Академический Документы
Профессиональный Документы
Культура Документы
Learning Handbook
TI 12B00A20-01E
Table of Contents
1 Introduction.................................................................................................................4
2 Basics on pH theory.................................................................................................... 4
3 Basics on ORP............................................................................................................ 28
www.yokogawa.com/us
3.8 Standard Maintenance and Calibration..............................................................................................33
8 Appendix 1:
Chemical Compatibility.........................................................................................................................60
9 Appendix 2:
Definitions...............................................................................................................................................64
10 Appendix 3:
www.yokogawa.com/us
1. INTRODUCTION
Measuring pH/ORP is very common, but taking true measurements and correct interpretation of the results is not
1. INTRODUCTION
self-evident. Certain effects can potentially cause problems if not taken into consideration.
The purpose of this book is to provide a comprehensive understanding of pH/ORP measurement and how to achieve
reliable results. Basic information on the principles of measuring pH/ORP, the construction of the sensing elements
and their basic use in process applications are provided.
A part of achieving accurate and reliable pH/ORP measurements requires sufficient and correct maintenance and
storage conditions. Prevention of common errors during maintenance and storage, as well as consistent detection of
loop failures is important. This book describes how these can be avoided and how failures can be detected.
This book is accompanied with a frequently asked question and answer section, as well as an appendix that includes
helpful information like a Chemical Compatibility Table and a Liquid Application Data Sheet, which can be used to
describe the user’s application.
2. Basics on pH theory
T
2.1 Concept of pH
he Danish scientist Sørensen other factors, such as:
defined the concept of pH as
follows: ■ The activity of other ions present
in the solution
pH equals the inverse of the logarithm ■ The temperature of the solution
to the base 10 of the hydrogen ion
concentration, as shown by the ■ The character of the solution.
pH = -10log aH+ = pH ) 3
(2) Born in Havrebjerg, Denmark,
The activity of the hydrogen H+ + H2O ↔ H3O+ Sørensen was a Danish chemist,
ions is not always linear with famous for the introduction of the
the concentration, since this concept of pH, a scale for measuring
activity is not only affected by the Consequently, a more correct acidity and basicity. From 1901 to 1938
concentration of ions, but also by definition for pH is: he was head of the prestigious Carls-
berg Laboratory, Copenhagen.While
working at the Carlsberg Laboratory
pH = -10log aH3O+ (3) he studied the effect of ion concen-
Note 1. The notation -10log .... can also tration on proteins,and because the
be written p ....
For clarity, the notation H+
concentration of hydrogen ions was
will be used in the book as the
particularly important, he introduced
Note 2. See appendix 2: Definitions. hydroxonium ion.
the pH-scale as a simple way of
Note 3. See Chapter 2.8: Buffer expressing it in 1909.
solutions.
www.yokogawa.com/us
2.2 the pH Scale
Your starting point for the pH scale
is pure water, which is said to be
neutral. Water dissociates1) into:
or:
2. BASICS ON pH THEORY
-log Kw = pKw = -log [H+] + -log
[OH-] = 14 log 10 (6)
pH Table
Some examples of the difference in These can be compared with the
the pH value of various liquids, foods pH values of common chemical
and fruit are shown in Figure 2.2b. compounds dissolved in water.
A L K A L I NI TY
Fig. 2.2b.
www.yokogawa.com 5
2.3 Measuring the pH Scale
The pH value can be measured by
different methods, e.g.:
A. Colorimetric pH measurement
B. Potentiometric pH measurement
6 www.yokogawa.com
2.4 PRINCIPLE OF POTENTIOMETRIC pH MEASUREMENT
The principle of potentiometric pH Nernst found that a potential The “normal potential” is the
measurement can be explained by difference occurs between a metal potential difference arising between
Nernst’s law. object and a solution containing metal and solution when this
ions of the same metal when the solution contains 1 mol Mn+/litre.
object is immersed in the solution.
Since the behavior of the gas
The potential difference E, caused hydrogen has a certain degree of
by the exchange of metal ions conformity with a metal (both have a
between metal and liquid, was positive ion formation), Nernst’s law
defined by Nernst as follows: can also be applied to a “hydrogen
electrode”1) immersed into a
solution containing hydrogen ions.
E = Eo + RT. In [Mn+] (7)
2. BASICS ON pH THEORY
nF The formula can be re-written as
follows:
R = Gas constant (R=8.314J/mol.K)
E = Eo 2) + RT. In [H+] (volt)
F = Farady number (F = 96493 C/ nF
mol.) or:
Walther Hermann Nernst
(1864-1941)
n = Valency of the metal With the constants:
Born in Briesen, West Prussia, in
1864. He spent his early school years
(Gymnasium) at Graudentz, and sub-
[Mn+] = Metal ion concentration E = Eo + 0,059 Ln [H+] (volt)
sequently went to the Universities of
Zurich, Berlin and Graz (Ludwig Boltz- T = Absolute temperature in Kelvin
mann and Albert von Ettinghausen),
studying physics and mathematics.
Eo = “Normal potential”
www.yokogawa.com 7
La+++ 115x10–10 m
Ba++ 1,35x10–10 m
Ca++ 0,99x10–10 m
2. BASICS ON pH THEORY
Cs+ 1,69x10–10 m
Li+ 0,6x10–10 m
O– – 1,4x10–10 m
Si++++ 0,4x10–10 m
When immersed in aqueous solutions, This depends on several factors, The movement of the ions will affect
all types of pH glass have the such as the composition of the glass the neutrality of the gel-layer. As a
particular property to exchange the and the temperature in which the result, a voltage will be developed
metal -ions of the glass texture against sensor is immersed. preventing the further transport of
the H+ ions in the solution. H+ ions.
The voltage development across
Since the H+ are bounded to a H2O the glass membrane is generally The value of the voltage depends
molecule and not free, the texture explained by means of the phase on the concentration of the
of the siliceous acid will be defound limit potential theory 1). hydrogen ions in the solution. Since
during the exchange by the bigger this voltage cannot be measured
H3+ -ion. After reaching the equilibrium the directly, it will be necessary to
hydrogen concentration (=activity) add a pH independent reference
As a result of this reaction a so-called outside the glass and inside potential in the measuring circuit.
“gel-layer” will be developed on the the gel-layer are equalized, and This addition allows measurement
surface of the glass membrane. This consequently, no transport of H+ of the potential differences across
gel-layer is the equivalent of the metal ions occurs. The voltage across the the glass membrane.
in the Nernst’s theory, and is therefore glass membrane is 0 volt.
essential for the functioning of the
glass sensor. If the concentration of hydrogen Note 1: For clarity, other theoretical
ions in the two phases differs from explanations like the theories
After one or two days the condition the concentration in the solution, of the adsorption potential,
reaches equilibrium and the resulting a transport of hydrogen ions takes membrane potential and
statistic mechanic will not be
gel-layer has a thickness between 10 place. explained.
and 40 nanometers.
8 www.yokogawa.com
2.5.1 COMPOSITION OF THE GLASS ELECTRODE
Normally the glass electrode has a membrane designed as follows.
bulb shaped membrane of specific
“pH glass” that is “welded” to the -B
all Shape (Shockproof)
Gold
2. BASICS ON pH THEORY
Consequently, it is impossible for
the buffer liquid to penetrate the - Flat shape
Contact
other parts of the electrode. This design is used in combined
sensors for application where
spring
Gold Nernst found that the combination A constant flow of electrolyte from
O connector of a metal and its insoluble salt in a the electrode prevents poisoning of
salt solution produces a constant mV the electrolyte around the reference
potential. pin.
Glass tube may be poisoned by the penetration the pH dependent potential of the
of “unwanted ions” in the salt glass electrode very accurately.
solution. To overcome this problem
Electrolyte
Diaphragm
www.yokogawa.com 9
2.5.3 THE MEASURING CIRCUIT WITH A GLASS AND A REFERENCE ELECTRODE
E3
as dual amplifier system. The
E4
diagrams show the potentials
which a ffect the final potential
R elecrolyte
R diaphragm
2. BASICS ON pH THEORY
E1
R
E5 E4 E1 E2 E3 liquid
The total sum (Et) of these potential Cz- = concentration of the salt solution The ideal conditions described
differences is measured by the pH- above cannot always be completely
Analyzer: The equation (8) will then be realized in practice. A small
Et = E 1 + E 2 + E 3 + E 4 + E 5 (8) simplified to: potential difference may exist
when the glass and the reference
Potential (E6) is the potential of pH Et = E1+ E2+ E5 (9) electrode are both immersed in a
electrode against solution ground: liquid of similar properties and pH
With correct selection of the
value to the electrolyte.
E6= E1-E2 -E3 electrolyte used in the reference
electrode and a good flow through
Potential (E7) is the potential Et = 0.05916 (pHinner- pHouter) + Easy (13)
the liquid junction, the potential
of Reference electrode against difference E5 can be neglected, so that This potential difference is called
solution ground:
the asymmetric potential of the
Et = E1 + E2 (10)
E7= E4 + E5 measuring system.
E= ENe
o
+ RT • In LmZ properties and consequently, the Note: 2. Errors resulting from Easy
F Cz- measured E1 now only depends on the are compensated for during
in which potential difference between the glass calibration using buffer solutions.
membrane and the process liquid.
Lmz = solubility product of sparingly
soluble salt
10 www.yokogawa.com
2.5.4 MORE INFORMATION ABOUT THE pH GLASS ELECTRODE
The correct selection of a glass ■ sensitivity of the glass electrode To facilitate this process, it is
electrode for a particular application (mV/pH) necessary for the pH sensitive glass
can only be made if details of the ■ alkaline error membrane to be “conditioned” by
components of the measuring loop ■ acid error allowing it to absorb a film of water or
and their significant properties are ■ chemical resistance of the glass gel-film.
known. The following points will be membrane
considered in detail: ■ electrical resistance of the glass Conditioning is achieved by soaking
membrane the electrode in water for a minimum
■ selection of the glass membrane of 24 hours.
2.5.4.1 Selection of
the glass membrane The selection of the correct type of
135
2. BASICS ON pH THEORY
100
complete pH measuring
70 loop. pH sensitive ■ “G” glass - This is used for the
glass has the particular membranes of electrodes in pro
property that alkali cesses where the nominal pH value
Temperature (ºC)
Note 1: T
he diagram shown in Figure2.5.4 is intended to assist with Note 2: Glass electrodes manufactured by Yokogawa are
selection of the most suitable type of glass electrode in “preconditioned” and may be used immediately without
conjunction with the application range shown for each type. soaking. To form and maintain the gel-film, the sensitive glass
The range of any particular glass type also depends on the bulb is protected with a rubber containing a small quantity of
membrane thickness. Three different thicknesses are available. water which forms a wet pocket for the membrane.
2.5.4.2 Sensitivity of the glass slope, at the analyzer. The decrease not reach an optimal value until the
electrode in sensitivity is usually caused by electrode has been conditioned for a
fouling of the glass membrane. It is number of hours.
The most important requirement of the greatest importance that the
in any electrode system for pH electrode is properly cleaned before When the decrease in sensitivity
measurement is that the actual mV/ “buffering” and adjustment for is caused by aging of the glass
pH ratio generated is as close as sensitivity is made. membrane the electrode can be re-
possible to the theoretical value. activated by etching the surface of the
If the electrode has been stored dry glass membrane.
The potential generated by a glass for a long period the sensitivity will
electrode is given by equation: This should be done by immersing the
electrode for 10 seconds in a solution
Note: The electrode must be cleaned
E=
RT. . 2,303 (pH – pHouter) mV carefully before and after of vinegar (1 mol.) and potassium
nF
fluoride (1 mol.). Ratio 1:1
inner
activation. (For cleaning and re-
At a temperature of 25°C the equation activation see: “Direction for use”
becomes: enclosed with each electrode). 2.5.4.3 Alkaline of the glass
Frequent re-activation decreases electrode
the life of the electrode.
