Вы находитесь на странице: 1из 28

SEE MORE AT SORInc.

com

Request Quote

Temperature Switches

The SOR® temperature switch utilizes a


vapor-pressure thermal system. Fluid vapor pressure changes
predictably according to the influence of temperature on the sensing
bulb. Process temperature changes cause proportional vapor pressure
changes in the temperature sensing bulb that act on the diaphragm/
piston assembly to actuate and deactuate a snap-action
electrical switching element at discrete process
temperatures. The instrument’s behavior is
determined by the vapor pressure principle.

Application Information
Basic models with direct and six-foot remote
temperature bulbs can be specified from the
quick selection guide on page 5.

More specific application


requirements can be met by
selecting optional components,
such as housings and electrical
switching elements,
from the balance
of the catalog.

Form 220 (08.16) ©SOR Inc. SORInc.com | 913-888-2630 | Registered Quality System to ISO 9001 1/28
Temperature
Switches Features
Features and Benefits

Robust Construction Snap-Action Electrical Switching Element


 Rugged, high-cycle rate tolerance,  Long life, high load capacity, high ambient

long life, not critical to vibration, temperature limit, insensitive to vibration,


high overrange and proof pressures, SPDT or DPDT switching action,
excellent corrosion resistance to optional “hermetically sealed” capsule
hostile environments. for hazardous locations and hostile
 Enclosure ratings: NEMA 1, 4, 4X, 7, environments.
or 9 available.
Shock/Vibration
 Ingress protection rating up to IP66.
 Select models tested to MIL-S-901D

(Navy) shock test.


Vapor Pressure Principle
 Select models tested to MIL-S-167
 Device’s behavior is predictable and
vibration test.
in accordance with the vapor pressure
principle. Factory Calibration
 Minimal ambient temperature  FREE! Calibrated to customer’s set
influence, fast response, high point, ready to install.
repeatability, narrow dead band.
Vapor Fill Fluid Agency Listings/Certification
 Excellent chemical and thermal
• Select models with ATEX, IECEx, CSA,
stability, predictable temperature- GOST R, INMETRO, Rostechnadzor
vapor pressure curve, nonflammable, (RTN), TestSafe, UL
low toxicity. • Meets most code and customer
requirements.
Direct Immersion Temperature
Sensing Bulbs Safety Certified to IEC 61508 (SIL)
 SOR products are certified to IEC 61508
 316SS can withstand 2300 psig

(1000 psig on 105 range) without for non-redundant use in SIL1 and
thermowell, faster response time, SIL2 Safety Instrumented Systems for
lower cost. most models. For more details or values
applicable to a specific product, see the
Remote Mount Sensing Bulbs Safety Integrity Level Quick Guide
 316SS capillary tube with 300 Series (Form 1528).
SS armor allows instrument to be
Warranty
panel mounted and bulb to be
 3 years from the date of manufacture.
remotely located.
 Standard 300 Series SS armor
protects capillary.

2/28 Registered Quality System to ISO 9001 | 913-888-2630 | SORInc.com Form 220 (08.16) ©SOR Inc.
Temperature
Switches Principle

How the SOR Temperature Switch Works


The SOR temperature switch consists of a pressure switch that has a sealed temperature sensing
bulb attached directly to the pressure port. (An optional remote temperature sensing bulb can be
connected to the pressure port with an armor-clad capillary.) The temperature sensing system
is partially filled with a fluid. Process temperature changes cause proportional vapor pressure
changes in the temperature sensing bulb that act on the diaphragm/piston assembly to actuate
and deactuate a snap-action electrical switching element at discrete process temperatures. The
instrument’s behavior is determined by the vapor pressure principle. (The 105 range unit is similar,
except the fill fluid is inert gas.)

Dual (HI-LO)
SOR temperature switches in this catalog may be specified with two set points. The two set points
may be set at either the same actuation point or split up to full scale with no interaction between set
points. The Dual HI-LO is available with hermetically sealed, explosion proof, UL Listed and CSA
Certified electrical switching elements or with a wide selection of UL Listed and CSA Certified
snap-action switching elements for both AC and DC service. The housing selection must be V1 or
V2. See page 9.

Drawing 0190316

Form 220 (08.16) ©SOR Inc. SORInc.com | 913-888-2630 | Registered Quality System to ISO 9001 3/28
Temperature
Switches How to Order

Model Number System

201 AH-EF 125-U9 -C7A-TT


Temperature Housing Switching Adjustable Diaphragm Pressure Port Accessories
Bulb Type Element Range System

Quick Selection Guide


Basic SOR temperature switches with standard parts are normally suitable for a variety of
industrial applications. Refer to the Quick Selection Guide section on page 5. Corrosive service
and particular customer requirements may require optional components. Refer to the How to
Order section on this page or the dedicated page to locate optional components, such as:
Housings, switching elements, and accessories. Each position in the model number, except
‘Accessories’, must have a designator.

Applications
SOR temperature switches in this catalog are suitable for a wide variety of process and fluid
power applications. Specific application requirements can normally be met by selecting optional
components, such as switching elements. Certain applications may require customized specials.
Consult area representative or the factory.
Weatherproof, conventional explosion-proof and hermetically sealed, explosion proof models are
presented in this catalog.

How to Order
Steps 1 through 4 are required; steps 5 and 6 are optional. Orders must have complete model
numbers, i.e., each component must have a designator.

Step 1: Determine if direct or remote (and capillary length) sensing is required. Select
temperature Sensing Bulb Designator from specifications (page 7).
Step 2: Determine the adjustable range required. Select Range Designator from specifications
(page 8).
Step 3: Select Housing for type of service (pages 9 and 10).
Step 4: Select Switching Element for housing and electrical service (pages 11 and 12).
Step 5: Select Diaphragm System (page 12).
Step 6: Select Accessories as required for service (pages 13 and 14).
Step 7: Determine if Thermowell is required. Select from tables on page 14 and order as a
separate item.
If Agency Listed, Certified or Approved temperature switches are required, see page 15 for
components that must be specified.

4/28 Registered Quality System to ISO 9001 | 913-888-2630 | SORInc.com Form 220 (08.16) ©SOR Inc.
Temperature
Switches Quick Selection Guide

Class II, Groups E, F & G; Divisions 1 & 2


Explosion Proof Class I, Groups C & D;

Class I, Groups A, B, C & D; Class II,


Explosion-Proof Switching Element
Weatherproof, NEMA 4, 4X, IP66

Groups E, F & G; Divisions 1 & 2


Contains Hermetically Sealed
Weatherproof-NEMA 4, 4X, IP66 Explosion Proof-Conventional Explosion Proof-Hermetically Sealed
(Direct Mount Shown) (Direct Mount Shown) (Direct Mount Shown)

