Вы находитесь на странице: 1из 18

FLAVOUR AND FRAGRANCE JOURNAL

Flavour Fragr. J. 2008; 23: 297–314


Published online 19 August 2008 in Wiley InterScience
(www.interscience.wiley.com) DOI: 10.1002/ffj.1887

Linear retention indices in gas chromatographic


John Wiley & Sons, Ltd.

analysis: a review
Barbara d’Acampora Zellner,1 Carlo Bicchi,3 Paola Dugo,4 Patrizia Rubiolo,3
Linear retention indices in gas chromatographic analysis

Giovanni Dugo1 and Luigi Mondello1,2*


1
Dipartimento Farmaco-chimico, Facoltà di Farmacia, Università di Messina, Viale Annunziata, 98168 Messina, Italy
2
Campus-Biomedico, Via Alvaro del Portillo 21, 00128 Roma, Italy
3
Dipartimento di Scienza e Tecnologia del Farmaco, Facoltà di Farmacia, Università di Torino, Via Pietro Giuria 9, 10125
Torino, Italy
4
Dipartimento di Scienze degli Alimenti e dell’Ambiente, Facoltà di Scienze MM.FF.NN., Università di Messina, Salita
Sperone, 98166 Messina, Italy
Received 9 January 2008; Revised 10 May 2008; Accepted 22 May 2008

ABSTRACT: The main purpose of any chromatographic analysis is to resolve mixtures of compounds into less complex
mixtures or ultimately into pure components. In addition to this function, the chromatographic system can provide retention
data which serve as complementary information for the positive identification of resolved components. The need to express
gas chromatographic retention data in a standardized system has long been recognized and retention index values
presented to be a valuable parameter. Those values are mainly calculated by applying the equations proposed by Kováts,
for isothermal analysis, and van den Dool and Kratz, for programmed gas chromatographic runs. In general, these indices
denote the retention behaviour of the compounds of interest according to a uniform scale determined by a series of
closely related standard substances. The use of retention indices in the flavour and fragrance field is well-documented,
and they are widely applied for the comparison of results between laboratories, as well as to characterize stationary
phases. Copyright © 2008 John Wiley & Sons, Ltd.

KEY WORDS: gas chromatography; gas chromatography/mass spectrometry; retention index; linear retention index; flavour
and fragrance materials

Introduction must be added that as gas chromatographic separations


are performed over a wide range of temperatures and on
Since the early days of gas chromatography (GC) an stationary phases of different polarities, no single sub-
enormous effort has been devoted to the standardization stance would fulfil the role of a universal standard.
of the methods used to determine retention data, with the Moreover, when the retention of the components of a
main target being a wider and reliable use of collated sample and the standard are grossly different, accuracy
published results for identification purposes. Already in would be impaired.
1952, in their first publication on gas–liquid chromato- In general, the most thoroughly studied, diffused and
graphy, A. T. James and A. J. P. Martin1 mentioned that accepted retention index calculation methods are based on
the retention volume of a pure substance on a certain gas the equation developed by Kováts in 19583 for isothermal
chromatographic column is a characteristic value which conditions, and on the equation propounded by van den
could be used for the identification of sample components. Dool and Kratz in 1963,4 used in the case of temperature-
Further, in 1955, Littlewood et al.2 introduced the specific programming conditions. Values calculated using the
retention volume as an attempt to make retention data latter approach are commonly denominated in literature
independent of temperature, stationary phase film thickness, as retention index (I), linear retention index (LRI) or
carrier gas flow rate and pressure drop across the column; programmed-temperature retention index (PTRI or IT), while
however, this parameter was not practically applied. From values derived from the former equation are usually
the many proposals a general consensus has emerged called retention index (I ) or Kováts index (KI). Retention
favouring the use of relative retention data expressed index calculations are applied to volatile compounds
with respect to the retention of a standard substance. It present in numerous extracts, such as flavour and fragrance
materials, foods, drugs, petrochemicals, environmental
samples, etc.
* Correspondence to: L. Mondello, Dipartimento Farmaco-chimico, Facoltà
di Farmacia, Università di Messina, Viale Annunziata, 98168 Messina, Italy. Nowadays the use of temperature-programmed GC is
E-mail: lmondello@unime.it adopted in a wide range of applications. Considering that

Copyright © 2008 John Wiley & Sons, Ltd.


298 B. D’ACAMPORA ZELLNER ET AL.

on a given capillary column stationary phase, the operational solution and dissolution process of the solute molecules
variables, such as carrier gas flow-rate, stationary phase inside and outside the stationary liquid phase. When
film thickness and linear temperature programming, can considering the average gas linear velocity as a constant
alter an analyte’s retention time, the problems of retention throughout an analysis, the solutes should spend an
index reproducibility between laboratories may still identical period of time in the mobile phase to elute from
represent a drawback. In spite of the considerable instru- the column. To this extent, only the differences in the
mental advances made, the acquisition of reliable retention time spent in the stationary phase are responsible for the
index values may still be arguable. When temperature solutes distinct retention time values.
programmes are applied with a constant heating rate The time that any solute spends in the mobile phase,
during the entire analysis, the calculated retention indices that is the unretained peak time (tM), summed with the so-
are reliable. However, when isothermal steps are included called adjusted retention time ( t′R ), which corresponds to the
in the temperature programme or the rate is changed time the solute spends distributed into the stationary phase,
during the analysis, the calculation of the indices is more is equivalent to the retention time (tR) [see equation (1)]:
troublesome.5–7 It is worthy of note that discrepancies
may arise from the fact that the retention index values are tR = tM + t′R (1)
reported by many different laboratories and the operating
conditions are not always reliable. Moreover, it should be The magnitude of retention depends on the partition
pointed out that differences may also arise from the use coefficient, or distribution constant (K), which is defined
of columns coated with identical stationary phase but as the ratio of the equilibrium concentrations of a solute
manufactured by distinct companies. in the stationary (cs) and mobile phases (cm) during parti-
During the 1980s in particular, various aspects of the tioning in the column [refer to equation (2)]:
retention index system were the subject of a great number
cs
of studies and several criteria of retention index systems KD = (2)
were, and are still, being developed to improve their cm
reliability. With time, numerous linear relationships between Hence the greater is the K value for a sample component,
the retention index and other fundamental properties, the higher is its solubility and the longer its retention in
such as boiling point,8 carbon number9 and molecular the stationary phase.
weight10,11 were derived. The prediction of programmed In general, the affinity of a solute for the stationary phase
temperature retention indices is also an in-depth discussed depends on its vapour pressure and the activity coefficient
subject,12 involving the calculation and theoretical deter- of the solute in that phase. Differences between those two
mination of the elution order of many compounds. properties result in the column’s ability to differentiate
Retention index systems have been exhaustively reviewed two solutes, eluting them with different retention times.
by Budahegyi et al.6 and Tarján et al.7 Solute retention can also be expressed in terms of
The present contribution provides an overview of the retention factor, k, which quantifies the ratio of the time
theoretical aspects and applications of retention indices spent in the stationary phase and that spent in the mobile
to gas chromatographic analysis. It is certain that the phase, as presented in equation (3). In practice, the latter
references cited in this review represent a very small part parameter is more utilized in association to column
of the studies performed on retention index theories and efficiency calculations:
their applications. Gas chromatographic retention para-
t′R
meters are briefly described; followed by retention index k= (3)
calculations and systems. The application of indices to the tM
Rohrschneider and McReynolds phase constant determi- A further parameter is the relative retention (r), which
nations and compound identification in flavour and expresses the degree of separation between two peaks,
fragrance analysis are reported. Furthermore, a critical not necessarily in adjacent positions; one represents a
viewpoint is discussed with respect to the indiscriminate standard (st) and the other the solute of interest. Relative
use of retention indices. retention can be expressed using standard solute retention
factors or adjusted retention times. The latter are used in
equation (4), where i refers to an individual solute. The
Principles of Gas Chromatographic relative retention of solutes eluting after the standard will
Retention Parameters be >1, while that of compounds eluting prior to the
standard <1 (see Figure 1):
As is well known, the retention of analytes in gas chro-
t′Ri
matographic capillary columns results from the differential ri = (4)
distribution (partition) of the solutes between the stationary t′R(st )
liquid and the mobile gas phases.13 Solute retention in the Also worthy of mention is the retention volume (VR),
column, and also their resolution, are a consequence of the which can be measured experimentally from the mobile