E = 59,16. 298 . (pHinner – pHouter)mV 298
273+25
In addition to a correct mV/pH ratio,
In the equation pHouter is the pH value it is essential that the generation of
of the liquid at the outer face of the potential difference is only
glass membrane and pHinner is the influenced by the activity
Sodium Error
0
www.yokogawa.com 11
In practice, the alkaline error measurements in either strong acid Bulb shaped (shock-proof):
is often caused by sodium and or strong basic liquids. In aggressive 50-100 MW
consequently the term “sodium solutions, a heavy duty electrode
error” is also used to describe this with a thick, dome-shaped glass Dome shaped (heavy duty): 120-200
effect. membrane is preferable. MW
The alkaline error can be 2.5.4.5 Electrical resistance of 2.5.4.6 The response time of
considerably reduced by making the glass membrane the glass electrode
certain additions to the pH The response time of a glass electrode
Since glass is a good insulator,
sensitive glass, which improves indicates the ability of an electrode to
potentiometric measurements
the selectivity of the electrode. In follow accurately any changes in the pH
cannot be obtained with normal
Fig 2.5.4.3 the alkaline error of the value. The response time is normally
glass, and constituents must
different types of glass at varying defined as the time taken to reach 63% of
be added which will reduce the
pH values are shown. the value of a step change in input. Since,
membrane resistance below
1000 MW to minimize the effect in practice, the response time depends
As well as affecting the selectivity,
of electrical disturbances on the on a lot of factors e.g.: the reference
2. BASICS ON pH THEORY
12 www.yokogawa.com
2.5.5.2 Reference system a silver wire electrolytically coated This paste is sealed into a tube by
with silver chloride. This metal - means of a plug wadding soaked in
Generally, the reference system metal salt combination is dipped in KCI.
used in reference electrodes, are: a potassium chloride solution (KCI).
This reference assembly is similarly
Silver/Silver chloride-Potassium A second type of construction for dipped in a KCI solution.
chloride: (Ag/ AgCI-KCI) this system consists of a silver wire
dipped in a paste of silver chloride,
This reference system consists of silver and potassium chloride.
2. BASICS ON pH THEORY
Silver chloride paste (AgCI) in 1 molal KCI +230 mV ±5 mV upto 120°C
2.5.5.3 Junctions of
the reference 1 2 3 4
Ceramic Ceramic PTFE Sleeve
The selection of the correct type of
junction of a reference electrode
depends on the process conditions
under which the electrode has to
function.
1.Ceramic junction.
2.Ceramic junction.
Fig. 2.5.6
3.P.T.F.E. junction.
because of its larger flow surface. diaphragm at equal speed
4.Glass sleeve capillary element. The sleeve can be easily cleaned
by first moving the ground ring ■ having a low electrical resistance
The purpose of the junction is to upwards and then wiping the The most common electrolyte used
maintain contact between the ground faces. in reference electrodes are:
reference system in the electrode
and the process liquid. Non-flowing reference electrodes ■ 1moal KCL solution (with or
with a porous P.T.F.E. junction can without gel)
When selecting the correct also be used in many dirty liquid
junction, consideration has to be ■ 3.3 molal KCI solution
applications. The dirt resistant
given to ensure that the process properties of P.T.F.E. will prevent ■ saturated KCI solution.
liquid does not penetrate into the complete fouling of the diaphragm.
electrode causing poisoning and
a consequential unstable liquid
2.5.5.4 Electrolytes in the
junction potential.
reference electrode
With the first two types of junction
The electrolyte in the reference
listed above, the KCI solution flows
electrode must satisfy the following
slowly into the process. The flow
requirements:
rate is dependent on the over-
pressure in the electrode and on the ■ chemically inert and neutral
process temperature.
■ no reaction with the process
The electrolyte flow rate increases liquid
with increasing temperature. ■ having a constant activity of ions
For use in very dirty liquids a glass ■ equitransferent, i.e. the ions of
sleeve capillary element is preferred the electrolyte must pass the
www.yokogawa.com 13
2.5.5.5 Pressure compensated The pressure inside the bellows equals
reference electrode the pressure outside and only the
elasticity of the bellows itself causes
In processes with pressure the over-pressure, which results in a
Reference variations, the composition of the flow of electrolyte.
system electrolyte may change as a result of
process liquid penetration into the When the bellows are fully “expanded”
electrode. the electrolyte is exhausted and
refilling is required. The bellows must
Any change in composition of the be compressed before refilling.
electrolyte may cause a measuring
error or even poisoning of the
reference system of the electrode.
14 www.yokogawa.com
2.5.6 Construction of the Temperature ELECTRODE
2. BASICS ON pH THEORY
system, whose purpose is to provide
Tube automatic compensation for the
effects of temperature variations on
the measuring system.
www.yokogawa.com 15
2.6 The effect of temperature
Nernst equation RT
13
F
14
It has been previously stated that at 100°C for pure water As the effect of temperature on any
the pH value of pure water at 0°C process liquid is highly dependent
differs from the pH value at 100°C. pK = 12,24 or pH = pOH = 6,12 on its composition, it is not possible
to accurately compensate for this
effect automatically.
16 www.yokogawa.com
2.6.3 Temperature effect on the application range of the glass electrode
MΩ
Furthermore, the chemical resistance 10
of the glass membrane is temperature
dependent and correct selection is 1
important.
0.1
A third factor is the membrane
resistance of the glass electrode. This
2. BASICS ON pH THEORY
increases considerably at lower process
0.01
9.5
pH at Reference Temperature
80 ºC
9
8.5
0 ºC
8
7.5
T = Tref = 25ºC
7
6.5
5.5
5
5 6 7 8 9
pH at Process Temperature
Fig. 2.6.4
www.yokogawa.com 17
2.7 Isolation resistance
In view of the relatively high to the electrodes as possible. At depending on the type of junction.
resistance of the pH sensitive all times the connections between
glass membrane, it is necessary electrodes and analyzer should be At higher resistance values the
to use a analyzer with a high input kept dry. The insulation resistance sensitivity of the measuring system
impedance. This impedance must decreases considerably when any will be reduced and may cause an
be at least a factor of 1000 higher moisture is present. instability of measurement.
than the membrane resistance
(the resistance of the reference The insulation resistance of the
electrode is much lower and can be reference electrode is less critical
neglected). as its resistance with respect to the
measuring liquid is much lower. Note 1:pH measurements in low
Insulation and screening of all Generally, an insulation resistance conductivity liquids with a
cables and connections between of 107W is adequate1). analyzer with two high input
2. BASICS ON pH THEORY
the measuring electrode and the impedances for both the glass
analyzer must be of the highest The resistance between reference and the reference electrode
require a good insulation
order. In industrial applications the electrode and liquid is usually resistance.
analyzer should be installed as near between 1 and 10kW at 25°C,
18 www.yokogawa.com
Temperature dependence an automatic adjustment for For accuracy it is recommended that
temperature variations will be a buffer should not be used for more
The temperature dependence of the
done. Any stated pH value is than a month after opening. Buffers
pH of a buffer solution is generally
only meaningful if the measuring should be stored in tightly sealed,
specified in terms of measured
temperature is also specified. preferably air-tight bottles made
pH values at certain discrete
of polyethene or borosilicate glass.
temperatures.
Be Aware Buffers should not be returned to the
Many buffer tables are pre- bottles once removed.
Buffers with a pH above 7 are
programmed in Yokogawa particularly sensitive to atmospheric
Analyzers. So if during calibration CO2. Buffer showing any sign
the temperature compensator of turbidity must be discarded
is immersed in the buffer liquid, immediately.
2. BASICS ON pH THEORY
2.9 PERIODIC MAINTENANCE AND CALIVON OF pH SENSORS
2.9.1 Why Is Maintenance The validiation control chart (Fig. These effects usually happen
Needed? 2.9.1) shows that frequency of slowly; therefore, calibration
maintenance of your measurement should not be necessary more
The selection of pH electrodes depends on the required accuracy. frequently than about once a
and holders (fittings) is based on A pH measurement was checked month in typical general purpose
the demands of the application daily in buffer solution without applications. If more frequent
where they will be used. The desire adjustment. This chart shows calibration is needed, it is usually
is to achieve an accurate, reliable that the reading measurement is because the cleaning process was
measurement with a reasonable swinging around the calibration not effective, the calibration was
life expectancy while minimizing value. To guarantee an accuracy not well executed, the pH readings
the required routine maintenance. of 5% you have to calibrate the are temperature dependent or
When a quality pH sensor system measurement at a minimum the wrong electrodes have been
is undamaged, clean and properly twice a week. If accuracy of 10 % selected. If a film remains on the
calibrated, it will provide a is accepted you can prolong the pH sensor after cleaning, then a
measurement that is accurate frequency to once in two weeks. measuring error can be interpreted
and reliable. This sounds simple as a need for re-calibration. Since,
enough, but ensuring the system is Periodic calibration is necessary to these changes are reversible with
clean and calibrated will sometimes ensure the highest measurement proper cleaning, it is a key step in
involve a significant amount of accuracy. Calibration adjusts for the maintenance process.
maintenance. The effect of dirty or the aging of the sensors and the
faulty electrodes can be anything non-recoverable changes to the Note: The periodic maintenance advice
from slow response to completely electrodes that take place. that follows is intentionally
erroneous measurements. general in nature because pH
sensor maintenance is highly
application specific.
Calibration Value
Days 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29
www.yokogawa.com 19
2.9.2 General maintenance procedure; clean – check – calibrate
Clean: electrodes can be put back on line. If with bottom of the calibration
The starting point for any pH not, then proceed to calibrate. vessel.
maintenance activity is to make sure
Calibrate: 2. A
djust the meter reading to the
the electrode system is properly
pH value of the buffer solution
cleaned. A variety of cleaning A pH measurement loop requires
according to the method selected
solutions can be used depending on regular calibration of the electrodes
(automatic or manual).
the coating effects of the process on to compensate for their aging and
the electrodes. Typically a 5 to 10% deterioration. Typically this is done 3. Rinse the electrodes with tap
solution of HCl works well. about once a month. Always use water to remove the traces of
fresh buffer solutions to avoid the the first buffer. Then, immerse
1. Rinse off the sensor in just tap possibility of introducing errors from the electrodes in the second
water to remove any heavy contaminated or aged solutions. buffer solution (e.g. 4.01 or 9.18,
process coating. Buffers supplied as liquids have a depending on the measuring
limited shelf life, especially alkaline range) and repeat step 1.
2. I mmerse the electrodes in the
2. BASICS ON pH THEORY
the electrodes in a second (different the measurement stabilize. For must be connected. The
buffer solution temperature
value) solution and allow them to a good calibration, avoid any must be within the technical
stabilize. If the readings in both contact of the glass membrane specification limits as
cases are within e.g. +/-0.1 pH, the indicated on the label.
