Вы находитесь на странице: 1из 43

Universal weight transfer

controller
Four relay universal type

user's Guide

VER-60.05
Please read this product manual carefully before use.
Please keep this product manual in a safe place for
your reference.

table of Contents
1 Overview
1.1 Selection and interface configuration
1.2 main features
1.3 technical parameters
2 installation
2.1 controller fixed
2.2 Front panel schematic
2.3 Rear panel wiring instructions
2.4 Sensor and controller wiring method
3 operating instructions
3.1 button description
3.2 Function Description
3.3 weight calibration FN1
3.4 Working parameter setting FN2
3.5 communication parameter setting FN3
3.6 analog output setting FN4
3.7 Adjust the bottom end of the analog output with the top FN6
3.8 Relay output and digital input setting FN5
3.9 cumulative ingredient count query and delete
3.10 controller parameter initialization operation
3.11 sensor internal code viewing function operation
4 appendix
4.1 Command mode
4.2 MODBUS compatible communication protocol description
4.3 serial port MODBUS communication protocol example
4.3.1 03 function example
4.3.2 16 function examples
4.3.3 06 function example
4.4 continuous transmission method
4.5 large screen output mode
4.6 calibration parameter modification instructions
4.7 Calibration parameters modification application examples
4.8 Sensitivity Calibration Application Example
4.9 Relay output mode description
4.9.1 Output Mode 0 (Disable Output) Description
4.9.2 Output Mode 1 (Real-time Fixed Output) Example
4.9.3 Output Mode 2 (Real-time Sorting Output) Example
4.9.4 Output Mode 3 (External Control Fixed Value Output) Example
4.9.5 Output Mode 4 (External Control Sorting Output) Example
4.9.6 Output Mode 5 (with zero return control output) example
4.9.7 Output mode 6 (External control 3 material batching with discharge
control) example
4.9.8 Output Mode 7 (External Control 2 Fast and Slow Batching Control)
Example
4.9.9 Output mode 8 (External control 1 material fast and slow batching belt
discharge control) example
4.9.10 Output Mode 9 ( Fast and Slow Reduction Batching with Material Overload
Control) Example
4.9.11 Output Mode 10 (Automatic Weight Sorting Control) Example
4.9.12 Output Mode 11 (External Control Trigger Weight Sorting Control)
Example
4.9.13 Output Mode 12 (External Control Instantaneous Peak) Example
4.9.14 Output Mode 13 (Input and Output Self Test) Description
4.9.15 Output Mode 14 (with time hold interval fixed value output) example
4.10 Information Tips
5 Maintenance and precautions

POLICE REPORT
The controller has an internal high
voltage alternating current, please
professionals commissioning,
inspectiontest and maintenance controller.

POLICE REPORT
2
To ensure operator safety and stability
control and reliable performance, be sure
to keep the
controller connectedto good good 
connectedto good good .

NOTE ITALY
In the feed line instrument table of the
electrical gasconnected connection ,
make advance to the power source cutoff,
in a device table on power before Please like wait 30seconds b
ell again to the instrument table connectedthrough power source.

NOTE ITALY
This instrument table as static electricity sensitive senseprovided a
pparatus, in that
the use and maintenance care ofPlease note intended to adopt to
take anti- staticelectricity measures applied .

NOTE ITALY
This controller is a weight transfer control device and cannot
be used as a measuring instrument for trade settlement.

1 Overview
The universal weight transfer controller is a transmission display controller
for industrial control. It integrates weight display, analog signal
output, Modbus-RTU communication, large screen communication, digital input and
relay output. High-speed 24 -bit Σ -Δ A/D converter, 16 -
bit D/A converter, all interfaces are optically isolated. Fully consider the
complexity of electromagnetic interference in industrial sites, carefully design
software and hardware, suitable for industrial field applications such as building
materials, chemicals, food, metallurgy.
1.1 Selection and interface configuration
model Sensor Input Output Serial Serial Analog
interface switch relay port 1 port 2 quantity
A 1 Ge 4 Ge 4 way Have no Have
B 1 Ge 4 Ge 4 way Have no no
C 1 Ge 4 Ge 4 way Have Have no
D 1 Ge 4 Ge 4 way no no no

3
1.2 main features
 Built-in anti- EMC interference circuit, strong anti-electromagnetic
interference capability, stable data, suitable for industrial field
applications.
 Three calibration methods: weight calibration, calibration parameter
modification, sensitivity input calibration (free weight calibration).
 RS232 or RS485 can be selected to support Modbus-RTU communication multi-
word continuous reading and writing function.
 Can set the analog output
type: 0 ~ 20mA , 4 ~ 20mA , 0 ~ 5V or 0 ~ 10V .
 Relay output mode can be set : sorting, setting, quantification, decrement,
batching, sorting, peak and other modes.
 actual feeding weight is out of tolerance, the jog feed is
automatically started.
 Support the host computer to calibrate the controller.
 number of cycles of ingredients can be set to support unlimited batching
of ingredients.
 total number of ingredients is accumulated, and the actual ingredient
weight is read by the upper machine.
 Support a series of operations such as zeroing, peeling, starting and stopping
of the host computer communication.
 Stainless steel anti-corrosion panel, all metal shielded casing.
 24BIT  - △ type A / D conversion, the maximum number of A /
D pulses: 1,000,000 .
 A/D rate is
optional: 15 times / second, 30 times / second, 60 times / second, 10
0 times / second, 200 times / second.
 Three calibration methods: weight calibration, calibration parameter
modification, sensitivity input calibration (free weight calibration)
 . 7 bit 0.56 inch LED digital display , 10 segments beam indicated .
7 th LED indicators.
1.3 technical parameters
 excitation voltage: 5.0VDC , can be driven . 8 only 350  an analog
sensor.
 Range signal range: 1.5 ~ 40mV .
 Zero signal range: - 40 ~ 38.5mV .
 Analog current output: Max 500 Ω.
 Analog voltage output: Min 10 K Ω.
 Relay contact capacity: AC3A/ 250V , DC3A/ 30V .
 Maximum sensitivity: 0.3uV/d .
 Nonlinear: better than 0.01% FS .
 Power supply voltage range: AC 200 ~ 242V , frequency 49 ~ 51Hz ,
maximum power consumption 8 watts. The controller requires a good grounding
wire and does not share a power supply with equipment such as motors, relays
or heaters that are prone to power supply noise.
 operating temperature is: -20 °C~50 °C, humidity is 10 %
~ 95 %, no condensation.
 storage temperature is: -30 °C ~60 °C, humidity is 10 % ~ 95 %, no
condensation.

2 An installation
2.1 controller fixed

4
The controller adopts the panel installation method, and the thickness of the
front wall of the installed cabinet does not exceed4 mmPreferably, the opening
dimensions on the cabinet are as follows: ( Dimensions are in millimeters )

Controller dimensions: panel length170mm,width85mmBody


length150mm,width75mm,deep100mm.
Before installation, please remove the top rods on both sides of the casing,
then install the weighing terminal into the cabinet, and fix the two ejector pins
on both sides of the weighing terminal. It is required to tighten the controller
casing to ensure a firm installation.
2.2 Front panel schematic

 Relay1 : Relay 1 action indicator.    Relay2 :


Relay 2 action indicator.
 Relay3 : Relay 3 action indicator.    Relay4 :
Relay 4 action indicator.
 Tare : The peak detection mode is the peak hold indicator, and the
other mode is the tare active indicator.
 Zero : Zero zone indicator.              Motion :
Controls the process run indicator.
2.3 Rear panel wiring instructions
The rear panel interface diagram ( type A is an example) is as follows:

5
1 , the upper row terminal wiring instructions (if no such actual output, the
lower portion of the tagging NULL ):
Serial Upper Lower Description Remarks
number label label
1. 220VAC L AC220V power line AC220V AC power input
2. N AC220V power supply
zero line
3. E AC220V power ground
 
1. Analog V+ Voltage output The controller can select
output the voltage or current
2. COM Voltage or current output, and the output type
  output common is selected by parameters.
3. I+ Current output
4. Serial GND Reference ground Communication interface 2
port2 for RS232communication is RS232output
interface 2 communication interface 2
5. TXD The transmitting end
of RS232communication
interface 2
6. RXD Receiver
of RS232communication
interface 2
7. Serial 485B RS485-B end Communication interface
port 1:RS232 output
8. 485A RS485-A end communication
9. interface 1 andRS485 out
GND RS232 communication
put communication
interface1 /
RS485 reference ground interface can be
selected , see Note 1
10. TXD The transmitting end
of RS232communication
interface 1(connected to
the receiving end of the
host computer)

6
11. RXD The receiving end
of RS232communication
interface 1(connected to
the sending end of the
host computer)

Note 1 : The communication interface of the controller has two communication


modes : RS232 and RS485 , but it cannot be output at the same time. The
communication mode is RS232 or RS485 through the internal jumper . The data
format of the two communication methods is the same. If there is no special
requirement, the factory default internal jumper selects RS485 communication
mode.
2 , the lower row terminal wiring instructions:
Serial Upper Lower Description Remarks
number label label
1. Relay NULL Unused Four sets of
output relay normally
2. COM4 Relay 4 common contact open contact
3. outputs
NO4 Relay 4 normally open
contact
4. NULL Unused
5. COM3 Relay 3 common contact
6. NO3 Relay 3 normally open
contact
7. COM2 Relay 2 common contact
8. NO2 Relay 2 normally open
contact
9. COM1 Relay 1 common contact
10. NO1 Relay 1 normally open
contact
11. Switch NULL Unused External passive
input switch or button
12. COM Input switch common input
13. SW1 Input switch 1
14. SW2 Input switch 2
15. SW3 Input switch 3
16. SW4 Input switch 4
17. Load EXE+ Sensor is stimulating The controller
cell selects the six-
18. SEN+ Positive sensor feedback wire sensor by
19. default. The four-
EXE- Sensor negative excitation
wire sensor needs
20. SEN- Sensor negative feedback to short the
positive and
twenty SIG- Sensor negative signal positive feedback,
one. negative and
twenty SIG+ Sensor positive signal negative feedback.

7
two.  
twenty SHLD Sensor shield
three. Note 2
 
Note 2 : ▲! The connection between the sensor and the controller must be
reliable, and the shielded wire and metal casing of the sensor must be reliably
grounded. The cable does not allow the controller to be plugged and unplugged
while the power is on, preventing static electricity from damaging the controller
or sensor.
! Sensors and controllers are all static-sensitive devices. Anti-static
measures must be taken during use. It is strictly forbidden to perform electric
welding operations or other strong electric operations on the weighing platform.
In the thunderstorm season, reliable lightning protection measures must be
implemented to prevent sensors from being struck by lightning. And the damage of
the controller ensures the personal safety of the operator and the safe operation
of the controller equipment and related equipment.

