PURELAB ®
Laboratory water purification systems
Page 3 of 46
Page73
Page 4 of 46
Page74
PURELAB
1-2
PU
R
pro ELAB
d
fin uct
der
Matching systems to applications
supporting you all the way
Water purification is often a complex process that incorporates multiple technologies into a single
purification system; therefore, we have developed the product finder to help you select the best
water purification system suited to your application needs. Our team of technical and application
specialists is always available to answer any questions, as well as providing friendly, expert advice
in choosing the most suitable water purification system.
1000
liters/day 100
Pulse (Page 11)
flex 3&4 (Page 7 & 8)
10
Water analysis
* For more information on applications and scientific water standards see our education guidebook
“The Pure LabWater Guide”. A free copy is available at www.elgalabwater.com
Page 5 of 46
Page75
Worldwide service
and support
For total confidence in your
pure water systems
Every PURELAB comes complete with one extra feature – first class service
and support wherever you are in the world. ELGA has installed thousands
of systems globally, and our service engineers will apply their expertise
to the installation, validation and maintenance of your water purification
systems in compliance with all the relevant codes.
Visit www.elgalabwater.com to find your nearest contact.
Page 6 of 46
Page76
Validation support
Water systems that are employed within a validated laboratory should Quality assured
always operate within specification. This should be evidenced through Our commitment to
documentation and trending should warn the user if performance is likely the highest quality
to fall below their requirements. ELGA offers a market-leading validation control processes
guarantees reliability
package with fully trained personnel using controlled equipment and and compliance with
documentation to support you throughout the validation process. both international,
environmental and
ELGA products are continually being developed to further support user organization
standards
validation requirements. Developments include:
Designed and
manufactured under
• PIN-coded access to software set points an ISO9001:2000
prevents unauthorized changes to operation total quality system.
or system settings Tested to comply
with CE, EMC, EN
• Cartridge Identification (CID) technology
61010 (UL CSA), PIRA,
provides full traceability of each cartridge for
WEEE Directive and
GLP and other validation requirements
other standards as
• RS232 data port to enable appropriate.
downloading of data to PC or printer
• Water quality data available with or
without temperature compensation
Page 7 of 46
Page77
Page 8 of 46
Page78
Ultra
5-6
9
1 Efficient fast rinse 1
10 Full spectrum UV treatment
10
ultrafiltration cartridge
• Low microbial and organic
• Filters out endotoxin, specification
proteins and nucleases
• High transmittance synthetic
• Rinses up to 7 times faster quartz sleeve for higher
than other endotoxin filters efficiency
• Significantly reduces costs • Lamp failure alarm
of sanitization
Page 9 of 46
Page79
PURELAB flex
Designed for today’s laboratories
The most intuitive and innovative system for all your
pure water needs
2
4
6
7 Data capture via USB for system
performance validation
3
Page 10 of 46
Page80
flex
7-8
1 5
4
1 Intuitive to use, ergonomic
3 4 Prioritized information
handset design
shown at all times (system
status, TOC, alarm)
2 Clear water purity display for
absolute confidence as you dispense
5 Flexible dispensing in four different ways:
• Variable flow – drop by drop or up to
3 Realtime TOC monitoring 2 liters per minute
for critical analytical • Autovolume dispense from 50ml to
applications 60 liters and repeat dispensing
• Hands free with optional foot pedal