0
close to pH 7.
pH 7 pH
The line is shifted from “1” to “2”
laterally so it passes through the
id zero point. Next, perform a span id
ea
adjustment
l using a standard ea
solution with a span pH (typically
l
20 www.yokogawa.com
2.9.4 One-point calibration
2. BASICS ON pH THEORY
ib r ra e
grab sample. The other is by using
ra tio
tio n
a sample solution being measured. n
The pH of the sample solution
is manually determined by, for
example, a portable pH analyzer,
and then the pH analyzer is adjusted
so that it reads the same pH value. Fig. 2.9.4 One-point calibration
www.yokogawa.com 21
2.9.6 Calibration Parameters
22 www.yokogawa.com
2.10 Differential Electrodes
A pH sensor measures the voltage the process may contain ions that The benefit of using a caption
that the pH membrane measures poison the reference element, and reference electrode, that it has NO
as function of the pH value of the these ions penetrate the element junction; there is NO path from the
process sample. This voltage is then by diffusion especially at high process to the internal element;
compared with the mV output of a temperatures. so NO poisoning can occur. Also
reference cell that is independent since there is NO junction, there is
on the pH value of the sensor. In a This means most of the problems NO plugging or coating problems
conventional pH measuring sensor are not related to the pH measuring to worry about; and there is NO
this reference electrode is like electrode, but almost exclusively to electrolyte depletion problem
described before a Metal/ Insoluble the reference electrode. Therefore, because there is NO electrolyte.
metal salt/salt solution combination the solution of the problem must
inserted in an electrolyte solution be sought in alternative ways of
(or gel) that is separated from generating a reference voltage
the process with a junction. This that is stable over time and is
2. BASICS ON pH THEORY
ensures that the electrolyte independent on the pH value.
composition does not vary with
process changes and that the mV The best solution is to have a
output does not change either. hermetically sealed reference
electrode, like a differential
However, the reference electrode sensor. A differential sensor
is still in open electrolytic still incorporates a pH sensitive
contact with the process and the measuring element to detect ph
composition of the electrolyte changes. However, for generating
changes over time by diffusion, so a stable reference voltage now a
regular calibration is required. Also, cation sensitive measuring element
is used.
www.yokogawa.com 23
2.10.2 How a salt-sensitive reference works, and the advantages
Fig. 2.10.2a
mV vs SCE
0
sensor, the analog signal to
the output of the reference
-50
mV
Fig. 2.10.2b
reading.
0 1 2 3 4
Fig. 2.10.2c
24 www.yokogawa.com
The graph (Figure 2.10.2d) shows
that output signal from the pH glass 500
element and the salt reference
pH glass and
element. The graph shows when
salt reference 400
the output signal of salt reference
elements
300
is stable and independent from pH glass element
changes of pH concentration and 200
can be used for pH measurement. 100
mV
0
pH
-100
600 g / l
-200
60 g / l
salt reference element
2. BASICS ON pH THEORY
6 g/ l -300
0, 6 g / l
-400
0 2 4 6 8 10 12 14 16
pH
Fig. 2.10.2d
Fig. 2.10.2f:
Differential pH measurement at 20Differential
ºC pH
pH- at
SCE60 g/l NaCl
pNa-SCE
mV
diff
www.yokogawa.com 450 25
400
0
0 1 2 3 4 5 6 7 8
pH
400
The sensor output is linear
over almost the complete pH
350
range with only a slight loss 300
of sensitivity below 1 pH. In
this scenario, the Asymmetry 250
Potential is now -50 mV at pH 7, 200
instead of the traditional 0 mV. In
this situation it is good practice 150
to put the ITP setting of the pH 100
analyzer to 5.6 pH to reflect
2. BASICS ON pH THEORY
-100
0 1 2 3 4 5 6 7 8
pH
NaCl. 350
-50
This can be seen as a problem
because in most applications this -100
concentration is not constant.
0 1 2 3 4 5 6 7
pH
However, there is no reason for
concern because a change in salt
concentration of +/- 25% only
causes an offset of 5 mV, or an
Fig. 2.10.2h: Differential pH at 6g/l NaCl
26 www.yokogawa.com
2.10.3 Calibration
Normally the pH standards that are Refer to the analyzer Instruction Manual for details.
pre-programmed in the pH analyzer
can be used for calibration, with the Standard buffers can be used for functionality check of sensors, however
pH analyzer set to “AUTOCAL”. The if they are used for calibration, the Asymmetry and Slope values will be
SC24V however is a differential pH incorrectly calculated.
sensor which needs pH buffers that
have the same ionic strength because Specific buffers for sodium-based reference are available:
the sodium reference will change as
pH 2.00 (Ionic strength – 1 mol NaCl)
the ionic strength changes. These
n
specific buffer solutions and the pH n pH 9.00 (Ionic strength – 1 mol NaCl)
analyzer set to “MANUAL” calibration.
2. BASICS ON pH THEORY
2.10.4 Buffer calibration recommended), stir the electrode measurement stabilize. Adjust the
for a few seconds to ensure that pH reading in the pH analyzer to
To calibrate the SC24V sensor, the gel layer of the pH glass the value indicated on the bottle
two buffer solutions with known membrane is completely wet (in this case 4.00). Quit the manual
pH values are required. It is by buffer solution and all other calibration routine by accepting
recommended that one buffer fluids are removed. Then keep the new calculated calibration
solution has a value near to pH 7.00. the electrode still and let the data of zero and slope.
Depending on the process value measurement stabilize. Adjust the After calibration, re-install the
to be measured, the second buffer pH reading in the pH analyzer to sensor into the process.
solution should be either acidic the value indicated on the bottle
(below pH 7.00) or alkaline (above pH (in this case 7.00). Go to solution 2
7.00). Available buffers are: pH 2.00, to do a two-point calibration;
pH 4.00, pH 7.00 and pH 9.00.
4. Rinse sensor thoroughly with tap
The following is a very general water;
2-point manual calibration Note: It is important to understand it
procedure with buffer solution: 5. Immerse the sensor in the is possible the SC24V differential
second buffer solution (pH sensor may not show the correct
1. Clean the sensor using a 5% 4.00 is recommended), stir the pH value after buffer calibration.
solution of HCl; electrode for a few seconds
The reason is that the ionic
strength of the buffers is 1 mol
to ensure that the gel layer NaCl, which can deviate from the
2. R
inse sensor thoroughly with tap
of the pH glass membrane is ionic strength of the process. For
water (DO NOT use demineralized
completely wet, soaked or an accurate pH reading, an extra
water); 1 point process calibration has
saturated by buffer solution and
to be done to compensate for
3. Immerse the sensor in the all other fluids are removed. Then differences in ionic strength.
keep the electrode still let the
first buffer solution (pH 7.00 is
www.yokogawa.com 27
3. Basics on ORP
3.1. Concept of ORP With redox reactions, we speak oxidation-reduction potential.
in terms of the strengths of the
Have you ever wondered what gives While a pH value can be obtained
oxidizing and reducing agents.
a sapphire its deep blue color? within seconds, a stable ORP value
Oxidizing agents have the capacity
It comes from a simple REDOX can take up to several minutes,
or potential to acquire electrons
reaction between the titanium (IV) if not hours, to reach the final
and become reduced. Reduction
and iron (II) impurities. The REDOX equilibrium due to the type of
means the gain of electrons by an
reaction can be seen as: reactions and their reaction rates.
atom, leading to a decrease in the
The ORP measurement behavior
oxidation state of the element.
is strongly influenced by the metal
Ti4+ + Fe2+ → Ti3+ + Fe3+
surface condition. For example, a
Cu2+ + 2e− → Cu
However in order for the reaction new, unconditioned ORP electrode
to occur energy must be supplied. Reducing agents donate electrons will show different values than
This is easily accomplished when and therefore become oxidized. an ORP electrode that has been
ordinary white light passes through Oxidation means the loss of conditioned and considered in use.
the crystals. The reaction between electrons from an atom, leading to
the titanium and iron absorbs the an increase in the oxidation state of 3.2. The ORP Scale
red, orange and yellow light regions the element.
of the spectrum to fuel the REDOX A simple working definition for ORP
reaction, thus allowing only the blue Fe → Fe2+ + 2e− is a solution’s capacity for electron
light to pass through the crystals, transfer known as oxidation or
resulting in the deep blue color seen Since reduction and oxidation reduction, given in millivolts. The
in sapphire crystals. reactions occur simultaneously, the measurement of ORP is the reading
formulas for the two half reactions of the voltage potential between
3. BASICS ON ORP
28 www.yokogawa.com
Soda Tap Water Bottled Water Vitamin C Green Tea Cod Liver Oil Living Water
3. BASICS ON ORP
Soda is known to have a pH value the Hydrogen ion concentration in Nernst formula:
of around 2.00; ;and, as shown solution, ORP only provides relative
here, the respective ORP value for measures of chemicals and cannot
discriminate one from another. Even Eh = E0 + RT/nF log A ox/A red
soda is approximately 400 mV. This
indicates that a positive mV reading though ORP is non-ion specific, it is
(or below a pH 7) is associated with an inexpensive and useful method In which…
the charge of the Hydrogen ion, H+, for controlling and monitoring the
activity of various compounds Eh = is the Oxidation Reduction
and the solution is said to be acidic.
such as chlorine, ozone, bromine, Potential value of the
By contrast, a negative mV (or above
cyanide, chromate, and many reaction
pH 7) is associated with the charge
of the Hydroxyl ion, OH–, and the others.
E0 = is the standard potential
solution is said to be alkaline or that is particular to the
basic. 3.3. Measuring the ORP Value reaction series and that
has a constant value, that
An ORP system can be defined in ORP is measured in milivolts (mV),
not is affected by Aox/Ared,
the same manner. Unlike pH, ORP with no correction for solution
temperature, etc.
values are affected by all oxidizing temperature. Similar to pH, ORP is
and reducing agents, not just acids not a measurement of concentration
and bases which only influence a directly, but of activity level. ORP RT/nF =is the Nernst Number
pH measurement. Since ORP is the is the measure of the ratio of the
direct measurement of electrons activities of the oxidizing and A ox = is the activity of the oxidant
in transit during Oxidation- reducing species in a solution. The
Reduction reactions, under oxidizing ORP value of a particular material A red = is the activity of the
conditions, the measuring probe results in either a positive or reductant
loses electrons to the solution, negative mV output; the value is
which creates a positive potential; in determined by the size of the atom
Note 1: A
half cell is a structure that
a reducing environment, electrons of the material and the number of contains a conductive electrode
are donated to the probe, producing electrons found in its outer electron and a surrounding conductive
a negative potential. shell. The response speed of the electrolyte separated by a
process measurement varies with naturally occurring Helmholtz
Since a reducing agent is capable the concentration of the REDOX in double layer. Chemical reactions
within this layer momentarily
of accepting an electron and an the system; higher concentrations pump electric charges between
oxidizing agent is capable of losing are faster and lower concentrations the electrode and the electrolyte,
an electron, it can be said that the are slower. resulting in a potential difference
stronger the reducing agent the between the electrode and the
more negative the ORP value, and Once again, the German physical electrolyte.
the stronger the oxidizing agent the chemist and physicist, Walther Note 2: A n electrochemical cell is a
more positive the ORP value. Hemann Nernst found that a device capable of either deriving
electrical energy from chemical
potential difference occurs between
reactions, or facilitating
For example: a metal object and a solution that chemical reactions through the
contains ions of the same metal, introduction of electrical energy.
Acid Permanganate solution is when the object is immersed in the
www.yokogawa.com 29
Some Standard Reduction Potentials in Aqueous Solution at 25°C
Na Na+ + e- → Na -2.710
Strongest
Ca Ca2+ + 2 e- → Ca -2.760
Reducing
K K+ + e- → Lk -2.292 Agent
Li Li+ + e- → Li -3.040
The reference point for all oxidation or reduction reactions are compared to the hydrogen ion/ hydrogen (H+/H2)
reaction, which has a standard potential, E0, of 0 mV.
Table 3.3a
2H+ + 2e - → H2
Tables for standard potentials, E0, normally, a silver/silver chloride • Since ORP is non-ion specific
as seen in Table 3.3a, for various wire in a potassium chloride measurement, the chemistry of
reactions and their half reaction electrolyte solution. It may be the redox reaction can be quite
can be found in various General either free flowing or gel filled. complex, especially if several ionic
Chemistry Textbook reference The measuring probe is typically species involving varying numbers
materials3). platinum though some other inert of electrons transferred contribute
metals have been tested. to the reaction/oxidation reduction
The tables are usually written as potential.
reduction reactions, showing the When measuring ORP an
free electrons and the oxidized important feature to remember
atom on the left and the reduced is that unlike pH, temperature
atom on the right hand side of the compensation is not normally • Most ORP measurements are done
reaction equation. used for ORP measurements. at constant temperatures, such
Temperature does have as in process measurement and
A typical industrial ORP two distinct effects on ORP control.
measurement loop is similar to measurements. However, it is not
that used for pH measurement. It compensated for because:
includes a high input impedance
analyzer, a reference electrode, • The isopotential point (the point
measuring electrode, and system of thermal independence) of an
ground. ORP system is only relative to
the particular redox reaction, Note 3: Oxtoby, Nachtrieb, Freeman.