Direct Mount Temperature Switches


SPDT Form “C” Contacts
Maximum Process Pressure 2300 psi
Electrical
Adjustable Overrange Typical Dead
Rating Model Number
Range Temperature Band
@250VAC
201NN-K115-U9-C7A l
150 to 375oF 520oF 15 amps 2.2oF 1.2oC
201L-K115-U9-C7A l
66 to 190oC 270oC
5 amps 6.6 F
o
3.7 C
o
201AH-EF115-U9-C7A l
201NN-K125-U9-C7A l
40 to 225oF 360oF 15 amps 1.2oF 0.7oC
201L-K125-U9-C7A l
5 to 107oC 182oC
5 amps 3.6 F
o
2.0 C
o
201AH-EF125-U9-C7A l
201NN-K135-U9-C7A l
-50 to 70oF 190oF 15 amps 1.4oF 0.8oC
201L-K135-U9-C7A l
-45 to 21oC 88oC
5 amps 4.2 F
o
2.3 C
o
201AH-EF135-U9-C7A l
Remote Mount Temperature Switches with 6’ Capillary
(not shown)
SPDT Form “C” Contacts
Maximum Process Pressure 2300 psi
Electrical
Adjustable Overrange Typical Dead
Rating Model Number
Range Temperature Band
@250VAC
203NN-K115-U9-C7A l
150 to 375oF 520oF 2.2oF 1.2oC
15 amps 203L-K115-U9-C7A l
66 to 190oC 270oC
3.3oF 1.8oC 203BA-KB115-U9-C7A l
203NN-K125-U9-C7A l
40 to 225oF 360oF 1.2oF 0.7oC
15 amps 203L-K125-U9-C7A l
5 to 107oC 182oC
1.8 F
o
1.0 C
o
203BA-KB125-U9-C7A l
203NN-K135-U9-C7A l
-50 to 70oF 190oF 1.4oF 0.8oC
15 amps 203L-K135-U9-C7A l
-45 to 21oC 88oC
2.1 F
o
1.2 C
o
203BA-KB135-U9-C7A l

Design and specifications are subject to change without notice. For latest revision, see sorinc.com.

Form 220 (08.16) ©SOR Inc. SORInc.com | 913-888-2630 | Registered Quality System to ISO 9001 5/28
Temperature
Switches Glossary of Terms

SOR recognizes that there is not an industry convention with respect to terminology and definitions pertinent
to temperature switches. The following list applies to SOR Temperature Switches.

Temperature Switch Maximum Process Pressure


A bi-stable electromechanical device that actuates/ The maximum process pressure to which the
deactuates one or more electrical switching temperature sensing bulb should be exposed without
element(s) at a predetermined discrete temperature being protected by a thermowell.
(set point) upon rising or falling temperature.
Repeatability
Adjustable Range
The ability of a temperature switch to successively
The span of temperature between upper and lower operate at a set point that is approached from a
limits within which the temperature switch can be starting point in the same direction and returns to the
adjusted to actuate/deactuate. It is expressed for starting point over consecutive cycles to establish a
increasing temperature. temperature profile. The closeness of the measured
set point values is normally expressed as percentage
Set Point
of full scale (maximum adjustable range temperature.)
That discrete temperature at which the temperature
switch is adjusted to actuate/deactuate on rising or Repeatability is 1% of full scale for ranges 135, 125
falling temperature. It must fall within the adjustable and 115. Range 105 has a repeatability of 2% of full
range and be called out as increasing or decreasing scale.
temperature.

Dead Band SPDT Switching Element


The difference in temperature between the increas- Single-Pole, Double-Throw (SPDT) has three
ing set point and decreasing set point. It is expressed connections: C-Common, NO-Normally Open and
as “typical,” which is an average with the increasing NC-Normally Closed, which allows the switch to be
set point at mid-adjustable range with the standard K electrically connected to the circuit in either NO or
switch element. It is normally fixed (not adjustable). NC state.

Hermetically Sealed DPDT Switching Element


A welded steel capsule with glass-to-metal, factory- DPDT is two synchronized SPDT switching elements
sealed electrical leads that isolates the electrical which actuate together at increasing set point and
switching element(s) from the environment. deactuate together at decreasing set point. Discrete
SPDT switching elements allow two independent
Overrange circuits to be switched; i.e., one AC and one DC.
Overrange temperature is that temperature to which
the sensing bulb can be continuously exposed The synchronization linkage is factory set, and is
without causing permanent change of set point or not field adjustable. Synchronization is verified by
distortion sufficient to cause leakage or significant connecting test lamps to the switching elements and
degradation of the fill fluid. Temperatures greater observing them go “On” simultaneously at actuation
than overrange could cause permanent damage and and “Off” simultaneously at deactuation.
render the device inoperative.

6/28 Registered Quality System to ISO 9001 | 913-888-2630 | SORInc.com Form 220 (08.16) ©SOR Inc.
Temperature Step 1: Sensing
Switches Bulb Designator
201AH-EF125-U9-C7A-TT
Temperature Bulb Type

Capillary Length
Designator Mounting Configuration Process Connection
feet meters
201 Direct - -
203 6.0 1.8
205 10.0 3.0 1/2” NPT(M)
Remote
207 15.0 4.5
209 20.0 6.0

Notes
1. For applications where a special length capillary 2. Special bulb dimensions are available. Contact
system is required, contact the factory or your the factory for details.
local representative for specifications and
delivery.

Form 220 (08.16) ©SOR Inc. SORInc.com | 913-888-2630 | Registered Quality System to ISO 9001 7/28
Temperature
Switches Step 2: Adjustable Range
201AH-EF125-U9-C7A-TT

Adjustable Ranges
Adjustable Range Overrange Maximum Process
Typical Dead Band
Designator Increasing Temperature Temperature Pressure
°F °C °F °C °F °C psi bar
135 -50 to 70 -45 to 21 1.4 0.8 190 88 23001 158
125 40 to 225 5 to 107 1.2 0.7 360* 182* 23001 158
115 150 to 375 66 to 190 2.2 1.2 520 270 2300 1
158
105** 300 to 1000 150 to 540 15 8.3 1100 590 1000 70

* Overrange temperature decreases to 250oF (120oC) when NB option is specified. See accessories on
page 13.
** Remote mount only.
1
Overrange is reduced to 1150 psi when the CV accessory is selected.

Dead Band Considerations

1. Dead band values are expressed as typical


Switching Element Designators Multiplier
expected at mid-range using the standard K
switching element. When optional switching K, KA, W 1.0
elements are specified, corresponding dead D, E, J, JR, KB, M, Y 1.5
band multipliers must be applied to the A, AD, B, EF, G 3.0
typical dead band values shown in the table L, JF, YY 3.5
whenever optional switching elements other
AF, EE 4.0
than K, KA or W are used.
BD, EB, JJ, S 5.0
2. Dead bands are fixed, except when T or H
switching elements are used. EG 5.5
3. Dead band can be widened by selecting an AA, BB, GG, JB, JG KK 6.0
optional switching element with a multiplier LL 6.5
greater than 1.0. AG 8.5
Example: Model 201NN-G125-U9-C7A
T 2.5 to 6.5
Typical standard dead band: 1.2°F
Switching Element G multiplier: 3 H 1.0 to 3.0
Corrected typical dead band: 1.2°(3) = 3.6°F

8/28 Registered Quality System to ISO 9001 | 913-888-2630 | SORInc.com Form 220 (08.16) ©SOR Inc.
Temperature
Switches Step 3: Housing
201AH-EF125-U9-C7A-TT

General Purpose NEMA 1


Electrical: 3/4” NPT(F) - Right Electrical: 3/4” NPT(F) - Left, Electrical: Exposed contacts
Material: Aluminum Right Material: Aluminum
See Agency Listings page 15 Material: Aluminum Open bracket with exposed switching
See Switching Element Groups 1, See Switching Element Groups element - does not meet NEMA 1
2, 3 & 4 on page 10.
PP P3 1, 2, 3 & 4 on page 10. H3 See Switching Element Groups 1 & 3 on
page 10.

