Академический Документы
Профессиональный Документы
Культура Документы
Contents
2 Installation …………………………………………….…….………………… 10
3 Operating …………………………………………….……...………………… 25
4.1 Ammoniac
4.1.1 Principle …………………………………………………………. 61
4.1.2 Test ………………………………………………………………. 62
4.1.3. Recalibration ……………………………………………………. 74
4.2 Chlorophyll A
4.2.1 Principle …………………………………………………………. 76
4.2.2 Test ………………………………………………………………. 76
4.5 Conductivity
4.5.1 Principle …………………………………………………………. 111
4.5.2 Test ………………………………………………………………. 112
4.5.3. Recalibration ……………………………………………………. 120
4.7 Hydrocarbons
4.7.1 Principle …………………………………………………………. 131
4.7.2 Test ………………………………………………………………. 132
4.7.3. Recalibration ……………………………………………………. 142
4.13 pH
4.13.1 Principle …………………………………………………………. 198
4.13.2 Test ………………………………………………………………. 199
4.13.3. Recalibration ……………………………………………………. 207
4.16 Rhodamine
4.16.1 Principle …………………………………………………………. 234
4.16.2 Test ………………………………………………………………. 235
4.16.3. Recalibration ……………………………………………………. 245
4.18 Temperature
4.18.1 Principle …………………………………………………………. 253
4.18.2 Test ………………………………………………………………. 254
4.18.3. Recalibration ……………………………………………………. 262
1 Quick start
Step 1
Unpack the analyser, check that nothing has been damaged during the transportation and fix it
on a wall. The dimensions are given on chapter 2.
Step 2
Connect the sample inlet and outlet with plastic tubing, ID 6.4 mm (1/4”) OD 9.6 mm (3/8”).
Refer to chapter 2 for inlet and outlet location if necessary.
Assemble the fitting elements as below:
With the peristaltic sampling pump, the inlet pressure must be limited to 1 Bar (equiv.
to 1000 hPa or 15 psi)
Step 3
Depending on the options, refer to chapter 2 to connect the 4-20 mA output or the RS232 cable
or the RS485 cable or the Ethernet cable. If a Wi-Fi link has been chosen as an option, check
that the antenna is properly fixed on the right side of the enclosure.
Step 4
Connect the power cable to a grounded power socket.
To avoid electric shock, the analyser must be connected to the earth via the power
socket.
Step 5
Check that the date and time on the bottom right of the screen are correct. If not, refer to
chapter 3 for setting date and time.
Step 6
Refer to chapter 3 to configure the optional interface (4-20 mA or RS232 or RS485 or Ethernet
or Wi-Fi).
Step 7
The default mode is the stopped mode (no automatic measurements, no automatic cleaning).
Set the measuring mode (continuous, periodic or extern) depending of the application (refer to
chapter no 3, timing screen).
The analyser is factory calibrated and is now ready to start the measurements.
For checking and recalibration, refer to chapter 4 for each individual parameter.
2 Installation
2.1 Fixing
Make 4 holes in the panel or in the wall for 8 mm diameter screws according to the pattern
below:
Refer to the drawing and table below for the hydraulic connections.
Flow diagrams are given on the chapter 2.3 for different measuring configurations.
The sample inlet and outlet must be connected with plastic tubing, ID 6.4 mm (1/4”) OD 9.6 mm
(3/8”).
Reagents or cleaning solution must be connected with plastic tubing, ID 3.2 mm (1/8”) OD 6.4
mm (1/4”)
D) PO4 only or PO4 with other optical measurement except NH4 and H2S
All the cable must be connected using the glands as show below.
The module OUT4-20-500 delivers an active 4-20 mA signal for PLC, recorders or SCADA
system. Each OUT4-20-500 module is affected to one measuring channel.
Refer to chapter 3 to set up the 4-20 mA module position and scale values for each channel.
Screw terminal J2
The Wi-Fi module follows the IEEE 802.11b standard and is internally connected to a SMA
connector on the right side of the analyser.
The 2.4 GHz antenna must be fixed on this SMA connector as represented below.
Refer to chapter 3 (communications) to enable the Wi-Fi communication and to set up the
network name (SSID) and the WEP key.
The Wi-Fi module follows the standards listed on the table below.
Standards
Emissions AS/NZS 3548
AS/NZS 3548 CISPR 22
Japan IOH
003NY04115 0000
003GZ04064 0000
FCC Part 15 Subpart C
(FCC ID: MCQ-50M880)
IC RSS 210 (IC:1846A-50M880)
FCC Part 15 Subpart B
ICES-003
EN 55022
EN 61000-4-2
EN 61000-4-3
EN 61000-4-6
EN 301 489-3
EN 300 328
VCCI
Immunity EN 55024
Safety UL 60950-1
CSA 22.2 No. 60950--1
EN 60950
The Ethernet module provide a RJ-45 jack that meets the ISO 8877 requirements for
10/100BASE-T. See the following figure and table for pin orientation and pin assignments.
Refer to chapter 3 (communications) to select the Ethernet port and to set up the IP address
and mask.
The RS232 port has a Sub-D 9 pins male connector and is located on the screen board
DSP500, position J15.
Refer to chapter 3 (communications) to select the communication mode and the baud rate.
The format is always 8 bit, no parity, 1 bit stop.
To connect this port to a laptop (directly or through or a USB adaptor), use a null cable (pin 2 to
pin 3, pin 3 to pin 2, pin 5 to pin 5).
The RS485 port has a 6-pin screw terminal located on the screen board DSP500, position J16
as shown below:
RS485
Refer to chapter 3 (communications) to select the communication mode and the baud rate.
The format is always 8 bit, no parity, 1 bit stop.
The module PH500 accepts the direct connection of a pH electrode on the screw terminal J2.
For automatic temperature compensation, the Pt100* sensor of the electrode must be
connected on J3.
The module COND500 accepts the direct connection of a conductivity electrode on the screw
terminal J2. Conductivity constant of K=0.01, K=0.1, K=1 and K=10 are accepted, refer to
chapter 4 to setup the right constant.
For automatic temperature compensation, the Pt100* sensor of the electrode must be
connected on J3. Refer to chapter 4 to adjust the temperature coefficient.
The module IN4-20-500 accepts the connection of a 4-20 mA output from any probe or external
measuring system. The floating input is insulated by optocouplers.
A 15V DC output (insulated, max 20 mA) is available on the connector J3. This output must be
used for any probe with electrical contact to the water sample (example: chlorine probe).
A 12V DC output (non insulated, max 200 mA) is available on the connector J4. Do not use this
output to supply probe with electrical contact to the water sample.
Refer to chapter 4 to set proper label, unit and scale factor of the measuring channel attached to
this module.
The module RELAY500 gives on J2 a normally closed (NC) and a normally open (NO) relay
contacts for different functions. Refer to chapter 4 to select the right function, threshold, etc....
A 24V DC output is available on the screw terminal J3 (total max current 1A). This output can be
used for example to power external pumps or valve for a multiplexing system.
The module LOGIC500 accepts two logical signals from a PLC to control the measurement
cycles.
The START input is used to start a measurement cycle for all the channels when a pulse (6 to
48V DC) is received. The minimum pulse width is 200 mS. The external measuring mode must
be selected.
The INHIBIT input is used to stop and inhibit any further automatic measurement (periodic mode
or continuous mode) as long as this input is maintained high (6 to 48 VDC).
Note: as these two input signals refer to different measuring mode, it has no sense to use both
of them on a same application.
Before connecting the power cord, check that the voltage written on the serial
number plate corresponds to the mains voltage.
To avoid electric shock, the analyser must be connected to the earth via the power
socket. Check that the power socked is earthed.
Disconnect the power cord or mains before any wiring, connection or maintenance
operation inside the analyser.
If all the conditions above are satisfied, plug the power cord into a grounded power socket.
If the power cord needs to be changed or removed, the mains must be connected to the screen
board DSP500 on the screw terminal J1 as follows:
Neutral
Earth
Line
2.4.12 USB
A USB key (memory stick) can be connected on the USB connector type A represented on the
drawing of the section 2.4 for measurements download, configuration download or upload, or for
upgrading the analyser software. Refer to the Operating section for using the USB key.
Disconnect the USB key as soon as the operation if finished as USB connection is
not watertight.
3 Operating
3.1 Overview
When powered on, the analyser displays the process screen after a fast hardware check.
Then the measurements start automatically and the process values and graphs are updated as
soon as necessary (except in stopped mode that is the default mode after delivery, see on the
timing screen to change the mode).
The timing, communication, regional setting and factory screens are accessible only
in administrator mode.
Process screen
Alarm screen
Timing screen
Communication screen
The process screen is the basic screen of the analyser. It is display a few seconds after power
on and remains displayed all the time unless another screen has been selected. The default
password for login is 123 in user mode and 456 in administrator mode.
This screen displays the very last measurement for all the active channels on the left side and
the graph of the recorded measurements on the right side.
If the screen saver mode has been selected, the backlight of the screen will be cut automatically
when no key has been activated for 15 minutes or during the night time, depending of the
screen saver mode. All the functions (measurement, cleaning, temperature control...) remain
active during the screen saving time. To reactivate the screen, press anywhere on the screen.
4 5 6 7
8 9 10 11 12
The label is automatically set when the channel is created. It can be modified in the check
screen.
The unit is automatically set when the channel is created. It can be modified in the check
screen.
3 Temperature
Display the auxiliary temperature measurement of the channel. It concerns only the channels
with automatic temperature compensation (NH4, H2S, PO4, PAH, Chl. A, pH and conductivity).
4 Channel value
This value is the very last measurement done for each channel. The number of digit can be
modified in the check screen. By default, negative values resulting from normal measuring
fluctuations or from a wrong zero are displayed. They can be eliminated and replace by zero by
changing a setting on the check screen.
5 Adjustment - checking
This key gives access to the check screen for each channel checking or calibration. Then the
value display area must be pressed to select the right channel.
Alternatively as a shortcut, the value display area can be pressed directly.
This key gives access to a full screen zoom of the historical graph. The channel must be
selected by pressing or on the value area or on the graph, depending on the total number of
channels.
This graph shows the last measurements over the selected time scale. The red dot line
corresponds to the high alarm threshold defined in the alarm screen, while the green dot line
correspond to the low alarm threshold.
The Y scale of the graph is automatic, but cannot be lower than the high threshold. If the high
threshold is not used as an alarm, it can be reduced in order to increase the Y scale for a better
resolution on low measurements.
To view a zoom of the graph, press anywhere on the graph area. Refer to the zoom screen for
more details.
This field displays the current date and time. To change the date or the date format, refer to the
regional setting screen.
Depending on the measuring mode, this field displays information about the measurements
schedule.
If an inhibit signal is send to the LOGIC500 logical input module (for any mode except the
stopped mode), this field displays “Held” otherwise display will be as follows:
a) For continuous mode, this field displays “Measuring...” followed by the label of the
channel being currently acquired
- During settling down time (if any), this field displays “Settling down...”
- If a measuring cycle command has been received from the LOGIC500 logical input
module or through MODBUS communication, a measuring cycle starts and the
display is one of the following:
- During settling down time (if any), this field displays “Settling down...”
Depending on the cleaning method, this field displays information about the cleaning schedule.
- Between cleaning cycle, the time remaining for the next cleaning cycle is displayed,
example: “Next cleaning: 01:03:24” means that the next cleaning cycle will occur in 1
hours 3 min and 24 sec.
- During purging by acid, the time remaining is displayed, example: “Purging... 00:00:15”
means that purging will continue during 15 sec.
- During the waiting time, the time remaining is displayed, example: “Waiting... 00:00:05”
means that waiting time will continue during 5 sec.
- During the recovery time, the time remaining is displayed, example: “Recovery...
00:00:20” means that recovery will continue during 20 sec.
c) Ultrasonic method
- Between cleaning cycle, the time remaining for the next cleaning cycle is displayed,
example: “Next cleaning: 01:03:24” means that the next cleaning cycle will occur in 1
hours 3 min and 24 sec.
- During cleaning by acid, the time remaining is displayed, example: “Cleaning... 00:00:15”
means that cleaning will continue during 15 sec.
- During the recovery time, the time remaining is displayed, example: “Recovery...
00:00:20” means that recovery will continue during 20 sec.
An open padlock means that a user is currently logged, should be in user mode or super user
mode. To logout, press on this icon. The analyser automatically logout after 15 minutes without
action on the touch screen.
A close padlock means that the analyser is currently logout. Press anywhere in the touch screen
to start a login procedure.
3.4 Alarm screen
The alarm screen shows for each channel the high and low alarm thresholds including eventual
boundaries as well as the “fault” relay number (if any) and “ready” relay number (if any).
12 Mute mode
This field select or deselect the mute mode. In mute mode, the loudspeaker is bared.
The alarm screen enables to set the high and low alarm threshold, as well as the boundaries.
The relay modules (if any) are displayed as a reminder for the external wiring.
9 10
5 6 7 8
Each measuring channel is listed in the alarm screen. The label can be modified in the check
screen if necessary.
2 Relay position
The relay position refers to one of the 12 module’s socket positions on the DSP500 main board.
Refer to installation section for socket drawing on the DSP500 board.
Note: this button remains empty if no relay module has been installed (option).
The high alarm level can be changed by pressing on this field. A default value is preset when
the channel is created. The high alarm level is primary used to fix the high limit represented by a
red dot line on the recorded measurement graph. The limit is used as the minimum graph scale.
The secondary use is to drive a high alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement overpass the high alarm value and will stay activated
until the measurement goes down below this value after subtracting the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
This button shows the current position of the optional alarm relay. It can be used to activate or
deactivate the concerned relay for test purpose only. The relay recovers its expected state after
5 seconds or when the alarm screen is left.
Note: this button exists only if a relay has been installed (option).
5 Relay position
The relay position refers to one of the 12 module’s socket positions on the DSP500 main board.
Refer to installation section for socket drawing on the DSP500 board.
Note: this button remains empty if no relay has been installed (option).
The low alarm level can be changed by pressing on this field. A default value is preset when the
channel is created. The low alarm level is primary used to fix the low limit represented by a
green dot line on the recorded measurement graph.
The secondary use is to drive a low alarm relay if the relay button has been activated and if a
relay module has been installed on the selected module socket. In this case, the relay will be
activated as soon as the measurement underpass the low alarm value and will stay activated
until the measurement goes down below this value after adding the boundary value.
Refer to installation section for the relay wiring. Both functions NO (Normally Open) and NC
(Normally Closed) are available.
This button shows the current position of the optional alarm relay. It can be used to activate or
deactivate the concerned relay for test purpose only. The relay recovers its expected state after
5 seconds or when the alarm screen is left.
Note: this button exists only if a relay has been installed (option).
8 Boundaries
The purpose of boundary is avoiding too frequent changes of state of the alarm relays what can
harm process devices like pumps or electric-valves. The boundary value is subtracted from the
high level value before leaving the high alarm state and reciprocally added to the low alarm level
before leaving the low alarm state as shown on the example below for the high level only:
The relay position refers to one of the 12 module’s socket positions on the DSP500 main board.
Refer to installation section for socket drawing on the DSP500 board.
The fault relay is activated as soon as a measurement error occurs on any of the channel and
stays activated as long as an error is still pending. The errors are automatically clear on the next
measuring cycle.
Note: this button exists only if a relay has been installed (option).
This button shows the current position of the optional alarm relay. It can be used to activate or
deactivate the concerned relay for test purpose only. The relay recovers its expected state after
5 seconds or when the alarm screen is left.
Note: this button exists only if a relay has been installed (option).
This screen allows selecting the proper measuring and cleaning mode and then to set or
change all the timing values, including for the multiplexing function if used.
6 5 14 15 16 17 10
1
11
2
12
3
13
7
18
9
19
4
16 20
17 21
23
22
24 25 26 27 28 29 23 24
1 Stopped mode
In this mode, all the automatic measurements or automatic cleaning cycles are stopped.
This mode can be use during maintenance operations or for doing manual measurements only.
In this mode, the measurements are continuously updated at the highest possible rate. The
display and the 4-20 mA output (if any) are updated every second.
The measurements are stored in the internal memory as an average during the period defined
in the “record period” field.
If some of the measuring channels are not compatible with the continuous mode (as NH4 or
PO4 for example) the continuous mode will not be available for selection or will be automatically
updated in periodic mode.
The measurement can be held at anytime if a high level voltage has been applied on the inhibit
input of the LOGIC500 module (if any).
In this mode, the measurements are done at a specific period of time define in the “period” field.
The display and the 4-20 mA output (if any) are updated at the end of the measurement cycle.
The measurements are stored in the internal memory at each cycle.
If the measuring cycle time is higher than the defined period, a new measurement will start
immediately after the previous one.
The measurement can be held at anytime if an high level voltage has been applied on the inhibit
input of the LOGIC500 module (if any).