2.4 Sensor and controller wiring method


Correspondence Correspondence between sensor output line and controller
category terminal
Sensor output Positive Positive Negative Negative Negative Positive Shielded
line incentive feedback incentive feedback signal signal wire
(red (black (White (green
line) line) line) Line)
Controller Positive Positive Negative Negative Negative Positive shield
terminal feed feedback letter letter

Four-wire Forward and Negative and Negative Positive Shielded


sensor positive feed short negative feed short letter letter wire
The wiring diagram of the sensor and controller is as follows:

8
 
 
 
 
 

3 operating instructions
3.1 button description
button Button name Function 1 Function 2 Function 3
[Settings] / Weigh state The currently set Pressing and
directly sets function parameters pressing other keys
relay are not saved in simultaneously in
parameters the setting state the weighing state
and the next to enter a function
function item is setting state.
entered.
[Right shift Set the digital Parameter reduction Show peeling and
key] / [Reduce flashing in FN6function cancel peeling
key] / [Tare position to the setting state
key] right
[Zero Zeroing Increase the Press [Setting Key]
key] /[Add function in parameter or number and [Zero Key] at
button] weighing state of the setting in the same time to
the setting state set the parameters.
[OK Sub-function Save the currently Quick setting of
button] /[Time item that set function time parameter in
button] enters the parameters in the weighing state
current setting state and
function in the enter the next
setting state function item
[Start and Relay output Exit the current Exit query
Stop starts or stops function item in parameters
Button] /[Exit the setting state
Button]
3.2 Function Description
The function item type is divided into a first-level function item and a second-
level function item, and the second-level function item is a sub-function item of
the first-level function item, and the first-level function item presses the [OK]
key to enter the second-level function item, and when the second-level function
item is selected, the second-level function item is selected. When the last
function item of the item is pressed, press [Set] or [Enter] to enter the next
level function item. For convenience of explanation, a list of function items is
listed for reference.
Serial Primary function Serial Secondary function item symbol (name)
number symbol (name) number
1. FN1 ( weight 1) FN1.1 X (calibration mode selection)
calibration ) 2) D XXX (division value setting)
3) CXXXXXX (range setting)
4) FN1.2 (zero calibration)

9
5) FN1.3 (load calibration)
6) XXXXXXX ( load weight input)
2. FN2 (working 1) FN2.1X ( ADC conversion rate setting)
parameter setting) 2) FN2.2X (filter mode setting)
3) FN2.3X (button clear parameter setting)
4) FN2.40X (automatic zero tracking range
setting)
5) FN2.41X (tracking time setting)
6) FN2.50X (dynamic detection parameter
setting)
7) FN2.51X (stability setting)
8) FN2.52X (return to zero division
setting)
9) FN2.6X (continuous filtering parameter
setting)
10) FN2.7X (power-on auto-zero range
parameter setting)
11) FN2.8X (intermittent filter parameter
setting)
3. FN3 (communication 1) FN3.1X (serial 1 baud rate setting)
parameter setting) 2) FN3.2X (serial port 1 communication mode
setting)
3) FN3.3XXX (Serial 1 communication address
setting)
4) FN3.4X (Serial 2 communication mode
setting)
4. FN4 (analog output 1) FN4.1X (analog output type setting)
setting) 2) FN4.2X (analog output corresponding
weight range setting)
3) LXXXXXX (the bottom of the analog
output corresponds to the weight value
setting)
4) HXXXXXX (the analog output top
corresponds to the weight value setting)
5. FN5 (relay output 1) FN5.1X (relay output mode setting)
parameter setting) 2) FN5.2X (alternate)
3) FN5.3X (start automatic zero setting)
4) FN5.4X (fast time relay mode setting)
5) FN5.5X (discharge mode setting)
6) FN5.6X (slave send settings immediately)
7) FN5.7X (Tare Enable Setting)
8) FN5.8X (cumulative count clear setting)
9) AXXXXXX (relay parameter A setting)
10) bXXXXXX (relay parameter B setting)
11) CXXXXXX (relay parameter C setting)
12) dXXXXXX (relay parameter D setting)
13) EXXXXXX (relay parameter E setting)
14) FXXXXXX (relay parameter F setting)
15) PXXXXXX (relay parameter P setting)
16) HXXXXXX (relay parameter H setting)
17) UXXXXXX (relay parameter U setting)
18) LXXXXXX (relay parameter L setting)
19) C1 X.X (start delay or peak hold time
parameter setting)
20) C2 X.X (stability delay or jog interval
time parameter setting)

10
twenty C3 X.X (preset comparison time setting)
one)
twenty C4X.X ( discharge delay time setting )
two)
twenty C5X.X ( jog time setting )
three)
twenty C6X.X ( cycle start delay time
four) setting )
25) C7XX ( cycle batching times setting )
6. FN6 (adjust the 1) AL-CAL1 (analog output bottom coarse
bottom and top of adjustment)
the analog output) 2) AL-CAL2 (fine adjustment at the bottom
of the analog output)
3) AL-CAL3 (analog output bottom fine
adjustment)
4) AH-CAL1 (analog output top coarse
adjustment)
5) AH-CAL2 ( simple adjustment of analog
output tip)
6) AH-CAL3 (analog output top fine
adjustment)
7. FN6-ESC (exit
setting)
8. LOADINC (analog
reset factory
value)
3.3 weight calibration FN1
Prepare the weight or weight of the corresponding weight before
calibration. Need to enter the correct password (when calibration weight 12111 ),
a correct password input 5 within minutes without entering again 5 minutes later if
the parameter is set to re-enter the password again.
The specific calibration steps are shown in the table below.
ste display Function Description
p name
(symbol)
1. 【 FN1 Weight In the weighing state, press [Set] and [Drop]
1】 calibration to enter FN1 .
(FN1 ) Press [OK] to enter the calibration password
input interface.
2. 【 P- Calibration The correct calibration password is 
is 12111
12111 
 .
00000 】 password Press [Right Shift Key] to select the modified
input bit;
(PASSWORD Press [Add] to modify the size of the flashing
) digits;
Press [OK] to enter the next step if the
password is correct.
Press [Exit] to return to the weighing state;
Press [Function Key] to cancel the current
operation and proceed to the next step;
If the password is incorrect, it will prompt ERR
06 .
3. 【 FN1.1 Calibration Where X is the set calibration mode.
X】 method X=0: 
X=0: Weight loading calibration
calibration(  X=0
X=0 
 is the
selection normal calibration mode, and steps steps 
 4-8
4-8 
 can
be 
be setto complete the calibration ).
calibration

11
X=1: The calibration parameters are modified.
After step 5 , go directly to step 9 .
X=2: Sensitivity input calibration. After
step 5 , go directly to step12 .
Refer to step 2 for key operation .
4. [ D XXX Graduation Where XXXX is the division value, the settable
3X】 value scale values
setting (D ​​are: 1 , 0.1, 0.01 , 0.001 , 2 , 0.2 , 0.02 , 0.002 , 5 ,0.
XXXX ) 5 , 0.05 , 0.005 , 10 , 20 , 50 .
Press [Add] to select the desired division value;
Press [OK] to save the set index value and
proceed to the next step.
5. 【 CXXXX Set the Where XXXXXX is the scale range (when
XX】 scale range calibration mode 0 and 1 ) or sensor total range
or sensor (calibration mode 2 ) , if the scale range
total is10000, enter0010000.The setup is completed
range (CXX automatically to the next step.
XXXX ) Refer to step 2 for key operation .
6. 【 FN1.2 Zero Indicates that the zero point will be calibrated.
5 】 calibration The requirement
requirement is to empty the scale and keep
( FN1.2 ) the weighing platform stable .
stable
Press [OK] to start zero calibration, the
controller will display “  ” and
the light column will be gradually extinguished.
During this period, the data will be unstable and
the light column will return to full
light. After the light column is completely
extinguished, the calibration zero point is
completed and the next step is automatically
entered.
7. 【 FN1.3 Weight Indicates that the calibration process will be
6 】 standard performed. First,
First, the weighing platform is
( FN1.3 ) required to load the weight or weight weight . It is
recommended that the loaded weight be greater
than 20 % of the  maximum weighing .
If the weight or weight is evenly distributed or
placed in the center of the weighing platform and
stabilized, press [OK] Key] Starts the range
calibration, the controller displays “ ” and
the light column is gradually extinguished.
During this period, the data is unstable and the
light column will return to full light.After the
beam is completely extinguished, the controller
automatically proceeds to the next step.
8. First【00 Enter the The controller displays “ 0000000”, at which
7 00000 】 weight point the actual weight value of the loaded
of the weight or weight is entered. If the weight value
weight is10000 , enter " 0010000 " and press the [OK]
After to be button. If the calibration is accurate, the
[0010000 ] calibra controller immediately displays the entered
ted weight. At this point, the weight calibration
( XXXXXXX step is completed and normal weighing control can
) be performed. Refer to step 2 for key
operation . Note Note that theweighing platform must
be kept stable after 5~7 steps.
Step 9~11 calibration parameters are modified ( restore the original parameters
recovery )
before modification for recovery 
For detailed application, see 4.7 Calibration Parameters Modifying Application
Examples! ! !
Examples ! ! !

12
9. 【 FN1.2 Zero Indicates that the zero calibration internal code
】 internal will be entered. Hmeans positive, - means
code input negative
(FN1.2 ) After pressing [OK], the controller displays the
【 H001000 current zero point code value, such as
】 “ H001000 ”, indicating that the current
【 H002000 internal code is “ 1000 ”.
】 At this point, the user can modify the zero
internal code value according to actual needs. If
it is changed to " 2000 ", enter
" H002000". Refer to step 2 for key operationand
press [OK] to enter the next step.
10. 【 FN1.3 Range Indicates that the span calibration internal code
!! 】 internal will be entered. Hmeans positive, - means
code input negative
(FN1.3 ) After pressing [OK], the controller displays the
【 H020000 current range code value, such as “ H020000 ”,
】 indicating that the current internal code is
【 H030000 “ 20000 ”.
】 At this point, the user can modify the zero
internal code value according to actual needs. If
it is changed to “ 30000 ”, enter
“H030000 ”. Refer to step 2 for key
operation and press [OK] to enter the next step.
11. 【 FN1.4 Calibration Indicates that the range calibration weight will
】 weight input be entered. H means positive, - means negative
(FN1.4 ) After pressing [OK], the controller displays the
current calibration weight value, such as
【 H003000 “ H003000 ”, indicating that the current
】 calibration weight is “ 3000 ”.
【 H001500 At this point, the user can modify the weight of
】 the weight according to actual needs. If it is
changed to “ 1500 ”, enter “ H001500 ”. Refer
to step 2 for key operation and press [OK] to end
the parameter input calibration.
Steps 12~13 are calibration for sensitivity
For detailed application, see 4.8 Sensitivity Calibration Application
Example! ! !
Example ! ! !
12. 【 LX.XX Sensitivity “ LX.XXXXX ” indicates the current
XXX】 input sensitivity value. The user can modify the
( LX.XXXXX sensitivity value according to actual needs. For
) example, the current sensitivity is “ 2.0 ”
【 L2.0000 or L2.00000 . If you want to change to 3.0, the
0】 input is L3.0000 , and then press the [OK] button.
【 L3.0000 】Yes. Generally, after completing this step,
0】 press [Exit] to return to the weighing state to
complete the sensitivity calibration . If the zero
point has deviation, press [Zero] to set the
zero. Refer to step 2 for key operation . To
modify the zero point internal code of the
sensitivity calibration, proceed to step 13 .
13. 【 F1.5 X Zero code [ F1.5 X 】 Sensitivity calibration zero point
】 internal internal code modification enable
code =0 does not enable the modification of the zero
【 F1.5 1 modification point; =1 enables the modification of the zero
】 enable point; =2zero point restores the factory zero
【 H001500 ( FN1.5 ) point, and updates the working zero point. Press
】 [OK]
【 H002500 After setting X=1 , you can modify the zero code

13
】 of the sensitivity calibration. For example, if
the original internal code is “ 1500 ”, you need
to change it to “ 2500 ”, then enter
“ 0002500 ”, and then press [OK]. At this
point, the sensitivity calibration operation is
all over.