Weatherproof - NEMA 4, 4X, IP66


Electrical: 3/4” NPT(F) - Right Electrical: 3/4” NPT(F) - Left, Electrical: 3/4” NPT(F) - Right
Material: Aluminum Right Material: Aluminum
See Agency Listings page 15 Material: Aluminum Cover: Heavy Duty with Viton gasket
See Switching Element Groups 1, See Agency Listings page 15 See Switching Element Groups 1, 2, 3 &
NN 2, 3 & 4 on page 10. N3 See Switching Element Groups N4 4 on page 10.
1, 2, 3 & 4 on page 10.

Electrical: 3/4” NPT(F) - Right Electrical- RN: Electrical- RT: 3/4” NPT(F) - Right
Material: Carbon Steel 3/4” NPT(F) - Right Electrical- RS: M20 x 1.5 - Right
Electrical- RM: M20 x 1.5- Standard terminal block
See Switching Element Right Material: 316SS
Groups 1 & 3 on page 10. Standard terminal block See Agency Listings page 15
N6 RN RT
Material: Aluminum RS See Switching Element Groups
RM See Agency Listings page 15 1, 2, 3 & 5 on page 10.
See Switching Element Groups
1, 2, 3 & 5 on page 10.

Electrical: 3/4” NPT(F) - Top Electrical: 90o conduit block Electrical: 3/4” NPT(F) - Right
Standard terminal block 1/2” NPT(M) Manual reset only
Material: Aluminum 18” free wire leads Standard terminal block
See Agency Listings page 15 Material: Series 2000 Material: Aluminum
See Switching Element Groups 1, 3 Aluminum See Agency Listings page 15
*V1 & 4 on page 10.
AC Switching Element AD only. RB See Switching Element Group 7 on page 10.

Hazardous Locations - Hermetically Sealed Switching Element NEMA 4, 4X, 7, 9, IP66


Contains UL Listed and CSA Certified ATEX/IECEx Certified Contains UL Listed and CSA Certified
hermetically sealed switching element. (Ex db IIC T6/T5 Gb) hermetically sealed switching elements.
Electrical: 1/2” NPT(M) - Top Electrical: 1/2” NPT(M) - Top Electrical: 3/4” NPT(F) - Top
Material: Copper-free** aluminum Material: Aluminum
Material: Copper-free** aluminum See Switching Element Group 5 on
See Switching Element Group 5 on See Agency Listings page 15
page 10.
AG page 10. BG See Switching Element Group 5 on
page 10.
*V2
Contains UL Listed and CSA Certified ATEX/IECEx Certified Contains UL Listed and CSA Certified
hermetically sealed switching element. (Ex db IIC T6/T5 Gb) hermetically sealed switching elements.
Electrical: 1/2” NPT(M) - Top Electrical: 1/2” NPT(M) - Top Electrical: 3/4” NPT(F) - Top
Material: 316SS Material: 316SS Material: Copper-free** aluminum
See Switching Element Group 5 on See Agency Listings page 15 Weatherproof: NEMA 4/4X
See Switching Element Group 6 on
AH page 10. BH See Switching Element Group 5 on
page 10.
*BA page 10.

* Not recommended for direct mount where vibration is expected. Housing should be securely mounted to a flat
surface (bulkhead or panel rack) or pipe stanchion.
** Consult the factory.

Form 220 (08.16) ©SOR Inc. SORInc.com | 913-888-2630 | Registered Quality System to ISO 9001 9/28
Temperature
Switches Step 3: Housing
201AH-EF125-U9-C7A-TT

Hazardous Locations - Conventional Explosion Proof NEMA 4, 4X, 7, 9, IP66


UL Listed Class I, Groups C & D; UL Listed Class I, Groups C & D;
Class II, Groups E, F & G; Divisions Class II, Groups E, F & G; Divisions 1
1 & 2 as an outlet box only & 2 as an outlet box only
Electrical: 3/4” NPT(F) - Right Electrical: 3/4” NPT(F) - Right Separate electrical and set point
Material: Cast Iron Material: Copper-free**aluminum adjustment compartments
*L Weatherproof with Option CG *LC Weatherproof
See Switching Element Groups 1, 2, Weatherproof
See Switching Element Groups 1 Standard terminal block
& 3 below. 3 & 4 below.
See agency listings on page 15.
UL Listed Class I, Groups C & D; UL Listed Class I, Groups C & D;
Class II, Groups E, F, & G; Divisions Class II, Groups E, F & G; Divisions
1 & 2 as an outlet box only 1 & 2 as an outlet box only
Electrical: 3/4” NPT(F) - Left, Right, Electrical: 3/4” NPT(F) - Left, Right, *B3 Electrical: 3/4” NPT(F) - Left, Right
Top Material: Aluminum
Top
*S Material: Cast Iron *SC Material: Copper-free**aluminum
Weatherproof with Option CG Weatherproof
See Switching Element Groups 1, See Switching Element Groups *B4 Electrical: M20 x 1.5 - Left, Right
3, & 7 below. 1, 2, 3, 4 & 7 below. Material: Aluminum

Class 1, Groups A, B, C, D;
Electrical: M20 x 1.5 - Left, Right
Class II, Groups E, F, & G, Divisions *B5 Material: Cast Iron
1 & 2 as an outlet box only
Electrical 3/4” NPT(F) - Left, Right,
Top
*TA Material (Housing): Aluminum *B6 Electrical: 3/4” NPT(F) - Left, Right
Material (Cover): Aluminum Material: Cast Iron
Line Mounted. Weatherproof with
Option CG.
See Switching Element Groups 1 Switching Element Groups 1, 2, 3, 4 & 5 below.
& 3 below.

* Not recommended for direct mount where vibration is expected. Housing should be securely mounted to a flat surface
(bulkhead or panel rack) or pipe stanchion.
** Consult the factory.

Switching Element Group / Housing Compatibility


Group 1 Group 2 Group 3 Group 4 Group 5 Group 6 Group 7
A, AA, B, BB, BD*,
GG, KK, AF, AG, EF, EG, EB, JB, JR,
C**, E, EE, G, J, JJ, T H D, M
LL, YY JF, JG KB
K, KA, L, S, W, Y

*BD only available with RN & RT housings **C micro switch is not available in L, S and TA housings

10/28 Registered Quality System to ISO 9001 | 913-888-2630 | SORInc.com Form 220 (08.16) ©SOR Inc.
Temperature Step 4: Switching
Switches Element
201AH-EF125-U9-C7A-TT

Cross reference compatibility chart on page 10 to ensure that switching element will fit in housing.

Switching Electrical Electrical Dead Band


AC Rating DC Rating Resistive Designator
Element Contact Connection Multiplier
Service Type Type Volts Amps Volts Amps Volts Amps SPDT DPDT SPDT DPDT
Normal
250 15 125 0.4* 30 5.0* 1.0 6.0 K KK
Service AC

Low Power 125 1 - - 28 1.0* 1.0 - KA N/A


Gold Contacts 125 1 - - 30 1.0 1.5 5.0 J JJ
Wide Dead Band AC 250 15 125 0.5 - - 3.0 6.0 G GG
Double Switching Element DPDT -(2) SPDT Synchronized actuation/deactuation at increasing/decreasing Set Points

AC or DC 250 11 125 0.5* 30 5.0 3.0 6.0 A AA


All other Switching Elements - 18” 18 AWG Color-Coded Wire Leads except when terminal blocks are specified.