In this mode, the measurements are started or by a pulse (>0.3 sec) send to the start input of
the LOGIC500 module, or by sending a value of 1 in the start measurement address by a
MODBUS communication.
In continuous mode only, the sampling pump is running every 15 seconds during a time defined
in this field. To preserve the tubing life time, put a value as low as possible like 2 or 3 seconds
to just renew the sample.
6 Record period
In continuous mode only, the recorded value is the average of the measurements during the
period defined in this filed.
In periodic mode only, the sampling pump is running before every measuring cycle during a time
defined in this field.
8 Measuring period
In periodic mode only, the measuring cycle occurs with a period of time defined in this filed.
In periodic mode only, it’s possible to add a settling down time defined in this filed after the
sampling and before the measurement to allow suspended solid to go down.
This mode concerns only a few specific applications, in all other cases, let the default value of 0.
In periodic mode only, it’s possible to synchronize a second external sampling pump through a
relay if this function is activated.
11 Off time
In periodic mode only and if the external sampling function is activated, this field set the off time
for the external pump starting at the beginning of internal sampling.
12 On time
In periodic mode only and if the external sampling function is activated, this field set the on time
for the external pump starting at the end of the off time.
13 Rest time
In periodic mode only and if the external sampling function is activated, this field set the rest
time for the external pump starting at the end of the on time.
The rest time allows suspended solids in the sample to settling down before the internal
sampling.
If activated, the acid cleaning will start automatically according to the defined period.
Acid cleaning is mandatory for NH4 and H2S measurements regarding the risk of deposit inside
the outlet tubing.
15 Auto zero
If acid cleaning is available, an auto-zero can be performed automatically on the acid cleaning
solution if selected.
If the auto-zero is selected, the cleaning solution must be prepared with pure water
and analysis grade acid to get a correct zeroing.
This field contains the period of time between two cleaning cycles.
The recommended period is 24 hours.
A too short cleaning time may cause a bad zeroing with the auto-zeroing function.
Keep the default value of 60 seconds for a good zeroing.
This field contains the waiting time between the end of the cleaning solution pumping and the
zeroing to enable any bubbles produced by the cleaning pump to go up.
The default value is 5 seconds. It is recommended to keep this default value unless authorized
by the manufacturer.
This field contains the recovery time before starting again the measurements on the sample
after a cleaning operation. The default value is 30 seconds. It is recommended to keep this
default value unless authorized by the manufacturer.
A too short recovery time may cause erratic measurements during the few seconds
that follow a cleaning cycle.
Keep the default value of 30 seconds for a good sample recovery.
This button shows the ultrasonic module position. It can be set or changed only by the
manufacturer.
This field contains the period of time between two ultrasonic cleaning cycles.
The recommended period is 2 hours. It is recommended to keep this default value unless
authorized by the manufacturer.
This button shows the relay module position that drives an electric-valve to collect the sample
after analysis when reagents are use (example PO4 measurement). It can be set or changed
only by the manufacturer.
This button shows the current position of the optional waste collect relay. It can be used to
activate or deactivate this relay for test purpose only. The relay recovers its expected state after
5 seconds or when the timing screen is left.
Note: this button exists only if a relay has been installed (option).
Theses buttons show the relay module positions that drives external electric-valve (not
provided) to select a specific sample stream over several streams. They can be set or changed
only by the manufacturer.
These buttons show the current position of the multiplexing relays. They can be used to activate
or deactivate theses relays for test purpose only. Each relay recovers its expected state after 5
seconds or when the timing screen is left.
Note: this button exists only if a relay has been installed (option).
These fields contains the sampling time of each stream for multiplexing purpose.
The default value is 15 seconds. It must be adjusted depending on the flow of each stream to
get a good renew of the sample.
A too short sampling time may cause erratic measurements as the measured stream
may contain a part of the previous stream.
This button shows the logical input module position that enable to inhibit a specific stream to be
measured (for example if no sample is available). It can be set or changed only by the
manufacturer.
These buttons show the current states of the inhibition inputs of the logical input modules for
testing purpose.
Note: this button exists only if an inhibition module (LOGIC500) has been installed (option).
This screen allows setting or changing the communication settings for the two internal serial
ports of the microcontroller, port A and port B. By default, no communication is activated.
25 3 6 7 8 9 10
11
12
13
14
20
15
16 17 18 19 22 23 24 25 21
1 Remote mode
When selected, the remote communication mode allows driving the analyser at a distance by a
PC laptop. A special software Netscan500 must be installed before on the PC.
The connection thought the RS232 allows a communication at a short distance only (<10
meters) for demonstration purpose with a beamer for example.
For longer distance, a phone modem or GSM modem must be connected on the RS232 port.
2 MODBUS mode
When selected, the MODBUS communication mode allows reading the last measurements or
the error codes as well as starting a new measuring or cleaning cycle. The MODBUS protocol is
a standard and simple protocol implemented on most of the SCADA systems. If necessary, the
full protocol can be downloaded at www.modbus.org .
Basic frames using the function no 3 as shown below allows to read the measurements:
Slave
number
3 Address of first byte NB of bytes CRC16
Slave ……
3 NB of read bytes 1st byte 2nd byte CRC16
number
The measurement values follows the floating point standard IEEE754, format single 32 bit (4
bytes), higher first, lower last. Invalid codes are 1 byte long.
Each device on a MODBUS network has a specific number. This number can be entered in this
field. The default value is 1.
Select the modem mode if and only if a phone or GSM modem is connected on the RS232 port.
The baud rate must be fixed strictly in accordance with the system connected on the RS232 port
and generally at the higher speed compatible with the media capacity.
Up to 3 Hayes initialization sequences can be entered in these fields. Theses sequences are
send to the modem each time that the analyser is power on.
After entering theses sequence, put off and on again the analyser to send them to the modem.
The 10 last received characters are displayed in this window for test purpose only. Non ASCII
characters are represented by a filled black square.
The 10 last transmitted characters are displayed in this window for test purpose only. Non ASCII
characters are represented by a filled black square.
9 Transmit test
Pressing on this key transmit “ABC” characters on the port for test purpose only.
10 MODBUS test
Pressing on this key displays the MODBUS test screen as below if MODBUS mode is enable.
It displays the diagnostic MODBUS counters updated continuously at each received frame as
well as the time and date of the last received frame.
Counter 1 displays the total number of valid frames received, whatever they are intended or not
for the analyser.
Counter 2 displays the number of frames received with a CRC error, whatever they are
intended or not for the analyser.
Counter 3 displays the number exception answers preceded. Exception answer is produced in
case of unrecognized command or wrong parameter on a received frame.
Counter 4 displays the number of valid frames received specifically intended for the analyser.
Counter 5 displays the number of valid broadcast frames received. Broadcast frame are not
usual.
Counter 8 displays the number of characters received with a format error, whatever they are
intended or not for the analyser.
Counter 9 displays the number of frames proceeds for which an answer has been sent to the
host.
The Modbus addresses of the channels and other functions can be visualized by pressing on
the “addresses” key as shown below:
11 Remote mode
When selected, the remote communication mode allows driving the analyser at a distance by a
PC laptop. A special software Netscan500 must be installed before on the PC.
The maximal cable length on the RS485 port is 1200 m (4000 ft).
12 MODBUS mode
Refer to the screen number 2 for the MODBUS protocol details and addresses.
Each device on a MODBUS network has a specific number. This number can be entered in this
field. The default value is 1.
The port B can be use through an RS485 interface located on the DSP500 board. Refer to the
installation section for wiring.
If the RS485 interface is selected, Wi-Fi interface or Ethernet interface are desactivated, even if
the optional module(s) has been installed on the DSP500 board.
The port B can be use through an optional Wi-Fi interface module located on the DSP500
board. The module output is connected to an external antenna on the side of the analyser.
If the Wi-Fi interface is selected, RS485 interface or Ethernet interface are deactivated, even if
the optional module has been installed on the DSP500 board.
The port B can be use through an optional Ethernet interface module located on the DSP500
board.
If the Ethernet interface is selected, RS485 interface or Wi-Fi interface are deactivated, even if
the optional module has been installed on the DSP500 board.
The baud rate must be fixed strictly in accordance with the system connected on the RS485 port
and generally at the higher speed compatible with the media capacity. This baud rate does not
apply for the Wi-Fi or for the Ethernet interface.
18 Wi-Fi SSID
The Wi-Fi network name (SSID) must be entered in this field before using the Wi-Fi
communication. All devices linked by the same Wi-Fi network must have the same SSID.
The Wi-Fi networks are generally encrypted using a WEP key to protect data privacy. The WEP
key of the network must be entered in this field before using the Wi-Fi communication.
Both 64 bits keys (10 characters) and 128 bits keys (26 characters) are accepted.
If no WEP key are used (unsecured network), leave this field blank.
The Wi-Fi module is reinitialised each time new parameters are entered, so the new parameters
are taken in account within a few seconds.
20 Ethernet IP address
Use this series of fields to enter the Ethernet IP address in accordance with the network
parameters. The Ethernet module is reinitialised each time new parameters are entered, so the
new parameters are taken in account within a few seconds.
21 Ethernet IP mask
Use this series of fields to enter the address mask in accordance with the network parameters.
The Ethernet module is reinitialised each time new parameters are entered, so the new
parameters are taken in account within a few seconds.
The 10 last received characters are displayed in this window for test purpose only. Non ASCII
characters are represented by a filled black square.
The 10 last transmitted characters are displayed in this window for test purpose only. Non ASCII
characters are represented by a filled black square.
Pressing on this key transmit “ABC” characters on the port for test purpose only.
25 MODBUS test
Pressing on this key displays the MODBUS test screen. Refer to the arrow number 10 above.
This screen allows setting or changing the regional settings like the language, date, decimal
separator, date format as well as miscellaneous parameters as screen saver and display
appearance.
1 2 3
4 6 7 5
1 Language selection
Choose the appropriate language using this series of button. The change is immediate.
2 Clock update
Update the date and time using theses series of fields. The change is immediate and all
engaged measuring cycle or cleaning cycle will be aborted and automatically rescheduled
according to the new date and time.
3 Screen saver
- No: there no screen saver. The screen stay back lighted all the time.
- 15 min: the screen back light turns off after 15 minutes of touch screen inactivity. All the
functions like measurements, cleaning, and communication remain active. Press anywhere in
the screen to reactivate the screen back light.
- Night: the screen back light turns off after 7 pm until 7 am. All the functions like
measurements, cleaning, and communication remain active. Press anywhere in the screen to
reactivate the screen back light.
4 User password
This field allows changing the user level password. The default password is 123.
The user level allows the access only to measurements display and basic parameters as alarm
levels.
5 Administrator password
This field allows changing the administrator level password. The default password is 456.
The administrator level allows the access to all the parameters.
6 Decimal separator
This button allows choosing a point or a comma as decimal separator for the recorded
measurement field.
7 Date format
This button allows choosing between the usual date format (day/month/year) or the American
date format (month/day/year).
It allows backup or restoring the complete analyser configuration internally in a flash memory or
externally on a USB key, to set or delete measuring channels, to change the thermal control
parameters, to test the 24V DC output and other hardware.
2 3 4 5 6 7 12 13
9
8
10 11
14
15
16
19
17
18
20 21
1 Delete all
This button reset completely the analyser configuration, including channels, calibration, alarms,
communications, timings and regional settings.
2 Backup on EEPROM
This button allows backup the complete analyser configuration, including channels, calibration,
alarms, communications, timings and regional settings on the internal flash memory (EEPROM).
This button allows backup the complete analyser configuration, including channels, calibration,
alarms, communications, timings and regional settings on a USB key.
After pressing this button, the analyser waits for the connection of a USB key. Press on any key
to cancel this function.
This button allows backup the complete analyser configuration, including channels, calibration,
alarms, communications, timings and regional settings from the internal flash memory
(EEPROM).
This button allows restoring the complete analyser configuration, including channels, calibration,
alarms, communications, timings and regional settings from a USB key.
After pressing this button, the analyser waits for the connection of a USB key. Press on any key
to cancel this function.
This button clears all the stored measurements inside the internal memory.
This area displays the software version and the DSP500 board number.
9 Channels method
Theses buttons display the method used for the measuring channels in case of different method
available, for example “SPECTRO” for spectrograph, “LED” for LED source.
10 Channel adding
Theses buttons allows adding a measuring channel if the corresponding hardware has been
installed before.
11 Module function
Theses buttons display the function and the position allocated by the manufacturer to each of
the 12 modules that can be plugged on the DSP500 board. For the 4-20 mA outputs, the
corresponding channel is display on the second line.
Theses buttons display the position allocated by the manufacturer to the photodiodes
connectors for each channel using a photodiode.
Theses buttons stay blank if no photodiodes are used.
13 LED or PM position
Theses buttons display the position allocated by the manufacturer to each channel using a
photomultiplier board (PM) or a LED board.
These buttons stay blank if no PM or LED are used.
Theses buttons display the position allocated by the manufacturer for the xenon lamps.
The top position is always allocated to the gas phase measurements (NH4, H2S) while the
second is for absorbance measurements on the water sample (NO3, UV245, Colour, PO4) and
the third one for fluorescence (PAH/Oil, Chlorophyll).
This field display the temperature measured on each of the 3 temperature channels.
This field allows setting or changing the temperature set point for each of the 3 temperature
control loop. They can be used or not depending of the configuration.
This field shows the present power output value in percent for each of the 3 temperature control
loop. 100% means the output is on all the time, 0% means the output is off all the time. For
intermediate values, the output is on during the percent value of a periodic cycle of about 2
seconds.
This button turn on or off the power output for each of the 3 temperature control loop for testing
purpose only.
Note: the output automatically resume to the normal state after 5 seconds.
This button displays the popup below where the PID parameters can be changed for each of the
3 temperature control loop.
Theses button display the current state of each of the 12 outputs. Theses outputs are driving
internal pumps and valves depending on the configuration as indicated above each button.
Pressing on one of these keys put ON or OFF the corresponding output for testing purpose
only.
21 ADC values
This area displays the 16 channels of the internal ADC (Analogue to Digital Converter) of the
DSP. These values are for the manufacturer only.
4.1 .1 Principle
The measuring principle is based on the analysis of the UV absorption spectrum of the
ammonia gas (NH3) in equilibrium with the dissolved ammonia gas in the water sample.
A small quantity of sodium hydroxide (NaOH) is added to the sample to increasing the pH above
11 where most of the NH4+ is transformed in NH3 (>99.9%).
The UV light is generated by a pulsed xenon lamp to reach an ultra long life time.
The absorption spectrum is performed by a concave grating coupled to a 2048 pixels CCD
(Charge-Coupled Device).
A Fast Fourier Transform (FFT) is applied on the absorption spectrum to extract the periodic
peaks specific to the ammonia gas.
Automatic temperature compensation is done using a Pt100 temperature probe on the sample.
4.1 .2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. It corresponds to the last measurement done
(automatic or manual) concerning the concentration or error code, sample temperature, lamp
spectrum, absorbance spectrum and FFT spectrum.
7 1 6 8 9 10 11
4 33
3
34
21
12
32
22
13 23
24
14 25
26
15 27
28
16 29
17 20 19 18 31 30
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit, as well as the sample temperature read
during the measurement.
During a measuring cycle, the last measured value is display in light green until replaced by the
new value in black.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or the cleaning pump or the ultrasonic cleaning (if any).
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the NaOH pump for testing purpose or to fill the tubing after
installation. The pump will stop automatically after 10 seconds.
It can also be stopped at any time by pressing on the stop button.
During a measuring cycle, this area displays the different steps of the cycle with an associated
countdown timer or counter.
During the ammonia measurements, the steps are:
4) Reagent: the NaOH is added to the sample inside the stripping pot
5) Stripping: the gas flow cell is placed in a close loop with the stripping pot
6) Measuring: the spectrum is acquired on the gas coming through the sample, and then the
concentration is computed and displayed
7) Purging: the gas flow cell is purged again by external ambient air to remove humidity
8 Scale button
9 Signal graph
This graph displays the lamp spectrum get during the last measuring cycle. It is updated during
the measuring cycle when the lamp is flashing.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The signal is normally saturated for wavelength above 215 nm as ammonia measurement is
performed at lower wavelength.
It’s recommended to have a signal at 210 nm above 1000 to get a good repeatability, but
measurement can still be done for signal as low as 100 if the repeatability is not critical as on
waste water for values above 1 mg/l.
If the light signal is too high, an alarm message is displayed on the top left corner.
The values can be read by moving the cursor with the vertical arrows.
10 Absorbance spectrum
This graph displays the absorbance spectrum get during the last measuring cycle. It is displayed
in light grey during a measuring cycle until updated.
For more details or to investigate about the absorbance spectrum, press anywhere on the graph
to zoom it as below:
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows located above the Y axis.