3.4 Working parameter setting FN2


The user can modify the controller operating parameters as needed. These
parameters include: ADC conversion rate, filtering mode, button clearing,
automatic zero tracking range, dynamic detection parameters, continuous filtering
parameters, power-on auto-zero range parameters, and intermittent filtering
parameters. When setting parameters, if the modified parameters do not need to be
saved, press [Set] to enter the next step. If the modified parameters need to be
saved, you need to press [OK] to save and automatically enter the next parameter,
press [Exit] Returns the weighing status. The specific setting steps are shown in
the table below.
step display Function name Description
(symbol)
1. 【 FN2 】 Working Press [Set] and [Drop] at the same
parameter time toenter FN1 , then press [Set] to
setting ( FN2 ) displayFN2 .
Press [OK] to enter the next step.
2. 【 FN2.1 X 】 ADC conversion The value that X can set is 0~5 .
rate setting 0 = 7.5 Hz ; 1 = 15 Hz ; 2 = 30 Hz ; 3 = 60
(FN2.1X ) Hz ; 4 = 100 Hz ; 5 = 200 Hz .
The factory default is 4 .
Press [Add] to select the
desired ADCconversion rate;
Press [OK] to save the changes and
proceed to the next step.
Press [Set] to save the changes and go
to the next step;
3. 【 FN2.2 X 】 Filter mode The values ​​that X can set are 0 , 1 .
setting 0= continuous filtering
( FN2.2X) mode(quantitative
(quantitative feeding control
mode) ;
mode)
1 = intermittent filtering
method.(fast
(fast weighing mode)
The factory default is 0 . Set the
method as in step 2 .
4. 【 FN2.3 X 】 Button clear The values ​​that X can set
parameter are 0 , 1 , 2 , 3 .
setting 0 = prohibited;
( FN2.3X ) 1 = allowable zero range ± 50%
FS ;
2 = allowable zero range ± 100%
FS ;
3= Allow zero range is unlimited
The factory default is 3 . Set the
method as in step 2 .
5. 【 FN2.40 X 】 Automatic zero The value that X can set is 0~9 .
tracking range 0 = forbidden
setting 1= Allow automatic zero
( FN2.40X) tracking 1d/ sec
3= Allow automatic zero
tracking 3d/ sec

14
9= Allow automatic zero
tracking 9d/ sec
The factory default is 3 . Set the
method as in step 2 .
6. 【 FN2.41 X 】 Automatic zero The value that X can set is 0~9 . Unit
tracking time seconds
setting 0 = disabled; 1 = allowed, automatic zero
( FN2.41X) tracking time 1 second
The factory default is 5 . Set the
method as in step 2 .
7. 【 FN2.50 X 】 Dynamic The value that X can set is 0~7 .
detection 0 = disable dynamic detection
parameter 1= Allow, dynamic detection
setting sensitivity 1d
( FN2.50X ) 2= Allow, dynamic detection
sensitivity 2d
7= Allow, dynamic detection
sensitivity 7d
The factory default is 2 . Set the
method as in step 2 .
8. 【 FN2.51 X 】 Settling time The value that X can set is 0~9 , the
setting unit is 0.1 second.
( FN2.51X) 0 = stable judgment is prohibited; 1
=allowed, stable judgment time
is 0.1 second
The factory default is 2 . Set the
method as in step 2 .
9. 【 FN2.52 X 】 Zero division X can be set to a value of 0~9d , unit
number setting division value
(FN2.52X ) 0= zero return processing is
prohibited; 1 = allowed, zero return
range 1 division value
The factory default is 4 . Set the
method as in step 2 .
10. 【 FN2.6 X 】 Continuous The value that X can be set is 0 to 9.
filtering Thenumber represents the filter
parameter strength. The larger the value, the
setting stronger the filtering degree, and the
( FN2.6X ) corresponding stabilization time will be
longer.
The factory default is 4 . Set the
method as in step 2 .
11. 【 FN2.7 X 】 Start auto zero The values ​​that X can set
range parameter are 0 , 1 , 2 , 3 ,4 .
setting 0 = disabled, the zero point is the
( FN2.7X) calibration zero
1= Automatic zero setting range
± 4% FS
2= Automatic zero setting range
± 10% FS
3= Automatic zero setting range
± 20% FS
4=The starting zero remains
unchanged, that is, the last zero is
set.
The factory default is 4 . Set the
method as in step 2 .
12. 【 FN2.8 X 】 Intermittent The value that X can set is 0 to 3.

15
Filter Thenumber represents the filter
Parameter strength. The larger the value, the
Settings stronger the filtering degree and the
( FN2.8X) corresponding stabilization time will be
longer. The factory default
is 1 . Set the method as in step 2 .
3.5 communication parameter setting FN3
The settable communication parameters include serial port 1 baud rate,
communication mode, communication address and serial port 2 communication
mode. The local address can be set when the controller communicates. The specific
setting steps are shown in the table below.
step display Function name Description
(symbol)
1. 【 FN3 】 Communication FN3 is a communication parameter setting
parameter setting function with sub-function items.
( FN3 ) Press [Set] and [Drop] at the same
time toenter FN1 , then press [Set] to
cycle to display FN3 .
Press [OK] to enter the next step.
2. 【 FN3.1 X 】 Serial port 1 baud The values ​​that X can set
rate setting are 0 , 1 , 2 , 3 , 4.
(FN3.1X ) 0 = 1200bps ; 1 = 2400bps ; 2 = 4800bps ;
3=9600bps ; 4=19200bps ;
The factory default is 4 .
Press [Add] to select the desired baud
rate;
Press [OK] to save the changes and
proceed to the next step.
Press [Set] to save the changes and go
to the next step;
3. 【 FN3.2 X 】 Serial The values ​​that X can set are 0 , 1 ,
1
port communicatio and 2 .
n mode setting 0 = MODBUS-
(FN3.2X ) RTU output; 1 = ASCIIcontinuous
output; 2 = Yaohua or Keli large screen
output;
The factory default is 0 . Set the
method as in step 2 .
4. 【 FN3.3XXX Serial XXX indicates the local address of the
】 port 1communicatio
multi-machine communication in command
n address mode. The value that can be set
setting is 0~255 .The factory default is 1 .
( FN3.3XXX ) Press [Right Shift Key] to select the
modified bit;
Press [Add] to modify the size of the
flashing digits;
Other setting methods are the same as
step2
5. 【 FN3.4 X 】 Serial The values ​​that X can set are 0 , 1 .
port 2communicatio 0 = no output; 1 = Yaohua or Keli large
n mode setting screen output;
(FN3.4X ) The factory default is 0 . Set the
method as in step 2 .

3.6 analog output setting FN4

16
When the controller has an analog output module, the analog output type can be
set by the FN4 function item and the weight transfer range of the analog
output can be selected. Analog Output Type 0 ~ 20mA , .
4 ~ 20mA , 0 ~ 5V and 0 ~ 10V . The weight transfer range of the
analog output can be from zero to maximum weighing or it can be the selected
weight range. When setting parameters, if the modified parameters do not need to
be saved, press [Set] to enter the next step. If the modified parameters need to
be saved, you need to press [OK] to save and automatically enter the next
parameter, press [Exit] Returns the weighing status. The specific setting steps
are shown in the table below.
step display Function name Description
(symbol)
1. 【 FN4 】 Analog output FN4 sub-function item
setting ( FN4 ) Press [Set] and [Drop] at the same
time toenter FN1 , then press [Set] to
cycle to display FN4 .
Press [OK] to enter the next step.
2. 【 FN4.1 X 】 Analog output The values ​​that X can set
type setting are 0 , 1 , 2 , 3 .
(FN4.1X ) Current output: 0 = 0 ~ 20mA ; 1 =
4 ~20mA ;
Voltage
output: 2=0 to 5V ; 3=0 to 10V .
The factory default is 1 .
Press [Add] to select the desired
analog output type;
Press [OK] to save the changes and
proceed to the next step.
Press [Set] to save the changes and go
to the next step;
(Note that the controller needs to be
powered on again after the analog
output type is changed)
3. 【 FN4.2 X 】 Analog output The values ​​that X can set are 0 , 1 .
corresponding 0 means that the weight transfer range
weight range is from zero to the maximum weighing,
setting when the zero point corresponds to the
( FN4.2X) bottom end of the analog output, and
when the weight is the maximum
weighing, it corresponds to the top of
the analog output;
1 indicates that the weight transfer
range is the selected weight range
interval.
The low end of the weight interval
corresponds to the bottom of the analog
output, and the high end corresponds to
the top of the analog output.
The factory default is 0 .
Press [Add] to select the corresponding
analog output corresponding weight
range; press [OK] to save the
corresponding analog output
corresponding weight range. If FN4.2=0 ,
enter FN5 ; if FN4.2=1 , go to the next
step.
Press [Set] to save the changes.
IfFN4.2=0 , enter FN5 ; if FN4.2=1 , go

17
to the next step.
4. 【 LXXXXXX 】 The bottom of Set the weight value corresponding to
the analog the bottom of the analog output.
output Press [Right Shift Key] to select the
corresponds to modified bit;
the weight Press [Add] to modify the size of the
value setting flashing digits;
(LXXXXXX ) Press [Set] to save the changes and go
to the next step;
Press [OK] to save the changes and
proceed to the next step.
5. 【 HXXXXXX 】 Analog output Set the weight value corresponding to
top the top of the analog output.
corresponding Set the method as in step 4 .
weight value
setting
(HXXXXXX )

3.7 Adjust the bottom end of the analog output with the
top FN6
Four analog outputs are calibrated before leaving the factory, and you only need
to select the type of analog output. If necessary, you can adjust the bottom and
top values ​​of the currently selected analog output in the FN6 function. For
example, the currently selected analog output type is 0 to 5V , and
the FN6 function can be adjusted to 0.5 to 4.5V . When adjusting, the
analog output terminal is connected to a high-precision voltage or ammeter, and
the steps are set according to the following table. When the requirement is
reached, the setting is exited. In addition, the controller also has a factory
output recovery function for analog output values. The specific setting steps are
shown in the table below. (Note: changes need to enter the correct password when
the analog output, first enter the correct password after 5 do not need to re-
enter within minutes, 5 minutes later if you modify the analog output again to
re-enter the password.)
step display Function name Description
(symbol)
1. 【 FN6 】 Adjust the In the weighing state, press [Setting
bottom and top Key] and [Target Key] simultaneously
of the analog to enterFN6 .
output (FN6 ) Press [OK] to enter the next step.
Press [Set] to enter ESC .
2. 【 P-00000 】 Password input The correct password is 13222
(PASSWORD ) Press [Right Shift Key] to select the
modified bit;
Press [Add] to modify the size of the
flashing digits;
Press [OK] to enter the next step if
the password is correct, and ERR 06 if
the password is incorrect .
3. [ AL-CAL1 】 Analog output The bottom of the analog output is
bottom coarse roughly adjusted, and a high-precision
adjustment voltmeter or ammeter is connected to the
( AL-CAL1 ) corresponding analog output according to
the current analog output type. Observe
the values ​​on the watch and adjust
them as follows:
Press [Add] to increase the bottom value

