Vapor-Phase Hydrogen Peroxide

Resistance of Environmental
Isolates
Robert R. Reich* and Ross A. Caputo

B
The authors examine the vapor-phase arrier technology has become the de facto standard in
hydrogen peroxide resistance of microbial aseptic processing and sterility testing, with more than
isolates recovered from the controlled 200 units in use worldwide. Isolators in aseptic pro-
environments of pharmaceutical and/or cessing applications were featured prominently in the
US Food and Administration good manufacturing practice
medical device companies and compared
(GMP) guidance about sterile drug products produced by asep-
them with commercially available biological tic processing (1) as well as in several presentations delivered
indicators under various test conditions. by FDA’s team leader for guidance and policy, Richard Fried-
man (2, 3). The Pharmaceutical Inspection Convention rec-
ommendation document, “Isolators Used for Aseptic Process-
ing and Sterility Testing,” (4) and the PDA technical report,
“Design and Validation of Isolator Systems for the Manufac-
turing and Testing of Health Care Products” also discuss using
isolators for these applications (5).
Although using isolators in the pharmaceutical and medical
device industries continues to increase—as does the generation
of technical reports, guidelines, and publications for their de-
sign and validation—the industries and regulatory agencies do
not have consensus on several issues. Two issues are germane
to the present study: (a) the air classification for the background
environment where an isolator is located, and (b) the level at
which microbial reduction is appropriate to demonstrate a val-
idated isolator decontamination process.
The isolator’s interior should meet Class 100 (ISO 5, Euro-
pean Union [EU] Class A) standards as a minimum, but the
classification of the environment surrounding an isolator is not
Figure 1: Typical vapor-phase hypdrogen peroxide concentration as clear cut. FDA has recently stated that a Class 100,000 (ISO
profiles. 8) background environment “can be appropriate, depending
on isolator design and manufacturing situations. An aseptic
processing isolator should not be located in an unclassified area”
(1). The EU guidance on GMPs states in Annex 1 that the back-
ground environment for an isolator “should be controlled and
for aseptic processing it should be at least Grade D” (6). PDA
has suggested that the background isolator environment be con-
Robert R. Reich is the president and trolled but not be classified (5).
Ross A. Caputo, PhD, is the founder The air classification of the isolator background environment
and CEO of PSI, 909 Orchard Street, will influence the microbial challenge to the isolator system. In-
Mundelein, IL 60060, tel. 847.566.9229, fax digenous flora, either airborne or operator- or personnel-
847.566.4960, rreich@pharmsystems.com.
related vectors, is a major potential source of contamination to
*To whom all correspondence should be addressed. the interior isolator environment. This contamination could
occur through integrity breaches in the gloves, rapid transfer
50 Pharmaceutical Technology AUGUST 2004 www.phar mtech.com
sure conditions of temperature, relative humidity (RH), and
VHP concentration, and delivers them to the test BIs in a square
wave fashion. This allows us to determine VHP D-values and
standardize BI performance testing. The equipment’s perfor-
mance is analogous to ANSI/AAMI BIER unit standards for
steam and ethylene oxide (9). Using this equipment, Caputo
examined the resistances of commercially available VHP BIs
(10). He reported a wide range of VHP resistance values among
commercial BIs, large disparities among the manufacturers, re-
sistance claims, and discrepancies with the actual experimen-
tal resistance values.