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj
LINEAR RETENTION INDICES IN GAS CHROMATOGRAPHIC ANALYSIS 299

This system was based on the fact that each analyte is


referenced in terms of its position between the two n-
paraffins that bracket its retention time. Furthermore, the
index calculation is based on a linear interpolation of the
carbon chain length of the two bracketing paraffins. In
the original definition, the retention index of a particular
substance was calculated using only n-paraffins with even
carbon atoms as references [as represented in equation
(7)]:
Isst. ph. (T ) = 200(log Xs − log Xz /
(7)
log X( z +2) − log Xz ) + 100 z
where I is the isothermal retention index at temperature
T, s is the compound of interest, st.ph. is the stationary
Figure 1. Relative retentions of multiple solutes phase, and X is, for instance, the retention time used for
the calculation, while z and z + 2 are n-alkanes with z and
z + 2 carbon numbers, respectively. By definition, the
phase volumetric flow rate (Fc) and the retention time, as retention index of the n-paraffins is equal to 100 times
shown in equation (5): their carbon number for any given stationary phase and at
any given column temperature, e.g. n-C6 has an index of
VR = t R ⋅ FC (5) 600.
At that time, the use of even-number n-paraffins was
However, since tR can be measured directly, its application justified by the possible occurrence of an oscillation in
is preferred over VR. the chromatographic properties of successive numbers of
the complete n-paraffin series. However, this statement
was experimentally confirmed to be erroneous and
Retention Index Theories Kováts redefined the first proposed fundamental equation
to:
A compound’s retention behaviour on a specific column
is characterized by the three parameters presented in the Isst. ph. (T ) = 100[z + (log Xs − log Xz / log X( z +1) − log Xz )]
previous section: retention time, retention factor and (8)
relative retention. In general, the retention time is influ-
enced by the applied linear velocity, temperature, phase where, as for equation (7), z and z + 1 are n-alkanes with
ratio and column length. As a consequence, retention z and z + 1 carbon numbers, respectively. Moreover, the
time is not useful for peak identification. Compared to closer setting of the reference paraffins enables a more
retention time, retention factor presents an advantage due accurate evaluation of a compound’s behaviour in the
to the inclusion of the unretained peak time, contouring uniform scale.
changes caused by linear velocity discrepancies and In general, the retention index system is a combination
column length differences. On the other hand, relative of two chromatographic parameters, relative retention
retention depends also on the phase ratio, so that the [equation (4)] and retention volume [equation (5)]. However,
results are comparable as long as an identical reference several drawbacks of this system could be observed. A
peak is chosen, and the same column temperature is mere comparison between retention indices calculated on
applied. However, the accuracy of this information can be columns of distinct stationary phases does not allow to
degraded for peaks eluting far from the reference peak. distinguish whether the difference in the values is related
In order to overcome this limit, É. Kováts3 introduced to an increased retention of the substance of interest or to
a retention index (I) system in which a homologous a decreased retention of the members of the reference
series of paraffins arranged in straight chains (normal series on the stationary phase in question, or vice versa.14
paraffins) were applied as reference peaks. As is well- In addition, the retention indices, as proposed by Kováts,
known, this homologous series elutes, when isothermal GC refer solely to data obtained under isothermal elution
conditions are applied, with retention times increasing conditions, while by applying a column temperature pro-
exponentially. Under such conditions a semilogarithmic gramme the series of n-paraffins elutes in a linear mode.
relationship exists between the adjusted retention times To each successive peak, a constant increment is added
( t′Ri ) of the n-paraffins and their carbon numbers (cn); a to the retention time of its predecessor, instead of a non-
and b are constants [see equation (6)]: linear increment, as could be observed under isothermal
conditions. A similar relationship was observed between
log t′Ri = a ⋅ cn ⋅ b (6) the programmed-temperature retention times ( t RT ) of the

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj
300 B. D’ACAMPORA ZELLNER ET AL.

indices were determined for six standard compounds


(limonene, linalool, linalyl acetate, acetophenone, naph-
thalene and cinnamyl alcohol) considered as representative
for the substances commonly detected in essential oils,
which were co-injected with even carbon-numbered n-
alkanes from C8 to C24. The g-pack value depends on the
type of stationary phase and, in general, alterations of the
carrier gas flow rate, temperature-programming rate and
initial temperature do not affect g-pack values. On the
other hand, modifications in the selectivity of a stationary
phase, such as those deriving from column bleeding, may
decrease the g-pack value, while the modifications of the
stationary phase due to oxidation may increase a g-pack
value. The latter values were calculated for two stationary
Figure 2. Scheme of the calculation of isothermal
retention indices
phases, a polyethylene glycol and a 100% dimethyl
polysiloxane, representative of polar and non-polar
columns, respectively. With polar columns, a satisfactory
upper temperature was considered to be 225 °C, so that
n-paraffins and their carbon numbers (cn); it can be the elution temperature of n-tetracosane was adjusted
expressed as in equation (9): to 225 °C. However, it was observed that the elution of
homologous series under temperature-programmed con-
log t RT = a′ ⋅ cn ⋅ b′ (9) ditions was not linear, leading van den Dool to suggest
the use of both even and odd carbon-numbered n-alkanes
where a′ and b′ are proportionality constants. to reduce possible errors. The above-mentioned report was
The programmed-temperature retention index calculation published in 1980, followed by a second one released in
is based on the following equation proposed by H. van 1981, also based on the reproducibility of chromatographic
den Dool and D. J. Kratz,4 which does not use the loga- retention data of essential oils analysed by temperature-
rithmic form, as presented in equation (10): programmed GC, on methyl silicone stationary phases.16
Both publications present an appendix comprising methods
I T = 100 [z + (t Ri
T − t T /t T
Rz R( z +1) − t Rz )]
T (10) recommended for the control of the reproducibility of the
system, from the establishment of column performance to
As previously cited, when the latter equation is applied the standardized tabulation of the results.
for the calculation of indices, these are commonly Retention index systems are not only represented by
denominated as linear retention indices (LRIs). Figure 2 the well-known proposals made by Kováts and, later on,
illustrates schematically the retention index calculation by van den Dool and Kratz; the continuous interest and
procedure. Moreover, when series of n-paraffins containing study on retention mechanisms, and also retention index
only even or odd carbon atoms as references are used, systems, led to the introduction of several other concepts,
equation (11) must be applied, where n represents the some of which are briefly reported here.
difference in carbon atom number of the two n-paraffins Several concepts were derived from the equation pro-
that bracket a solute’s retention time: posed by Kováts, such as the molecular retention index,
developed by M. B. Evans and J. F. Smith in 1961,10,11
I T = 100n(t Ri
T
− t Rz
T
/ t RT( z +1) − t Rz
T
) + 100 z (11) which was defined as the molecular weight of a hypothetical
n-alkane presenting the identical retention of the compound
It has to be pointed out that the retention indices calculated being investigated. The formula of that hypothetical n-
for isothermal and temperature-programmed conditions alkane equivalent to the target compound is CIH2I′+2 and
commonly present proximate values, but are not identical. will possess an effective molecular weight according to
Furthermore, there is the possibility that IT values vary the following relation:
according to the applied temperature rate and the initial
temperature. Moreover, since the reproducibility of Me = 14.026 I ′ + 2.016 = 0.1406 I + 2.016 (12)
temperature-programmed retention indices is dependent
on working variables, such as carrier gas flow rate, where I = 100I′, whereas I′ is the number of carbon
stationary phase film thickness and linear temperature atoms and Me is the molecular weight of the hypothetical
programming rate, those parameters should be standardized. n-alkane. The difference between that hypothetical and
In 1974 van den Dool described the g-pack value,15 a the real molecular weight may be a useful parameter to
method for the characterization of GC columns using correlate retention and chemical structure. A further system,
relative retention indices, as proposed by Kováts. The defined as the theoretical nonane value, and developed

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj
LINEAR RETENTION INDICES IN GAS CHROMATOGRAPHIC ANALYSIS 301

by the aforementioned researchers, was propounded, also largely attributed to random errors, led to the develop-
in 1961.17 According to that system, the retention value ment of the unified retention index concept by N. Dimov.24
was first determined as proposed by Kováts and then Considering that the function dI/dT, which represents the
transferred to a system in which n-nonane is the standard. variation of the indices according to the temperature, is a
However, the accuracy of the determination decreased hyperbolic function, the unified retention index (UIT) was
when compounds with retention values far from that of proposed, which may be defined as a statistical treatment
the reference standard were analysed. of data by means of linear regression [see equation (16)]:
Another system based on the equation proposed by
Kováts is the standard retention index developed by P. G. UIT = UI0 + (dUi/dT )T (16)
Robinson and A. L. Odell in 1971.8 In that system the
reference parameter is neither the retention time nor the where UI0 is the value of UIT at 0 °C and dUi/dT is the
adjusted retention time, but the boiling points of the target index increment with the analysis temperature (the slope
analyte and the reference standards. Moreover, the of the curve).
generalized retention index, developed by J. Novák and The index itself and the temperature increment of the
J. Ružiková in 1974,18 also consisted of a modification of unified retention index have the advantage of being
the equation proposed by Kováts, which may be applied statistical values and, thus, more reliable than individual
also when homologous series other than n-alkanes are used experimental retention values. Furthermore, it is charac-
as reference. The identical name, generalized retention terized by a standard deviation (SD), and the calculation
index, was given to a further system created by I. G. of the confidence interval at any desired level is possible.
Zenkevich and B. V. Ioffe in the 1980s19,20 that combines The concept was proposed not only as an easy and con-
the logarithmic system of Kováts with the retention indices venient mode of determining the retention index value at
based on the linear relationship between I and (t′R + qlog t′R ), any temperature within the investigated range, but also as an
where t′R is the adjusted retention time and q a coefficient evaluation method of the proper column temperature for
related to the analytical conditions. the analysis of complex mixtures, demonstrating that the
Based on the fact that the retention indices should be elution sequence of the sample components can change.25
independent of the effects of the stationary phase, V. G. The virtual carbon number was a further proposal studied
Berezkin reported, in 1974, the invariant retention index for the identification of compounds without applying any
(I0), proposing the following relationships between the index calculation, but using solely the retention behaviour.9
retention indices and the invariant retention index:21 According to this system, an homologous series is not
necessary for the calculation of retention indices, since the
I = I0 + a/ PL (13) analysed matrix certainly contains a series of compounds
with well-known retention indices; the latter, which do not
I = I0 + b/kS (14) necessarily belong to a homologous series, could be used
as index references. The index of the other compounds
I = I0 + c/VL (15) shall then be calculated as with n-alkanes, although instead
of the carbon numbers of the n-alkanes, virtual carbon
where a, b and c are constants, PL is the percentage of numbers of the reference compounds should be used.
stationary phase present in the column, VL is the volume A further method for compound identification using
of stationary phase in the column, and kS is the capacity only retention behaviour without index calculation is
factor of the reference substance. retention time locking (RTL) for programmed tempera-
A further system, the homologous index, was intro- ture analysis.26 In the RTL method, the inlet pressure is
duced in 1985 and may be defined as the retention index adjusted to provide an identical retention time for the same
obtained using a homologous series, consisting of com- compound in any system equipped with the same nominal
pounds similar to those to be characterized.22 Rasanen column. This method applies constant pressure and only
et al.23 measured the homologous retention indices of small deviations in column size and stationary phase loading
alkaline drugs through software able to automatically parameters are tolerated; moreover, it cannot be used
assign the so-called secondary retention index standard under constant flow-rate conditions.
series (I*) by a pattern recognition algorithm; the I* The reduced Kováts index (RKI) was also proposed for
standards were then identified as members of a pattern retention data determination of wax esters. According to
rather than individual peaks. Afterwards the same software Stránský et al.27 this index system should get round the
calculated the I* values of other peaks through linear drawbacks arising from the large KI values generated by
interpolation using absolute retention times, compared the multiplying factor of 100. RKIs are calculated based
the obtained values to those contained in libraries and on equation (17), where NCA is the number of carbon
reported the identified compounds. atoms:
Moreover, the investigation on the variation in the
retention index of hydrocarbons on squalene, which are RKI = I − 100 ⋅ NCA (17)