18
of the analog output;
Press [Reduce Key] to decrease the
bottom value of the analog output;
Once adjusted to meet the requirements,
press [OK] to save the adjustment value,
the controller automatically enters the
next step.
4. 【 AL-CAL2 】 Fine adjustment Fine adjustments are made at the bottom
of the bottom of the analog output.
of the analog The adjustment method is the same as
output ( AL- step 3.
CAL2 )
5. 【 AL-CAL3 】 Analog output Minor adjustments are made at the bottom
bottom fine of the analog output.
adjustment The adjustment method is the same as
( AL-CAL3 ) step 3.
6. 【 AH-CAL1 】 Analog output The analog output tip is roughly
top coarse adjusted.
adjustment The adjustment method is the same as
( AH-CAL1 ) step 3.
7. 【 AH-CAL2 】 Analog output Fine adjustments are made at the top of
top fine the analog output.
adjustment The adjustment method is the same as
(AH-CAL2 ) step 3.
8. 【 AH-CAL3 】 Analog output Minor adjustments are made at the top of
top fine the analog output.
adjustment The adjustment method is the same as
(AH-CAL3 ) step 3.
9. [ ESC ] Exit setting Press [Set] to enter LOADINC ;
status ( FN6- Press [OK] to exit the setting and
ESC) return to the weighing state.
10. 【 LOADINC 】 Analog to The four analog output values ​​are
factory default restored to the factory values.
(LOADINC ) Press [Set] to return to FN6 .
Press [OK] to restore the four analog
output values ​​to the factory defaults,
then display OFF-ON , wait for the user
to shut down and restart.
3.8 Relay output and digital input setting FN5
Table 1 : Output Control Mode
Setting Relay mode Relay 1 Relay 2 Relay 3 Relay 4
s
(FN5.1
)
=0 Prohibit output shut down shut down shut down shut down
=1 Real- Fixed value1 Setting 2 Setting 3 Setting 4
time fixed
output
=2 Real- Lower limit Upper limit qualified Unused
time sorting
output
=3 External Fixed value1 Setting 2 Setting 3 Setting 4
control fixed
value output
=4 External Lower limit Upper limit qualified Unused
control sorting

19
output
=5 With zero Fixed value1 Setting 2 Setting 3 Setting 4
return self-
locking value
output
=6 3 material Material 1 Material 2 Material 3 Discharge
batching with
material
discharge
=7 2 material Material 1fa Material 1sl Material 2fa Material 2 sl
speed st ow plus st ow plus
ingredients
=8 1 material fast Quick add Slow plus Discharge Unused
and slow
ingredient belt
discharge
=9 Fast and slow Fast Slow Short overload
reduction decrease reduction material
batching with feeding
material
overload
control
=10 Automatic Lower limit Maximum Qualified Unused
weight sorting of inspection
inspection limit
=11 External Lower limit Maximum Qualified Unused
control trigger of inspection
weight division inspection limit
=12 External Peak lower Peak ceiling
Peak Unused
control limit qualified
instantaneous
peak
=13 Relay output Switch 1 Switch 2 Switch 3 Switch 4
self test
=14 With holding Interval 1 Interval 2 Interval 3 Interval 4
interval fixed
value output
Table 2 : Switch input function
Setting Relay mode Switch 1 Switch 2 Switch 3 Switch 4
method
(FN5.1 )
=0 Prohibit output no no no Zero
=1 Real-time fixed no no no Zero
output
=2 Real- no no no Zero
time sorting
output
=3 External control Start /sto no no Zero
fixed value p
output
=4 External control Start /sto no no Zero
sorting output p
=5 With zero return no no no Zero
self-locking
value output
=6 3 material Start /sto Suspend /continu Discharge no
batching with p e
material

20
discharge
=7 2 material speed Start /sto Suspend /continu no no
ingredients p e
=8 1 material fast Start /sto Suspend /continu Discharge no
and slow p e
ingredient belt
discharge
=9 Fast and slow Start /sto Suspend /continu no no
reduction p e
ingredients
=10 Automatic weight Start /sto no no no
sorting p
=11 External control Start /sto Trigger switch
trigger weight p
division
=12 External control Start /sto no no no
instantaneous p
peak
=13 Relay output self Relay 1 Relay 2 Relay 3 Relay 4
test
=14 With holding no no no no
interval fixed
value output
Table 3 : Relay parameter setting steps
Remark: Press [Recipe Key] to enter the following step 9 in the shortcut
mode . You can query and set a total of 16 relay parameters from Step 9 to
Step 25 .
step display Function name Description
(symbol)
1. 【 FN5 】 Relay output Press [Set] and [Drop] at the same
setting ( FN5 ) time toenter FN1 , then press [Set] to
cycle to display FN5 .
Press [OK] to enter the next step.
2. 【 FN5.1 XX 】 Relay output The value that XX can be set is 0~14 ,
mode setting see Table 1 above .
(FN5.1XX ) If the user does not use the relay
output function, it is recommended to
set the relay output mode to 0 .
The factory default is 1 .
Press [Add] to select the desired relay
output mode;
Press [Right Shift Key] to select the
modified bit;
Press [OK] to save the changes and
proceed to the next step.
Press [Set] to save the changes and go
to the next step;
3. 【 FN5.2 X 】 Alternate The value that X can set
parameter is 0~2 . Alternate parameter
setting The factory value is 0 . Set the method
( FN5.2X ) as in step 2 .
4. 【 FN5.3 X 】 Start automatic The values ​​that X can set
zero setting are 0 , 1 , and 2. Set the method as
(FN5.3X ) in step 2 .
The tare weight is automatically
calculated before X=0 is fed, and
peeling is performed.

21
Zero is set before X=1 is added .
Zeroing and peeling
are not performedbefore X=2 feeding 
. Factory default is 1
5. 【 FN5.4 X 】 Quick add mode The value that X can set is 0~1 .
setting =0 : When adding quickly, the fast and
( FN5.4X) slow plus relays are output at the same
time.
=1 : When feeding fast, the output is
added quickly, and the slow plus relay
does not output.
The factory value is 0 . Set the method
as in step 2 .
6. 【 FN5.5 X 】 Discharge mode The value that X can set is 0~1 .
setting =0 : After the ingredients are completed,
( FN5.5X) the discharge is automatically started.
=1 : After the ingredients are finished,
you must manually press [Discharge
Switch] before starting the discharge.
The factory value is 0 . Set the method
as in step 2 .
7. 【 FN5.6 X 】 Slave The value that X can set is 0~1 .
immediately send =0 : After the MODBUS-
settings RTU communication slave receives the
(FN5.6X ) command, it will respond with a delay
of 3T time.
=1 : The MODBUS-RTU communication
slave responds immediately after
receiving the command.
The factory value is 1 . Set the method
as in step 2 .
8. 【 FN5.7 X 】 Peeling enable The value that X can set is 0~1 .
setting =0 : The controller does not allow the
( FN5.7X) tare operation.
=1 : The controller allows the tare
operation.
The factory value is 0 . Set the method
as in step 2 .
9. 【 AXXXXXX Relay Setting the relay parameters A .
】 parameter Asetti
Press [Right Shift Key] to select the
ng modified bit;
( AXXXXXX ) Press [Add] to modify the size of the
flashing digits;
Press [Set] to save the changes and go
to the next step;
Press [OK] to save the changes and
proceed to the next step.
10. 【 bXXXXXX Relay Set the relay parameter b .
】 b
parameter settiSet the method as in step 9 .
ng
( bXXXXXX )
11. 【 CXXXXXX Relay Setting the relay parameters C .
】 parameter Csetti
Set the method as in step 9 .
ng
( CXXXXXX )
12. [ dXXXXXX ] Relay Set the relay parameter d .
parameter dsetti
Set the method as in step 9 .
ng

22
( dXXXXXX )
13. [ EXXXXXX Relay Set relay parameters E .
】 parameter Esetti
Set the method as in step 9 .
ng
( EXXXXXX )
14. 【 FXXXXXX Relay Set relay parameters F .
】 parameter Fsetti
Set the method as in step 9 .
ng
( FXXXXXX )
15. 【 PXXXXXX Relay Set the relay tolerance value 1 .
】 P
parameter settiSet the method as in step 9 .
ng
( PXXXXXX )
16. 【 HXXXXXX Relay Set the relay tolerance value 2 .
】 parameter Hsetti
Set the method as in step 9 .
ng
( HXXXXXX )
17. 【 UXXXXXX Relay Set the relay tolerance value 3 .
】 parameter Usetti
Set the method as in step 9 .
ng
( UXXXXXX )
18. 【 LXXXXXX Relay Set the zero zone value.
】 parameter Lsetti
Set the method as in step 9 .
ng
( LXXXXXX )
19. [ C1 X.X 】 Delay Set the start delay or peak hold time,
parameter 1setti which ranges from 0.0 to 9.9 seconds;
ng Set the method as in step 9 .
( C1 XX )
20. [ C2 X.X 】 Delay Set the stabilization time or jog
parameter 2setti interval time, the value range
ng is 0.0~9.9 seconds;
( C2 XX ) Set the method as in step 9 .
twenty [ C3 X.X 】 Delay Set the prohibition comparison time, the
one. parameter 3setti value range is 0.0~9.9 seconds;
ng Set the method as in step 9 .
( C3 XX )
twenty 【 C4 X.X 】 Delay Set the discharge delay time, the value
two. parameter 4setti range is 0.0~9.9 seconds;
ng Set the method as in step 9 .
( C4 XX )
twenty [ C5 X.X 】 Delay Set the jog feed time, which ranges
three. parameter 5setti from0.0 to 9.9 seconds;
ng Set the method as in step 9 .
( C5 XX )
twenty [ C6 X.X 】 Delay Set the cycle ingredient delay, the
four. parameter 6setti value range is 0.0~9.9 seconds;
ng Set the method as in step 9 .
( C6 XX )
25. 【 C7 XX 】 Cycle number Set the number of batching cycles, the
setting value range is 00~99 , if set to 99, it
( C7 XX ) will loop indefinitely;
Set the method as in step 9 .
3.9 cumulative ingredient count query and delete
The specific operation steps are shown in the table below.
step display Function name Description

23
(symbol)
1. 【 P-00000 】 Password input In the weighing state, press [OK]
( PASSWORD ) and [Exit] to enter the password
input interface.
The correct password is 12111
Press [Right Shift Key] to select
the modified bit;
Press [Add] to modify the size of
the flashing digits;
Press [OK] to enter the next
step if the password is correct;
Press [Exit] to return to the
weighing state.
If the password is incorrect, it
will prompt ERR 06 .
2. 2, 【 CXXXXXX】 Show cumulative Where XXXXXX is the cumulative
counts (CXXXXXX number, if the cumulative number
) is10 , C000010 is
displayed . Press [OK] to enter
the next step.
Key operation reference step 1
3. 5, 【 FN5.8 X】 Cumulative number Where X is the cumulative number
of deletions of times to delete the enable bit
enabled =0 does not delete the
( FN5.8 X ) accumulated number of times;
=1 delete the cumulative number
of times;
To delete the cumulative number,
enter " 1" and press [Enter].
3.10 controller parameter initialization operation
step display Function name Description
(symbol)
1 【 P-00000 】 Password input When the controller is powered on the
( PASSWORD ) self-test status , press
status [OK] and
[Exit] at the same time to enter the
password input interface.
Note: The initialization work parameter
password is 22222 , and the
initialization calibration parameter
password is 33333 .
Press [Right Shift Key] to select the
modified bit;
Press [Add] to modify the size of the
flashing digits;
Press [Exit] to return to the weighing
state.
If the password is incorrect, it will
prompt ERR 06 .
After entering the correct password,
press [OK] to complete the
initialization. Note: The parameter
initialization must be performed when the
controller is powered on.
3.11 sensor internal code viewing function operation
When the controller shows that the weight is unstable (or the display 0 does
not change, or the abnormal value such as overload is displayed), the sensor

24
failure factor is eliminated by checking whether the sensor internal code is
stable.
step display Function name Description
(symbol)
1 【 xxxxxx 】 Display the In the weighing state, press [ Setting
code value Key ] and [ OK ] simultaneously
inside the toenter the internal code display
sensor interface. At this time, the
outputindicators 1 , 2 , 3 ,
and 4 are on, and the internal code
value of the sensor is in the range of
about350,000 . Whether the stability and
linearity change can detect whether the
sensor is good or bad. Then press
the [ Exit ] or OK key ] or []
weighing mode setting key.