Materials and methods


Environmental isolates. Microbial isolates were obtained from
the controlled environments of various US pharmaceutical and
Figure 2: Typical vapor-phase hypdrogen peroxide and water medical device facilities from the contract PSI environmental
concentration profiles. monitoring and microbial taxonomy program. Identical mi-
croorganisms, isolated from various locations, were selected for
ports, or isolator structure or through mouse holes in open iso- evaluation to ascertain whether there were any differences in
lators. Therefore, indigenous organisms challenge isolator de- resistance. Bacterial isolates were identified using the Ribo-
contamination procedures. Decontamination cycles must allow Printer technology (DuPont Qualicon, Wilmington, DE), which
an appropriate safety margin to provide confidence in the cycle’s is based on the DNA sequence of the 16S RNA gene (11). Molds
robustness. were identified using traditional methods. Representative
Although the definition of the “appropriate” level is open to colonies were selected for identification and study.
discussion, FDA’s latest draft aseptic-processing guideline says, A total of 26 isolates—19 bacterial and 7 fungal species—
“Normally a four- to six-log reduction can be justified, de- were examined and are listed below. The number in parenthe-
pending on the application” (1). Yet PDA contends that a three- ses represents the number of isolates of that species evaluated.
log reduction is sufficient (5). The FDA guideline further ad- Bacterial species:
vises that the specific biological indicator (BI) spore titer should ● Bacillus subtilis (2)
be known and justified. Neither the FDA nor PDA document ● Bacillus cereus (2)
addresses the resistance of the BI, which is an equally impor- ● Bacillus licheniformis (2)
tant performance parameter to total count. The total resistance ● Staphyloccus aureus (5)
of the BI is a function of its population and organism resistance ● Staphyloccus warneri (3)
(D-value). This principle has been stressed by Pflug (7). ● Micrococcus luteus (2)
Vapor-phase hydrogen peroxide (VHP) is currently the most ● Staphyloccus epidermitis (3).
widely used isolator decontamination agent. Many BIs for mon- Fungal species:
itoring the effectiveness of the VHP decontamination cycle are ● Candida albicans (1)
commercially available. Unfortunately, no VHP BI monograph ● Penicillium species (3)
or performance standard specifies the minimum acceptable re- ● Aspergillus species (2)
sistance values, standard test conditions, or the testing equip- ● Aspergillus niger (1).
ment. These standards are available for commercial BIs designed After each organism was identified, it was grown on the ap-
to monitor other antimicrobial processes—such as saturated propriate culture medium, harvested, washed in a physiological
steam, ethylene oxide, or dry heat. As a result, BI manufactur- buffer, and prepared in a test-suspension. For resistance deter-
ers certify their monitors at various conditions using different mination, approximately 103–106 cells were inoculated onto a
test apparatuses. In fact, some manufacturers only certify the 316 stainless steel substrate. Conidia spores were used to chal-
population of their BIs and do not address VHP resistance at lenge the resistance of the mould cultures and endospores were
all. This situation makes it difficult for users to select an ap- used for the challenge of bacteria of the genera Bacillus and
propriate BI for their particular application and to generate Geobacillus.
meaningful performance comparison data. Ultimately, one val- The resistance testing of the isolates was performed in a VHP
idates the isolator decontamination process to demonstrate mi- BIER unit (VHyPer, PSI) using the following exposure condi-
crobial control of the interior environment in support of asep- tions, unless otherwise noted:
tic manufacturing and/or sterility testing activity. The correlation ● VHP: 1.25 mg/L

of the BI resistance to the potential indigenous microbial con- ● RH: 25%

tamination must be understood to make the BI an effective tool. ● temperature: 30 C

Khorzad reported on the design, performance, and valida- ● time: 0.5–2.5 min, in 0.5-min increments

tion of a VHP biological indicator evaluator resistometer (BIER) ● pre- and post-temperatures: 30 C

unit (8). This unit precisely and accurately controls the expo- ● packaging: no packaging.

