MS21neo V

Medical Scale
USER MANUAL

Please keep the instruction manual at hand all the time for future reference.
TABLE OF CONTENTS

PREFACE......................................................................................................2
GENERAL INFORMATION........................................................................2
SAFETY INSTRUCTION............................................................................2
ENVIROMENTAL........................................................................................3
CLEANING...................................................................................................3
MAINTENANCE..........................................................................................3
WEIGHING OPERATION............................................................................4
WARRANTY-LIABILITY.............................................................................4
DISPOSING OF THE SCALE......................................................................5
EXPLANATION OF THE GRAPHIC SYMBOLS.......................................6
EMC GUIDANCE AND MANUFACTURER’S DECLARATION.............8
SPECIFICATION........................................................................................13
Medical Scale Models Approval List..........................................................14
POWER ADAPTOR STANDARDS...........................................................15
PANEL.........................................................................................................17
KEY FUNCTION........................................................................................18
LCD SYMBOL DESCRIPTION................................................................19
TIME SETTING..........................................................................................20
USING SCALE...........................................................................................20
USING HOLD FUNCTION.......................................................................21
USING BMI FEATURE..............................................................................22
USING TARE FUNCTION.........................................................................23
STORING AND RE-CALLING PRE-TARE WEIGHT.............................23
SET UP YOUR DEVICE............................................................................26
PRINTING FUNCTION.............................................................................27
INSTRUCTION FOR CHARGING AND CONNECTING.......................29
SCALE ASSEMBLING..............................................................................30
INSTRUCTION FOR REPLACING BATTERIES....................................34
ERROR MESSAGE....................................................................................37
TROUBLESHOOTING..............................................................................38

1
PREFACE
Manufacturer’s Declaration of Conformity
Thank you for choosing CHARDER MEDICAL product. All features
This product has been manufactured in accordance with
of this product were designed to state of the art and are optimized
the harmonized European standards, following the
for simple and straightforward use. If you have any queries or
provisions of the below stated directives:
experience any problems not addressed in the operating instructions,
93/42/EEC as amended by please contact your CHARDER MEDICAL service partner, or visit us
2007/47/EC on the Internet at www.chardermedical.com
Medical Device Directive

We strongly recommend you use the scales on flat and hard surface.
2014/31/EU Any soft surface, like carpet will cause inaccuracy.
Non-automatic Weighing Instruments
X
Directive
important instructions for installation, proper use and maintenance of
Please see separate document showing on sticker of device
the device.
for above CE marking.
Authorized EU Representative:
The manufacturer shall not be liable for damages arising out of
failure to heed the following instructions:

• When using electrical components under increased safety


requirements, always comply with the appropriate regulations.
Manufactured by: • Improper installation will render the warranty null and void.
Charder Electronic Co., Ltd. • Ensure the voltage marked on the power supply unit matches
No.103, Guozhong Rd., Dali Dist., your mains power supply.
Taichung City 412 ,Taiwan (R.O.C.) • This device is designed for use indoors.
• Observe the permissible ambient temperatures for use
FDA no.: D072479 • The device meets the requirements for electromagnetic
compatibility. Do not exceed the maximum values specified in the
applicable standards.
• Expected Service Life: 5 years
• These batteries should be kept away from small children. If

IN-1223 [12184D] 04/2016

-1- - 40 -
swallowed, promptly seek medical assistance. • Key buttons failure - Front key panel damaged or disconnected

If you have any problem, contact your local CHARDER 3. Buzzer malfunction
MEDICAL service partner. • Wrong welding of PVC wire
• Key buttons & control panel damaged or disconnected.

• A
l
l

b
a
t
t
e
r
i
e
s

c
o
n
t
a
i
n

t
o
x
i e
c r
i
c e
o s
m
p s
o h
u o
n u
d l
s d
;
b
d e
i
s d
p e
o l
s e
a g
l a
t
o e
f d

b t
a o
t
t a
• Please do not incinerate batteries.
c • The optimum operating temperature for the scale is 5℃ to +35
o ℃
m ;
p
e a
t l
e t
h
n
o
t
u
g
o
h
r
g
i
a
t
n
i w
s i
a l
t l
i
o o
n p
, e
r
c a
o t
m e
p
l a
t
t
h h
i e
g
h s
e c
r a
l
a e
n s
d
b
l a
o t
w t
e e
r r
y
t
e l
m i
p f
e e
r
a w
t i
u l
r l
e
s b
e e
c
a o
d m
v m
e e
r n
s d
e
l u
y
s
i
e
n
f
g
f
e
a
c
l
t
c
e
d o
. h
o
l
• CLEANING
We
b
w
a
o
s
u
e
l
d
d