11 FFT spectrum
This graph displays the FFT spectrum get during the last measuring cycle. It is displayed in light
grey during a measuring cycle until updated.
For more details or to investigate about the FFT spectrum, press anywhere on the graph to
zoom it as below:
The peak value is the average FFT value inside the NH3 band (red dot lines).
The ref value is the average FFT value outside the NH3 band.
They are represented by the green lines.
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows located above the Y axis.
12 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
13 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
14 Display mode
This button allows selecting the concentration display mode as NH4 (default mode) or as N of
NH4.
If N-NH4 is selected, the NH4 concentrations are multiplied by 14/18=0.7777 before displaying
and the label is automatically updated to “N-NH4”.
The measurements previously stored in the internal memory will keep their values in NH4.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
16 Number of decimal
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
17 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
18 Temperature adjustment
The temperature of the sample given by the internal temperature probe can be adjusted by this
field. This adjustment is not important as any difference of temperature is taken in account
during the calibration. But if this value is changed, a new calibration must be done.
19 Temperature coefficient
The sample temperature affects directly the NH3 liquid/gas ratio. This coefficient is used for the
temperature compensation. It can be change only by the manufacturer.
20 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
21 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
Note: this factor must be set to 1.0 before doing a new linearization curve
22 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
23 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
24 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
25 Reaction time
26 Reagent time
27 Wavelength mini
This field allows changing the minimal wavelength taken in account for the measurement.
28 Wavelength max
This field allows changing the maximal wavelength taken in account for the measurement.
This field allows changing the minimal frequency off the FFT taken in account for the
measurement.
This field allows changing the maximal frequency off the FFT taken in account for the
measurement.
31 Background factor
This field allows changing the multiplying background factor of the FFT off-band signal.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Beer-Lambert law used on most of the measurements is not linear for high concentrations
due to a light mask effect. Therefore it’s necessary to introduce a correction before displaying
the final concentration. This correction is based on the linearity curve represented on the top.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0,0
as starting values for S1 and M1. The final value is linearly extrapolated between these points.
34 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.1 .3 Recalibration
- Prepare a 10 mg/l NH4 standard. For this weight 2.944 g of dried ammonium chloride (ClNH4)
and dilute in 1litre of distilled water to obtain a mere solution of 1g/l NH4. Then take 10 ml of this
mere solution and complete to 1 litre of distilled water.
- Do at least 3 manual measurements on this 10 mg/l standard using the check screen. The last
two measurements must be close together with less than 5% of difference.
- Press on the “scale” button of the check screen (as describe above) and then enter the
standard value on the keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
4.2 Chlorophyll A
4.2 .1 Principle
The measuring principle is based on the red light fluorescence that is a typical characteristic of
the chlorophyll A
The light is generated by a pulsed xenon lamp to reach an ultra long life time.
The excitation filter limits the bandwidth of the excitation light, while the emission filter selects
the specific fluorescent light.
A second detector (optional) is used to compensate the eventual absorbance of the sample due
to turbidity.
4.2 .2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The concentration or error code corresponds to the
last measurement done, while signal values and graph are updated continuously.
1 7 8 9 10 11 12
6
13 21 30
14 22
23
15
16
17
18
19 20 24 25 26 27 28 29
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit, as well as the sample temperature and
standard deviation (sigma) obtained on the elemental acquisitions.
During a measuring cycle, the last measured value is display in light green until replaced by the
new value in black.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
7 Scale button
To avoid a wrong calibration, check the zero or redo a zero before doing a
calibration.
8 Zero button
Put first pure water inside the flow cell before pressing on this button.
Note: an auto-zero is performed automatically during each acid cleaning if this option is selected
on the timing screen.
9 PM signal
This window displays the PM signal continuously. This signal corresponds to the fluorescent
light re-emitted by the chlorophyll A.
On pure water, this signal is normally close to zero with fluctuations of a few tens of mV.
When the light signal added to the darkness signal reach 3000 mV, a message “OVL!” in red is
displayed to indicate a saturation of the acquisition system.
It might be caused by an over-range concentration of the sample or by ambient light if the door
of the analyser is opened with a high level of ambient light (sun, daylight).
This window displays the photo detector signal continuously (optional). This signal corresponds
to the excitation light of the sample. It is use to compensate an eventual absorption of the
excitation blue light by the sample, for example due to turbidity.
When the light signal added to the darkness signal reach 3000 mV, a message “OVL!” in red is
displayed to indicate a saturation of the acquisition system.
It might be caused by ambient light if the door of the analyser is opened with a high level of
ambient light (sun, daylight).
11 PM Signal graph
This graph displays the PM signal continuously updated. This signal corresponds to the
fluorescent light re-emitted by the chlorophyll A.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The average signal value in mV and its standard deviation (sigma) on the last 500 acquisitions
are displayed on the bottom right of the screen.
The values can be read by moving the cursor with the vertical arrows. After reading, restart the
acquisitions by pressing on the “play” button if necessary.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
This graph displays the photo detector signal continuously updated. This signal corresponds to
the excitation light of the sample.
For more details or to investigate about the signal, press anywhere on the graph to zoom it.
Refer to PM signal graph for the zoom display.
13 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
14 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
16 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
17 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
18 Temperature adjustment
The temperature of the sample given by the internal temperature probe can be adjusted by this
field. This adjustment is not important as any difference of temperature is taken in account
during the calibration. But if this value is changed, a new calibration must be done.
19 Temperature coefficient
The fluorescence is affected by the temperature. This coefficient is used for the temperature
compensation.
20 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
21 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
Note: this factor must be set to 1.0 before doing a new linearization curve
22 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
23 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
24 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
25 Turbidity factor
This field enable a compensation of the excitation light loses due to turbidity.
A value of 1.0 corresponds to a basic compensation.
26 Offset
This field show the internal offset applied on the signal for the zeroing.
It is updated at each zeroing and has not to be changed.
It can be used for troubleshooting by the manufacturer.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last zeros done with the date, time and zero value adjustment,
whatever manually or by an automatic cleaning cycle.
Note: a strong zero adjustment value might reveal an abnormal zeroing operation.
This area concerns the linearity compensation applied on the rough measurements.
The Beer-Lambert law used on most of the measurements is not linear for high concentrations
due to a light mask effect. Therefore it’s necessary to introduce a correction before displaying
the final concentration. This correction is based on the linearity curve represented on the top.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
30 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.2 .3 Recalibration
- Press on the “scale” button of the check screen and then enter the standard value
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
4.3 .1 Principle
The measuring principle is based on the light absorption on the blue area according to the Beer-
Lambert law:
[C ] = k ⋅ log I 0
I1
A second light beam is used to compensate the turbidity of the sample or an eventual deposit
inside the flow cell in the yellow area.
The light is generated alternatively by LEDs.
4.3 .2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The concentration or error code corresponds to the
last measurement done, while signal values and graph are updated continuously.
1 7 8 9 10 11 12
6
13 19 27
14 20
21
15
16
17
18
22 23 24 25 26
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit, as well as the standard deviation (sigma)
obtained on the elemental acquisitions.
During a measuring cycle, the last measured value is display in light green until replaced by the
new value in black.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
7 Scale button
To avoid a wrong calibration, check the zero or redo a zero before doing a
calibration.
8 Zero button
Put first pure water inside the flow cell before pressing on this button.
Note: an auto-zero is performed automatically during each acid cleaning if this option is selected
on the timing screen.
9 LED 1 signal
This window displays the LED1 signal continuously. This signal corresponds to the light that can
be absorbed by the yellow colour of the sample.
On pure water, the signal must normally be in a 1000 to 2000 mV range (value below 500 mV
might deteriorate significantly the repeatability while values above 2500 mV might cause an
error for light saturation).
When the light signal added to the darkness signal reach 3000 mV, a message “OVL!” in red is
displayed to indicate a saturation of the acquisition system.
It might be caused by ambient light if the door of the analyser is opened with a high level of
ambient light (sun, daylight) or by an optical misalignment.
10 LED 2 signal
This window displays the LED2 signal continuously. This signal corresponds to the reference
light that is not absorbed by the colour of the sample.
On pure water or with a sample without turbidity, the signal must normally be in a 1000 to
2000 mV range (value below 500 mV might deteriorate significantly the repeatability while
values above 2500 mV might cause an error for light saturation).
When the light signal added to the darkness signal reach 3000 mV, a message “OVL!” in red is
displayed to indicate a saturation of the acquisition system.
It might be caused by ambient light if the door of the analyser is opened with a high level of
ambient light (sun, daylight) or by an optical misalignment.
This graph displays the LED1 signal continuously updated. This signal corresponds to the light
that can be absorbed by the yellow colour of the sample.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The average signal value in mV and its standard deviation (sigma) on the last 500 acquisitions
are displayed on the bottom right of the screen.
The values can be read by moving the cursor with the vertical arrows. After reading, restart the
acquisitions by pressing on the “play” button if necessary.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
This graph displays the LED2 signal continuously updated. This signal corresponds to the
reference light that is not absorbed by the organic matter.
For more details or to investigate about the signal, press anywhere on the graph to zoom it.
Refer to LED1 signal graph for the zoom display.
13 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
14 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
16 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
17 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
18 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
19 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
Note: this factor must be set to 1.0 before doing a new linearization curve
20 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
21 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
22 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
23 Offset
This field show the internal offset applied on the signal for the zeroing.
It is updated at each zeroing and has not to be changed.
It can be used for troubleshooting by the manufacturer.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last zeros done with the date, time and zero value adjustment,
whatever manually or by an automatic cleaning cycle.
Note: a strong zero adjustment value might reveal an abnormal zeroing operation.
This area concerns the linearity compensation applied on the rough measurements.
The Beer-Lambert law used on most of the measurements is not linear for high concentrations
due to a light mask effect. Therefore it’s necessary to introduce a correction before displaying
the final concentration. This correction is based on the linearity curve represented on the top.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
27 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.3 .3 Recalibration
- Do a manual measurement on this standard using the check screen (as describe above).
- Press on the “scale” button of the check screen and then enter the standard value on the
keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
4.4.1 Principle
The measuring principle is based on the blue light absorption to the Beer-Lambert law:
[C ] = k ⋅ log I 0
I1
The light is generated by a pulsed xenon lamp to reach an ultra long life time.
The absorption spectrum is performed by a concave grating coupled to a 2048 pixels CCD
(Charge-Coupled Device).
A double-wavelength algorithm allows compensating the turbidity or a deposit inside the flow
cell.
4.4.2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The concentration or error code corresponds to the
last measurement done, as well as the signal and absorbance graphs.
7 1 6 8 9 10 11
2
18
12 29
19
13 26
20
14
21
15
16
17 22 23 24 25 27 28
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit.
During a measuring cycle, the last measured value is display in light green until replaced by the
new value in black.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or the cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
During a measuring cycle or a zeroing cycle, this area displays a counter of the elemental
acquisitions preceded by the message “measuring…” or “zeroing…” respectively.
8 Zero button
Put first pure water inside the flow cell before pressing on this button.
Note: an auto-zero is performed automatically during each acid cleaning if this option is selected
on the timing screen.
9 Scale button
To avoid a wrong calibration, check the zero or redo a zero before doing a
calibration.
10 Signal graph
This graph displays the lamp spectrum get during the last measuring cycle. It is updated during
the measuring cycle when the lamp is flashing.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
With pure water inside the flow cell, the signal must not be saturated (below 2000 mV) on the
two wavelengths bands used for the absorbance calculation defined as “Peak wavelength” and
“Ref wavelength” on the check screen. Otherwise, an error will occur during the measuring
cycle.
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
11 Absorbance spectrum
This graph displays the absorbance spectrum get during the last measuring cycle. It is displayed
in light grey during a measuring cycle until updated.
For more details or to investigate about the absorbance spectrum, press anywhere on the graph
to zoom it as below:
The average absorbance for the peak wavelength band and for the reference wavelength band
are represented by a green line and displayed above the curve.
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
12 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
13 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
15 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
16 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
17 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
18 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
Note: this factor must be set to 1.0 before doing a new linearization curve
19 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
20 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
21 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
22 Peak wavelength
This field allows changing the peak wavelength taken in account for the measurement. It
corresponds to a wavelength absorbed by yellow colour of the sample.
The default value is 350 nm. It can be changed exceptionally to avoid a known interference.
23 Peak width
This field allows changing the width band of the peak wavelength taken in account for the
measurement.
24 Reference wavelength
This field allows changing the reference wavelength taken in account for the measurement.
It corresponds normally to a wavelength where the colour sample doesn’t absorb.
The default value is 590 nm. It can be changed exceptionally to avoid a known interference.
25 Reference width
This field allows changing the width band of the reference wavelength taken in account for the
measurement.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Beer-Lambert law used on most of the measurements is not linear for high concentrations
due to a light mask effect. Therefore it’s necessary to introduce a correction before displaying
the final concentration. This correction is based on the linearity curve represented on the top.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
28 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.4.3 Recalibration
- Do a manual measurement on this standard using the check screen (as describe above).
- Press on the “scale” button of the check screen and then enter the standard value on the
keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
4.5 Conductivity
4.5.1 Principle
The conductivity is the reciprocal of resistance and is determined by the ratio current/voltage
according to Ohm’s law.
1 l amps
C= =
R l volt
Since the measured conductivity depends on the electrode geometry, the measured
conductivity must be multiplied by the cell constant to obtain the specific conductivity expressed
in Siemens per centimeter (S/cm).
d
The cell constant is defined by: k=
A
The conductivity of a solution is proportional to its ion concentration, as long as there is no ionic
interaction.
The default coefficient is 2% per ºC but may be modified according to the solution typical
coefficient obtain experimentally.
4.5.2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The conductivity (or error code), signals and history
graph are continuously updated.
1 6 7 8 9 10
2
19
11 26
20
12 23
21
13
22
14
15
16
17 24 25
18
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit.
During an automatic measuring cycle, the measured value is display in light green until
replaced by the new value in black.
2 Start button
3 Stop button
Pressing on this button stops the sampling pump or the cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
7 Zero button
8 Scale button
Wait that the electrode stabilizes on the standard solution before calibrating. It may
take several minutes, especially if the electrode was dry before using.
To avoid a wrong calibration, calibrate first the electrode on the zero point (as
described above).
9 Signal value
This area displays the conductivity in mV read on the channel of the conductivity module.
This area displays the conductivity value over the last 500 acquisitions, starting when the check
screen appears.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
11 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
12 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
14 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
15 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
16 Temperature adjustment
The temperature of the sample given by the internal temperature probe can be adjusted by this
field. This adjustment is not important as any difference of temperature is taken in account
during the calibration. But if this value is changed significantly, a new calibration must be done.
17 Temperature coefficient
The conductivity is affected by the temperature. This coefficient is used for the temperature
compensation. The default value is 2%. It may be adjusted depending of the solution.
18 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
19 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
20 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
21 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
22 Offset
This field show the internal offset applied on the signal for the zeroing.
It is updated at each zeroing and has not to be changed.
It can be used for troubleshooting by the manufacturer.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
26 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.5.3 Recalibration
The conductivity electrode must be checked on a regular base depending on the condition of
use, and recalibrated if necessary.
- Put the electrode in pure water (or alternatively disconnect the electrode)
- Do manual measurement using the check screen (as describe above), wait for the stabilisation
of the conductivity value, then press on the ZERO key.
- Do manual measurement on this buffer using the check screen, wait for the stabilisation of the
conductivity, then press on the SCALE key and enter the standard value on the keypad.
The calibration is finished. New offset and calibration factors has been calculated and has been
recorded inside the zero and calibration history displayed on the check screen.
These new calibration factors will be taken in account for all the further measurements.
4.6.1 Principle
In presence of oxygen, the fluorescence decreases that decreases the signal on the red
detector.
The dissolved oxygen probe delivers a 4-20 mA signal that is connected to a 4-20 mA input
module. This module also delivers the 12V DC voltage to supply the probe.
4.6.2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The dissolved oxygen value, signals and history graph
are continuously updated.
1 6 7 8 9 10
2
17
11 24
18
12 21
19
13
20
14
15
16 22 23
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit.
During an automatic measuring cycle, the measured value is display in light green until replaced
by the new value in black.
2 Start button
3 Stop button
Pressing on this button stops the sampling pump or the cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
7 Zero button
8 Scale button
This button is used for calibrating the DO electrode on a high value (air for example).
To avoid a wrong calibration, calibrate first the connected device on the zero point
(as described above).
9 Signal value
This area displays the signal in mV read on the input channel of the 4-20 mA input module.
This area displays the dissolved oxygen values over the last 500 acquisitions, starting when the
check screen appears.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
11 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
12 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
14 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
15 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
16 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
17 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
18 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
19 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
20 Offset
This field show the internal offset applied on the signal for the zeroing.
It is updated at each zeroing and has not to be changed.