4 appendix
4.1 Command mode
Baud rate: 1200/2400/4800/9600/19200 (optional) 8 -bit data bit, 1 start
bit, 1 stop bit, no parity.
Note: The value transmitted by the actual weight value in the command mode is a
signed integer number, that is to say, it does not contain a decimal point. For
example, to transmit the gross weight value, the weight is1.386 kg(ton, gram,
kilogram), this value is also the controller display value, the transmitted weight
value is actually 1386 hexadecimal number, ie 0X56A, with a signed 16- bit
number, the high byte is 0X05 , the low byte is 0X6A; is represented by a
signed long integer, from high byte to low byte are 0X00 , 0X00 ,0X05 , 0X 6A.

4.2 MODBUS compatible communication protocol


description
When the parameter [FN3.2=0] , the Modbus compatible communication mode is
selected . The bus can only select RS232 or RS485 mode and select by internal
jumper. MODBUS is a master-slave network communication protocol. This weighing
terminal is called as a slave in the MODBUS network and is called by the host
system. The data format is RTU mode and
supports 03 , 06 and 16 functions. Holding register 40001 , the data
address in the message is register 0000 . The function code area is the
operation specified by the hold register type, therefore, "4XXXX" is the default
address type. For example: holding register 40001 addressable register
address value 0000hex ( + decimal 0) ; Holding Register 40011 addressable
register address 000Ahex (10 decimal 10) . Up to 30 internal registers can
be read continuously at a time using the 03 function . The 16 functions
write up to 20 registers at a time. Weigh the data mapping
address in modbus :
Content Description Remarks
address

40001 Controller display weight ( signed 16 - Read only (function


bit, short ) -32768~32767 code 03 )

40002 Decimal point value Read only (function


( 0 , 1 , 2 , 3 ) code 03 )

25
40003- Controller display weight (signed 32 -bit, Read only (function
40004 long) code 03 )

40005- Tare weight (signed long) Read only (function


40006 code 03 )

40007 The status output word is defined as Remark: When the actual
follows: weight storage mark of the
ingredientchanges from 0 to
Bit 0: Display weight peeling mark  1 , the upper machine can
read the actual weight for
Bit 1: Display weight stability sign storage.
Bit 2: Display weight zero zone sign

Bit 3: Discharge process flag

Bit 4: Relay 1 output flag

Bit 5: Relay 2 output flag

Bit 6: Relay 3 output flag

Bit 7: Relay 4 output flag

Bit 8: Ingredients actual weight


preservation mark

Bit 9 to Bit 15: Invalid

40008- Relay parameter A value, written data is Read and write (function


40009 simultaneously written to the internalEEPROM code03,16 )

40010- Relay parameter B value, the written data Read and write (function
40011 is simultaneously written into the code03,16 )
internalEEPROM

40012- Relay parameter C value, written data is Read and write (function


40013 simultaneously written to the internalEEPROM code03,16 )

40014- Relay parameter D value, the written data Read and write (function
40015 is simultaneously written into the code03,16 )
internalEEPROM

40016- Relay parameter E value, written data is Read and write (function


40017 simultaneously written to the internalEEPROM code03,16 )

40018- Relay parameter F value, the written data Read and write (function
40019 is simultaneously written into the code03,16 )
internalEEPROM

40020 Tolerance P value, written data is Read and write (function


simultaneously written to the internalEEPROM code03,16 )

40021 Tolerance H value, written data is Read and write (function


simultaneously written to the internalEEPROM code03,16 )

40022 Tolerance U value, written data is Read and write (function


simultaneously written to the internalEEPROM code03,16 )

40023 Zero zone weight L value, the written data Read and write (function
is simultaneously written into the code03,16 )
internalEEPROM

40024 Cumulative number Read only (function code 03,


06, 16 )

40025- Actual weight Read only (function


40026 code 03,16)

40027 Bit  Clear (1 valid ) Write only (function


0 code06,16 )

26
Bit  Peeling (1 effective ) Write only (function
1 code06,16 )

Bit  Start and stop control bit (1 valid ) Write only (function


2 code06,16 )

40028 Calibration handshake Write (function code 06 )


signal ( signed 16bits ) , write 0x88
is valid

40029 Zero calibration ( signed 16 bits ) , Write (function code 06 )


write0x00 is valid

40030 Range calibration ( signed 16 bits ) , Write (function code 06 )


write the actual calibration weight

4.3 serial port MODBUS communication protocol


example

4.3.1 03 function example


The following controller addresses are set in FN3.3 .
1. The upper machine-readable controller displays the weight (signed 16 -bit, ie
read save register 40001 )
information Controller function Register start Number of CRC check code
address code address to be registers read ( 16bit)
16
read ( bits) 16
( bits)
High Low byte High Low byte High Low byte
byte byte byte
HEX ( 16decimal) 01 03 00 00 00 01 84 0A
Controller is responsive to: ( assuming that the display wt =
1000 ( 16 hexadecimal: 0x03E8 ) )
information Controller function Controller Returns the data CRC check code
address code returns of ( 16bit)
the number the 40001 register
of bytes (16 bits)
High Low High Low byte
byte byte byte
HEX ( 16decimal) 01 03 02 03 E8 B8 FA
2 , the upper machine read controller display weight (signed long integer, ie read
and hold register 40003-40004 )
information Controller function Register start Number of CRC check code
address code address to be registers read ( 16bit)
read (16 bits) ( 16 bits)
High Low byte High Low byte High Low byte
byte byte byte
HEX ( 16decimal) 01 03 00 02 00 02 65 CB
Controller is responsive (
to: assuming that the display wt =
80,000 ( 16 hexadecimal: 0x00013880 ) )
information Control functi Control Returns the Returns the CRC check
ler on ler data of data of code
address code returns the 40003regis the 40004regi ( 16 bit)
the ter ( 16 bits) ster
number ( 16bits)
of High Low High Low Hig Low
bytes byte byte byte byte h byte
byt

27
e
HEX ( 16deci 01 03 04 00 01 38 80 B9 93
mal)

4.3.2 16 function examples


1. The upper computer writes the relay parameter A value (that is, the write holding
register 40008~40009 , and the written number is a signed long integer number, for
example, setting C value =70000 (0X00011170) )
informati Contr func Register Number Numb Write dat Write data CRC check
on oller tion start of er a to  to code
addre code address register of the 40008re the 40009reg ( 16 bit)
ss to be s to byte gister ister
written write s ( 16bits) ( 16bits)
(16 bits ( 16bits writ
) ) ten
Hi Low Hi Low High Low High Low Hi Low
gh byt gh byt byte byte byte byte gh byte
by e by e by
te te te
HEX ( 16d 01 10 00 07 00 02 04 00 01 11 70 E 3D
ecimal) E

Controller response:
information Controller function Register start Number of CRC check code
address code address to be registers to ( 16bit)
written write ( 16 bits)
( 16 bits)
High Low byte High Low byte High Low byte
byte byte byte
HEX ( 16decimal) 01 10 00 07 00 02 F0 09
2 , the host computer uses 16 functions, from the address 40008 unit to
write 13 words, write relay parameters A = 1000 , b = 2000 , C = 3000 , D = 4000 , E =
60 , F = 50 , P = 20 , PC to send 16 hexadecimal code is as follows: 01 10 00 07 00
0D1a 00 00 03 e8 00 00 07 d0 00 00 0b b8 00 00 0f A0 00 00 00 3c 00 00 00 32 00 14 68
2d

4.3.3 06 function example


1. The upper computer writes the clear command (ie, write hold register 40027 )
information Controller function Register Write data to CRC check code
address code address to be the40027 register ( 16bit)
written ( 16bits)
( 16 bits)
High Low byte High Low High Low byte
byte byte byte byte
HEX ( 16decimal) 01 06 00 1A 00 01 69 CD

Controller response:

28
information Controller function Register Write data to CRC check code
address code address to be the40027 register ( 16bit)
written ( 16bits)
( 16 bits)
High Low byte High Low High Low byte
byte byte byte byte
HEX ( 16decimal) 01 06 00 1A 00 01 69 CD
2 , the host computer write start batching command (ie write hold register 40027 )
information Controller function Register Write data to CRC check code
address code address to be the40027 register ( 16bit)
written ( 16bits)
( 16 bits)
High Low byte High Low High Low byte
byte byte byte byte
HEX ( 16decimal) 01 06 00 1A 00 04 A9 CE

Controller response:
information Controller function Register Write data to CRC check code
address code address to be the40027 register ( 16bit)
written ( 16bits)
( 16 bits)
High Low byte High Low High Low byte
byte byte byte byte
HEX ( 16decimal) 01 06 00 1A 00 04 A9 CE

3. The host computer can use the 06 function to perform weight calibration on the
controller. It is necessary to send 4 commands with the 06 function , first
perform zero calibration, and then perform range calibration. The calibration
steps are as follows:
A. Send the calibration handshake instruction
( write 0x88 in the 40028 unit with
the 06 function ): 01 06 00 1b 0088 f9 ab
B. Send the zero calibration command when the sensor is
in no-load state ( write 0x00 in the 40029 unit with
the 06 function ):
01 06 00 1c 00 00 48 0c
C, then send the calibration handshake instruction
( write 0x88 in the 40028 unit with
06
the  function ):  01 06 00 1b 0088 f9 ab
D. Send the range calibration command under the sensor loading state
(calibration weight
is 4500 , write 0x1194 in 40030 unit with 06 function ): 01 06 00 1d
11 94 14 33

4.4 continuous transmission method


Baud rate: 1200/2400/4800/9600/19200 (optional)
Each frame of data consists of 10 bits, 8 data bits, 1 start bit, 1 stop
bit, no parity.
The data appears on both the RS232 and RS485 buses. The data is consistent

29
with the weight content displayed by the controller. Each group of data
contains 10 frames. The first frame is the data start character frame “=”,
followed by 7 data frames, and the high-order invalid 0 is filled with “ ”
(space). If the displayed value is negative, the highest bit of the data frame
will be sent "-". If the meter displays a decimal point, the weight data contains
" . " decimal point characters, and finally the 2 frame end character.
Starter Display weight data Terminator
= High           Low 0D 0A
position position

For example: the controller displays: “ 12345 ”, the serial port sends the
data “=  12345 ”.
Starter Display weight data Terminator
=     1 2 3 4 5 0D 0A

The controller displays: “ 1234.5 ”, the serial port sends data


“= 1234.5 ”.
Starter Display weight data Terminator
=   1 2 3 4 . 5 0D 0A

The controller shows: "-1234.5 ”, serial port sends data "=-1234.5 ”.