The reference electrode is typically and therefore there is no 1994. Chemistry Science
a standard pH reference electrode, “standard” isopoint for the
of Change. Philadelphia:
Saunders College Publishing.
overall ORP reaction.
30 www.yokogawa.com
3.4. Composition of the that insulate the platinum surface common, consists of a noble
Measuring Electrode from the solution will decrease the metal in contact with a
speed of response. solution containing both the
ORP/Redox is a potentiometrical
oxidized and reduced forms
measurement of the oxidizing/ The basic measurement principle of an oxidation-reduction
reducing power of a liquid. An ORP is that the measuring electrode will system. This is typically a
measuring electrode is similar to that give up electrons to an oxidant or platinum.
of a pH measuring electrode, except accept electrons from a reductant,
it is normally constructed of an inert without interfering with the An ORP measuring electrode can
(noble) metal. chemical reactions that are taking either be a separate electrode (as
place within the solution. seen in Figure 3.4a), a combination
The most common metal used
ORP/Reference electrode (as seen in
is platinum. Platinum, which is The metallic electrode can be Figure 3.4b), or a combination ORP/
considered the standard, has classified into three distinct types of pH measuring electrode (as seen in
excellent chemical resistance but ORP electrode. Figure 3.4c). The choice depends on
suffers slightly from Chemisorption1)
the application as well as customer
of oxygen, which slows down the i) The first consists of a metal installation requirements.
measurement response time. THis in contact with a solution
means that the surface can absorb of the same metal ions, i.e.
organic compounds and it may be a silver electrode placed
attacked by sulfides and cyanides in in a solution of silver Note 1: Chemisorption (or chemical
strongly reducing solutions such as nitrate, which will develop adsorption) is adsorption in
solutions with redox potentials less a potential proportional to which the forces involved are
than -500 mV. the silver ion activity. valence forces of the same
kind as those operating in
3. BASICS ON ORP
The oxygen bonds to the surface ii) The second consists of a the formation of chemical
in strong oxidizing solutions and metal electrode coated compounds.
hydrogen bonds to the surface in Chemisorption means to
with a sparingly soluble
take up and chemically bind
strongly reducing solutions. Some salt of metal, in contact
(substance); in this case,
anti-corrosion chemicals added to with a solution containing oxygen, Oxygen, onto the
cooling towers and pasteurization the anion of the metal salt, surface of another substance.
processes perceive the electron i.e. a silver-silver chloride
active surface of the platinum as electrode in a potassium
corrosion and passivates it. All of chloride solution.
these surface reactions can result in
slow response. Any surface coatings iii) The third kind, and most
Reference
system
Platinium wire
band
Glass housing
pH Glass Ag/AgCl Ag/AgCl
www.yokogawa.com 31
3.5. Composition of the Et
Reference Electrode
Et
E3
ORP can be measured relative E4
to any practical or theoretical R elecrolyte
reference electrode such as Ag/
AgCl, or SHE (Standard Hydrogen R elecrolyte
E2
Electrode) like described in section
2.5.2 and 2.5.5. E5
R membrane
3.6. The Measuring Circuit R diaphragm
E1
An ORP probe is really a millivolt R
meter, measuring very tiny
Metal Pin
liquid connected
voltages generated across a circuit Fig. 3.6a Fig. 3.6b E3 E2 E1 to Platinum
formed by a measuring electrode
(the positive pole of the circuit The Figures 3.6a/b shows that
Et =E1 + E2 + E3
normally platinum), and a reference following potentials are of the most
electrode (the negative pole), with significance: It is also important that within the
the process solution in between. analyzer being used for detection,
The difference in voltage between E1= Potential between the ORP
there must be a high impedance
the two electrodes is what is metal surface and the process
(resistance) in order to measure the
actually being measured by the very tiny voltages (or charge build
analyzer. E2= Potential between reference
electrode and the electrolyte up) being generated by the constant
acceptance and giving up of electrons
These voltages give us an indication
E3= Potential that develops at the on the ORP electrode.
of the ability of the oxidizers or
3. BASICS ON ORP
3.7. Standard ORP vs pH the effectiveness of the chlorine to monitor the diatomic Cl2 level,
Compensated ORP (rH) as an oxidizer. ORP can be used to we would have a large mV reading
indicate the activity of chlorine in a to start with, but as it oxidizes with
ORP measures the ratio of the solution. Since addition of chlorine water reacting with the Calcium
activities of the oxidizing and increases the oxidizing capability Thiosulfate to form a salt, the mV
reducing species in a solution. of water, measurement of the ORP reading would decrease.
This is a measure of the solution's provides a useful indicator of the
ability to oxidize or reduce another quantity of active chlorine present. Since ORP varies with pH changes,
substance. As an oxidizer is This is very important when the as well as changes in the chlorine
added to the process, it “steals” chlorine is being used as a biocide levels, we must compensate for the
electrons from the surface of the agent to control algae growth effects of any pH changes. This can
ORP measuring electrode, causing in the process. One drawback, be done by measuring the pH and
it to become more positively however, is that pH changes also ORP independent of one another
charged. Continuing to add oxidizer affect the oxidizing potential of and then calculating the effect of
generates a higher and higher the available chlorine and the the pH change on the ORP using
positive voltage. The role of an ORP resultant ORP value. formulas and graphs.
system is to measure these tiny
voltages generated across a circuit Below a value of 1.9 pH, chlorine A simpler and more direct method
formed by a measuring electrode exists as a diatomic molecule is to compensate for the pH changes
(the positive pole of the circuit, (Cl2) in water. As the pH increases by replacing the standard Ag/AgCl
normally platinum), and a reference above 1.9, chlorine oxidizes water reference electrode normally used
electrode (the negative pole, to produce HOCl and the ORP with a pH measuring electrode. This
reference electrode), immersed in millivolt reading will go down. As is known as pH Compensated ORP
the solution. the pH value continues to increase, (rH).
HOCl further dissociates into OCl-
ORP systems are typically rugged, above a pH of 7.3. Since the pH measuring electrode
but do have some limitations. output changes as the pH of the
For example, when ORP is used HOCl being more active that OCl- process changes, it acts as a moving
with a chlorine-based sanitation has a higher ORP value. So, as reference effectively cancelling out
system, it will not indicate the the pH increases, an ORP sensor any change in pH and leaving only the
chlorine concentration in parts per detects a decrease in value, which mV value, which is due to changes in
million. It will, however, indicate reflects the decrease in HOCl. diatomic chlorine (Cl2) levels.
Therefore, if we were using ORP
32 www.yokogawa.com
3.8. Standard Maintenance and After any cleaning procedure, the practical if pH loops are also being
Calibration electrode should be allowed to soak maintained. Preparation and use of
in clean tap water for at least 30 both types of solutions are discussed
Maintenance and calibration for an minutes to allow residual chemicals below:
ORP system has similarities with to dissipate and the sensor to
those methods and procedures recover. After calibration, when the Quinhydrone1) Solution
used for both pH and conductivity electrodes are placed back in the
process, they should be allowed to To create an ORP solution using a
systems. The problems and
equilibrate for at least 15 minutes. pH buffer (either 4.0 or 7.0) stir in
maintenance procedures associated
For optimal operation, the sensors a small amount, approximately <
with the ORP reference electrode,
should be installed in an area with 0.5 gm, of quinhydrone into 200
for example, are handled the
good agitation as process flows mls of solution. Quinhydrone is
same as they are for a pH loop.
past the electrode help to keep the not very soluble, so only a small
The maintenance of the ORP
platinum sensing element clean. amount will dissolve in the buffer
measuring electrode, however, is
changing the solution to an amber
handled in much the same way as a
ORP electrodes should need less color. If all of the quinhydrone
conventional conductivity sensor.
frequent calibration than a typical dissolves, then continue to add small
Calibration of the ORP system is also
pH sensor since the redox potential amounts and stir again. Saturation
similar to how a conductivity loop is
is a characteristic of the interaction is achieved when a small amount of
calibrated.
between the platinum and the quinhydrone remains undissolved
Cleaning the measuring electrode redox equilibrium. However, it is after mixing.
will improve accuracy and the prudent to periodically verify the
performance of the measuring Whether it is a 4.0 or a 7.0 buffer
sensors response time. The
system by placing the electrode in a you are using, Table 3.8a shows
frequency of maintenance, which
solution with a known potential and the mV reading you should obtain
includes cleaning and calibration,
calibrate as needed to correct for depending on which reference
is determined by how the process
the reference side of the sensor. electrode is being used. As an
3. BASICS ON ORP
affects the electrode. Methanol,
example, a quinhydrone/pH 4.0
isopropyl alcohol or a detergent can
Standard ORP Solutions solution should give a 253 mV (± 30
be used to remove oily or organic
mV) at 25°C for a reference electrode
coatings. While those applications
When verification or calibration of that has 3M KCl internal fill.
where removal of any anti-corrosion
an ORP sensor is required, there are
chemicals or mineral deposits,
two types of Standard Solutions
soaking the electrode in 10% nitric Note 1: The quinhydrone powder
that are commonly used. The first
acid for about 10-15 minutes is a poses a moderate health risk,
are premade solutions designed to
good starting point. causing irritation of the lungs
provide a specific stable mV value, with prolonged exposure. The
If the various forms of chemical typically one that falls within the premade calibration solutions
cleaning are not sufficient to achieve process ORP range. The second type are fairly innocuous unless
an accurate measurement and of solutions, and probably the most ingested in large amounts.
response time, a last resort would common, are those that are made Both types should be handled
using the standard pH 4 and pH 7 carefully following good
be to polish the platinum surface laboratory practices.
with a 600 grit wet-dry emery cloth buffers with quinhydrone crystals
or a 1-3 micron alumina polishing mixed in until saturation is reached. Note 2: SCE = Saturated Calomel
Either of these pH buffer solutions Electrode
powder to remove any surface
pitting or stubborn coatings. can be used for calibration of an Note 3: SHE = Standard Hydrogen
ORP measuring system and are very Electrode
Table 3.8a mV value of ORP solution made with pH Buffers and Quinhydrone
www.yokogawa.com 33
Pre-made Stabilized ORP Solutions Solution was prepared against. To at 25° C. This is the 250 mV value on
use the chart below, you have to the solution minus the 231 mV value
Reference electrodes with different know what (1) reference solution shown as the difference between
internal fill solutions will have is used in the reference electrode the SHE and the 1M KCl references.
different mV outputs when they and (2) what reference solution This would be 19 mV.
are put in the same Standard the known pre-made solution is
Solution. This is because the being compared to. For example,
Standard Solution was prepared if you have a pre-made 250 mV
with one specific reference fill solution that is referenced to SHE
solution in mind. Table 9.2 lists in (Standard Hydrogen Electrode)
the left-most column, some of the and the reference electrode in your
most commonly used reference measuring loop uses a 1 M KCl fill Note 1: S
CE = Saturated Calomel
Electrode
electrode fill solutions. Across the solution, then on the transmitter
top of the table are the possible you would NOT read 250 mV, but Note 2: SHE = Standard Hydrogen
Electrode
reference fill solutions that Standard instead you would read only 19 mV
20°C 25°C 30°C 20°C 25°C 30°C 20°C 25°C 30°C 20°C 25°C 30°C 30°C 30°C 30°C
68°F 77°F 86°F 68°F 77°F 86°F 68°F 77°F 86°F 68°F 77°F 86°F 86°F 86°F 86°F
3. BASICS ON ORP
From SCE1) – – – +34 +35 +36 +45 +45 +46 +13 +14 +16 +241 +241 +241
From Ag/
AgCI -34 -35 -36 – – – +11 +10 +9 -21 -22 -23 +205 +205 +205
(3M KCL)
FromAg/
AgCI -45 -45 -45 -11 -10 -9 – – – -32 -31 -30 +202 +199 +196
(sat KCL)
From Ag/
AgCI -13 -14 -16 +21 +22 +23 +32 +31 +30 – – – +234 +231 +228
(1M KCL)
From SHE2) -247 -244 -241 -213 -209 -205 -202 -199 -196 -234 -231 -228 – – –
Table 3.8b
Proper calibration 3. Perform a single (1) point If a short span is found (less than a
calibration. +150 mV change between the first
The following steps are commonly solution and the second solution),
used for calibration of an ORP loop. A. P
ick a solution with a mV value the platinum/ gold measuring
closest to the control point. surface may be coated and the
1. Clean the ORPand Reference electrode should be re-cleaned and
electrodes FIRST. B. If possible adjust (heat) re-calibrated.
the solution to one of the
2. M
ake FRESH Buffer Solutions temperatures shown in the mV
with either Quinhydrone Crystals tables above.