It can be used for troubleshooting by the manufacturer.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
24 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.6.3 Recalibration
The DO probe must be checked on a regular base depending of the application, and
recalibrated if necessary.
- Put the probe on zero-oxygen solution. A zero solution can be prepared by dissolving 10 g of
sodium sulfite into 300 ml of pure water and adding "a shake" of cobalt chloride, which acts as
a catalyst.
- Do a manual measurement using the check screen (as describe above), wait that the value
stabilizes (it may takes a few minutes) then press on the ZERO key.
- Do a manual measurement using the check screen, wait that the value stabilizes then press
on the SCALE key and enter the value as explained below on the keypad.
If the unit is in mg/l, refer to the table below to enter the appropriate value depending of
the probe temperature:
The calibration is finished. New offset and calibration factors has been calculated and has been
recorded inside the zero and calibration history displayed on the check screen.
These new calibration factors will be taken in account for all the further measurements.
4.7 Hydrocarbons
4.7 .1 Principle
The measuring principle is based on the UV light fluorescence that is a typical characteristic of
the mono or poly aromatics hydrocarbons (PAH)
The UV light is generated by a pulsed xenon lamp to reach an ultra long life time.
The excitation filter limits the bandwidth of the excitation light, while the emission filter selects
the specific fluorescent light.
A second detector (optional) is used to compensate the eventual absorbance of the sample due
for example to organic matter.
4.7 .2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The concentration or error code corresponds to the
last measurement done, while signal values and graph are updated continuously.
1 7 8 9 10 11 12
6
13 22 31
14 23
24
15
16
17
18
19 21 20 25 26 27 28 29 30
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit, as well as the sample temperature and
standard deviation (sigma) obtained on the elemental acquisitions.
During a measuring cycle, the last measured value is display in light green until replaced by the
new value in black.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
7 Scale button
To avoid a wrong calibration, check the zero or redo a zero before doing a
calibration.
8 Zero button
Put first pure water inside the flow cell before pressing on this button.
Note: an auto-zero is performed automatically during each acid cleaning if this option is selected
on the timing screen.
9 PM signal
This window displays the PM signal continuously. This signal corresponds to the fluorescent
light re-emitted by the aromatics hydrocarbons contained in the sample.
On pure water, this signal is normally close to zero with fluctuations of a few tens of mV.
When the light signal added to the darkness signal reach 3000 mV, a message “OVL!” in red is
displayed to indicate a saturation of the acquisition system.
This window displays the photo detector signal continuously (optional). This signal corresponds
to the excitation light of the sample. It is use to compensate an eventual absorption of the
excitation light at 254 nm by the sample, for example organic matter.
When the light signal added to the darkness signal reach 3000 mV, a message “OVL!” in red is
displayed to indicate a saturation of the acquisition system.
It might be caused by ambient light if the door of the analyser is opened with a high level of
ambient light (sun, daylight).
11 PM Signal graph
This graph displays the PM signal continuously updated. This signal corresponds to the
fluorescent light re-emitted by the aromatics hydrocarbons contained in the sample.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The average signal value in mV and its standard deviation (sigma) on the last 500 acquisitions
are displayed on the bottom right of the screen.
The values can be read by moving the cursor with the vertical arrows. After reading, restart the
acquisitions by pressing on the “play” button if necessary.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
This graph displays the photo detector signal continuously updated. This signal corresponds to
the excitation light of the sample.
For more details or to investigate about the signal, press anywhere on the graph to zoom it.
Refer to PM signal graph for the zoom display.
13 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
14 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
15 Display mode
This button allows selecting the concentration display mode as PAH or as OIL (default mode).
The measurements previously stored in the internal memory keep their original values.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
17 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
18 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
19 Temperature adjustment
The temperature of the sample given by the internal temperature probe can be adjusted by this
field. This adjustment is not important as any difference of temperature is taken in account
during the calibration. But if this value is changed, a new calibration must be done.
20 Temperature coefficient
The fluorescence is affected by the temperature. This coefficient is used for the temperature
compensation.
21 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
22 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
Note: this factor must be set to 1.0 before doing a new linearization curve
23 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
24 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
25 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
26 Turbidity factor
This field enable a compensation of the excitation light loses due to turbidity, organic matter.
A value of 1.0 corresponds to a basic compensation.
27 Offset
This field show the internal offset applied on the signal for the zeroing.
It is updated at each zeroing and has not to be changed.
It can be used for troubleshooting by the manufacturer.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last zeros done with the date, time and zero value adjustment,
whatever manually or by an automatic cleaning cycle.
Note: a strong zero adjustment value might reveal an abnormal zeroing operation.
This area concerns the linearity compensation applied on the rough measurements.
The Beer-Lambert law used on most of the measurements is not linear for high concentrations
due to a light mask effect. Therefore it’s necessary to introduce a correction before displaying
the final concentration. This correction is based on the linearity curve represented on the top.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
31 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.7 .3 Recalibration
The analyser is factory calibrated on phenol. Depending on the aromatic ratio of the sample a
recalibration based on a laboratory measurement is generally recommended.
- When the result is obtained, do a manual measurement on the remaining sample using the
check screen describe above.
- Press on the “scale” button of the check screen and then enter the sample value determined
by the laboratory on the keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
4.8 .1 Principle
The measuring principle is based on the analysis of the UV absorption spectrum of the
hydrogen sulphide gas (H2S) in equilibrium with the dissolved hydrogen sulphide gas in the
water sample.
A small quantity of hydrochloric acid (HCl) is added to the sample to decreasing the pH below 1
where most of the S2- is transformed in H2S (>99.9%).
The UV light is generated by a pulsed xenon lamp to reach an ultra long life time.
The absorption spectrum is performed by a concave grating coupled to a 2048 pixels CCD
(Charge-Coupled Device).
Automatic temperature compensation is done using a Pt100 temperature probe on the sample.
4.8 .2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. It corresponds to the last measurement done
(automatic or manual) concerning the concentration or error code, sample temperature, lamp
spectrum, absorbance spectrum and FFT spectrum.
7 1 6 8 9 10
4 31
3
32
20
11
30
21
12 22
23
13 24
25
14 26
27
15 28
16 19 18 17 29
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit, as well as the sample temperature read
during the measurement.
During a measuring cycle, the last measured value is display in light green until replaced by the
new value in black.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or the cleaning pump or the ultrasonic cleaning (if any).
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the HCl pump for testing purpose or to fill the tubing after
installation. The pump will stop automatically after 10 seconds.
It can also be stopped at any time by pressing on the stop button.
During a measuring cycle, this area displays the different steps of the cycle with an associated
countdown timer or counter.
During the ammonia measurements, the steps are:
4) Reagent: the HCl is added to the sample inside the stripping pot
5) Stripping: the gas flow cell is placed in a close loop with the stripping pot
6) Measuring: the spectrum is acquired on the gas coming through the sample, and then the
concentration is computed and displayed
7) Purging: the gas flow cell is purged again by external ambient air to remove humidity
8 Scale button
9 Signal graph
This graph displays the lamp spectrum get during the last measuring cycle. It is updated during
the measuring cycle when the lamp is flashing.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The signal must not be saturated (below 2000 mV) for the wavelengths used by the
measurement, that are below 225 nm.
It’s recommended to have a signal at 210 nm around 500 mV to get a good repeatability, but
measurement can still be done for signal as low as 100 mV if the repeatability is less critical.
If the light signal is too high, an alarm message is displayed on the top left corner.
The values can be read by moving the cursor with the vertical arrows.
10 Absorbance spectrum
This graph displays the absorbance spectrum get during the last measuring cycle. It is displayed
in light grey during a measuring cycle until updated.
For more details or to investigate about the absorbance spectrum, press anywhere on the graph
to zoom it as below:
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows located above the Y axis.
11 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
12 Unit
This field allows changing the channel unit displayed with the measurement. The unit size is
limited to 6 characters.
13 Display mode
This button allows selecting the concentration display mode as H2S (default mode) or as S of
H2S.
If S-H2S is selected, the H2S concentrations are multiplied by 32/34=0.941 before displaying
and the label is automatically updated to “S-H2S”.
The measurements previously stored in the internal memory will keep their values in H2S.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
15 Number of decimal
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
16 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
17 Temperature adjustment
The temperature of the sample given by the internal temperature probe can be adjusted by this
field. This adjustment is not important as any difference of temperature is taken in account
during the calibration. But if this value is changed, a new calibration must be done.
18 Temperature coefficient
The sample temperature affects directly the H2S liquid/gas ratio. This coefficient is used for the
temperature compensation. It can be change only by the manufacturer.
19 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
20 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
Note: this factor must be set to 1.0 before doing a new linearization curve
21 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
22 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
23 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
24 Reaction time
25 Reagent time
26 Peak wavelength
This field allows changing the peak wavelength taken in account for the measurement. It
corresponds to the absorbance peak of H2S.
27 Peak width
This field allows changing the width band of the peak wavelength taken in account for the
measurement.
28 Reference wavelength
This field allows changing the reference wavelength taken in account for the measurement.
It corresponds normally to a wavelength where the H2S doesn’t absorb.
29 Reference width
This field allows changing the width band of the reference wavelength taken in account for the
measurement.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Beer-Lambert law used on most of the measurements is not linear for high concentrations
due to a light mask effect. Therefore it’s necessary to introduce a correction before displaying
the final concentration. This correction is based on the linearity curve represented on the top.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0,0
as starting values for S1 and M1. The final value is linearly extrapolated between these points.
32 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.8 .3 Recalibration
- Prepare a 10 mg/l H2S standard. For this weight 2.434 g of dried Na2S and dilute in 1litre of
distilled water to obtain a mere solution of 1g/l. Then take 10 ml of this mere solution and
complete to 1 litre of distilled water.
- Do at least 3 manual measurements on this 10 mg/l standard using the check screen. The last
two measurements must be close together with less than 5% of difference.
- Press on the “scale” button of the check screen (as describe above) and then enter the
standard value on the keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
4.9.1 Principle
The 4-20mA input give in percent the current value on a 4-20 mA loop connected to another
instrument. The load resistance is 100 ohm.
4.9.2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The input value, signals and history graph are
continuously updated.
1 6 7 8 9 10
2
17
11 24
18
12 21
19
13
20
14
15
16 22 23
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit.
During an automatic measuring cycle, the measured value is display in light green until replaced
by the new value in black.
2 Start button
3 Stop button
Pressing on this button stops the sampling pump or the cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
7 Zero button
8 Scale button
This button is used for calibrating the connected device on a high value.
To avoid a wrong calibration, calibrate first the connected device on the zero point
(as described above).
9 Signal value
This area displays the signal in mV read on the input of the 4-20 mA input module.
This area displays the 4-20 mA input value over the last 500 acquisitions, starting when the
check screen appears.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
11 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
12 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
14 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
15 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
16 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
17 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
18 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
19 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
20 Offset
This field show the internal offset applied on the signal for the zeroing.
It is updated at each zeroing and has not to be changed.
It can be used for troubleshooting by the manufacturer.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
24 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.9.3 Recalibration
The 4-20 mA input can be recalibrated to take in account a drift of the connected device.
- Assume that the connected device is measuring a zero standard (or alternatively replace by a
4 mA generator)
- Do a manual measurement using the check screen (as describe above) then press on the
ZERO key.
- Assume that the connected device is measuring a known standard (or alternatively replace by
a 20 mA generator)
- Do a manual measurement using the check screen, then press on the SCALE key and enter
the standard value on the keypad.
The calibration is finished. New offset and calibration factors has been calculated and has been
recorded inside the zero and calibration history displayed on the check screen.
These new calibration factors will be taken in account for all the further measurements.
4.10.1 Principle
The measuring principle is based on the UV light absorption around 220 nm according to the
Beer-Lambert law:
[C ] = k ⋅ log I 0
I1
The UV light is generated by a pulsed xenon lamp to reach an ultra long life time.
The light detection is performed by two photodiode detectors through interferential filters; one
centred on the nitrate absorption band while the other is used as reference to compensate
organic matter, turbidity or deposit inside the flow cell.
4.10.2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The concentration or error code corresponds to the
last measurement done, while the signals are continuously updated.
7 1 6 8 9 10 11 12 13
2
21
14 29
22
15 26
23
16
24
17
18
19 20 25 27 28
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit.
During a measuring cycle, the last measured value is display in light green until replaced by the
new value in black.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or the cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
During a measuring cycle or a zeroing cycle, this area displays a counter of the elemental
acquisitions preceded by the message “measuring…” or “zeroing…” respectively.
8 Zero button
Put first pure water inside the flow cell before pressing on this button.
Note: an auto-zero is performed automatically during each acid cleaning if this option is selected
on the timing screen.
9 Scale button
To avoid a wrong calibration, check the zero or redo a zero before doing a
calibration.
This window displays the photo detector 1 (PH1) signal continuously. This signal corresponds to
the light absorbed by the nitrate.
When the light signal added to the darkness signal reach 3000 mV, a message “OVL!” in red is
displayed to indicate a saturation of the acquisition system.
This window displays the photo detector 2 (PH2) signal continuously. This signal corresponds to
the reference light that is not absorbed by the nitrate. It is use to compensate turbidity or organic
matter or deposit on the flow cell.
When the light signal added to the darkness signal reach 3000 mV, a message “OVL!” in red is
displayed to indicate a saturation of the acquisition system.
This graph displays the PH1 signal continuously updated. This signal corresponds to the light
absorbed by the nitrate.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The average signal value in mV and its standard deviation (sigma) on the last 500 acquisitions
are displayed on the bottom right of the screen.
The values can be read by moving the cursor with the vertical arrows. After reading, restart the
acquisitions by pressing on the “play” button if necessary.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
This graph displays the photo detector no 2 (PH2) signal continuously updated. This signal
corresponds to the reference light through the sample.
For more details or to investigate about the signal, press anywhere on the graph to zoom it.
Refer to photo detector no 2 signal graph for the zoom display.
14 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
15 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
16 Display mode
This button allows selecting the concentration display mode as NO3 (default mode) or as N of
NO3.
If N-NO3 is selected, the NO3 concentrations are multiplied by 14/62=0.2258 before displaying
and the label is automatically updated to “N-NO3”.
The measurements previously stored in the internal memory will keep their original values.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
18 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
19 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
20 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
21 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
Note: this factor must be set to 1.0 before doing a new linearization curve
22 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
23 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
24 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
25 Offset
This field show the internal offset applied on the signal for the zeroing.
It is updated at each zeroing and has not to be changed.
It can be used for troubleshooting by the manufacturer.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Beer-Lambert law used on most of the measurements is not linear for high concentrations
due to a light mask effect. Therefore it’s necessary to introduce a correction before displaying
the final concentration. This correction is based on the linearity curve represented on the top.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
29 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.10.3 Recalibration
- Prepare a 50 mg/l NO3 standard. For this weight 1.629 g of dried KNO3 and dilute in 1litre of
distilled water to obtain a mere solution of 1g/l. Then take 50 ml of this mere solution and
complete to 1 litre of distilled water.
- Do a manual measurement on this standard using the check screen (as describe above).
- Press on the “scale” button of the check screen and then enter the standard value on the
keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
4.11.1 Principle
The measuring principle is based on the UV light absorption around 220 nm according to the
Beer-Lambert law:
[C ] = k ⋅ log I 0
I1
The UV light is generated by a pulsed xenon lamp to reach an ultra long life time.
The absorption spectrum is performed by a concave grating coupled to a 2048 pixels CCD
(Charge-Coupled Device).
A double-wavelength algorithm allows compensating the turbidity or a deposit inside the flow
cell.
4.11.2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The concentration or error code corresponds to the
last measurement done, as well as the signal and absorbance graphs.
7 1 6 8 9 10 11
2
19
12 30
20
13 27
21
14
22
15
16
17 18 23 24 25 26 28 29
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit.
During a measuring cycle, the last measured value is display in light green until replaced by the
new value in black.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or the cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
During a measuring cycle or a zeroing cycle, this area displays a counter of the elemental
acquisitions preceded by the message “measuring…” or “zeroing…” respectively.
8 Zero button
Put first pure water inside the flow cell before pressing on this button.
Note: an auto-zero is performed automatically during each acid cleaning if this option is selected
on the timing screen.
9 Scale button
To avoid a wrong calibration, check the zero or redo a zero before doing a
calibration.
10 Signal graph
This graph displays the lamp spectrum get during the last measuring cycle. It is updated during
the measuring cycle when the lamp is flashing.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
With pure water inside the flow cell, the signal must not be saturated (below 2000 mV) on the
two wavelengths bands used for the absorbance calculation defined as “Peak wavelength” and
“Ref wavelength” on the check screen. Otherwise, an error will occur during the measuring
cycle.
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
11 Absorbance spectrum
This graph displays the absorbance spectrum get during the last measuring cycle. It is displayed
in light grey during a measuring cycle until updated.