Starter Display weight data Terminator
= - 1 2 3 4 . 5 0D 0A

4.5 large screen output mode


When the parameter [F3.2=2] is selected, the large screen output mode is
selected. The large screen signal is RS232 signal, serial output is binary code,
the baud rate is fixed at 600 , and each frame data has 11 bits and one start bit
( 0 ), 8 data bits (lower bit first), 1 flag bit, 1 stop bit ( 1 ). The meter sends a
set of data at regular intervals. Each set of data contains 3 frames of data. The
meaning of the 3 frames of data is shown in the table below.
First . 1 frame of data:
The content Comment
first X -bit
0 Start bit Binary 0
1 D0 D0 , d1 , and d2 are
2 D1 decimal places ( 0~3)
3 D2
4 D3 Weight symbol
( 1negative 0 positive)
5 D4 Unused
6 D5 Weighing data G16
7 D6 Weighing data G17
8 D7 Weighing data G18
9 Flag bit Binary 0
10 Stop bit Binary 1
Remarks: G0~G18 consists of low-to-high 19 -digit
binary code
Of 2 frame data:

30
The content Comment
first X -bit
0 Start bit Binary 0
1 D0 Weighing data G8
2 D1 Weighing data G9
3 D2 Weighing data G10
4 D3 Weighing data G11
5 D4 Weighing data G12
6 D5 Weighing data G13
7 D6 Weighing data G14
8 D7 Weighing data G15
9 Flag bit Binary 0
10 Stop bit Binary 1
Remarks: G0~G18 consists of low-to-high 19 -digit
binary code
Of . 3 frame data:
The content Comment
first X -bit
0 Start bit Binary 0
1 D0 Weighing data G0
2 D1 Weighing data G1
3 D2 Weighing data G2
4 D3 Weighing data G3
5 D4 Weighing data G4
6 D5 Weighing data G5
7 D6 Weighing data G6
8 D7 Weighing data G7
9 Flag bit Binary 1
10 Stop bit Binary 1
Remarks: G0~G18 consists of low-to-high 19 -digit
binary code
4.6 calibration parameter modification instructions
1. Calibration parameter modification input directly affects the
weight calibration parameters. Therefore, it is recommended to back up
the original calibration parameters first.
2. The calibration parameter is modified to adjust the working zero
point. To set the zero to zero, set FN2.7=0 (turn off the power to
zero).
3, the scale work zero point should be increased, then the zero point
code value should be reduced, the scale work zero point should be
lowered, then the zero point code value should be increased.
4. The weight of the working zero of the scale can be adjusted by
modifying the inner code value of the zero point. The calculation
formula of the adjustment is as follows:
Zero code internal code modification = (range internal code / calibration
weight) * (zero weight adjustment amount)
Remark: "(range internal code / calibration weight)" is the reciprocal of
the calibration factor
5, the calculation formula of the  calibration
coefficient : calibration coefficient = calibration weight / range internal code
Therefore, the calibration factor can be modified by adjusting the range

31
internal code and the calibration weight. When the calibration factor is
increased, the display weight becomes larger. The calibration factor is reduced
and the display weight is reduced.

4.7 Calibration parameters modification application


examples
Example 1 : The method of adjusting the working zero point by modifying the zero
code value.
For example: a tank is filled with materials, according to the volume
estimate80000kgCan actually show81000kgIf you continue to add1000kgMaterial,
display will also increase at the same time1000kg, indicating that the weight
is not caused by the zero point change, you can modify the zero point inner
code correction, lower the zero point1000kg. For example, the initial zero
internal code is 3045 , the calibration weight is 20000 , the range internal
code is 100000 , and the calibration coefficient
is 20000/100000=0.2.1000kgZero code internal code modification amount =
1000/0.2=5000 . According to the scale
scale work zero point to be lowered, the
zero code value should be raised. The internal code is increased by 5000 , that
is, the internal code is changed to3045+5000=8045 .
Example 2 : Method of fine-tuning the weight by the coefficient correction
method.
For example: suppose the weight on the weighing platform1000kg, controller
display998kg, the calibration coefficient needs to be increased by 100 0
÷ 998 times to be equal to 1.002 times , the original calibration weight
is 2000 , and the range internal code is10000 . If the internal code of the
range is kept unchanged, the calibration weight can be modified
to 2000*1.002=2004 .
Example 3 : Manual calibration of weight calibration parameters.
For example: after the user completes the calibration of the scale with the
weight calibration method, in order to prevent the loss of the calibration
parameters in the future, the calibration parameters can be recorded in the
calibration parameter modification input option after the weight calibration,
if the calibration is performed later. If the parameter is lost, the previously
recorded parameter can be entered in the calibration parameter modification
input option to complete the calibration parameter recovery.
4.8 Sensitivity Calibration Application Example
Example 1 : Weight-free calibration
For example: 4 scales with a sensitivity of 3.0 for a scale, the range
is100 kgThe sensor, the division value d=0.01 , the input parameters in the
sensor total range = 100 * 4 = 400 , input range =
sensitivity calibration: sensor
C0400.00 ; input sensitivity = L3.00000
Example 2 : Detect if the sensor is faulty
For example: the excitation voltage of the controller is DC5V . If the
sensor total range setting = sensitivity *5000 and the division value d=1 , the
weight value displayed by the controller is the microvolt voltage of the
sensor. The stability of the value can directly detect the quality of the
sensor.

4.9 Relay output mode description


According to the input and output control mode in Section 3.8 , specific
application examples of each control mode are given here to further understand the
application (in the following
following examples, the relay uses a normally open contact to
illustrate the problem, and the output indicates a normally open contact action to

32
close the relay. Off means the normally open contact is open )
open

4.9.1Output mode 0 (disable output) description


The 4- way relay is turned off in any state.

4.9.2Output mode 1 (real-time fixed value output)


example
Assume that the relay parameters A=500 (set value 1 ), B=2000 (set
value 2 ), C=3000 (set value 3 ), D=4000 (set value 4 ).
Display weight >=500 , relay 1 output, otherwise off.
Display weight >=2000 , relay 2 output, otherwise off.
Display weight >=3000 , relay 3 output, otherwise off.
Display weight >=4000 , relay 4 output, otherwise off.

4.9.3Output mode 2 (real-time sorting output) example


Assume that the relay parameters are A=500 (lower limit) and B=2000 (upper
limit). Relay 1 : Lower limit relay, Relay 2 : Upper limit relay, Relay 3 :
Qualified relay.
Display weight <500 , lower limit relay output, otherwise off.
Display weight > 2000 , upper limit relay output, otherwise off.
500<= display weight <=2000 , qualified relay output, otherwise closed.

4.9.4Output mode 3 (external control fixed value output)


example
When the start / stop switch (switch 1 ) is activated, the relay function is
the same as output mode 1 .
When the start / stop switch (switch 1 ) is stopped, the 4- way relay output is
always off.

4.9.5Output mode 4 (external control sorting output)


example
When the start / stop switch (switch 1 ) is activated, the relay function is
the same as output mode 2 .
When the start / stop switch (switch 1 ) is stopped, the 4- way relay output is
always off.

4.9.6Output mode 5 (with zero return control output)


example
Note : Press [Setting] to set the relay parameters directly.
Note 
Assume that the relay parameters A=500 (set value 1 ), B=2000 (set
value 2 ), C=3000 (set value 3 ), D=4000 (set value 4 ).
E=20 (zero zone weight).
Display weight >=500 , relay 1 output is valid.
Display weight >=2000 , relay 2 output is valid.
Display weight >= 3000 , relay 3 output is valid.

33
Display weight >= 4000 , relay 4 output is valid.
Display weight <20 , relay 1 , relay 2 , relay 3 , relay 4 output is off.

4.9.7Output mode 6 (external control 3 material batching


with discharge control) example
Note 
Note  : Press [Setting] to set the relay parameters directly.
Variable Comment Variable Comment
setting setting
FN5.1=6 Mode 6 , 3 ingredients with material discharge
FN5.3=1 Before feeding, =1 is set to zero, =0 is peeled, =2 is not set to zero, no
peeling
FN5.5=1 FN5.5=1 discharge mode is manual, such as FN5.5=0 discharge mode is
automatic.
C1=1.0 Start delay time is 1.0 second C2=2.0 Stable delay time
is 2.0seconds
C3=1.5 Prohibition comparison time C4=2.5 The discharge delay time
is 1.5 seconds is 2.5 seconds
C5=2.5 Jog feed time is 2.5 seconds C6=2.0 The cycle ingredient delay
is 2 seconds
C7=3 Cycle the number of ingredients 3 , start once, automatically cycle the
ingredients 3 times. Such as C7 = 99 , the infinite loop Ingredients
Relay is material 1 , relay 2 is material 2 , relay 3 is material 3 , and relay 4 is
1
discharging material.
The indicator outputs 1 , 2 , 3 , and 4 correspond to relays 1 , 2 , 3 , and 4,
respectively.
Switch 1 is the start / stop switch, switch 2 is the pause / resume switch,
switch 3 is the discharge switch, and the [Run] indicator lights when the
ingredients are activated.
A=1000 Feed 1 ingredient target value is 1000 ;
b=10 Feeding a furnish of advance 10 ; material 1 fed ( 1000-10 ) = 990 , the
material 1 relay closes.
C=2000 Material 2 ingredient target value is 2000 ;
D=20 Feed 2 Ingredients advance of 20 ; material 2 fed ( 2000-20 ) = 1980 ,
the material 2 relay closes.
E=3000 The target value of material 3 is 3000 ;
F=30 Feed 3 ingredients advance of 30 ; material 3 fed ( 3000-30 ) = 2970 ,
the material 3 relay closes.
P=0 Feed 1 ingredients allowable error of 0 ; therefore, feed 1 does not
perform jogging after the end of feed.
H=10 Feed 2 Ingredients tolerance of 10 ; material 2 after the end of the
addition, feed 2 actual addition weight of less than ( 2000-10 )
=1990 when the jog feed. The jog feed time is C5 and the jog feed
interval is C2 .
U=20 Feed 3 ingredients with tolerance 20 ; material 3 after the end of the
addition, the material 3 actual addition weight of less than ( 3000-20)
= 2980 when the jog feed. The jog feed time is C5 and the jog feed
interval is C2 .
L=50 The weight of the zero zone is 50. When the weight of the discharge
is less than 50 , the weight of the controller has returned to the zero

