythropoiesis-
cells secrete erythropoeitin in an oxygen lating agents, or HIF receptor blockers. stimulating agents, or HIF receptor
delivery–dependent manner regulated Iron supplementation required for blood blockers
by HIF lost in the dialyzer
Acid–base balance The kidney reabsorbs bicarbonate and Patients adhere to a protein-restricted A renal tubular cell bioreactor will excrete
excretes acid by an ammonium-based diet to reduce acid generation, and some acid, but oral bicarbonate will be
buffering system and hydrogen ion dialysate bicarbonate concentration is necessary to completely buffer dietary
exchange for other cations adjusted to control pH acid load, and certainly for a synthetic
‘tubular’ membrane
Abbreviations: HIF, hypoxia inhibitory factor; RAD, renal assist device.

Figure 4 | Scanning electron microscopy images showing explanted darts taken from rat femoral vein showing debris on the uncoated
(left) and clean surfaces on polyethylene glycol–coated (right) darts (adapted from ref. 18).

proteins at the dialyzer membrane surface effectively to dialyzer design is required for a long-term implanted
increases the protein concentration, which may cause some device.
proteins to come out of solution as a gel, or may allow The mammalian nephron accomplishes the majority of
progression of the coagulation cascade to thrombin metabolic waste excretion through a two-stage tandem
deposition and fiber clotting. Protein fouling of a process of nonspecific size-based filtration in the glomerulus,
biomaterial is a complex sequence of events driven by very followed by transcellular and paracellular transport of water
high–frequency contact interactions between solutes and and a small number of solutes (primarily electrolytes, urea,
surface, followed by reversible binding and then a conforma- glucose, amino acids, and small proteins), so that waste
tional change in the protein with exposure of hydrophobic products present in about 100–140 l of filtered blood each day
epitopes and irreversible loss of solubility.22,23 As such, are concentrated into a fluid volume small enough (1–2 l)
to achieve adequate device longevity, a different approach that it can be balanced by oral intake. In contrast, standard

Kidney International (2013) 84, 256–264 261


review WH Fissell et al.: Wearable and implantable dialysis devices

intermittent hemodialysis requires B240 l of ultrapure water size.26 Although in typical log-normal pore size distributions
for each treatment, a volume and mass that is impractical for large pores are relatively few in number, the largest pores
implanted or portable therapies. Dialysate regeneration by contribute significantly to the water volume flux through the
sorbents has long been used to minimize water requirements, dialyzer membrane and thus disproportionately increase the
but requires fresh reagents for each treatment as they become leakage of plasma proteins into the ultrafiltrate. For standard
saturated, thus limiting the applicability of sorbent technol- dialysis membranes, this effect is ameliorated by shifting the
ogy to implantation devices. entire distribution of pores toward smaller pore sizes, and
Therefore, a new type of dialyzer membrane was required thus reducing the number of large pores sufficiently to limit
for an implantable device with a resistance low enough to protein leakage. This then reduces hydraulic permeability and
allow 300 ml/min of well-mixed blood to flow from the compromises clearance of so-called ‘middle molecules’
arterial side of the device through to the venous return, yet between 5 and 25 kD.
only driven by a native cardiac perfusion of 80–100 mm Hg. Silicon micromachining is a technological development
At the same time, the dialyzer membrane also had to have capable of producing very high numbers of miniaturized
sufficient hydraulic permeability such that lower hydrostatic electromechanical devices with extraordinary repeatability
pressures of 30–50 mm Hg within the dialyzer could drive with low unit cost,27 and adaptation of this technology to
30 ml/min of small solute clearance by convection. It would filtration allows nearly complete control over pore size and
appear that nature, as well as man, has chosen elongated pore shape within Angstrom-level fidelity across several centimeters
structures for filtration in many settings, from the baleen of silicon wafer, to produce flat-sheet membranes of highly
plates of filter-feeding whales to the slots in storm drains in regular elongated slit-shaped pores of 5–10 nm (Figure 6).
city streets, elongated structures provide steric hindrance to In vitro and in vivo testing of these prototype membranes
large objects without blocking fluid flow. This would also showed that their performance matched predicted hydraulic
appear to hold true in the kidney, where the glomerular slit permeability and steric and electrostatic hindrances,28 and
diaphragm appears to consist of elongated slots of at least outperformed conventional polymer membranes with round
two very different length scales,24,25 whereas the standard pores.29 Blood–materials interactions were examined both
dialyzer polymer membranes have irregular, round pore in vitro and in vivo. Polyethylene glycol surface-modified
structures with considerable variability in individual pore silicon nanopore membranes were tested in an ultrafiltration
cell using citrated bovine blood, and hydraulic permeability
and sieving curves remained stable over 96 h of filtration, with
no detectable change from the initial first hour to that during
the 96th hour of the study.
For a successful implantable artificial kidney, the ‘glo-
merular’ membrane has to be linked to a ‘tubule’ membrane
designed to reabsorb ultrafiltrate, so that patients only pass
2 l of urine daily, yet excrete an adequate amount of waste
products. Although a ‘tubule’ membrane that can reabsorb
water and solutes could be engineered similarly, in some
ways, to the membranes used in reverse osmosis water
systems, one that could differentiate which solutes to
reabsorb and which to excrete has not yet been developed.
100 µm
One option would be to consider using a bioreactor to solve
problems that engineering alone could not resolve. Renal
Figure 5 | Light microscopy of vein removed from a rat that had
an uncoated silicon dart inserted showing thrombus formation tubule cell bioreactors have been maintained in perfused
(adapted from ref. 18). tissue culture for extended periods of time before clinical use.

