Академический Документы
Профессиональный Документы
Культура Документы
Agilent Confidential
25 October 2018
1
Agilent 1260 Infinity II LC
Maintenance
Confidentiality Label
October 25, 2018
3
Confidentiality Label
October 25, 2018
4
Confidentiality Label
October 25, 2018
5
Confidentiality Label
October 25, 2018
6
Confidentiality Label
October 25, 2018
7
Confidentiality Label
October 25, 2018
8
Confidentiality Label
October 25, 2018
9
Confidentiality Label
October 25, 2018
10
Confidentiality Label
October 25, 2018
11
Confidentiality Label
October 25, 2018
12
Confidentiality Label
October 25, 2018
13
Pump Maintenance
Confidentiality Label
October 25, 2018
14
Confidentiality Label
October 25, 2018
15
Confidentiality Label
October 25, 2018
16
Confidentiality Label
October 25, 2018
17
Confidentiality Label
October 25, 2018
18
Auto sampler inject sequence
Confidentiality Label
October 25, 2018
19
Confidentiality Label
October 25, 2018
20
Sampler Maintenance
Confidentiality Label
October 25, 2018
21
Sampler Maintenance
Confidentiality Label
October 25, 2018
22
Confidentiality Label
October 25, 2018
23
Confidentiality Label
October 25, 2018
24
HPLC connectors from different manufacturers.
Confidentiality Label
October 25, 2018
25
Confidentiality Label
October 25, 2018
26
Detector Maintenance
Confidentiality Label
October 25, 2018
27
Confidentiality Label
October 25, 2018
28
Confidentiality Label
October 25, 2018
29
Confidentiality Label
October 25, 2018
30
Cleaning a reversed phase silica column
Sometimes you can alleviate a high pressure or peak tailing issue by cleaning your column. Before cleaning your
column, we recommend that you disconnect it from your detector and run your wash solvents into a beaker or other
container. In some cases, it is advisable to backflush the column while cleaning it. This keeps the column
contamination from flowing through the column. Check with your column manufacturer, or review the instructions that
came with your column before doing this.
Cleaning the column requires solvents that are stronger than your mobile phase. You should use at least ten column
volumes (see figure 29) of each solvent for cleaning analytical columns. See the following procedure for cleaning
reversed phase columns. When finished cleaning your column, return your column to the flow direction recommended
by the manufacturer.
Confidentiality Label
October 25, 2018
31
Agilent 1260 Infinity II LC
Troubleshooting
LC possible troubles
Confidentiality Label
October 25, 2018
33
Troubleshooting Steps
Confidentiality Label
October 25, 2018
34
System Check
• Solvent filters and Bottle heads installed?
Confidentiality Label
October 25, 2018
35
System Check
Often a simple operator error is the root cause of a problem: gather information!
• Has something been changed on the LC system?
• Is the correct column installed?
• Are the correct solvents being used (including modifier)?
• Is the flow rate set properly?
• Is it the right sample in the vial, dissolved in the correct solvent?
• Has the method been changed?
• ...
Visual inspection for
• Correct connections
• Air bubbles
• Leaks
•…
Check pressure: increased pressure points to a restriction of flow in the system, most
commonly a blocked frit
8
Confidentiality Label
October 25, 2018
36
Troubleshooting: Leaks
Confidentiality Label
October 25, 2018
37
Troubleshooting: Leaks
Confidentiality Label
October 25, 2018
38
Troubleshooting: Frits Blocking
• Absence/dirty solvent filters. Never use system without solvent filters. This
will clog degasser and will create issue in the gradient valve.
• Degasser Do not leave water in degasser, this will result in algae growth and
contamination will result in frit blocking.
• Never use old\regenerated frits. Always use a new frit. Otherwise particles
will block tubing after the pump.
Confidentiality Label
October 25, 2018
39
Troubleshooting: Fluctuating Pressure
Confidentiality Label
October 25, 2018
40
Troubleshooting: Degasser
Pressure Fluctuations
• Solvent Inlet Filters – Block or contaminated
• Change degasser channels/ check for solvent flow
• Bypass degasser
Confidentiality Label
October 25, 2018
41
Troubleshooting: Pump
Pressure Fluctuations:
• Use pressure plot
• Use pump leak test (system flushed?!)
• Check purge valve frits
• Binary pump: check channel by channel
• Pump maintenance – All parts should be clean and leak free
• Leak
• Symptomes:
• Ripple > 10% → AIV cartridge
• Ripple 3% to 4% → Outlet ball valve/pump seal
Gradient Problem:
• Pump – Mixing issue (Gradient valve)
• Filters – Blocked or contaminated
• Pressure variation
• Column stabilization
• Change of solvent brand
Confidentiality Label
October 25, 2018
42
Troubleshooting: Autosampler
Pressure Fluctuations
• Switch between main pass and bypass. Pressure difference
should be 3 to 5 bar.
• Use needle-up in main pass (needle vs. needle seat)
• Check waste volume corresponding to injection volume
• Use Agilent materials (vials, plates, mats)
• Perform a tray alignment/autoreferencing
Carry Over
• Washing vial Should be without septa
• Washing solvent Choose correct solvent
• Needle seat Should be clean, leak free and without
scratches
• Injection valve parts Rotor seal age, number of
injections
Area variation
• ALS maintenance Inj. Vol. variation/carry over
• Waste line location
• Sample degration, check cooling performance
• Injection volume Check injection volume by calibration
Confidentiality Label
October 25, 2018
43
Troubleshooting: Detector
Confidentiality Label
October 25, 2018
44
Overview poor retention reproducibility
Confidentiality Label
October 25, 2018
45
Troubleshooting: Decreasing Retention Time
Confidentiality Label
October 25, 2018
46
Troubleshooting: Increasing Retention Time
Confidentiality Label
October 25, 2018
47
Troubleshooting: Fluctuating Retention Time (1)
Confidentiality Label
October 25, 2018
48
Troubleshooting: Fluctuating Retention Time (2)
Confidentiality Label
October 25, 2018
49
Troubleshooting: Temperature and pH
• Change in column temperature can change retention time and peak spacing
• Causes for fluctuating temperature:
• Temperature in lab circulates in course of the day (without column oven)
• Increased solvent temperature due to friction through clogged frit
• Peltier elements, sensors, heaters, board… fail (with column oven)
• Change in temperature can lead to change in pH of buffer
• Small changes in pH lead to huge effects on tr
Confidentiality Label
October 25, 2018
50
Peak Shape: Peak Splitting\Tailing\Fronting
Confidentiality Label
October 25, 2018
51
Confidentiality Label
October 25, 2018
52
Agilent Lab Advisor
Confidentiality Label
October 25, 2018
53
Confidentiality Label
October 25, 2018
54