Академический Документы
Профессиональный Документы
Культура Документы
lnter1aboratury Coopera · e S dy of
e rec1s on
of the Measu ement
- of ~~
PARTICULATE MATTER
IN THE ATMOSPHERE
(Optical Density of Filtered Deposit)
•
using
ASTM Method 01704
-
0
DS 55-83
Cf1
0
O'I
U'I
§
......
I
""
-
Gfiltt AMERICAN SOCIETY FOR TESTING AND MATERIALS
FINAL REPORT
on
NOTE
INTRODUCTION . . . 1
SUMMARY OF RESULTS 1
EXPERIMENTAL PROGRAM . 2
Reproducibility . . . 29
Repeatability . . . . 30
Analysis of Variance . . 30
Components of Variance. 31
Analysis of Tests • . • 43
Relationship of Standard Deviation to Mean. 43
Analysis of Duplicate Measurements. 43
RECOMMENDATIONS . . 51
ACKNOWLEDGEMENTS 52
REFERENCES 54
APPENDIX
i
LIST OF FIGURES
Figure
LIST OF TABLES
Table
ii
LIST OF TABLES
(Continued)
Table
iii
DS55S3-EB/Apr. 1974
by
INTRODUCTION
SUMMARY OF RESULTS
where, sb' and, m, are expressed in Cohs per 1000 linear feet.
• The standard deviation of 86 duplicate measurements of the
optical density of the same spot by one laboratory at one
site (Manhattan), is 0.55 Cohs per 1000 linear feet.
EXPERIMENTAL PROGRAM
Instrumentation
Calibration of Instruments
The tape samplers were placed at sill level next to a row of windows
on the north side of the sixth floor of Cooper Union, 51 Astor Place, New York
City. The probes extended through the windows and hung downward a short distance
at the building wall which bordered Ninth Street. Sampling was performed
during the period of January 9 to January 15, 1972.
The length of each test period was selected with the following
objectives:
5
Sampling
Test Duration, SamEling Date-Time(a)
Number hr (a) Start Stop
In analysis of the Site 1 data, the 8 and 12-hour tests and the 23 and
24-hour tests are designated short and long duration tests, respectively.
In performing the tests, it was not possible to synchronize all sampling
periods exactly. However, slight variations of a half hour or so in the length
of simultaneous sampling periods should be immaterial since the sampling periods
were long (8-24 hours) and comparable data, Cohs per 1000 linear feet, are
obtained even though sampling duration and volumes are different.
Table 2 gives the test schedule at Bloomington, where particulate
loading was so low that 24-hour samples were necessary to achieve measurable
optical densities of the filtered deposits. The same considerations as to the
exact synchronization of tape changes applied at Bloomington as at Los Angeles.
Sampling
Test Duratto~, SamEling Date-Time(a)
Number hr a Start Stop
Sampling
Test Du~~ttg~, Sam:eling Date -Time(a)
Number Start Stop
the long periods were two hours each, because this was the maximum change
period setting on several timers. An approximately equal number of short and
long test periods were specified. Short periods were varied to include
sequences of 30 minutes, 40 minutes, or 60 minutes each in attempting to match,
optical density of the collected sample with anticipated atmospheric variations
at the test site. On a few occasions there were instrument or timer malfunc-
tions, so that synchronization was incomplete. The data are therefore separated
into 44 approximately isochronous groups for statistical analysis as described
in detail in following sections.
11
Participating Laboratories
Sampling Sample
Test Sampling Time, Volume, Coh/1000
Laboratory Number Duration hr cu ft Linear Feet
TABLE 4. (Continued)
Sampling Sample
Test Sampling Time, Volume, Coh/1000
Laboratory Number Duration hr cu ft Linear Feet
£b)
E 1 L 16.9 (a) 0.5!+8(c)0.558 (c)
2 s 1.1 ditto 1.oqg 1.oqq
3 s 1. & " 0.555 0. 527
4 s 8. I+
II
0.665 0.620
5 L 2 3. 8 II
0.112 0.673
6 s 10.4 II
1.211 1.339
7 s 13.7 " 0.41+5 (i.381
~ L 23.5 " o.5q9 o.sso
Sampling Sample
Test Time, Volume, Coh/1000
Laboratory Number hr cu ft Linear Feet
Sample
Test Start Sam2 ling Time Volume, Coh/1000
Laboratory Number Day Time hr min min cu ft Linear Feet
TABLE 6. (Continued)
Sample
Test Start SamEling Time Volume, Coh/1000
Laboratory Number Day Time hr min min cu ft Linear Feet
TABLE 6. (Continued)
Sample
Test Start SamElinB Time Volume, Coh/1000
Laboratory Number Day Time hr min min cu ft Linear Feet
R i 9 i720 ilt ltO 880 59.0 i.598
8 2 ill 0~00 2 30 i50 iO.i 5.54i
TABLE 6. (Continued)
Sample
Test Start SamE ling Time Volume, Coh/1000
Laboratory Number Day Time hr min min cu ft Linear Feet
TABLE 6. (Contini.Jed)
Sample
Test Start Sam:eling Time Volume, Coh/1000
Laboratory Number Day Time hr min min cu ft Linear Feet
c 1 9 1700 15 00 qoo • 1.15
c 2 1C 0800 2 30 150 • 3.88
c 3 10 1100 0 40 40 • 6.33
c 4 10 111t0 0 r.o 40 • 6.02
c 5 1(\ 1220 0 1+0 40 • 5.6Q
c 6 10 1300 0 40 40 • 6.56
c 7 10 1340 0 40 40 • 5. 41
c 8 10 1420 0 40 40 • 7.65
c q 10 1sao 0 40 40 • 7.40
c 10 10 1540 0 40 40 • 7.95
c 11 10 1620 0 40 40 • 7.04
c 20 11 0900 0 30 30 • 7.93
c 21 11 oq30 0 30 30 • 5.88
c 22 11 10 (i 0 0 30 30 • 5.19
c 23 11 1030 0 30 30 • 6.33
c 24 11 1100 0 40 40 • 5.32
c 25 11 1140 0 40 40 • 4.08
c 26 11 1220 c ltO 40 • 2.75
c 27 11 1300 0 '•0 40 • 4.90
c 28 11 13Ct0 0 40 40 • 3.27
c 29 11 1420 0 40 40 • 3.23
c 30 11 1500 0 ltO 40 • 2.30
c 31 11 1540 1 20 80 • 2.2s
c 40 12 oqoo 1 co 60 • c...18
c 41 12 1000 0 40 40 • 3.53
c 42 12 1040 0 ltO 40 • 2.38
c 43 12 1120 0 40 40 • 2.98
c 44 12 1200 G 40 40 • 2. 78
c lt5 12 121t0 1 00 60 • 2.39
c 46 12 13Ct0 1 00 60 • 2.64
20
TABLE 6. (Continued)
Sample
Test Start SamEling Time Volume, Coh/1000
Laboratory Number Day Time hr min min cu ft Linear Feet
Sample
Test Start SamEling Time Volume, Coh/1000
Laboratory Number Day Time hr min min cu ft Linear Feet
1 g 1700 15 00 900
0 • 1. 01
0 2 10 0800 2 30 150 • 3.36
0 3 10 1100 0 I+ 0 40 • 5.23
0 I+ 1C 111+0 0 40 l+O • 3.20
0 5 10 1220 0 40 40 • 2.61+
0 6 10 1300 0 40 40 • 3.20
0 7 1G 1340 0 40 40 • 2.61+
0 8 10 1420 0 40 40 • 3.78
0 q 10 1500 1j 40 l+O • 2.92
0 10 10 151+0 0 20 20 • 4.15
1(l 1600 20
0 11 0 20 • 4.15
0 12 10 1620 0 40 40 • 3.26
TABLE 6. (Continued)
Sample
Test· Start SamEling Time Volume, Coh/1000
Laboratory Number Day Time hr min min cu ft Linear Feet
0 49 12 1441 1 59 119 • 1.85
0 50 12 1540 2 00 120 • 1.99
n 51 12 1840 2 li 0 120 • 2.31
0 52 12 2040 2 00 120 • 2.05
0 53 12 2240 2 00 120 • 1. 40
0 54 13 0040 2 fJ 0 120 • 1.1+1t
0 55 13 0240 2 co 120 • 1.85
0 56 13 0440 2 () 0 120 • 3.32
0 57 13 0640 2 00 120 • 3.03
0 58 13 OR40 0 40 40 • 4.29
0 59 13 OCJ20 0 40 40 • 2.09
0 60 13 1000 0 36 36 • 2.37
n 61 13 1036 0 40 40 • 1.58
0
0
62
63
13
13
1116
1156
0
IJ
40
40
40
40 ..