w
r
• P s
l
e o
a f
s
e w
a
d t
o e
r
n
o w
t h
e
u n
s
e c
l
l e
a a
r n
g i
e n
g
a
m t
o h
u e
n
t s
• A f
l r
MAINTENANCE o
w m
a
y t
s h
e
d
i m
s a
c i
o n
n s
n
e p
c o
t w
e
t r
h
e s
u
s p
c p
a l
l y
e
s b
e
f o
o t
r
e r
e
c q
l u
e i
a r
n e
i
n a
g n
. y

• T
r
h
o
e
u
t
s
i
c
n
a
e
l
e
m
a
d
i
o
n
e
t
s
e
n
n
• Interference due to RF disturbance, ground vibration…etc. • Load cells with faulty grinding standard.
• Unstable platform feet adjustments according to bubble level
indication
• Incorrect position or other external objects within weighing
area
• The weighing-scale is not put in a solid & firm ground area,
such as carpet floor or lawn.

3. Connection failure for data transmission to PC or printer


• Wrong connection wires or faulty wires for transmission between
the digital indicator & load cells.
• Wrong indicator models
• Wrong internal wiring or wire broken

In case of the following defective mode occurs, it is suggested to


contact your nearest Authorized Dealer for further technician service
& repair:

1. POWER switch-on failure :


• Push-button faulty
• Short circuit wires - Wire broken
• Safety fuse burnt out
• Wire connection problem
• Main power adaptor faulty – Parts Replacement
2. LCD display faulty
• Possible hardware defects include: Uneven brightness in the LCD
display screen & texts color blurred, smeared rainbow screen,
incorrect decimal display
• LCD PIN broken or short circuit
• PCB cooper foil broken & loosed welding
• Unable to save or read data – IC or transistor faulty, internal parts
broken.
• LCD showing “ERRL” after switch on - Load cell damaged
• Overload may cause the weigh to malfunction.
• Software system crash
• Resonator faulty
-1- - 38 -
occur, please contact your local dealer or electrochemical or electrical interference or humidity, unless this is
CHARDER MEDICAL service partner. attributable to negligence on the part of CHARDER MEDICAL.

Before reading detailed instructions on how to use all the


weighing functions that are built into your scale, please read
the following important guidelines:
• Always be sure that the display shows `Zero` before
use, if it does not then please press the ZERO key.
• The device is designed to detect when a stable
weight is achieved, the indicator will `bleep` twice to
indicate a stable weight value, your reading should
be taken at this point.

• If a fault or defect is present on receipt of the unit


which is within CHARDER MEDICAL’s scope of responsibility,
CHARDER shall have the right to either repair the fault or
supply a replacement unit. Replaced parts shall be the
property of CHARDER. Should the fault repairs or
replacement delivery not be successful, the statutory
provisions shall be valid. The period of warranty shall be two
years, beginning on the date of purchase. Please retain your
receipt as proof of purchase. Should your scale require
servicing, please contact your dealer or CHARDER MEDICAL
Customer Service.
• No responsibility shall be accepted for damage caused
through any of the following reasons: Unsuitable or improper
storage or use, incorrect installation or commissioning by the
owner or third parties, natural wear, changes or modifications,
incorrect or negligent handling, overuse, chemical,
- 39 - -2-
• If operating, climatic or any other influences lead to a major
change in conditions or material quality, the treaty for perfect unit Troubleshooting for defective modes:
functioning shall be rendered null and void. If CHARDER provides Original purchaser can enjoy the benefits under the effective Warranty
against functional defects in material and workmanship subject to the
and individual warranty, this means that the unit supplied will be free terms and conditions listed in the yearly Warranty Program & Return
of faults for the length of the warranty period. Policy.
Our warranty service program includes the following:
1. Technician repair service under warranty or at a service
DISPOSING OF THE SCALE
maintenance charge depending on the workmanship for
• This product is not to be treated as regular household waste, the defective functionality or cause of damage covered by
but should be handed in to an electrical/electronic equipment the warranty.
recycling centre. 2. Parts replacement from the manufacturing factory under
• You can obtain further details from your local council, your the warranty or at a certain cost for the replaced parts plus the
municipal waste disposal company or the firm which you purchased workmanship charge if not covered under the warranty.
the product.
Before you contact our Authorized Dealer in your country for technician
repair service, please read through the following section carefully:

Self-checking Tips:
Some functional defects can be identified and maintained by
users as listed below:

1. Power-on failure
• Check if the main power adaptor has not plugged onto the
scale properly
• Check if the battery power is running low - Replace with new
batteries
2. Indicator showing “0000” ZERO SPAN out of range
• Incorrect weighing result - Avoid damages by external
environment force such as free-drop to the ground, collision by
external objects, etc.
• Proper re-calibration procedure required to correct the setting of
weighing accuracy.
SN-T13000001 Designation of the serial number of
every device, applied at the device.

(Number as an example)

“Please note the accompanying


documents“
or “Observe operating instructions”

Identification of manufacturer of
medical product including address

Charder Electronic Co., Ltd.


No.103, Guozhong Rd., Dali Dist.,
Taichung City 412, Taiwan (R.O.C.)

“Electro-medical appliance“
with attachment for type B

Dispose of old appliances


separately from your household
waste !l!
Instead, take them to communal
collection points.

Carefully read this operation manual


before setup and commissioning,
even if you are already familiar with
Charder scales.

-1- - 36 -
Transport and storage temperature 5.
+60° limit indicating the upper and the
C lower limit
(Transport and storage temperature
-20°C on packaging)

Place the battery housing back and make sure that the housing pin is connecting to the right point

6.

7.
X

Place the housing back


3. EMC GUIDANCE AND MANUFACTURER’S
DECLARATION

Guidance and manufacturer’s declaration-electromagnetic


emissions
The MEDICAL SCALE MS21NEOV is intended for use in the
electromagnetic environment specified below.
The customer or the user of the MEDICAL SCALE MS21NEOV should
assure that it is used in such an environment.

Electromagnetic
Emission test Compliance
environment-guidance
RF emissions Group 1 The MEDICAL SCALE
CISPR 11 MS21NEOV uses RF energy
only for its internal function.
Rechargeable Battery and AA-size Battery Housing
Therefore, its RF emissions
are very low and are not likely
to cause any interference in
4. nearby electronic equipment.
X Class B The MEDICAL SCALE
RF emissions
MS21NEOV is suitable for
CISPR 11
use in all establishments,
Class A including domestic
Harmonic emissions establishments and those
IEC 61000-3-2 directly connected to the
public low-voltage power
Voltage fluctuations Compliance
/flicker emissions supply network that supplies
IEC 61000-3-3 buildings used for domestic
purposes.

-9- 34
Electrostatic ± 6 kV contact ± 6 kV contact Floors should be wood,
discharge(ESD) ± 8 kV air ± 8 kV air concrete or ceramic tile.
IEC 61000-4-2 If floors are covered with
1.
synthetic material, the
relative humidity should
be at least 30%
Electrical fast ± 2kV for power ± 2kV for power Mains power quality 2.
transient/burst supply lines + supply lines Not should be that of a
IEC 61000-4-4 1kV for applicable typical commercial or
input/output lines hospital environment.
Surge IEC ± 1kV line(s) to ± 1kV differential Mains power quality
61000-4-5 line(s) ± 2kV mode Not should be that of a
line(s) to earth applicable typical commercial or
hospital environment.
Voltage Dips, <5% UT(>95% <5% UT(>95% Mains power quality Open the battery housing cover
short dip in UT) for 0,5 dip in UT) for 0,5 should be that of a
interruptions cycle 40% cycle 40% typical commercial or
and voltage UT(60% dip in UT(60% dip in hospital environment. If
variations on UT) for 5 cycles UT) for 5 cycles the user of the
power supply 70% UT(30% dip 70% UT(30% dip MEDICAL SCALE
input lines IEC in UT) for 25 in UT) for 25 MS21NEOV requires
61000-4-11 cycles <5% cycles <5% continued operation
UT(>95% dip in UT(>95% dip in during power mains
UT) for 5 s UT) for 5 s interruptions, it is
recommended that the Take out the battery housing
MEDICAL SCALE
MS21NEOV be
powered from an
uninterruptible power
supply or a battery.
Power 3 A/m 3 A/m The MEDICAL SCALE
frequency(50/60 MS21NEOV power
Hz) magnetic frequency magnetic
field IEC 61000- fields should be at levels
4-8 characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
X