For more details or to investigate about the absorbance spectrum, press anywhere on the graph
to zoom it as below:
The average absorbance for the peak wavelength band and for the reference wavelength band
are represented by a green line and displayed above the curve.
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
12 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
13 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
14 Display mode
This button allows selecting the concentration display mode as NO3 (default mode) or as N of
NO3.
If N-NO3 is selected, the NO3 concentrations are multiplied by 14/62=0.2258 before displaying
and the label is automatically updated to “N-NO3”.
The measurements previously stored in the internal memory will keep their original values.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
16 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
17 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
18 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
19 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
Note: this factor must be set to 1.0 before doing a new linearization curve
20 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
21 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
22 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
23 Peak wavelength
This field allows changing the peak wavelength taken in account for the measurement. It
corresponds to a wavelength absorbed by nitrate.
The default value is 220 nm. It can be changed exceptionally to avoid a known interference or to
reduce the sensitivity for high range measurements.
24 Peak width
This field allows changing the width band of the peak wavelength taken in account for the
measurement.
25 Reference wavelength
This field allows changing the reference wavelength taken in account for the measurement.
It corresponds normally to a wavelength where the nitrate doesn’t absorb.
The default value is 254 nm to compensate the organic matter. It can be changed exceptionally
to avoid a known interference.
26 Reference width
This field allows changing the width band of the reference wavelength taken in account for the
measurement.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Beer-Lambert law used on most of the measurements is not linear for high concentrations
due to a light mask effect. Therefore it’s necessary to introduce a correction before displaying
the final concentration. This correction is based on the linearity curve represented on the top.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
27 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.11.3 Recalibration
- Prepare a 50 mg/l NO3 standard. For this weight 1.629 g of dried KNO3 and dilute in 1litre of
distilled water to obtain a mere solution of 1g/l. Then take 50 ml of this mere solution and
complete to 1 litre of distilled water.
- Do a manual measurement on this standard using the check screen (as describe above).
- Press on the “scale” button of the check screen and then enter the standard value on the
keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
4.12.1 Principle
The reduction potential is a measure of the tendency of the solution to either gain or lose
electrons when it is subject to change by introduction of a new species.
A solution with a higher (more positive) reduction potential than the new species will have a
tendency to gain electrons from the new species (i.e. to be reduced by oxidizing the new
species) and a solution with a lower (more negative) reduction potential will have a tendency to
lose electrons to the new species (i.e. to be oxidized by reducing the new species).
The method of measurement is based on the potential between two half-cells (reference and
sensing) containing a conductor immersed in an appropriate electrolyte solution and ended by a
conductive glass membrane. These two half-cells are generally combined in a single body
electrode.
4.12.2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The concentration (or error code), signals and history
graph are continuously updated.
1 6 7 8 9 10
2
17
11 24
18
12 21
19
13
20
14
15
16 22 23
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit.
During an automatic measuring cycle, the measured value is display in light green until
replaced by the new value in black.
2 Start button
3 Stop button
Pressing on this button stops the sampling pump or the cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
7 Zero button
Put first pure water inside the flow cell before pressing on this button.
Alternatively, replace the electrode by a strap.
8 Scale button
Wait that the electrode stabilizes on the standard before calibrating. It may take
several minutes.
To avoid a wrong calibration, calibrate first the electrode on the low value point (as
described above).
9 Signal value
This area displays the ORP signal in mV read on the input of the pH module.
This area displays the pH value over the last 500 acquisitions, starting when the check screen
appears.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
11 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
12 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
14 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
15 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
16 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
17 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
18 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
19 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
20 Offset
This field show the internal offset applied on the signal for the zeroing.
It is updated at each zeroing and has not to be changed.
It can be used for troubleshooting by the manufacturer.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
24 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.12.3 Recalibration
The ORP electrode must be checked on a regular base depending on the condition of use, and
recalibrated if necessary.
- Put the electrode in a zero buffer solution (or alternatively replace the electrode by a strap)
- Do manual measurement on this buffer using the check screen (as describe above), wait for
the stabilisation of the ORP value, then press on the ZERO key.
- Do manual measurement on this buffer using the check screen, wait for the stabilisation of the
pH value, then press on the SCALE key and enter the standard value on the keypad.
The calibration is finished. New offset and calibration factors has been calculated and has been
recorded inside the zero and calibration history displayed on the check screen.
These new calibration factors will be taken in account for all the further measurements.
4.13 pH
4.13.1 Principle
The pH of a solution is the negative logarithm of the molar concentration of dissolved hydronium
ions (H3O+).
The method of measurement is based on the potential between two half-cells (reference and
sensing) containing a conductor immersed in an appropriate electrolyte solution and ended by a
conductive glass membrane. These two half-cells are generally combined in a single body
electrode.
The voltage between the half-cell follows the Nernst equation:
V = V0 + 2,3
RT [
H +m
⋅ log +
]
nF [H ref ]
With V0: voltage of the reference half-cell, R: ideal gas constant, T: absolute temperature, n
:valence, F: Faraday constant, H+m: concentration of H+ in the measured solution, H+ref:
reference concentration of H+.
At 25ºC, one pH unit corresponds to 59.16 mV. At pH 7, the output is 0V (for all temperatures).
4.13.2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The concentration (or error code), signals and history
graph are continuously updated.
1 6 7 8 9 10
2
18
11 25
19
12 22
20
13
21
14
15
16
17 23 24
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit.
During an automatic measuring cycle, the measured value is display in light green until replaced
by the new value in black.
2 Start button
3 Stop button
Pressing on this button stops the sampling pump or the cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
7 Low button
This button is used for calibrating the measurement on a low value, for example a buffer
solution of 4.0 pH.
Wait that the electrode stabilizes on the standard before calibrating. It may take
several minutes.
8 High button
This button is used for calibrating the measurement on a high value, for example a buffer
solution of 10.0 pH.
Wait that the electrode stabilizes on the standard before calibrating. It may take
several minutes.
To avoid a wrong calibration, calibrate first the electrode on the low value point (as
described above).
9 Signal value
This area displays the pH signal in mV read on the input of the pH module. It corresponds to a
theoretical value of 0 mV for a pH of 7, -414 mV for a pH of 14 and +414 mV for a pH of 0.
10 pH history graph
This area displays the pH value over the last 500 acquisitions, starting when the check screen
appears.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
11 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
12 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
14 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
15 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
16 Temperature adjustment
The temperature of the sample given by the internal temperature probe can be adjusted by this
field. This adjustment is not important as any difference of temperature is taken in account
during the calibration. But if this value is changed significantly, a new calibration must be done.
17 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
18 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
19 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
20 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
21 Offset
This field show the internal offset applied on the signal for the zeroing.
It is updated at each zeroing and has not to be changed.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
25 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.13.3 Recalibration
The pH electrode must be checked on a regular base depending on the condition of use, and
recalibrated if necessary.
- Do manual measurement on this buffer using the check screen (as describe above), wait for
the stabilisation of the pH value, then press on the LOW key and enter the standard value on
the keypad.
- Do manual measurement on this buffer using the check screen, wait for the stabilisation of the
pH value, then press on the HIGH key and enter the standard value on the keypad.
The calibration is finished. New offset and calibration factors has been calculated and has been
recorded inside the zero and calibration history displayed on the check screen.
These new calibration factors will be taken in account for all the further measurements.
4.14 .1 Principle
The measuring principle is based on colorimetry. A small quantity of reagent is injected on the
sample and after a fixed reaction time the developed colour is measured by the light absorption
according to the Beer-Lambert law:
[C ] = k ⋅ log I 0
I1
A second light beam is used to compensate the turbidity of the sample or an eventual deposit
inside the flow cell.
The light is generated alternatively by two LEDs.
4.14 .2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The concentration or error code corresponds to the
last measurement done, while signal values and graph are updated continuously.
1 7 8 9 10 11 12
6
3
28
2
13 22 30
14 23
24
15
25
16
26
17
27
18
21 20 19 29
Measurement window
1
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit, as well as the standard deviation (sigma)
obtained on the elemental acquisitions.
During a measuring cycle, the last measured value is display in light green until replaced by the
new value in black.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the reagent pump for testing purpose or to fill the tubing after
installation. The pump will stop automatically after 10 seconds.
It can also be stopped at any time by pressing on the stop button.
7 Scale button
To avoid a wrong calibration, check the zero or redo a zero before doing a
calibration.
8 LED 1 signal
This window displays the LED1 signal continuously. This signal corresponds to the light that can
be absorbed by the colour developed inside the sample.
On pure water, the signal must normally be in a 1000 to 2000 mV range (value below 500 mV
might deteriorate the repeatability while values above 2500 mV might cause an error for light
saturation).
When the light signal added to the darkness signal reach 3000 mV, a message “OVL!” in red is
displayed to indicate a saturation of the acquisition system.
It might be caused by ambient light if the door of the analyser is opened with a high level of
ambient light (sun, daylight).
9 LED 2 signal
This window displays the LED2 signal continuously. This signal corresponds to the reference
light that is not absorbed by the colour developed inside the sample.
On pure water or with a sample without turbidity, the signal must normally be in a 1000 to
2000 mV range (value below 500 mV might deteriorate significantly the repeatability while
values above 2500 mV might cause an error for light saturation).
When the light signal added to the darkness signal reach 3000 mV, a message “OVL!” in red is
displayed to indicate a saturation of the acquisition system.
It might be caused by ambient light if the door of the analyser is opened with a high level of
ambient light (sun, daylight) or by an optical misalignment.
This graph displays the LED1 signal continuously updated. This signal corresponds to the light
that can be absorbed by the colour developed inside the sample.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The average signal value in mV and its standard deviation (sigma) on the last 500 acquisitions
are displayed on the bottom right of the screen.
The values can be read by moving the cursor with the vertical arrows. After reading, restart the
acquisitions by pressing on the “play” button if necessary.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
This graph displays the LED2 signal continuously updated. This signal corresponds to the
reference light that is not absorbed by the colour developed inside the sample.
For more details or to investigate about the signal, press anywhere on the graph to zoom it.
Refer to LED1 signal graph for the zoom display.
12 Reaction graph
This graph shows the PO4 measured concentration all along the reaction time for checking the
reaction development.
13 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
14 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
15 Display mode
This button allows selecting the concentration display mode as P of PO4 (default mode) or as
PO4.
If P-PO4 is selected, the PO4 concentrations are multiplied by 32/96=0.3333 before displaying
and the label is automatically updated to “P-PO4”.
The measurements previously stored in the internal memory will keep their original values.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
17 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
18 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
19 Temperature adjustment
The temperature of the sample given by the internal temperature probe can be adjusted by this
field. This adjustment is not important as any difference of temperature is taken in account
during the calibration. But if this value is changed significantly, a new calibration must be done.
20 Temperature coefficient
The colorimetric result is slightly affected by the temperature. This coefficient is used for the
temperature compensation.
21 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
22 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
Note: this factor must be set to 1.0 before doing a new linearization curve
23 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
24 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
25 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
26 Reaction time
Precautions must be taken before changing this value as an incomplete reaction time can affect
the result.
27 Reagent time
Precautions must be taken before changing this value as improper reagent concentration might
affect the result.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Beer-Lambert law used on most of the measurements is not linear for high concentrations
due to a light mask effect. Therefore it’s necessary to introduce a correction before displaying
the final concentration. This correction is based on the linearity curve represented on the top.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
30 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.14 .3 Recalibration
- Prepare a 1 mg/l P-PO4 standard. For this weight 4.39 g of dried potassium phosphate
(KH2PO4) and dilute in 1litre of distilled water to obtain a mere solution of 1g/l NH4.
Then take 1 ml of this mere solution and complete to 1 litre of distilled water.
- Do a manual measurement on this standard using the check screen (as describe above).
- Press on the “scale” button of the check screen and then enter the standard value on the
keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
4.15 .1 Principle
The measuring principle is based on colorimetry. A small quantity of reagent is injected on the
sample and after a fixed reaction time the developed colour is measured by the light absorption
according to the Beer-Lambert law:
[C ] = k ⋅ log I 0
I1
The light is generated by a pulsed xenon lamp to reach an ultra long life time.
The absorption spectrum is performed by a concave grating coupled to a 2048 pixels CCD
(Charge-Coupled Device).
4.15 .2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The concentration or error code corresponds to the
last measurement done, as well as the signal graph and reaction graph.
1 7 8 9 10
6
3
30
2
11 20 32
12 21
22
13
23
14
24
15
25
16
26
31
17 19 18 29 28 27
Measurement window
1
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit, as well as the standard deviation (sigma)
obtained on the elemental acquisitions.
During a measuring cycle, the last measured value is display in light green until replaced by the
new value in black.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the reagent pump for testing purpose or to fill the tubing after
installation. The pump will stop automatically after 10 seconds.
It can also be stopped at any time by pressing on the stop button.
7 Scale button
To avoid a wrong calibration, check the zero or redo a zero before doing a
calibration.
8 Signal graph
This graph displays the lamp spectrum get during the last measuring cycle. It is updated during
the measuring cycle when the lamp is flashing.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
With pure water inside the flow cell, the signal must not be saturated (below 2000 mV) on the
two wavelengths bands used for the absorbance calculation defined as “Peak wavelength” and
“Ref wavelength” on the check screen. Otherwise, an error will occur during the measuring
cycle.
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
9 Absorbance spectrum
This graph displays the absorbance spectrum get during the last measuring cycle. It is displayed
in light grey during a measuring cycle until updated.
For more details or to investigate about the absorbance spectrum, press anywhere on the graph
to zoom it as below:
The average absorbance for the peak wavelength band and for the reference wavelength band
are represented by a green line and displayed above the curve.
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
10 Reaction graph
This graph shows the PO4 measured concentration all along the reaction time for checking the
reaction development.
11 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
12 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
13 Display mode
This button allows selecting the concentration display mode as P of PO4 (default mode) or as
PO4.
If P-PO4 is selected, the PO4 concentrations are multiplied by 32/96=0.3333 before displaying
and the label is automatically updated to “P-PO4”.
The measurements previously stored in the internal memory will keep their original values.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
15 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
16 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
17 Temperature adjustment
The temperature of the sample given by the internal temperature probe can be adjusted by this
field. This adjustment is not important as any difference of temperature is taken in account
during the calibration. But if this value is changed significantly, a new calibration must be done.
18 Temperature coefficient
The colorimetric result is slightly affected by the temperature. This coefficient is used for the
temperature compensation.
19 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
20 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
Note: this factor must be set to 1.0 before doing a new linearization curve
21 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
22 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
23 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
24 Reaction time
Precautions must be taken before changing this value as an incomplete reaction time can affect
the result.
25 Reagent time
Precautions must be taken before changing this value as improper reagent concentration might
affect the result.
26 Peak wavelength
This field allows changing the peak wavelength taken in account for the measurement. It
corresponds to a wavelength absorbed by developed colour inside the sample.
27 Peak width
This field allows changing the width band of the peak wavelength taken in account for the
measurement.
28 Reference wavelength
This field allows changing the reference wavelength taken in account for the measurement.
It corresponds normally to a wavelength where developed colour inside the sample doesn’t
absorb.
29 Reference width
This field allows changing the width band of the reference wavelength taken in account for the
measurement.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Beer-Lambert law used on most of the measurements is not linear for high concentrations
due to a light mask effect. Therefore it’s necessary to introduce a correction before displaying
the final concentration. This correction is based on the linearity curve represented on the top.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
32 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.15 .3 Recalibration
- Prepare a 1 mg/l P-PO4 standard. For this weight 4.39 g of dried potassium phosphate
(KH2PO4) and dilute in 1litre of distilled water to obtain a mere solution of 1g/l NH4.
Then take 1 ml of this mere solution and complete to 1 litre of distilled water.
- Do a manual measurement on this standard using the check screen (as describe above).
- Press on the “scale” button of the check screen and then enter the standard value on the
keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
4.16 Rhodamine
4.16 .1 Principle
The measuring principle is based on the orange light fluorescence that is a typical characteristic
of the Rhodamine.
The light is generated by a pulsed xenon lamp to reach an ultra long life time.
The excitation filter limits the bandwidth of the excitation light, while the emission filter selects
the specific fluorescent light.
A second detector (optional) is used to compensate the eventual absorbance of the sample due
to turbidity.
4.16 .2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The concentration or error code corresponds to the
last measurement done, while signal values and graph are updated continuously.
1 7 8 9 10 11 12
6
13 21 30
14 22
23
15
16
17
18
19 20 24 25 26 27 28 29
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit, as well as the sample temperature and
standard deviation (sigma) obtained on the elemental acquisitions.
During a measuring cycle, the last measured value is display in light green until replaced by the
new value in black.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
7 Scale button
To avoid a wrong calibration, check the zero or redo a zero before doing a
calibration.