• 2.36
2.61
0 64 13 1236 0 40 40 • 2.61
0 65 13 1316 0 40 40 • 2.95
0 66 13 1356 0 44 44 • 2.63
0 67 13 1440 J 42 42 • 6.11
0 68 13 1522 0 40 40 • 5.96
0 69 13 1602 0 38 38 • 3.78
0 70 13 1640 0 40 40 • 2.99
0 71 13 1720 0 40 40 • 2.oa
TABLE 6. (Continued)
Sample
Test Start Sam:eling Time Volume, Coh/1000
Laboratory Number Day Time hr min min cu ft Linear Feet
E 1 q 1100 15 (a)
00 900 • t.76 1. 6«;
E 2 10 0800 1 53 113 • r..11 5.1.:
E 3 10 1300 0 40 40 • 7.54 7.86
E r. 10 1340 0 40 40 • 6.43 7.62
E 5 10 11t20 Q 40 40 • 1.01 8. 49
E 6 10 1c;oo 0 40 40 • 7.30 6.27
E 1 10 15ft0 0 40 r.o • 9.53 8.97
E 8 10 1620 0 40 40 • 7.36 8. t>J
E 9 10 1700 2 00 120 • 6.66 6.86
E 10 10 1<l00 2 00 120 • 6.03 6.17
E 11 10 2100 2 on 120 • 1.12 6.75
E 12 10 2300 2 00 120 • 6.56 5.56
E 13 11 0100 2 00 120 • 5.71+ 4.87
E 14 11 0300 2 00 120 • 6.88 6.q&
E 15 11 ()500 2 co 120 • 1.20 1.12
E 16 11 0100 2 00 120 • CJ.60 8.67
E 17 11 0900 0 30 30 • 6.36 5.07
E 18 11 0930 0 3 !) 30 • 7.65 8.95
E 19 11 1000 il 30 30 • 1.11 8.08
E 20 11 1030 0 30 30 • 7.87 5.82
E 21 11 1100 0 40 40 • 5.66 6.22
E 22 11 1140 0 40 40 • s.90 6.30
E 23 11 1220 a 40 40 • r..61 5.98
E 24 11 1300 0 40 40 • r..11 5.66
E 25 11 13ft(j 0 i+O 40 • r..11 4.sz
E 26 11 1420 a 40 40 • 5.09 4.93
E 27 11 1500 0 ftO 40 • r..69 5.09
E 29 11 1615 0 lt5 45 • r..31 4.1r.
E 30 11 1700 2 co 120 • l+.52 t+.30
E 31 11 1900 2 00 120 • 5.24 s.2r.
E 32 11 2100 2 co 120 • 5.53 s.13
E 33 11 2300 2 00 120 • 4.57 ... r.9
E 34 12 0100 2 co 120 • 5.24 4.sr.
E 35 12 0300 2 fl 0 120 • r.. r. 0 3.92
E 36 12 0500 2 00 120 • r.. tl3 3.68
E 37 12 0700 2 00 120 • s.oq 4.85
E 38 12 0900 1 00 60 • 4.91 4.16
E 39 12 1000 0 40 40 • 6.64 E>.16
E ltO 12 1040 0 ftf) r.o • 2.2 .. 1 .....
E 41 12 1120 0 fc.Q ltO • 3.92 3.76
E lt2 12 1200 0 It 0 40 • 3.52 4.32
E lt3 12 12r.o 1 00 60 • 1.60 2.2 ..
E 41t 12 13ft0 1 00 60 • 3.63 4.06
TABLE 6. (Continued)
Sample
Test Start Sam;eling Time Volume, Coh/1000
Laboratory Number Day Time hr min min cu ft Linear Feet
(a)
E 45 12 141+0 2 00 120 • 4.29 3.57
E 46 12 1640 2 00 120 • 4.93 4.00
E 47 12 1'\40 2 00 120 • 4.97 4.73
E lt 8 12 2040 2 00 120 • 4.54 5.5&
E 49 12 2240 2 00 120 • 6.04 4.22
E 50 13 0040 2 0 i) 120 • 4.01 3.61
E 51 1~ 0240 2 00 120 • 4.20 4.57
E 52 13 044u 2 00 120 • 5.53 5.10
E 53 13 0640 2 00 120 • 5.97 6.16
Sample
Test Start SamEling Time Volume, Coh/1000
Laboratory Number Day Time hr min min cu ft Linear Feet
F 3 10 1100 0 40 40 • 1.31
F 4 10 1140 0 40 40 • 1.G8
F 5 10 1220 0 40 40 • 1. 31
F 6 1C! 1300 0 40 40 • o.94
F 7 10 131+0 0 40 40 • 0.58
F 8 10 1420 0 40 40 • 0.10
F 9 10 1500 0 40 40 • o.58
F 10 1C 154~ 0 40 40 • 0.94
F 11 10 1620 0 40 40 • 1.43
F 28 11 1340 Q 40 40 • 1.98
F 29 11 1420 0 40 40 • 1.12
F 30 11 1500 0 40 40 • 1.46
F 31 11 1540 1 16 7G • 1.15
F 32 11 1700 2 00 120 • 1. 73
F 33 11 1900 2 00 120 • 2.4G
F 34 11 2100 2 c~ 120 • 2.04
F 35 11 2300 2 00 120 • 1.Glt
F 36 12 0100 2 00 120 • h51t
F 37 12 0300 2 00 120 • 1.26
F 38 12 0500 2 00 120 • 1.1+9
F 39 12 0700 2 00 120 • 1. 78
F 40 12 oqoo 1 00 60 • 2.42
F 1+1 12 1000 0 40 40 • 1.80
F 42 12 1040 0 40 40 • 1.75
F 43 12 1120 0 40 ltO • 1.&8
F 44 12 1200 0 40 40 • 1.43
F 45 12 1240 1 00 GO • 1.37
F 1+6 12 1340 1 00 GO • 1.28
F 47 12 1440 2 00 120 • 1. GO
F 48 12 1640 2 00 120 • 2.00
F 49 12 161+0 2 00 120 • 2.10
F 50 12 2040 2 00 120 • 1.e5
F 51 12 2240 2 00 120 • 1.G5
F 52 13 0040 2 00 120 • 1.e5
F 53 13 0240 2 00 120 • 2.00
F 54 13 0440 2 fj 0 120 • 2.10
F 55 13 0&1+0 1 45 105 • 2.GCJ
26
TABLE 6. (Continued)
Sample
Test Start SamEling Time Volume, Coh/1000
Laboratory Number Day Time hr min min cu ft Linear Feet
F 56 13 0840 0 40 40 • 4.71
F 57 13 og20 !) 40 40 • 2.96
F 58 13 1IJ0 0 0 40 40 • 2.44
F 59 13 1040 0 40 40 • 2.96
F 60 13 1120 0 40 40 • 3.23
F 61 13 1200 0 40 40 • 3.50
F 62 13 1240 Q 40 40 • 2.96
F 63 13 1320 0 40 40 • 2.70
F 64 13 1400 0 40 40 • 3.36
F 65 13 1440 0 40 40 • 4.48
F 66 13 1520 0 40 40 • 5.14
F 67 13 1600 0 40 r.o • 4.48
F 68 13 1640 0 40 40 • 3.72
F 69 13 1720 0 40 40 • 2.10
F 70 13 18 00 2 00 120 • 1.31
F 71 13 20 00 2 00 120 • 1.69
F 72 13 2200 2 00 120 • t.50
F 73 14 ori o o 2 (l 0 120 • 0.63
F 74 14 0200 2 00 120 • 0.42
F 75 14 01+00 2 00 120 • o.sq
F 76 14 0600 2 00 120 • o.93
F 77 1l+ 0'3 0 0 2 co 120 • 1.29
F 78 14 10 00 2 00 120 • 0.99
F 79 14 1200 2 00 120 • o.s6
F 80 14 140() 2 00 12n • 0.99
F 81 14 1oon 2 OD 120 • 1.04
F 82 14 1800 2 00 120 • 1.26
F 83 14 2000 2 00 120 • 1.40
F 84 14 2200 2 Cu 120 • 1.36
F 85 15 OOOJ 2 0
(! 120 • 1.01+