• Locked dispense for glassware filling
PURELAB flex 2 PURELAB flex 3&4 PURELAB flex 1 PURELAB flex 3&4 PURELAB flex 1
Daily volume >10 Liters <10 Liters >10 Liters <10 Liters >10 Liters
Water quality Type I ultrapure water Type II/Type II+ pure water Type III RO water
Page 11 of 46
Page81
PURELAB Classic
Economy without compromise for high performance,
ultrapure water production
2
Efficient fast rinse
ultrafiltration cartridge
• F ilters out endotoxins, proteins
3 5 Single purification pack
and nucleases
1 design, easily upgraded
to dual pack
•R
inses up to 7 times faster than 2
Page 12 of 46
Page82
“Ultrapure
water systems
to match your
budget”
PURELAB Option-Q
Ultrapure water direct from potable tap water
2
Fully recirculating UV ensures the 4
2
highest microbial quality
6 Optional ELGA Biofilter –
when fitted to the Option-Q
produces water which is
free from biologically active
1
3 Height adjustable point-of-use impurities
dispense tap for a choice of wall
or bench mounting
Page 13 of 46
Page83
PURELAB Pulse
The only Electrodeionization Pulse Technology
4
1 Space-saving wrap-around
reservoir with smooth sides
and vented filtration for
optimized water purity 4 Full recirculation
3
guarantees water quality
at point-of-use
2 Fully recirculating UV 2
5
ensures the highest 5 Pulse Module
microbial quality (EDI purification)
Page 14 of 46
Page84
PURELAB Option-R
Designed to meet The PURELAB solution for overcoming your water
and exceed Type II+/II purification challenges
1 Space-saving wrap-
5
around reservoir with
3
smooth sides and
vented filtration for
optimized water purity 4 Height adjustable point-
of-use dispense tap for
choice of wall or bench
2 Fully recirculating 1 mounting
UV ensures the
highest microbial 2
quality
5 Full recirculation
guarantees water
4
quality at point-of-use
3 Deionization
purification
(ion-exchange)
cartridge
Page 15 of 46
Page85
“The compact
solution for small
distribution
loops”
PURELAB
PURELAB 7000
The PURELAB 7000 dispenses up to 4 liters/min
of Type II+ water (>10 MΩ-cm) and can be
connected to small distribution loops to feed
an ultra pure system and two or three dispense
points, as well as your main application.
Designed to meet Type II or Type III water quality The system has an in-built reservoir to ensure
your required pure water capacity is met at
peak times.
When general laboratory and primary grade water is
needed at high flow rates the PURELAB 3000 & 7000
offer a reliable and consistent supply with the
flexibility to meet your usage needs.
1
Reliable water quality
3 4
Economic Purity
•U
nique sanitization 1
3 Unique system
3 operation protection
• Easy-to-use Building Management System
• Automated alarms
Page 16 of 46
Page86
PURELAB
PURELAB Prima
The ultimate in flexibility for primary grade water
The compact PURELAB Prima systems provide Type III water and are
available in a choice of flow rates, 7, 15, 30, 60, 90, or 120 liters/hr, to meet
your specific needs.
Page 17 of 46
Page87
Contact us: ELGA is the global laboratory water brand name of Veolia Water Solutions
and Technologies. The information contained in this document is the
property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied
ELGA offices and distributors are located in more than
without liability for errors or omissions. © VWS (UK) Ltd. 2011 – All
60 countries and are fully trained in all ELGA systems.
rights reserved. ELGA®, PURELAB®, ADEPT® Labpure® and PureSure® are
To find your nearest ELGA representative, go to registered trademarks of VWS (UK) Ltd.
www.elgalabwater.com and select your country for Your local ELGA representative
contact details.
ELGA Global Operations Centre
Tel: +44 1494 887 500
Fax: +44 1494 887 505
Email: info@elgalabwater.com
Web: www.elgalabwater.com
www.elgalabwater.com