Wide Dead Band DC 250 15 125 0.5 30 10.0* 3.5 6.5 L LL


Narrow Dead
250 5 125 0.5* 30 5.0* 1.5 4.0 E EE
Band DC
250 5 125 0.3 - - 3.0 6.0 B BB
Hi-Ambient
Temperature 250 5 125 0.5* - - 1.5 3.5 Y YY
Rating - 400°F
250 5 125 0.3 - - 1.0 - W N/A
Potted Wire Leads
1/2” NPT(M) 250 11 125 0.5* 30 5.0 3.0 - AD N/A
Condition Connection
Wide Adjustable Dead 2.5 to
250 15 125 0.4* - - - T N/A
Band 6.5
Narrow Adjustable
250 15 - - - - 1 to 3 - H N/A
Dead Band
Manual Reset -
K, KA, G, L, C, N, S, Y, W Switching Elements - Screw Terminals.

Decreasing Temperature
D N/A
(Automatic Actuation-
Increasing Temperature)
Manual Reset - 250 15 125 0.5 - - 1.5 -
Increasing Temperature
T & H Switching Elements - Consult the factory.

(Automatic Actuation- M N/A


Decreasing
Single Switching Element SPDT - (1) SPDT

Temperature)
Corrosion 250 15 125 0.4* 30 5.0* 1.5 - KB N/A
Resistant Explosion- 250 5 125 0.5* 30 5.0* - 5.0 N/A EB
Proof Hermetically
Sealed Switching 250 11 125 0.5* 30 5.0 4.0 8.5 AF AG
Element 250 5 125 0.5* 30 5.0 3.0 5.5 EF EG
Corrosion Resistant, 125 1 - - 28 1.0* 1.5 - JR N/A
Explosion Proof, Lower-
125 1 - - 30 1.0 - 6.0 N/A JB
Power Service
Hermetically Sealed
125 1 - - 30 1.0 3.5 6.0 JF JG
Gold Contacts
Explosion-Proof EEx d
250 7 250 0.25 30 7.0 5.0 - BD N/A
IIC T6

Form 220 (08.16) ©SOR Inc. SORInc.com | 913-888-2630 | Registered Quality System to ISO 9001 11/28
Temperature Step 4: Switching
Switches Element
201AH-EF125-U9-C7A-TT
Notes
1. Double switching elements have wire leads 5. Switching Elements W & Y have Elgiloy
except when supplied in housings RN, RT, springs.
RB, B3, B4, B5, B6 and V1. Terminal blocks 6. Certain switching elements can handle
are standard in these housings. greater voltage and/or amperage. Consult
2. Dead band multipliers must be applied to the the factory should your requirements
typical dead band figures given in the exceed catalog values. All switching
specification tables on page 8. elements above except BD are UL Listed
3. Switching element ambient temperature and CSA Certified. The DC current ratings
limits: marked with an asterisk (*) are not UL Listed
-65 to 400 F (-54 to 200 C) B, Y, W
o o
but have been verified by testing and/or
-65 to 250oF (-54 to 120oC) A, E & J experience.
-40 to 167oF (-40 to 75oC) AF, AG, 7. Cross reference compatibility chart at the
EB, EF, EG, bottom of page 10 to ensure that switching
JB, JF, JG, element will fit in housing.
JR, KB
-13 to 158oF (-25 to 70oC) BD
-65 to 180oF (-54 to 80oC) All others
4. The hermetically sealed switching element
capsule is UL Listed, CSA Certified and
TestSafe Approved as an explosion proof
snap switch according to the table with con-
ditions and exceptions specified in Note 3.

Hazardous Location
Agency Designator
Conditions
UL Listed Class I, Groups A, B C AF, EF, AG, EG,
CSA & D; Class II, Groups E, KB, EB, JB, JF,
Certified F & G; Divisions 1 & 2 JG, JR
TestSafe Ex s Zone 1 IIC T4 IP65 AF, EF, AG, EG,
Approved Ex tD A21 T105°C IP65 KB, EB

Step 5: Diaphragm System


201AH-EF125-U9-C7A-TT

U8 - Only available on units with CS, CL, ML, NM or WV options or BG/BH housings.
U9 - Available on all units.

12/28 Registered Quality System to ISO 9001 | 913-888-2630 | SORInc.com Form 220 (08.16) ©SOR Inc.
Temperature
Switches Step 6: Accessories
201AH-EF125-U9-C7A-TT

Accessory/Option & Description Designator


Neoprene cover gasket (o-ring) to make L, S and TA explosion-proof housings weathertight. CG
ATEX/IECEx approved temperature switch. See Agency Listings on page 15 for details. CL
CSA Certified temperature switch. Available with PP, NN, RB, RN, RT, B3, B6 & V1. Housing has earth (ground) lug.
CS
See Agency Listings on page 15 for details.
Canadian Registration Number (CRN) - Process ratings may be affected. Consult the factory for details. CV
Cemented cover gasket on weathertight housings. GC
Sealed electrical lead adapter. Provides protection to housing interior, switching element and dry side of pressure sensing
GG
assembly from condensate in electrical conduit and corrosive atmospheres. (Protrudes approximate 2” from housing.)
Universal terminal box. Stainless steel. 1/2” NPT(F). ATEX/IECEx Certified Ex db IIC T4, T5 & T6 Gb. HB
Universal terminal box. Stainless steel. M20 x 1.5(F). ATEX/IECEx Certified Ex db IIC T4, T5 & T6 Gb. HBME
Universal terminal box. Stainless steel. 1/2” NPT(F). FM Approved and CSA Certified Explosion-proof Class I, Groups A, B,
HT
C, & D; Class II, Groups E, F, & G, Class III; Divisions 1 & 2 (NEMA 4X IP65)
Crouse Hinds ECD-15 for Hazardous Locations Class I, Groups C & D; Class II, Groups E, F & G;
Breather Drain on S or SC housings only. KK
Sintered metal plug in weathertight housing.
Terminal block. 6-place compression type standard in B and R series housings. Optional in LC and SC housings. Not
LL
available with all housings. Consult the factory.
Multi-Listed temperature switch. ATEX/IECEx, CSA & UL. Available with B3 & B6 housings. See Agency Listings on page
ML
15 for details.
Special construction to minimize the effect of across ambient temperature changes. Available on Ranges 135 and 125
NB
only. Reduces overranges to 250°F (120°C) on Range 125.
Compliance to NACE Certification MR-01 1-75. NC
INMETRO approved temperature switch. See Agency Listings on page 15 for details. NM
Pipe (stanchion) mounting kit for (1-1/2 to 2” pipe). Order as a separate line item for UL Listed and CSA Certified
PK
temperature switches.
Tag, fiber. Attached with plastic wire to housing. Stamped with customer-specified tagging information. PP
Powder coat epoxy coating. No coating on stainless steel parts or plated screws. (500 hours-salt spray) PY
Tag, stainless steel. Attached with stainless steel wire to housing. Stamped with customer-specified tagging information.
RR
(2 lines, 18 characters and spaces per line.)
Explosion-proof and weathertight electrical junction box with screw terminals. Aluminum 3/4” NPT(F) top or right conduit
connections as required. UL Listed and CSA Certified Class I, Groups A, B, C & D; Class II, Groups E, F & G; TB
Divisions 1 & 2, (AG, AH, BA, L, LC, S, SC & TA housings). Includes cover o-ring for weathertight applications.
Factory set and potted to prevent future adjustment. This option results in permanent Set point. Available only on housing
TP
AC, AG, AH, BG and BH.
Oversize stainless steel nameplate or separate stainless steel tag. Permanently attached to housing. Stamped with
TT
customer-specified tagging information.
Fungicidal varnish. Covers exterior and interior except working parts. VV
UL Listed temperature switch available with B3 or B6 housing. See Agency Listings on page 15 for details. WV
“X” is used as a suffix to the model number for special requirements. Each “X” must by completely identified in the text of
the order or inquiry. When more than one “X” is required, use “X” followed by the number of such items. For example, “X3” X
means three separate otherwise unidentifiable requirements.
Epoxy coating. Exterior only. Polyamide epoxy with 316SS pigment. (200 hours-salt spray) YY
Chained cover with captive screws to conform to former JIC specification. ZZ

