Вы находитесь на странице: 1из 5

Available online at www.sciencedirect.

com

Forensic Science International: Genetics 2 (2008) 226–230


www.elsevier.com/locate/fsig

Comparison of five DNA quantification methods


Karsten Nielsen a,*, Helle Smidt Mogensen a, Johannes Hedman b,
Harald Niederstätter c, Walther Parson c, Niels Morling a
a
Section of Forensic Genetics, Department of Forensic Medicine, Faculty of Health Sciences, University of Copenhagen, Denmark
b
Swedish National Laboratory of Forensic Science, Linköping, Sweden
c
Institute of Legal Medicine, Innsbruck Medical University, Innsbruck, Austria
Received 24 December 2007; received in revised form 11 February 2008; accepted 26 February 2008

Abstract
Six commercial preparations of human genomic DNA were quantified using five quantification methods: UV spectrometry, SYBR-Green dye
staining, slot blot hybridization with the probe D17Z1, QuantifilerTM Human DNA Quantification kit and RB1 rt-PCR. All methods measured
higher DNA concentrations than expected based on the information by the manufacturers. UV spectrometry, SYBR-Green dye staining, slot blot
and RB1 rt-PCR gave 39, 27, 11 and 12%, respectively, higher concentrations than expected based on the manufacturers’ information. The DNA
preparations were quantified using the QuantifilerTM Human DNA Quantification kit in two experiments. The measured DNA concentrations with
Quantifiler were 125 and 160% higher than expected based on the manufacturers’ information. When the QuantifilerTM human DNA standard (Raji
cell line) was replaced by the commercial human DNA preparation G147A (Promega) to generate the DNA standard curve in the QuantifilerTM
Human DNA Quantification kit, the DNA quantification results of the human DNA preparations were 31% higher than expected based on the
manufacturers’ information. The results indicate a calibration problem with the QuantifilerTM human DNA standard for its use with the
QuantifilerTM Human DNA Quantification kit. The possible reasons for the problem are discussed and a solution is suggested. The results
emphasise the need for standard reference DNA material and standard methods for DNA quantification.
# 2008 Published by Elsevier Ireland Ltd.

Keywords: Human DNA; Quantification; Quantifiler; Slot blot; SYBR-Green; UV spectrometry

1. Introduction PCR quantification method of nuclear and mitochondrial DNA


based on the TaqMan assay validated for forensic samples was
Reliable quantification of human DNA in a forensic genetic reported by Andréasson et al. [5]. The nuclear target was the
sample makes it possible to adjust the concentration of template human retinoblastoma susceptibility gene (RB1) which is a
DNA used for STR analysis in order to obtain optimal PCR single copy gene located on chromosome 13. Based on this assay,
reactions and STR typing results with as small amounts of DNA Niederstätter et al. [6] developed a modular real-time (rt) PCR
as possible. Various methods have been developed for system for absolute quantification of human nuclear (n) and/or
quantification of human DNA in forensic samples, e.g. a slot mitochondrial (mt) DNA with an optional internal amplification
blot method using a 40 bp long DNA-probe designed to match positive control. Others have developed real-time PCR assays
the human specific alpha satellites that are present in 500–1000 based on multi copy number DNA targets like the Alu sequences
copies on chromosome 17 (D17Z1) [1,2]. The AluQuant Human [7,8]. The advantage of using a multi copy number target DNA is
Quantification System [3] includes an enzyme linked DNA probe the potential of developing more sensitive assays, but the
that is complementary to human specific Alu sequences, which variation of the DNA quantification results from person to person
are repetitive elements with a consensus sequence of approxi- may be higher due to variations in the numbers of copies of the
mately 280 bp [4]. Alu sequences are found in 500,000– DNA targets. Applied Biosystems (Foster City, CA) introduced
1,000,000 copies per human genome. A human specific real-time the QuantifilerTM Human DNA Quantification kit for forensic
samples based on the single copy gene hTERT (human
telomerase reverse transcriptase) that is located on chromosome
* Corresponding author. Tel.: +45 3532 6110; fax: +45 3532 6120. 5 [9]. This assay has an internal positive control to ascertain PCR
E-mail address: karsten.nielsen@forensic.ku.dk (K. Nielsen). amplification.