(See section 3.8), or pour a fresh
sample of a pre-made stabilized C. Allow stabilization time then
ORP solution. adjust to the correct value.
A. O
RP Buffers are best made just
4. Rinse the electrode between
before using.
calibration measurements if
B. Never keep (store) ORP checking at a second buffer value.
calibration solutions.
34 www.yokogawa.com
Notes
3. BASICS ON ORP
www.yokogawa.com 35
4. Products and Features
Number of sensor inputs: Single or
dual sensor measurement.
cutting-edge technology.
It includes information on
electronics (converter/transmitter),
electrodes, and their features and
functionalities.
36 www.yokogawa.com
Mounting: IP66/NEMA 4X 1/2DIN Display: Main screen displays B) 2–Wire type analyzer:
enclosure for field mounting and containing primary variable in
(Yokogawa model: FLXA21 series)
panel mounting large font (user selectable), other
process variable(s) in small font,
Trending: Trending display available Unit symbols, Tag number (user
up to 2 weeks programmable), Process description
(user programmable), Status of
Data storing: On-screen logbooks alarm output(s), Status indicator
store calibration data, configuration during HOLD and WASH situation,
changes and events Main function keys.
Input ranges: Analyzer accepts
ranges such as pH: -2 to 16 pH, ORP:
-1500 to 1500 mV, rH: 0 to 100 rH and
Temperature - Pt1000: -30 to 140ºC,
Pt100: -30 to 140ºC, 350Ω (DKK): -30
to 140ºC, 5k1: -30 to 140ºC, 6k8: -30
to 140ºC, PTC10k: -30 to 140ºC, NTC
8k55: -10 to 120ºC, 3kBalco: -30 to
140ºC. Fig. 4.08 FLXA21 outside
Advanced temperature
compensation: Analyzer offers three
additional modes for advanced
temperature compensation viz; Full
compensation for strong acids and
alkali’s using NEN6411 algorithm,
linear temperature coefficient
setting, and Matrix temperature
compensation.
www.yokogawa.com 37
Sensor 1
Fails Fig. 4.09 Sensors check
Output Signal: Bi-directional by enclosures (IP Code) Type For all protection type,
HART digital communication of protection II 1G Ex ia IIC Ga,
superimposed on mA (4-20mA) Group: II, Category: 1G, T4: T4: for ambient temperature:
signal. for ambient temperature:–20 -20 to 55°C
to 55ºC, T6: for ambient
Ambient conditions: Temperature temperature:–20 to 40ºC, 6: for ambient temperature:
T
between -20 to +55ºC (-5 - 130ºF) Atmosphere pressure: 80kPa -20 to 40°C
and storage temperature -30 to (0.8bar) to 110kPa (1.1bar),
+70ºC (-20 - 160ºF). Humidity is 10 Atmosphere pressure: 80 kPa (0.8
Degree of Protection of the
to 95% RH at 40ºC (100ºF) (non- bar) to 110 kPa (1.1 bar)
Enclosure: IP66.
condensing).
Degree of Protection of the
Safety Standards: In line with Enclosure: NEMA Type 4X and
b) IEC Ex intrinsically safe
EN61010-1, UL 61010-1. IP66
approval Applicable standards:
IEC 60079-0: 2007 Part 0: General
EMC standards: EN61326-1 Class requirements, IEC 60079-11:
A, Table 2 (For use in industrial d) CSA Intrinsically safe and
2006 Part 11: Intrinsic safety nonincentive approval:
locations), EN61326-2-3, AS/NZS “i”, IEC 60079-26: 2006 Part 26:
CISPR11, Korea Electromagnetic Construction, test and marking Type of protection:
Conformity Standard. of Group II Zone 0 electrical
apparatus IEC 60529: 2001 Class I, Division 1, Groups A, B, C
Degrees of protection provided and D (Intrinsically Safe)
ADDITIONAL POINTS DUAL
4.PRODUCT & FEATURES
1) Combined pH/ORP sensor for 2) Combined pH/ORP sensor 3) Combined pH/ORP sensor
general purpose applications: for harsh applications with in applications where chemical
(Yokogawa model: FU20 series) fluctuating pressure and/or resistance is a vital issue:
temperature: (Yokogawa model: PH20 series)
Reference system: (example: pure water applications):
Long life saturated Ag/ Reference system: Long life
(Yokogawa model: FU24 series)
AgCl reference system saturated Ag/AgCl reference system
Reference system: Long
Reference junction: Reference junction: PTFE reference
life saturated Ag/AgCl
PTFE reference diaphragm to prevent
reference system.
diaphragm to prevent fouling and reduce
fouling and reduce Reference junction: PTFE measurement error.
measurement error. reference diaphragm Double junction
Double junction to prevent fouling and combined with ion-trap
combined with ion-trap reduce measurement to prolong the life of the
to prolong the life of the error. Double junction reference probe even in
reference probe even in combined with ion-trap chemically unfavorable
chemically unfavorable to prolong the life of the environments
environments reference probe even in
Temperature
chemically unfavorable
Temperature compensation: Integral
environments.
compensation: Pt1000 element for
Integral Pt1000 Temperature accurate temperature
element for accurate compensation: Integral compensation.
temperature Pt1000 element for
Additional
compensation. accurate temperature
measurement:
compensation.
Additional measurement: Equipped Equipped with platinum
with platinum ORP/LE electrode for Additional measurement: ORP/LE electrode for
accurate simultaneous pH- and ORP Equipped with platinum accurate simultaneous
measurements. ORP/LE electrode for pH- and ORP
accurate simultaneous measurements.
Electrolyte: Polymerized electrolyte pH- and ORP measurements.
Electrolyte:
to extend the sensors lifetime.
Electrolyte: Polymerized electrolyte Polymerized electrolyte
40 www.yokogawa.com
7) Combined pH sensor in a 12 increases the sensor lifetime Conductivity: Measuring in
mm design with an external significantly compared with pH applications from 10 μS/cm
solution ground: electrodes.
Connector: Variopin connector
(Yokogawa model: SC25V series) Material of solution ground:
Titanium Applications:
Sensor diameter and design: 12
mm. Temperature sensor: Pt1000 • SC25V-ALP25 for chemically
integration in pH compartment harsh applications and high
SC25V giving highly accurate temperature temperatures
compensation.
• SC25V-AGP25 for all general
Certification: ATEX II1G Ex ia IIC T3... purpose applications
Design: Sensor design with an T6 Ga
external solution ground. The
design made it possible to create CIP: CIP and Steam cleaning
a large electrolyte volume. This possible
Type Membrane Resistance pH range Temp. Pressure Reference Reference Diaphragm Flow
in MΩ/25ºC range (ºC) range kPa liquid system
SC21-AGP24 Universal 50 - 100 0 - 14 0 - 80 1-500 3.3 m. KCI Ag/AgCl (wire) Porous PTFE 0
pH-glass bulb Thickened Silver-silverchloride
SC21(D)-AAP26 Chem. res. pH-glass bulb 250 - 400 0 - 14 0 - 110 1-500 Oversaturated Ag/AgCl (wire) Porous PTFE 0
steam-sterilisable / bulb
3
4 KCI thickened Silver-silverchloride
SC21-ALP26 Chem. res. 500 - 900 0 - 14 10 - 120 1-500 Oversaturated Ag/AgCl (wire) Porous PTFE 0
pH-glass dome KCI thickened Silver-silverchloride
SC21-AGP26 Universal 120 - 200 0 - 14 -10 - 100 1-500 Oversaturated Ag/AgCl (wire) Porous PTFE 0
pH-glass bulb KCI thickened Silver-silverchloride
SC21-AGP64 Universal 50 - 100 0 - 14 0 - 80 1-500 3.3 m. KCI Ag/AgCl (wire) Porous PTFE 0
pH-glass bulb Thickened Silver-silverchloride
www.yokogawa.com 41
9) Combined flow type pH Reference system: Flowing reference “L” glass: The application of “L”
electrode: system for pollution resistance, and glass is for measurements in
highly stable reference potential. alkaline media with high process
(Yokogawa model: SC21C-AGC55)
temperatures. Low resistance glass
Electrode connection: PG13.5 for fast response and sensitivity.
standard DIN electrode connection. L-glass sensors for high temperature
SC21C-AGC55 chemically harsh applications.
Reservoir: Use in combination with
the pressurized electrolyte reservoir Design: Dimensions and design meet
to obtain a positive flow towards the the requirements of DIN 19263.
process
Type of sensor: Flowing Isothermal point of intersection: pH
electrolyte sensor uses a ceramic 7 (nominal value at 0 mV).
reference junction
10) Single glass electrode:
(Yokogawa model: SM21 series) Maximum pressure: 1000 kPa (10
Electrolyte: 3.3 molal KCl bar).
Two types of glass are available, as
Application: For tough applications described here: Thickness of the glass membrane.
where pollution of the reference
system is to be expected. Low ”G” glass: This is used for the Shockproof: Bulb membrane for
ionic applications where the membranes of electrodes in general purpose. A bulb shaped
possitive flow of electrolyte processes where the nominal pH glass membrane suitable for general
provides the conductivity needed value varies between pH0 and purpose.
to measure pH(< 50 µS/cm) pH14. Since this type of glass has
Heavy duty: This dome shaped glass
a wide application range it is also
Glass type: Heavy duty pH membrane is suitable for harsh and
termed “general purpose” glass.
sensitive glass abrasive applications.
High resistance glass for chemical
resistance. G-glass sensors for all
general applications.
SM21(D)-AL4 shockproof 300-450 15-130 10 High Temperature T cont. ≥ 70°C, high chem. Resistance
SM21(D)-AL6 heavy duty 600-900 25-130 10 High T + abbresive T cont. ≥ 70°C, high chem. Resistance
* all pH glass sensors have a sodium error of pH 0.17 above ph=13. This is because the glass will see the Na+ ion as H+ ion.
The pH reading is lower than it is in reality.
1. Ceramic junction
2. PTFE junction
3. Glass sleeve capillary element
Reference electrode
• Flowing type sensors for dirty • Ceramic and PTFE junction: Non-flow type
applications or (Ultra) pure electrolyte flows slowly into the
A non-flow type reference electrode
water applications (often needs process. PTFE resists dirt
can be used for processes that
a electrolyte reservoir) don’t contain components that
• Glass sleeve: very dirty applications
or for (Ultra) pure water poison the reference system. The
• Non-flow type sensors for all gel-type electrodes have a large
general applications applications because of the larger
flow-rate. area of porous PTFE junction for
optimal resistance against electrode
pollution.