Form 220 (08.16) ©SOR Inc. SORInc.com | 913-888-2630 | Registered Quality System to ISO 9001 13/28
Temperature
Switches Step 6: Accessories

Test Certificates
Certificates C1 C3 C4 C5 C6 C8 B5 B6 B7 A1
Calibration ¿ ¿ ¿ ¿ ¿
Inspection Report ¿ ¿ ¿ ¿
Compliance / Conformance ¿ ¿ ¿
Dielectric Test ¿ ¿
Insulation Resistance ¿ ¿ ¿
Typical Material of Wetted Parts ¿ ¿

Step 7: Thermowell
1. Determine insertion length from specification table.
2. Specify thermowell for either direct or remote mounted temperature switches from specifications
tables.
3. Specify process connection threading from specification table below.
4. The thermowell must be ordered as a separate item. Thermowells are 316SS (347SS on
275TW-NF100). Consult the SOR representative in your area or the factory for special material.
5. Special sensing bulb diameter and lengths are available. Consult the SOR representative in your
area or the factory to discuss your requirements.

Specifications and Dimensions


U A B Q
Available Sensing
Insertion Element Bore Insertion Process Maximum
Thermowell Bulb(s)
Mounting Length Length Diameter Diameter Connection Process
Model Number
in NPT(M) Pressure
201

205

209
203

207

mm in. mm in. mm in. mm in.

245 TW-DM 075 u 3/4


Direct 114.3 4.5 152.6 6 10.4 0.41 19.1 0.75
245 TW-DM 100 u 1
245 TW-RM 075 u u 3/4 6200 psi
Remote 114.3SALES 4.5 152.6 6 10.4 0.41 19.1 0.75
245 TW-RM 100
MODEL #
u u
ORDER # LINE ITEM # PURCHASE ORDER #
1 @ 500°F
U A B Q MAXIMUM
MODEL PART PROCESS
MATERIAL INSERTION ELEMENT BORE INSERTION PROCESS
NUMBER NUMBER CONNECTION
275 TW-RM 075 u u u u LENGTH LENGTH DIAMETER DIAMETER PRESSURE
3/4
245 TW-DM 075
245 TW-DM 100
0029070
0029072
Remote
316SST 190.5 152.6
114.3
4.50 6.00
7.5 228.6
10.4
0.41
9
19.1
0.75
9.9
3/4 NPTM
1 NPTM
0.39 19.1 0.75
275 TW-RM 100 u u
245 TW-RM 075
u u
0029070 114.3 152.6 10.4 19.1 3/4 NPTM
1
316SST 6200 PSI @ 500°F
245 TW-RM 100 0029072 4.50 6.00 0.41 0.75 1 NPTM
275 TW-RM 075 0029071
316SST
190.5 228.6 9.9 19.1 3/4 NPTM
4700 psi
*275 TW-NF 100 275uTW-RM
u100 u0029073
u Remote 190.5 9.007.5
7.50 228.6
0.39 9
0.75 16.81 NPTM 0.66 26.9 1.06 1
275 TW-NF 100 0029074 347SST
190.5 228.6 16.8 26.9
1 NPTM 4700 PSI @ 1000°F
@ 1000°F
7.50 9.00 0.66 1.06

*Model 275TW-NF100 must be used with Range 105.


SCALE 1

44.5 U INSERTION LENGTH


1.75
25.4 B BORE
1.00

Q 14685 W 105TH ST LENEXA, KS 66215 USA


913-888-2630
ISO-9001
SORINC.COM
DRAWN BY
1/2 NPTF K MITCHELL
PRODUCT CERTIFICATION DRAWING CHECKED BY
ALL DIMENSIONS ARE ±1/16 IN
M SMITH
PROCESS CONNECTION UNLESS OTHERWISE SPECIFIED ENGINEER APPROVAL
MM
LINEAR = J MODIG
IN

Drawing 0190020
DATE
A ELEMENT LENGTH 6.4 14 FEB 2011
0.25 THIS DRAWING IS THE EXCLUSIVE PROPERTY OF SOR.
NO USE WHATSOEVER OF THE INFORMATION CONTAINED
HEREON, NOR REPRODUCTION IN WHOLE OR PART MAY BE
MADE WITHOUT THE EXPRESS WRITTEN PERMISSION OF SOR.
TITLE

DIM DWG
THERMOWELL
DRAWING NUMBER REV
EO NUMBER: 4991

| 913-888-2630 | SORInc.com
0190020 7
14/28 Registered Quality System to ISO 9001 Form 220 (08.16)
SCALE: 1.50
©SORSHEET
Inc.1
DO NOT SCALE PRINT
OF 1
DWG SIZE

B
Model Name: 0190020.PART/7/1
Reset Form
Temperature
Switches Agency Listings

CSA For Hazardous Locations - Class 1, Groups B, C & D; Class II, Groups E, F & G
Pressure Port
Bulb Housing Switching Element Range Diaphragm Material & Accessories
Conn. Size

A, AA, AF, AG, B, BB, C, E, CS or ML Required.


EE, EF, EG, G, GG, H, J, JF, All except CG, GC,
ALL B3, B6 ALL ALL ALL
JG, JJ, K, KA, KK, L, LL, P, S, GG, HB, HT, KK, LL,
T, W, Y, YY ME, TB, TP, ZZ
General Purpose and Weatherproof (CSA Type 4)
FP (General A, AA, B, BB, C, E, EE, G,
Purpose GG, H, J, JJ, JL, K, KK, KA, L,
NN (Type 4) LL, S, T, W, Y, YY
A, AA, AF, AGT, B, BB, C, E,
RN (Type 4) EE, EF, EG, G, GG, GA, H, J,
C7A
RT (Type 4) FJ, JG, JJ, JL, K, KK, KA, L, LL, CS Required. All
ALL ALL U9 Standard Others
S, T, W, Y, YY except LL, GC
as Required
RB (type 4)
D, DA, M (Manual reset only)
RH (Type 4)
A, AA, B, BB, C, E, EE, G,
V1 (Type 4) GA, H, J, JJ, K, KA, L, LA, S,
SA, T, W, Y
ATEX/IECEx or INMETRO Ex db IIC T6/T5 Gb
Pressure Port
Bulb Housing Switching Element Range Diaphragm Material & Accessories
Conn. Size
CL (for all Hsgs)or ML
(for B3/B6 Hsgs)
A, AA, AF, AG, B, BB, C, Req’d for ATEX/IECEx
B3, B4, B5, E, EE, EF, EG, G, GG, H, NM Required for
ALL ALL ALL ALL
B6 J, JF, JG, JJ, JL, K, KA, KK, INMETRO
L, LL, P, N, S, T, W, Y, YY All except CG, GC,
GG, HB, HT, KK, LL,
ME, TB, TP, ZZ
BB, PP, RR, TT, TP,
VV, YY, HB, HBME
ALL BG, BH AF, AG, EF, EG, JF, JG ALL ALL ALL
NM Required for
INMETRO
Ex ia IIC T6...T4 Gb
PK, NC, X
CL Required for
RN, RM, RT,
ALL J, JJ, JF, JG ALL ALL ALL ATEX/IECEx
RS
NM Required for
INMETRO
UL For Hazardous Locations - Class I, Groups B, C, & D; Class II, Groups E, F, & G
Pressure Port
Bulb Housing Switching Element Range Diaphragm Material & Accessories
Conn. Size

A, AA, AF, AG, B, BB, C, WV or ML Required.