Tray Assembly

Recommended separation distance between portable and mobile RF


communications equipment and the MEDICAL SCALE
The MEDICAL SCALE MS21NEOV is intended for use in an electromagnetic
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies. environment in which radiated RF disturbances are controlled. The customer
NOTE2 These guidelines may not apply in all situations. Electromagnetic or the user of the MEDICAL SCALE MS21NEOV can help prevent
propagation is affected by absorption and reflection from structures, objects and electromagnetic interference by maintaining a minimum distance between
people. portable and mobile RF communications equipment (transmitters) and the
a Field strengths from fixed transmitters, such as base stations for radio MEDICAL SCALE MS21NEOV as recommended below, according to the
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and maximum output power of the communications equipment.
FM radio broadcast and TV broadcast cannot be predicted theoretically with Rated maximum output Separation distance according to frequency of transmitter m
power of transmitter
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
measured field strength in the location in which the MEDICAL SCALE d =1,2√P d =1,2√P d =2,3√P
MS21NEOV is used exceeds the applicable RF compliance level above, the
MEDICAL SCALE MS21NEOV should be observed to verify normal operation. 0,01 0,12 0,12 0,23
If abnormal performance is observed, additional measures my be necessary,
3.such as re-orienting or relocating the MEDICAL SCALE MS21NEOV. 0,1 0,38 0,38 0,73
b Over the frequency range 150 kHz to 80 MHz, field strengths should be les than 1 1,2 1,2 2,3
3 V/m.
10 3,8 3,8 7,3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where p is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.

NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic
33 propagation is affected by absorption and
- 10 -reflection from structures, objects
and people.
Model MS 21neo V

Capacity 15kg x 2g <6kg>5g

Accuracy ±3g < 6kg > ±7.5g


1.
Weight Unit kg

LCD Display 1.0” LCD display with 5 digits

Dimension 550 x 290 mm (Tray)

ON/OFF, ZERO, PRINT, BMI, HOLD,


Key
PRETARE, TARE , CLEAR, ENTER, 0~9 , 2.
Functions
INSTRUCTION CONNECTINGFOR M1-5 CHARGING AND
Y 6-AA size batteries.
Power Supply Y Rechargeable battery pack (optional).
Y 12V / 15Vswitching adaptor. The Medical Scales MS21NEOV listed below are the certified
If prompt Adjustable
displays onFeetthe LCD, please
x 1 , Flat headcharge the scale
machine bolt x models under the Council Directive 2007/47/EC for standard of
with Standard
MS 21neo V exclusive adaptor or replace the batteries. safety, health protection and accurate metrology characteristics
CAUTION: 3 , Flat head machine bolt x 4 , Plate x 1,
Locate adaptor plug on the right side of indicator.
Accessories
Y Always connect the User manualwith
AC adaptor x 1 ,the
Screw driver
indicator x1,
before of medical scales:
RS232
connecting to the mainscable
powerxsupply.
1 , 12V 2A Adaptor x 1
Y Please disconnect the adaptor from main power supply PRODUCT DESCRIPTION
Mechanical
before taking out the ACHeight
adaptorMeasurement HM80M
pin from indicator.
Options
Digital Height Measurement HM80D
Medical Scale
Operation 5℃ / 35℃
Temperature

RS232 Connector
AC Jack Connector

- 11 - 32
Height Measurement

AMP DRAWING NO.: CE APPROVED TYPE NO. TYPE


VOLTAGE / MODEL NO.:
9V DC 100mA AD-038A D41W1090100-13/1 EU
9V DC 100mA AD-0484 D35W090100-23/1 US
9V DC 100mA AD-0484 SP35-90100 US
9V DC 100mA AD-037A D41WK090100-13/2 UK
9V 200mA AD-8082A UE05WCP-0900205PC AU