F 86 15 0200 2 00 120 • "1.59
F 87 15 0400 2 co 120 • 1.40
F 88 15 0600 1 53 113 • 1.48
27
TABLE 6. (Continued)
Sample
Test Start SamEling Time Volume, Coh/1000
Laboratory Number Day Time hr min min cu ft Linear Feet
G 1 9 1705 14 55 895 312.9~2 o.&s
G 2 10 0800 2 30 150 52.590 2.37
G 3 10 1100 0 40 40 14.040 2.74
G 4 10 1140 0 40 40 14.044 2.73
G 5 10 1220 0 40 40 14.056 2.73
G 6 10 1300 0 40 40 14.056 3.13
G 7 10 1340 0 40 40 14.068 2.73
G 6 1(1 1420 0 40 40 1lt.06B 3.33
G 9 1G 1500 0 ltO 40 14.068 3.13
G 10 10 1540 0 40 40 14.061+ 3.75
G 11 10 1620 0 40 40 14.060 3.55
G 11A 10 1700 0 I+ IJ 40 14.060 4.18
G 11B 10 1740 0 40 40 14.068 4.18
G 11C 10 1823 0 40 40 14.068 4.39
G 12 10 1900 2 00 120 42.204 2.86
G 13 10 2100 2 00 120 42.192 2.95
G 14 1(1 2300 2 00 120 42.180 1.68
G 15 11 0100 2 (l 0 120 42.156 1.46
G 16 11 0300 2 QO 120 42.168 1.61
G 17 11 0500 2 co 120 42.156 3.05
G 18 11 070\l 2 00 120 42.180 3.88
G 19 11 0900 0 30 30 10.551 2.86
G 2C 11 0930 0 30 30 10.551 2.86
G 21 11 1100 0 30 30 10.551 2.61
G 22 11 1030 0 30 30 10.551 2.86
G 23 11 1100 0 40 40 14.068 2.53
G 24 11 1140 0 40 40 14.076 2.14
G 25 11 1220 a 40 40 14.036 2.54
G 26 11 1300 0 40 40 1c..012 2.54
G 27 11 1340 0 40 40 14.020 1.11
G 28 11 1420 il 40 40 14.046 1.11
G 29 11 1500 0 40 40 14.076 1.40
G 30 11 1540 1 20 80 28.160 1.36
G 31 11 1700 2 00 120 42.252 2.01
G 32 11 1900 2 00 120 42.216 2.95
G 33 11 2100 2 00 120 42.228 2.08
G 34 11 2300 2 00 120 42.226 1.61
G 35 12 0100 2 00 120 42.204 1.61
G 36 12 0300 2 00 120 42.144 1.18
G 37 12 0500 2 en 120 42.0GO 1.39
G 36 12 0700 2 00 120 42.03& 1. 77
G 39 12 0900 1 GO GO 21.030 2.37
G 4C 12 1000 0 40 40 14.026 1.11
G 41 12 1040 0 40 40 14.036 1.1+0
G 42 12 1120 0 40 l+O 14.036 1.11
G 4"3 12 1200 D 40 40 14.036 1.40
G 44 12 1240 1 co 60 21.060 1.30
G 45 12 1340 1 00 60 21.042 1.43
28
TABLE 6. (Continued)
Sample
Test Start SamEling Time Volume, Coh/1000
Laboratory Number Day Time hr min min cu ft Linear Feet
G 46 12 11+40 2 00 120 1+2. 084 1.51t
G 47 12 1f)l+0 2 00 120 42.072 1.84
G 1+8 12 181+0 2 00 120 42.06J 2.25
G 1+9 12 2040 2 GO 120 1+2. Q6() 2.00
G 50 12 221+0 2 00 12C 1+2.012 1.39
G 51 13 00 l+O 2 co 120 42.081+ 1.39
G 52 13 021+0 2 00 120 42.108 1.61
G 53 13 04 I+ 0 2 !J 0 120 1+2. 11+1+ 2.86
G 54 13 G61+0 2 00 120 1+2.156 2.59
G 55 13 081+0 0 l+O l+O 11+.024 5.31
G 56 13 0920 0 l+O l+O 13.976 2.75
G 57 1"3 10 0 0 0 40 40 13.952 2.75
G 58 13 1'340 0 40 40 13.961+ 2.95
G 59 13 1120 0 l+O 40 13.980 2.95
G 60 13 1200 0 l+O 40 13. 988 3.15
G 61 13 124;] Q l+O 40 13.980 2.95
G 62 13 1320 0 l+O 40 13.980 3.15
G 63 13 11+00 0 40 l+O 13.992 3.90
G 64 13 11t40 0 40 40 14.008 l+.86
G 65 13 1520 0 l+O l+O 11t.016 4.41
G 66 13 1600 0 40 I+ 0 14.056 s.,3o
G 67 13 161+0 a l+O l+O 11+.092 3.33
G 68 13 1720 0 l+O 40 14.10!) 2.14
G 69 13 1800 2 00 120 1+2.336 1.21+
G 70 13 20 0 'l 2 00 120 42.1+06 1.38
G 71 13 2200 2 00 120 42.396 1.29
G 72 1 t+ 0'.)00 2 ro 120 42.312 0.1+0
G 73 14 0200 2 co 120 1+2.252 c.28
G 71+ 14 040 0 2 00 120 42.180 G.59
G 75 14 0600 2 00 120 1+2. 0 96 0.81+
G 76 14 0800 2 cl) 120 42.01+8 1. 0 I+
G 77 11+ 1000 2 00 120 41. 988 0.78
G 78 14 1200 2 00 120 41.964 0.91
G 79 11+ 11+00 2 00 120 1+1. 9il4 0.98
G 80 14 1600 2 co 120 41.81+4 1.12
G 81 11+ 1800 2 00 120 41.820 0.92
G 82 14 20 00 2 00 120 1+1.796 1. 05
G 83 14 2200 2 00 120 41.772 1.12
G 81+ 15 0000 2 00 120 1+1.760 0.92
G 85 15 0200 2 00 120 1+1.71+8 1.26
G 66 15 01+00 2 00 120 1+1.736 1.31
G 87 15 0600 2 00 120 41.736 1.12
29
Laboratory:
A B c D E F G Total
Sunday 1 1 1 1 2 1 1 8
Monday 14 14 14 15 22 14 16 109
Tuesday 20 20 20 21 40 12 20 153
Wednesday 16 16 16 16 32 16 16 128
Thursday 21 21 21 22 44 21 21 171
Friday 12 12 11 12 24 12 12 95
Reproducibility:
Repeatability
Analysis of Variance
ComQonents of Variance
Total 5.93152 69
(a) Column headings: SSD, sum of squared deviations from the mean;
df, degrees of freedom; MS, mean square; s, standard deviation.
- - 2
I: n.]. (x.]. -x) ,
were, x.]. denotes the measurement value obtained by the ith laboratory,
and x denotes the arithmetic mean of the measurements obtained by all
laboratories for a given test period. This expression is also swmned over
test periods to obtain the number given in the table. For the laboratory
making duplicate measurements, x.]. is taken to be the average of the two
measurements, and the mean is given a weight of 2 so that n.]. = 2. For the
laboratories which made a single determination in a given test period, xi
is equal to that determination and n.]. is 1.