52 Pharmaceutical Technology AUGUST 2004 www.phar mtech.com


from three manufacturers, pack-
Table I: Comparative resistance Table II: Comparative resistance of multiple aged as they were received and chal-
of multiple lots of BIs (packaged). lots of BIs (unpackaged). lenged under identical test condi-
D-Value (min) D-value (min) tions: 1.7 mg/L H2O2, packaged, 35
Manufacturer 15% RH C, and D-value in minutes.
1.0 1.3 1.5 1.7
AS 1.2 Table II lists the comparative re-
Manufacturer mg/L mg/L mg/L mg/L
AS 1.3 sistance of multiple lots of BIs, chal-
BR — — 1.3 —
SB 2.8 lenged without primary packaging,
BR — — 1.3 —
AB 0.5 under identical test conditions:
BR — — 2.1 —
AB 0.5 10–15% RH, 35 C, and D-value in
AB — — — 0.1
AB 0.5 minutes. Data again show a great
SB 19.6 1.9 — —
AB 0.8 variation in resistance among var-
SB 10.1 3.1 — —
ious manufacturers.
OB 5.8 2.8 — 5.3
The exposure sequence was as fol- Table III demonstrates the effect
OB 7.3 2.2 — 1.7
lows: of testing the same lot of BIs within
OB 9.8 3.0 — 3.6
● pretreatment for 5 min at 30 C their primary packaging (as received
● exposure to square wave conditions, ● temperature: 30–35 C from the manufacturer) at the following
which is defined as moving from zero- ● time: 0–5.5 min, in 0.5-min increments test conditions: 1.5 mg/L H2O2, 15% RH,
VHP concentration to the desired VHP ● pre- and post-temperatures: 30–35 C 35 C, and D-value in minutes.
concentration and conditions in 3 s. ● packaging: nonwoven high-density poly- Table IV demonstrates the effect of test-
Flow velocity of the defined gaseous ex- ethylene (1073B Tyvek, DuPont, Wilm- ing the same lot of BIs removed from the
posure mixture to the test sample was ington, DE) or unpackaged packaging at the following test conditions:
30 ft/min. ● substrate: 316 stainless steel or glass 1.7 mg/L H2O2, 15% RH, 35 C, and D-
● post-treatment for 5 min at 30 C to aer- cover slips. value in minutes.
ate samples and ensure that residual VHP The exact testing conditions will be in- Table V shows RH’s affect on the resis-
does not kill additional microorganisms dicated in the results section for each data tance of the same lot of BIs, challenged
and skew the killing time results. set. The VHP BIER unit exposure sequence under identical conditions except for the
We assessed resistance by one of two was similar to that described for the envi- % RH: 1.0 mg/L H2O2, 35 C, and D-value
methods: (a) a fraction/negative (F/N) ronmental isolates. in minutes.
screening method to indicate whether a As described for the environmental iso- Table VI demonstrates the effect of in-
particular isolate’s D-value was smaller lates, we conducted resistance by either the oculating the same suspension of G.
than that of a conventional commercial F/N approach or the survivor curve method. stearothermophilus spores onto various
BI, or (b) a five-point survivor curve Incubation of the G. stearothermophilus BIs carrier substrates on the resistance of the
method. F/N D-values were calculated was conducted at 55–60 C and plate counts suspension. The SB lot was run under the
using the Stumbo, Murphy, and Cochran were performed after 48–72 h. following conditions: 0.9 mg/L H2O2,
or Spearman–Karber (12) methods. A packaged, 10% RH, 30 C. The BR lot was
total of 10 samples were used for each F/N Results run under the following conditions: 1.5
screening time. For the survivor curves, This section summarizes the results of mg/L H2O2, packaged, 15% RH, 35 C.
quadruplicate samples were used for each BIER performance, commercial BIs, and The humidity difference between the two
time cut. Each dilution was plated in du- environmental isolates. The resistance val- test conditions may have contributed to
plicate. A least-squares analysis was used ues are presented as relative values for the both the high resistance values in Table VI
to determine the line of best fit. Samples purpose of illustrative experimental analy- and the resistance disparity between lots