Page 18 of 46
Page88
PURELAB
Analytical Research
Water Outlet to
Quality Docking Vessel
Sensor or Reservoir
Drain
PURELAB
Analytical Research
Feedwater Requirements
Source quality Potable mains water supply Potable mains water supply Potable mains water supply
Fouling index - maximum 10 10 10
Conductivity 2000 µS/cm* 2000 µS/cm* 2000 µS/cm*
Free chlorine - maximum 0.5 ppm 0.5 ppm 0.5 ppm
Heavy metals - maximum 0.05 ppm 0.05 ppm 0.05 ppm
Silica - maximum 30 ppm 30 ppm 30 ppm
Temperature 1 - 35°C 1 - 35°C 1 - 35°C
Flowrate (maximum requirement) 145 l/hr 150 l/hr 155 l/hr
Drain requirements (gravity fall 70 l/hr 70 l/hr 80 l/hr
with air gap). Maximum during
service
Feedwater Pressure
Maximum - without internal boost pump 6.0 bar (90 psi) 6.0 bar (90 psi) 6.0 bar (90 psi)
Minimum - without internal boost pump 4.0 bar (60 psi) 4.0 bar (60 psi) 4.0 bar (60 psi)
Maximum - with internal boost pump 2.0 bar (30 psi) 2.0 bar (30 psi) 2.0 bar (30 psi)
Minimum - with internal boost pump Flooded Suction Flooded Suction Flooded Suction
* Deionization cartridge life may vary with feedwaters >1400 µS/cm
Electrical Requirements
Mains input 100-240V ac, 50-60Hz all models 100-240V ac, 50-60Hz all models 100-240V ac, 50-60Hz all models
System voltage 24V dc 24V dc 24V dc
Power consumption with boost pump 52VA 52VA 52VA
Power consumption without boost 28VA 28VA 28VA
pump
Fuses 2 x T3.15 Amp 2 x T3.15 Amp 2 x T3.15 Amp
Reservoir level connection Jack Plug 3.5mm Jack Plug 3.5mm Jack Plug 3.5mm
Noise level <45dBA <45dBA <45dBA
ELGA LabWater
Tel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505
Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained
in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or
omissions. © VWS (UK) Ltd. 2012 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR 38751-05
PURELAB
Analytical Research
Break Tank
Feedwater
Inlet Pre-Filter Outlet
to Reservoir
Water Water
Quality Boost Quality
Sensor Pump Sensor
Reverse Reverse
Osmosis Osmosis
Drain
Page 21 of 46
Page91
PURELAB
Analytical Research
1
MW > 100 Dalton 2 With suitable pretreatment, if required, depending on the feedwater
3
Subject to correct operating and maintenance procedures
Height 740mm
Width 570mm
Depth 320mm
Weight (operational) 47kg 48kg 54kg
Feedwater Requirements
Source quality Potable mains water supply
Fouling index – maximum 10
Conductivity – maximum 2000 µS/cm
Free chlorine – maximum <0.5 ppm
Heavy metals – maximum 0.05 ppm
Temperature 1 – 35°C
Flowrate (maximum requirement) 9 l/hr
Feedwater pressure maximum 6.0 bar (90 psi)
Feedwater pressure minimum 1.4 bar (20 psi)
Electrical Requirements
ELGA LabWater
Tel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505
Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this
document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions.
© VWS (UK) Ltd. 2011 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR 38752-04
Page 22 of 46
Page92
PURELAB
Analytical Research
• Features ‘Reverse Osmosis feed optimized’ resin mixes to increase the ion exchange
capacity of consumables and minimize running costs The only fully
deionization
• Quick and easy sanitization and replacement of consumables to reduce
maintenance time recirculating Type II
pure water system
• PURELAB Option systems are designed to be easy to access whether wall or bench
mounted with a convenient dispense tap. They can be used with our wrap-around
reservoirs to save space whilst optimizing purity
Process Flow PURELAB Option-R 7/15
Process Flow PURELAB Option-R 7/15
Water Quality Sensor
Outlet Outlet
Pre- Dispense Tap
Treatment
Reverse Reverse
Osmosis Osmosis Water Outlet to
Quality Docking Vessel
Sensor or Reservoir
UV Lamp Deionization
Cartridge
Drain
Page 23 of 46
Page93
PURELAB
Analytical Research
Feedwater Requirements