The agreement between laboratories within a test period is asso-
ciated with a mean square equal to 0.07502. This mean square is algebraically
2
represented bys
w
+ K sb 2 , wheresw2 denotes the within-laboratory variance,
2
sb denotes the between-laboratory variance, and K denotes a weighted average
33
number of runs per lab(S). An estimate of the variability (sb 2 ) among labora-
tories can be derived by taking the difference between these two mean squares,
2 2 2
(sw + K sb ) - (sw ), and dividing by K. Application of this procedure to
the results in Table 8 gives (0.07502 - 0.00096)/K = 0.0676 for the variance
among laboratories, and -.../o.0676 = 0.260 Cohs per 1000 linear feet for the
corresponding standard deviation (sb). This standard deviation represents
the reproducibility of the pooled tape sampler determinations at Los Angeles.
Although large variations were expected to occur between tests,
and were possibly expected between short and long time spans, these variations
were not of primary interest in this program. They are included in the table
simply to show the relationship of variations due to laboratories and to
time differences between test periods.
Table 9 shows an analysis of variance for measurements of particulate
matter at Bloomington (Site 2). This analysis is similar to the analysis
for Los Angeles described above, with one exception. In that analysis, there
was a component of variance between short and long duration tests which is absent
in the analysis for Bloomington. The between-measurement variation in particulate
matter is given by '1f5.00152 = 0.039 Cohs per 1000 linear feet. The between-
laboratory variation in the measurement of particulate matter is given by
(0.02856 - 0.00152)/K = 0.0240 for the variance among laboratories, and
~= 0.155 Cohs per 1000 linear feet for the corresponding standard
deviation (sb) of the pooled measurements.
Table 10 shows an analysis of variance for measurements of particulate
matter at Manhattan (Site 3). The statistical treatment is similar to the
analysis for the Los Angeles and Bloomington data described above. Only
about half of the data from the Manhattan tests were used in the analysis
which is summarized in Table 10. Data from the Manhattan tests were initially
divided into 88 groups. In these groupings, not all laboratories started
or terminated sampling at exactly the same time; therefore these data include
changes in particulate concentration with time. To eliminate this shortcoming,
the Manhattan data were selected and compiled into 44 tests, as shown in
34
Total 0.91217 38
(a) Colllll!n headings: SSD, slllll of squared deviations from the mean;
df, degrees of freedom; MS, mean square; s, standard deviation.
(b) This variation could not be calculated because the averages for
different spots were more nearly equal than the measurements by
different laboratories. The model for computing "s" does not
hold in this case.
(a) Coltmm headings: SSD, slllll of squared deviations from the mean;
df, degrees of freedom; MS, mean square; s, standard deviation.
35
Table 11, in which ail laboratories started and terminated at the same time
(within ten minutes) in each group. It is these 44 tests which were used to
obtain the results presented in Table 10.
Sam2le
Labor a- Test Start Saun> 1ing Time Linetfo3) Coh/1000
tory Number Day T.ime hr min min cu ft Feet Linear Feet
SaJ!!i? le
..,
~--
teat
Humber Day
Start
Time
Sa!!mling Time
hr min min cu ft
Linear
Feet(lO)
Coh/1000
Linear Feet
G
10
1(1
10
.2100
2100
2100
2
2
2
00
00
00
120
120
120
•
•
42.192
•
•
7.742
1.12
1.08
2.95
6.75
•
•
17 11 03110 2 00 120 24.74 • 2.63 •
A
8 11 C300 2 00 120 8.04 • 4.153 •
c 11 Cl250 2 00 ~20 • • 3.90 •
0 11 0257.5 1 40 100 • • 2.69 •
E 11 03DD 2 00 120 • • 6.88 6.96
F 11 0300 2 00 120 • 0.73 •
11 C30 il 2 00 120 42.166 7.737 1.61 •
G
37
Sa!!!J.?le
Labor a- Test Start Sam:eling Iime Linear Coh/1000
tory Number Day Time hr min min cu ft Feet(lO) Linear Feet
...
Sam2le
Wbor•- Test Start Sam2Hns Iime Linear Coh/1000
tory Number Day Time hr min min cu ft Feet(lO) Linear Feet
A 't1 12 1000 0 40 40 9.71 • 2.8 .. •
.8 1Z 1000 0 40 40 2.68 • 1.744 •
c 1.2 1000 0 40 40 • • 3.53 •
b 12 1000 0 40 40 • • 3.34 •
E 12 1000 0 40 40 • • 6.64 6.16
y: 12 1QOO 0 40
40.
40 • • 1.80 •
G 12 10'00 0 40 14.028 2.574 1.77 •
A 42 12 1040 0 40 40 9.71 • 1.51 •
B 12 10 40 0 40 40 2.68 • 1.958 •
c 12 1040 0 40 40 • • 2.38 •
0 12 10 40 0 30 30 • • 1. 40 •
E 12 10 40 0 40 40 • • 2. 2 4 1. 44
F 12 10 40 0 40 40 • • 1.75 •
G 12 10 40 a 40 40 14.036 2.575 1.40
•
A 43 12 1120 fl 40 40 9.71 • 1. 51 •
B 12 1120 0 40 40 2.68 • 1.121
c 12 1120 0 40 40 • • 2.98 •
0 12 1110 0 35 35 • • 0.90 •
E 12 1120 u 40 ltO • • 3.92 3.76
.F 12 1120 0 40 40 • • 1.68 •
G 12 1120 0 40 40 14.036 2.575 1.77 •
A 44 12 1200 (j 40 40 9.11 • 2.11 •
B 12 1200 0 40 40 2.68 • 1.406 •
c 12 1200 0 40 40 • • 2.78 •
D 12 1200 0 44 44 • • 0.96 •
E 12 1200 0 40 40 • • 3.52 4. 32
F 12 1200 0 40 40 • • 1.43 •
G 12 1200 0 40 40 14.036 2.575 1.40 •
A 45 12 12'+J 1 00 60 14.56 1.53 •
B 12 1241J 1 ca 60 4.02 • 1.368 •
c 12 1240 1 00 60 • • 2.39 •
D 12 1244 0 59 59 • • 1.26 •
E 12 1240 1 00 68 • • 1.60 2.24
F 12 1240 1 to 60 • • 1.37
G 12 1240 1 co 60 21.060 3.864 1.30 •
A 46 12 134Y 1 00 60 14. 56 • 1.11 •
8 12 134J 1 00 60 4.02 • 1.993 •
c 12 1340 1 00 60 • • 2.84 •
D 12 1343 0 58 58 • • 1.42 •
E 12 1340 1 00 60 • • 3.63 4. 06
F' 12 13 4 (I 1 00 60 • • 1.28 •
G 12 1340 1 00 60 21.042 3.861 1.43 •
39
Sam:ele
Labor a- Test Start SamElins Time Linear Coh/1000
tory Number Day Time hr min min cu ft Feet(lO) Linear Feet
Sam2le
Labor a- Test Start Sam21ing Time Linear Coh/1000
tory Number Day Time hr min min cu ft Feet Linear Feet
A ~3 13 4'240 2 00 120 24. 84 • 2.37 •
B 13 0240 2 00 128 e. 04 • 3.21t3 •
13 0240 z 00
00
120 • • 3.lt9 •
0 13 0240 2 120 • • 1.a5 •
E 13 0240 2 00 120 • • ... 20 ... 57
F 13 0240 2 00 120 • • 2.00 •
G 13 0240 2 00 120 42.108 7. 726 1e61 •
A 54 13 0440 2 00 120 24. 84 • 3.82 •
B 13 0441) 2 00 120 8.04 • 5.557 •
c 13 0440 2 00 120 • • s.os •
0 13 044 ll 2 00 120 • • 3.32 •
E 13 0440 2 00 120 • • 5.53 5.10
F 13 0440 2 LO 120 • • 2.10 •
G 13 04411 2 00 120 42.144 7.733 2.86 •
A 55 13 G64G 2 00 120 24. 84 • ... 25 •
B 13 0640 2 00 120 8e04 • 5.0lt3 •
c 13 0640 2 co 120 • • ... 86 •
0 13 0640 2 00 120 • • 3.03 •
E 13 0640 2 00 120 • • s.97 6.16
F 13 064() 1 45 105 • • 2.69 •
G 13 0640 2 DO 120 42.156 7.735 2.59
A 56 13 0840 0 40 40 9.66 • 7.5 .. •
B 13 G84J 0 40 40 2.68 • 7.357 •
c 13 084U il 40 40 • • 8.98 •
0 13 0840 0 40 40 • • ...29 •
E 13 0840 il 40 40 • • 8.91 8.27
F 13 0840 0 40 40 • • ... 11 •
G 13 0840 0 40 40 14.024 2.573 5.31 •
A 57 13 092u i) 40 40 9.66 • 4.11 •
B 13 0920 0 C.0 40 2.68 • 4.069 •
c 13 oq20 0 40 40 • • 5.32 •
0 13 0920 0 40 40 • • 2.09 •
E 13 0920 0 40 40 • • 6.89 1.~9
F 13 oq20 a C.0 40 • • 2.96 •
G 13 0920 0 40 40 13.976 2.564 2.15
A 58 13 10 ll (I 0 40 40 9.66 • 3.27 •
B 13 1000 0 40 40 2.68 • 4.218 •
c 13 10 00 0 40 40 • • 5.50 •
0 13 10 0 Cl G 36 36 • • 2.37 •
E 13 10 00 0 .40 40 • • 6.24 6 .. :89
F 13 1:J 00 0 40 40 • • 2.44 •
G 13 10 oa 0 40 40 13.952 2.560 2.75 •
41
TABLE 11. (Continued)
SamEle
Labor a- Test Start Sa!!!l?ling Time Linear Coh/1000
tory Number Day Time hr min min cu ft Feet(lO) Linear Feet
A 5q 13 10 40 0 40 4tJ 9.66 ..... 2
u ftO ltO
• •
B 13 1040 Z.68 • 4.153 •
c 13 1040 D 40 40
• • lt.67 •
0 13 1036 0 40 40 • • 1.5&
0 40 40
•
E 13 1il40 • • 9.16 9.2 ..