34
zone, and the discharge ends.
The control process is as follows: the start / stop switch is jogged once, and
the ingredients are automatically cycled 3 times according to the number of
times set by C7 . After the 1.0 second start delay, the controller performs zero
setting and enters the material 1 ingredient. When the material 1 is added to 990 ,
the material 1 relay is turned off. After a stable delay of 2.0 seconds, the
material 2 is fed, and the material 2 is added to 1980 . Material 2 relay is closed.
After 2.0seconds of stable delay, start material 3 is fed. When material 3 is added
to 2970 , material 3 relay is closed. After 2.0 seconds of stable delay, the feeding
ends, then wait for discharging, and the controller detects the discharging. When
the switch is valid, the discharge relay output is effective for discharging.
When the weight of the scale is less than 50 , the weight of the controller has
returned to zero, and then the delay is 2.5 seconds, the discharge ends, the
batching is completed, and then the delay is 2 seconds and then the cycle is
entered. Sub-ingredient, the batching cycle is automatically exited
after 3batches of ingredients are completed. (The corresponding indicator lights
when the relay output is valid)

4.9.8Output mode 7 (external control 2 fast and slow


batching control) example
Note : Press [Setting] to set the relay parameters directly.
Note 
Variable Comment Variable Comment
setting setting
FN5.1=7 Mode 7 , 2 fast and slow ingredients
FN5.3=1 Before feeding, =1 is set to zero, =0 is peeled, =2 is not set to zero, no
peeling
FN5.4=1 When FN5.4=0 fast feeding, the relay will output at the same time. For
example, FN5.4=1 will only add the relay output quickly.
C1=1.0 Start delay time is 1.0 second C2=2.0 Stable delay time
is 2.0seconds
C3=1.5 Prohibition comparison time C4=2.5 The discharge delay time
is 1.5 seconds is 2.5 seconds
C5=2.5 Jog feed time is 2.5 seconds C6=2 The cycle ingredient delay
is 2 seconds
C7=3 Cycle the number of ingredients 3 , start once, automatically cycle the
ingredients 3 times. Such as C7 = 99 , the infinite loop Ingredients
Relay 1 is material 1 fast, relay 2 is material 1 slow, relay 3 is material 2fast,
relay 4 is material 2 slow plus
The indicator outputs 1 , 2 , 3 , and 4 correspond to relays 1 , 2 , 3 , and 4,
respectively.
Switch 1 is the start / stop switch, switch 2 is the pause / resume switch, and the
[Run] indicator lights when the ingredients are activated.
A=1000 Feed 1 ingredient target value is 1000 ;
B=200 Feed 1 is quickly added with an advance amount of 200 ; when feed 1 is
fed to ( 1000-200 ) = 800 , feed 1 is quickly closed.
C=20 Feed 1 ingredients are slowly added with an advance amount of 20 ;
when feed 1 is fed to ( 1000-20 ) = 980 , feed 1 is slowly added and
closed.
D=2000 Material 2 ingredient target value is 2000 ;
E=200 Feed 2 is quickly added with an advance amount of 200 ; when feed 2 is
fed to ( 2000-200 ) = 1800 , feed 2 is quickly closed.

35
F=20 Feed 2 is slowly added with an advance amount of 20 ; when feed 2 is fed
to ( 2000-20 ) = 1980 , feed 2 is slowly added and closed.
P=0 Feed 1 ingredients allowable error of 0 ; therefore, feed 1 does not
perform jogging after the end of feed.
H=10 Feed 2 Ingredients tolerance of 10 ; material 2 after the end of the
addition, feed 2 actual addition weight of less than ( 2000-10 )
=1990 when the jog feed. The jog feed time is C5 and the jog feed
interval is C2 .
L=50 The weight of the zero zone is 50. When the weight is less than 50 ,
the weight of the controller has returned to the zero zone for the
next batching.
The control process is as follows: the start / stop switch is jogged once, and
the ingredients are automatically cycled 3 times according to the number of
times set by C7 . After the 1.0 second start delay, the controller performs zero
setting and feeds material 1 to quickly feed. When material 1 is added to 800 , it is
quickly closed, and material 1 enters slow feeding. When material 1 is fed to 980 ,
material 1 is slowly added and closed. After a stable delay of 2.0 seconds, the
starting material 2 is quickly fed. When the material 2 is added to the 1800 , the
material 2 is quickly closed, and the material 2 is fed into the slow feeding. When
the material 2 is added to the 1980 , the material 2 is slowly added and closed, and
after 2.0 seconds , the material is stabilized. After the delay, it is judged
whether the actual feeding weight of the material 2 is less than 1990 , if less
than 1990 , the material 2 is jogging, the jog feeding time is C5=2.5 seconds, the
jog feeding time is C2=2.0 seconds, and the jog feeding ends. After that, the
delay is 2 seconds to cycle into the next batching, and the batching cycle is
automatically exited after 3 batches of ingredients are completed.

4.9.9Output mode 8 (external control 1 fast and slow


batching with discharge control) example
Note : Press [Setting] to set the relay parameters directly.
Note 
 
Variable Comment Variable Comment
setting setting
FN5.1=8 8 1
Mode , material fast and slow ingredient belt discharge
FN5.3=1 Before feeding, =1 is set to zero, =0 is peeled, =2 is not set to zero, no
peeling
FN5.4=1 When FN5.4=0 fast feeding, the relay will output at the same time. For
example, FN5.4=1 will only add the relay output quickly.
C1=1.0 Start delay time is 1.0 second C2=2.0 Stable delay time
is 2.0seconds
C3=1.5 Prohibition comparison time C4=2.5 The discharge delay time
is 1.5 seconds is 2.5 seconds
C5=2.5 Jog feed time is 2.5 seconds C6=2 The cycle ingredient delay
is 2 seconds
C7=3 Cycle the number of ingredients 3 , start once, automatically cycle the
ingredients 3 times. Such as C7 = 99 , the infinite loop Ingredients
Relay is fast, relay 2 is slow, relay 3 is discharged.
1
The indicator outputs 1 , 2 , 3 , and 4 correspond to relays 1 , 2 , 3 , and 4,
respectively.
Switch 1 is the start / stop switch, switch 2 is the pause / resume switch,
switch 3 is the discharge switch, and the [Run] indicator lights when the
ingredients are activated.

36
A=1000 The target value of the ingredients is 1000 ;
B=200 The ingredients are quickly added with an advance amount of 200 ;
when the feed is ( 1000-200 ) = 800 , it is quickly closed.
C=20 The ingredients are slowly added with an advance amount of 20 ; when
the material 1 is fed to ( 1000-20 ) = 980 , it is slowly added and
closed.
P=10 The allowable error of the ingredients is 10 ; therefore , when
theactual weight of the feed is less than 990 , the jog feed is
started.
L=50 The weight of the zero zone is 50. When the weight is less than 50 ,
the weight of the controller has returned to the zero zone for the
next batching.
The control process is as follows: the start / stop switch is jogged once, and
the ingredients are automatically cycled 3 times according to the number of
times set by C7 . After the 1.0 second start delay, the controller performs zero
setting and enters the fast feeding. When it is added to 800 , it is quickly
turned off and enters the slow feeding. When feeding to 980 , it is slowly added
and closed. After a stable delay of 2.0 seconds, the actual judgment is made.
Whether the feeding weight is less than 990 , such as less than 990 , jog feeding,
jog feeding time is C5=2.5 seconds, jog feeding time is C2=2.0 seconds, after
jogging feeding, waiting for discharging, the controller detects When the
discharge switch is valid, the discharge relay output is effective for
discharging. When the weight of the scale is less than 50 , the weight of the
controller has been returned to zero, and then the delay is 2.5 seconds, the
discharge is finished, one batch is completed, and the delay is 2 seconds. Sub-
ingredient, the batching cycle is automatically exited after 3 batches of
ingredients are completed.

4.9.10Output mode 9 (fast and slow reduction batching


with material overload control) example
Note 
Note  : Press [Setting] to set the relay parameters directly.
Variable Comment
setting
FN5.1=9 Mode 9 , fast and slow reduction ingredients
FN5.4=0 FN5.4=0 fast and slow relays output at the same time, such
asFN5.4=1, only reduce the relay output.
C1=1.0 Start delay time is 1.0 second C2=2.0 Stable delay time
is 2.0seconds
C3=1.5 Prohibition comparison time C5=2.5 Jog feed time is 2.5seconds
is 1.5 seconds
C6=2.0 The cycle batching delay is 2.0 seconds.
C7=3 Cycle the number of ingredients 3 , start once, automatically cycle the
ingredients 3 times. Such as C7 = 99 , the infinite loop Ingredients
Relay 1 is fast material reduction, relay 2 is slow speed reduction, relay 3 is
short material feed, relay 4 is overload warning
The indicator outputs 1 , 2 , 3 , and 4 correspond to relays 1 , 2 , 3 , and 4,
respectively.
Switch 1 is the start / stop switch, switch 2 is the pause / resume switch, and the
[Run] indicator lights when the ingredients are activated.
A=1000 The target value of the reduced feed ingredients is 1000 ;

37
b=100 The rapid material reduction advance is 100 ; when the material
isquickly reduced to ( 1000-100 ) = 900 , the material 1 relay is turned
off.
C=20 The slow feed reduction advance amount is 20 ; when the slow feed
isreduced to ( 1000-20 ) = 980 , the material 2 relay is closed;
D=1200 The material lower limit weight, when the controller weight is less
than this value, the material is fed, and the material 3 relay is
turned on.
E=5000 The upper limit weight of the material is missing. When the
controller weight is greater than or equal to this value, the feeding
ends and the material 3 relay is closed.
F=6000 The overload weight value, the controller weight is greater than
this value, and the material 4 relay is turned on to alarm.
P=10 The allowable error of the material reduction is 10 ; when
the actual feed weight of the material 2 is less than ( 1000-10 ) =
990 , the jog reduction is performed. The jog reduction time is C5 and
the jog reduction time is C2 . If set to 0 , jog reduction is not
activated.
The control process is as follows: the start / stop switch is jogged once, and
the ingredients are automatically cycled 3 times according to the number of
times set by C7 . After the 1.0 second start delay, the controller performs rapid
material reduction. When the material is fast reduced, the relay 1 and the relay 2
are simultaneously turned on. When the material is reduced to 900 , the relay 1
isturned off to slow down, and the slow reduction is reduced to 980 . The relay 2
isturned off. After a stable delay of 2.0 seconds, the controller judges whether
the actual material reduction error is out of tolerance. For example, if the
actual weight loss is greater than 990 after the over-differential start , the
delay is 2after one reduction of the ingredients. The second cycle enters the
next reduction batching, and the batching cycle is automatically exited
after 3batches of ingredients are completed. (The corresponding indicator lights
when the relay output is valid )

4.9.11Output mode 10 (automatic weight sorting control)


example
Note 
Note  : Press [Setting] to set the relay parameters directly.
Variable Comment
setting
FN5.1=10 Mode 10 , automatic weight sorting control
Relay 1 is the lower limit output, relay 2 is the upper limit output, and
relay 3 is the qualified output.
The indicator outputs 1 , 2 , 3 , and 4 correspond to relays 1 , 2 , 3 , and 4,
respectively.
Switch 1 is the split start / stop switch, and the [Run] indicator lights when
the split check is started.
A=1000 The upper limit weight value of the sorting result is 1000.
The weightof the goods after the inspection is greater than this
value. It is considered that the current cargo is overweight and
the upper limit relay output is valid.
b =900 The lower limit weight value of the sorting result is 900.
The weightof the goods after sorting is less than this value. It is

