Enter “E”
Press “+” and select Basic Function & press “E”
Press “+” and select reference parameter and press “E”
Select Cal. Ref Density and press “E”
Now density is +0.5000 e-3
Now press”+”
Access code shall be asked. Which is 83. Now cursor will be on first digit. Press + to select number 8 as
first number and press “E” . Now press + to select number 3 as second digit and theren press “E” and
Now press “+” and “-“ together to come out. Press together many times (4 to 5 time) till you can view
main screen.
FOR EXTRA DENSITY PARAETER CONFIGURATION
Enter “E”
Press “+” and select Basic Function & press “E”
Press “+” and select MODBUS 485 and press “E”
Select CONFIGURE and press “E”
TAG NAME and press “E”
ADDRESS and press “E”
BOUD RATE and press “E”
PROG MODE and press “E”
STOP BITS and press “E”
1-8-3-2 and press “E”
DELAY and press “E”
Write prob ofer ma
Scan rey 1 to 6
7th press “+” set to 02015 sing + & - key.
If Access code shall be asked. Which is 83. Now cursor will be on first digit. Press + to select number 8 as
first number and press “E” . Now press + to select number 3 as second digit and theren press “E” and
again press”E”. Message will appear that u are in programming mode.
If Access code shall be asked. Which is 83. Now cursor will be on first digit. Press + to select number 8 as
first number and press “E” . Now press + to select number 3 as second digit and theren press “E” and
again press”E”. Message will appear that u are in programming mode.Now change any number by using
+, - and “E”.
After access code press + both option shall be available for yes & no.
Now enter
Pres “+” & “-“ simultaneously contionusly till you get main screen.
Select System and press “E” (Press + for up & down for options)
Press “+”
If Access code shall be asked. Which is 83. Now cursor will be on first digit. Press + to select number 8 as
first number and press “E” . Now press + to select number 3 as second digit and theren press “E” and
again press”E”. Message will appear that u are in programming mode.Now change any number by using
+, - and “E”.
Cabling
4,5 –
6,7-
8- ground
9,10- temp
11, 12 – density
41,42 –
Troubleshooting on 3.08.15 =
Following error was observed, Fluid temp max, range freq out, noise limit CH0, Range freq out.
There are three touch-buttons provided just below the display of mass flow-meter. ‘-/+’ are for scrolling
up/down or pervious/next, and ‘E’ is for enter command. Stepwise procedure for resetting the totalizer
reading to zero is given below:
Press ‘E’
Press ‘+’ to scroll down till you reach Totalizer Option
Press ‘E’
Press ‘+’ to scroll down till you reach Handling Totalizer
Press ‘E’
Reset All Total is displayed
Press ‘+’
Option for process code (****) will appear
Press ‘+’ and digit ‘0’ will appear in place of first *
Digit value can be changed (0-9) by pressing ‘+’
Process code is 83
Press ‘+’ till first digit is ‘8’
Press ‘E’ to go to next *
Press ‘+’ till second digit is ‘3’
Press ‘E’ to go to next *
Press ‘E’ to go to next *
Press ‘E’
Both Yes and No options will be displayed
Press ‘+’ to scroll to Yes
Press ‘E’
Are You Sure option will be displayed
Press ‘+’ to scroll to Yes
Press ‘E’
All totals will be reset and only No option will be displayed
Press ‘+’ & ‘-’ simultaneously to go back till you get main display showing MFM reading
SIMULATION OFF
Main menu-Output -Pulse frequency output 1- Configuration- Frequency - Then press “E”