1872-4973/$ – see front matter # 2008 Published by Elsevier Ireland Ltd.


doi:10.1016/j.fsigen.2008.02.008
K. Nielsen et al. / Forensic Science International: Genetics 2 (2008) 226–230 227

The DNA quantification methods mentioned have in Department of Forensic Medicine, Faculty of Health Sciences,
common that that they are indirect measures of the DNA University of Copenhagen, Denmark.
concentration in a sample. A standard curve based on various
concentrations of known concentrations of reference DNA is 2. Materials and methods
needed for the interpretation of test results. Standardized, high
quality reference DNA is needed to ensure reproducible DNA 2.1. UV spectrometry and SYBR-Green fluorometry
quantification results within and between laboratories. UV
spectrometry directly detects the DNA and RNA in a sample. UV spectrometry was performed in a PerkinElmer UV/VIS
The absorbance at 260 nm of 1 in a 1 cm path-length cuvette Lambda-16 spectrometer (PerkinElmer) with double beam. The
approximately equals a DNA concentration of 50 ng/ml for blank solution was distilled water.
double stranded DNA (dsDNA) and 40 ng/ml for single SYBR-Green fluometry was performed in a Lightcycler
stranded DNA (ssDNA) and RNA [10]. UV spectrometry, (Roche) with excitation: 494 nm and emission: 521 nm. A
however, is not suitable for quantification of DNA in forensic standard curve was made from known amounts of calf thymus
samples because UV spectrometry is not sensitive enough to DNA (Amersham Pharmacia Biotech).
detect the low amounts of DNA in many crime case stains. The Six commercial human DNA preparations (Table 1) were
DNA concentration must be at least 3 ng/ml in order to give subjected to agarose gel electrophoresis (1.2% (V/cm) 2 h) to
reliable results with UV spectrometry. In many forensic stain test for the presence of RNA and to test if the DNA preparations
samples, the DNA concentration is below 1 ng/ml. Forensic were degraded. The DNA preparations were diluted to 3, 5 and
stain DNA samples may be contaminated with UV-absorbing 10 ng/ml before quantification with UV spectrometry and dye
contaminants like proteins and phenol, which may interfere staining with SYBR-Green (Table 2). Samples were scanned
with the quantification results. Furthermore, UV spectrometry from 200 to 350 nm using UV spectrometry to reveal UV
measures all human and non-human DNA as well as RNA in a absorbing contaminants. The DNA concentration was calcu-
sample, which reduces the predictive value for successful lated as: DNA concentration = absorbance at 260 nm  dilu-
DNA-profiling of human STRs. dilution factor  50 ng/ml.
A number of fluorescent dyes interact with DNA. The dye,
e.g. may on binding to DNA become fluorescent or its 2.2. Comparison of slot blot and QuantifilerTM H
emission spectrum can change from that in its unbound state.
Dyes can have different binding affinities to dsDNA, ssDNA The DNA preparations were diluted from 50 to 0.023 ng/ml
and RNA. PicoGreen, Hoechst dyes and SYBR-Green before quantification with slot blot hybridization using the D17Z1
preferentially detects dsDNA whereas OliGreen preferen- DNA probe (Table 2) and QuantifilerTM Human DNA
tially detects ssDNA [11,12]. DNA quantification based on Quantification kit (QuantifilerTM H) following the manufacturer’s
fluorescent dyes like SYBR-Green is very sensitive. instructions (Table 2). K562 DNA (Promega) was used as
However, fluorometric methods are not well suited for reference DNA for the slot blot method. At the Section of Forensic
DNA quantification if the origin of the DNA is uncertain, Genetics in Copenhagen, four replicates of each of the DNA
because all kinds of DNA from animals, plants, bacterias, preparations were quantified with QuantifilerTM H using both the
etc. are detected in a sample. For a more detailed review of QuantifilerTM DNA and G147A as reference DNA. Some of the
methods used for quantification of DNA in forensic genetics, DNA preparations were quantified in the laboratories in Innsbruck
see Nicklas and Buel [12]. and Linköping. The DNA was diluted to 5 or 10 ng/ml before
Forensic laboratories certified under the EN/ISO/IEC 17025 shipping in PCR clean Safe-Lock 1.5 ml Eppendorf tubes by
standard [13] shall ensure that: ‘‘purchased manufacturers and ordinary mail. Three or four replicates of each DNA preparation
reagents and consumable materials that affect the quality of was quantified using RB1 rt-PCR with QuantifilerTM DNA as
tests and/or calibrations are not used until they have been reference DNA at the Innsbruck Institute of Legal Medicine
inspected or otherwise verified as complying with standard (GMI, Table 2). The DNA preparations G147A and G152A were
specifications or requirements defined in the methods for the additionally quantified with RB1 rt-PCR using the GMI DNA
tests and/or calibrations concerned’’. The results presented here standard [6]. Two replicates of each D3160, D3035 and HG DNA
are part of the validation of the QuantifilerTM Human DNA were quantified at the Swedish National Laboratory of Forensic
Quantification kit at the Section of Forensic Genetics, Science, Linköping, using the QuantifilerTM H kit (Table 2).