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj
302 B. D’ACAMPORA ZELLNER ET AL.

These values should express the influence of the presence halides, alkyl acetates, alkanoic acid methyl esters13 and
and position of the ester functionality or position of para-substituted n-alkylphenols.32 These compounds are
double bond on the stationary phase. However, it has to less likely to be retained by interfacial adsorption on polar
be pointed out that, although used for temperature- stationary phases than the n-alkanes, resulting in more
programmed GC analyses, retention index calculations reliable retention index values, and so more suitable
were logarithmic-based. qualitative identification. On the other hand, Shibamoto33
A further index system, the molecular topological suggested the use of polar series, such as ethyl esters, as
index, has been shown to be a very important structural an alternative. In the characterization of volatiles, the most
parameter for describing the chromatographic behaviour commonly applied reference series are the n-alkanes and
of a compound.28 The determination of these indices and the fatty acid ethyl esters (FAEEs), which are chosen
their correlation with retention indices of linear alkylben- according to the stationary phase used.
zene isomers with C10 to C14 linear alkyl chains were Standard series may also be selected according to the
investigated.29 characteristics of the matrix, so that the chosen standards
Moreover, there has also been a great deal of interest in are normally comprised in the sample under investigation,
the prediction of retention indices. The development of or present a chemical structure and a behaviour with the
predictive relationships would represent a key for the chromatographic system, similar to the compounds to be
reliability of retention data, the prediction of the retention determined. Several series have already been reported in
index of an analyte that was unavailable for study, as also the literature, such as the linear fatty acid methyl esters
the extrapolation of values of an analyte’s specific physical (FAMEs), which provide an equivalent chain length of
properties that could be correlated through its retention unsaturated and branched fatty acid esters in lipid
index. More general approaches to retention index predic- analysis,34,35 or polychlorobiphenyl (PCB) congeners
tion are based on the generation of topological, geometric possessing distinct numbers of chlorine atoms to identify
and electronic molecular descriptors, which are subse- the highly toxic PCB isomers.14 A further example is the
quently fitted to the retention index using multiple linear polynuclear aromatic hydrocarbons (PAH) retention index
regression or artificial neural networks. Developments in system, which uses the numbers of rings in unsubstituted
artificial intelligence were represented by Bruchmann, PAHs to characterize the retention behaviour of their
Zinn and Haffer30 with neural networks which were homologues and related substituted compounds. The
explored to predict GC retention index data based on determination of over 200 PAH structures by using naph-
electrotopographic indexes of monofunctional compounds, thalene, phenanthrene, chrysene and picene as reference
such as acyclic and cyclic monoterpenes and a mixed set standards has been reported by Lee et al.36
of monosubstituted compounds and terpenes. According Further homologous series are related to drug screening gas
to the authors, predictions by neural networks are gener- chromatographic analysis, such as the 1,4-benzodiazepine
ally in good agreement with predictions done by multiple index standard series for benzodiazepine screening37 and the
linear regression techniques. In addition, the prediction four series FA (N,N-dialkyl-4-fluoroanilines), FB (N,N-dialkyl-
of retention indices of homologous series was proved to 4-fluorobenzylamines), BA (N,N-dialkyl-4-benzylamines)
be effective. and TA (trialkylamines), applied in the screening of
alkaline drugs in blood samples.38 A similar approach was
also performed using MH (3-alkyl-1-methylhydantoins)
Alternatives to n-Alkanes as Reference and H (3-alkylhydantoins) series for the screening of
Series nitrogenous acidic and neutral drugs.23
The introduction into the market of element-specific
In general, retention index systems are based on the detectors is also a consistent reason for the selection
incremental structure–retention relationship, viz. that any of most suitable standard series, as is the case of the
regular increment in a series of chemical structures should electron-capture detector (ECD) and the flame photometric
provide a regular increment in corresponding retention detector (FPD), which are not sensitive to n-alkanes.13,31
times. This means that the retention index concept is not Homologous series of iodoalkanes, chloroalkanes,
restricted to the use of n-alkanes as standards, as is well bromoalkanes, n-alkanols or, as mentioned above, PCB
reported in the exhaustive review written by Castello.31 congeners can be exploited when an ECD is used;36 a
In practice, any homologous series presenting a linear standard series of sulphur-containing molecules, such as
relationship between the logarithm of the adjusted reten- dialkylsulphides, can be used with a FPD, and trialkylamines
tion time and the carbon number may be used. with a nitrogen phosphorus detector (NPD).39 Further
In general, n-alkanes present fluctuating behaviour on series, such as alkylbis(trifluoromethyl)phosphine sulfites
polar stationary phases. In consideration of the fact that (M series), are employed for several detectors – ECD,
retention index values are correlated to retention mecha- FID, FPD, alkali flame ionization detector (AFID),
nisms, alternative standard series of intermediate polarity photoionization detector (PID)40 and mass spectrometry
were introduced, such as 2-alkanones, alkyl ethers, alkyl (MS).41