f
G
13
13
1040
104u 0
0 40
40
40
40
•
13.964
•
2.562
2. C)'6
2.95 ..•
A 6~ 13 112U 0 40 40 9.66 • 4.29
0 40 40 •
B 13 1120 2. 68 • 3.933 •
c 13 1120 0 40 40 • • 5.81 •
0 13 1116 0 40 40 • Z.36
0 ltO 40 • •
f 13 1120 • • 6.16 5.92
'F 13 112u 0 40 40 3.23
0 40 40
• • •
G 13 1120 13.980 2.565 2.95 •
A 61 13 1200 J 40 40 9.66 4.42
a 40 40
• •
B 13 1200 2.68 • 4.153 •
c 13 1200 0 40 ltO
• • 5.1 .. •
0 13 1156 0 40 40 2.61
0 40 40
• • •
E 13 1200 • • 6.97 7.54
F 13 1200 0 40 40 • 3.50
0 40 40
• •
G 13 1200 13.988 2.567 3.15 •
A 62 13 1240 0 40 40 9.66 ... 21
0 40 40
• •
B 13 1240 2.68 • ... 764 •
c 13 124() 0 40 40
• • 5.99 •
0 13 1236 0 40 40 • 2.61
0 40 40
• •
E 13 1240 • • 5.92 1.18
F 13 1240 0 40 40 • 2.96
0 40 40 • •
G 13 1240 13.980 2.565 2.95 •
A 63 13 1320 0 40 40 9.66 • 5.31 •
8 13 1320 0 40 40 2.68 5.832
40
• •
c 13 1320 0 40 • • 7.06 •
13 1316 0 40 40 2.95
D
il c.o 40
• • •
E 13 1320 • • 11.10 8.02
f' 13 1320 0
0
40
40
40
ftO
.. • 2.70 •
G 13 1320 13.980 2.565 3.15 •
A 64 13 1400 J 40 40 9.66 • 5.16 •
B 13 1400 0 40 40 2.68 • 6.170 •
c 13 1400 a 40 40
• • 7.91 •
0 13 1356 0 44 44 • • 2.63 •
.E 13 14il0 0 40 ltO
• • 6.81 7.29
~ 13 1400 0 40 40 • • 3.36 •
G 13 1400 0 40 ftO 13.992 2.567 3.90 .
42
Sam:ele
f,ji;il)C)r-a'"' Test Start SamEling Time Linear Coh/1000
t:Ky Humber Day Time hr min min cu ft FeetClO) Linear Feet
•
~
65 13
13
141t0
141tD
0
0
..a
40
40
40
9.66
2.ei6
•
•
&.17
7.535
•
•
c0 13 141tll 0 40 40 • • 8.53 •
13 t~40 0 42 42 &.11 •
£ 13 tt.40 • •
40 40
.,
F' 13
13
1440
1440
0
0
0
40
40
40
40
•
•
14.008
•
•
2.570
9.32
4.48
4.86
9.89
•
•
A 66 13 1520 0 48 40 9. E>E> • 6.23 •
B 13 1520 0 40 40 2.68 • 7.179 •
c 13 1520 IJ 40 40 • • 8.97 •
0 13 1522 0 40 40 • • 5.96 •
E 13 1520 0 40 40 • • 12.40 12.24
F 13 1520 0 40 i+o • • 5.14 •
G 13 1520 0 40 40 14.01& 2.572 4.41
A 67 13 1E>OO 0 ftO 40 9. E>E> • 6.89 •
8 13 1E>OO 0 40 40 2.ei8 • 8.840 •
c 13 1600 0 40 40 • • 8.90 •
0 13 1E>02 0 38 38 • • 3.78
13 1600
•
E 0 40 40 • • 11.51 11. 75
F 13 1E>OO 0 40 40 • • 4.48 •
G 13 1600 0 40 40 14.056 2.573 5. 30 •
A 68 13 1E>40 0 40 40 9. E>E> 4.34
13 1&40
•
8 0 40 40 2.E>8 • 5. E>90 •
c 13 1E>40 0 40 40 • • 1.0 .. •
0 13 1E>40 0 40 40 • • 2.99 •
E 13 1E>40 0 40 40 • E>.73 6.19
F 13 1E>40 •
0 40 40 • • 3.72 •
G 13 1E>40 0 40 40 14.092 2.5ee, 3.33 •
A E>9 13 1720 0 40 l+O 9. E>6 • J.13 •
8 13 1720 0 ltO 40 2.68 3.311 •
:C 13 1720
•
2 00 120 • • 2.75 •
0 13 1720 0 40 40 • 2.08 •
E 13 1720 •
0 40 40 • • El. 0 8 5.92
F 13 1720 0 40 40 2.10 •
G 13 1720 • •
0 40 40 14.100 2.587 2.14 •
43
Analysis of Tests
Tables 12, 13, and 14 show a detailed analysis of tape sampler deter-
mination for each test at each of the three test sites. Each row, corresponding
to a given test at a given site, contains the mean value of particulate matter,
the between-laboratory standard deviation, and the coefficient of variation
relating these two statistics. In test 2 at Los Angeles, for example, the
table shows that number of measurements (n) is equal to 9; the mean tape sampler
determination (m) for these measurements is 0.639 Cohs per 1000 linear feet;
the between-laboratory variation (reproducibility) is given by sb = 0.401 Cohs
per 1000 linear feet based on 7 degrees of freedom (df); and the associated
coefficient of variation (CV), obtained from the ratio sb/m, is 63 percent.
The summary rows that appear at the bottom of each table give the associated
totals for n and df, and the average values form, sb' and CV.
0.193 + 0.367 m.
(b) Laboratory E was excluded from the statistical analysis because its
sampling period for test number 1 did not correspond with the other
laboratories.