42 www.yokogawa.com
The SR20-AC22 electrode has
a flexible PTFE tube. Therefore SR20-AC22 SR20-AP24 SR20-AP26
this electrode can be applied
in processes with frequent G 1/4" G 1/4" G 1/4"
temperature and pressure
fluctuations. Ø 17 23 Ø 17 23 Ø 17 23
Features
• Easy maintenance
• No reference liquid wastage
• Maximum process pressure: 1000
43
compensation.
• To be used in non-polluting fluids.
• Saturated KCl-solution (pellets). 3
• For low ionic applications and high
13
temperatures. Ø5
• Temperature range: 0 to 120ºC.
Ø 8,5
Two-phase electrode Gel-type electrode Double junction gel-type
• Diaphragm resistance (25ºC) <5kΩ.
* In applications where high process temperature occur together with very low (<2) or very high (>12) pH levels, the lifetime is shortened.
www.yokogawa.com 43
12) Single pNa electrode:
(Yokogawa model: SM23 series)
SR20-AC32 Unit: mm SR20-AS52 Unit: mm
120 120
Features
Ø 12 • hock-proof glass
S
membrane with fast Unit: mm Ø 12 SM23-AN6 Unit: mm
• Dimensions and design meet the
SM23-AN4
response.
requirements of DIN 19263.
4.PRODUCT & FEATURES
G 1/4" G 1/4"
characteristics for
Ø 17 23
type SM23-AN6
shaped) membranes.
• pNa range < 0 to 4 • Glass resistance @
25°C 450 - 700MΩ.
• pH 2 units higher than the pNa (pH- Ø 12
pNa>2). • Heavy-duty
Ø 12
processes. For
Ø 11
• Suitable (Ultra-, pure-) water example brine Shock proof AN4 Heavy duty AN6
application. applications.
44 www.yokogawa.com
13) Temperature electrodes: Unit: mm Unit: mm
SM60-T1 SM60D-T2
(Yokogawa model: SM60 series)
G 1/4"
For accurate pH measurement Wrench 17
temperature compensation is Ø 17 23 26
120 120
14) Differential pH Sensor: describe the sensor as pH control Output voltage: The output voltage
sensor rather than pH measuring of the sensor depends only on the
(Yokogawa model: FU20-FTS)
sensor. salt concentration of the process.
Differential electrode: The
Reference system: pNa, salt- Effect of salt: A rule of thumb is that
differential measuring principle
sensitive reference glass. The a change in salt concentration of +/-
combines the normal potential
reference is not a (liquid) junction 25% has an effect of less than 0.1pH
generated by the pH glass with the
but a glass sensor which responds on the pH reading.
potential from a Sodium pNa glass.
to the sodium level in the sample
In applications where Sodium-,
rather than to pH changes (within
Potassium- or Calcium-salt is
the applicable range of the sensor).
present the glass will generate a
stable reference voltage. This means
the measurement can be done
without a conventional reference
www.yokogawa.com 45
5. Lifespan and storage of electrodes
Most customers would like to know means that proper performance the electrode has to be confirmed
what the expected lifetime of a is guaranteed at first installation by calibration as described in the
sensor is when used under process of the product in process if stored calibration chapter applicable to the
conditions. Customers are also under prescribed conditions. The sensor type.
interested in the shelf life for a proper storage conditions for each
sensor when stored in inventory. type of sensor are mentioned in the Single reference electrode
appropriate instruction manual. On
Since the life of a sensor is the box the sensor is delivered in the When a reference electrode is to
dependent on various parameters acceptable storage temperature and remain unused for a long period
like temperature, pressure, service position is indicated by means of it is necessary to fill the electrode
composition, maintenance activity pictograms. completely with electrolyte
taken and sensor handling, it is solution, to close the refill opening,
difficult to define a particular life. Lifetime is defined as the time the and to protect the diaphragm
Even the same application life may product shows good performance against drying out by fitting the
vary from site to site and will vary in the processes. Because of the protection sleeve. This protection
from application to application. variations in process conditions we sleeve must also be filled with
cannot be held liable for this period. electrolyte solution.
The best answer to these questions However as a guide we can state
is given by recommending that if the lifetime of a particular Electrodes of type SR20(D)-AC32
guidelines for storing and sensor has been confirmed in mild have an inner duct (see page 43)
precautions during usage of the conditions the lifetime in more which contact the bellow with the
electrode to maximize the sensor's harsh conditions will limit the life of process liquid. This duct must be
lifetime. the sensor under such conditions. sealed with the rubber plug.
(See Table in 5.1a)
5.1. Recommended storage With the electrodes (types SR20-
guidelines pH electrodes 5.2. Recommended storage AS52) the sleeve must be loosened
guidelines and lifted. The hole in the fixed
Generally pH sensors consist of part must be closed by means of a
a glass electrode and reference protective sleeve or tape.
electrode (junction), both of which Single pH electrode
are consumable parts. Each part has Combined pH sensor
When storing a pH sensor, the glass
an acceptable shelf life and lifetime.
electrode should be protected with To guarantee optimal performance
Lifetime is influenced by many a cap containing KCl solution to of combined electrodes
factors such as sample components, prevent performance degradation from Yokogawa the sensors
pH value, temperature, and due to drying of the glass electrode’s should be stored following the
pressure. Tables 5.1a and 5.1b shows glass membrane. recommendations given in the
the approximate shelf time and applicable instruction manuals and
The sensors can be stored for following the indications on the
lifetime of various electrodes. This
approximately two years, but exterior of the product's original
is just a reference for calculating
this is only a guideline and is not box..
the necessary consumable amount
guaranteed. If the moisture content
when purchasing them. The
of the KCl solution evaporates, In the table below the shelf life
shelf time and lifetime are not
causing the glass membrane to for specific products are given,
guaranteed.
become dry during this storage however these periods can't be
period, the performance can be considered guaranteed.
Shelf life and lifetime restored by re-immersing the glass
5. LIFESPAN & STORAGE
46 www.yokogawa.com
Productgroup Maximum shelf life
Reference
Modelcode(s) in months
Reference-electrodes
Combined-electrodes
Glass electrodes
SM21*-****(all types) 15 Month-code on package
Table 5.1b
Notes
www.yokogawa.com 47
6. Troubleshooting and diagnostic
6.1 Errors resulting from measured potential is generated But often a periodic frequent cleaning
cracked membranes on the surface of the pH sensitive is needed to prevent building of a layer.
glass membrane it is of the greatest In these cases automatic systems help
A minute crack in the membrane to reduce maintenance efforts and
importance that these membrane
of a glass electrode is not always help to prolong the sensor lifetime.
surface remains clean at all times.
visible to the naked eye. Frequent Afterwards you find example of
shocks may give minute cracks in When the junction becomes clogged, automatic cleaning methods.
the membrane with consequently contact between the reference
measuring errors. In this case the • Mechanical brushing:
system in the electrode and the
analyzer reading is 0 mV and the A brush actuated by a mechanism
process liquid will be interrupted.
analyzer shows a reading of pH 7, electrically or pneumatically driven,
This will reduce the speed of
because most pH sensors using an passes periodically across the
response and may result in unstable
internal buffer with pH 7. membrane.
measurement or complete failures.
• Jet Cleaner:
For neutralization processes In most cases it is sufficient to clean This jet cleaner gives the opportunity
(setpoint is typically pH 7) this is only the electrode before calibration. to avoid build up of deposits and
a potentially dangerous situation If it is necessary to hand clean the consequent fouling on the sensor
for the environment and plant life. electrodes periodically, suitable by spraying water (or air) onto the
Without additional diagnostics, the cleansers may be used, e.g: membrane of the sensor.
error will not be detected.
• Diluted solution of hydrochloric • Chamber:
Yokogawa uses a patented online acid (10% HCI solution) Sensor is lifted away from process
impedance check. The analyzer Deposits of hydroxides, lime etc. at specified intervals and washed
frequently checks the impedance of can be removed by immersing the in cleaning chamber. Often possible
the pH membrane (and reference) electrode in a diluted solution of to keep the surface of the glass
via the solution ground. In case of a hydrochloric acid and then rinsing electrode clean by periodically
broken membrane an alarm can be with water. spraying it with a cleanser. The
generated.. cleanser used, its concentration and
•A
solution of water with the cleaning frequency, depends on
hydrochloric acid and pepsin the application. Often it is necessary
Albuminous depositions can be to sterilize the electrodes with steam.
Note: For details see Chapter 4,
"Products and Features" removed by means of a solvent By means of steam the temperature
containing: 0,5 ml concentrated of the electrodes is kept at a certain
hydrochloric acid, 10 gr pepsin and tempereature (e.g. over 120°C ) for a
1 litre demineralized water. After short time.
6.2 Fouling of the pH sensor
treatment the electrode should be
For industrial applications and thoroughly rinsed with clean water. •Ultra-sonic cleaning:
particularly for automatic process With the ultra-sonic cleaning method,
•H
ot water: the liquid around the electrodes is
measurements it is of the greatest
Depositions of oil and fat can be vibrated. This is preventing and/or
importance that the sensor is kept
removed by means of hot water. removing any deposit formed on the
clean.
If the result is unsatisfactory, a electrodes. With this method optimal
Fouling of the glass membrane is domestic abrasive may be carefully cleaning depends on the vibration
the most common cause of error in used. frequency and the velocity of the
any pH measuring system. As the process liquid past the electrodes.
Cleaning system
Cleaning system Process involved
Jet (Water/air) Brush Ultrasonic Chamber
Note 1: W
hen polar solvents are used for cleaning, these affect the Note 2: If possible increase the flow rate of the liquid to be measured past
gel-layer and it is necessary to soak the electrode for some the electrodes. This may result in a reduction of deposit on the
hours to allow the layer to reform. When a-polar solvents are electrodes (typically flow rates of 2m/sec. are necessary). This
used first re-clean the electrode with a-polar solvent and method is particularly suitable when measuring pH in liquids
after that soak the electrode to allow the layer to reform. containing solids that settle out quickly.
48 www.yokogawa.com
6.3 Prevention of the reference 6.4 Poisoning of the reference d.An excessive formation of
electrode fouling electrode crystals in the electrode may
restrict the discharge, and allow
Clogging or fouling of the reference As well as the possibility of fouling,
process liquid penetration.
electrode can be prevented or problems will occur if the reference
To overcome this the reference
reduced by: electrode becomes poisoned.
electrode must be filled with
When the flow of electrolyte in the
a.The correct choice of the junction an electrolyte which has a
reference electrode is insufficient,
type for the application. In saturation point higher than the
the composition of the electrolyte
highly polluted liquids the sleeve maximum temperature of the
may change as a result of process
junction gives better results than process liquid.
liquid penetrating the electrode.
the smaller ceramic capillary
type, alternatively, the reference Any change in composition
electrode with P.T.F.E. capillary of the electrolyte, in
may be used to advantage. the electrode reference
chamber, may cause a
b.Increasing the flow of electrolyte. measuring error.
By increasing the pressure on
the electrolyte in the reference Some common causes
electrode the flow will increase. of reference electrode
This will reduce the rate of fouling. poisoning are:
The reference electrode with a
flexible viton bellows provides this a.The reference electrode
feature. For applications where has been allowed to run
the process pressure fluctuates dry allowing the process
widely, and the chance of process liquid to penetrate into
liquid penetrating some types of the electrode. This
reference electrodes is increased, problem mainly occurs
the “Bellomatic” electrode should with sleeve junction
be used. electrodes by incorrect
treatment, lack of care
c.Cleaning (See section 6.2 for types in cleaning the ground
of cleaning) faces, or by not re-filling
the electrode early
Clogging of the junction can enough.
be remedied by soaking the
reference electrode in 3.3 m b.Another frequent cause
electrolyte solution. Slowly of poisoning is due to
increase the temperature of the higher than expected
solution to a maximum 80°C and process pressures
cool down in the same solution. or short-term, high-
pressure fluctuations.