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj
LINEAR RETENTION INDICES IN GAS CHROMATOGRAPHIC ANALYSIS 303

Application of Retention Index Values for Retention Index Compilations


Phase Constant Determinations
A myriad of retention index databases have been published
The selectivity of a stationary phase can be estimated by and, since the majority of such retention index data have
determining the degree to which polar compounds are not been generated under conditions in which stationary
retained by that phase in relation to their retention on a film polarity and thickness, carrier gas flow rate and
non-polar phase. The system of stationary phase constants temperature-programming rate have been standardized,
based on retention index differences (ΔI), introduced by their precision, comparability, reliability and usefulness
Rohrschneider,42 was the first widely adopted approach may be limited.
for the systematic classification of stationary phases. The Worthy of note, among the several retention index
system of phase constants is formulated on the assumption datasets, are the compilations also published as books and
that intermolecular forces are additive, and provide their widely used as references, as those authored by Jennings
individual contributions on retention. The relative ΔI and Shibamoto44 and by Adams45 (the 4th edition was
values are obtained by measuring the degree to which published in 2007). The latter book includes mass spectra,
the stationary phase to be characterized retains several chemical structures and retention indices of 2205 compounds
probe compounds (benzene, ethanol, methyl ethyl ketone, obtained on a 5% diphenyl:95% dimethyl polysiloxane
nitromethane and pyridine), relative to their retention on column (30 m × 0.25 mm i.d., 0.25 μm film thickness).
a non-polar hydrocarbon phase (squalane), both at 120 °C The use of KI and AI made by Adams is noteworthy, with
[equation (18)]: AI standing for arithmetic index, the calculation of which
is based on the equation proposed by van den Dool and
ΔI = I polarphase − Isqualane (18) Kratz.4 The former book reports the spectra of over 1150
flavour and fragrance volatiles and their respective KIs
where ΔI represents the retention index difference; the on methyl silicone and on polyethylene glycol stationary
latter is equivalent to the contribution from polar interactions. phases. Another retention index dataset is the Sadtler
The Rohrschneider constant is commonly represented by Standard Gas Chromatography Retention Index Library.46
the ratio of the index difference and 100: The latter comprises a four-volume reference set that
identifies over 2000 compounds by their retention indices
ΔI/100 = Rohrschneider constant (19) on three different stationary phases. Further books are
also available, such as the compilation of sesquiterpene
In order to evaluate the phase constants numerically, the hydrocarbons authored by Joulain and König47, compris-
McReynolds constants,43 derived from the Rohrschneider ing the mass spectra of 307 compounds along with their
method, are commonly used. Those constants enable to retention indices determined on a 25 m fused silica capil-
compare polarities of stationary phases based on their lary column coated with 100% dimethyl polysiloxane. A
selectivity for specific solute interactions; for instance, further interesting issue is the toxicological database
benzene has a constant of zero for squalane, while on a published by Maurer et al.,48 containing 7840 datasets of
100% dimethyl polysiloxane stationary phase the con- drugs, poisons, pesticides, pollutants and their metabolites.
stant equals 0.16, and on a polyethylene glycol column it Notwithstanding these publications, a considerable
has a value of 3.22. The larger the value of the McReynolds’ problem may be still observed in the inter-laboratory
phase constant, the stronger is the interaction expressed comparison of retention index values, since the applied
by that constant. The McReynolds values are used to chromatographic conditions are frequently not strictly
evaluate the degree of similarity between stationary phases, comparable. Richmond49 made an attempt to circumvent
to rank the phases by degree of polarity and to predict the this problem by merging data contained in different
elution order of the analytes. datasets using a so-called GC calibration approach,50
The Rohrschneider concept and related calculation which is just empirical and has no thermodynamic or
methods represent a very important research area,6,7,13 and quantitative structure–retention relationship adjustment.
both the aforementioned approaches are widely applied The merging of compilations led to the creation of the
to characterize and classify stationary phases. However, Pseudo-Sadtler database, which comprises data obtained
the McReynolds system may often contain erroneously from Sadtler46 and Jennings and Shibamoto.44 One successful
measured data; an additional drawback is that the constants example of the calibration approach is 2-octene, whose
are determined only at 120 °C. The combination of the retention index of 811, as reported by Jennings and
Rohrschneider and McReynolds systems supplemented Shibamoto, was equivocal due to ambiguous cis–trans
with data measured at 80 °C and 140 °C would ensure a geometry. The cis and trans isomers are separable and
greater development of this research area. The field of present retention indices of 803 and 811, respectively.51
structure-driven prediction of retention in gas chromato- Compilations of compounds and their respective retention
graphy is not well developed at present and new approaches indices can be widely seen in several publications, e.g.
should be faced. Lee et al.36 published the retention indices of over 200

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj
304 B. D’ACAMPORA ZELLNER ET AL.

Table 1. Some flavour and fragrance compounds whose spectra and linear retention indices are included in the
FFNSC ver. 1.3 MS Library (Chromaleont, Messina, Italy)

LRIa Common name CAS name CAS No. MW Formula

608 Vinyl dimethyl carbinol 3-Buten-2-ol, 2-methyl- 115-18-4 86 C5H10O


864 α-Angelicalactone 2(3H)-Furanone, 5-methyl- 591-12-8 98 C5H6O2
972 Pyromucic acid methyl ester 2-Furancarboxylic acid, methyl ester 611-13-2 126 C6H6O3
1030 Limonene Cyclohexene, 1-methyl-4-(1-methylethenyl)- 138-86-3 136 C10H16
1076 n-Octanol 1-Octanol 111-87-5 130 C8H18O
1101 Linalool 1,6-Octadien-3-ol, 3,7-dimethyl- 78-70-6 154 C10H18O
1115 Sotolone 2(5H)-Furanone, 3-hydroxy-4,5-dimethyl- 28664-35-9 128 C6H8O3
1152 Nona-(2E,6E)-dienal 2,6-Nonadienal, (2E,6E)- 17587-33-6 138 C9H14O
1153 Nona-(2E,6Z)-dienal 2,6-Nonadienal, (2E,6Z)- 557-48-2 138 C9H14O
1163 Furfurylthiol acetate Hexanoic acid, 2-furanylmethyl ester 13678-68-7 156 C4H8O2S
1180 Abhexone 2(5H)-Furanone, 5-ethyl-3-hydroxy-4-methyl- 698-10-2 142 C7H10O3
1200 Florosa 2H-Pyran-4-ol, tetrahydro-4-methyl-2-(2-methylpropyl)- 63500-71-0 172 C10H20O2
1250 Linalyl acetate 1,6-Octadien-3-ol, 3,7-dimethyl-, acetate 115-95-7 196 C12H20O2
1250 Thymoquinone 2,5-Cyclohexadiene-1,4-dione, 2-methyl-5-(1-methylethyl)- 490-91-5 164 C10H12O2
1368 4-tert-Butyl-cyclohexanol acetate Cyclohexanol, 4-(1,1-dimethylethyl)-, acetate 32210-23-4 198 C12H22O2
1375 α-Copaene Tricyclo[4.4.0.02,7]dec-3-ene, 1,3-dimethyl-8-(1-methylethyl)- 138874-68-7 204 C15H24
1651 α-Muurolol 1-Naphthalenol, 1,2,3,4,4a,7,8,8a-octahydro-1,6-dimethyl-4- 19435-97-3 222 C15H26O
(1-methylethyl)-, (1R,4S,4aR, 8aS)-
1651 Vanillyl butyl ether Phenol, 4-(butoxymethyl)-2-methoxy- 82654-98-6 210 C12H18O3
1767 Amberlyn Naphtho[2,1-b]furan, dodecahydro-3a,6,6,9a-tetramethyl- 3738-00-9 236 C16H28O
1960 Neocembrene 1,5,9-Cyclotetradecatriene, 1,5,9-trimethyl-12- 31570-39-5 272 C20H32
(1-methylethenyl)-, (1E,5E,9E,12R)-

a
Linear retention index calculated on a 5% diphenyl:95% dimethyl polysiloxane column (30 m × 0.25 mm i.d., 0.25 μm df) programmed from 50 °C to 250 °C
(10 min) at 3.0 °C/min.

PAHs, and Stránský et al.27 reported Kováts indices of 205 This database contains about 3000 molecules including
wax ester standards. With regards to flavour and fragrance their Kováts retention index on distinct stationary phases
materials, Davies52 published an extended compilation extracted from publications in that research field.
reporting a summary of retention index values obtained on With regard to GC–MS databases, several mass spectral
methyl silicone and on Carbowax 20M stationary phases, library software are available on the market; however,
and Bianchi et al.53 presented the retention indices of 250 only a few of them comprise retention index information
food volatile chemicals on a polar stationary phase. A to facilitate peak assignment;57,58 this is the case of the
further collection of n-alkanes and ethyl ester retention FFNSC MS Library ver. 1.3 (Chromaleont, Messina, Italy),
indices (IE) of 183 common constituents of essential oils the MS library developed by Adams (Allured Publishing,
was published by Shibamoto.33 Carol Stream, IL, USA), and the NIST 05—NIST/EPA/
The World Wide Web is also a widely used source of NIH Mass Spectral Library (National Institute of Standards
retention index values; however, their reliability may be and Technology, Gaithersburg, MD, USA). With respect
doubtful. Among the trustworthy and well-accepted to the FFNSC MS Library, this commercially available
references are the Flavornet54 and the LRI & Odour product contains 1827 compounds. For the sake of
Database.55 The latter contains the retention indices of over demonstration, Table 1 lists 20 compounds whose spectra
5000 volatile compounds identified in different food and linear retention indices are contained in the FFNSC
matrices. The database is a compilation of LRI data, using MS Library ver. 1.3. The spectra were acquired on a
n-alkanes as reference series, and extracted from the GCMS-QP2010 system (Shimadzu, Milan, Italy) equipped
scientific literature on food flavour. The former database with a 5% diphenyl:95% dimethyl polysiloxane column
comprises odour-active compounds detected by means of (30 m × 0.25 mm i.d., 0.25 μm df; SLB-5MS, Supelco,
GC–olfactometry (GC–O) applying quantitative odour Bellefonte, PA, USA). The oven temperature was pro-
measurement techniques. The compilation reports the grammed from 50 °C to 250 °C at 3.0 °C/min and held for
retention indices of 738 compounds, calculated by applying 10 min. This library, dedicated to Shimadzu GC–MS sys-
the equation proposed by Kováts using n-paraffins and tems, possesses not only information on retention indices
ethyl esters as reference series on four different stationary but also is equipped with retention index interactive tools,
phases. The first-mentioned database also reports the whose application is enabled when used in association with
compound’s odour description, while the second only the data-collecting software GCMS Solution (Shimadzu,
reports this for about 1500 compounds. A further reference Milan, Italy),59 since the latter software possesses an
is the Pherobase56 developed by the Chemical Ecology automated retention index calculation function. To per-
Group at the Hort Research in Lincoln, New Zealand. form the automatic retention index calculation, a series of

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj
LINEAR RETENTION INDICES IN GAS CHROMATOGRAPHIC ANALYSIS 305

Figure 3. GC–MS chromatogram of the Cistus ladanifer L. essential oil with indicated linalool peak (A), qualitative
parameters window of the GCMS Solution software, illustrating the retention index allowance and column type
selection tools (B), and the similarity search result window attained for this peak (C)

homologous standards, commonly n-alkanes, has to be as illustrated in Figure 3B, a column type selection tool is
injected prior to analysis, and the retention times, as well available, so that only retention index values determined
as retention indices, obtained for each standard inserted on that stationary phase are displayed in the similarity
in the index table of standards. Based on this information search result window (Figure 3C).
set, the software is able to calculate the retention index Also of great interest is the AMDIS (Automated Mass
for the detected compounds. The library search algorithms Spectral Deconvolution and Identification System) program,
incorporating LRIs as part of the match criteria assist in frequently used in combination with NIST Mass Spectral
the analyses and characterization of flavour and fragrance Libraries (both developed by the National Institute of
compounds, enhancing the credibility of the identification Standards and Technology). This software not only
through MS. The GCMS Solution software also permits deconvolutes GC–MS files but also enables the use of
the insertion of five distinct retention index values for each mass spectra libraries containing retention index information,
compound whose mass spectra was appended to a mass which were previously converted to the AMDIS format.61
spectral library. Figure 3 illustrates the library search of a A retention index-matching window is also available as
component of Cistus ladanifer L. essential oil.60 Moreover, an optional setting of the software. An example of the

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj
306 B. D’ACAMPORA ZELLNER ET AL.