were incubated for a minimum of 72 h at sis rather than values to be taken literally. SB and BR.
a temperature appropriate for the indi- BIER performance. Figure 1 shows typi- Environmental isolates. Table VII tabu-
vidual isolate. cal VHP concentration profiles for BIER lates the VHP resistance of the bacterial
Commercial BIs. A total of five commer- unit exposures at three VHP concentra- isolates under the following conditions:
cial G. stearothermophilus BIs were eval- tions. Figure 2 demonstrates typical VHP 1.25 mg/L H2O2, 28% RH, and 30 C. The
uated. For the purpose of this study, the and water concentration profiles for a resistance (D-value) of these isolates var-
manufacturers were designated by the BIER unit exposure. ied by as much as one log and all the S.
codes AB, AS, BR, OB, and SB. Resistance During installation, the BIER unit was aureus strains were catalase positive. On
was evaluated at various exposure condi- challenged 30 times at each of the two op- the basis of these results, the production
tions designed to mimic actual isolator erating conditions to establish interrun of catalase did not afford these organisms
decontamination conditions. Resistance and intrarun variations. These data any unique VHP resistance. Finally, the S.
testing was conducted in the VHP BIER demonstrated the precise and accurate de- warneri strains, which were catalase neg-
unit using the following conditions: livery of challenge parameters (8). ative, exhibited higher resistance than the
● VHP: 0.9–1.8 mg/L Commercial BIs. Table I lists the com- Bacillus spores. This result can probably
● RH: 10–30% parative resistance of multiple lots of BIs be attributed to clumping when drying on
54 Pharmaceutical Technology AUGUST 2004 www.phar mtech.com
Table III: Same-lot testing (packaged Table IV: Same-lot testing (packaged Table V: RH effect.
and unpackaged). and unpackaged). D-value (min)
D-value (min) D-value (min) Manufacturer 10% RH 35% RH
Manufacturer Packaged Unpackaged Manufacturer Packaged Unpackaged OB 5.8 2.8
BR 3.9 1.3 AB 0.5 0.1 OB 7.3 3.7
BR — 2.1 AB 0.5 — OB 9.8 3.4
BR — 1.3 AB 0.5 — SB 19.6 2.8
AB 0.8 — SB 10.1 2.4
the carrier, which tends to result in artifi-
cially high resistance. post-exposure sterility testing. These pro- Table VI: Inoculation effect.
Table VIII tabulates the VHP resistance cedures should not be performed without
of the fungal isolates under the following realizing the potential ramifications. D-value (min)
Manufacturer Steel Glass
conditions: 1.25 mg/L H2O2, 28% RH, and FDA has suggested, in its draft aseptic
SB 28.1 57.8
30 C. As with the bacterial isolates, the processing guideline, that when evaluat-
BR 3.9 3.1
resistance (D-value) varied by as much as ing the efficacy of an isolator decontam-
1 log between the fungal isolates. ination procedure, “an appropriate, quan-
tified BI challenge should be placed on Table VII: Resistance summary of
Discussion various materials ...” (1). The present data bacterial isolates.
The data indicate that the VHP BIER unit indicate that the supporting substrate can
Estimated D-value
provides precise and accurate delivery of influence the BI resistance. This phe-
standardized challenge conditions, thereby nomenon has been observed with gaseous Organism (min)
providing a standardized method of eval- ethylene oxide processes (18). Although B. subtilis (1) 0.7
uation of BI resistance. The data further many factors, including the texture of the B. subtilis (2) 0.8
suggest that a wide variation in the resis- surface, the porosity of the surface, the B. cereus (1) 0.1
tance of VHP commercially available BIs surface tension of the substrate—which B. cereus (2) 0.1
exists. Depending on the methodology used could lead to clumping of the inoculum— B. licheniformis (1) 0.1
for resistance determination, these values and the cleanliness of the carrier mater- B. licheniformis (2) 0.1