Table 1
Commercial DNA preparations used in this study
DNA preparation Manufacturer Source Source Sex DNA concentration (ng/ml)
D3160 Sigma Placenta Single Male 100
D3035 Sigma Placenta Single Female 140
Human genomic DNA Roche Blood (buffy-coat) Multiple Mix 200
G147A Promega Blood from multiple anonymous donors Multiple Male 177
G152A Promega Blood from multiple anonymous donors Multiple Female 109
QuantifilerTM DNA ABI Raji cell line Single Male 200
228 K. Nielsen et al. / Forensic Science International: Genetics 2 (2008) 226–230

Table 2
DNA quantification methods used in this study
Quantification method Instrument Principle Manufacturer References
a
UV spectrometry PerkinElmer Lambda 16 OD-260 – [10]
Dye staining Lightcycler (Roche) SYBR-Green I Bie & Berntsen [11]
Slot blot hybridization – D17Z1 DNA probe DNA-technology [1,2]
TaqMan real-time PCR ABI 7000 SDS, ABI 7300 SDS QuantifilerTM Human ABI [9]
TaqMan real-time PCR ABI 7700 SDS RB1 rt-PCR ABI [5,6]
a
Samples were scanned from 200 to 350 nm.

Table 3
UV spectrometry and SYBR-Green measurements
DNA preparation DNA stock (ng/ml)a OD-260 (ng/ml) SYBR-Green (ng/ml)
b
D3160 100 182  10 137  45
D3035 140 181  15 161  50
Human genomic DNA 200 446  18 300c
G147A 177 170  3 212
G152A 109 107  32 163
QuantifilerTM DNA 200 206  10 182  2
a
Manufacturers’ information.
b
Mean  Standard deviation.
c
Average of two measurements.