38
considered that the current cargo weight is light and the lower
limit relay output is valid.
C=400 The threshold value of the inspection table on the cargo
is 400. When the weight detected by the controller is greater than
this value, the controller considers that the cargo has started to
go to the inspection station.
d=300 The threshold weight of the sub-testing station is 300. When the
weight detected by the controller is less than this value, the
controller considers that the goods have begun to leave the
inspection station.Note: C> d must be met when setting
parameters ! ! !
E=1.00000 The weight correction coefficient of the sorting is 1.00000 , and
the coefficient defaults to 1.0000 . When the weight of the
controller is different from the actual one, the weight is adjusted
by modifying the sorting weight coefficient.
C1=0.2 C1 is the time of the inspection station on the goods. When the
weight detected by the controller is greater than the C value, the
controller delays the C1=0.2 second time and considers that the
goods are completely on the branching table. This value is based on
the actual time of the inspection station on the goods. Make
settings.
C2=0.5 C2 is the time when the goods stay at the inspection station, and
the effective weight of the inspection is collected within the
time ofC2=0.5 seconds.
C3=0.2 C3 is the sub-test time of the goods. When the weight detected by
the controller is less than the value of d , the controller delays
theC3=0.2 seconds and considers that the goods completely leave the
branch. This value is based on the actual time of the sub-test.
Settings.
C4=1.0 C4=1.0 seconds is the sorting result holding time, which is the
controller's sorting weight display time and also the relay output
result holding time.
C7=90 C7 is the effective data ratio of the controller sampling during the
retention time of the inspection station. The setting range is 0-
99% , and the general setting is about 90% .
The control process is as follows: When the detection is stopped, the [Run]
indicator is off, and the start / stop switch is activated once, and the
controller enters the automatic sorting operation state. At this time, the
[Run] indicator lights. The controller first checks whether the weight is
greater than the weight of the upper inspection table. When the weight is
greater than 400 , the controller considers that the goods start on the
inspection station, and then delays C1=0.5seconds , and then considers that
the goods are completely on the inspection table, and then In C2=0.5 seconds,
the controller starts to collect the weight data of the processed goods. When
the weight of the controller is less than 300 , the controller thinks that the
goods will completely leave the branching station within C3=0.2seconds. At
this time, the controller displays the actual The weight is divided and the
relay outputs the actual inspection result. After the C4=1.0 second time, the
current cargo sorting process ends and the next cargo inspection is performed.

4.9.12Output mode 11 (external control trigger weight


sorting control) example

39
Note 
Note  : Press [Setting] to set the relay parameters directly.
Variable Comment
setting
FN5.1=11 Mode 11 , external control weight sorting control
Relay is the lower limit output, relay 2 is the upper limit output, and
1
relay 3 is the qualified output.
The indicator outputs 1 , 2 , 3 , and 4 correspond to relays 1 , 2 , 3 , and 4,
respectively.
The switch 1 is the split start / stop switch, and the switch 2 is the external
control trigger switch, which triggers a cargo sorting once, and the [Run]
indicator lights when the split detection starts.
A=1000 The upper limit weight value of the sorting result is 1000.
The weightof the goods after the inspection is greater than this
value. It is considered that the current cargo is overweight and
the upper limit relay output is valid.
b =900 The lower limit weight value of the sorting result is 900.
The weightof the goods after sorting is less than this value. It is
considered that the current cargo weight is light and the lower
limit relay output is valid.
C=400 The threshold value of the inspection table on the cargo
is 400. When the weight detected by the controller is greater than
this value, the controller considers that the cargo has started to
go to the inspection station.
d=300 The threshold weight of the sub-testing station is 300. When the
weight detected by the controller is less than this value, the
controller considers that the goods have begun to leave the
inspection station.Note: C> d must be met when setting
parameters ! ! !
E=1.00000 The weight correction coefficient of the sorting is 1.00000 , and
the coefficient defaults to 1.0000 . When the weight of the
controller is different from the actual one, the weight is adjusted
by modifying the sorting weight coefficient.
C1=0.2 C1 is the time of the inspection station on the goods. When the
controller receives the trigger signal, the controller delays
theC1=0.2 second time and considers that the goods are completely on
the distribution platform. This value is set according to the
actual time of the inspection station on the goods.
C2=0.5 C2 is the time that the goods stay on the sorting table, and the
effective weight of the sorting is collected during the period
ofC2=0.5 seconds.
C3=0.2 C3 is the sub-test time of the goods. When the weight detected by
the controller is less than the value of d , the controller delays
theC3=0.2 seconds and considers that the goods completely leave the
branch. This value is based on the actual time of the sub-test.
Settings.
C4=1.0 C4=1.0 seconds is the sorting result holding time, which is the
controller's sorting weight display time and also the relay output
result holding time.
C7=90 C7 is the effective data ratio of the controller sampling during the
retention time of the inspection station. The setting range is 0-
99% , and the general setting is about 90% .
The control process is as follows: When the detection is stopped, the [Run]
indicator is off, and the start / stop switch is activated once, and the
controller enters the sorting operation state. At this time, the [Run]

40
indicator lights. The controller first waits for the detection trigger signal
to be valid. When the trigger signal is valid, the controller considers that
the goods start on the inspection station, and then delays C1=0.2 seconds, and
considers that the goods are completely on the inspection station, and then
at C2= Within 0.5 seconds, the controller starts to collect the weight data of
the processed goods. When the weight detected by the controller is less
than 300 , the controller thinks that the goods will completely leave the
branching station within C3=0.2 seconds. At this time, the controller displays
the actual The weight is divided and the relay outputs the actual inspection
result. After the C4=1.0 second time, the current cargo sorting process ends and
the next cargo inspection is performed.

4.9.13Output mode 12 (external control instantaneous


peak) example
Note : Press [Setting] to set the relay parameters directly.
Note 
Variable Comment
setting
FN5.1=12 Mode 12 , external control instantaneous peak
Relay 1 is the peak lower limit output, relay 2 is the peak upper limit output,
and relay 3 is the peak qualified output.
The indicator outputs 1 , 2 , 3 , and 4 correspond to relays 1 , 2 , 3 , and 4,
respectively.
Switch 1 is the peak start / stop switch, and the [Run] indicator lights when
the peak detection is activated.
A=1000 The upper limit of the peak result is 1000 , and the actual peak
value is greater than this value. It is considered that the current
peak exceeds the upper limit and the upper limit relay output is
valid.
b=900 The peak result has a lower limit of 900. The actual peak value is
less than this value. It is considered that the current peak value
is less than the lower limit and the lower limit relay output is
valid. The actual peak value is between 900 and 1000 , and the
qualified relay output is valid.
c=500 The peak start capture threshold is 500. After the peak capture is
started, after the controller displays the value greater than 500 ,
the controller starts to pick and save the peak value.
d=300 The peak end capture threshold is 300. After the effective peak is
captured, the controller displays the value less than 300 and the
controller ends the peak capture. Note that when setting
parameters, you need to: c>d .
C1=2.0 The peak hold time is 2 seconds. The C1 value setting is not 0 , the
peak is automatically released after C1 seconds, and then
automatically enters the next peak detection.
When the C1 value is set to 0 , you need to press the start / stop
switch or [Exit] key to cancel the peak hold. If you need to enter
the peak detection again, you need to press the start / stop switch
or [Exit] button again.
The control process is as follows: press the start / stop switch or [Exit]
key to start the peak detection. At this time, the [Run] indicator lights up.
After the controller displays the value greater than 500 , the controller starts
to pick the peak value, and the controller displays the value less than 300 and
ends the peak value. Capture, display window for peak hold for 2 seconds, relay

41
output for 2seconds, after 2 seconds, once peak detection ends, automatically
start the next peak detection.

4.9.14Output mode 13 (input and output self-test)


description
In this mode, . 4 switches and amount . 4 relays corresponding to the corresponding
relay output corresponding relay output is closed off when the switch active,
inactive, switch time.

4.9.15 Output Mode 14 (with time hold interval fixed


value output) example
Note : Press [Setting] to set the relay parameters directly.
Note 
Assume that the relay parameters A=500 (set value 1 ), B=2000 (set
value 2 ), C=3000 (set value 3 ), D=4000 (set value 4 ).
L = 20 (zero zone weight).
Display weight 20 < display weight <=500 , relay 1 output is valid , hold time
is C3 .
Display weight 500 < display weight <=2000 , relay 2 output is valid , hold time
is C3 .
Display weight 2000 < display weight <=3000 , relay 3 output is valid , hold time
is C3 .
Display weight 3000 < display weight <=4000 , relay 4 output is valid , hold time
is C3 .
4.10 Information Tips
Controller prompt Prompt message description
message
Err 01 Data conversion error
Err 06 wrong password
AdcErr ADC chip failure
ON-OFF The controller is powered on
again.
OUER Weight overload
FFFF Ingredient discharge prompt
ESC Cancel setting

42
5 Maintenance and precautions
 order to ensure the clarity and service life of the controller, the
controller should not be used under direct sunlight, and the place should be
relatively flat.
 should not be used in places with dust and vibration, avoid using it in
humid environment !
 sensor and controller must be connected reliably. The system should have
good grounding, away from strong electric field and strong magnetic field. The
sensor and controller should be away from strong corrosive objects and away
from inflammable and explosive materials ! Do not use flammable gas or
flammable steam. For use on occasions, do not use the canned system in a
pressure vessel !
 In areas where lightning strikes frequently occur, reliable lightning arresters
must be installed to ensure the personal safety of operators and prevent
lightning damage to the controller and corresponding equipment !
 Sensors and controllers are all static-sensitive devices. Anti-static measures
must be taken during use. It is strictly forbidden to perform electric welding
operations or other strong electric field operations on the weighing platform.
During the thunderstorm season, reliable lightning protection measures must be
implemented to prevent lightning strikes. Damage to the sensor and controller
to ensure the operator's personal safety and the safe operation of the
weighing equipment and related equipment !
 Do not use strong solvents (such as benzene or nitro oil) to clean the case !
 not inject liquid or other conductive particles into the controller to
prevent damage and electric shock to the controller!
 Before plugging and unplugging the controller and external device cables, you
must cut off the power of the controller and corresponding equipment! The
controller must be powered off before plugging in or connecting the sensor
cable! The controller and the large screen power must be turned off before
plugging in the large screen cable! Before plugging and unplugging the
communication cable, you must cut off the power of the controller and the host
computer!
 external interface of the controller must be used strictly according to
the method marked in the instruction manual. The connection must not be
changed without authorization. If the controller fails during the use,
immediately unplug the power supply and send it to the professional factory
for maintenance. The manufacturer should not repair it by itself to avoid
causing more damage. The controller is not allowed to open at will, otherwise
it will not be guaranteed.
 This controller is within one year from the date of sale. Under normal
conditions of use, non-human failure is covered by the warranty.
 company advises customers: The controller should be tested and accepted
before using the controller of the company. The company is only responsible
for the quality of the controller itself and is not responsible for the system
problems in which the controller is located.

43

Вам также может понравиться