Source Quality Potable mains water Potable mains water Potable mains water Potable mains water
supply supply supply supply
Fouling index - maximum 10 10 10 10
Total dissolved solids - maximum <2000 µS/cm* <2000 µS/cm* <2000 µS/cm* <2000 µS/cm*
Free chlorine - maximum 0.5 ppm 0.5 ppm 0.5 ppm 0.5 ppm
Heavy metals - maximum 0.05 ppm 0.05 ppm 0.05 ppm 0.05 ppm
Silica - maximum 30 ppm 30 ppm 30 ppm 30 ppm
Temperature 1 - 35°C 1 - 35°C 1 - 35°C 1 - 35°C
Flowrate (maximum requirement) 78 l/hr 85 l/hr 78 l/hr 85 l/hr
Drain requirements (gravity fall 70 l/hr
with air gap). Maximum during
service
Feedwater Pressure
Maximum - without internal boost pump 6.0 bar (90 psi) 6.0 bar (90 psi) 6.0 bar (90 psi) 6.0 bar (90 psi)
Minimum - without internal boost pump 4.0 bar (60 psi) 4.0 bar (60 psi) 4.0 bar (60 psi) 4.0 bar (60 psi)
Maximum - with internal boost pump 2.0 bar (30 psi) 2.0 bar (30 psi) 2.0 bar (30 psi) 2.0 bar (30 psi)
Minimum - with internal boost pump Flooded Suction Flooded Suction Flooded Suction Flooded Suction
* Deionization cartridge life may vary with feedwaters >1400 µS/cm
Electrical Requirements
Mains input 100-240V ac, 50-60Hz 100-240V ac, 50-60Hz 100-240V ac, 50-60Hz 100-240V ac, 50-60Hz
System voltage 24V dc 24V dc 24V dc 24V dc
Power consumption with boost pump 43VA 43VA 80VA 80VA
Power consumption without 19VA 19VA 50VA 50VA
boost pump
Fuses 2 x T3.15 Amp 2 x T3.15 Amp 2 x T6.3 Amp 2 x T6.3 Amp
Reservoir level connection Jack Plug 3.5mm Jack Plug 3.5mm Jack Plug 3.5mm Jack Plug 3.5mm
Noise level <45dBA <45dBA <45dBA <45dBA
ELGA LabWater
Tel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505
Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this
document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions.
© VWS (UK) Ltd. 2011 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR 38753-03/1
Page 24 of 46
Page94
PURELAB
Analytical Research
• Features ‘RO feed optimized’ resin mixes to increase the ion exchange capacity of
consumables and minimize running costs
Type II pure water
• Quick and easy sanitization and replacement of consumables to reduce
maintenance time
on demand at
higher flow-rates
• PURELAB Option systems are designed to be easy to access whether wall or bench
mounted with a convenient dispense tap
Break Tank
Water
Quality Outlet
Feedwater Sensor to Reservoir
Inlet Pre-Filter
Pump Water
Quality
Sensor
Reverse
Osmosis
Dispense
Tap
Return Deionization
to Break Tank Cartridge
Drain UV Lamp
Pump Feed
from Reservoir
Recirculation
Pump
Page 25 of 46
Page95
PURELAB
Analytical Research
Feedwater Requirements
Source quality Potable mains water supply
Fouling index - maximum 10
Total dissolved solids - maximum 2000µS/cm*
Free chlorine - maximum <0.5 ppm
Heavy metals - maximum 0.05 ppm
Temperature 1 - 35°C
Flowrate (maximum requirement) 9 l/min
Feedwater pressure maximum 6.0 bar (90 psi)
Feedwater pressure minimum 1.4 bar (20 psi)
* Deionization cartridge life may vary with feedwaters >1400 µS/cm
Electrical Requirements
Mains input 230V ac, 50Hz
115V ac, 60Hz
100V ac, 50/60Hz
Power consumption with 700VA
boost pump
Fuses 2 x 3.15 Amp – 230V ac, 50Hz
2 x 6.3 Amp – 115V ac, 60Hz
Reservoir level connection Din plug 6 way
ELGA LabWater
Tel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505
Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this
document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions.
© VWS (UK) Ltd. 2011 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR 38755-04
Page 26 of 46
Page96
PURELAB
Analytical Research
PURELAB Option-Q
Tap to type I ultrapure water direct from a potable tap water supply. The PURELAB Option-Q is
ideal for laboratories who need up to 100 liters of 18.2 MΩ.cm ultrapure water. With its simple
design and ease of use, water can be dispensed at a flow rate of 1.0 l/min.