8 Zero button
Put first pure water inside the flow cell before pressing on this button.
Note: an auto-zero is performed automatically during each acid cleaning if this option is selected
on the timing screen.
9 PM signal
This window displays the PM signal continuously. This signal corresponds to the fluorescent
light re-emitted by the Rhodamine.
On pure water, this signal is normally close to zero with fluctuations of a few tens of mV.
When the light signal added to the darkness signal reach 3000 mV, a message “OVL!” in red is
displayed to indicate a saturation of the acquisition system.
It might be caused by an over-range concentration of the sample or by the ambient light if the
door of the analyser is opened with a high level of ambient light (sun, daylight).
This window displays the photo detector signal continuously (optional). This signal corresponds
to the excitation light of the sample. It is use to compensate an eventual absorption of the
excitation blue light by the sample, for example due to turbidity.
When the light signal added to the darkness signal reach 3000 mV, a message “OVL!” in red is
displayed to indicate a saturation of the acquisition system.
It might be caused by ambient light if the door of the analyser is opened with a high level of
ambient light (sun, daylight).
11 PM Signal graph
This graph displays the PM signal continuously updated. This signal corresponds to the
fluorescent light re-emitted by the Rhodamine.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The average signal value in mV and its standard deviation (sigma) on the last 500 acquisitions
are displayed on the bottom right of the screen.
The values can be read by moving the cursor with the vertical arrows. After reading, restart the
acquisitions by pressing on the “play” button if necessary.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
This graph displays the photo detector signal continuously updated. This signal corresponds to
the excitation light of the sample.
For more details or to investigate about the signal, press anywhere on the graph to zoom it.
Refer to PM signal graph for the zoom display.
13 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
14 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
16 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
17 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
18 Temperature adjustment
The temperature of the sample given by the internal temperature probe can be adjusted by this
field. This adjustment is not important as any difference of temperature is taken in account
during the calibration. But if this value is changed, a new calibration must be done.
19 Temperature coefficient
The fluorescence is affected by the temperature. This coefficient is used for the temperature
compensation.
20 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
21 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
Note: this factor must be set to 1.0 before doing a new linearization curve
22 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
23 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
24 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
25 Turbidity factor
This field enable a compensation of the excitation light loses due to turbidity.
A value of 1.0 corresponds to a basic compensation.
26 Offset
This field show the internal offset applied on the signal for the zeroing.
It is updated at each zeroing and has not to be changed.
It can be used for troubleshooting by the manufacturer.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last zeros done with the date, time and zero value adjustment,
whatever manually or by an automatic cleaning cycle.
Note: a strong zero adjustment value might reveal an abnormal zeroing operation.
This area concerns the linearity compensation applied on the rough measurements.
The Beer-Lambert law used on most of the measurements is not linear for high concentrations
due to a light mask effect. Therefore it’s necessary to introduce a correction before displaying
the final concentration. This correction is based on the linearity curve represented on the top.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
30 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.16 .3 Recalibration
- Press on the “scale” button of the check screen and then enter the standard value
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
4.17.1 Principle
The measuring principle is based on the full UV-visible light absorption according to the Beer-
Lambert law:
[C ] = k ⋅ log I 0
I1
The light is generated by a pulsed xenon lamp to reach an ultra long life time.
The absorption spectrum is performed by a concave grating coupled to a 2048 pixels CCD
(Charge-Coupled Device).
4.17.2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
7 1 6 8 9 10
11 13
12
14
15
1 Measurement window
This window displays only the label and unit as no measured value is attached to the full UV-
VIS spectrum. In case of a measuring error, the error code is written, refer to troubleshooting
section.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or the cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
During a measuring cycle or a zeroing cycle, this area displays a counter of the elemental
acquisitions preceded by the message “measuring…” or “zeroing…” respectively.
8 Zero button
Put first pure water inside the flow cell before pressing on this button.
Note: an auto-zero is performed automatically during each acid cleaning if this option is selected
on the timing screen.
9 Signal graph
This graph displays the lamp spectrum get during the last measuring cycle. It is updated during
the measuring cycle when the lamp is flashing.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
With pure water inside the flow cell, the signal must not be saturated (below 2000 mV) over the
full spectrum, otherwise, an error will occur during the measuring cycle.
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
10 Absorbance spectrum
This graph displays the absorbance spectrum get during the last measuring cycle. It is displayed
in light grey during a measuring cycle until updated.
For more details or to investigate about the absorbance spectrum, press anywhere on the graph
to zoom it as below:
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
11 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
12 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
13 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
14 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
4.18 Temperature
4.18.1 Principle
The table below shows the Pt 100 resistance in ohm for temperature between 0°C and 100°C.
°C 0 1 2 3 4 5 6 7 8 9
0 100.00 100.39 100.78 101.17 101.56 101.95 102.34 102.73 103.12 103.51
10 103.90 104.29 104.68 105.07 105.46 105.85 106.24 106.63 107.02 107.40
20 107.79 108.18 108.57 108.96 109.35 109.73 110.12 110.51 110.90 111.28
30 111.67 112.06 112.45 112.83 113.22 113.61 113.99 114.38 114.77 115.15
40 115.54 115.93 116.31 116.70 117.08 117.47 117.85 118.24 118.62 119.01
50 119.40 119.78 120.16 120.55 120.93 121.32 121.70 122.09 122.47 122.86
60 123.24 123.62 124.01 124.39 124.77 125.16 125.54 125.92 126.31 126.69
70 127.07 127.45 127.84 128.22 128.60 128.98 129.37 129.75 130.13 130.51
80 130.89 131.27 131.66 132.04 132.42 132.80 133.18 133.56 133.94 134.32
90 134.70 135.08 135.46 135.84 136.22 136.60 136.98 137.36 137.74 138.12
4.18.2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The temperature (or error code), signals and history
graph are continuously updated.
1 6 7 8 9 10
2
18
11 25
19
12 22
20
13
21
14
15
16
17 23 24
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit.
During an automatic measuring cycle, the measured value is display in light green until replaced
by the new value in black.
2 Start button
3 Stop button
Pressing on this button stops the sampling pump or the cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
7 Zero button
Put first the temperature element at 0°C in a Dewar filled with ice and water.
8 Scale button
This button is used to calibrate the temperature element with a reference thermometer.
To avoid a wrong calibration, check the zero or redo a zero before doing a
calibration.
9 Signal value
This area displays the signal in mV read on the temperature channel of the electrode module.
This area displays the temperature over the last 500 acquisitions, starting when the check
screen appears.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
11 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
12 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
14 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
15 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
16 Temperature adjustment
The temperature can be adjusted with a reference thermometer without doing a complete
calibration by entering an offset value in this field.
17 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
18 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
19 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit.
It is has no reason to be changed except if internal elements are replaced.
20 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
21 Offset
This field show the internal offset applied on the signal for the zeroing.
It is updated at each zeroing and has not to be changed.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
Normally, no linearity correction is necessary for the temperature measurement on the water
temperature range 0 to 100°C.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
25 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.18.3 Recalibration
The temperature probe is factory calibrated with an accuracy of +/- 1°C. It can be recalibrated if
a higher accuracy is requested.
- Do manual measurement using the check screen (as describe above), wait for the stabilisation
of the temperature value, and then press on the ZERO key.
- Put the temperature probe in a Dewar filled of water at a temperature between 50°C and 80°C
with a reference thermometer inside
- Do manual measurement using the check screen, wait for the stabilisation of the temperature
value, then press on the SCALE key and enter the value given by the reference thermometer on
the keypad.
The calibration is finished. New offset and calibration factors has been calculated and has been
recorded inside the zero and calibration history displayed on the check screen.
These new calibration factors will be taken in account for all the further measurements.
4.19.1 Principle
The measuring principle is based on the UV light absorption according to the Beer-Lambert law:
[C ] = k ⋅ log I 0
I1
The UV light is generated by a pulsed xenon lamp to reach an ultra long life time.
The light detection is performed by a photodiode detector through an interferential filter that
generally reuse to the reference light for the nitrate or UV254 measurement.
4.19.2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The concentration or error code corresponds to the
last measurement done, while the signal and signal graph are continuously updated.
7 1 6 8 9 10 11
2
18
12 25
19
13 22
20
14
21
15
16
17 23 24
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit.
During a measuring cycle, the last measured value is display in light green until replaced by the
new value in black.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or the cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
During a measuring cycle or a zeroing cycle, this area displays a counter of the elemental
acquisitions preceded by the message “measuring…” or “zeroing…” respectively.
8 Zero button
Put first pure water inside the flow cell before pressing on this button.
Note: an auto-zero is performed automatically during each acid cleaning if this option is selected
on the timing screen.
9 Scale button
To avoid a wrong calibration, check the zero or redo a zero before doing a
calibration.
This window displays the photo detector 1 (PH1) signal continuously. This signal corresponds to
the light absorbed by the nitrate.
On pure water, the signal must normally be in a 500 mV to 2000 mV range.
When the light signal added to the darkness signal reach 3000 mV, a message “OVL!” in red is
displayed to indicate a saturation of the acquisition system.
11 Signal graph
This graph displays the light signal absorbed by the turbidity continuously.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The average signal value in mV and its standard deviation (sigma) on the last 500 acquisitions
are displayed on the bottom right of the screen.
The values can be read by moving the cursor with the vertical arrows. After reading, restart the
acquisitions by pressing on the “play” button if necessary.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
12 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
13 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
15 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
16 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
17 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
18 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
Note: this factor must be set to 1.0 before doing a new linearization curve
19 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
20 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
21 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Beer-Lambert law used on most of the measurements is not linear for high concentrations
due to a light mask effect. Therefore it’s necessary to introduce a correction before displaying
the final concentration. This correction is based on the linearity curve represented on the top.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
25 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.19.3 Recalibration
- Take a turbidity standard (formazine for example) representative of the measuring range
- Do a manual measurement on this standard using the check screen (as describe above).
- Press on the “scale” button of the check screen and then enter the standard value on the
keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
4.20.1 Principle
The measuring principle is based on the light absorption to the Beer-Lambert law:
[C ] = k ⋅ log I 0
I1
The light is generated by a pulsed xenon lamp to reach an ultra long life time.
The absorption spectrum is performed by a concave grating coupled to a 2048 pixels CCD
(Charge-Coupled Device).
4.20.2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The concentration or error code corresponds to the
last measurement done, as well as the signal and absorbance graphs.
7 1 6 8 9 10 11
2
18
12 27
19
13 24
20
14
21
15
16
17 22 23 25 26
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit.
During a measuring cycle, the last measured value is display in light green until replaced by the
new value in black.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or the cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
During a measuring cycle or a zeroing cycle, this area displays a counter of the elemental
acquisitions preceded by the message “measuring…” or “zeroing…” respectively.
8 Zero button
Put first pure water inside the flow cell before pressing on this button.
Note: an auto-zero is performed automatically during each acid cleaning if this option is selected
on the timing screen.
9 Scale button
To avoid a wrong calibration, check the zero or redo a zero before doing a
calibration.
10 Signal graph
This graph displays the lamp spectrum get during the last measuring cycle. It is updated during
the measuring cycle when the lamp is flashing.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
With pure water inside the flow cell, the signal must not be saturated (below 2000 mV) on the
two wavelengths bands used for the absorbance calculation defined as “Peak wavelength” and
“Ref wavelength” on the check screen. Otherwise, an error will occur during the measuring
cycle.
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
11 Absorbance spectrum
This graph displays the absorbance spectrum get during the last measuring cycle. It is displayed
in light grey during a measuring cycle until updated.
For more details or to investigate about the absorbance spectrum, press anywhere on the graph
to zoom it as below:
The average absorbance for the peak wavelength band is represented by a green line and
displayed above the curve.
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
12 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
13 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
15 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
16 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
17 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
18 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
Note: this factor must be set to 1.0 before doing a new linearization curve
19 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
20 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
21 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
22 Peak wavelength
This field allows changing the peak wavelength taken in account for the measurement. It
corresponds to a wavelength band used to determine the turbidity of the sample.
The default value is 350 nm. It can be changed exceptionally to avoid a known interference.
23 Peak width
This field allows changing the width band of the peak wavelength taken in account for the
measurement.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Beer-Lambert law used on most of the measurements is not linear for high concentrations
due to a light mask effect. Therefore it’s necessary to introduce a correction before displaying
the final concentration. This correction is based on the linearity curve represented on the top.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
27 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.20.3 Recalibration
- Take a turbidity standard (formazine for example) representative of the measuring range
- Do a manual measurement on this standard using the check screen (as describe above).
- Press on the “scale” button of the check screen and then enter the standard value on the
keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
4.21.1 Principle
The light beam is generated by a pulsed laser diode while the light detection is performed by a
photodiode detector placed at 90° from the light beam as represented below:
4.21.2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The concentration or error code corresponds to the
last measurement done, while the signal and signal graph are continuously updated.
7 1 6 8 9 10 11
2
18
12 26
19
13 23
20
14
21
15
22
16
17 24 25
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit.
During a measuring cycle, the last measured value is display in light green until replaced by the
new value in black.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or the cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
During a measuring cycle or a zeroing cycle, this area displays a counter of the elemental
acquisitions preceded by the message “measuring…” or “zeroing…” respectively.
8 Zero button
Put first pure water inside the flow cell before pressing on this button.
Note: an auto-zero is performed automatically during each acid cleaning if this option is selected
on the timing screen.
9 Scale button
To avoid a wrong calibration, check the zero or redo a zero before doing a
calibration.
This window displays the photo detector signal continuously. This signal corresponds to the light
re-emitted by the turbidity of the sample from the laser diode.
On pure water, the signal must normally stay below 100 mV.
When the light signal added to the darkness signal reach 3000 mV, a message “OVL!” in red is
displayed to indicate a saturation of the acquisition system. It might be due to an over-ranged
sample.
11 Signal graph
This graph displays the light signal re-emitted by the turbidity of the sample from the laser diode.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The average signal value in mV and its standard deviation (sigma) on the last 500 acquisitions
are displayed on the bottom right of the screen.
The values can be read by moving the cursor with the vertical arrows. After reading, restart the
acquisitions by pressing on the “play” button if necessary.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
12 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
13 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
15 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
16 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
17 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
18 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
Note: this factor must be set to 1.0 before doing a new linearization curve
19 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
20 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
21 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
22 Offset
This field show the internal offset applied on the signal for the zeroing.
It is updated at each zeroing and has not to be changed.
It can be used for troubleshooting by the manufacturer.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Beer-Lambert law used on most of the measurements is not linear for high concentrations
due to a light mask effect. Therefore it’s necessary to introduce a correction before displaying
the final concentration. This correction is based on the linearity curve represented on the top.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
26 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.21.3 Recalibration
- Take a turbidity standard (formazine for example) representative of the measuring range
- Do a manual measurement on this standard using the check screen (as describe above).
- Press on the “scale” button of the check screen and then enter the standard value on the
keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
4.22.1 Principle
The measuring principle is based on the UV light absorption at 254 nm according to the Beer-
Lambert law:
[C ] = k ⋅ log I 0
I1
With [C ] : Sample concentration
k: Absorption coefficient (molecule specific)
I0: Light intensity before the sample
I1 : Light intensity after of the sample
The UV light is generated by a pulsed xenon lamp to reach an ultra long life time.
The light detection is performed by two photodiode detectors through interferential filters; one
centred on 254 nm while the other is used as reference to compensate the turbidity or deposit
inside the flow cell.
The UV light absorption at 254 nm is generally well correlated with the organic matter contents
on natural water (rivers, lakes) and municipal wastewater as well as wastewater on food and
paper industries.
After checking the correlation and after having introduced the right calibration factor, the
measurements can be displayed as COD measurement, or even BOD or TOC measurement.
4.22.2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The concentration or error code corresponds to the
last measurement done, while the signals are continuously updated.
7 1 6 8 9 10 11 12 13
2
21
14 29
22
15 26
23
16
24
17
18
19 20 25 27 28
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit.
During a measuring cycle, the last measured value is display in light green until replaced by the
new value in black.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or the cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
During a measuring cycle or a zeroing cycle, this area displays a counter of the elemental
acquisitions preceded by the message “measuring…” or “zeroing…” respectively.
8 Zero button
Put first pure water inside the flow cell before pressing on this button.
Note: an auto-zero is performed automatically during each acid cleaning if this option is selected
on the timing screen.
9 Scale button
To avoid a wrong calibration, check the zero or redo a zero before doing a
calibration.
This window displays the photo detector 1 (PH1) signal continuously. This signal corresponds to
the light absorbed by the organic matter.
When the light signal added to the darkness signal reach 3000 mV, a message “OVL!” in red is
displayed to indicate a saturation of the acquisition system.
This window displays the photo detector 2 (PH2) signal continuously. This signal corresponds to
the reference light that is not absorbed by the organic matter. It is use to compensate turbidity or
deposit on the flow cell.