Figure 4. AMDIS analysis of a GC–MS chromatogram of Cistus ladanifer L. essential oil detecting linalool (calcu-
lated RI of 1099.4 and RI-RI(lib) of –1.6)

use of AMDIS on a Cistus ladanifer L. oil analysis however, this analytical task was made troublesome
applying the FFNSC MS Library ver. 1.3 converted to due to the complexity of those real-world samples. In this
AMDIS format is presented in Figure 4. respect, the utilization of retention indices, when com-
It is certain that the above-cited retention index compi- bined with other characterization methods, such as mass
lations represent a very small part of the studies performed spectrometry, was shown to be an effective tool for the
on flavour and fragrance materials. identification of flavour and fragrance compounds.
In general, apart from the GC hardware configuration
applied, the retention index determination may be per-
Retention Indices and Chromatographic formed through the injection of the homologous series as
Systems internal or external standard. When solutions of standard
compounds or simple extracts are analysed, the use of an
As is well-known, flavour and fragrance materials are internal standard series is a valid approach, since both
characterized by a wide diversity of components belonging analytes and references are subjected to identical factors
to several chemical classes, and are commonly added to influencing retention time measurement. However, when
a great variety of products, such as foods (ice creams, complex samples have to be analysed, the use of an internal
alcoholic and non-alcoholic beverages, baked goods, etc.), standard series is not feasible. The work developed by
cosmetics, toiletries, functional products, tobacco products A. Tóth and L. Praszna63 on retention index reproducibility
and pharmaceutical preparations among others.62 The reports values obtained for analyses carried out during a
advent of GC marked a real turning point for flavour and period of 2 months. These researchers observed that at
fragrance research. In the early stages of research in this isothermal temperatures the changes in retention times
field, attention was devoted to the development of methods over the analysis period may be negligible, but in general
to acquire deeper knowledge on the profiles of the volatiles; the measured retention times decreased according to the

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj
LINEAR RETENTION INDICES IN GAS CHROMATOGRAPHIC ANALYSIS 307

use of the column. On the other hand, when applying introduction of electronic pressure control devices and
programmed temperatures, retention data presented a improved column oven heat distribution in GC systems.
deviation, which was attributed to the programming rate The use of autosamplers and autoinjectors also enhanced
or carrier gas velocity. To overcome these limits, one of retention time precision and, in consequence, the reliability
the n-alkanes contained in the series of standard alkanes of calculated retention index values increased. Furthermore,
was used as internal standard. The difference of the the improved quality of stationary phases has also played
retention times of that n-alkane analysed as internal and a role in the increased column stability and long-term
as external standards was then applied as a correction reproducibility of analysis. Additionally, according to the
factor, normalizing the retention times that were used in the previously mentioned report published by the Analytical
calculation of the so-called corrected retention indices. Methods Committee,16 it is desirable to use two columns
Concerning the stationary phases, in the case of signifi- with different stationary phases to attain two retention
cant interfacial adsorption of solutes or standards, wide index values and enhance confidence in assignments.16,52
variations in retention index values can occur. Those It is accepted that the reproducibility of retention indices
variations may be related to sample size effects, in particular between different laboratories is usually within a few index
the concentration ratio of the analysed substance to the units for non-polar stationary phases and within 10 –50
n-alkane, and with the phase loading or the column’s film index units for polar phases.13 The large range of units
thickness. A remark need to be made on the influence of for polar columns could be related to the heterogeneity
the stationary phase film thickness (df); index values between columns manufactured by different companies, the
calculated on a thinner film tend to be higher than those misuse of the gas chromatographic technique, a possible
measured for a thicker one, since the residual surface reactivity between the analyte and the stationary phase,
activity will be shielded by the thicker film. Moreover, as well as the amount of sample injected, which can induce
the dependence of the index is hyperbolic, so that the deviation. A further aspect of great importance, which is
application of linear extrapolation may lead to unreliable frequently overlooked, is the analytical reproducibility;
retention values.64 The fact that retention index values are most publications do not report data on analysis performed
specific to a stationary phase at a determined temperature in duplicate or triplicate. Moreover, some publications
has also to be considered.65 report the use of an initial isothermal hold in the GC
The LRI reliability in identifying an essential oil oven temperature programme, even though in practice it
component at an intra-laboratory level was investigated was found not to give any additional resolution, as reported
in depth by Bicchi et al.66 They analysed three essential in an investigation on the monoterpene hydrocarbon
oils with different columns [polysiloxane (OV-1) and contents of essential oils.5
polyethylene glycol (CW 20M)] that had distinct charac- Publications on flavour and fragrance materials which
teristics (producer and inner diameter) and under differ- report calculated retention indices and make reference to the
ent analytical conditions (mainly temperature programme works published by Kováts3 and van den Dool and Kratz 4
and flow rate). The data obtained were used to evaluate are countless. However, publications that inappropriately
the influence of analytical conditions on retention index use the retention indices are also innumerable, e.g. apply-
variation, showing that retention index repeatability with ing the equation proposed by Kováts in temperature-
CW 20M was mainly conditioned by the temperature programmed GC analysis. Although there is general and
programme. constant interest in the application of retention indices in
The application of two sufficiently different temperature GC, their standardization is still a target to be achieved.
gradients may cause the alteration in the elution of an To better illustrate the divergences observed in the literature
analyte on identical polar stationary phases, e.g. the and investigate the reliability of retention indices, the
application of an oven temperature rate of 2 °C/min or values reported in the Flavour and Fragrance Journal
10 °C/min may alter the calculated retention index value. In publications of the year 2006, for the monoterpene limonene
this respect, Hérent et al. proposed the double indexation and the oxygenated monoterpene linalool on polar
method,67 which consisted of the determination of two (polyethylene glycol) and non-polar (5% diphenyl, 95%
retention index values for an identical compound analysed dimethyl polysiloxane) stationary phases, were collected.
with two distinct oven temperature rates, viz. 6 °C/min and Tables 2 and 3 present an overview of the retention index
10 °C/min, on the same polar column. FAMEs from C5 to values of limonene and linalool, respectively, measured
C20 were chosen as reference series, due to the previously using n-alkanes series throughout 2006. During that year,
mentioned low solubility of n-alkanes in several polar the retention index of limonene was reported in 31 and
stationary phases. According to the authors, the method 55 articles, on polar and non-polar phases, respectively,
presented increased the reliability of identification and while that of linalool was in 28 and 64 articles. The dis-
might be very useful for a rapid identification of major crepancies between the values demonstrate that published
constituents of essential oils. retention indices, when originated from poor analytical
Noteworthy is the increased reliability of retention procedures, might not always be used as complementary
times obtained under programmed temperature since the information for the reliable identification of unknown

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj
308 B. D’ACAMPORA ZELLNER ET AL.

Table 2. Retention index values of limonene and their frequency of citation in


Flavour and Fragrance Journal published throughout the year 2006

Polar stationary phase Non-polar stationary phase

LRIa Frequency of citation LRIb Frequency of citation

Limonene 1183 3 1010 1


1190 2 1013 1
1195 1 1022 5
1199 2 1023 1
1201 1 1025 3
1202 3 1026 2
1203 10 1029 3
1204 4 1030 10
1205 5 1031 17
1032 8
1033 1
1035 1
1038 2

a
Linear retention index obtained on polyethylene glycol column.
b
Linear retention index obtained on 5% diphenyl, 95% dimethyl polysiloxane column.