can vary by as much as one log. This find- ial, can influence BI resistance on a par- S. aureus (1) 0.2
ing is consistent with data from studies that ticular substrate, the observed substrate S. aureus (2) 0.5
indicate a wide variation in reported BI effect appears to be real. S. aureus (3) 0.5
VHP resistance values (e.g., D-value of 2.0 The VHP resistance of environmental S. aureus (4) 0.1
min at 1.5 mg/L VHP concentration for B. isolates varies greatly. The data showed S. aureus (5) 0.1
[now G.] stearothermophilus [13] and D- differences among organism types as well S. warneri (1) 1
values in the 5–10 second range for the as the same organism isolated for various S. warneri (2) 0.8
same organism [14]). This widely reported areas of the country. Most isolates had a S. warneri (3) 0.2
resistance variation is probably at least par- lower VHP resistance than that reported S. epidermidis (1) 0.3
tially caused by the testing methodology. If for most BIs. Some isolates, however, had S. epidermidis (2) 0.5
these values are used to develop deconta- a greater VHP resistance than that re- S. epidermidis (3) 0.13
mination cycles or to calculate or support ported for some commercially available M. luteus (1) 0.2
decontamination safety factors, greatly di- VHP BIs. This fact emphasizes the need M. luteus (2) 0.2
vergent cycles could result. to know and then calibrate the resistance
Both the testing of RH as well as the BI of the BI selected to conduct isolator cycle tion of the indigenous flora of the envi-
packaging also have been shown to greatly development and validations. This fact ronment enclosed by the isolator.
influence the D-value. The RH situation further highlights the need for BI manu- A reduction of six logs of a certified BI
appears very analogous to that of ethylene facturers and the pharmaceutical indus- is achievable when the conditions such as
oxide, in which the resistance of organisms try to work together to establish consen- VHP concentration, RH, load configura-
is increased at RH levels below 30% sus on appropriate D-value ranges for tion, or substrates are controlled. If any of
(15–17). In addition, the nonwoven, high- commercially available BIs used with VHP. these variables are uncontrolled or mis-
density, polyethylene pouch typically used In addition, the typical indigenous flora understood and a BI positive is generated,
to package BIs increases the resistance of of the isolator environment and its corre- then too often the cause is immediately
the BI unit. Unpackaged BIs demonstrate lation to BI resistance must be well un- attributed to “bad” BIs. Microbiologists
a considerably lower resistance. Isolator- derstood. Finally, proper calibration of the and engineers experienced in sterilization
validation practitioners must appreciate BIs would also help to eliminate some of science understand the concept of ap-
and understand this fact. BIs are some- the observed variability. It is therefore rec- proximate first-order kinetics: the more
times removed from their package for easy ommended that VHP resistance be con- robust the decontamination/sterilization
suspension in the isolator or to facilitate sidered as part of the initial characteriza- cycle, the straighter the line. If this logic
56 Pharmaceutical Technology AUGUST 2004 www.phar mtech.com
Table VIII: Resistance summary of References 7. I.J. Pflug, “Microbiology and Engineering of
1. FDA, “Guideline for Industry. Sterile Drug Sterilization Processes,” in Environmental
fungal isolates. Products Produced by Aseptic Process- Sterilization Laboratory, I.J. Pflug (University
ing–Current Good Manufacturing Practice,” of Minnesota, Minneapolis, MN, 4 ed., 1980).
Estimated D-value draft guidance (FDA, Rockville, MD, 2003). 8. D. Khorzad et al., “Design and Operational
Organism (min) 2. R.L. Friedman, “Revision of the 1987 Asep- Qualification of a Vapor-Phase Hydrogen