This ensures that any electrolyte To alleviate this
that may have crystallized in the problem the electrolyte
junction will become soluble reservoir should be
again. maintained at a higher
level to increase the
If the outside of a ceramic junction electrolyte pressure.
has been fouled by the process An alternative
liquid the junction can be scoured solution is to replace
slightly using fine emery paper, the electrode with
this increases the flow capacity. an integral pressure
compensation type,
An alternative cleaning method
such as “Bellomatic”
is to use a suitable solvent in
(See page 43/44).
which the particular solids, that
have fouled in the junction, will c.When a reference
dissolve. The electrode should be electrode filled
rinsed in clean water before use.
6.TROUBLESHOOTING &
www.yokogawa.com 49
6.5 Errors caused by damped or To prevent "inside radiation" a full study of which is too complex
electrode cable and connection of disturbances, the cables are to be covered in this book. It is
supplied with a special layer of essential therefore, that the liquids
It is of the greatest importance
graphite for screening. This layer are grounded at the point where
to protect the electrode cables
can be difficult to remove. It is the pH value is to be measured.
and connections from ingress of
not uncommon to see measuring
moisture. This may reduce the This can be easily achieved when
errors (e.g. instability) due to
insulation resistance of the cable metal fittings are used. When
improper shortening and finishing
connections causing a partial short plastic fittings are used they should
of the cable. The cause of the
circuit and consequently an erro- be fitted with an solution ground
measuring error was the result
neous reading. Often it happens that electrode of suitable metal.
of a strongly decreased isolation
the cable is broken.
resistance between the core and
screen of the electrode cable.
The online impedance control allows 6.8 Errors caused by poor
the user to detect errors caused by installation
For the glass electrode cable, this
moisture or broken cable.
isolation resistance must be more When commissioning a pH
Note: In case of moisture problems
than 1,000 times the resistance measuring system, care should be
Yokogawa can offer easily across the glass membrane. taken that the electrode connectors
solutions to solve this item, like do not become wet, or fouled
SENCOM (SENsors with digital Shortening cables is strongly
with fine deposits. Particular care
COMmunication) discouraged.
should be taken if the electrodes
are removed for tests, or pressure
6.7 Interferences by stray testing of the pipe, and are left
6.6 Errors caused by shortening
unprotected.
the electrode cables voltages in the liquid
Often, it is usual for installation In many industrial applications If this occurs any dirt should
technicians to cut off surplus the voltage potential of process be carefully removed from the
cable lengths. This is done despite liquids is significant and cannot electrode connections and the
the availability of different cable be neglected when making pH cables removed and, either
lengths. measurements. This voltage level thoroughly dried, or preferably
may be due to a number of causes, replaced.
Notes
6.TROUBLESHOOTING &
DIAGNOSTIC
50 www.yokogawa.com
7. Frequently Asked Questions & Answers
I. RELATED TO SENSORS: that the measurement can be done pressure, ambient temperature)
www.yokogawa.com 51
7.10. What are the advantages The viscosity of the electrolyte The sensor must be cleaned and
of 3.3 molar KCl ? decreases with increasing washed carefully before being
temperature. When you have a inserted into the buffer.
All pH sensor suppliers use a
Bellomatic that lasts two months
7.FREQUENTLY ASKED Q&A
ºC 0 1 2 3 4 5 6 7 8 9
0 100.000 100.391 100.781 101.172 101.562 101.953 102.343 102.733 103.123 103.513
10 103.902 104.681 104.681 105.071 105.460 105.849 106.238 106.627 107.016 107.404
20 107.793 108.181 108.570 108.958 109.346 109.734 110.122 110.509 110.897 111.284
30 111.672 112.059 112.446 112.833 113.220 113.607 113.994 114.380 114.767 115.153
Resistance in ohm
40 115.539 115.925 116.311 116.697 117.083 117.469 117.854 118.240 118.625 119.01
50 119.395 119.780 120.165 120.550 120.934 121.319 121.703 122.087 122.471 122.855
60 123.239 123.623 124.007 124.390 124.774 125.157 125.540 125.923 126.306 126.689
70 127.072 127.454 127.837 128.219 128.602 128.984 129.366 129.748 130.130 130.511
80 130.893 131.274 131.656 132.037 132.418 132.799 133.180 133.561 133.941 134.322
90 134.702 135.083 135.463 135.843 136.223 136.603 136.982 137.362 137.741 138.121
100 138.500 138.879 139.258 139.637 140.016 140.395 140.773 141.152 141.530 141.908
52 www.yokogawa.com
7.16. Can we use a glass Advantages are: amplifier by jumping terminal 13
electrode as reference? If yes, (input 2) with 14 (common voltage of
a:The sensor can be calibrated in the
how and where? the analyzer).
laboratory and stored. As soon as
www.yokogawa.com 53
from this feature. Also applications sensor measures at the surface thermal expansion of the internal
corrosive to glass like hot alkaline and regular cleaning coupled with buffer there is always a considerable
solutions with high salt content or the dynamic response diagnostic air bubble in the sensor.
processes with risk of HF (Fluoride function is recommended.
When the sensor is mounted upside
7.FREQUENTLY ASKED Q&A
54 www.yokogawa.com
7.30. Can we select jet cleaning output from the second input The pH sensor has an accurate
along with 2-wire analyzer? module? temperature sensor for temperature
compensation. With PH450 it is easy to
This combination is possible with In a dual input FLXA21 analyzer you
select one mA output for temperature.
PH201 distributor. The jet cleaning can use a P&F HART® to convert the
III. COMMON (related to both sensor and analyzers) QUERIES AND ANSWERS:
7.34. What feature does an Assume that this is all Okay. Then FLEXA displays only known pH so we
analyzer need to connect an ion we connect the ISE to the input read pNa for pH.
selective electrode? 1 terminals of the FLEXA and a
reference electrode to the input 2 Assume the range of the analyzer
The two important features/ terminals. When we use the pNa is requested to be 1- 100 g/l NaCl
functions are the possibility to sensor (SM23-AN4), then we know that and then the output range is -0.22
change the Isopotential point the ITP is 0. (The inner buffer solution to 1.78 pNa/pH. See Q3. Now we
and the possibility to generate a is 1 mol/l NaCl) so we program ITP as use the concentration mode for salt
concentration table to have output 0 and calibrate the analyzer with a 1 concentration and use a 21 point
linear to concentration. mol solution (pNa= 0) and a 0.01 mol table function.
solution (pNa= 2). The display in the
7.35. Can I connect an ion Next you define the
selective electrode to concentration scale. Here
Flexa? the range is reversed
because the higher the
The name Ion Selective
pNa value is the lower the
Electrode (ISE) is incorrect
concentration is and you
because every ISE is also
want to have a range of
sensitive to other ions
0- 100 g/l which is 0- 100
than the one mentioned
ppt (parts per thousand=
on the specification
g/kg).
sheet. Take for example
the pNa electrode. This
responds to Sodium ion
concentration if the pH is
higher than pNa +2 and
if no other salt ions are
present. So the name fig. 7.35:
Ion Sensitive Electrode is linerarisation
better. function of analyzer
www.yokogawa.com 55
in-line measurement at process 7.37. What do you recommend 7.40. Can pH and ORP be
temperature. But this can only be to use as a laboratory measured simultaneously?
done when the in-line analyzer instrument? Most of our pH sensors have a
is properly compensated for
The PH72 is an excellent device for noble material solution ground
7.FREQUENTLY ASKED Q&A
traces of ammonia or morpholine process and diagnostic data on that no weak acids are in the water,
so ingress of acid will immediately
to increase the pH. This sample a personal computer? If so what change the pH value.
is completely unbuffered. So as are the methods?
soon as it is exposed to ambient This is difficult and can only be
air the pH will drop due to done through additional device
absorption of carbon dioxide from like FieldMate through HART
the air. communication.
7.41. When do we recommend description of the failure, lifetime error in the pH buffers, but they are
automatic cleaning/calibration? stand time and diagnostic data different from the process. An easy
from analyzer (slope, asymmetric check is to look at the diagnostic
Good pH sensors only need monthly potential, glass impedance, information on the pH analyzer to
validation, so no need to do reference impedance). see if the Asymmetry Potential is
automatic calibration as long as the high or the Slope is low. Most likely,
customer has access to the sensor. Then if it is clear from the details this is the problem.
In some applications the sensor gets that sensor needs to be sent to the
fouled seriously during usage and factory for investigation. Ask for Another cause can be the infamous
then automatic cleaning devices the document called “Clean Sensor ground loop current. This can be
like chemical spray unit or auto Declaration” to protect the health the case when you use pH sensors
retractable fittings can be installed and safety of our colleagues in the without solution grounding.
successfully. laboratory. Additional relevant
information other than the queries 7.44. Can a reference sensor be
Typical customer experience with asked for is also helpful. For
fully automatic auto calibration stored in demi-water?
example, provide the validation
systems is that the system needs results, or make note of what the Sensors that are not used need to be
more maintenance than the pH sensor reads in three different fresh stored in a solution that guarantees
sensor used to need. Plus the buffer solutions. that the sensor is ready for use.
autocal procedure often results
in miscalibration of the sensors, When the sensor is stored in demi
7.43. How can the process water the salt will be washed-out of
mixing buffers, poor washing of
calibration chamber. empty buffers reading be wrong when it is the junction of the reference cell. So
bottles. correct in the pH buffers? it is not a good solution. Yokogawa
This is the well-known pH problem uses the same salt concentration in
that we call Diffusion Potential. If the wet pocket as is used inside the
7.42. When someone states
the sensor junction is plugged, then sensor.
that a pH sensor has failed,
there is no good electrical contact
what details are needed for For combination sensors we add
between electrolyte and process.
evaluation? a trace of acid to keep the glass
This results in diffusion potential
When a sensor has failed, then the membranes active. So the best
that is directly measured as error.
following information is needed method is to keep the it and store the
The chemical composition of pH
for evaluation: model code, serial sensor back in the wet pocket when
buffers is different from process
number, application details, not in use.
liquids. So when the junction is in
bad condition, you calibrate for this
56 www.yokogawa.com
7.45. Is there any special All (FL) EXA analyzers have pH value, so they cannot be used to
consideration for pH sensor independent calibration routines for calibrate ORP analyzers when the
storage & transportation? pH and for ORP. pH sensor is used as reference cell.
REMOVE
ELECTRODE CAP
YOKOGAWA
Electrolyte
BELLOW DOWN
7.50. How to refill a bellomatic Step 1: Remove the electrode cap: Step 4: Fill the sensor with the
sensor? same as first drawing in fig. 7.50. proper electrolyte leaving 3 cm from
the top without electrolyte.
There are several methods Step 2: Push the bellow down with
described in the manual and other the backside of the pencil and insert Step 5: Assemble the electrode cap.
publications. the rubber plug in the inner tube at
the bottom of the Bellomatic.
There is also an easy method that
requires only a pencil. Assuming the Step 3: Remove the pencil. The
customer kept the original package, bellow will stay down if the plug is
there is a rubber plug that sealed the in place.
inner tube prior to installation:
58 www.yokogawa.com
Notes
www.yokogawa.com 59
Appendix 1: Chemical Compatibility
It is the customer’s responsibility These tables should be used only The result may differ if a sample
tto decide what material to use. The as a guide with no guarantee.. The contains more than one chemical. It
material used in the sample should tables show the chemical resistance is recommended to refer to multiple
be confirmed. of material to individual chemicals. tables.