Table 3. Retention index values of linalool and their frequency of citation in


Flavour and Fragrance Journal published throughout the year 2006

Polar stationary phase Non-polar stationary phase

LRIa Frequency of citation LRIb Frequency of citation

Linalool 1509 2 1079 1


1521 4 1088 1
1522 1 1092 2
1539 1 1096 1
1542 3 1097 5
1543 2 1098 22
1544 3 1099 9
1546 1 1100 7
1547 2 1101 4
1549 1 1102 1
1551 2 1103 10
1553 4 1108 1
1556 2

a
Linear retention index obtained on polyethylene glycol column.
b
Linear retention index obtained on 5% diphenyl, 95% dimethyl polysiloxane column.

compounds. As previously mentioned, by convention, peak assignment. The application of the retention index
calculated values which are comprised within a determined system has been extended to other gas chromatographic
range may be considered as acceptable, i.e. ±5 units for approaches, such as olfactometry. The latter technique
methyl silicone stationary phases and ±10 units for was a breakthrough in analytical aroma research, enabling
polyethylene glycol phases. the differentiation of a multitude of volatiles, previously
separated by GC, into odour-active and non-odour-active
compounds, related to their concentrations existing in the
Specific Applications of Retention Indices matrix under investigation. Especially when working with
GC–O systems equipped with detectors that do not provide
Usefulness of Retention Indices in GC and structural information, a compound’s retention index is
GC–MS Analyses then commonly associated to its odour description, sup-
porting reliable peak assignment.
Considering gas chromatographic analyses using flame Moreover, with the growing popularity of GC coupled
ionization (FID), thermal conductivity (TCD) or other to mass spectrometry (GC–MS), a large number of studies
detectors which do not provide structural information of has exploited mass spectral information for peak identifi-
the analysed molecules, retention times, and more precisely cation; the most frequent and simple identification method
retention indices, are used as the primary criterion for consisting in the comparison of the acquired unknown

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj
LINEAR RETENTION INDICES IN GAS CHROMATOGRAPHIC ANALYSIS 309

Figure 5. Representation of the similarity between mass spectra of monoterpenes: (A) α-thujene (A) and α-
phellandrene (B), and sesquiterpenes: (E)-β-farnesene (C) and (E)-β-bergamotene (D)

mass spectra with those contained in a reference MS considered as the most structurally varied class of natural
library. However, as is well-known, isomers or similar plant products, derived by repetitive fusion of branched
substances, when analysed by means of GC–MS, may five-carbon units based on the isopentane skeleton
have their identity imprecisely or unreliably assigned, a (isoprene units).68 As the mass spectra of these compounds
drawback which is often observed in essential oil analyses. are usually very similar, peak assignment becomes a
As is widely acknowledged, the composition of essential difficult and sometimes impracticable task. In Figure 5A
oils is mainly represented by terpenes. The latter can be and B the mass spectra similarity of monoterpenes is

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj
310 B. D’ACAMPORA ZELLNER ET AL.

represented through α-thujene and α-phellandrene. For the sesquiterpene is usually sufficient to ensure its identification,
former compound, a retention index of 927 is commonly provided the reference data have been recorded using
obtained on 5% diphenyl stationary phase, while for the authentic samples. Indeed, for this class of compounds
latter the retention index is 1007. The same can be there would be no need to adopt a polar stationary phase,
observed for the sesquiterpenes (E)-β-farnesene and (E)- which could even lead to misinterpretations caused by
β-bergamotene, whose mass spectra are showed in Figure possible changes in the retention behaviour of sesquiter-
5C and D, respectively, and whose retention indices of pene hydrocarbons as a result of column ageing or deteri-
1452 and 1483 are obtained on a 5% diphenyl stationary oration, and depending on the manufacturer. Moreover,
phase. In order to increase the reliability of the analytical in agreement with the authors’ experience, attention should
results and to address the qualitative determination of be paid to the retention index and the mass spectrum
compositions of complex samples by GC–MS, retention registration of each individual sesquiterpene, since many
indices can be an effective tool. The use of retention indices compounds with rather similar mass spectra may be eluted
in conjunction with the structural information provided in a narrow range; more than 160 compounds can be
by GC–MS is widely accepted and routinely used to con- eluted within 100 KI units on a non-polar column, e.g.
firm the identity of compounds. 1400 –1500.
It is worthwhile to mention the series of collaborative In addition, it has to be brought to light that in 1997,
trials performed by the Working Group on Methods of the Analytical Methods Committee reported a method
Analysis of the International Organization of the Flavor which has been developed from a collaborative study
Industry (IOFI), aiming to investigate the reproducibility performed by 16 laboratories that applied temperature-
of retention indices of some flavour compounds and programmed GC using capillary columns coated with
to establish quality control parameters.69 As is widely non-polar stationary phases.70 In this study the previously
acknowledged, most of the methods applied to identify described g-pack value, proposed by van den Dool15 for
components of flavour and fragrance samples rely on packed columns, has been specified for each investigated
chromatographic techniques, although the use of an capillary column, and reproducible relative retention
additional confirmatory method is of extreme importance. indices have been obtained for the six standard com-
The trials consisted of the study of 21 compounds com- pounds (limonene, linalool, linalyl acetate, acetophenone,
monly detected in flavours or essential oils, or whose naphthalene and cinnamyl alcohol). The study consisted
concentration in food products has been limited by of the analysis of the standards mixture, an even carbon-
regulatory legislation. These compounds were analysed numbered n-alkanes solution (C8–C24), a Spanish rosemary
on polar and non-polar stationary phases, under identical oil sample, and the latter oil co-injected with the afore-
operating parameters, using n-alkanes as reference series mentioned n-alkanes solution, on different 100% dimethyl
for the retention index values calculation. For all com- polysiloxane columns with dimensions in the ranges
pounds investigated the relative standard deviation (RSD) 25 – 60 m length, 0.20 –0.32 mm i.d. and 0.25–1.05 μm
was <1%, except for coumarins, which presented an stationary phase film thickness, produced by different
RSD of 1.3% on the polar column. In general, the RSD manufacturers. In all analyses the oven temperature pro-
values varied from 0.2% to 1.3% and 0.2% to 0.9% for gramme was set from 50 °C to 250 °C at a rate of 4 °C/min,
polar and non-polar columns, respectively. According to and the carrier gas control could be either constant flow
the IOFI, the use of retention index data obtained in two or pressure. Initial results demonstrated that there were
analytical columns coated with stationary phases of distinct differences between different column manufacturers, but
polarities is highly suggested, and identifications made similar results were obtained between laboratories using
on a single column may only be accepted if used in com- similar columns purchased from the same manufacturer.
bination with mass spectral information. Furthermore, it The recommended method should be applied to columns
is outlined that the most accurate identifications are made in the ranges 25–30 m length, 0.22–0.32 mm i.d. and
by using authentic standard compounds, and that neither 0.25 μm film thickness, using the identical above-described
GC nor MS are satisfactory when applied as stand-alone oven temperature programme. The standards mixture is
techniques. In some cases, even when using GC–MS, a co-injected with the n-alkanes solution, and the retention
further method may still be required to satisfactorily indices and the g-pack values are calculated. After the
characterize a compound, such as nuclear magnetic reso- column’s performance within a chromatographic system
nance (NMR) or infra-red (IR) spectroscopy. has been satisfactorily established, it can be used under
The above-mentioned considerations may be exemplified identical conditions to analyse the essential oil, and the
by the identification of sesquiterpene hydrocarbons, as latter co-injected with the aforementioned n-alkanes solution.
highlighted by Joulain and König.47 As can be widely The two chromatograms obtained are compared to check
observed, a few of the large number of published works the reproducibility of the system, as also to confirm whether
report a reliable identification of known sesquiterpenes. there are interferences due to co-elutions of an essential
It could be observed that when associated with its non- oil’s component with an n-alkane peak. It is worthy of
polar GC retention index, the mass spectrum of a given note that, according to the committee, the operating