• Optional ELGA Biofilter - when fitted to the Option-Q produces water which is free
from biologically active impurities. This makes it suitable for use with applications
which require endotoxin free ultrapure water, bacteria free water, and nuclease
free ultrapure water Ultrapure type I
water direct
• Type III / RO water available from reservoir
from a tapwater
• Front-entry service doors for easy and quick access to consumables supply
• Data collection capabilities through RS232 interface – for compliance with
GLP guidelines
Process Flow
Process PURELAB
Flow Option-Q
PURELAB Option-Q
Water Quality Sensor
Outlet Outlet
Dispense
Pre- Tap
Treatment
Reverse Reverse
Osmosis Osmosis Water Outlet to
Quality Docking Vessel
Sensor or Reservoir
185/254 nm Deionization
UV Lamp Cartridge
Drain
PURELAB
Analytical Research
Feedwater Requirements
Source quality Potable mains water supply Potable mains water supply
Fouling index - maximum 10 10
Conductivity* <2000 μS/cm <2000 μS/cm
Total chlorine 0.5 ppm 0.5 ppm
Heavy metals - maximum 0.05ppm 0.05ppm
Silica - maximum 30 ppm 30 ppm
Temperature 1 - 35°C 1 - 35°C
Flowrate (maximum requirement) 78 l/hr 85 l/hr
Drain requirements (gravity fall 70 l/hr 70 l/hr
with air gap). Maximum during
Service
Feedwater Pressure
Maximum - without internal boost pump 6 bar (90 psi) 6 bar (90 psi)
Minimum - without internal boost pump 4 bar (60 psi) 4 bar (60 psi)
Maximum - with internal boost pump 2 bar (30 psi) 2 bar (30 psi)
Minimum - with internal boost pump Flooded suction Flooded suction
* A restriction on the daily output may be necessary for feedwater >1400 µS/cm
Electrical Requirements
Mains input 100 - 240V ac, 100 - 240V ac,
50-60Hz all models 50 - 60Hz all models
System voltage 24V dc 24V dc
Power consumption with 90VA 90VA
boost pump
Power consumption without 60VA 60VA
boost pump
Fuses 2 x T6.3 Amp 2 x T6.3 Amp
Reservoir level connection Jack Plug 3.5mm Jack Plug 3.5mm
Noise level <45dBA <45dBA
ELGA LabWater
Tel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505
Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained
in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or
omissions. © VWS (UK) Ltd. 2011 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR 38757-04
PURELAB
Analytical Research
PURELAB Classic
The PURELAB Classic system combines high performance with economy to deliver 18.2 MΩ-cm
water at a very cost-effective price. Despite its budget price the PURELAB Classic contains many
market leading features.
Water Quality
Sensor
Ultrafiltration
Membrane
Feedwater
Inlet
(Pre-purified Water) Pump
185/254 nm Deionization
UV Lamp Cartridge Dispenser
Purified Water
Drain Outlet
Page 29 of 46
Page99
PURELAB
Analytical Research
1
With POU filter fitted. 2 <1 CFU/ml without point-of-use filter.
Feedwater Requirements
Parameter Limits
Source - originally from potable Preferably reverse osmosis (RO) or filtered service deionization (SDI) or distilled.
supply, then pre-treated Note: mixed bed or twin bed deionized supplies should be cation limited at exhaustion.
Fouling index (max) 1 for all models. A 0.2 micron membrane prefilter is recommended for all non-RO feeds.
Service deionization (SDI) - MΩ-cm 1 MΩ-cm minimum resistivity at exhaustion.
Reverse osmosis (RO) - µS/cm Recommended < 30 µS/cm
Free chlorine 0.05 ppm max.
TOC Recommended 50 ppb max.
Carbon dioxide 30 ppm max.
Silica 2 ppm max.
Particulates Filtration down to 0.2 micron advisable to protect internal and/or point of use filters.
Temperature 1 - 40°C Recommended 10 - 15°C
Flowrate (maximum requirement) 130 l/hr
Drain requirements (gravity fall Up to 2 l/min
with air gap). Maximum during
service
Feedwater pressure 0.7 bar (10 psi) maximum, 0.07 bar (1 psi) minimum
Electrical Requirements
Mains input 100 - 240V ac, 50 - 60Hz all models
System voltage 24V dc
Power consumption during 60VA
recirculation
Power consumption during 75VA
dispense
Fuses 2 x T6.3 Amp
Reservoir level connection Jack Plug 3.5mm
Noise level during recirculation <40dBA
ELGA LabWater
Tel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505
Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this
document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions.
© VWS (UK) Ltd. 2010 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR 38748-03
Page 30 of 46
Page100
PURELAB
Analytical Research
PURELAB Ultra
The PURELAB Ultra is the intelligent Type I ultrapure water purification system for your most
critical applications. It incorporates many unique and innovative features that set it apart from
other systems and guarantees high water purity of 18.2 MΩ-cm.