a b c

7 nm pore

Figure 6 | Scanning electron microscope images showing silicon nanopore membranes. (a) Low magnification showing an array of
rectangular membranes; (b) higher magnification showing the pores on a single membrane; and (c) tilted, high magnification showing
a close-up of the slit pore.

262 Kidney International (2013) 84, 256–264


WH Fissell et al.: Wearable and implantable dialysis devices review

During and after deployment in an extracorporeal circuit,


when cell nutrition and respiratory gas exchange was entirely
dependent on a combination of ultrafiltrate flow from a
hemofilter through the hollow fibers coated with renal
tubular cells, harvested from renal transplant discards, and
blood flow outside the hollow fibers of the bioreactor, human
renal tubular cells not only maintained their viability but
also demonstrated their epithelial integrity by low-inulin
leak rates.30–34 Similarly, when grown to confluence on
silicon nanopore membranes,35 indirect immunofluorescence
imaging showed localization of zona occludens-1 to cell–cell
junctions and acetylated tubulin in cilia, suggesting a
differentiated phenotype (Figure 7). In addition, the transe-
pithelial resistance of these tubular epithelial monolayers grown
on silicon membranes was similar to that of monolayers grown Figure 7 | Fluorescent microscopy image of human renal
proximal tubule cells cultured from discarded renal transplant
on conventional tissue culture inserts. More importantly, when kidneys on silicon nanopore membranes. Blue stain refers to
grown in a calibrated shear-stress bioreactor, these renal cell nuclei, green refers to tight junctions, and red refers to cilia
proximal tubule cells showed shear-dependent rearrangement (adapted from ref. 34).
of cytoskeletal proteins and shear-dependent solute trans-
port.36,37 The question arises, as with other bioartifical devices,
as to what cell mass is required to achieve required function.
As such, mass balance considerations dictate the operating
characteristics of the renal cell bioreactor. A target small solute
clearance of 30 ml/min implies that a very similar volume must
be reabsorbed in the bioreactor. Isolated perfused tubules have
been shown to reabsorb filtrate between 20 and 40 ml/min/m2,
suggesting that a bioreactor surface area of less than a square
meter should suffice to match reabsorption to filtration.38,39
Further work is needed to determine the interplay between
bioreactor geometry and the minimum fractional excretion of
sodium required.
As such, a renal tubular–based bioreactor linked to a
synthetic ‘glomerular’ membrane could potentially provide a
reabsorptive capacity for the ultrafiltrate, as well as providing
some additional tubular metabolic control. This would be
surgically implanted using iliac vessel access (Figure 8).
However, as the device would not be able to replace normal Figure 8 | Schematic design for an implantable artificial kidney
device, using iliac vessels for arterial blood inflow and venous
renal function, patients would still rely on medications return, with ultrafiltratate draining into the bladder.
(table 1) to maintain homeostasis.
proof-of-concept experiments have demonstrated the feasi-
SUMMARY bility of each component of an implanted artificial kidney
Although it has taken more than 50 years to develop a comprising a novel ‘glomerular’ membrane technology,
prototype of a truly wearable artificial kidney for the optimizing membrane separation process coupled with a
treatment of patients with CKD5d, two devices based on bioreactor of living cells harvesting energy, and solute
current hemodialysis and PD paradigms are currently reabsorption from the ‘glomerular’ membrane’s plasma
embarking on clinical trials. How successful these devices ultrafiltrate, recapitulating nephron anatomy.
are will depend not only on their efficacy in terms of solute We are therefore potentially at a crossroads for new
removal but also their ability to maintain electrolyte, paradigms for the treatment of CKD5d patients with
acid–base and volume homeostasis, patient acceptance, and wearable and implantable devices on the near and far
costs. It may well be that these treatments are not suitable for horizon, respectively, that could in the future not only
all patients, but could potentially offer more patients the improve patient survival but also the quality of life.
advantages of both more frequent and longer dialysis
treatments than current in-center-based hemodialysis pro-
DISCLOSURE
grams, with improved quality of life for patients. Implantable Drs Fissell and Roy are primary inventors on one or more patents
devices on the other hand potentially offer a treatment related to implantable artificial kidneys, and principals of Silicon
solution for the majority, and over the past few years key Kidney, LLC.