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj
LINEAR RETENTION INDICES IN GAS CHROMATOGRAPHIC ANALYSIS 311

conditions and the chromatographic system may be con- homologous series to be applied as references; the use
sidered as satisfactory when the g-pack value is 1.0 ± 0.005. of FAMEs and FAEEs series is advisable rather than
Moreover, a much wider variation in elution tempera- n-alkanes.74 –76
tures was observed on capillary columns than on packed
ones.
Retention Indices Applied to Multidimensional
Gas Chromatographic Analysis
Gas Chromatographic Enantiomer
Characterization Supported by Retention Indices As is well known, flavour and fragrance matrices are
complex multi-component mixtures with many overlapping
Enantioselective GC (Es-GC) became an essential tool peaks in the gas chromatograms. The determination of
for stereochemical analysis mainly after the introduction the structure of a component directly from such mixtures
of cyclodextrin (CD) derivatives as chiral stationary is difficult, if not impossible. The analysis is then more
phases about 25 years ago.71,72 The main application of reliable when multidimensional gas chromatography
Es-GC to flavour and fragrance materials concerns the (MDGC) is applied.
precise determination of enantiomeric compositions and The performance of MDGC analysis enables higher
ratios, which characterize the matrix, determining its quality resolution and better characterization of the components
and, in the case of natural materials, its geographic origin, of flavour and fragrance matrices. Briefly, in MDGC key
biogenesis, authenticity and extraction technique applied.73 fractions of a sample are selected from the first column
The calculation of linear retention indices of optically and re-injected onto the second one, where, ideally, they
active compounds can be considered troublesome due to should be fully resolved. Two chromatographic columns
complex retention mechanisms on chiral stationary phases of differing polarities, but generally of identical dimen-
(CSPs). In this case the latter are made up of CD derivatives sions, are employed. Furthermore, when the heart-cut
and the mechanism of retention is based on the host–guest operations are not carried out, the primary column elutes
interaction, and if a homologous series, such as the n- normally in the first dimension GC system, while the pre-
alkanes, is used, the hydrocarbons randomly occupy positions viously re-injected fractions are analysed in the second
in the chiral cavity. As a consequence, n-alkanes can be dimension GC system, equipped with a detector.77
considered as unsuitable for retention index determina- A heart-cut MDGC approach based on retention index
tions on CSPs. Nevertheless, other reference series can theory was demonstrated for the analysis of FAMEs
be utilized on CD stationary phases, such as linear chain mixtures.78 The reported method used predetermined
FAMEs and FAEEs. However, it can be affirmed that retention index data of the target compound to estimate
retention indices are seldom reported for optically active the retention time range in which this compound would
compounds, and publications refer to retention times elute and so precisely define the linear retention window
rather than indices. for the heart-cut to be set. In this experiment the equation
For many years, GC–MS has been the benchmark tech- proposed by van den Dool and Kratz was applied in the
nique for the qualitative analysis of flavour and fragrance following mode:
volatiles. However, when considering Es-GC–MS, the
identification of a target compound can be hindered due I ( x ) − Iref 2
t R ( x ) = t R,ref 2 + (t R,ref 2 − t R,ref 1) ⋅ (20)
to the similar fragmentation pattern of enantiomers. The Iref 2 − Iref 1
automation of library search algorithms incorporating
linear retention indices as part of the match criteria, A similar concept was adopted for the screening of
briefly described in the section on Retention Index allergens in flavour and fragrance materials.79 The proposed
Compilations above, can also be extended to enantiose- method consisted, firstly, of the determination of the linear
lective GC–MS analysis, assisting in the discrimination of retention index of each allergen in the first dimension
enantiomers.74–76 (5% diphenyl stationary phase) of the MDGC system in
An enantioselective MS library is characterized by two stand-by mode. Secondly, cut windows were derived from
features, the mass spectrum, which could be useful to the application of ±10 LRI units with respect to the
determine the position of enantiomers in a chromatogram, experimental LRI value of each compound. The ±10 LRI
and the retention index, which can be considered as a unit values were applied in equation (20) and in heart-cut
tool to discriminate these isomers. Moreover, considering time windows for the target compounds were established.
that a universal chiral selector with widespread capability Through such an approach, the injection of an allergen
for enantiomer separation is not available, and thus effective solution prior to each MDGC application would not be
optical separation of all chiral compounds present in a required; it would only be necessary to derive the LRIs
matrix may be unachievable on a single chiral column, the through preliminary analysis with n-hydrocarbons. The
use of a set of columns coated with distinct CD derivatives allergens standard solution would have to be injected
as chiral selectors is necessary. The same is true for the only periodically for calibration purposes.

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj
312 B. D’ACAMPORA ZELLNER ET AL.

A further MDGC technique is comprehensive two- composition. Furthermore, ordered structures were observed
dimensional gas chromatography (GC × GC), which within a 2D contour plot; a monoterpene hydrocarbons
promotes an orthogonal two-column separation, with region could be observed, as was also a similar region at
complete sample transfer executed by means of a modulator; higher 1tR (1D retention time) corresponding to sesquiter-
the latter entraps, refocuses and releases fractions of the pene hydrocarbons. The oxygenated derivatives of both
GC effluent from the first dimension (1D, column 1), onto these groups are generally encountered closely after the
the second dimension (2D, column 2), in a continuous elution of the main groups in the 1D.
mode; this method enables an accurate screening of In order to improve peak identification in GC × GC,
complex matrices, offering very high resolution and enhanced retention index databases can be used for optimization
detection sensitivity.80,81 and provisional peak identification.85 At present, only 1D
The use of a retention index system in GC × GC analysis retention index databases derived from monodimensional
has also been widely discussed in the literature and, although GC are available and several papers report their use in
several investigations have been focused on the develop- GC × GC. Western et al.86,87 proposed a method to generate
ment of calculation methods, a standardized method has a retention index database also for the 2D separation. The
not yet been introduced. The establishment of accurate repeated analyses during a GC × GC run of reference
retention indices from GC × GC retention data is still an compounds, such as n-alkane, 2-methyl ketones and
emerging topic of research. FAMEs, were carried out by timed sequential solution
Dallüge et al82 used 1D retention indices and information injection, and the 2D retention indices were further
obtained from 2D retention times to improve the identifi- interpreted by the authors as pseudo-isothermal retention
cation of volatiles detected in cigarette smoke by means indices, based on the isovolatility curves discussed by
of GC × GC coupled to a time-of-flight mass spectrometer Beens et al.88,89
(GC × GC–TOF–MS). Out of a total of 2500 volatiles,
reference retention indices were found for only 238 com-
pounds, and out of these a group of 152 volatiles had Concluding Remarks
their experimentally determined index in good agreement
with the literature values. Retention data can be considered a robust tool for
Retention index calculation using data derived from qualitative gas chromatographic analysis. In many cases,
unconverted comprehensive GC chromatograms is also the identification of an unknown compound is supported
among the proposed retention index calculation methods.83 by correlation between retention times, or retention tem-
In a GC × GC–qMS analysis on a highly complex perfume, peratures, for that unknown and a series of closely related
the assignment of peak identifications was based on the standard substances on two different stationary phases
injection of pure standard compounds, MS library matching presenting distinct polarities. Besides, the narrow peaks
and 1D linear retention index data. The latter was used as a inherent to capillary gas chromatographic applications are
filter during MS library searches, with the elimination of well suited to qualitative analysis through retention data.
matches outside a predefined index window. As described It is certainly true that the utilization of selective or spectro-
by the authors, retention indices were calculated through scopic detectors provides additional and indispensable
data derived from unconverted GC × GC chromatograms, information for positive identity confirmation.
in which the retention time of the central component was Obviously, it is not feasible to report all the research
considered in the case of an odd number of modulated work carried out on retention index systems, as also all the
analyte peaks, while in the case of an even number of results that have been obtained by individual research
modulated peaks the central retention time between the groups. The widespread use of retention indices in flavour
two most internal peaks was considered. Furthermore, by and fragrance materials analysis is a known fact, although
using a column set consisting of a non-polar 30 m × 0.25 mm the need is towards a uniform expression through the
i.d. primary column and a polar 1 m × 0.25 mm i.d. normalization of those indices. Particular attention should
secondary capillary, the authors considered the influence be paid not only to repeatability, but also to reproducibility,
of the polar 2D column negligible for the less polar which is far from acceptable in many cases.
analytes, in terms of retention, but relevant for the more As a consequence, Flavour and Fragrance Journal
polar ones, which were strongly retained. enforces in the instructions for authors the need for replicated
A different approach was used in the analysis of nine sample preparation and analysis, as well as the use of
lavender oils (Lavandula angustifolia);84 the correlation appropriate statistical treatment to the resultant data. The
of the linear retention indices from GC–MS analysis and demand for reliable compound identification and quantifi-
the GC × GC retention data was successfully used to identify cation is also of utmost importance. Compound identification
unknown compounds in the oils. About 70% of the must be validated by at least two different methods, with
total number of components in each sample was reliably the most commonly applied being GC and MS. Moreover,
characterized and 85 components were identified. The latter assuming the use of GC data for identification, calculated
compounds accounted for more than 95% of the oil’s and reference retention indices of each identified compound