45
1 8 0.292 6 0.092 32
2 8 0.399 6 0.206 52
3 (b) 7 0.290 5 0.136 47
4 8 0.279 6 0.154 55
5 8 0.330 6 0.162 49
All(b) 39 o. 319 29 0.155 49
1 7 1.098 6 0.446 41
2 7 3.273 6 1.630 so
6 7 4.163 6 2.2S9 S4
7 7 3.92S 6 2.320 S9
8 7 4.741 6 2.381 so
9 7 4.603 6 2.620 S7
10 7 S.374 6 2.876 S3
14 7 4.609 6 2.103 46
17 7 3.228 6 2.006 62
28 7 2.636 6 1.126 43
29 7 2.143 6 1.264 46
30 7 2.230 6 1.216 SS
32 7 2.826 6 1.022 36
33 7 3.660 6 l.OS3 29
34 7 3.122 6 l.2SS 40
41 7 3.09S 6 1.739 S6
42 7 l.80S 6 0.400 22
43 7 2.069 6 1.024 so
44 7 l.9S2 6 0.918 47
4S 7 l.S4S 6 0.392 2S
46 7 2.043 6 0.811 43
47 7 2.482 6 0.991 40
48 7 2.878 6 1.130 39
49 7 3.2S3 6 1.090 34
so 7 2.911 6 1.017 3S
Sl 7 2.621 6 1.638 62
S2 7 2.226 6 0.892 40
S3 7 2.680 6 0.973 36
S4 7 4.124 6 1.246 30
SS 7 4.062 6 1.316 32
S6 7 6.728 6 l.9S4 29
S7 7 4.113 6 1.674 41
S8 7 3.827 6 l.S43 40
S9 7 4.270 6 2.408 S6
60 7 4.lOS 6 1.434 3S
61 7 4.278 6 l.4S4 34
62 7 4.208 6 1.419 34
63 7 S.443 6 2.994 SS
64 7 S.134 6 1.938 38
6S 7 6. 71S 6 1.819 27
66 7 7.184 6 2.729 38
67 7 7.100 6 2.79S 39
68 7 4.834 6 l.6SO 34
69 7 3.170 6 1.362 43
.µ
Cl)
Q)
4-l
1-4
CC!
Q) 5. 0
i::
-rl
r-f
0
n 57
0
0
r-f r = 0.80
1-1
Q)
p,.
Sb = 0.193 + 0.367 m
Cl)
..c: ". 0
8
.......
,.Cl
Cl)
.._,
i::
0
·r-l
.µ 3.0 ..
CC!
·r-l
>
..
<II
Q ..
.. .. ..
"c:I
1-1
CC!
"c:I
..
i::
CC!
.µ
C/'l 2.0 .. ..
:>...
1-1 .. .. ..
.µ
0
CC!
..
1-1
......
.0
0
CC!
....
H
I
i::
Q) I .0
<II
~
.µ
<II
i:Q
0.0 --~~~....~~~~""-~~~-'-~~~~i....-~~~...a...~~~......i'"-~~~.....~~~--'
0. 0 I .0 2. 0 3.0 ". 0 5. 0 6. 0 7. 0 8. 0
.j.J
Q)
I2. 5
Q)
44
1-1
Cl!
Q)
i:::
-r-1
r-l
I0.0
0
0
0
•
r-l
1-1
Q)
• ••
p. • •
tll
,.c:
7. 5 • •
8
,....,,
N
i:i:::
'.J'
• •
00 •• • •
i:::
·r-1 5.0 • .. .,f. •
••• • ,.• "'
"O
Cll
Q)
i:i:::
• • •
"O
i:::
"'¥ ••
0
cJ
Q)
Cf)
2.5 •• •• "'•
... •
~
0. 0 .....~~~~--'~~~~~-'-~~~~~......~~~~~.........~~~~~--~~~~---
0. 0 2. 5 5.0 7. 5 l 0 . 0 l 2 . 5 l 5. 0
Sb = 0.193 + 0.367m,
where, sb, and, m, are the reproducibility and mean soiling index, respectively,
in Cohs per 1,000 linear feet. The observed coefficient of variation (100 sb/m)
of between-laboratory measurements over a soiling index range of 0.27 to 7 Cohs
per 1,000 feet varied from about 64 to 22 percent. These statistical measures
indicate that Method D 1704, as tested, does not yield highly reproducible
between-laboratory measurements of particulate matter in the atmosphere.
sampling period. Due to the characteristics of the filter media and variations
in particulate concentration with time, the flow rate probably does not vary
linearly with time as assumed. Consequently, errors can occur in the flow
measurements which contribute to the lack of reproducibility.
RECOMMENDATIONS
(2) The sampling probe should be free of sharp bends, connections, or any
restrictions which might trap particulates.
ACKNOWLEDGEMENTS
Fred Bergman
Nick Stich
Frank Hanis
Cliff Decker
Denny Wagoner
53
John Driscoll
Jim Becker
Roland Hebert
REFERENCES
(1) Annual Book of ASTM Standards, Part 23, American Society for Testing
and Materials, 1915 Race Street, Philadelphia, Pennsylvania 19103.
(5) Kempthorne, O., "The Design and Analysis of Experiments", John Wiley
and Sons, Inc., pgs 103-110 (1952).
(6) Scheffe, H., "The Analysis of Variance", John Wiley and Sons, Inc.,
pgs 248-258 (1959).
APPENDIX
REPRINT OF ASTM
57
D 1704
and the area of the circular spot or to the size 4.4 Flow Meters-Gas meters of any suit-
of the air sample in linear units. Such a scale able type, such as calibrated orifice or capil-
serves to describe the quantity of aerosols in a lary flow meters; one to measure air flow in
linear unit of air in terms of its staining or the range of 0.10 to 0.50 ft 3 (0.003 to 0.014
soiling properties. It is related only indirectly m 3 ) min for stain or spot samples; one to
to the mass of deposit. however, the recom- measure air flow of about I 0 to 50 ft 3 (0.28 to
mended simultaneous collection, using a high 1.40 m 3 )/min for collection of particulate
volume sampler, of a weighable sample of the matter for weight determination and analysis.
aerosol, permits a direct conversion of the 4.5 Pumps-For the spot samples, a recip-
optical densities to mass concentration units. rocating diaphragm pump, or one of the con-
stant displacement vacuum type, with an ini-
3. Definitions
tial capacity of about 2.0 ft 3 (0.056 m 3 )/min
3.1 For definitions of terms used in this is employed. Mass samples are collected by
method, refer to ASTM Definitions D 1356, means of a relatively large capacity pump of
Terms Relating to Atmospheric Sampling and the multivane fan or compressor type, with
Analysis. 3 initial delivery of about 10 to 50 ft 3 (0.28 to
1.40 m 3 )/min.
4. Apparatus•
4.6 Clamping Device and Interval Timer-
4.1 General Description-The instrument The inlet head shall be equipped with a
consists of an assembly containing a sampling clamping device to fasten securely over a con-
probe with filter screen for removing coarse stant exposed area of filter paper. This area is
material, clamping device or head to hold the exposed for I, 2, 3 h or more for a given spot
inlet plug securely to the paper, paper holder sample, depending upon atmospheric and
to accommodate a roll of filter paper, interval meteorological conditions. The timing device
timer actuated by a clock, calibrated flowme- at the end of this period releases the paper,
ter, electrically driven air pump, and timer which is moved to expose a fresh area and the
control device for adjustment of the sampling cycle of operations is repeated. The timing
cycle (see Figs. I and 2). The optical density cycle must be capable of adjustment in order
of the deposit or circular stain on the filter to avoid sample spots of excessive optical
paper is determined by measurement of light density. Parts that come in contact with the
transmittance or reflectance in a laboratory filter paper shall be of stainless steel, plastic,
optical instrument of conventional design, or other corrosion-resistant noncontaminating
such as a spectrophotometer or photoelectric materials. Filter holder shall be adequate to
colorimeter. provide area and flow rate for spot samples
NOTE I-In order to characterize the deposits 0.25 to 0.40 ft 3 (0.007 to 0.011 m 3 )/min
more definitely, with every set of paper spot sam-
ples, a parallel sample of air should be filtered through a circle of paper I in. in diameter or,
through a fibrous filter at a measured flow rate, alternatively, through an area of I in. 2 (6.4
over the same period. The mass of the filtered ma- cm').