The data were log transformed and analyzed with ANOVA degraded DNA, because single stranded DNA absorbs 20–30%
using the free software package R (http://www.r-project.org). more UV light at 260 nm than double stranded DNA does. In
Because of incomplete experimental design, the ANOVA was these three samples, contaminants that absorb UV light in the
performed as three pair wise comparisons of methods: (1) range of 200–230 nm with a maximum of 6 units at
QuantifilerTM H (Copenhagen) vs. QuantifilerTM H (G147A, approximately 210 nm were observed (results not shown). The
Copenhagen); (2) RB1 rt-PCR vs. QuantifilerTM H (G147A, presence of UVabsorbing contaminants may to some extent have
Copenhagen); and (3) QuantifilerTM H (Copenhagen) vs. interfered with the measurements at 260 nm. UV-measurements
QuantifilerTM H (Linköping). Model: measured DNA = - of DNA concentrations should not be relied upon unless one is
method + DNA preparation + method  DNA preparation + - extremely confident of the quality and condition of the DNA. A
residual. Quantification results of QuantifilerTM DNA were not UV-measurement can be misleading unless the DNA is pure and
included in the ANOVA. dissolved in a ‘UV-clean’ buffer solution. In general, DNA
quantification with SYBR-Green resulted in measurements of
2.3. Test of QuantifilerTM H batch variation 15–50% higher DNA concentrations than expected based on the
declarations of the manufacturers, except for QuantifilerTM DNA
Two different batches of the QuantifilerTM H kit (L/N: that was measured 9% lower than expected.
0404011 and 0411017) were tested using a 5 ng/ml dilution of Fig. 2 shows a log–log scatter plot of the measured DNA
each of the six DNA preparations shown in Table 1. Each DNA concentrations obtained with the slot blot and the QuantifilerTM
preparation was measured twice. H (Raji cell reference DNA) quantification measurements of all
the DNA preparations (ordinate) compared to the expected
3. Results and discussion

The six DNA preparations mainly contained DNA of high


molecular weight although D3160, D3035 and Human genomic
DNA (HG DNA) also contained a minor component of partly
degraded DNA (Fig. 1). There was no sign of RNA. The DNA
concentration measured by UV spectrometry of the DNA
preparations G147A, G152A and QuantifilerTM DNA were all
very similar (4%) and compared well to the expected DNA
concentrations declared by the manufacturers. The DNA
concentration of the DNA preparations D3160, D3035 and
HG DNA were 82, 29 and 123%, respectively, higher than the
expected DNA concentration (Table 3). The higher DNA Fig. 1. Agarose gel electrophoresis of DNA preparations. M1: ladder; 1:
concentration measurements for the samples D3160, D3035 and D3160; 2: 3035; 3: human genomic DNA; 4: G147A; 5: G152A; 6: quantifiler
HG DNA may to some extent be explained by the presence of DNA.
K. Nielsen et al. / Forensic Science International: Genetics 2 (2008) 226–230 229