Kidney International (2013) 84, 256–264 263


review WH Fissell et al.: Wearable and implantable dialysis devices

REFERENCES 20. Hofmann CL, Fissell WH. Middle-molecule clearance at 20 and 35 ml/kg/h
1. Davenport A, Gura V, Ronco C et al. A wearable hemodialysis device for in continuous veno-venous haemodiafiltration. Blood Purif 2010; 29:
patients with end-stage renal failure: a pilot study. Lancet 2007; 370: 259–263.
2005–2010. 21. Messer J, Mulcahy B, Fissell WH. Middle-molecule clearance in CRRT: in-
2. Bomar JB, Decherd JF, Hlavinka DJ et al. The elucidation of maximum vitro convection, diffusion, and dialyzer area. ASAIO J 2009; 55: 224–226.
efficiency-minimum cost peritoneal dialysis protocols. Trans Am Soc Artif 22. Holland NB, Qiu Y, Ruegsegger M et al. Biomimetic engineering of non-
Intern Organs 1974; 20A: 120–129. adhesive glycocalyx-like surfaces using oligosaccharide surfactant
3. Moncrief JW, Popovich RP, Nolph KD. The history and current status of polymers. Nature 1998; 392: 799–801.
continuous ambulatory peritoneal dialysis. Am J Kidney Dis 1990; 16: 23. Vacheethasanee K, Marchant RE. Surfactant polymers designed to
579–584. suppress bacterial (Staphylococcus epidermidis) adhesion on
4. Ronco C, Fecondini L. The Vicenza wearable artificial kidney for peritoneal biomaterials. J Biomed Mat Res 2000; 50: 302–312.
dialysis (ViWAK PD). Blood Purif 2007; 25: 383–388. 24. Blantz RC, Gabbai FB, Peterson O et al. Water and protein permeability is
5. Ronco C, Davenport A, Gura V. A wearable artificial kidney: dream or regulated by the glomerular epithelial slit diaphragm. J Am Soc Nephrol
reality? Nat Clin Pract Nephrol 2008; 4: 604–605. 1994; 4: 1957–1964.
6. Davenport A. Role of dialysis technology in the removal of uraemic 25. Rodewald R, Karnovsky MJ. Porous substructure of the glomerular slit
toxins. Hemodial Int 2011; 15(Suppl 1): S49–S53. diaphragm in the rat and mouse. J Cell Biol 1974; 60: 423–433.
7. Lee DBN, Roberts M. A peritoneal based automated wearable artificial 26. Zeman LJ, Zydney AL. Microfiltration and Ultrafiltration. Marcel Dekker:
kidney. Clin Exper Nephrol 2008; 12: 171–180. New York, 1996, pp 180–253.
8. Maroni BJ, Steinman TI, Mitch WE. A method for estimating nitrogen 27. Davenport A. Membrane designs and composition for haemodialysis,
intake of patients with chronic renal failure. Kidney Int 1985; 27: 58–65. haemofiltration and haemodialfiltration: past, present and future. Minerva
9. Roberts M, Lee DBN. Wearable artificial kidneys. A peritoneal dialysis Urol Nefrol 2010; 62: 29–40.
approach. Dial Transplant 2006; 36: 780–782. 28. Fissell WH, Dubnisheva A, Eldridge AN et al. High-performance silicon
10. Gura V, Davenport A, Beizai M et al. Beta 2-microglobulin and phosphate nanopore hemofiltration membranes. J Memb Sci 2009; 326: 58–63.
clearances using a wearable artificial kidney: a pilot study. Am J Kidney Dis 29. Kanani DM, Fissell WH, Roy S et al. Permeability-selectivity analysis for
2009; 54: 104–111. ultrafiltration: Effect of pore geometry. J Memb Sci 2010; 349: 405–410.
11. Gura V, Ronco C, Davenport A. The wearable artificial kidney, why and 30. Fissell WH, Kimball J, MacKay SM et al. The role of a bioengineered
how: from holy grail to reality. Semin Dial 2009; 22: 13–17. artificial kidney in renal failure. Ann N Y Acad Sci 2001; 944: 284–295.