terial shall be determined by weighing on a labora- 4.7 Spot Sampling Probe-A small cyclone
tory type analytical balance. This collected material
may be analyzed for carbon content, or for metallic chamber gravity dust trap or funnel covered
and non-metallic constituents by spectrograph1c and with a 325-mesh screen and connected with a
microchemical methods. suitable length of glass tubing, the other end
4.2 Filter Paper, Fibrous-Any suitable of which is attached to the sample inlet head
type of filter paper of relatively low resistance by short rubber connections. This is designed
to air flow and good filtering efficiency. It is to exclude coarse particles of dust larger than
essential to specify the type of paper used in about 40 µm in diameter.
sampling, for example soft, open-texture, 'The AISI Automatic Smoke Sampler designed by
rapid filtering filter paper. Hemeon et al of the Mellon lnslilute and manufactured by
Research Appliance Co .. Pittsburgh. Pa. is a typical ex-
4.3 Filter, Fibrous, for Weight Determina- ample of equipment that has been found 10 be satisfactory
tion and Analysis-Suitable types for this for this purpose. A number of other salisfactory inslru-
ments are also available commercially.
purpose are purified asbestos and prepared ·'A 1106 Fiberglass nash-fired web (Mine Safety Appli-
fiberglass of high filtration efficiency and low ances Co., Penn and Braddock Aves., Pittsburgh, Pa.
15208) with a DOP efficiency of 99.97 percent at 28 fpm
resistance to air flow.• has been found satisfactory.
58
D 1704
4.8 Transmission or Reflectance Meter- equivalent area of clean filter paper. Then
Any conventional laboratory instrument is determine the corresponding value for the
suitable if equipped with a stable light source, spot sample. Record the readings in terms of
preferably having an adjustment for wave the optical density scale of the spectropho-
length to yield substantially monochromatic tometer.
light, and photoelectric optical system suitable 6.4 Calculate the optical density and op-
for determining incident, transmitted, or re- tical reflectance, as follows:
flected light from clean paper in comparison Optical density = log (/ 0 / I)
with spot samples. For convenience, the in- where:
>trument may be provided with an optical / 0 the intensity of transmitted light
density scale. The monochromatic light source through the clean paper, and
shall provide light of wavelength about 400 I the intensity of transmitted light
nm. If monochromatic light is not available, through the sample.
white light may be employed or a set of suit- Log refers to the logarithm to base IO.
able filters may be used to pass a band of Optical reflectance = log (R 0 / R)
light in the region of 375 to 450 nm.
where:
5. Sampling R0 the intensity of the light reflected from
5.1 For information on sampling, refer to the clean paper, and
ASTM Recommended Practice D 1357, Plan- R the intensity of the light reflected from
ning the Sampling of the Atmosphere. 3 the sample.
NOTE 2-The reflectance may be determined in a
6. Procedure for Spot Samples spectrophotometer, preferably at a constant wave-
length of 400 nm with the light reflected at angles
6.1 After assembling the apparatus de- between 3S and SS deg, and the readings taken on
scribed in Section 4, place a roll of clean filter the density scale. It has been shown that there is an
approximately linear relationship between optical
paper tape in the holder and pass the atmos- density and optical reflectance.
phere under test through the sampling probe
at 0.25 ft 3 (0.007 m 3 )/min if the exposed cir- 7. Procedure for Determination of Mass
cular area of the paper under the inlet head is Concentration
equivalent to a I-in. (25-mm) diameter circle. 7. I For collection of samples for mass
For larger exposed areas the flow rates are concentration 6 and analysis, pass the atmos-
proportionately greater. The time cycle, on phere under test through the high volume fil-
which the sampler is operated, is set to avoid tration apparatus, using an asbestos or min-
stains or spots of optical density exceeding eral fiber paper or fiberglass filter previously
0.30. This specification is met usually by set- dried and weighed. The sampling period is
ting the time rotor to yield a spot sample usually 24 h or a sufficient period of time to
every 2 h, although in summer weather under collect from 50 to 250 mg of particulate
normal meteorological conditions it can be 1i1atter.
lengthened to 3 h. Under winter conditions 7.2 Read the flowmeter of the high volume
where consumption of coal or other fuel is sampler at the start and at the end of each
high and smoke conditions may be dense, the sampling period. It is also advisable to take
time cycle can be reduced accordingly. readings at intervals during each test period if
6.2 Measure the air flow frequently enough there is an appreciable drop in flow rate with
with the gas flow meter to establish the av- increasing resistance of the filter. Record the
erage air sampling flow rate. Check the gas average flow rate under the test conditions as
flow meter periodically by calibration with a the flow rate.
standard wet test gas meter. 7.3 For flow rates of about IO to 50 ft~
6.3 At convenient intervals remove the (0.28 to 1.40 m 3 )/min the filter is usually re-
exposed paper tape from the rewind roller and
' The types of high-volume samplers employe? by the
measure each spot for optical density by U. S. Public Health Service, Robert A. Taft Sanitary En-
transmittance or reflectance. Adjust the spec- gineering Center, Cincinnati, Ohio; the U: S. Atomic _En-
ergy Commission, or the International Joint Comm1ss1on,
trophotometer or colorimeter scale to read Windsor-Detroit area, have been found to be satisfactory
100 percent reflectance or transmittance of an for this purpose.
59
D 1704
placed every 24 h. Dry the test filter in an NOTE 4: Example-Suppose measurements with
oven at 105 C for several hours to constant the gas flow meter have established the average
flow rate through each spot area of paper as 0.22
weight, and determine the weight by means of ft' (0.006 m')/min, the area exposed is a sampling
a laboratory analytical balance. circle I in. (25 mm) in diameter or 5.45 x 10-a ft',
7.4 Record the increase in weight of the and the timing spindles of the instrument are set for
2-h samples, then,
filter in terms of micrograms per cubic meter
of air. l = (0.22 x 120)/(1000 x 5.45 x 10-•)
= 4.85 (thousands of feet)
7.5 If desired, analyze the deposit for car-
If a given spot shows a light transmission value of
bon, heavy metals, etc. 50 percent, then,
NOTE 3-For those samples which are collected Optical density = log (100/50) = log 2 = 0.301
in the atmosphere of industrial areas where carbon Cohs/1000 ft= (0.301 x 100)/4.85 = 6.2
and dark tarry materials represent a considerable
portion of the aerosol contaminants, the carbon 8.3 The Km unit is defined as the amount
content of the individual samples may be estimated
by calibration of the optical density values with the of deposit which produces an optical reflec-
carbon content of the deposit as determined by a tance of 0.10 on an area of 1 cm 2 through
representative number of chemical analyses. Where which 1 m' (35.4 ft') of air has been passed.