0.5 1.0 O O O O X O O O O
0.05 2.0 O O O O X O O O O
COMPATIBILITY
Zinc Chloride 5 O O O O X O O O O
Iron (III) Chloride 5 O O O O X O O O O
Iron (III Nitrate) 5 1.3 O O O O O O O O O
Sodium Sulfite 5 O O O O * O O O O
Sodium Carbonate 5 11.8 O O O O O O O O O
Basic
Salt
Sodium Phosphate 5 O O O O * O O O O
Potassium Chloride 5 O O O O * O O O O
Sodium Sulfate 5 O O O O * O O O O
Neutral Salt
Calcium Chloride 5 O O O O * O O O O
Sodium Nitrate 5 8.2 O O O O * O O O O
Aluminium Chloride 5 O O O O X O O O O
Hydrogen Peroxide 1 O O O O * * * O O
Organic Oxidizing
Chlorinated Lime 1 * O O O * * * O *
Potassium Dichromate 5 4.5 O O O O * O O O O
Alcohol 10 O O O O O O O * O
Organic solvent or oil
(excluding alcohol) * O * O O O * O O
Chlorinated Solvent X O X O * * * O *
60 www.yokogawa.com
8. Chemical Resistance Table
This table should be used only as a guide and no guarantee is given. The table shows
Table 9-2
the(from TI 12B07A03-01E)
chemical resistance
of material to individual chemicals. The result may differ if a
sample contains more than one chemical.
: Excellent
: Good
: Not so good
X : Unusable Electrode
Holder material Ultrasonic transducer material Seal O-ring
Electrode solution ground tip material body
material
Polypropylene SUS 316 Hastelloy C Titanium Viton Ryton Remarks
Concentʼn Temp Judge Concentʼn Temp Judge Concentʼn Temp Judge Concentʼn Temp Judge Concentʼn Temp Judge
APPENDIX1: CHEMICAL
37 90 X
20 20 10 b 20 30 10 b 20 20
Chromic acid
COMPATIBILITY
Inorganic acid
20 40 X
10 20 14 30 X 15 43 20 40 5 20
Hypochlorous Acid
10 40 40 X
Hydrobromic acid 40 30 Strong acid
10 20 10 30 10 30 10 100 Weak acid 5 20
Nitric acid 60 X
10 80 10
57 20 57 25 X 57 30
Hydroiodic acid
57 70
3 20 6 30 5 30 5 30 90 20
Sulfuric acid
3 100 5 100 X 5 70 5 100 X 30 90
30 60 15 30 5 30 5 30 85 90
Phosphoric acid
30 100 5 b 5 b 5 60
15 80 10 b 10 b 10 b 15 30
Ammonia water
15 100 28 65 20 65 20 65
10 b 10 b 10 b 10 20
Potassium hydroxide
25 b 25 b 25 b 10 90
Sodium 20 80 20 30 20 30 20 30 10 20
Alkali
sat. 80 25 b 25 b 25 b
Calcium chloride
sat. 100
20 40 30 b X 30 b X 30 b 20 60
Ferric chloride
20 60
Sodium chloride, 20% 100 90 X 90 X 90 20
+ Saturated Cl2
(Electrolysis solution)
Seawater, Magnesium 24 24 24 24
chloride sat. 80 42 b 42 b 40 b 80
5 60 20 b 20 b 20 b 10 90 Polypropylene may
Ammonium sulfate sometimes be eroded
Nitrates Sulfates
2
20 80
Potassium bichromate 10 b 10 b 10 b
Sodium sulfide 60 80 10 b 10 b 10 90
Sodium bisulfate 10 b 10 b
20 30 X 30 30 20 X
Wet chlorine gas 40
60 X
Gases
www.yokogawa.com 61
<Toc> <Ind> <8. Chemical Resistance Table > 20
96 70 100 b 100 90
Ethyl alcohol
COMPATIBILITY
COMPATIBILITY
100 100
100 20 100 20
Chlorophenol 100 70
100 100 X
Xylene 100 20 X 100 20
100 20 X
Chlorobezene
100 100 X
Chloroform 100 20 X 100 b 100 b 100 b 100 90
100 20
100 70
Dioxane 100 90
Organic substances
100 100 X
Dichloroethare 100 70 X
100 20
Ethyl nitrate 100 105 100 90
100
Carbon tetrachloride 100 20 X 90 b 90 b 100 24 X
Trichloroethylene 100 20 X 100 b 100 b 100 b 100 90 X
Toluene 100 20 X 145 100 90
Benzophenone
100 20
Benzaldehyde 100 70 100 20
100 100 X 100 90 X
Benzyl alcohol
100 20 100 30 100 30 100 25 100 90
benzene
10 70 37 b 37 b 37 b
Fomaldehyde
10 100
Methylnaphthelen
100 20
Methyl ethyl ketone 100 90
70
Methyl alcohol 100 20 100 25 100 25
100 20 100 90 X
Nitrobenzene 100 70
100 100 X
100 20 10 b 10 b
Acetic acid 100 70
100 100 X
100 20 95 30 95 30 95 30 100 90
Phenol
100 100
Benzonic acid 100
100 20 100 20
Motor oil 100 70
100 100
Petroleum ether 100 20 100 20
100 20 101 100 20
Kerosene
100 70 X
10 40 50 100 50 100 50 100
Tartaric acid 10 60
10 80
100 70 100 25 100 180 100 180
Oil and fats
Carbon sulfide 100 20 X 100 25 100 25
Note: "b" refers to the boiling point. T0802.EPS
62 www.yokogawa.com
63
COMPATIBILITY
www.yokogawa.com
Notes
APPENDIX 2: DEFINITIONS Buffer solution Dilution value
A buffer solution is a solution with a The dilution value of a buffer solution is
Activity constant, accurate fixed pH value. defined as the variation of the pH value
The use of this solution is necessary that occurs when the solution is diluted
The activity is a measure of the to determine any deviations in the with an equal volume of water.
amount of ions chemically active in a measuring system and to allow accurate
concentration of the ions in a solution.
This means that the participation of
calibration. The most important feature Divalent
of a buffer solution is that the pH value
ions in a chemical reaction is not only remains constant when it is diluted with A divalent ion is an ion with a diva lent of
determined by the concentration but water, acid or base. positive or negative charges.
also by the presence of other ions in the
Examples are:
solution. In concentrated solutions the
activity of the ions is considerably less
Calomel
AB ↔ A+B
than the total concentration. Calomel is an insolule salt (H92CI2) used K= [A]•[B]
as part of a reference system in the [AB]
Activity coefficient reference electrode.
The activity coefficient (f) shows the Dissociation
ratio between the active concentration Concentration Dissociation is separating into positive
and the total concentration. or negative ions.
The concentration of a solution is
active concentration determined by the quantity of matter
f=
total concentration dissolved per volume or per weight of Dissociation constant (K)
the solution.
The number gives the ratio between the
An ions concentration of the separated ions in
Concentration units a matter and the concentration of the
An ions are negatively charged ions, e.g.
G per litre (or kg/m3). unseparated matter.
CI , N0 -, OH .
– 3 –
DEFINITIONS
APPENDIX 2:
determines the position of this point. two positive or negative charges, e.g.
Usually this point is at pH 7 and 0 mV. P04 3 , Cr , Fe .
– 6+ 3+
Strong acid
Note: The position of the isothermal point
Reference electrode A strong acid is one which completely
of intersection is important for
the accuracy of the temperature dissociates:
correction of the electrode system.
A reference electrode is used in
HCI H + CI
+ –
combination with an ion-specific
electrode. Its function is to generate
stable potentials independent of Titration curve
Monovalent
DEFINITIONS
APPENDIX 2:
the composition of the liquid to be
The titration curve indicates the pH
A monovalent ion contains a single measured.
curve obtained when a quantity of
positive or negative charge (examples:
reagent is added to a predetermined
H , Na , Cl etc.).
Reference system
+ + –
quantity of a reagent necessary to
Nernst equation A reference system is a combination of a neutralization a process liquid.
metal, an insoluble salt of this metal and
This equation is used to determine the electrolyte. A reference system is used in Silver chloride
potential of an ion-specific electrode. both the glass and reference electrode.
Silver chloride (AgCI) is an insoluble salt
E=E0+ nF • In A used in a reference system for both glass
RT Response time and reference electrodes.
The response time of a glass electrode
E = measured potential indicates the time it takes to follow a Salt bridge
R = gas constant (8,314 J/mol. K) T = step change in pH. Usually the reponse
absolute temperature in oK A salt bridge is used in applications
time is defined as the time taken to
n = valency of the ion where the electrolyte of a reference
reach 63% of the final value of a step
F = Faraday number (96493 Coulomb) In electrode interacts with the process. A
change.
= 2,303 log (logarithm) salt bridge is a barrier solution which is
A = ion activity Under process conditions the response compatible with the composition of the
E0 = electro-chemical normal potential time of the measuring system depends process liquid.
on a number of factors versus the
positioning of the electrode in the Weak base
Nernst factor process stream, the reference electrode
used, etc.; and, consequently, the times A weak base only partly dissociates into
The description “Nernst factor” is
stated may not always be achieved in ions.
used to describe the combination of
constants in the Nernst equation. pratice. Example:
Cu (OH)2 Cu + + 20H .
2 –
RT 8,314 x 278
x 2,303= x 2,303 = 0,0591 Selectivity constant
nF 1 x 96493
The selectivity constant indicates Weak acid
pH electrode increase in output of the ion specific
A weak acid only partly dissociates into
electrode. Concentration of strange ions
ions.
An ion-specific electrode for the required to produce the same output as
measurement of the hydrogen activity in those to which the electrode is specific. Example:
a solution. H2CO3 2H + CO3 .
+ 2–
Sensitivity
The sensitivity of a pH glass electrode is
the mV output voltage per pH change.
Theoretically, the sensitivity of a pH
electrode is 59,16 millivolts at 25°C.
www.yokogawa.com 65
APPENDIX 3: LIQUID APPLICATION DATA SHEET
Please place checkmarks in the appropriate boxes and fill in the necessary information in the blanks.
Liquid Application Data Sheet [ QMQI CY C
Customer Data / General
Project/Reference Inquiry No. :
Customer TAG :
City Street:
Telephone FAX :
e-Mail
Expected
Final Destination Month :_________ Year : __________
Delivery
needed Documention Quoted Commissioning: Yes No
(Certification ...) Services Training: Yes No
PROCESS CONDITIONS
Industry Segment Power Chemical Petrochemical Water /Waste Water
Solids (Types):
Concentration
Organics (Types):
Inorganics (Types):
Installation Data
Installation Indoor Outdoor Amb. Temp.: _________ Haz. Area Class.: __________ IP: ________
Type cleaning system Jet Cleaning Brush Chemical /Spray Ultrasonic __________
Analyzer Non Ex Intrinsic Non Incendive | Certification: _________ Distance to Sensor: _____
2 wire | 4 wire (24 - 30 VDC) 4 wire (80 - 240 VAC) | 4-20 mA/Hart PB FF
Remarks:
66 www.yokogawa.com
Notes
DEFINITIONS
APPENDIX 2:
www.yokogawa.com 67
Notes
www.yokogawa.com/us
Notes
www.yokogawa.com/us
www.yokogawa.com/us
Notes
www.yokogawa.com/us
Yokogawa Electric Corporation Yokogawa Corporation of America Yokogawa Europe B.V.
World Headquarters 12530 W. Airport Blvd., Euroweg 2, 3825 HD
9-32, Nakacho 2-chome, Sugar Land, TX 77478 Amersfoort, The Netherlands
Musashino-shi, Tokyo 180-8750 USA yokogawa.com/eu
Japan
yokogawa.com/us
yokogawa.com/jp
Yokogawa Engineering Asia PTE. LTD Yokogawa China Co., Ltd. Yokogawa Middle East &
Africa B.S.C.(c)
5 Bedok South Road, Singapore 469270, 3F Tower D, Cartelo Crocodile Building,
Singapore No.568 West Tianshan Road, Shanghai P.O. Box 10070,
200335, China Manama Building 577,
yokogawa.com/sg
Road 2516, Busaiteen 225,
yokogawa.com/cn
Muharraq, Bahrain
yokogawa.com/bh
The contents of this document are subject to change without prior notice. All rights reserved. Copyright © 2017 Yokogawa Corporation of America.
[Ed:03] Printed in USA TI 12B00A20-01E 2017-01-2020