12. Gura V, Ronco C, Nalesso F et al. A wearable haemofilter: first human 31. Humes HD, Fissell WH, Weitzel WF et al. Metabolic replacement of kidney
study of slow continuous ambulatory ultrafiltration. Kidney Int 2008; 74: function in uremic animals with a bioartificial kidney containing human
497–502. cells. Am J Kidney Dis 2002; 39: 1078–1087.
13. Davenport A. Anticoagulation for continuous renal replacement therapy. 32. Tumlin J, Wali R, Williams W et al. Efficacy and safety of renal tubule cell
Contrib Nephrol 2004; 144: 228–238. therapy for acute renal failure. J Am Soc Nephrol 2008; 19: 1034–1040.
14. Cafazzo JA, Leonard K, Easty AC et al. Patient-perceived barriers to the 33. Humes HD, Buffington DA, MacKay SM et al. Replacement of renal
adoption of nocturnal home haemodialysis. Clin J Am Soc Nephrol 2009; function in uremic animals with a tissue-engineered kidney. Nat
4: 784–789. Biotechnol 1999; 17: 451–455.
15. Sergeyeva O¸, Gorodetskaya I, Ramos R et al. Frequent Hemodialysis 34. Humes HD, Weitzel WF, Bartlett RH et al. Initial clinical results of the
Network Trials Group. Challenges to enrollment and randomization of the bioartificial kidney containing human cells in ICU patients with acute
Frequent Hemodialysis Network (FHN) Daily Trial. J Nephrol 2012; 25: renal failure. Kidney Int 2004; 66: 1578–1588.
302–309. 35. Fissell WH, Manley S, Westover A et al. Differentiated growth of human
16. Carson RC, Juszczak M, Davenport A et al. Is maximum conservative renal tubule cells on thin-film and nanostructured materials. ASAIO J
management an equivalent treatment option to dialysis for elderly 2006; 52: 221–227.
patients with significant comorbid disease? Clin J Am Soc Nephrol 2009; 4: 36. Ferrell N, Desai RR, Fleischman AJ et al. A microfluidic bioreactor with
1611–1619. integrated transepithelial electrical resistance (TEER) measurement
17. Caplin B, Kumar S, Davenport A. Patients’ perspective of haemodialysis- electrodes for evaluation of renal epithelial cells. Biotechnol Bioengin
associated symptoms. Nephrol Dial Transplant 2011; 26: 2656–2663. 2010; 107: 707–716.
18. Melvin ME, Fissell WH, Roy S et al. Silicon induces minimal 37. Ferrell N, Ricci KB, Groszek JJ et al. Albumin handling by renal tubular
thromboinflammatory response during 28-Day intravascular implant epithelial cells in a microfluidic bioreactor. Biotechnol Bioengin 2011; 109:
testing. ASAIO J 2010; 56: 344–348. 797–803.
19. Vernon K, Peasegood J, Riddell A et al. Dialyzers designed to increase 38. Burg MB, Orloff J. Control of fluid absorption in the renal proximal tubule.
internal filtration do not result in significantly increased platelet J Clin Invest 1968; 47: 2016–2024.
activation and thrombin generation. Nephron Clin Pract 2010; 117: 39. Kokko JP, Burg MB, Orloff J. Characteristics of NaCl and water transport in
c403–c408. the renal proximal tubule. J Clin Invest 1971; 50: 69–76.

264 Kidney International (2013) 84, 256–264


Copyright of Kidney International is the property of Nature Publishing Group and its content
may not be copied or emailed to multiple sites or posted to a listserv without the copyright
holder's express written permission. However, users may print, download, or email articles for
individual use.

Вам также может понравиться