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj
LINEAR RETENTION INDICES IN GAS CHROMATOGRAPHIC ANALYSIS 313

are to be included; the use of two columns with different 41. Kostiainen R, Nokelainen S. J. Chromatogr. 1990; 513: 31–37.
42. Rohrschneider L. J. Chromatogr. 1966; 22: 6 –22.
stationary phases is also advised. With regards to quan- 43. McReynolds WO. J. Chromatogr. Sci. 1970; 8: 685 – 691.
tification, raw area percentages obtained by GC are not 44. Jennings W, Shibamoto T. Qualitative Analysis of Flavour and
acceptable for reporting the composition of a matrix; Fragrance Volatiles by Glass Capillary Gas Chromatography.
Academic Press: New York, 1980.
standardization methods are requested. Moreover, the 45. Adams RP. Identification of Essential Oil Components by Gas
isolation or synthesis of the compounds present at more than Chromatography/Mass Spectrometry. Allured: Carol Stream, IL,
20% are suggested for calibration purposes and structural 2007.
46. Sadtler Research Laboratories. The Sadtler Standard Gas
confirmation by other spectroscopic techniques. Chromatography Retention Index Library. Sadtler Research
Laboratories: Philadelphia, PA, 1984.
47. Joulain D, König WA. The Atlas of Spectral Data of Sesquiterpene
Hydrocarbons. E.B.–Verlag: Hamburg, 1998.
References 48. Maurer HH, Pfleger K, Weber A. Mass Spectral and GC Data of
Drugs, Poisons, Pesticides, Pollutants and Their Metabolites.
1. James AT, Martin AJP. Biochem. J. 1952; 50: 679 – 690. Wiley–VCH Verlag: Weinheim, 2007.
2. Littlewood AB, Phillips CSG, Price DT. J. Chem. Soc. 1955; 49. Richmond R. J. Chromatogr. A 1996; 724: 131–134.
1955: 1480. 50. Richmond R. J. Chromatogr. A 1996; 724: 229 –234.
3. Kováts E. Helv. Chim. Acta 1958; 41: 1915 –1932. 51. Hayes PC, Pitzer EW. J. High Resolut. Chromatogr. Chromatogr.
4. van den Dool H, Kratz PD. J. Chromatogr. 1963; 11: 463– 471. Commun. 1985; 8: 230 –242.
5. Royal Society of Chemistry, Analytical Methods Committee. 52. Davies NW. J. Chromatogr. 1990; 503: 1–24.
Analyst 1980; 105: 262–273. 53. Bianchi F, Careri M, Mangia A et al. J. Sep. Sci. 2007; 30: 563 –572.
6. Budahegyi MV, Lombosi ER, Lombosi TS et al. J. Chromatogr. 54. Acree T, Arn H. Flavornet. Cornell University, NYSAES:
1983; 271: 213 – 307. New York: http://www.flavornet.com (accessed 13 May
7. Tarján G, Nyiredy Sz, Györ M et al. J. Chromatogr. 1989; 472: 2007).
1– 92. 55. Mottram R. The LRI and Odour Database. Flavour Research
8. Robinson PG, Odell AL. J. Chromatogr. 1971; 57: 1–10. Group, School of Food Biosciences, University of Reading, UK:
9. Harangi J. J. Chromatogr. A 2003; 993: 187–195. http://www.odour.org.uk (accessed 13 May 2007).
10. Evans MB, Smith JF. Nature 1961; 190: 905– 906. 56. El-Sayed AM. The Pherobase: Database of Insect Pheromones
11. Evans MB, Smith JF. J. Chromatogr. 1962; 8: 303 –307. and Semiochemicals. HortResearch, Lincoln, New Zealand: http://
12. Gerbino TC, Castello G. J. Chromatogr. A 1995; 699: 161–171. www.pherobase.com (accessed 15 May 2007).
13. Poole CF. The Essence of Chromatography. Elsevier: Amsterdam, 57. Mondello L, Dugo P, Basile A et al. J. Microcol. Sep. 1995; 7:
2003. 581– 591.
14. Castello G, Testini G. J. Chromatogr. A 1996; 741: 241–249. 58. Shellie R, Marriott P, Zappia G et al. J. Essent. Oil Res. 2003; 15:
15. van den Dool H. Standardization of GC Analysis of Essential Oils. 305–312.
Proefschrift, Rijksuniversiteit te Groningen, Rotterdam, 1974. 59. Costa R, De Fina MR, Valentino MR et al. Nat. Product Commun.
16. Royal Society of Chemistry, Analytical Methods Committee. 2007; 2: 413 – 418.
Analyst 1981; 106: 448 – 455. 60. Costa R, De Fina MR, Valentino MR et al. Application of a new
17. Evans MB, Smith JF. J. Chromatogr. 1961; 5: 300 –307. GC–MS library, designed with a retention index filter tool, to the
18. Novák J, Ružiková J. J. Chromatogr. 1974; 91: 79 – 88. analysis of the essential oil of Cistus ladanifer. In Proceedings of
19. Zenkevich IG. Zh. Anal. Khim. 1984; 39: 1038 [in Russian]. the First International Medicinal and Aromatic Plants Conference
20. Zenkevich IG, Ioffe BV. J. Chromatogr. 1988; 439: 185 –194. on Culinary Herbs, Antalya, Turkey, 2007.
21. Berezkin VG. J. Chromatogr. 1974; 91: 559 –582. 61. van Asten A. Trend. Anal. Chem. 2002; 21: 698 –708.
22. Luo X, Wong X, Cheng L. Gaodeng Xuexiao Huaxue Xuebao 62. Bauer K, Garbe D, Surburg H. Common Fragrance and Flavor
1985; 5: 61. Materials – Preparation, Properties and Uses. Wiley–VCH:
23. Rasanen I, Ojanperä I, Vuori E. J. Chromatogr A 1995; 693: 69 – Weinheim, 2001.
78. 63. Tóth A, Praszna L. Flavour Fragr. J. 1998; 13: 196 –202.
24. Dimov N. J. Chromatogr. 1985; 347: 366 –374. 64. Kováts E. Adv. Chromatogr. 1965; 1: 229 –247.
25. Škrbic BD. J. Chromatogr. A 1997; 764: 257–264. 65. Hinshaw JH, Ettre LS. Introduction to Open-tubular Column Gas
26. Blumberg LM, Klee M. Anal. Chem. 1998; 70: 3828 –3839. Chromatography. Advanstar Communication: Cleveland, 1994.
27. Stránský K, Zarevúcka M, Valterová I et al. J. Chromatogr. A 66. Bicchi C, Binello A, D’Amato A et al. J. Chromat. Sci. 1999; 37:
2006; 1128: 208 –219. 288–294.
28. Balaban AT. Chem. Phys. Lett. 1982; 89: 399 – 404. 67. Hérent MF, De Bie V, Tilquin B. J. Pharm. Biomed. Anal. 2007;
29. Heinzen VEF, Yunes RA. J. Chromatogr. A 1996; 719: 462– 467. 43: 886 – 892.
30. Bruchmann A, Zinn P, Haffer CM. Anal. Chim. Acta 1993; 283: 68. Croteau R, Kutchan T, Lewis NG. Biochemistry and Molecular
869 – 880. Biology of Plants, Buchanan GW, Gruissen W, Jones R (eds).
31. Castello G. J. Chromatogr. A 1999; 842: 51– 64. ASPB and Wiley: Hoboken, NJ, 2000.
32. Mjøs SA, Meier S, Boitsov S. J. Chromatogr. A 2006; 1123: 98–105. 69. International Organization of the Flavor Industry (IOFI). Z. Lebensm.
33. Shibamoto T. In Capillary Gas Chromatography in Essential Oil Unters. Forsch. 1991; 192: 530.
Analysis, Bicchi C, Sandra P (eds). Huethig Verlag: Heidelberg, 70. Royal Society of Chemistry, Analytical Methods Committee.
1987. Analyst 1997; 122: 1167–1174.
34. Woodford EP, van Gent CM. J. Lipid Res. 1960; 1: 188–190. 71. König WA. Gas Chromatographic Enantiomer Separation with
35. Miwa TK, Mikilajczak KL, Earle FR et al. Anal. Chem. 1960; 32: Modified Cyclodextrins. Hüthig: Heidelberg, 1992.
1739 –1742. 72. Schurig V, Novotny HP. J. Chromatogr. 1988; 441: 155 –163.
36. Lee ML, Vassilaros DL, White CM et al. Anal. Chem. 1979; 51: 73. Dugo G, Lamonica G, Cotroneo A et al. Perf. Flav. 1992; 17: 57–
768 –773. 74.
37. Rasanen I, Ojanperä I, Vuori E et al. J. High Resol. Chromatogr. 74. d’Acampora Zellner B, Casilli A, Crupi ML et al. Assessment and
1993; 16: 495 –500. Enantiomeric Differentiation of Chiral Compounds by means of
38. Rasanen I, Ojanperä I, Vuori E. J. Chromatogr. A 1995; 693: 69 –78. GC–Enantio/O and GC–Enantio/MS/O. In Proceedings of Enantiosep
39. Hall GL, Whitehead WE, Mourer CR et al. J. High Res. Chromatogr. ’07, Ferrara, Italy, 2007.
1986; 9: 266 –271. 75. Rubiolo P, Liberto E, Cagliero C et al. Linear retention indices
40. Manninen A, Kuitunen ML, Julin L. J. Chromatogr. 1987; 394: in enantioselective GC–Mass spectrometry (Es–GC–MS) as a
465– 471. tool to identify enantiomers in flavour and fragrance fields. In

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj
314 B. D’ACAMPORA ZELLNER ET AL.

38th International Symposium on Essential Oils, Graz, Austria, 83. Mondello L, Casilli A, Tranchida PQ et al. J. Chromatogr. A
2007. 2004; 1067: 235 –243.
76. Rubiolo P, Liberto E, Cagliero C et al. (in preparation). 84. Shellie R, Mondello L, Marriott P et al. J. Chromatogr. A 2002;
77. Deans DR. Chromatographia 1968; 1: 18 –22 970: 225 –234.
78. Himberg K, Sippola E, David F et al. J. High Res. Chromatogr. 85. Vendeuvre C, Bertoncini F, Thiébaut D et al. J. Sep. Sci. 2005;
1993; 16: 645– 650. 28: 1129 –1136.
79. Mondello L, Casilli A, Tranchida PQ et al. LCGC Europe 2008; 86. Western RJ, Marriott PJ. J. Chromatogr. A 2003; 1019: 3 –14.
21: 130 –137. 87. Western RJ, Marriott PJ. J. Sep. Sci. 2002; 25: 832– 838.
80. Phillips JB. J. Chromatogr. A 1999; 856: 331–347. 88. Beens J, Tijssen R, Blomberg J. J. Chromatogr. A 1998; 822:
81. Liu Z, Phillips JB. J. Chromatogr. Sci. 1991; 1067: 227–231. 233 –251.
82. Dallüge J, van Stee LLP, Xu X et al. J. Chromatogr. A 2002; 974: 89. Beens J, Tijssen R, Blomberg J. J. High Resol. Chromatogr.
169 –184. 2000; 21: 63 – 64.

Copyright © 2008 John Wiley & Sons, Ltd. Flavour Fragr. J. 2008; 23: 297–314
DOI: 10.1002/ffj

Вам также может понравиться