DISCUSSION
For experiment a,a depth of the dead zone, in liner probe, we can determine the value of
measurement for 9 lines targets at the dead zone which is D1,D2,D3,D4,D5,D6,D7,D8 and D1(D9). All
the 9 line targets were valued 10.2 mm,9.0 mm, 8.1 mm,6.9 mm,5.9 mm, 4.8 mm, 3.9 mm, 2.6 mm
and 1.8 mm respectively.so,the dead zone depth we got was exceed 0.2mm from the actual depth
which is 10mm. For the curve probe, the value that we obtained for 9 line targets are 11.8 mm,9.3
mm,7.8 mm,6.7 mm,5.9 mm,5.1 mm,4.1 mm,3.9 mm and 4.0 mm. As we can see, the depth of the
dead zone that we got was exceed 1.8 mm which is 11.8 mm from the actual depth. Often,
a
real-time unit has a dead zone greater than 10 inm, which means that the
test object will not be useful in evaluating the dead zone. The depth at
which " ID="I4.18.1">"normal tissue " ID="I4.18.3">texture" begins to be
seen determines the dead zone when one is using a TE phantom (Fig. 4B).
Electronic calipers or other distance indicators (marker dots) are used
to quantitate the depth of the dead zone. . Performance Limits. If the
depth of the dead zone exceeds 15 mm, an attempt should be made to correct
the problem. Unfortunately,our results shows that the readings of the dead
zone for two type of probe don’t showmore then 15mm.
For experiment D, we do the axial literal resolution, Eleven line targets with a curved array are clearly
visible separately. The distances between the line targets are from 1 mm in the central area to 5 mm
in the peripheral area. The curved array of line targets is used for the test of the axial and lateral
resolution. In this image, all 11 line targets are visualized separately and clearly (arrows). The central
part of the line targets is zoomed in the inset. The readings that we got from the liner probe are
5.2mm(D1),4.3MM(D2),2.8MM(D3),2.5MM(D4),1.6MM(D5)1.4MM(D6),2.2MM(D7),3.1MM(D8)4M
M(D9)5.0MM(D10) andfor the curve probe
5.0mm(D1),4.3MM(D2),2.9MM(D3),2.2MM(D4),1.6MM(D5)1.4MM(D6),2.5MM(D7),4.0MM(D8)4M
M(D9)5.1MM(D10). so we can conclude it as we success because our readings was near with the actual
rnage readings.