Intermediate Post UV
Water Quality Quality
Sensor Sensor
Water
Quality Sensor
Ultrafiltration
Membrane
Feedwater
Inlet
Pump
(Pre-purified Water) Primary 185/254 nm Polishing
Cartridge UV Lamp Cartridge Dispenser
Feedwater
Drain Drain Outlet
Page 31 of 46
Page101
PURELAB
Analytical Research
Feedwater Requirements
Parameter Limits
Source - originally from potable Preferably reverse osmosis (RO) or filtered service deionisation (SDI) or distilled.
supply, then pre-treated Note: mixed bed or twin bed deionised supplies should be cation limited at exhaustion.
Fouling Index (max) 1 for all models. A 0.2 micron membrane prefilter is recommended for all non-RO feeds
Service Deionization (SDI) - MΩ-cm 1 MΩ-cm minimum resistivity at exhaustion
Reverse Osmosis (RO) - µS/cm Recommended <30 µS/cm
Free chlorine 0.05 ppm max
TOC Recommended 50 ppb max
Carbon dioxide 30 ppm max
Silica 2 ppm max
Particulates Filtration down to 0.2 micron advisable to protect internal and/or point of use filters
Temperature 1 - 40°C - Recommended 10 - 15°C
Flowrate (maximum requirement) 130 l/hr
Drain requirements (gravity fall Up to 2 l/min
with air gap). Maximum during
service
Feedwater Pressure 0.7 bar (10 psi) maximum, 0.07 bar (1 psi) minimum
Electrical Requirements
Mains input 100 - 240V ac, 50 - 60Hz all models
System voltage 24V dc
Power consumption during 60VA
recirculation
Power consumption during 75VA
dispense
Fuses 2 x T6.3 Amp
Reservoir level connection Jack Plug 3.5mm
Noise level during recirculation <40dBA
ELGA LabWater
Tel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505
Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this
document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions.
© VWS (UK) Ltd. 2011 – All rights reserved. ELGA®, PURELAB® and PureSure® are registered trademarks of VWS (UK) Ltd.
LITR 38750-04
Page 32 of 46
Page102
Page 33 of 46
Page103
PURELAB flex 2 PURELAB flex 3&4 PURELAB flex 1 PURELAB flex 3&4 PURELAB flex 1
Daily volume >10 Liters <10 Liters >10 Liters <10 Liters >10 Liters
Water quality Type I ultrapure water Type II/Type II+ pure water Type III RO water
Prioritized information
• Hands free with optional foot pedal
shown at all times (system • Locked dispense for glassware filling
status, TOC, alarm)
Optional accessories
• Wall mounting bracket
• Foot pedal
• Thermal printer
• Point of Use filters
• Leak sensor
Page 35 of 46
Page105
Page 36 of 46
Page106
Water Purity
Sensor
Water
Purity and
185/254 nm Purification Temperature Dispense
UV Lamp Pack Sensor Handset
Recirculation
Pump
Water Purity
Reverse Sensor
Osmosis CVF Composite Water
Vent Filter Purity and
185/254 nm Purification
Pack Temperature Dispense
Reservoir UV Lamp Sensor Handset
0.2µm Point
of Use Filter
Water Purity and
or Biofilter
Temperature
(optional)
Sensor
Page 37 of 46
Page107
Feedwater requirement
Source Originally from potable supply, then pre treated. Potable Tap Water Originally from potable
Preferably reverse osmosis (RO) or filtered service supply, then pre treated.
deionization (SDI) or distilled. Preferably reverse
osmosis (RO) or filtered
service deionization
(SDI) or distilled.