E, EE, EF, EG, G, GG, H, All except CG, GC,
ALL B3, B6 ALL ALL ALL
J, JF, JG, JJ, K, KA, KK, GG, HB, HT, KK, LL,
L, LL, P, S, T, W, Y, YY ME, TB, TP, ZZ
Rostechnadzor (RTN) Certificate
Permit for instruments used and operated in hazardous industrial facilities in Russia. Standard on most
models. Certificate available on request.
Form 220 (08.16) ©SOR Inc. SORInc.com | 913-888-2630 | Registered Quality System to ISO 9001 15/28
Temperature
Switches Approximate Weights

Weight Add
Housing (kgs) Accessories (kgs)
(lbs) (lbs)
AC 1 .5 PK Pipe Kit 1.5 0.7
AG, BG, H3 1.5 .75 TB Junction Box with
5 2.25
AH, BH, NN, N3, N4, PP, P3 2 1 Terminal Block

RM, RN 2.5 1.25 HB, HBME or HT


2.5 1.1
Universal Terminal Box
BA, N6, RB, V1 3 1.5
RT 3.5 1.75 Actual shipping weights may vary from the
L, LC, SC 4 2 charted values because of product material,
TA 4.5 2.25 configuration and packaging requirements.
V2 5 2.5
B3, B4 8 3.5
B5, B6 10 4.5

16/28 Registered Quality System to ISO 9001 | 913-888-2630 | SORInc.com Form 220 (08.16) ©SOR Inc.
Temperature
Switches Dimensions
Dimensions in this catalog are for reference only. They may be changed without notice. Contact the factory
for certified drawings for a particular model number. All dimensions shown are expressed as millimeters over
inches. (Linear = mm/in.)

201 Direct Mount


Drawing 0190317

203-209 Remote Mount


Drawing 0190318

Dimensions
C with NB
Feature A-B C D E Diameter
Option
135, 125, 135, 125,
Range All 135, 125, 115 105 135, 125 105 105
115 115
Bulb m ft. mm in. mm in. mm in. mm in. mm in. mm in. mm in.
201 N/A N/A 105.7 4.16 - - 107.2 4.22 - - - - 9.7 0.38 - -
135 5.3 170 6.7
203 1.8 6.0 112.0 4.41 148.3 5.84 112.0 4.41 to to to to 9.7 0.38 16.0 0.63
396 15.6 433 17.1
147 5.8 170 6.7
205 3.0 10.0 124.7 4.91 148.3 5.84 112.0 4.41 to to to to 9.7 0.38 16.0 0.63
409 16.1 433 17.1
185 7.3 170 6.7
207 4.5 15.0 162.8 6.41 148.3 5.84 112.0 4.41 to to to to 9.7 0.38 16.0 0.63
447 17.6 433 17.1
216 8.5 170 6.7
209 6.0 20.0 194.6 7.66 148.3 5.84 112.0 4.41 to to to to 9.7 0.38 16.0 0.63
480 18.9 433 17.1

* With NB option, dimension D is: 135 to 396mm (5.3 to 15.6 in.) - Remote Mount

Form 220 (08.16) ©SOR Inc. SORInc.com | 913-888-2630 | Registered Quality System to ISO 9001 17/28
Temperature
Switches Dimensions
Dimensions in this catalog are for reference only. They may be changed without notice. Contact
the factory for certified drawings for a particular model number. Dimensions in this catalog are
expressed as millimeter over inches (Linear = mm/in.).

Weatherproof-Non-Hazardous
MODEL # Service
SALES ORDER # (NEMA 4,LINE
4X,ITEM # IP66)PURCHASE ORDER #
ST

NN & N3 = (SHOWN) 56.5 BULB


108.1 4.6 2.22 MODEL # A-B
4.25 N4 = ADD
0.18 1829
1.5 203
69.9 PP & P3 = SUBTRACT 28.6 72
2.75 0.06 1.13 3048
205
33.3 120
1.31 26.1 4572
1.03 207
180
6096
209
240
ELECTRICAL
CONNECTION W/NB OPTION SEE ABO
3/4 NPTF STD 125, 135 PER PRO
1/2 NPTF OPT RANGE ONLY MODEL
(BOTH SIDES ON SEE ABO
N3 & P3 HSGS) 40.5 HI-TEMP
PER PRO
1.59 105 RANGE
159.1 MODEL
6.26 NON-STANDA

CLEARANCE SLOTS
66.9 6.4
2.64 FOR 7.1
0.25
HARDWARE WITH 0.28
63.5 76.2
MIN TO MAX
2.50 3.00
MOUNTING CENTERS
52.3 POINT A
2.06

306.3
12.06
TEMPERATURE
PRO
SENSING BULB
Housing: NN, N3, N4, PP, P3 1/2 NPTM
GLAND NUT
Drawing 0190157 A
U

N
H
MA
POINT B C E
TITLE

DIMEN
22.4
ARMORED BULB HEX D (TO WRENCH TIGHT) NN/N3
0.88
CAPILLARY EXTENSION EO NU

SCALE

Model Name: 0190157.ASSEM/4/0

Housing: N6 Drawing 0190173

18/28 Registered Quality System to ISO 9001 | 913-888-2630 | SORInc.com Form 220 (08.16) ©SOR Inc.
Temperature
Switches Dimensions

Dimensions in this catalog are for reference only. They may be changed without notice. Contact
the factory for certified drawings for a particular model number. Dimensions in this catalog are
expressed as millimeter over inches (Linear = mm/in.).