12V 2A AD-8057 UE24WV-120200SPA EU


Connecting with PC
12V 2A AD-8058 UE24WU-120200SPA US
12V 2A AD-8056 UE24WB-120200SPA UK 1. Start Hyper Terminal
12V 2A AD-8074 UE24W4-120200SPAS AU Start Hyper Terminal program from clicking
12V 1A N/A UE24WV-120100SPA EU
Start Menu → Programs → Accessories → Communication →
12V 1A N/A UE24WV-120100SPA UK
Hyper Terminal.
15V300mA AD-8079A UE05WCP-150030SPC EU 2. New Connection Description
15V300mA AD-016D D41W150300-13/1 US Give new connection a name then click OK.
15V300mA AD-8064 MTP121UL-120100A US
3. Select Your COM Port
15V300mA AD-8079B UE05WCP-150030SPC UK
15V300mA AD-8079C UE05WCP-150030SPC AU Click Connect to select your COM port. Usually there’s only one
15V 300mA AD-0420 D41WI150300-13/1 EU option for select. Then click OK.
15V 300mA AD-0370 D41WK150300-13/2 UK
4. Port Settings
15V 300mA AD-0482 D41WA150300-13/2 AU
Click Bits per second to set up rate at 9600, Data bits at 8, Parity at
None, Stop bits at 1 and Flow control at Hardware. Then click OK to
complete your setting.
5. Output Data
When the patient has been weighed and BMI calculated, simply

33 - 10 -
press the PRINT key to output data from scale to PC or an Optional
Printer.

PRINTING FUNCTION
Standard Accessories
The Weighing, BMI and Height results can be printed-out for records No. Accessories Item Spec. Qty
using RS232 interface cable (included in accessory kit), which is
connected with the round plug at the back terminal. 1 Adjustable Feet SW-1111 1
After weighing and calculating BMI simply press PRINT key to print
out the results. Flat head machine
2 M4*0.7*8 4
bolt
The format presented below is the standard format of results print-
out and cannot be changed. Flat head machine
3 M4*0.7*15 3
bolt(for plate)
GROSS WEIGHT 60.00kg TARE WEIGHT 30.00kg NET WEIGHT 30.00kg PATIENT HEIGHT 100.0cm PATIENT B.M.I 37. 5
4 Plate(for column) SS-3771 1

5 User manual IN-1223 1

6 Screw driver 1

Parameters of RS232 interface 7 RS232 cable WR-8159 1


Set parameters of the scale interface on the connected device. It is
not possible to change the scale parameters.
• Baud rate: 9600 bps
8 12V 2A Adaptor AD-0522 1
• Parity check: None
• Data length: 8 bits
• Stop bit: 1 bit
• Handshake: RTS/CTS
- 11 - 32
• Data code: ASCII

kg-only face plate


Switch ON the scale and long press [ZERO] key for 3 seconds, first
“SETUP” and then “A.OFF” will display successively.

AUTO-OFF TIME SETUP


This enables operator to select the auto turn OFF time of the device.

BUZZER ON/OFF SETUP


Y This enables operator to select beep sound ON/OFF.
Y Y
Y Auto off time: 120 sec/180 sec/240 sec/300 sec/off
Y Buzzer: On/Off
Y

240 S 240 S
*Please Press ‘Tare’ key to confirm the setting. If you want to select as auto off time, press ‘Tare’ key when displays on

the screen.
** To apply the changes proceed till END displays on the screen
17
33 - 10 -
- 26 -
KEY FUNCTION

Press numeral key 1 ~ 5 to assign Key Description


the position to pre-tare weight.
ON/OFF Turn ON/OFF the scale

PRINT Data transmission via interface(Print-out the results)

Press ENTER key to store pre- BMI To determine Body Mass Index
tare weight; the instrument will
make beep sound. To hold weighing result on the display / determine stable
HOLD
weighing value

C. RECALL PRE-TARE VALUE To reset the display to 0.0kg display / Zero the scale (±2%
ZERO
DESCRIPTION EXAMPLE of full capacity)

Switch weighing unit between kg and lb (non approval


Long press PRE-TARE key for 3 UNIT
model)
seconds; the display will show pre-
tare value- m1 firstly. The pre-tare M1-5 To store 5 pre-tare values (approval model).
value will blink.
PRE-TARE function for subtraction of a known container
Press numeral key 1 ~ 5 to choose pre-tare value PRE-TARE weight or any other object before weighing process.

Press ENTER key to confirm the Tare allows the user to zero the instrument to cancel the
pre-tare weight; the instrument weight of a container/ clothes from the reading of the
TARE
will automatically perform tare instrument, thus giving the true weight of the product/
function. person being tested, while weighing.

CLEAR To clear the wrong entry while entering digits.

ENTER To confirm the performing functions.

0-9 Entering digits


- 11 - 32
Press PRE-TARE key to return to
Normal Mode.