Conductivity <1 µS/cm 3 <1 µS/cm 3 <1 µS/cm 3 <2000 µS/cm 4 <30 µS/cm
Contaminant
Hardness N/A N/A N/A <350 ppm as CaCO3 TBC
Free Chlorine <0.05 ppm Cl2 <0.05 ppm Cl2 <0.05 ppm Cl2 <0.5 ppm Cl2 <0.05 ppm Cl2
Chloramine <0.05 ppm Cl2 <0.05 ppm Cl2 <0.05 ppm Cl2 <0.2 ppm Cl2 <0.05 ppm Cl2
Total Chlorine <0.05 ppm Cl2 <0.05 ppm Cl2 <0.05 ppm Cl2 <0.5 ppm Cl2 <0.05 ppm Cl2
Silica <2 ppm <2 ppm <2 ppm <30 ppm SiO2 <2 ppm SiO2
Carbon Dioxide Maximum <0.1 ppm <0.1 ppm <0.1 ppm <30 ppm <30 ppm
(recommended <20 ppm) (recommended <20 ppm)
Fouling index <1 <1 <1 <10 <1
Iron / Manganese N/A N/A N/A <0.05 ppm N/A
Organics (TOC) N/A N/A N/A <2 ppm <50 ppb
recommended
Particulates A 0.2 micron membrane pre filter is recommended for all non-RO feeds to extend N/A
point-of-use filter life
Temperature 4 - 40°C (Recommended 10 - 25°C)
Flowrate (requirement at 15°C) <2 l/min <2 l/min <2 l/min Up to 75 l/hr Up to 60 l/hr
Drain requirements (gravity fall with air gap) None required None required None required >90L/hr >70L/hr
3 4
If feedwater is in the range of 1 μS/cm – 30 μS/cm you will need a pre-conditioning cartridge Purification pack life may vary with feedwaters >1400 μS/cm
Feedwater pressure
Maximum 22psi (1.5bar) 22psi (1.5bar) 22psi (1.5bar) 90psi (6bar) 5 90psi (6bar)
Minimum Flooded suction Flooded suction Flooded suction 30psi (2bar) 5 1psi (0.0 7bar)
5
If <60psi (4 bar) a separate boost pump is recommended
Electrical requirements
Mains Input 100-240V ac, 50-60Hz
System control voltage (not including 24V dc
pumps and UV)
Power consumption during peak demand 120VA 120VA 120VA 100VA 100VA
Noise level <40dBa
ELGA LabWater
Tel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505
Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS
(UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions.
© VWS (UK) Ltd. 2010 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR38975-03
Page 38 of 46
Page108
PURELAB
Analytical Research
PURELAB 3000
The PURELAB 3000 series is ideal for all high usage applications requiring Type III water. The
flexibility of the system means that it can be easily upgraded to provide Type II pure water or
increased capacity should requirements change. Pure water can be used direct from the internal
50 liter reservoir or fed to much larger vessels as required.
• Auto rinse and simple sanitization procedures maintain the purity of water for the life of the system
• A variety of connections and distribution pumps are available to cater for a diverse range of applications
PURELAB 7000
The PURELAB 7000 series is designed for applications requiring higher flows of Type II pure water
or as a pure water feed system for small distribution loops. The system can supply water from the
internal 50 liter reservoir at varying flow rates or can feed into much larger vessels as required.
• Small high purity distribution system with variable flow rates
• Auto rinse and simple sanitization procedures maintain the purity of water for the life of the system
Process Flow PURELAB 7000 Series
Process Flow PURELAB 7000
(Optional)
Water E-Cartridge
Quality Return
Sensor From Loop
CVF Composite
Pump Vent Filter
Reservoir
To
Distribution
Reverse Loop
Osmosis Delivery/ Water
Modules Recirculation Quality
Pump Sensor
Boost
Pump Drain
Ultra
Microfiltration
Category 5 UV Lamp
Feedwater Air Gap
Inlet
Labpure (Optional)
Deionization Labpure
Protek Cartridge Deionization
Pretreatment Cartridge
Cartridge
Page 39 of 46
Page109
PURELAB
Analytical Research
Feedwater Requirements
Source Tap water as detailed below
Contaminant
Conductivity <2000 µS/cm*
Hardness <250 ppm as CaCO3
Free chlorine <4 ppm Cl2
Chloramine <1 ppm as Cl2
Silica <30 ppm SiO2
Fouling Index <10 FI
Iron/manganese <0.1 ppm Fe/Mn
Organics <3 ppm TOC
Temperature 4 – 40°C (Recommended 15 – 25°C)
Flowrate 9 l/min
(maximum requirement @ 15°C)
Drain requirements 20 l/min
(gravity fall with air gap)
Feedwater pressure 6 bar (90 psi) maximum, 2 bar (30 psi) minimum
Failure to comply with the feedwater pretreatment recommendations will affect the life and performance of key components and may invalidate the warranty.