Weatherproof-Non-Hazardous Service (NEMA 4, 4X, IP66)


MODEL # SALES ORDER # LINE ITEM # PURCHASE ORDER #
STANDARD MO
BULB LEN
107.3
4.22 MODEL # A-B C
52.0 65.6 1829 112.0
203
2.05 2.58 72 4.41
69.9 28.6 3048 124.7
205
2.75 1.13 120 4.91
33.3 26.2 4572 162.8
207
1.31 1.03 180 6.41
6096 194.6
209
240 7.66
W/NB OPTION SEE ABOVE
112.0
125, 135 PER PROBE
4.41
RANGE ONLY MODEL #
ELECTRICAL
CONNECTION SEE ABOVE
HI-TEMP 148.3
SEE CHART PER PROBE
105 RANGE 5.84
MODEL #
184.1
40.5
NON-STANDARD MOD
7.25
1.59

6.4 ELECTRICAL
CLEARANCE SLOTS FOR HOUSING
66.9 0.25 CONNECTION
2.64 HARDWARE WITH
63.5 76.2 7.1 3/4 NPTF STD
MIN TO MAX
2.50 3.00 0.28 RL 1/2 NPTF OPT
MOUNTING CENTERS LH & RH
306.3 C M20 X 1.5 F
52.3 POINT A RM
12.06 RH ONLY
2.06 14685 W
BULB 3/4 NPTF STD
EXTENSION TEMPERATRUE RN 1/2 NPTF OPT ISO-9001
SENSING BULB RH ONLY
1/2 NPTM
PRODUCT CERTIFICA
GLAND NUT M20 X 1.5 F
RS ALL DIMENSIONS
RH ONLY UNLESS OTHERW

3/4 NPTF STD LINEAR =


RT 1/2 NPTF OPT
RH ONLY
THIS DRAWING IS
ARMORED POINT B E 3/4 NPTF STD
NO USE WHATSOEVE
HEREON, NOR REPRO
CAPILLARY RU 1/2 NPTF OPT MADE WITHOUT THE E

22.4 LH & RH TITLE

HEX D (TO WRENCH TIGHT)


Housing: RM, RN, RS, RT 0.88
Drawing
RW 0190136 M20 X 1.5 F
LH & RH
DIM DWG 203/209 R-S
EO NUMBER: 5090
MODEL # SALES ORDER # LINE ITEM # PURCHASE ORDER #
M20 X 1.5 F SCALE: 0.50
RY STANDARD MODELS
LH & RH
139.9 DO NOT SCALE PRINT

Model Name: 0190136.ASSEM/5/0+


5.51 BULB LENGTH
114.3 70.0 MODEL # A-B C
4.50 2.76
ELECTRICAL 1829 112.0 125.
57.2 CONNECTION 27.0 203
72 4.41 4.95
2.25 3/4 NPTF (STD) 1.06
1/2 NPTF (OPT) 3048 124.7 138.
205
120 4.91 5.45
4572 162.8 176.
207
180 6.41 6.95
6096 194.6 208.
209
240 7.66 8.20
W/ NB OPTION SEE ABOVE
112.0 125.
125, 135 PER PROBE
4.41 4.95
RANGE ONLY MODEL NO.

SEE ABOVE
HI-TEMP 148.3 162.
PER PROBE
186.2 105 RANGE 5.84 6.38
MODEL NO.
7.33 9.5
0.38
NON-STANDARD MODEL

2X 7.1
0.28
68.6 MOUNTING
2.70 HOLES

14685 W 10

ISO-9001

52.4 POINT A 306.3


2.06 12.06 PRODUCT CERTIFICATI
27.8 ALL DIMENSIONS AR
1/2 NPTM 1.09 UNLESS OTHERWIS
GLAND NUT M
Housing: V1 BULB Drawing 0190094
TEMPERATURE
LINEAR =
I

EXTENSION SENSING BULB


THIS DRAWING IS TH
NO USE WHATSOEVER O
HEREON, NOR REPRODU

| 913-888-2630 | Registered Quality System to ISO 9001


MADE WITHOUT THE EXP
Form 220 (08.16) ©SOR Inc. SORInc.com E 19/28
TITLE

DIM DWG 203/209 V1 D


ARMORED POINT B 22.4 C
HEX EO NUMBER: 4991
CAPILLARY 0.88 D (TO WRENCH TIGHT)
Temperature
Switches Dimensions

Dimensions in this catalog are for reference only. They may be changed without notice. Contact
the factory for certified drawings for a particular model number. Dimensions in this catalog are
expressed as millimeter over inches (Linear = mm/in.).

Weatherproof-Non-Hazardous Service (NEMA 4, 4X, IP66)

Housing: RB - Manual Reset Drawing 0190230

20/28 Registered Quality System to ISO 9001 | 913-888-2630 | SORInc.com Form 220 (08.16) ©SOR Inc.
Temperature
Switches Dimensions

Dimensions in this catalog are for reference only. They may be changed without notice. Contact
the factory for certified drawings for a particular model number. Dimensions in this catalog are
expressed as millimeter over inches (Linear = mm/in.).

Conventional Explosion Proof

Housing: L Drawing 0190026

Housing: B3, B4, B5, B6 Drawing 0190312

Form 220 (08.16) ©SOR Inc. SORInc.com | 913-888-2630 | Registered Quality System to ISO 9001 21/28
Temperature
Switches Dimensions

Dimensions in this catalog are for reference only. They may be changed without notice. Contact
the factory for certified drawings for a particular model number. Dimensions in this catalog are
expressed as millimeter over inches (Linear = mm/in.).

Conventional Explosion Proof

Housing: S Drawing 0190028


MODEL # SALES ORDER # LINE ITEM # PURCHASE ORDER #
STANDARD MODELS
150.6 LENGTH
5.93 BULB
MODEL # A-B C
73.7
2.90 1829 112.0 12
106.4 203
72 4.41 4.
4.19
108.0 3048 124.7 13
4.25 ELECTRICAL 34.0 205
120 4.91 5.
TYP CONNECTION 1.34
3/4 NPTF STD 4572 162.8 17
54.0 8.7 207
1/2 NPTF OPT 180 6.41 6.
2.13 0.34
TYP M20 X 1.5 F OPT 6096 194.6 20
209
240 7.66 8.
W/NB OPTION SEE ABOVE
112.0 12
125, 135 PER PROBE
4.41 4.
RANGE ONLY MODEL #
SEE ABOVE
HI-TEMP 148.3 16
PER PROBE
105 RANGE 5.84 6.
MODEL #
NON-STANDARD MODEL D
212.4
8.36

131.2
5.17 CLEARANCE
6.4
77.3 FOR
0.25
3.04 MOUNTING
HARDWARE

14685 W 105TH
91
ISO-9001 S
52.3 POINT A
2.06
PRODUCT CERTIFICATION D
306.3 ALL DIMENSIONS ARE ±1
12.06 UNLESS OTHERWISE SPE
MM
LINEAR =
POINT B 1/2 NPTM IN

Housing: LC, SC GLAND NUT TEMPERATURE


SENSING BULB
Drawing 0190003
THIS DRAWING IS THE EXCL
NO USE WHATSOEVER OF THE
HEREON, NOR REPRODUCTION
MADE WITHOUT THE EXPRESS W
ARMORED
22/28 Registered Quality System to ISO 9001
CAPILLARY | 913-888-2630
BULB | SORInc.com C Form 220
E (08.16) ©SOR Inc.
TITLE

DIMENSION DRAWING
EXTENSION 203/205/207/209 LC
22.4 EO NUMBER: 5090 DRAW
HEX D (TO WRENCH TIGHT)
0.88
SCALE: 0.50
Temperature
Switches Dimensions

Dimensions in this catalog are for reference only. They may be changed without notice. Contact
the factory for certified drawings for a particular model number. Dimensions in this catalog are
expressed as millimeter over inches (Linear = mm/in.).

Conventional Explosion Proof

Housing: S Manual Reset Drawing 0190308

Housing: TA Drawing 0190184

Form 220 (08.16) ©SOR Inc. SORInc.com | 913-888-2630 | Registered Quality System to ISO 9001 23/28
Temperature
Switches Dimensions

Dimensions in this catalog are for reference only. They may be changed without notice. Contact
the factory for certified drawings for a particular model number. Dimensions in this catalog are
expressed as millimeter over inches (Linear = mm/in.).