- 25 - 18

lb B. Enter Weight Using 0~9 Keys:


DESCRIPTION EXAMPLE

Press PRE-TARE key, and the


extreme left digit will blink.

Enter pre-tare weight using 0~9


keys. For example- 5kg.

Definitions
Stable symbol: To indicate that the weight is stable.
Minor weight: Weight under zero. Press ENTER key to confirm the
pre-tare weight; the display will
Zero symbols: Weight is at zero point.
show the minus sign on left of
Low battery: Battery need to charge or replaced.
the pre-tare weight value.

Press M1-5 key; the blinking m


sign will appear on the display.

33 - 10 -
Tare allows the user to zero the instrument to cancel the weight of a container/ clothes from the reading of the instrument, thus giving the true weight of
the product/ person being tested.
Y Place the object need to be tare on the tread platform.
Y Press TARE key after the weight stabilizes and stable sign displays.
Y Place the object need to be weighed (without removing the tare object)
To delete the saved tare value, remove the tare object from the tread platform and press [TARE] key.

MS 21neo V Approval type can store 5 sets of pre-tare values, and there are two methods to store pre-tare value – Using Dead Weight or Enter weight

STORING AND RE-CALLING PRE-TARE WEIGHT


using 0~9 keys:
A. Using Dead Weight:
DESCRIPTION EXAMPLE

Press M1-5 key after loading the


weight on the platform; the display
shows blinking m sign.

- 11 - 32
Press numeral key 1 ~ 5 to assign
the position to pre-tare weight.

- 23 -
Long press HOLD key for 3 seconds to enter the TIME SETTING mode, beginning with the top row, with the flashing digit. Press HOLD key after
successful change to move to next step.
EX: To input Dec 25, 2008, 8:00am.
Enter year using keys from 0~9.
Press HOLD key after successful
entry to move to next.
Enter date.
Ex: 12/25 enter “12.25”
Press HOLD key after successful
entry to move to next.

Enter time.

Display
Format YYYY→MM.DD→TT:SS

• Switch on the scale using key. The diagnostic scale self- check is performed and the software version is displayed.
• The „0,00 kg” weight displays on the screen, now the scale is ready for weighing.

Note: If „0,00 kg” won’t display on the screen, press key to

zero the scale, can be used any time to zero the scale.
• Place a person in the middle of the scale. Wait until the scale stabilizes and stable sign (ο) shows on the screen
Direction:
If a person is heavier than the scale capacity, the display will show the „Err” prompt (= overload).
33 - 10 -
- 20 -
1. In normal mode, press BMI key to enter into BMI mode.
MS 21neo V is provided with the integrated hold function 2. The display will show last height and the extreme left digit of
(determination of average value). It enables people to be weighed figure will blink.
3. Enter the height using (0~9) number keys. (ex:170cm)
accurately although they are not still on the scale plate.
• MS21neo V can be connected with automatic height
Note: Determination of average value is not possible when a person moves too
much. measuring rod, after connecting; automatically extracts the
• Switch ‘ON’ the scale using [ON/OFF] key. The diagnostic self- height measurements and displays the height of the person.
check will be performed and after that „0.0 kg” will display on 4. Proceed to weigh as normal. The instrument will show the
weight, height, and BMI value.
the screen with ‘stable’ and ‘zero sign’.
5. At this stage, the weight and height can be changed, and the
• Move the object/ person on the scale platform.. MS21neo V will automatically calculate the BMI value according
• Press the [HOLD] key. The ‘HOLD’ will display on the screen to the changed weight and height.
with blinking triangle and after few seconds the average of 6. HOLD feature can also be used to lock the weight, height, and
BMI results. Press HOLD key again to disable the lock function.
fluctuating weight will display on the screen and will lock on the
7. Press BMI button to return normal weighing mode and the
display.
person can now take off the scale.
• To release the locked weight on display simply press the
[HOLD] key again or remove the object/ person from the scale Body Mass Index Categories
platform and make sure there is no weight on platform: the Classification of weight for adults over 18 years on the basis of Body
display will return to „0.0 kg” display automatically. Mass Index according to WHO, 2000 EK IV and WHO 2004 (WHO -
• HOLD key function can be activated before or after putting the World Health Organization).
weight on the tread platform. But in case of weighing unstable
person it is recommended to press HOLD key after the person
moves on the tread platform.

- 11 - 32