Feedwater requirements are specified for units fitted with a Protek L2 pretreatment cartridge. If feedwater purity is variable or values are outside any of the
above ranges, additional pretreatment is recommended to be installed in the feedwater supply to the unit.
* Deionization cartridge life may vary with feedwater >1400 µS/cm
If in doubt seek advice from Technical Support at ELGA LabWater.
Electrical Requirements
Mains input 230V ac, 50Hz 115V ac, 60Hz
System control voltage (not 24V dc
including pumps and UV)
Power consumption (peak demand) 650VA
Electrical protection rating 10 amps
ELGA LabWater
Tel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505
Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this
document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions.
© VWS (UK) Ltd. 2011 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR 38749-04
Page 40 of 46
Page110
PURELAB
Research & Testing
PURELAB Pulse
The PURELAB Pulse is the cost-effective choice for laboratories that require higher output
volumes especially in hard water areas. Applications range from sample dilution and reagent
preparation, cell culture, media preparation and glassware washing.
UV Lamp
Drain Drain
PURELAB
Research & Testing
1
Standard conditions are 4 bar inlet pressure, 0 bar back pressure, fed with potable water and a clean pre-treatment cartridge.
2
Subject to correct operating and maintenance procedures and use of POU filter. 3 Optimum performance will be achieved with moderate use on moderate feedwaters.
At high usage, (>100 l/day) with high Total Conductivity and CO2 feedwaters (>700μS/cm, 20 ppm CO2) some reduction in product water resistivity may occur.
Feedwater Requirements
Source Potable Mains water supply as detailed below
Conductivity* <2000 µS/cm <2000 µS/cm
Temperature 1 – 35oC 1 – 35oC
Flow rate required (maximum) 80 l/hr 80 l/hr
Drain requirements (minimum) 80 l/hr 80 l/hr
Contaminant
Hardness <350 ppm as CaCO3 <350 ppm as CaCO3
Free Chlorine <0.5 ppm <0.5 ppm
Chloramine <0.2 ppm <0.2 ppm
TOC <3 ppm <3 ppm
Silica <30 ppm <30 ppm
Fouling Index <10 <10
Iron/Manganese <0.2 ppm <0.2 ppm
CO2 – maximum** <30 ppm <30 ppm
Feedwater Pressure
Without internal boost pump 90 psi (6.0 bar) maximum, 60 psi (4.0 bar) minimum
With internal boost pump 30 psi (2.0 bar) maximum, flooded suction minimum
* A restriction on the daily output may be necessary for feedwater >1400 µS/cm. ** Contact ELGA LabWater for feed water >30ppm
Electrical Requirements
Mains input 100 - 240V ac, 50 - 60Hz 100 - 240V ac, 50 - 60Hz
System voltage 24V dc 24V dc
Power consumption with boost pump 110VA 110VA
Power consumption without 85VA 85VA
boost pump
Fuses 2 x T6.3 Amp 2 x T6.3 Amp
Reservoir level connection Jack Plug 3.5mm Jack Plug 3.5mm
Noise level <45dBA <45dBA
ELGA LabWater
Tel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505
Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS
(UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2011 – All rights reserved. ELGA® and PURELAB® are
registered trademarks of VWS (UK) Ltd.
LITR 38754-03
PURELAB ACCESSORIES
Remote Dispensing and Monitoring
Bench-top Dispensers
and Control Stations
Remote dispensing and monitoring enables you to position your PURELAB water purification
system under the bench or on a wall to free up valuable space on the bench-top.
• Dispense Gun clips securely into a wall-mounting bracket for fixed dispensing or can
be unclipped for extended reach (1 meter range from the bracket)
Page 43 of 46
Page113
PURELAB ACCESSORIES
Remote Dispensing and Monitoring
ELGA LabWater
Tel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505
Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this
document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions.
© VWS (UK) Ltd. 2010 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR38918-02
Page 44 of 46