Hermetically Sealed-Explosion Proof SALES PAGE 1


MODEL # SALES ORDER # LINE ITEM # PURCHASE ORDER #

76.2 MIN STANDARD MODELS


3.00 BULB LENGTH
INSTALLATION E
MODEL # A-B C D
FACTORY SEALED WIRE LEADS CLEARANCE
1829 112.0 125.7 381.8
COLOR CODED AND MARKED 203 TO
38.1 72 4.41 4.95 15.03
457.2
MINIMUM LENGTH 1.50 3048 124.7 138.4 394.5
18.00 ELECTRICAL 205 TO
CONNECTION 120 4.91 5.45 15.53
1/2 NPTM 4572 162.8 176.5 432.6
207 TO 9.7
180 6.41 6.95 17.03
0.38
6096 194.6 208.3 464.3
209 TO
240 7.66 8.20 18.28
W/NB OPTION SEE ABOVE
WEATHERTIGHT 112.0 125.7 381.8
125, 135 PER PROBE TO
COVER SCREW 4.41 4.95 15.03
RANGE ONLY MODEL #
FLEXIBLE SEAL SEE ABOVE
HI-TEMP 148.3 162.1 418.1 16.0
RETAINER PER PROBE TO
105 RANGE 5.84 6.38 16.46 0.63
MODEL #

168.9 NON-STANDARD MODEL DIMENSIONS


6.65
EXTERNAL GROUND
BH, BG, & JH HOUSING

34.9
1.38

14685 W 105TH ST LENEXA, KS 66215 USA


913-888-2630
52.3 POINT A ISO-9001 SORINC.COM
2.06 306.3
DRAWN BY

K MITCHELL
12.06 PRODUCT CERTIFICATION DRAWING CHECKED BY
ALL DIMENSIONS ARE ±1/16 IN
M SMITH
UNLESS OTHERWISE SPECIFIED
1/2 NPTM MM
ENGINEER APPROVAL
LINEAR = J MODIG
GLAND NUT IN
DATE

POINT B BULB TEMPERATURE 04 APR 2012


EXTENSION SENSING BULB THIS DRAWING IS THE EXCLUSIVE PROPERTY OF SOR.
NO USE WHATSOEVER OF THE INFORMATION CONTAINED
HEREON, NOR REPRODUCTION IN WHOLE OR PART MAY BE
MADE WITHOUT THE EXPRESS WRITTEN PERMISSION OF SOR.
TITLE
E DIM DWG 203/209 MINI-HERMET
ARMORED
CAPILLARY C EO NUMBER: 4991 DRAWING NUMBER REV

0190175 6
22.4 D (TO WRENCH TIGHT) SCALE: 0.69
HEX DWG SIZE
0.88 SHEET 1 OF 1
DO NOT SCALE PRINT B
Model Name: 0190175.ASSEM/6/0
Reset Form
Housing: AG, AH, BG, BH Drawing 0190175
SALES
MODEL # SALES ORDER # LINE ITEM # PURCHASE ORDER #

121.0 STANDARD MODELS


4.77 BULB LENGTH
60.5 MODEL # A-B C D
2.38 80.7 1829 112.0 125.7 381
3.18 203 TO
108.0 72 4.41 4.95 15.0
4.25 FACTORY SEALED LEADS 37.4 3048 124.7 138.4 394
COLOR CODED AND MARKED 1.47 205 TO
54.0 ELECTRICAL 120 4.91 5.45 15.5
457.2
2.13 MINIMUM LENGTH CONNECTION 4572 162.8 176.5 432
18.00 207 TO
3/4 NPTF STD 180 6.41 6.95 17.0
1/2 NPTF OPT 6096 194.6 208.3 464
M20 X 1.5 F OPT 209 TO
240 7.66 8.20 18.2
W/NB OPTION SEE ABOVE
112.0 125.7 381
125, 135 PER PROBE TO
4.41 4.95 15.0
RANGE ONLY MODEL #
SEE ABOVE
HI-TEMP 148.3 162.1 418
PER PROBE TO
105 RANGE 5.84 6.38 16.4
MODEL #
NON-STANDARD MODEL DIMENSIO
1 7.1
220.9 CLEARANCE
8.70 6.4 0.28
FOR 41.6
0.25 1 ADD FOR M20 X 1.5 F ELECTRICAL CONN
MOUNTING 1.64
HARDWARE

87.3
3.44
14685 W 105TH ST LENEXA, KS
913-888-2630
ISO-9001 SORINC.COM
TEMPERATURE DR

SENSING BULB K
PRODUCT CERTIFICATION DRAWING CH
52.3 POINT A ALL DIMENSIONS ARE ±1/16 IN
M
2.06 306.3 UNLESS OTHERWISE SPECIFIED EN
12.06 MM
LINEAR = JM
IN
DA

POINT B 1/2 NPTM 1


GLAND NUT THIS DRAWING IS THE EXCLUSIVE PROPERTY O
NO USE WHATSOEVER OF THE INFORMATION CO
54.7 HEREON, NOR REPRODUCTION IN WHOLE OR PA
2.15 MADE WITHOUT THE EXPRESS WRITTEN PERMISSIO

Housing: BA Drawing 0190002 TITLE

DIMENSION DRAWING
203/205/207/209 BA
ARMORED BULB EO NUMBER: 5095 DRAWING NUMBER

CAPILLARY EXTENSION 22.4 C E 0190002


HEX SCALE: 0.57

| 913-888-2630 | SORInc.com
0.88 SHEET 1 OF
24/28 Registered Quality System to ISO 9001
Model Name: 0190002.ASSEM/7/0
D (TO WRENCH TIGHT)
Form 220 (08.16) ©SOR Inc. DO NOT SCALE PRINT
Temperature
Switches Dimensions

Dimensions in this catalog are for reference only. They may be changed without notice. Contact
the factory for certified drawings for a particular model number. Dimensions in this catalog are
expressed as millimeter over inches (Linear = mm/in.).

Hermetically Sealed-Explosion Proof

Housing: V2 Drawing 0190107

Form 220 (08.16) ©SOR Inc. SORInc.com | 913-888-2630 | Registered Quality System to ISO 9001 25/28
Temperature
Switches Dimensions

Dimensions in this catalog are for reference only. They may be changed without notice. Contact
the factory for certified drawings for a particular model number. Dimensions in this catalog are
expressed as millimeter over inches (Linear = mm/in.).

General Purpose

Housing: AC (NEMA 1) Drawing 0190005

Housing: H3 Drawing 0190008

26/28 Registered Quality System to ISO 9001 | 913-888-2630 | SORInc.com Form 220 (08.16) ©SOR Inc.
Temperature
Switches Notes

Form 220 (08.16) ©SOR Inc. SORInc.com | 913-888-2630 | Registered Quality System to ISO 9001 27/28
SOR Inc. | Lenexa, KS USA | 913-888-2630 | Fax 913-888-0767 | SORInc.com

REGIONAL OFFICES
China Middle East
SOR China | Beijing, China | china@SORInc.com SOR Measurement & Control Equipment Trading DMCC | Dubai, UAE
+86 (10) 5820 8767 | Fax +86 (10) 58 20 8770 middleeast@SORInc.com | +971 4 363 3637 | Fax + 1 913 312 3596

28/28 Registered Quality System to ISO 9001 | 913-888-2630 | SORInc.com Form 220 (08.16) ©SOR Inc.

Вам также может понравиться