carbon is the principal coloring material, the color
of the spot will vary with a given weight of deposit This air sample represents 33,000 linear ft
in direct ratio to the proportion of carbon. The (10,058 linear in). Optical reflectance, R, of
carbon values estimated in this manner would be 0. 1 is equivalent to 91 percent reflectance:
valid only for the particular industrial area in which
the calibration was carried out. R =log (100/91) =log I.I= 0.1
Km = (0.1 X 100)/33 = 0.3 Cohs/1000 ft
8. Calculation
8.4 The third unit, Ruds, "reflectance unit
8.1 For comparison purposes with similar
of dirt shade,'' is defined as a deposit which
samples collected in other sampling areas and
produces an optical reflectance of 0.01 due to
under sampling conditions where the area of
10,000 linear ft (3,050 linear m ), of air:
paper exposed and the volume flow rate may
R =log (100/99) =log 1.01 = 0.01
be different, the optical density values for
transmission are converted to a unit scale Ruds = (0.01 x 100)/10 = 0.1 Cohs/1000 ft
called the Coh unit. Similarly, the optical re-
9. Report
flectance values are expressed in Km or R uds
units. The Coh unit is defined as 100 times 9.1 Report the results of the optical tests
the optical density of the deposit, so that an and observations described in Section 6 and
optical density of 0.301 is 30.1 Coh units. the calculations given in Section 7 as follows:
Furthermore, since the intensity of the filter 9.1.1 Date of test,
paper stain is determined by the total volume 9.1.2 Air sampling volume and area of
of air aspirated through the filter and the area stain or deposit. Specify type of paper filter
of the exposed filter, the size of the air sample employed,
is properly expressed in linear units of air, 9.1.3 Size of air sample in linear feet (lin-
that is, the volume of air aspirated divided by ear meters), expressed in multiples of 1000
the area of the filter. For convenience, the linear ft (305 linear m ),
results are reduced to multiples of 1000 linear 9.1.4 Type of optical measurement, that is,
ft (305 linear m) of air sample, so that the transmission or reflectance at specified wave-
soiling for any time period is described in length of light in millimicrons,
terms of Coh units/ 1000 linear ft of air. 9.1.5 Values in Coh units/ 1000 ft of air
8.2 If the quantity of air sampled during sample (or other units) for 1, 2-h, or other
each spot sample is equal to L, in thousands time cycle,
of linear feet, then, 9.1.6 Average Coh units for each 24-h pe-
riod,
l = flow (ft'/min) x sampling time in min
9.1.7 Monthly average Coh units for each
1000 x circular area of spot (ft') hour of the day, or for each sampling cycle of
the day,
optical density x 100 9.1.8 Monthly minimum, maximum, and
Cohs/1000 f t = - - - - - - -
l average Coh units for each month of the year,
60
D 1704
9.1.9 Report the weight collected by filtra- material are discharged into the air. A few
tion on a fibrous filter in a high-volume sam- locations may show a light gray deposit, but
pler or other paper filtration unit as mass such colors are not the rule and should be
concentration in micrograms per cubic meter regarded as a local source of contamination
of air and record the time and date of the or local problem only.
sampling period, and 10.4 The values based on optical density of
9.1.10 Relate the mass concentration the deposit are indicative of the soiling char-
values to the corresponding Coh or other acteristics of particles of less than about 10
units over the same sampling· period. µ,/ m to below 0.1 µ/ m in size and, therefore,
describe the smoke or haze content of the
10. Limitations atmosphere. There is no direct relation between
JO. I The size of the air sample when de- such values and mass concentration of particu-
scribed in linear units is by specification com- lates per unit volume of air, although empirical
pletely independent of the type of filter paper relations may be calculated for a specific case
sampler employed. Thus air may be sampled if the particle size, shape, refractive index,
at any desired rate using any convenient area etc., and other factors are reasonably con-
of paper. stant. These empirical relations should be
10.2 In the evaluation of the spot sample useful in comparing deposits from different
the magnitude <>'f the result is dependent on types of sources and in different locations. In
the characteristics of the filter paper, and on order to have the results of different determi-
the particular method of optical measurement nations at different locations under different
employed, that is, light transmission, light re- conditions comparable with each other, every
flectance, and wavelength of light, whether effort shall be exerted to have the independent
monochromatic or polychromatic. The rela- variables as consistently uniform as possible.
tion between transmittance and reflectance
measurements of spot samples is only approx-
imately linear and depends upon the nature of 11. Precision and Accuracy
the deposit. The ratio of optical densities by 11.1 The precision and accuracy of this
reflectance and transmittance may vary in method depend upon the accuracy of calibra-
progressing from light to heavy deposits. This tion of the gas flow meter, extent of variation
ratio also varies with the wavelength of light of aspiration rate through the paper during
employed so that it is advisable to use a sub- each sampling cycle, and the filtration effi-
stantially monochromatic light source. Trans- ciency of the paper. All these factors should
mittance is more sensitive, especially for light be controlled and the extent of variation kept
deposits. However, variations in filter paper within reasonably narrow limits. The optical
transmittance as a source of error may be measurements can be made with a high de-
avoided by use of reflectance. The use of a gree of precision if a steady source of mono-
white porcelain backing plate for the spot is chromatic light is used, or light from a stable
also helpful in eliminating variations in reflec- source is passed through neutral filters whose
tance due to the paper itself. transmission remains unchanged indefinitely.
10.3 Normal filter paper deposits are es- Periodic checks of the sampling instrument
sentially various shades of black due pri- and photoelectric colorimeter or spectropho-
marily to carbon. They may be tinted brown tometer should be carried out for satisfactory
or reddish in areas where iron oxides or other operation.
REFERENCES
(I) Clayton, George D., Determination of Atmos- Ass., JPCAA Vol 4, 1954, p. 143-150, 171.
pheric Contaminants, Am. Gas Assn., PC-53- (3) Davidson, W. F., and Masters, Warren, "Au-
12, May, 1953. tomatic Dust Sampling and Analyzing Instru-
(2) Gruber, Charles W., and Alpaugh, Edwin L., ments for Atmospheric-Pollution Surveys,"
"The Automatic Filter Paper Sampler in an Monthly Weather Review, MWREA Vol 69,
Air Pollution Measurement Program.·' Air 1941, p. 257.
Repair, Journal of the Air Pollution Control (4) Hall, S. R., "Evaluation of Particulate Con-
61
D 1704
cc1H rnti(ln> wilh Collect inf! Appar<il U\, " Ana· tinuou, Detc1 mi nation of Air Cont:11uinanb, · ·
lyrical Clzemi.11r.1•, ANCHA Vol 24. 19)2, p. l'roceedi11gs. Am . Soc. Testing Mat~ .. Vol 53.
996. 1953, pp. 1136 lo 1159.
IS) I leinwn. W . C. L.. · ·]n,truments fo1 Auto· (9 J Katz, .\.f .• Sanderson. H. P .. and Ferguson,
rnal ic Ai1 Pollution :v!ta>UI cmell\\," Proceed· M. B.. "l'valual ion of Pai ticulate Mattei in
ings, Air Pollmion ~nd Snwkc P1cvcntion Air Pollution Studic>," ..Jnalytical Ch,'mistrr.
Assn ., p. 115, 19.~ I. ANCHA Vol 30, 1958.
(6) Hcme<ln, W. C. L.. Srnscnbaugh, J. D .. uml ( 10) Mcetham, /I.. R .. "Atrnosphc1ic Pollution ...
ll ai ncs, Jr., Ci . F.. "Design aml U sc of Equip· London, Pergamon Pre" Ltd .. 1952.
mcnt in Air Pollution Mcasurcmcnb." l n.<tm· (I I) Shaw and Owen,, "The S1uokc Prohlem of
ments, INSTA, Vol 26, p. 566, Ap1il. 1953. G real C 1ties." London. Constable and Co.
(7) Hemeon. W. C. L., 1laine<;, J 1.. George F .. Ltd., 192).
und Ide. 1larold M .. · · Dctcrminuti(in llf 11 an 112) Thoma,, M. D.. "The Present S1<11 us of I he
und Smoke C'nnccntnitions by Filler Paper Development llf I nstrumcn1ation fo1 the St 11dy
Sa1uplcrs,'' Journal uf the A it Pol/11tio11 Con· ol' Air Pollution, '' Stanfo1d Research Inst.,
tro/ Ass11., JPCAA, Vol 3, No. I, p. 22. Au· Los Angeles. 1952. pp. 16 to 2.'\.
gnst, 1953. ( 13) ''on 13 rand , E. K.. ''Applications of a Pon a blc
18) Kati., M .. untl Clayton. G. 0 .. "lnstrnmcn- Continuous S1uokc Re~ordc1," Mt•ch011ical
Wtion and Analytical Tcchnir.iues for lhe Con· Engineering. M EEN A Vol 72. 19':>0, p. 479.
62
~~l~ D 1704
Timing
Mechanism
Inlet
Pump Motor
By publication of this standard no position is taken with respect to the validity of any patent rights in connection there-
with, and the American Society for Testing and Materials does not undertake to insure anyone utilizing the standard
against liabilitr for infringement of any lel/ers Patent nor assume any such liability.
63