C. albicans 0.7 tic Processing Guideline,” paper presented at Peroxide Biological Indicator Evaluator Re-
Penicillium species (1) 0.1 the 2002 PDA annual meeting, New Orleans, sistometer Unit,” Pharm. Technol. 27 (11),
Louisiana, 10–12 December 2002. 84–90 (2003).
Penicillium species (2) 0.1 9. Association for the Advancement of Medical
3. R.L. Friedman, “FDA Aseptic Processing
Penicillium species (3) 0.9 Guideline Update,” paper presented at the Instrumentation (AAMI), “Resistometers
Aspergillus species (1) 0.2 International Society for Pharmaceutical En- Used for the Characterization of the Perfor-
Aspergillus species (2) 0.1 gineering (ISPE) Conference, 12th Annual mance of Biological and Chemical Indica-
Aspergillus niger 1.3 Barrier Technology Forum, Arlington, Vir- tors,” ANSI/AAMI, ST44 (2003).
ginia, 2–5 June 2003. 10. R.A. Caputo, “Comparative Resistance of Va-
4. Pharmaceutical Inspection Convention porized Hydrogen Peroxide Biological Indi-
of linear extrapolation is accepted, then the (PIC), “Isolators Used for Aseptic Process- cators from Various Vendors,” paper pre-
first solution to a BI positive should not be ing and Sterility Testing,” (PIC, Geneva, sented at the ISPE Barrier Technology Forum,
to decrease the acceptance criteria for a suc- Switzerland, 2004). Zurich, Switzerland, 23–24 September 2002.
cessful cycle from a six-log to a three-log 5. PDA Technical Report #34, “Design and Val- 11. M. Sethi, “Ribotyping with the RiboPrinterR
idation of Isolator Systems for the Manufac- Microbial Characterization System,” in Rapid
reduction. If the decontamination/ Analytical Microbiology: The Chemistry and
turing and Testing of Health Care Products,”
sterilization conditions are less than robust, (PDA, Bethesda, MD, 2002). Physics of Microbial Identification, Wayne
the answer should not be to reduce the 6. European Commission, “Manufacturing of Olsen, Ed. (David Horwood International
challenge but to fix the exposure parame- Sterile Medicinal Products,” in The Rules Gov- Publishing Ltd., Godalming Surrey, UK,
ters. The microbial load of the product and erning Medicinal Products in the European 2002), pp. 81–122.
Union, Medicinal Products for Human and 12. I.J. Pflug and R.G. Holcomb, “Principles of
environment will continue to be unknown Thermal Destruction of Microorganisms,”
Veterinary Use: Good Manufacturing Practices
and must be managed and addressed (Office for Official Publications of the Eu- in Disinfection, Sterilization and Preservation,
through a deep understanding of the cycle ropean Communities, Luxembourg, Vol. 4, S.S. Block (Lea and Febigen, Philadelphia,
parameters and their interrelationships. 1997), Annex 1. PA, 3d ed., 1983), pp. 751–810.
13. S.M. Davenport, “Meeting the Goals of In-
creased Sterility Confidence for Advanced
Aseptic Processing: Implementation of a Re-
stricted Areas Barrier System,” paper pre-
sented at the Barrier Users Group Sympo-
sium, Bethesda, Maryland, 31 May 1995.
14. J. Akers and J. Agalloco, “Isolators: Validation
and Sound Scientific Judgment,” J. Pharm.
Sci. Technol. 54 (2), 110–111 (2000).
15. J.E. Whitbourne and R.R. Reich, “Ethylene
Oxide Biological Indicators: Need for Stricter
Qualification Testing Control,” J. Parenteral
Drug Assoc. 33, 132–143 (1979).
16. R.R. Reich and L.L. Morien,“Influence of En-
vironmental Storage Relative Humidity on
Biological Indicator Resistance, Viability and
Moisture Content,” Appl. Environ. Microbiol.
43 (3), 609–614 (1982).
17. R.A. Caputo and T.E. Odlaug, “Sterilization
with Ethylene Oxide and Other Gases,” in
Disinfection, Sterilization and Preservation,
S.S. Block, Ed. (Lea and Febigen, Philadel-
phia, PA, 3d ed., 1983), chapter 2.
18. D.J. Burgess and R.R. Reich, “Industrial Eth-
ylene Oxide Sterilization,” in Sterilization
Technology: A Practical Guide for Manufac-
turers and Users of Health Care Products, R.
Morrissey and G.B. Philips, Eds. (Van Nos-
trand Reinhold, New York, NY, 1993), pp.
152–195. PT