When the light signal added to the darkness signal reach 3000 mV, a message “OVL!” in red is
displayed to indicate a saturation of the acquisition system.
This graph displays the PH1 signal continuously updated. This signal corresponds to the light
absorbed by organic matter.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
The average signal value in mV and its standard deviation (sigma) on the last 500 acquisitions
are displayed on the bottom right of the screen.
The values can be read by moving the cursor with the vertical arrows. After reading, restart the
acquisitions by pressing on the “play” button if necessary.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
This graph displays the photo detector no 2 (PH2) signal continuously updated. This signal
corresponds to the reference light through the sample.
For more details or to investigate about the signal, press anywhere on the graph to zoom it.
Refer to photo detector no 2 signal graph for the zoom display.
14 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
15 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
16 Display mode
This button allows selecting the concentration display mode as Abs/m (default mode) or as COD
if the correlation between UV absorbance and COD has been settled.
If COD is selected, the calibration factor is preset to 0.5 (typical value for river water), the
linearization curve is taken in account and the label is automatically updated to “COD”.
The measurements previously stored in the internal memory will keep their values in Abs/m.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
18 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
19 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
20 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
21 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
Note: this factor must be set to 1.0 before doing a new linearization curve
22 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
23 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
24 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
25 Offset
This field show the internal offset applied on the signal for the zeroing.
It is updated at each zeroing and has not to be changed.
It can be used for troubleshooting by the manufacturer.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Beer-Lambert law used on most of the measurements is not linear for high concentrations
due to a light mask effect. Therefore it’s necessary to introduce a correction before displaying
the final concentration. This correction is based on the linearity curve represented on the top.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
29 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.22.3 Recalibration
As measurements are often given in COD unit, a specific calibration directly on the sample must
be done. The default calibration factor of 0.5 applies only for river water. Municipal wastewater
or industrial wastewater may have a very different factor.
- Take 2 litres of the sample at an expected significant concentration according to the process
- Proceed to a COD laboratory measurement with one litre only of the sample (keep the second
litre on a cold place if the laboratory measurement is not done the same day)
- Use the second litre for the analyser and do at least two measurements on this sample using
the check screen. The last two measurements must be close together with less than 2% of
difference.
- Press on the “scale” button of the check screen and then enter the laboratory COD value on
the keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen. This new calibration factor will be
taken in account for all the further measurements.
4.23.1 Principle
The measuring principle is based on the UV light absorption at 254 nm according to the Beer-
Lambert law:
[C ] = k ⋅ log I 0
I1
The UV light is generated by a pulsed xenon lamp to reach an ultra long life time.
The absorption spectrum is performed by a concave grating coupled to a 2048 pixels CCD
(Charge-Coupled Device).
A double-wavelength algorithm allows compensating the turbidity or a deposit inside the flow
cell.
The UV light absorption at 254 nm is generally well correlated with the organic matter contents
on natural water (rivers, lakes) and municipal wastewater as well as wastewater on food and
paper industries.
After checking the correlation and after having introduced the right calibration factor, the
measurements can be displayed as COD measurement, or even BOD or TOC measurement.
4.23.2 Test
From the process screen, press on the “tools” icon and then on the channel to test, or directly to
the channel area as a shortcut.
If you are logged in user mode, the parameters are masked. You must be logged in
administrator mode to access to the parameters (default password is 456).
The check screen as below is displayed. The concentration or error code corresponds to the
last measurement done, as well as the signal and absorbance graphs.
7 1 6 8 9 10 11
2
19
12 30
20
13 27
21
14
22
15
16
17 18 23 24 25 26 28 29
1 Measurement window
This window displays the last measured value (or the error code in case of error, refer to
troubleshooting section) with proper label and unit.
During a measuring cycle, the last measured value is display in light green until replaced by the
new value in black.
2 Start button
Pressing on this button starts a manual measuring cycle. It can be aborted at any time by
pressing on the stop button. At the end of the cycle, a new value will be displayed on the
measurement window.
The process value on the process screen will not be updated by this measurement to avoid any
interference with the plant supervisor system, same concerning the MODBUS channel’s value,
4-20 mA output and alarm relays (if any).
3 Stop button
Pressing on this button stop and aborts any engaged measuring cycle. It can also be used to
stop the sampling pump or the cleaning pump.
Pressing on this button starts the sampling pump. The pump will stop automatically after 10
seconds. It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the cleaning pump. The pump will stop automatically after the
time defined in the acid “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
Pressing on this button starts the ultrasonic cleaning. The cleaning will stop automatically after
the time defined in the ultrasonic “cleaning time” field of the timing screen.
It can also be stopped at any time by pressing on the stop button.
During a measuring cycle or a zeroing cycle, this area displays a counter of the elemental
acquisitions preceded by the message “measuring…” or “zeroing…” respectively.
8 Zero button
Put first pure water inside the flow cell before pressing on this button.
Note: an auto-zero is performed automatically during each acid cleaning if this option is selected
on the timing screen.
9 Scale button
To avoid a wrong calibration, check the zero or redo a zero before doing a
calibration.
10 Signal graph
This graph displays the lamp spectrum get during the last measuring cycle. It is updated during
the measuring cycle when the lamp is flashing.
For more details or to investigate about the signal, press anywhere on the graph to zoom it as
below:
With pure water inside the flow cell, the signal must not be saturated (below 2000 mV) on the
two wavelengths bands used for the absorbance calculation defined as “Peak wavelength” and
“Ref wavelength” on the check screen. Otherwise, an error will occur during the measuring
cycle.
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
11 Absorbance spectrum
This graph displays the absorbance spectrum get during the last measuring cycle. It is displayed
in light grey during a measuring cycle until updated.
For more details or to investigate about the absorbance spectrum, press anywhere on the graph
to zoom it as below:
The average absorbance for the peak wavelength band and for the reference wavelength band
are represented by a green line and displayed above the curve.
The values can be read by moving the cursor with the vertical arrows.
If necessary, use the horizontal arrows to displace the graph or press on the graph to zoom it.
The Y scale is automatic but may be override by the small arrows placed above the Y axis.
12 Label
This field allows changing the channel label displayed with the measurement. The label size is
limited to 6 characters.
13 Unit
This field allows changing the channel unit displayed with the measurement. The label size is
limited to 6 characters.
14 Display mode
This button allows selecting the concentration display mode as Abs/m (default mode) or as COD
if the correlation between UV absorbance and COD has been settled.
If COD is selected, the calibration factor is preset to 0.5 (typical value for river water), the
linearization curve is taken in account and the label is automatically updated to “COD”.
The measurements previously stored in the internal memory will keep their values in Abs/m.
Negatives values are possible for most of the measurements as they result of normal
fluctuations if the measurement is close to zero.
By default, the display of negative values is allowed as negative values may also result from a
wrong zero and consequently it may be useful to be aware of such problem when seeing
strongly negative values.
But as negative values have no theoretical meaning, it’s possible to replace them by zero by
deselecting this function.
16 Number of decimals
This field allows changing the number of decimal displayed for the measurement value to
remove not significant decimals.
Note: this choice does not affect the measurements stored in the internal memory.
17 Number of averaging
The measurement as displayed on the process screen (or read under MODBUS protocol or
available on the 4-20 mA output) can be formed by the average over a number of basic
measurements defined by the field.
It can be useful to improve the repeatability of the measurements if the response time is not
critical.
18 4-20 mA output
This button indicates if a 4-20 mA output has been installed or not for this channel (option).
If a module has been installed, its position is displayed on the right. Pressing on this button
shows the popup below that enable to test the 4-20 mA output by pressing on the “test” button.
The current value of the 4-20 mA output is displayed on the top, based on the last process
measurement (but not on the test measurement to avoid disturbances on the process
supervisor system).
The “scan” key allows scanning the 4-20 mA output mA by mA for testing purpose only.
The scale value for 20 mA and 4 mA are displayed below. These preset values can be changed
by pressing on these fields.
If the output value doesn’t mach exactly with the theoretical value, it’s possible to calibrate the
output by pressing on the “calibration” key. Then real values for 4 and 20 mA must be entered.
Note: position M12 is not available for the 4-20 mA output modules.
19 Calibration factor
This factor is normally changed automatically while doing a calibration by pressing on the
“scale” button and then recorded in the calibration history. This multiplying factor is taken in
account as the last step before displaying the measurement value.
Note: this factor must be set to 1.0 before doing a new linearization curve
20 Scale factor
This factor is normally determined in factory to transform the basic measurement to a scaled
value in the final unit, mg/l for example.
It is has no reason to be changed except if internal elements are replaced.
21 Number of readings
This field allows changing the number of basic readings averaged to obtain one final
measurement.
Increasing this value improve the repeatability but increase also the measuring time.
Depending on the application, it can be exceptionally increased or decreased.
22 Minimum signal
This field fix the minimum light signal value over the wavelength range taken in account for the
measurement. An error will be displayed if the signal over the wavelength range is lower than
this value.
23 Peak wavelength
This field allows changing the peak wavelength taken in account for the measurement. It
corresponds to a wavelength absorbed by organic matter.
The default value is 254 nm. It can be changed exceptionally to avoid a known interference or to
improve the COD correlation if a study has been carry on previously.
24 Peak width
This field allows changing the width band of the peak wavelength taken in account for the
measurement.
25 Reference wavelength
This field allows changing the reference wavelength taken in account for the measurement.
It corresponds normally to a wavelength where the organic matter doesn’t absorb.
The default value is 350 nm. It can be changed exceptionally to avoid a known interference or to
improve the COD correlation if a study has been carry on previously.
26 Reference width
This field allows changing the width band of the reference wavelength taken in account for the
measurement.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This window displays the 10 last calibrations done with the date, time and new calibration factor.
This area concerns the linearity compensation applied on the rough measurements.
The Beer-Lambert law used on most of the measurements is not linear for high concentrations
due to a light mask effect. Therefore it’s necessary to introduce a correction before displaying
the final concentration. This correction is based on the linearity curve represented on the top.
The Y-axis corresponds to the rough measurements entered on the M1 to M10 fields while the
X-axis corresponds to the standard or final measurement, entered on the S1 to S10 fields.
The M1-M10 and S1-S10 values must be strictly increasing. It’s highly recommended to put 0.0
and 0.0 as starting values for S1 and M1. The final value is linearly extrapolated between these
points.
27 Linearity ON/OFF
This button allows deactivating temporary the linearity curve (OFF), for example to redo a new
linearity curve.
The linearity curve is automatically reactivated (ON) when leaving the check screen.
4.23.3 Recalibration
As measurements are often given in COD unit, a specific calibration directly on the sample must
be done. The default calibration factor of 0.5 applies only for river water. Municipal wastewater
or industrial wastewater may have a very different factor.
- Take 2 litres of the sample at an expected significant concentration according to the process
- Proceed to a COD laboratory measurement with one litre only of the sample (keep the second
litre on a cold place if the laboratory measurement is not done the same day)
- Use the second litre for the analyser and do at least two measurements on this sample using
the check screen. The last two measurements must be close together with less than 2% of
difference.
- Press on the “scale” button of the check screen and then enter the laboratory COD value on
the keypad.
The calibration is finished. A new calibration factor has been calculated and has been recorded
inside the calibration history displayed on the check screen.
This new calibration factor will be taken in account for all the further measurements.
The maintenance is limited to the replacement of the peristaltic pump tubing and to the refilling
of the cleaning or reagent solution, depending on the configuration.
The period of replacement of the pump tubing depends on the use. A weekly checking is
recommended. The typical lifetime with Pharmed® tubing is estimated to about 1000 hours of
continuous operation.
General troubleshooting:
The screen remains totally black after - Check the power socket
connecting the power cord.
- Check J1 connector (mains input, high voltage!)
AND
- Check J2 connector on the DSP500 (mains input
The red LED D1 on the DSP500 board is for the power supply, high voltage!)
OFF.
- Check J3 connector (24V output from the power
supply)
The screen is lighted but nothing is - Check the screen connector J8 on the rear of the
displayed. DSP500 board
5.2 Ammonia
Maintenance
The maintenance is limited to refilling of the 10% NaOH solution and the 5% H2SO4 cleaning
solution.
Troubleshooting
Symptoms Origin
value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
- Condensation inside the gas flow cell (no trace of water must be
visible on the gas circuit)
Unstable value - Deposit or dirty on the optical parts (lens and/or windows), check the
light level at 210 nm (zoom on the signal graph of the check screen)
Measuring errors
1 The temperature of the - Wait about 15 minutes after power on. The temperature is
heated parts is lower displayed on the bottom line of the screen
than 5 °C of the set
point - Check the temperature probe connection on J19 “PT100-3”
of the DSP500 board (no connection gives a temperature of
about -273 °C)
2 No signal from the - Check the spectrograph connection on J20 of the DSP500
spectrograph board
3 The light level on the - Check the light level (zoom on the signal graph of the check
spectrograph is too screen), the light level can be slightly adjusted by moving the
high on the lamp holder or the flow cell
wavelengths used for
the measurement - Failure on the CCD500 board (replace)
4 The light level on the - Water inside the gas flow cell (dismount and dry all the gas
spectrograph is too low circuit, check that there is no clogging on the outlet)
on the wavelengths
used for the - Failure on the xenon lamp circuit if no flashes are visible
measurement during the measurement, check the connection of the
XENON500 board on J26 “XENON1” of the DSP500 board
(the orange neon lamp on the XENON500 board must always
be on, if not replace the XENON500 board) and check the
lamp connection on the XENON500 board (J2)
5.3 Chlorophyll A
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure).
- No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
1 No signal from the - Check the photomultiplier connection on one of the position
photomultiplier J30 to J34 (“LED1” to “LED5”) of the DSP500 board according
to the position displayed on the factory screen
2 No signal from the - Check the photo detector connection on one of the position
photo detector J21 to J25 (“IN1” to “IN5”) of the DSP500 board according to
the position displayed on the factory screen
3 The light level - Sample concentration over range, check again with pure
coming from the water
photomultiplier is too
high - Check that there is no bubble inside the flow cell, if yes check
the fittings
4 The light level - Check that there is no bubble inside the flow cell, if yes check
coming from the the fittings
reference photo
detector is too high - Failure on the photo detector cable (repair or replace)
5 Reserved
6 The light level - Sample concentration over range, check again with pure
coming from the water
reference photo
detector is too low - Sample turbidity too high, check again with pure water
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure).
- No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
1 No signal from the - Check the photo detector connection on one of the position
photo detector J21 to J25 (“IN1” to “IN5”) of the DSP500 board according to
the position defined on the check screen in the position box
2 Reserved
3 The light level - Check that there is no bubble inside the flow cell, if yes check
coming from the peak the fittings
LED is too high
- Failure on the photo detector cable (repair or replace)
4 The light level - Check that there is no bubble inside the flow cell, if yes check
coming from the the fittings
reference LED is too
high - Failure on the photo detector cable (repair or replace)
5 The light level - Sample concentration over range, check again with pure
coming from the peak water
LED is too low
- Sample turbidity too high, check again with pure water
6 The light level - Sample concentration over range, check again with pure
coming from the water
reference LED is too
low - Sample turbidity too high, check again with pure water
7 The temperature - Failure on the peak LED board cable (repair or replace)
compensation signal
of the peak LED is - Failure on the peak LED board (replace)
invalid
8 The temperature - Failure on the reference LED board cable (repair or replace)
compensation signal
of the reference LED - Failure on the reference LED board (replace)
is invalid
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure).
- No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
1 reserved
2 No signal from the - Check the spectrograph connection on the connector J20 of
spectrograph the DSP500 board
3 The light level on the - Check that there is no bubble inside the flow cell, if yes check
used bands of the fittings
wavelengths is too
high - Failure on the spectrograph cable (repair or replace)
4 The light level on the - Sample concentration over range, check again with pure
used bands of water
wavelengths is too
low - Sample turbidity too high, check again with pure water
5.6 Conductivity
Maintenance
The conductivity probe must be cleaned on a regular base (at least weekly) depending of the
application.
Troubleshooting
Symptoms Origin
Unstable value - Electrical interference on the probe cable, put the probe cable away
from power cables
Measuring errors
Maintenance
The dissolved oxygen probe must be cleaned on a regular base (at least weekly) depending of
the application.
Troubleshooting
Symptoms Origin
Measuring errors
Reserved
5.8 Hydrocarbons
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure).