of G152A using RB1 rt-PCR were 56% lower than declared by


the manufacturers, and the results were significantly different
from those obtained with QuantifilerTM H with G147A as
reference. The reason for these results is not known. RB1 rt-
PCR quantifications of G147A and G152A were also performed
using a different DNA standard (GMI DNA [6]) and no
significant difference (t-test, p > 0.05) was observed between
the DNA concentrations measured using QuantifilerTM DNA
and GMI DNA as reference DNA (data not shown).
Gel electrophoresis and UV spectrometry (scanning from
200 to 350 nm) of the QuantifilerTM DNA showed that the DNA
was mainly of high molecular weight and that it contained
neither UV absorbing agents nor RNA in detectable amounts.
UV spectrometry (absorbance 260 nm), SYBR-Green fluoro-
Fig. 2. Log–log scatter plot of DNA measurements using slot blot and metry, slot blot and RB1 rt-PCR measurements of DNA
QuantifilerTM H. Regression lines were forced through (0,0) since all negative confirmed that the DNA concentration was 200 ng/ml DNA as
DNA controls gave negative results for both methods. Expected DNA con- declared by the manufacturer. In average, QuantifilerTM H
centration: as informed by the manufacturers.
measured 100% higher DNA concentrations than QuantifilerTM
H with G147A as reference DNA demonstrating a calibration
DNA concentration according to the declarations of the problem when the QuantifilerTM DNA standard is used together
manufacturers (abscissa). Both regression lines go through with the QuantifilerTM H kit. The hTERT gene encodes the
(0,0) since the negative controls for both assays always gave catalytic component of the human telomerase that is inactivated
negative results. The slopes of the regression lines represent the in most normal cells, but is activated in 85% of all tested
ratios between the measured and expected DNA concentra- immortalized cells [14]. This indicates that genetic differences
tions. In general, QuantifilerTM H measured the DNA in the hTERT gene between wild type DNA and cell line DNA
concentration 2.25 times higher than expected, while slot blot exist. Mutations in the primer binding sites of one or both
in general measured 1.11 times higher than expected. The hTERT alleles in the QuantifilerTM DNA standard might lower
regression coefficients of the QuantifilerTM H kit (R2 = 0.9486) the single-molecule PCR efficiency for affected gDNA target
and the slot blot method (R2 = 0.7236) demonstrate that molecules in early cycles, and result in the worst case in a total
QuantifilerTM H has less unexplained variation than the slot blot loss of PCR product derived from the affected allele(s).
method, i.e. QuantifilerTM H is more reproducible but less Sequence variation within the probe binding site could result in
accurate than slot blot. As supported by the supplementary a lowered number of fluorochrome molecules liberated per
investigations presented below, the most likely explanation for PCR cycle causing a reduced ability (or even complete failure)
the inaccurate results obtained with QuantifilerTM H is the use to detect the affected PCR products. Both possibilities would
of the DNA from the Raji cell line as reference DNA.
In order to explore the unexpected measurements of DNA
concentrations with QuantifilerTM H, a collaborative exercise
was performed with the forensic genetic laboratories in
Innsbruck and Linköping. Fig. 3 shows the measured DNA
concentrations of the DNA preparations D3160, D3035, HG
DNA, G147A, G152A and QH DNA analyzed in three
laboratories using (1) QuantifilerTM H (Copenhagen and
Linköping), (2) QuantifilerTM H with G147A as reference
DNA (Copenhagen) and (3) RB1 rt-PCR with QuantifilerTM
DNA as reference DNA (Innsbruck). The method QuantifilerTM
H (Copenhagen) in average measured 100% higher than
QuantifilerTM H (G147A, Copenhagen). The result of the
ANOVA showed no interaction ( p = 0.5906) but highly
significant effects of method ( p < 0.001) and of DNA
preparation ( p < 0.001). In average, QuantifilerTM H (Linköp- Fig. 3. DNA quantification of samples with three methods in three forensic
ing) measured 19% higher DNA concentrations than Quanti- genetic laboratories. The bars represent the means of two to four replicates and
filerTM H (Copenhagen), and the result of the ANOVA showed the error bars represent the standard deviations. Quantifiler H (G147A): DNA
no interaction ( p = 0.6937) but significant effects of method preparation G147A (Promega) was used as reference DNA. Using ANOVA
(methods: Quantifiler H and Quantifiler H (G147A); DNA preparations: D3160,
( p = 0.003) and of DNA preparation ( p = 0.002). When DNA
D3035, HG DNA, G147A and G152A) no interaction was observed
preparation G152A was excluded from the analysis, significant ( p = 0.5906). Highly significant effects of both method ( p < 0.001) and
interaction was observed ( p = 0.003), but no significant effect DNA preparation ( p < 0.001) were observed. In average, Quantifiler H mea-
of method ( p = 0.27) was observed. The quantification results sured 100% higher DNA concentrations than Quantifiler H (G147A).
230 K. Nielsen et al. / Forensic Science International: Genetics 2 (2008) 226–230