- No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
1 No signal from the - Check the photomultiplier connection on one of the position
photomultiplier J30 to J34 (“LED1” to “LED5”) of the DSP500 board according
to the position displayed on the factory screen
2 No signal from the - Check the photo detector connection on one of the position
photo detector J21 to J25 (“IN1” to “IN5”) of the DSP500 board according to
the position displayed on the factory screen
3 The light level - Sample concentration over range, check again with pure
coming from the water
photomultiplier is too
high - Check that there is no bubble inside the flow cell, if yes check
the fittings
4 The light level - Check that there is no bubble inside the flow cell, if yes check
coming from the the fittings
reference photo
detector is too high - Failure on the photo detector cable (repair or replace)
5 Reserved
6 The light level - Sample concentration over range, check again with pure
coming from the water
reference photo
detector is too low - Sample turbidity too high, check again with pure water
Maintenance
The maintenance is limited to refilling of the 10% HCl solution and the 5% H2SO4 cleaning
solution.
Troubleshooting
Symptoms Origin
value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
- Condensation inside the gas flow cell (no trace of water must be
visible on the gas circuit)
Unstable value - Deposit or dirty on the optical parts (lens and/or windows), check the
light level at 210 nm (zoom on the signal graph of the check screen)
Measuring errors
1 The temperature of the - Wait about 15 minutes after power on. The temperature is
heated parts is lower displayed on the bottom line of the screen
than 5 °C of the set
point - Check the temperature probe connection on J19 “PT100-3”
of the DSP500 board (no connection gives a temperature of
about -273 °C)
2 No signal from the - Check the spectrograph connection on J20 of the DSP500
spectrograph board
3 The light level on the - Check the light level (zoom on the signal graph of the check
spectrograph is too screen), the light level can be slightly adjusted by moving the
high on the lamp holder or the flow cell
wavelengths used for
the measurement - Failure on the CCD500 board (replace)
4 The light level on the - Water inside the gas flow cell (dismount and dry all the gas
spectrograph is too low circuit, check that there is no clogging on the outlet)
on the wavelengths
used for the - Failure on the xenon lamp circuit if no flashes are visible
measurement during the measurement, check the connection of the
XENON500 board on J26 “XENON1” of the DSP500 board
(the orange neon lamp on the XENON500 board must always
be on, if not replace the XENON500 board) and check the
lamp connection on the XENON500 board (J2)
Maintenance
No maintenance is required.
Troubleshooting
Symptoms Origin
Measuring errors
Reserved
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure).
- No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
1 No signal from the - Check the photo detector connection on one of the position
peak photo detector J21 to J25 (“IN1” to “IN5”) of the DSP500 board according to
the position defined on the check screen in the position box
2 No signal from the - Check the photo detector connection on one of the position
reference photo J21 to J25 (“IN1” to “IN5”) of the DSP500 board according to
detector the position defined on the check screen in the position box
3 The light level - Check that there is no bubble inside the flow cell, if yes check
coming from the peak the fittings
photo detector is too
high - Failure on the peak photo detector cable (repair or replace)
4 The light level - Check that there is no bubble inside the flow cell, if yes check
coming from the the fittings
reference photo
detector is too high - Failure on the reference photo detector cable (repair or
replace)
5 The light level - Sample concentration over range, check again with pure
coming from the peak water
photo detector is too
low - Sample turbidity too high, check again with pure water
6 The light level - Sample concentration over range, check again with pure
coming from the water
reference photo - Sample turbidity too high, check again with pure water
detector is too low
- Failure on the reference photo detector board cable (repair or
replace)
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure).
- No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
1 reserved
2 No signal from the - Check the spectrograph connection on the connector J20 of
spectrograph the DSP500 board
3 The light level on the - Check that there is no bubble inside the flow cell, if yes check
used bands of the fittings
wavelengths is too
high - Failure on the spectrograph cable (repair or replace)
4 The light level on the - Sample concentration over range, check again with pure
used bands of water
wavelengths is too
low - Sample turbidity too high, check again with pure water
Maintenance
The ORP probe must be cleaned on a regular base (at least weekly) depending of the
application.
Troubleshooting
Symptoms Origin
Unstable value - Electrical interference on the probe cable, put the probe cable away
from power cables
Measuring errors
5.14 pH
Maintenance
The pH probe must be cleaned on a regular base (at least weekly) depending of the application.
Troubleshooting
Symptoms Origin
- pH probe disconnected
Unstable value - Electrical interference on the probe cable, put the probe cable away
from power cables
Measuring errors
Maintenance
The maintenance is limited to refilling of the one or two reagent bottles and the 5% H2SO4
cleaning solution.
Troubleshooting
Symptoms Origin
Negative value - Very high turbidity on the sample that is not totally compensated
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
1 No signal from the - Check the photo detector connection on one of the position
photo detector J21 to J25 (“IN1” to “IN5”) of the DSP500 board according to
the position defined on the check screen in the position box
2 Reserved
3 The light level - Check that there is no bubble inside the flow cell, if yes check
coming from the peak the fittings
LED is too high
- Failure on the photo detector cable (repair or replace)
4 The light level - Check that there is no bubble inside the flow cell, if yes check
coming from the the fittings
reference LED is too
high - Failure on the photo detector cable (repair or replace)
5 The light level - Sample concentration over range, check again with pure
coming from the peak water
LED is too low
- Sample turbidity too high, check again with pure water
6 The light level - Sample concentration over range, check again with pure
coming from the water
reference LED is too
low - Sample turbidity too high, check again with pure water
7 The temperature - Failure on the peak LED board cable (repair or replace)
compensation signal
of the peak LED is - Failure on the peak LED board (replace)
invalid
8 The temperature - Failure on the reference LED board cable (repair or replace)
compensation signal
of the reference LED - Failure on the reference LED board (replace)
is invalid
Maintenance
The maintenance is limited to refilling of the one or two reagent bottles and the 5% H2SO4
cleaning solution.
Troubleshooting
Symptoms Origin
Negative value - Very high turbidity on the sample that is not totally compensated
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
1 reserved
2 No signal from the - Check the spectrograph connection on the connector J20 of
spectrograph the DSP500 board
3 The light level on the - Check that there is no bubble inside the flow cell, if yes check
used bands of the fittings
wavelengths is too
high - Failure on the spectrograph cable (repair or replace)
4 The light level on the - Sample concentration over range, check again with pure
used bands of water
wavelengths is too
low - Sample turbidity too high, check again with pure water
5.17 Rhodamine
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure).
- No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
1 No signal from the - Check the photomultiplier connection on one of the position
photomultiplier J30 to J34 (“LED1” to “LED5”) of the DSP500 board according
to the position displayed on the factory screen
2 No signal from the - Check the photo detector connection on one of the position
photo detector J21 to J25 (“IN1” to “IN5”) of the DSP500 board according to
the position displayed on the factory screen
3 The light level - Sample concentration over range, check again with pure
coming from the water
photomultiplier is too
high - Check that there is no bubble inside the flow cell, if yes check
the fittings
4 The light level - Check that there is no bubble inside the flow cell, if yes check
coming from the the fittings
reference photo
detector is too high - Failure on the photo detector cable (repair or replace)
5 Reserved
6 The light level - Sample concentration over range, check again with pure
coming from the water
reference photo
detector is too low - Sample turbidity too high, check again with pure water
Maintenance
Measuring errors
1 reserved
2 No signal from the - Check the spectrograph connection on the connector J20 of
spectrograph the DSP500 board
3 The light level on - Check that there is no bubble inside the flow cell, if yes check
some wavelengths is the fittings
too high
- Failure on the spectrograph cable (repair or replace)
4 The light level on - Sample concentration over range, check again with pure
some wavelengths is water
too low
- Sample turbidity too high, check again with pure water
5.19 Temperature
Maintenance
No maintenance is required.
Troubleshooting
Symptoms Origin
Measuring errors
2 Probe short-circuited - Check the probe with an ohmmeter, the value must be in a
100 to 120 ohm range, if not replace the probe
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure).
- No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
1 No signal from the - Check the photo detector connection on one of the position
peak photo detector J21 to J25 (“IN1” to “IN5”) of the DSP500 board according to
the position defined on the check screen in the position box
2 No signal from the - Check the photo detector connection on one of the position
reference photo J21 to J25 (“IN1” to “IN5”) of the DSP500 board according to
detector the position defined on the check screen in the position box
3 The light level - Check that there is no bubble inside the flow cell, if yes check
coming from the peak the fittings
photo detector is too
high - Failure on the peak photo detector cable (repair or replace)
4 The light level - Check that there is no bubble inside the flow cell, if yes check
coming from the the fittings
reference photo
detector is too high - Failure on the reference photo detector cable (repair or
replace)
5 The light level - Sample concentration over range, check again with pure
coming from the peak water
photo detector is too
low - Sample turbidity too high, check again with pure water
6 The light level - Sample concentration over range, check again with pure
coming from the water
reference photo - Sample turbidity too high, check again with pure water
detector is too low
- Failure on the reference photo detector board cable (repair or
replace)
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure).
- No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
1 reserved
2 No signal from the - Check the spectrograph connection on the connector J20 of
spectrograph the DSP500 board
3 The light level on the - Check that there is no bubble inside the flow cell, if yes check
used bands of the fittings
wavelengths is too
high - Failure on the spectrograph cable (repair or replace)
4 The light level on the - Sample concentration over range, check again with pure
used bands of water
wavelengths is too
low - Sample turbidity too high, check again with pure water
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Value is too high - Bubbles in the sample (check peristaltic pump inlet fitting).
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
1 No signal from the - Check the photo detector connection on one of the position
peak photo detector J21 to J25 (“IN1” to “IN5”) of the DSP500 board according to
the position displayed on the factory screen
2 The light level - Check that there is no bubble inside the flow cell, if yes check
coming from the peak the fittings
photo detector is too
high - Failure on the photo detector cable (repair or replace)
3 No reference light - Check the LED connection on one of the position J30 to J34
from the laser diode (“LED1” to “LED5”) of the DSP500 board according to the
board position displayed on the factory screen
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure).
- No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
1 No signal from the - Check the photo detector connection on one of the position
peak photo detector J21 to J25 (“IN1” to “IN5”) of the DSP500 board according to
the position defined on the check screen in the position box
2 No signal from the - Check the photo detector connection on one of the position
reference photo J21 to J25 (“IN1” to “IN5”) of the DSP500 board according to
detector the position defined on the check screen in the position box
3 The light level - Check that there is no bubble inside the flow cell, if yes check
coming from the peak the fittings
photo detector is too
high - Failure on the peak photo detector cable (repair or replace)
4 The light level - Check that there is no bubble inside the flow cell, if yes check
coming from the the fittings
reference photo
detector is too high - Failure on the reference photo detector cable (repair or
replace)
5 The light level - Sample concentration over range, check again with pure
coming from the peak water
photo detector is too
low - Sample turbidity too high, check again with pure water
6 The light level - Sample concentration over range, check again with pure
coming from the water
reference photo - Sample turbidity too high, check again with pure water
detector is too low
- Failure on the reference photo detector board cable (repair or
replace)
Maintenance
Troubleshooting
Symptoms Origin
Negative value - Bad zero or auto-zero: the cleaning solution has been prepared with
polluted water or fault in the acid cleaning system (no flow, pump
failure).
- No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Value is too low - No water in the flow cell (clogging, bad inlet connection, fault in the
peristaltic pump)
Unstable value - Deposit or dirty on the flow cell, check the light level on the check
screen
Measuring errors
1 reserved
2 No signal from the - Check the spectrograph connection on the connector J20 of
spectrograph the DSP500 board
3 The light level on the - Check that there is no bubble inside the flow cell, if yes check
used bands of the fittings
wavelengths is too
high - Failure on the spectrograph cable (repair or replace)
4 The light level on the - Sample concentration over range, check again with pure
used bands of water
wavelengths is too
low - Sample turbidity too high, check again with pure water
6 Specifications
6.1 General
Display: Colour TFT LCD, Size: 10.4”, resolution: 640 x 480 pixels
LED backlight with screen saver
Resistive touch screen
Memory: 4000 records (up to 16 measurement channels) with date and time
USB port For USB memory keys, any format (FAT16, FAT32)
Standard USB connector type A with IP68 protective cap
- Recorded measurement downloads (compatible with Excel®)
- Complete configuration backup/restore (proprietary format)
- Screen copy in BMP format (compatible with Windows®)
Outputs: Active 4-20 mA (optional), load 500 ohm maxi, resolution 0.005 mA,
Individual galvanic isolation
Connection on removable screw terminals
Operating limits: 0 to 50 ºC
6.2 Ammonia
Specifications
Accuracy: +/- 5%
Repeatability as sigma
- At 10 mg/l:
- At 0.5 mg/l:
- Around zero:
Interferences
6.3 Chlorophyll A
Specifications
Calibrated range:
Measuring range:
Accuracy: +/- 5%
Repeatability as sigma
Interferences
Specifications
Flow cell 30 mm 10 mm 1 mm
Scale factor
Calibration factor
Calibrated range
Measuring range
Accuracy
Measuring time
Repeatability (sigma): (1,2)
At full calibrated range
At 50% of calibrated range
At 10% of calibrated range
Around zero
Interferences
No interference
Specifications
Flow cell 30 mm 10 mm 1 mm
Scale factor
Calibration factor
Calibrated range
Measuring range
Accuracy
Measuring time
Repeatability (sigma): (1,2)
At full calibrated range
At 50% of calibrated range
At 10% of calibrated range
Around zero
Interferences
No interference
6.6 Conductivity
Specifications
Accuracy: +/- 5%
Repeatability as sigma
Interferences
No known interferences
Specifications
Accuracy: +/- 5%
Repeatability as sigma
Interferences
No known interferences
6.8 Hydrocarbons
Specifications
Accuracy: +/- 5%
Repeatability as sigma
Interferences
Specifications
Calibrated range:
Measuring range:
Accuracy: +/- 5%
Repeatability as sigma
Interferences
Specifications
Measuring range: 0 – 20 mA
Accuracy: +/- 5%
Repeatability as sigma
Specifications
Accuracy: +/- 5%
Repeatability as sigma
Interferences
The measurement is the sum of nitrite (NO2) and nitrate (NO3) ions.
Specifications
Accuracy: +/- 5%
Repeatability as sigma
Interferences
The measurement is the sum of nitrite (NO2) and nitrate (NO3) ions.
Specifications
Accuracy: +/- 5%
Repeatability as sigma
- At -2000 mV:
- At zero:
- At +2000 mV:
Interferences
6.14 pH
Specifications
Measuring range: 0 to 14
Accuracy: +/- 5%
Repeatability as sigma
- At pH 7:
Interferences
Measuring range:
Accuracy: +/- 5%
Repeatability as sigma
Interferences
Measuring range:
Accuracy: +/- 5%
Repeatability as sigma
Interferences
6.17 Rhodamine
Specifications
Calibrated range:
Measuring range:
Accuracy: +/- 5%
Repeatability as sigma
Interferences
Specifications
Resolution: 0.3 nm
6.19 Temperature
Specifications
Calibrated range:
Accuracy: +/- 1 °C
Specifications
Flow cell
Scale factor
Calibration factor
Calibrated range
Measuring range
Accuracy
Measuring time
Repeatability (sigma): (2,3)
At full calibrated range
At 50% of calibrated range
At 10% of calibrated range
Around zero
Interferences
Specifications
Flow cell
Scale factor
Calibration factor
Calibrated range
Measuring range
Accuracy
Measuring time
Repeatability (sigma): (2,3)
At full calibrated range
At 50% of calibrated range
At 10% of calibrated range
Around zero
Interferences
Specifications
Calibrated range:
Measuring range:
Accuracy: +/- 5%
Repeatability as sigma
Interferences
Unknown.
Specifications
Flow cell
Mode Abs/m COD river Abs/m COD river Abs/m COD ww(1)
Scale factor
Calibration factor 1.0 0.5 1.0 0.5 1.0 8.0
Calibrated range
Measuring range
Accuracy
Measuring time
Repeatability (sigma): (2,3)
At full calibrated range
At 50% of calibrated range
At 10% of calibrated range
Around zero
Interferences
1) Abs/m mode
Possible interference with high concentrations of metallic ions when measuring organic matter
2) COD mode
Each molecule has a specific UV/COD ratio. Display as COD can be done only on solutions
having a relatively constant molecules ratio like river water, municipal waste water, food and
paper industries.
It is recommended to carry out a correlation study before using the analyser in COD mode.
Specifications
Flow cell
Mode Abs/m COD river Abs/m COD river Abs/m COD ww(1)
Scale factor
Calibration factor 1.0 0.5 1.0 0.5 1.0 8.0
Calibrated range
Measuring range
Accuracy
Measuring time
Repeatability (sigma): (2,3)
At full calibrated range
At 50% of calibrated range
At 10% of calibrated range
Around zero
Interferences
1) Abs/m mode
Possible interference with high concentrations of metallic ions when measuring organic matter
2) COD mode
Each molecule has a specific UV/COD ratio. Display as COD can be done only on solutions
having a relatively constant molecules ratio like river water, municipal waste water, food and
paper industries.
It is recommended to carry out a correlation study before using the analyser in COD mode.
---------