lead to an overestimation of the DNA content in the DNA [2] J.S. Waye, H.F. Willard, Structure, organization, and sequence of alpha
satellite DNA from human chromosome 17: evidence for evolution by
preparations tested.
unequal crossing-over and an ancestral pentamer repeat sharing with the
In PCR assays for DNA quantification where genetic human X chromosome, Mol. Cell Biol. 6 (1986) 3156–3165.
differences in DNA template between wild type DNA and cell [3] M.N. Mandrekar, A.M. Erickson, K. Kopp, B.E. Krenke, P.V. Mandrekar,
line DNA can be expected, like hTERT, the use of cell line DNA R. Nelson, K. Peterson, J. Shultz, A. Tereba, N. Westphal, Development of
as reference DNA is not an optimal solution. However, the a human DNA quantification system, Croat. Med. J. 442 (2001) 336–339.
QuantifilerTM cell line DNA standard works well with the RB1 [4] A.J. Mighell, A.F. Markham, P.A. Robinson, Alu sequences, FEBS Lett.
417 (1997) 1–5.
rt-PCR kit, and the QuantifilerTM H kit works well with the [5] H. Andréasson, U. Gyllensten, M. Allen, Real-time DNA quantification of
DNA preparation G147A (DNA extracted from blood samples nuclear and mitochondrial DNA in forensic samples, Biotechniques 33
from 80–100 males). This demonstrates that cell line DNA can (2002) 402–411.
be used as reference DNA and parts of potential oncogenes can [6] H. Niederstätter, S. Köchl, P. Grubwieser, M. Pavlic, M. Steinlechner, W.
Parson, A modular real-time PCR concept for determining the quantity
be used as DNA templates for quantification, but the
and quality of human nuclear and mitochondrial DNA, Forensic Sci. Int.:
combination of cell line DNA and oncogene based DNA Genet. 1 (2007) 29–34.
quantification assays may create difficulties. [7] M.E. Sifis, K. Both, Burgoyne LA, A more sensitive method for quanti-
The method on which the QuantifilerTM H kit is based works fication of genomic DNA by Alu amplification, J. Forensic Sci. 47 (2002)
well. However, the measurements are inaccurate – although 589–592.
reproducible – due to the fact that Raji cell DNA is used as [8] J.A. Nicklas, E. Buel, Development of an Alu-based, real-time PCR
method for quantification of human DNA in forensic samples, J. Forensic
reference DNA. Many forensic laboratories still use the kit at Sci. 48 (2003) 936–944.
AB 7000 instruments. Therefore, it would be an improvement if [9] Quantifiler kits user manual, Applied Biosystems, Foster City, US, 2003.
the Raji cell DNA were substituted with better suited DNA for [10] J. Sambrook, D.W. Russell, Molecular Cloning: A Laboratory Manual,
the purpose. 3rd ed., Cold Spring Harbor Laboratory Press, Cold Spring Habor, NY,
The results also demonstrate the need for internationally USA, 2001.
[11] K. Rengarajan, S.M. Cristol, M. Mehta, J.M. Nickerson, Quantifying DNA
agreed standard reference DNA materials and improvements of concentrations using fluorometry: a comparison of fluorophores, Mol. Vis.
the methods for DNA quantification [15,16]. 8 (2002) 416–421.
[12] J.A. Nicklas, E. Buel, Quantification of DNA in forensic samples, Anal.
Acknowledgment Bioanal. Chem. 376 (2003) 1160–1167.
[13] General requirements for the competence of testing and calibrating
laboratories (ISO/IEC 17025:2005), CEN Management Centre, rue de
We thank Poul Svante Eriksen, M.Sc., Department of Strassart 36, B-1050 Brussels, Belgium.
Mathematical Sciences, University of Aalborg, Denmark, for [14] A. Zhang, et al., Frequent amplification of the telomerase reverse tran-
help with the ANOVA analysis. scriptase gene in human tumors, Cancer Res. 60 (2000) 6230–6235.
[15] M.C. Kline, D.L. Duewer, J.W. Redman, J.M. Butler, Results from the
NIST 2004 quantification study, J. Forensic Sci. 50 (2005) 570–578.
References [16] P.M. Vallone, M.C. Kline, D.L. Duewer, A.E. Decker, J.W. Redman, J.C.
Travis, M.V. Smith, J.M. Butler, Development and usage of a NIST
[1] P.S. Walsh, J. Varlaro, R. Reynolds, A rapid chemiluminiscent method for Standard Reference Materials for real-time PCR quantitation of human
quantification of human DNA, Nucleic Acids Res. 20 (1992) 5061–5065. DNA, Forensic Sci. Int.: Genet., in press.

Вам также может понравиться